Results of system suitability.

Parameter Value
Retention time (RT, min) 4.0
Theoretical plate number (N) 36,594
HETP (mm) 0.0027
Capacity factor (k’) 5.67
Symmetry factor 1.23
System precision (%RSD, N = 6) Area 1.17
Retention time 0.13