Skip to main content
. 2002 Jul;40(7):2357–2362. doi: 10.1128/JCM.40.7.2357-2362.2002

TABLE 4.

Testing profiles of seven specimens found discordant when tested by two technologists in replicates of three in the LCx C. pneumoniae RUO assay

Identification no. (dilution) No. of positives/total no. of replicates tested by laboratory B in-house PCR Assay valuea for the indicated replicate tested by:
Technologist 1
Technologist 2
1 2 3 1 2 3
T2219 (10−7) 3/6 354.0 (+) 55.3 (−) 10.7 (−) 9.4 (−) 9.5 (−) 9.6 (−)
VR-1310 (10−7) 2/6 482.7 (+) 782.8 (+) 710.6 (+) 9.5 (−) 9.9 (−) 9.8 (−)
A03 (10−7) 2/6 9.7 (−) 10.3 (−) 10.0 (−) 239.4 (+) 9.6 (−) 9.7 (−)
379 CBD (10−6) 2/6 540.7 (+) 10.0 (−) 9.7 (−) 10.7 (−) 10.3 (−) 10.1 (−)
TNFαI 3I (10−5) 1/6 11.8 (−) 9.5 (−) 11.9 (−) 10.8 (−) 10.6 (−) 521.1 (+)
PS32 (10−7) 0/6 10.0 (−) 273.2 (+) 10.4 (−) 10.0 (−) 9.8 (−) 9.6 (−)
TNFαI 2I (10−5) 0/6 366.9 (+) 10.4 (−) 11.3 (−) 9.8 (−) 9.6 (−) 9.7 (−)
a

The cutoff for positivity in the LCx RUO PCR is 100.0, and positive (+) and negative (−) results are indicated by symbols in parentheses.