Skip to main content
. 2025 Mar 31;32(Pt 3):506–523. doi: 10.1107/S1600577525001389

Table 2. Comparison between the von Hamos (VH) and Laue spectrometers.

The efficiency values of the von Hamos spectrometer in Kα and Kβ measurements are set to 1.

  VH Laue-Si Laue-SiO2
Crystal Silicon Silicon Quartz
Radius (m) 0.5 1.5 1.5
Open window 3 cm × 11 cm 3 cm × 8 cm 3 cm × 8 cm
Index (111) (111) (110)
Efficiency at Nb Kα 1 ∼4 ∼11
Efficiency at Nb Kβ 1 ∼8 ∼22
Resolution at Nb Kβ (eV) 1.9 2.4 2.2