Table 3.

Diagnostic accuracy of plasma calprotectin and CRP for diagnosing remission in IBD patients

UC (n = 99) CD (n = 209)
plasma calprotectin, EliATM (automated) plasma calprotectin, CalprestTM (manual) hs-CRP plasma calprotectin, EliATM (automated) plasma calprotectin, CalprestTM (manual) hs-CRP
AUROCa 0.999b 0.986b 0.960 0.675c 0.662c 0.716
 SE 0.001 0.01 0.02 0.05 0.05 0.05
 95% CI 0.962–1.000 0.939–0.999 0.900–0.989 0.607–0.738 0.594–0.726 0.649–0.776
Optimal threshold ≤10.5 ≤8.1 ≤5.9 ≤10.2 ≤13.3 ≤11.8
Sensitivity 98.63 87.67 90.41 48.04 67.6 62.01
 95% CI 92.6–100.0 77.9–94.2 81.2–96.1 40.5–55.6 60.2–74.4 54.5–69.1
Specificity 100 100 96.15 83.33 63.33 76.67
 95% CI 86.8–100.0 86.8–100.0 80.4–99.9 65.3–94.4 43.9–80.1 57.7–90.1
PPV 100 100 98.5 94.5 91.7 94.1
 95% CI 95.0–100.0 94.4–100.0 92.0–100.0 87.6–98.2 85.6–95.8 88.2–97.6
NPV 96.3 74.3 78.1 21.2 24.7 25.3
 95% CI 81.0–99.9 56.7–87.5 60.0–90.7 14.2–29.7 15.6–35.8 16.7–35.5
p valuea <0.0001 <0.0001 <0.0001 0.0003 0.001 <0.0001

CRP, C-reactive protein; 95% CI, 95% confidence interval; PPV, positive predictive value; NPV, negative predictive value; SE, standard error.

aDiagnostic accuracy calculated via ROC analysis according to DeLong et al. [17].

bComparison of AUROC values for calprotectin accuracy between the two assays (EliA™ vs. Calprest™, p = 0.10).

cComparison of the accuracy of the AUROC values for calprotectin between the two assays (EliA™ vs. Calprest™, p = 0.3).