Table 1.

Sequence parameters

Sequence TR/ms TE/ms Flip angle/° Reconstructed FOV/mm Reconstructed matrix
Ventilation imaging 3.2 1.6 ~10 384 × 384 × 384 96 × 96 × 96
Dissolved-phase imaging 18 Echo train 1: 0.61/2.23/3.85 Echo train 2: 1.42/3.04 21 (dissolved)/0.4 (gas) 320 × 320 × 320 48 × 48 × 48
Fixed-TR dynamic spectroscopy 36 1.2 60 n/a n/a
CSSR Variable 0.7 ~27 (excitation using maximum RF power) n/a n/a
Diffusion-weighted imaging 57 41.6 7.4 400 × 400 × 240 64 × 64 × 6

The sequence parameters for dissolved-phase imaging are comparable to those previously established for multi-echo techniques and to 129Xe clinical trial consortium recommendations for multi-site trials [15, 28]. Compared to the recommendations, ventilation imaging uses a radial trajectory facilitating acceleration. Diffusion-weighted imaging uses shorter diffusion times to facilitate the estimation of surface-volume ratio. The impact of frequency-selective excitation in fixed-TR dynamic spectroscopy on dissolved-phase ratios is assessed in the supplemental material