Minimum requirements for continuous glucose monitors: What does it mean?

Jan S Krouwer 1,✉
PMCID: PMC12085550  PMID: 40380801

The U.S. Food and Drug Administration (FDA) 510(k) approval of laboratory medicine assays has always been consistent. 1 Manufacturers submit data according to a series of protocols. The two most important parameters are bias and precision. Clinical & Laboratory Standards Institute protocols are typically used (EP9 for bias and EP5 for precision). Approval is granted if the estimated performance is similar to existing assays (eg, substantial equivalence to a predicate device).

This paradigm changed for glucose meters in 2003 with the publication of the glucose meter standard ISO 15197 with a minimum acceptability requirement (revised in 2013). International Federation of Clinical Chemistry and others have proposed similar minimum requirements where a high percentage of results must meet a limit to be acceptable.2,3

The FDA has adopted several minimum requirements for continuous glucose monitors (CGMs). As one of several examples,

For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent. 4

One can view requirements like this in two ways:

  1. Failure to meet the requirement is grounds to reject the device’s approval.

  2. Meeting the requirement suggests that the device should be approved, but there is a concern.

The concern is that approving the device means that unacceptable results are potentially allowed (eg, 100% – X% results that exceed the limit). Since a CGM produces 100 000 annual glucose results, up to 1000 results (or about three results per day) will be in the B zone or higher in an error grid. 5

The disapproval of a device that does not meet minimum requirements is reasonable but the approval of devices that do meet minimum requirements allows for unacceptable results which is against the notion that approval implies a device that is safe and effective.

Whereas the above discussion relates to the approval process, one can examine results for devices postapproval by examining the adverse event database. Adverse events for the Dexcom G7 and Abbott Libre 3 were analyzed for the year 2024 with the results for inaccuracy or all causes: inaccuracy: malfunction 26 682, injury 2244, death one; all causes: malfunction 161 696, injury 6338, and death four.

Clearly, unacceptable results have been produced by these CGMs that have met minimum FDA requirements. It within the FDA’s purview to remove a device from the market which typically involves a 2 × 2 table as shown in Table 1.

Table 1.

Consequences for Keeping or Removing a Product.

Decision Good consequence Bad consequence
Keep the CGM device on market Users can monitor their glucose using CGMs Users get injured from device inaccuracy or other problems
Remove the CGM device from market No injury to users because the device is not on the market Users cannot monitor their glucose using CGMs

A rational decision is the one that provides the most benefit and the least harm to users.

To approve a device that meets minimum requirements implies that it is safe and effective. Adverse event data from real-world data suggest this is not the case. Devices that meet minimum requirements in spite of producing unacceptable results nevertheless merit approval when consequences are considered of keeping the device on the market versus removing it. Disapproval of a device that does not meet minimum requirements is reasonable in reducing the number of adverse events.

Acknowledgments

None.

Footnotes

Abbreviations: CGM, continuous glucose monitor; CLSI, Clinical & Laboratory Standards Institute; MAUDE, manufacturer and user facility device experience; IFCC, International Federation of Clinical Chemistry.

The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

ORCID iD: Jan S. Krouwer Inline graphic https://orcid.org/0000-0003-2300-076X

References


Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society

RESOURCES