Optimizing Osseodensification Drilling for Dental Implant Placement: An In Vitro Study

Xingru Tao 1,2,3, Jie Yang 4, Tai Ma 4, Ming Chen 4, Qinglong An 5,, Dedong Yu 1,2,3,5,
PMCID: PMC12152765  PMID: 40497493

ABSTRACT

Objectives

This study aims to optimize the parameters of osseodensification drilling (OD) to improve the primary stability of dental implants in low‐density bone.

Material and Methods

Polyurethane foam blocks (PFB) of 0.160 g/cm3 were used to simulate low‐density bone (Type IV bone). Two drills, i.e., bone level tapered drills and Densah drills, were used in conventional drilling (CD) and OD, respectively. Drilling was performed on a DMU machine, varying spin speed, feed per tooth, and irrigation with or without 4°C saline. Thrust force, temperature, and entrance characteristics of implant sites in CD and OD were compared. The primary stability of implants was assessed via insertion torque (IT), removal torque (RT), and implant stability quotient (ISQ).

Results

OD demonstrated higher thrust force than CD with maximum values at 300 rpm (2.99 ± 0.22 N vs. 2.77 ± 0.17 N, p < 0.01). Temperature elevation was lower in OD than that in CD under irrigation (3.35°C vs. 4.67°C). Despite comparable ISQ values (CD: 46.71 ± 8.56 vs OD: 47.08 ± 5.95, p = 0.86), OD achieved higher IT (11.73 ± 0.45 N·m vs. 7.77 ± 0.21 N·m, p < 0.001) and higher RT (9.28 ± 0.45 N·m vs. 6.65 ± 0.19 N·m, p < 0.001). Morphological analysis revealed fewer defects (tears/potholes) in OD than CD.

Conclusions

OD drills may avoid iatrogenic damage of the morphology of implant sites and enhance primary stability in Type IV bone at 1500 rpm and 0.04 mm/z with irrigation to prevent thermal damage. OD also outweights CD in increased bone density, thrust force, and torque.

Clinical Implications

For Type IV alveolar bone, OD at 1500 rpm and 0.04 mm/z is recommended for improving he primary stability of dental implants. Sufficient irrigation is crucial in both CD and OD for circumventing the thermal damage to bone.

Keywords: dental implant, osseodensification drilling, polyurethane foam blocks, primary stability, Type IV bone

Summary

What Is Known?

  • Osseodensification drilling (OD) techniques can enhance the primary stability of dental implants in low‐density bone, but optimal drilling parameters remain unveiled.

What This Study Adds?

  • Identified optimized OD parameters (1500 rpm, 0.04 mm/z feed, irrigation) to provent the impaired implant sites and improve torque values (IT/RT) versus CD.

  • Demonstrated OD increases bone density and thrust force while maintaining comparable ISQ, offering a reliable method for improving the primary stability of implants in Type IV bone.


Abbreviations

CD

conventional drilling

ISQ

implant stability quotient

IT

insertion torque

OD

osseodensification drilling

PFB

polyurethane foam blocks

RT

removal torque

1. Introduction

Dental implants have been predominantly utilized for oral rehabilitation in partial or complete edentulous cases owing to the pronounced long‐term success rate. Appropriate density of bone around inserted implants is vital for this long‐term success as it may significantly facilitate the firm and direct integration between bone and dental implants, i.e., osteointegration (Frisch et al. 2020; Chauhan et al. 2018; Kotsakis and Romanos 2022). For dental implant placement, the holes with expected diameters and depths in the jawbone should be prepared by drills. Conventional drilling (CD) techniques employ drills with sequential diameters to create cylindrical osteotomies or conical holes, and the diameter of the final drill is usually equal to the implants. However, in osteoporotic areas, the diameter of prepared sites is usually larger than that of the final drills and the implants, resulting in inadequate osteointegration and losing the primary stability (Lekholm UJTip 1985; Putra et al. 2024). To address this clinical challenge, some strategies were proposed. For instance, osteotome technique was introduced in cases with insufficient bone height and density for dental implant insertion in post‐maxilla areas (Summers 1994). It increases bone density by grafting a layer of bone or bone substitutes on the surface of residual bone, ultimately improving the primary stability of the inserted dental implant. However, this procedure may cause severe side effects, including unexpected trauma to the patients, accidental fractures, and bone displacement.

Alternatively, Huwais and Meyer first introduced a method, i.e., “osseodensification” drilling (OD), for preparing implant sites in bone with inferior density (Huwais and Meyer 2017). This technique can achieve lateral bone extrusion and increase bone density at the contact interface with the implant (Mello‐Machado et al. 2021; Tabassum et al. 2010; Bergamo et al. 2021), thereby finally increasing the primary stability of dental implants. In addition, OD reduces heat generation, diminishing the potential thermal damage to bone during the implant site preparation (Mullings et al. 2021; Bhargava et al. 2023; Yeh et al. 2021; Slete et al. 2018).

A wealth of studies have confirmed superior penetration forces, torque, and primary stability of dental implants after using OD than using CD techniques. Bettach et al. documented that OD in low‑density bone generated a significantly higher torque than CD (Bettach et al. 2023). As with increased torque, the improved primary stability of dental implants and more heat production were also reported. A systematic review revealed that a significant increase in the insertion torque (IT) of the implants positioned through the OD protocol compared to the CD (Inchingolo et al. 2021). Additionally, compared with the osteotome technique, OD is less invasive and less likely to induce fracture, thereby circumventing some complications, such as dizziness and infection.

However, Huwais and other related studies did not recommend exact parameters used in OD for clinical practice. Hence, this study is designed to explore the potentially optimal parameters, including thrust force, real‐time temperature around the bone during drilling, IT, and removal torque (RT) in OD, which may guide clinical practice.

2. Materials and Methods

In this study, polyurethane foam blocks (PFB) (Sawbones, Vashon Island, Washington, D.C., USA) of 0.160 g/cm3 (PCF 10) were used to simulate low‐density bone (Type IV bone), as standardized by the American Society for Testing and Materials (Standard Specification for Rigid Polyurethane Foam for Use as a Standard Material for Testing Orthopaedic Devices and Instruments n.d). The dimensions of the PFB used in the experiments were 80×40×20mm, and SEM images of the PFB material are shown in Figure 1B. BLT dental implants with 4.1 mm diameter and 10 and 16 mm length (BLT, Straumann, Switzerland) were used in this study (Figure 1C,D). Saline was stored in 4°C for irrigation during drilling. Ethics approval was waived for this In Vitro study.

Figure 1.

Figure 1

Involved materials in the study. (A) PFB; (B) SEM image of PFB; (C) 10 mm long implant; (D) 16 mm long implant; (E–G) bits of BLT drills; (H–J) bits of Densah drills.

Based on a previous study that used 40 samples of PFB models (de Carvalho Formiga et al. 2021), we calculated the sample size using G*Power (Standard Edition, Universität Düsseldorf, Düsseldorf, Germany). After a comprehensive consideration of various factors, 48 samples were recruited in this experiment (effect size = 0.92, α = 0.05, power = 0.8). Osteotomies were created in the PFBs using two different drills: bone‐level tapered (BLT) drills for CD (BLT, Straumann, Switzerland) and Densah drills (Versah, Jackson, Michigan, USA) for OD. The BLT drills contains three different drill bits: a 2.2 mm diameter pilot drill, 2.8 and 3.5 mm diameter drills. Three Densah drills were employed in the present study: a 1.6 mm diameter pilot drill bit, 1828 (cone shape; 18 means the secondary 1.8 mm diameter; 28 means the main diameter of the crown is 2.8 mm), and 2838 drills. The materials used in this study are shown in Figure 1.

Osteotomy experiments were performed using a DMU computer numerical control machine. Three levels of spin speeds (N), two levels of feed per tooth (Fz), and two levels of irrigation were set (Table 1). According to Versah protocols, we used three spin speed: 300, 800, and 1500 rpm in this study. A drilling force measurement system (KISTLER 9272 dynamometer, Kistler Group, Switzerland) was used to measure the dynamic thrust force.

Table 1.

Experimental parameters in this study.

Material Spin speed (N) (rpm) Feed per tooth Fz (mm/z) Implant depth (mm) Irrigation
PCF 10 300, 800, 1500 0.04, 0.08 10, 16 without saline; saline

To collect the internal temperature of the material, a K‐type thermocouple was embedded in two diameter holes (Φ 1 mm) to collect the real‐time temperature at depths of 5, 10, and 16 mm (Figure 2C). Therefore, a thermocouple temperature‐measuring system was designed using NI LabView to analyze the temperature inside the hole. The surface temperatures of the PFB and drills were measured using an infrared thermal imager (FLIR A615, Teledyne FLIR, Wilsonville, OR, USA). The hole surface was evaluated using a material‐type upright laser confocal microscope (VK‐X3000, KEYENCE, Osaka, Japan) and a Raman image scanning electron microscope (RISE‐MAGNA, Magna, Canada). The IT and RT were measured using a torque‐testing machine (Tohnichi Mfg, Tohnichi, Japan), and the implant stability quotient (ISQ) was measured using a resonance frequency analysis system (Osstell, Gothenburg, Sweden). All involved instruments in this study are illustrated in Figure 2.

Figure 2.

Figure 2

Drilling experimental setup. (A) machining center; (B) drilling platform; (C) schematic showing thermocouple distribution.

3. Statistical Analysis

SPSS 26.0 (IBM, Armonk, NY, USA) was used for the statistical analysis. Shapiro–Wilk test was used to conduct data normality analysis, and independent samples t‐test was used to compare differences between these two groups. All data were shown as mean ± standard deviation, and a statistically significant difference was set as p < 0.05.

4. Results

4.1. IT, RT, and ISQ Values

The primary stability of inserted implants in the two groups was evaluated by immediate IT, RT, and ISQ after their successful placement in the PFBs, and the details are listed in Table 2. Accodring to the Shapiro–Wilk test analysis results, the data are normally distributed. As shown in Table 2, the IT values in the OD group significantly increased. The IT values in CD and OD groups are 7.77 ± 0.21 and 11.73 ± 0.45, respectively, with a significant difference (p < 0.001). Likewise, the OD group has a higher RT value than the CD group (9.28 ± 0.45 and 6.65 ± 0.19, respectively), demonstrating a significant difference (p < 0.001). However, ISQ values in the four directions (buccal, lingual, mesial, and distal) show no significant differences between the CD and OD groups. The mean ISQ values for the CD and OD groups were 46.71 versus 47.08 (buccal), 45.67 versus 46.96 (lingual), 49.08 versus 49.71 (mesial), and 48.17 versus 49.92 (distal). The OD group consistently showed a slightly higher average ISQ value compared to the CD group.

Table 2.

IT, RT, and ISQ in CD and OD groups.

Groups Min. (N·m) Max. (N·m) SD Average (N·m) 95%CI S‐W test p‐value
IT
CD 5.80 9.50 1.04 7.77 7.33, 8.21 0.61 p < 0.001
OD 8.00 14.80 2.22 11.73 10.79, 12.67 0.05
RT
CD 4.80 9.20 0.92 6.65 6.26, 7.04 0.20 p < 0.001
OD 6.00 14.10 2.22 9.28 8.35, 10.22 0.26
ISQ
BLT‐buccal 28.00 64.00 8.56 46.71 43.10, 50.32 0.13 0.86
OD‐buccal 35.00 57.00 5.95 47.08 44.57, 49.60 0.19
BLT‐lingual 21.00 65.00 9.39 45.67 41.70, 49.63 0.07 0.55
OD‐lingual 35.00 57.00 4.95 46.96 44.87, 49.05 0.32
BLT‐mesial 25.00 65.00 9.32 49.08 45.15, 53.02 0.42 0.66
OD‐mesial 42.00 63.00 5.86 49.71 47.23, 52.18 0.09
BLT‐distal 28.00 65.00 8.74 48.17 44.48, 51.86 0.17 0.22
OD‐distal 42.00 63.00 5.83 49.92 47.45‐52.38 0.143

Abbreviations: CD, conventional drilling; IT, insertion torque; ISQ, implant stability quotient; OD, osseodensification; RT, removal torque.

4.2. Entrance Characteristics of Prepared Implant Sites

The entrance characteristics of implant sites were assessed using a material‐type upright laser confocal microscope. In Figure 3, the entrance characteristics were varied when using different drills to prepare the implant sites under the same condition (N = 300 r/min, Fz = 0.08 mm/z, depth = 10 mm, without irrigation). In the CD group, burrs, tears, and potholes were observed at the prepared sites. However, only a few burrs appeared in the OD group. During clockwise drilling, the cutting action of the tool generated burrs on the prepared surface. When the implant was inserted, burrs were further displaced due to torque, leading to tearing or the formation of depressions. In the counterclockwise mode, the tool compresses the material, causing localized fractures and adhesion of material to the prepared site walls. This compression increased the material's resistance to deformation during implant insertion, resulting in fewer occurrences of tearing or potholes, and only some burrs.

Figure 3.

Figure 3

Characteristics of prepared implant sites. (A) implant sites prepared via BLT drills; (B) implant sites prepared via Densah drills.

4.3. Spin Speed and Thrust Force

During drilling, the thrust force (in the axial direction) reflects different material machining stages. Figure 4a shows the thrust force of BLT drills under dry drilling during the following routine clinical procedures. With this method, the osteotomy is generally performed using bits of three different diameters (2.2, 2.8, and 3.5 mm). A 2.2 mm pilot drill bit was used to form a pre‐hole. The 2.8 and 3.5 mm drill bits were used to form the final implant hole to complete the osteotomy. The A‐I dashed line indicates the critical state when the pilot drill bit is ready to enter the PFB. The A‐II dashed line indicates the state at which the drill bit reaches the bottom of the PFB. Moreover, A‐Ⅲ represented the state of the drill bit as it exited the PFB. The thrust force produced by the BLT drill bits was relatively small, in the range of 0–6 N. The thrust force was 1.53 (pilot drill bit), 2.22 (2.8 mm drill bit), and 2.77 N (3.5 mm drill bit), respectively. However, using the same drilling parameters, the thrust force of the Densah drill bits is higher than that of the BTL drill bits (Figure 4b). The thrust force of the 1.6 mm pilot drill bit was 0.97–1.42 N (49.20–85.49%) higher than that of 2.2 mm pilot drill bit under the clinical routine method. The working mode of the pilot drill bit was clockwise cutting. The 2.2 mm BLT drill bit was built with more spiral grooves than the 1.6 mm Densah drill bit. As shown in Figure 4b, the thrust force has an evident tendency to increase. The thrust force of OD VT‐1828 and OD VT‐2838 was 0.32–0.67 N (21.10–61.58%) and 0.18–4.44 N (7.64–134.86%) higher than that of the 2.8 and 3.5 mm BLT drill bits, respectively. For drilling parameters: N = 300 r/min; Fz= 0.04 mm/z; implant depth: 16 mm. The thrust force produced by the BLT drill bits was 0–6 N. The thrust force was 1.53 (pilot drill bit), 2.28 (2.8 mm drill bit), and 2.65 N (3.5 mm drill bit). All three drill bits were rotated clockwise. Subtractive manufacturing has also been used. Force analysis of the PFB was conducted according to the theory of the three deformation zones. Because of the structural characteristics of the material, the cutting forces in the three deformation zones were relatively small. Therefore, the thrust force from the deformation zones was small.

Figure 4.

Figure 4

(A) Thrust force signals showing the bone level tapered (BLT) drills under drilling without irrigation during clinical routine method: Pilot, 2.8 mm, and 3.5 mm drills (drilling parameters: N = 300 r/min; F z = 0.08 mm/z; implant hole depth = 16 mm). (B) Thrust force signals showing the Densah drills under drilling without irrigation during osseodensification method: Pilot, 1828, and 2838 drills (drilling parameters: N = 300 r/min; F z = 0.08 mm/z; implant hole depth = 16 mm). (C) Thrust force signals showing the Densah drills under drilling without irrigation during osseodensification method: Pilot, 1828, and 2838 drills (drilling parameters: N = 300 r/min; F z = 0.04 mm/z; implant hole depth = 16 mm). (D) Thrust force signals showing the Densah drills under drilling without irrigation during osseodensification method: Pilot, 1828, and 2838 drills (drilling parameters: N = 1500 r/min; F z = 0.04 mm/z; implant hole depth = 16 mm). (E) Effect of spin speed and feed per tooth on the thrust force under drilling without irrigation (implant depth = 10 mm and 16 mm). E‐a: 10 mm implant, BLT and Densah groups, thrust force and spin speed (N from 300 to 1500 r/min); E‐b: 10 mm implant, BLT and Densah groups, thrust force and feed per tooth (Fz from 0.04 to 0.08 mm/z; E‐c: 16 mm implant, BLT and Densah groups, thrust force and spin speed (N from 300 to 1500 r/min); E‐d: 16 mm implant, BLT and Densah groups, thrust force and feed per tooth (Fz from 0.04 to 0.08 mm/z).

However, the thrust force of Densah drill bits was 2.24 (pilot drill bit), 2.34 (OD VT‐1828), and 2.99 N (OD VT‐2838) (Figure 4c). In the osseodensification method, the tool is changed from clockwise to counterclockwise, and the material is squeezed by the tool instead of cutting. The thrust force increased by 46.41, 2.63, and 12.83%, respectively. For drilling parameters: N = 1500 r/min, Fz = 0.04 mm/z; Implant depth = 16 mm. Thrust force was 0.44 (2.2 mm drill bit), 1.86 (2.8 mm drill bit), and 2.46 N (3.5 mm drill bit), respectively (Figure 4d). With an increase in the spin speed, the extrusion between the tool and material was reduced. Therefore, the thrust forces were reduced by 80.36, 20.51, and 17.73%, respectively.

The analysis in Figure 4e indicates that the thrust force initially increases and then decreases with rising spin speed (300–1500 r/min), for both BLT and Densah drill bits. At 300 r/min, the thrust force of Densah drill bits was approximately 47.22% higher than that of BLT drill bits, while at 800 r/min, it was 13.00% higher. At 1500 r/min, the thrust force for BLT drill bits was marginally higher (0.08%) than for Densah drill bits. The thrust force increased with feed per tooth for both BLT and Densah drill bits. At a 10 mm implant depth, the thrust force for BLT drill bits increased by 47.21% and 12.12%, while for Densah drill bits, the increase was approximately 22.73% and 23.72%, respectively.

4.4. Temperature

The temperature data collected in various intervals, i.e., 0–70 s, were used to analyze the heat product affected by different machining parameters. Figure 5A,B show the temperature changes under certain drilling parameters (300 rpm, 0.04 mm/z and 1500 rpm, 0.04 mm/z); Without irrigation, the average maximum temperature (Tmax) of BLT drilling reached 12.66°C at 300 rpm and 0.04 mm/z; at 300 rpm and 0.08 mm/z, the Tmax slightly increased to 12.71°C. Conversely, at 1500 rpm and 0.04 mm/z, Tmax increased to 10.91°C, demonstrating a 13.82% reduction attributed to decreased friction at higher spin speeds. During drilling with irrigation, as shown in Figure 6C, the Tmax reached 10.67°C under 300 rpm and 0.04 mm/z, while under 300 rpm, 0.08 mm/z and 1500 rpm, 0.04 mm/z, the Tmax reached 11.57°C and 10.39°C, respectively, reflecting the cooling effect of irrigation. The same trend was observed with increasing feed per tooth and spin speeds. Notably, irrigation reduced Tmax by 15.72% (12.66°C–10.67°C) at 300 rpm and 0.04 mm/z, and by 8.97% (12.71°C–11.57°C) at 300 rpm and 0.08 mm/z.

Figure 5.

Figure 5

(A–F) Heat map showing the hole temperature under bone‐level tapered (BLT) drill bits (Implant depth: 10 mm; color level: 9–14). (G–L) Heat map showing the hole temperature under Densah drill bits (implant depth = 10 mm; color level: 10–15). The “1” and “2” on the vertical axis represent the pilot drill of BLT, “3” and “4” represent the 2.8 mm drill bit of BLT, and “5” and “6” the 3.5 mm drill bit of BLT. The odd number represented 5 mm depth, and the even number represented 10 mm depth.

Figure 6.

Figure 6

Effect of spin speed and feed per tooth on the Tmax under drilling without irrigation: (A, B) bone‐level tapered BLT drill bits; (C, D) Densah drill bits; implant depth: 10 mm.

For Densah drills, Figure 5G–L illustrates the same trends. Under drilling without irrigation, Tmax was 12.74°C at 300 rpm and 0.04 mm/z, slightly increasing to 12.81°C at 300 rpm and 0.08 mm/z. However, at 1500 rpm and 0.04 mm/z, Tmaxdecreased to 12.62°C. This trend suggests that while higher feed rates slightly increase the heat due to thicker chips, higher spin speeds reduce friction, resulting in lower Tmax. With irrigation, the Tmax reached 12.68, 12.07, and 12.05°C, respectively. The Tmax for drilling with saline showed 5.26 (from 12.74°C to 12.68°C), 5.78 (from 12.81°C to 12.07°C), and 4.52% (from 12.62°C to 12.05°C) decreased. The use of saline effectively reduced the heat; however, less temperature decrease in the OD group was identified.

Figure 6 highlights temperature differences (Tmax) at various depths. For BLT drills, at a depth of 5 mm under drilling without irrigation, Tmax decreased by 3.52% with increasing spin speed, attributed to reduced tool‐material friction. A similar trend was observed at 10 mm depth. Under irrigation, at 5 mm depth, Tmax decreased by 13.59%, 15.22%, and 16.49%, with the maximum reduction of 21.12% observed at 10 mm depth (from 3.5°C to 0.90°C). However, increasing feed per tooth slightly increased Tmax, attributed to the greater material thickness removed per revolution. For Densah drills, Tmax followed a different trend under drilling without irrigation. Initially, it increased with spin speed due to extrusion friction, rising by 0.70% (from 3.94°C to 4.03°C at 5 mm depth). Beyond a threshold speed, Tmaxdecreased by 3.07% (from 4.02°C to 3.62°C), as reduced contact time lessened the extrusion effect. Under irrigation, Tmax decreased by 14.51% (from 4.03°C to 2.14°C) at 5 mm depth. While the trend of temperature reduction with increasing spin speed was similar to BLT drills, the decrease for Densah drills was less significant.

As summarized in Table 3, both BLT and Densah drills benefited from saline irrigation in reducing thermal damage during dental implant site preparation. However, BLT drills showed greater temperature reduction efficiency compared with Densah drills. This highlights the importance of optimizing drilling parameters and irrigation methods to minimize thermal damage and improve clinical outcomes.

Table 3.

The ∆Tmax during the implant site preparation with 16 mm depth.

No. Drill bit Spin speed N (r/min) Feed per tooth Fz (mm/z) Cold fluid Tmax(°C)
1 BLT 300 0.04 Without irrigation 5.13
2 BLT 300 0.08 Without irrigation 6.00
3 BLT 800 0.04 Without irrigation 5.01
4 BLT 800 0.08 Without irrigation 5.52
5 BLT 1500 0.04 Without irrigation 4.65
6 BLT 1500 0.08 Without irrigation 4.67
7 BLT 300 0.04 Saline 3.43
8 BLT 300 0.08 Saline 3.64
9 BLT 800 0.04 Saline 3.19
10 BLT 800 0.08 Saline 3.44
11 BLT 1500 0.04 Saline 3.19
12 BLT 1500 0.08 Saline 3.25
13 Densah 300 0.04 Without irrigation 4.46
14 Densah 300 0.08 Without irrigation 4.78
15 Densah 800 0.04 Without irrigation 4.80
16 Densah 800 0.08 Without irrigation 4.88
17 Densah 1500 0.04 Without irrigation 3.69
18 Densah 1500 0.08 Without irrigation 4.34
19 Densah 300 0.04 Saline 3.84
20 Densah 300 0.08 Saline 3.93
21 Densah 800 0.04 Saline 3.84
22 Densah 800 0.08 Saline 4.32
23 Densah 1500 0.04 Saline 3.35
24 Densah 1500 0.08 Saline 3.85

4.5. Optimized Drilling Parameters

Comparative analysis revealed distinct mechanical differences between OD and CD. The OD protocol generated elevated thrust forces and cutting resistance during cortical bone preparation, accompanied by increased thermal generation. However, this technique demonstrated superior performance in implant stability metrics, showing significantly enhanced insertion and removal torque compared to CD.

A three‐dimensional function involving speed, feed per tooth, and heat production was established in this study to explore the optimal parameters, aiming to identify the optimal parameters that minimize heat production during drilling. Based on the results, this study suggested a set of parameters could be used in clinical OD performance: the spin speed and feed per tooth were set as 1500 rpm and 0.04 mm/z, respectively (Figure 7).

Figure 7.

Figure 7

(A) Three‐dimensional (3D) image showing the effect of spin drilling speed and feed per tooth on temperature. (B) Contour map showing the effect of spin drilling speed and feed per tooth on temperature.

5. Conclusion

OD can improve thrust force, IT, and RT during and after implant site preparation in light of more compacted surrounding bone compared with CD. The potential optimal spin speed and feed rate per tooth in OD are 1500 rpm and 0.04, respectively. In addition, irrigation is vital to eschew thermal injury to bone during drilling in both CD and OD.

6. Discussion

After the first introduction of “osseodensification” by Huwais and Meyer in 2017, this strategy has been receiving increasing attention when preparing implant sites in bone with compromised density. Meanwhile, exploring the optimal parameters for OD to improve the clinical outcomes in implant dentistry remains ongoing.

Regarding the primary stability of implants, our data align with previous studies that OD leads to better primary stability than CD (Kalra et al. 2024; Fontes Pereira et al. 2023). To evaluate the primary stability of implants, besides IT, the most used parameter, RT that is rarely used in the clinical due to the ethical consideration can also quantify the mechanical interlock strength at the bone‐implant interface as a secondary assessment metric In Vitro (Orhan and Ciğerim 2024; Kim et al. 2020). According to our study, OD, compared with CD, significantly enhances the primary stabilily of implants, validated by increased IT and RT. Although this result agrees with the data from Kalra et al. we offer a different interpretation of the underlying mechanism (Kalra et al. 2024). Kalra et al. suggest that elevated primary and secondary stability are attributed to bone densification along the osteotomy walls and the presence of residual bone chips when using OD, while we suspect the differences in diameter between OD and CD drills may explain this phenomenon. The interfacial frictional forces at the implant‐osteotomy interface exhibit diameter‐dependent variations: larger implant diameters generate significantly greater bone thread engagement resistance, thereby directly modulating the biomechanical coupling efficacy, which is critical for primary stability establishment. The average diameter of CD drills in the present study is 3.5 mm; however, based on the diameter calculation, the diameter of 2838 drill at 10 mm is 3.44 mm.

Clinically, it is generally recommended that the initial IT of an implant should be more than 35 N·m and the value of ISQ should be above 70 in. immediate loading cases. Research indicated that there is a positive correlation between IT and initial ISQ (Park et al. 2012). However, excessive IT during implant placement in dense bone can contribute to implant fracture. Exceed IT may also induce other adverse events. Alqahtani et al. documented that in D4 bone (serverly osteoporotic bone), implants with varying diameters and thread configurations were also associated with abnormal stress levels under immediate loading conditions (Alqahtani et al. 2023). It has been well‐acknowledged that in temperatures over 53°C lasting more than 1 min, the viability of various cells in bone is significantly impaired (Eriksson et al.1982). The temperature during drilling is affected by some factors, such as the bone type, drill diameters, spin speed, and irrigation (Vaidya et al. 2023; Soldatos et al. 2022; Chen et al. 2021; Gaspar et al. 2013; Gehrke et al. 2013; Salomó‐Coll et al. 2021; Chen et al. 2019; Strbac et al. 2014; Alam et al. 2023). In our study, the temperatures during CD and OD were both below 53°C; however, the profiles of temperature changes in CD and OD groups are different. The temperature during CD consistently decreased with increasing spin speed. As the spin speed and material removal rate increase, the contact between the drill and the material decreases, resulting in reduced friction and the heat generation. By contrast, in OD, the temperature rose and then fell with increasing spin speed. Intringuingly, in Huwais and Meyer's study, the temperature in OD consistently increased and was higher than CD. We assume this discrepancy is linked to the higher spin speed used in our study. In OD, the drill rotated counterclockwise, and the material was crushed and broken, subsequently adhering to the wall of the hole. At low spin drilling speed, the squeezing between the tool and material was gentle. Nonetheless, with increased spin speed, the extrusion friction between the tool and material surged. Of note, when the spin speed exceeded a certain threshold, the contact time between the tool and the material decreased, and the extrusion friction effect became small accordingly.

The temperature was also affected by irrigation. Irrigation can drastically decrease bone temperature during drilling, and a low‐temperature irrigation media can further reduce heat generation during implant site preparation (Saxena et al. 2024; Bulloch et al. 2012; Boa et al. 2016). In our study, external irrigation was employed, significantly decreasing temperature during OD and CD, which was not investigated in Huwais's study. Comparative studies on these irrigation techniques have been conducted. A study by Srtbac et al. found that combined irrigation can provide sufficient reduction in temperature changes during CD. However, the optimal irrigation method for OD remains under exploration (Strbac et al. 2014). In this study, we did not evaluate the effect of these two irrigation modes on temperature decreasement in OD. In future, a more comprehensive research will be scheduled to provide detailed information on this parameter and optimize the irrigation practice in clinical in OD.

Moreover, the characteristics of the dental implant site preparation varied from OD to CD in our study. It was discovered that only the powdery debris attached to the hole wall were identified in OD groups. In contrast, entrance morphology of prepared sites in CD groups features burrs, tearing, and potholes in the clockwise cutting mode. Studies demonstrated that implant designs affected osseointegration (Causey et al. 2021). Sergio et al. suggested that implants with different macrogeometries affect the IT and ISQ (Gehrke et al. 2023). However, the feature of the hole entrance morphology was not reported in studies before. However, our experiments did not prove that burrs have the same effect on animal bones. This result encourages us to schedule further experiments to explore whether the surrounding morphology affects further osseointegration or whether the powdery debris and burres around the entrance improve implant success.

The PFB used in the study may also be responsible for distinguishing results, as Huwais and Meyer prepared implant sites in porcine tibial plateau bones. Although this PFB can simulate Type IV bone, the density of actual bone is heterogeneous due to the distinct distance from the cortical bone to the bone marrow cavity. Polyurethane blocks with homogeneous density provide reproducibility and control, making them suitable for In Vitro comparisons. This study provided practicable parametric guidance for different osteotomies of human Type IV bone. Besides, polyurethane foam models have mechanical properties similar to bone tissue, and the mechanical properties allow the standardization of procedures by eliminating existing anatomical and structural differences in bone (Gehrke et al. 2018; Romanos et al. 2018). Thus, it can serve as a bone substitute In Vitro alternatives from a biomechanical perspective. Bano et al. revealed that polyurethane foam can be preferred as a bone substitute in biomechanical studies for it can estimate strain in the peri‐implant region (Bano and Shaukat 2025). Our findings on the thrust force, temperature, implant stability, IT and RT, and hole entrance morphology shed light on the clinical practice. While polyurethane foam models offer standardized testing platforms, their inherent material characteristics fundamentally differ from the biomechanical behavior of human bone. The synthetic substances' homogeneous structure and static viscoelastic properties fail to replicate the anisotropic architecture and spring‐back effect inherent to osseous tissue, potentially introducing deviations in implant stability assessments that may not fully correlate with clinical scenarios. In addition, polyurethane foam models cannot provide histological evidence. In the future, to collect stronger evidence, we plan to conduct animal studies, using cone‐beam computed tomography on animal bones before the experiment to ensure consistent bone density conditions. For instance, a study conducted by Frizzera et al. used pig mandibles to evaluate osseodensification, providing valuable insights into bone behavior during implant procedures (Frizzera et al. 2022).

Author Contributions

Xingru Tao: writing – original draft, writing – review and editing, methodology, conceptualization, visualization. Jie Yang: writing – original draft, writing – review and editing, visualization, formal analysis. Tai Ma: investigation. Ming Chen: supervision, project administration. Qinglong An: writing – review and editing, resources, supervision, conceptualization. Dedong Yu: resources, supervision, funding acquisition.

Ethics Statement

The authors have nothing to report.

Consent

The authors have nothing to report.

Conflicts of Interest

The authors declare no conflicts of interest.

Acknowledgments

This study was supported by the National Natural Science Foundation of China (Grant No. 52175422) and Clinical Research Project of Health Inductry of Shanghai Health Commission (Grant No. 202240194), and we thank Dr. Chunfeng Xu for his valuable guidance on this manuscript and language editing.

Xingru Tao and Jie Yang equally to the work and should be regarded as co‐first authors.

Contributor Information

Qinglong An, Email: qlan@sjtu.edu.cn.

Dedong Yu, Email: yudedong@sjtu.edu.cn.

Data Availability Statement

The data supporting the findings of this study are available from the corresponding author upon reasonable request.

References

  1. Alam, K. , Qamar S. Z., Iqbal M., et al. 2023. “Effect of Drill Quality on Biological Damage in Bone Drilling.” Scientific Reports 13, no. 1: 6234. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Alqahtani, A. R. , Desai S. R., Patel J. R., et al. 2023. “Investigating the Impact of Diameters and Thread Designs on the Biomechanics of Short Implants Placed in D4 Bone: A 3D Finite Element Analysis.” BMC Oral Health 23, no. 1: 686. 10.1186/s12903-023-03370-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Bano, N. Z. , and Shaukat A.. 2025. “Comparison of Epoxy Resin and Rigid Polyurethane Foam as a Bone Substitute to Study Implant Biomechanics.” Cureus 17, no. 2: e79124. 10.7759/cureus.79124. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Bergamo, E. T. P. , Zahoui A., Barrera R. B., et al. 2021. “Osseodensification Effect on Implants Primary and Secondary Stability: Multicenter Controlled Clinical Trial.” Clinical Implant Dentistry and Related Research 23, no. 3: 317–328. 10.1111/cid.13007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Bettach, R. , Boukhris G., De Aza P. N., et al. 2023. “New Strategy for Osseodensification During Osteotomy in Low‐Density Bone: An In Vitro Experimental Study.” Scientific Reports 13, no. 1: 11924. 10.1038/s41598-023-39144-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Bhargava, N. , Perrotti V., Caponio V. C. A., Matsubara V. H., Patalwala D., and Quaranta A.. 2023. “Comparison of Heat Production and Bone Architecture Changes in the Implant Site Preparation With Compressive Osteotomes, Osseodensification Technique, Piezoelectric Devices, and Standard Drills: An Ex Vivo Study on Porcine Ribs.” Odontology/the Society of the Nippon Dental University 111, no. 1: 142–153. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Boa, K. , Barrak I., E. Varga, Jr. , Joob‐Fancsaly A., Varga E., and Piffko J.. 2016. “Intraosseous Generation of Heat During Guided Surgical Drilling: An Ex Vivo Study of the Effect of the Temperature of the Irrigating Fluid.” British Journal of Oral & Maxillofacial Surgery 54, no. 8: 904–908. 10.1016/j.bjoms.2016.06.004. [DOI] [PubMed] [Google Scholar]
  8. Bulloch, S. E. , Olsen R. G., and Bulloch B.. 2012. “Comparison of Heat Generation Between Internally Guided (Cannulated) Single Drill and Traditional Sequential Drilling With and Without a Drill Guide for Dental Implants.” International Journal of Oral & Maxillofacial Implants 27, no. 6: 1456–1460. [PubMed] [Google Scholar]
  9. de Carvalho Formiga, M. , Gehrke A. F., De Bortoli J. P., and Gehrke S. A.. 2021. “Can the Design of the Instruments Used for Undersized Osteotomies Influence the Initial Stability of Implants Installed in Low‐Density Bone? An In Vitro Pilot Study.” PLoS One 16, no. 10: e0257985. 10.1371/journal.pone.0257985. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Causey, G. C. , Picha G. J., Price J., Pelletier M. H., Wang T., and Walsh W. R.. 2021. “The Effect of a Novel Pillar Surface Morphology and Material Composition Demonstrates Uniform Osseointegration.” Journal of the Mechanical Behavior of Biomedical Materials 123: 104775. 10.1016/j.jmbbm.2021.104775. [DOI] [PubMed] [Google Scholar]
  11. Chauhan, C. J. , Shah D. N., and Sutaria F. B.. 2018. “Various Bio‐Mechanical Factors Affecting Heat Generation During Osteotomy Preparation: A Systematic Review.” Indian Journal of Dental Research: Official Publication of Indian Society for Dental Research 29, no. 1: 81–92. [DOI] [PubMed] [Google Scholar]
  12. Chen, C. H. , Coyac B. R., Arioka M., et al. 2019. “A Novel Osteotomy Preparation Technique to Preserve Implant Site Viability and Enhance Osteogenesis.” Journal of Clinical Medicine 8, no. 2: 170. 10.3390/jcm8020170. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Chen, J. , An Q., Zou F., Yu D., and M. J. M. Chen. 2021. “Analysis of Low‐Frequency Vibration‐Assisted Bone Drilling in Reducing Thermal Injury.” Materials and Manufacturing Processes 36, no. 1: 27–38. 10.1080/10426914.2020.1813891. [DOI] [Google Scholar]
  14. Eriksson, A. , Albrektsson T., Grane B., and McQueen D.. 1982. “Thermal Injury to Bone.” International Journal of Oral Surgery 11, no. 2: 115–121. 10.1016/s0300-9785(82)80020-3. [DOI] [PubMed] [Google Scholar]
  15. Fontes Pereira, J. , Costa R., Nunes Vasques M., Salazar F., Mendes J. M., and Infante da Câmara M.. 2023. “Osseodensification: An Alternative to Conventional Osteotomy in Implant Site Preparation: A Systematic Review.” Journal of Clinical Medicine 12, no. 22: 7046. 10.3390/jcm12227046. [DOI] [PMC free article] [PubMed] [Google Scholar]
  16. Frisch, E. , Wild V., Ratka‐Krüger P., Vach K., and Sennhenn‐Kirchner S.. 2020. “Long‐Term Results of Implants and Implant‐Supported Prostheses Under Systematic Supportive Implant Therapy: A Retrospective 25‐Year Study.” Clinical Implant Dentistry and Related Research 22, no. 6: 689–696. 10.1111/cid.12944. [DOI] [PubMed] [Google Scholar]
  17. Frizzera, F. , Spin‐Neto R., Padilha V., et al. 2022. “Effect of Osseodensification on the Increase in Ridge Thickness and the Prevention of Buccal Peri‐Implant Defects: An In Vitro Randomized Split Mouth Pilot Study. Article.” BMC Oral Health 22, no. 1: 233. 10.1186/s12903-022-02242-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Gaspar, J. , Borrecho G., Oliveira P., Salvado F., and Martins dos Santos J.. 2013. “Osteotomy at Low‐Speed Drilling Without Irrigation Versus High‐Speed Drilling With Irrigation: An Experimental Study.” Acta Medica Portuguesa 26: 231–236. [PubMed] [Google Scholar]
  19. Gehrke, S. A. , Cortellari G. C., de Oliveira Fernandes G. V., et al. 2023. “Randomized Clinical Trial Comparing Insertion Torque and Implant Stability of Two Different Implant Macrogeometries in the Initial Periods of Osseointegration.” Medicina (Kaunas, Lithuania) 59, no. 1: 168. 10.3390/medicina59010168. [DOI] [PMC free article] [PubMed] [Google Scholar]
  20. Gehrke, S. A. , Guirado J. L. C., Bettach R., Fabbro M. D., Martínez C. P. A., and Shibli J. A.. 2018. “Evaluation of the Insertion Torque, Implant Stability Quotient and Drilled Hole Quality for Different Drill Design: An In Vitro Investigation.” Clinical Oral Implants Research 29, no. 6: 656–662. 10.1111/clr.12808. [DOI] [PubMed] [Google Scholar]
  21. Gehrke, S. A. , Loffredo Neto H., and Mardegan F. E.. 2013. “Investigation of the Effect of Movement and Irrigation Systems on Temperature in the Conventional Drilling of Cortical Bone.” British Journal of Oral & Maxillofacial Surgery 51, no. 8: 953–957. [DOI] [PubMed] [Google Scholar]
  22. Huwais, S. , and Meyer E. G.. 2017. “A Novel Osseous Densification Approach in Implant Osteotomy Preparation to Increase Biomechanical Primary Stability, Bone Mineral Density, and Bone‐to‐Implant Contact.” International Journal of Oral & Maxillofacial Implants 32, no. 1: 27–36. 10.11607/jomi.4817. [DOI] [PubMed] [Google Scholar]
  23. Inchingolo, A. D. , Inchingolo A. M., Bordea I. R., et al. 2021. “The Effectiveness of Osseodensification Drilling Protocol for Implant Site Osteotomy: A Systematic Review of the Literature and Meta‐Analysis.” Materials (Basel, Switzerland) 14, no. 5: 1147. 10.3390/ma14051147. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Kalra, J. , Dhawan P., and Jain N.. 2024. “Implant Stability and Crestal Bone Level in Osseodensification and Conventional Drilling Protocols: A Systematic Review and Meta‐Analysis.” Journal of Prosthetic Dentistry: S0022‐3913(24)00524‐9. 10.1016/j.prosdent.2024.07.036. [DOI] [PubMed] [Google Scholar]
  25. Kim, G. T. , Jin J., Mangal U., et al. 2020. “Primary Stability of Orthodontic Titanium Miniscrews Due to Cortical Bone Density and Re‐Insertion.” Materials (Basel, Switzerland) 13, no. 19: 4433. 10.3390/ma13194433. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Kotsakis, G. A. , and Romanos G. E.. 2022. “Biological Mechanisms Underlying Complications Related to Implant Site Preparation.” Periodontology 2000 88, no. 1: 52–63. 10.1111/prd.12410. [DOI] [PubMed] [Google Scholar]
  27. Lekholm UJTip. 1985. Patient Selection and Preparation. 199–209.
  28. Mello‐Machado, R. C. , Sartoretto S. C., Granjeiro J. M., et al. 2021. “Osseodensification Enables Bone Healing Chambers With Improved Low‐Density Bone Site Primary Stability: An In Vivo Study.” Scientific Reports 11, no. 1: 15436. 10.1038/s41598-021-94886-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Mullings, O. , Tovar N., Abreu de Bortoli J. P., et al. 2021. “Osseodensification Versus Subtractive Drilling Techniques in Bone Healing and Implant Osseointegration: Ex Vivo Histomorphologic/Histomorphometric Analysis in a Low‐Density Bone Ovine Model.” International Journal of Oral & Maxillofacial Implants 36, no. 5: 903–909. 10.11607/jomi.8828. [DOI] [PubMed] [Google Scholar]
  30. Orhan, Z. D. , and Ciğerim L.. 2024. “Evaluation of Effect of Different Insertion Speeds and Torques on Implant Placement Condition and Removal Torque in Polyurethane Dense D1 Bone Model.” Polymers 16, no. 10: 1361. 10.3390/polym16101361. [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. Park, K. J. , Kwon J. Y., Kim S. K., et al. 2012. “The Relationship Between Implant Stability Quotient Values and Implant Insertion Variables: A Clinical Study.” Journal of Oral Rehabilitation 39, no. 2: 151–159. 10.1111/j.1365-2842.2011.02255.x. [DOI] [PubMed] [Google Scholar]
  32. Putra, R. H. , Cooray U., Nurrachman A. S., et al. 2024. “Radiographic Alveolar Bone Assessment in Correlation With Primary Implant Stability: A Systematic Review and Meta‐Analysis.” Clinical Oral Implants Research 35, no. 1: 1–20. 10.1111/clr.14195. [DOI] [PubMed] [Google Scholar]
  33. Romanos, G. E. , Delgado‐Ruiz R. A., Sacks D., and Calvo‐Guirado J. L.. 2018. “Influence of the Implant Diameter and Bone Quality on the Primary Stability of Porous Tantalum Trabecular Metal Dental Implants: An In Vitro Biomechanical Study.” Clinical Oral Implants Research 29, no. 6: 649–655. 10.1111/clr.12792. [DOI] [PubMed] [Google Scholar]
  34. Salomó‐Coll, O. , Auriol‐Muerza B., Lozano‐Carrascal N., Hernández‐Alfaro F., Wang H.‐L., and Gargallo‐Albiol J.. 2021. “Influence of Bone Density, Drill Diameter, Drilling Speed, and Irrigation on Temperature Changes During Implant Osteotomies: An In Vitro Study.” Clinical Oral Investigations 25: 1047–1053. [DOI] [PubMed] [Google Scholar]
  35. Saxena, V. , Dhawan P., and Rani S.. 2024. “Effect of Guided Implant Drilling on Bone Temperature Changes During Implant Osteotomy: A Comprehensive Systematic Review.” Cureus 16, no. 9: e70216. 10.7759/cureus.70216. [DOI] [PMC free article] [PubMed] [Google Scholar]
  36. Slete, F. B. , Olin P., and Prasad H.. 2018. “Histomorphometric Comparison of 3 Osteotomy Techniques.” Implant Dentistry 27, no. 4: 424–428. 10.1097/id.0000000000000767. [DOI] [PubMed] [Google Scholar]
  37. Soldatos, N. , Pham H., Fakhouri W. D., et al. 2022. “Temperature Changes During Implant Osteotomy Preparations in Human Cadaver Tibiae Comparing MIS(®) Straight Drills With Densah(®) Burs.” Genes 13, no. 10: 1716. 10.3390/genes13101716. [DOI] [PMC free article] [PubMed] [Google Scholar]
  38. Standard Specification for Rigid Polyurethane Foam for Use as a Standard Material for Testing Orthopaedic Devices and Instruments . n.d. 10.1520/f1839-08r21. [DOI]
  39. Strbac, G. D. , Unger E., Donner R., Bijak M., Watzek G., and Zechner W.. 2014. “Thermal Effects of a Combined Irrigation Method During Implant Site Drilling. A Standardized In Vitro Study Using a Bovine Rib Model.” Clinical Oral Implants Research 25, no. 6: 665–674. 10.1111/clr.12032. [DOI] [PubMed] [Google Scholar]
  40. Summers, R. B. J. C. 1994. "A New Concept in Maxillary Implant Surgery: The Osteotome Technique." Compendium 15, no. 2: 154‐156. [PubMed] [Google Scholar]
  41. Tabassum, A. , Meijer G. J., Wolke J. G., and Jansen J. A.. 2010. “Influence of Surgical Technique and Surface Roughness on the Primary Stability of an Implant in Artificial Bone With Different Cortical Thickness: A Laboratory Study.” Clinical Oral Implants Research 21, no. 2: 213–220. [DOI] [PubMed] [Google Scholar]
  42. Vaidya, P. V. , Dutta A., Rooj S., et al. 2023. “Design Modification of Surgical Drill Bit for Final Osteotomy Site Preparation Towards Improved Bone‐Implant Contact.” Heliyon 9, no. 6: 16451. [DOI] [PMC free article] [PubMed] [Google Scholar]
  43. Yeh, Y.‐T. , Chu T. G., Blanchard S. B., and Hamada Y.. 2021. “Effects on Ridge Dimensions, Bone Density, and Implant Primary Stability With Osseodensification Approach in Implant Osteotomy Preparation.” International Journal of Oral & Maxillofacial Implants 36, no. 3: 474–484. 10.11607/jomi.8540. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data supporting the findings of this study are available from the corresponding author upon reasonable request.


Articles from Clinical and Experimental Dental Research are provided here courtesy of Wiley

RESOURCES