Skip to main content
. 2025 May 22;17(10):651–660. doi: 10.1080/17576180.2025.2506352

Table 2.

System suitability data for method A in serum.

    Ethylene glycol
Diethylene glycol
Triethylene glycol
SST
Parameter
Specification Day
1
Day
2
Day
3
Day
1
Day
2
Day
3
Day
1
Day
2
Day
3
RT (min) N/A 6.83 6.83 6.83 9.76 9.76 9.76 11.83 11.83 11.83
RSD <2.0% 0.00 0.02 0.03 0.00 0.02 0.02 0.01 0.01 0.01
Area RSD <5.0% 0.38 3.39 1.77 0.33 3.56 1.91 1.28 3.18 2.43
Theoretical Plates >2000 68735 80536 88853 101511 104160 107720 306727 337349 302414
USP
Tailing
Factor
<2.0 0.97 0.98 1.02 0.97 0.96 0.95 1.02 1.01 1.01
RSD <10.0% 3.15 4.84 9.58 4.43 7.88 7.90 4.45 5.95 3.56
Resolution >2.0 >2.0 >2.0 >2.0 >2.0 >2.0 >2.0 >2.0 >2.0 >2.0
Capacity
Factor
k’ > 3 6.18 6.18 6.17 9.25 9.25 9.25 11.42 11.42 11.42
RSD <2.0% 0.00 0.03 0.03 0.00 0.02 0.02 0.01 0.01 0.01

N/A = Not applicable. RT = Retention time. SST = System Suitability Test. AUC = Area Under the Curve. RSD = Relative Standard Deviation. k’ = capacity factor.