Skip to main content
. 2025 Jun 13;120:e240282. doi: 10.1590/0074-02760240282

TABLE II. Performance parameters obtained for the 18 kits licensed for use and commercially available in Brazil, as used in this study.

Kits SEN % (IC95%) SPE % (IC95%) ACC % (IC95%) PLR NLR DOR K (IC95%)
EIABio 100 (96.6-100) 86.8 (77.4-92.7) 94.6 (90.3-97.0) 7.6 0.0001 71,226 0.89 (0.82-0.95)
EIADia 99.0 (94.8-99.8) 90.5 (81.7-95.3) 95.5 (91.5-97.7) 10.5 0.0105 994 0.91 (0.84-0.97)
EIARec 99.1 (95.1-99.8) 80.8 (70.7-88.0) 91.5 (86.7-94.7) 5.2 0.0112 462 0.82 (0.74-0.90)
EIALis 99.1 (95.0-99.8) 78.2 (67.8-85.9) 90.4 (85.4-93.9) 4.5 0.0116 391 0.80 (0.71-0.89)
EIASym 100 (96.6-100) 88.5 (79.5-93.8) 95.2 (91.1-97.5) 8.7 0.0001 84,260 0.90 (0.84-0.96)
EIAEur 92.9 (86.0-96.5) 89.5 (80.6-94.6) 91.4 (86.3-94.7) 8.8 0.0798 111 0.82 (0.74-0.91)
EIAGol 94.5 (88.5-97.5) 86.8 (77.4-92.7) 91.4 (86.4-94.6) 7.2 0.0634 113 0.82 (0.74-0.90)
IHAWie 93.7 (87.6-96.9) 70.9 (60.1-79.7) 84.2 (78.4-88.7) 3.2 0.0890 36 0.67 (0.56-0.78)
IHAGol 71.2 (62.1-78.8) 94.9 (87.7-98.0) 81.1 (74.9-86.0) 14.1 0.3037 46 0.63 (0.52-0.74)
IIFVir 100 (96.5-100) 62.1 (50.1-72.8) 85.5 (79.5-90.0) 2.6 0.0002 17,547 0.67 (0.55-0.79)
IIFCon 87.4 (79.9-92.3) 88.6 (79.7-93.9) 87.9 (82.5-91.8) 7.7 0.1423 54 0.75 (0.66-0.85)
IIFBio 90.1 (83.1-94.4) 73.4 (62.8-81.9) 83.2 (77.2-87.8) 3.4 0.1350 25 0.65 (0.54-0.76)
RDTOns 92.8 (86.3-96.8) 87.3 (78.0-93.8) 90.5 (85.5-93.9) 7.3 0.08 89 0.80 (0.72-0.78)
RDTSdb 95.5 (89.9-98.5) 89.9 (81.0-95.5) 93.2 (88.6-96.0) 9.43 0.05 188 0.86 (0.78-0.93)
RDTWlc 97.3 (92.3-99.4) 92.4 (84.2-97.2) 95.3 (91.2-97.5) 12.8 0.03 438 0.90 (0.84-0.96)
RDTBio 100 (96.7-100) 78.5 (68.2-86.1) 91.1 (86.1-94.3) 4.65 <0.001 40,469 0.81 (0.72-0.90)
CMIAAr 97.3 (92.4-99.1) 81.4 (69.6-89.3) 91.8 (86.7-95.0) 5.2 0.03 157 0.81 (0.72-0.91)

SEN: sensitivity; SPE: specificity; ACC: accuracy; PLR: positive likelihood ratio; NLR: negative likelihood ratio; DOR: diagnostic odds ratio; k: Cohen’s Kappa index. Note: to determine the performance parameters, the number of samples with indeterminate results was disregarded.