Skip to main content
Arthroplasty Today logoLink to Arthroplasty Today
. 2025 Jun 4;33:101732. doi: 10.1016/j.artd.2025.101732

Mid-Flexion Sagittal Stability of Total Knee Arthroplasty Implanted With Kinematic Alignment: A Quantitative Radiographic Laxity Study With Single-Radius Posterior-Stabilized and Condylar-Stabilized Implants

Lauren K Holbrook a, Emma N Horton a, David F Scott a,b,∗
PMCID: PMC12172297  PMID: 40530303

Abstract

Background

Stability between full extension and 90° of flexion is crucial for successful outcomes in total knee arthroplasty (TKA). Thus, we compared the sagittal/anteroposterior (AP) stability of knees implanted with single-radius condylar-stabilized (CS) vs posterior-stabilized (PS) devices from a randomized, prospective study. We hypothesized that the implant group with superior sagittal/AP stability would exhibit better clinical outcomes.

Methods

Sixty-five patients underwent primary TKA with a PS (n = 28) or CS (n = 37) device. Outcomes collected 5-7 years postoperatively included Knee Society (KS) Scores, range of motion (ROM), the Lower Extremity Activity Scale, the Forgotten Joint Score, and mechanical and anatomic alignment. Sagittal/AP stability was evaluated in 45° and 90° of flexion via stress radiographs.

Results

The CS and PS groups averaged 85.8 and 77.3 months post-TKA, respectively, at evaluation. The CS group demonstrated greater mean AP laxity, especially in 45° mid-flexion. Alignment, ROM, and patient-reported outcome scores were similar. However, women in the CS group demonstrated inferior mean KS Function (P = .02) and Total Composite scores (P < .01) compared to women in the PS group. These scores were not significantly different for men or the combined groups.

Conclusions

Mid-flexion sagittal/AP laxity was substantial in both implant groups. With the limited population available, the greater laxity observed in patients with CS implants did not correspond to differences in Forgotten Joint Score, Lower Extremity Activity Scale, ROM, or alignment; however, the difference between the KS Function and Total Composite scores in women revealed an inverse relationship between sagittal laxity and clinical outcomes.

Level of evidence

I.

Keywords: Total knee arthroplasty, Kinematic alignment, Mid-flexion laxity

Introduction

Mid-flexion instability between full extension and 90° of flexion compromises optimal clinical outcomes in total knee arthroplasty (TKA) [[1], [2], [3], [4], [5], [6], [7], [8], [9]]. Patients with mid-flexion instability may experience soft-tissue tenderness, joint effusion, sensations of instability without the knee giving way, and pain, especially when the joint is loaded while ascending/descending stairs or rising from a seated position [5,6,[10], [11], [12]]. In contrast, mid-flexion stability may allow for more natural movement through a full range of motion (ROM) and reduce hamstring recruitment [13,14]. Furthermore, greater sagittal stability corresponded to superior clinical outcomes in a randomized trial [15].

Various surgery-related and patient-related factors can affect mid-flexion stability, including implant design [16]. For example, the post of a posterior-stabilized (PS) implant does not engage until 45°-60° of flexion, which may contribute to mid-flexion instability by creating abnormal kinematics with paradoxical motion in <60° of flexion [17]. Conforming polyethylene insert designs—like the dished, condylar-stabilized (CS) implant—were introduced as alternatives to post and cam PS devices and aim to provide anteroposterior (AP) stability throughout a full ROM [18]. However, Sur et al. found that moderately congruent CS TKA implants exhibit greater laxity in 90° than PS implants, as measured by stress radiographs [19]. They observed equivalent Western Ontario and McMaster Universities Osteoarthritis Index scores between the PS and CS groups but a better combined Total Composite Knee Society (KS) Score (Pain/Motion plus Function) in the PS group, which had less laxity. The publication of this experimental protocol encouraged the principal investigator to perform a similar comparison study evaluating laxity in both 45° (mid-flexion) and 90° of flexion.

The primary objective of the present study was to compare the sagittal/AP stability of knees implanted with CS and PS devices. We defined AP knee stability as minimal anterior-posterior displacement in millimeters on stress radiographs and hypothesized that these 2 implants would demonstrate varying degrees of stability. The secondary objective was to compare the Forgotten Joint Score (FJS), Lower Extremity Activity Scale (LEAS), and KS Scores between the 2 groups. Our secondary hypothesis was that the implant design group with greater AP stability would have better clinical outcome measures (KS scores) and patient-reported outcome measures (PROMs) than those with less AP stability.

Material and methods

Institutional review board approval was obtained for this study, which was conducted between January 2016 and November 2016. Participants were selected from a cohort who participated in a randomized prospective trial (the “parent study”) comparing the Triathlon CS knee and the Triathlon PS knee (Stryker, Mahwah, NJ, USA) [20]. The parent study defined the following inclusion and exclusion criteria:

Parent study inclusion criteria

  • 1.

    Patients willing and able to provide informed consent.

  • 2.

    Patients able to comply with follow-up requirements, including postoperative weight-bearing restrictions and self-evaluations.

  • 3.

    Male and nonpregnant female patients aged 21-80 years at the time of surgery.

  • 4.

    Patients requiring a primary total knee replacement.

  • 5.

    Patients with a diagnosis of osteoarthritis, traumatic arthritis, or avascular necrosis.

  • 6.

    Patients with intact collateral ligaments.

Parent study exclusion criteria

  • 1.

    Patients with inflammatory arthritis.

  • 2.

    Patients with a body mass index >40 kg/m2.

  • 3.

    Patients with a history of total or unicompartmental reconstruction of the affected joint.

  • 4.

    Patients who have had a high tibial osteotomy or femoral osteotomy.

  • 5.

    Patients with neuromuscular or neurosensory impairment that would limit the ability to assess the performance of the implant.

  • 6.

    Patients with a systemic or metabolic disorder leading to progressive bone deterioration.

  • 7.

    Patients who are immunologically compromised or receiving chronic steroids (>30 days).

  • 8.

    Patients with compromised bone stock by disease or infection, which cannot provide adequate support and/or fixation to the prosthesis.

  • 9.

    Patients with knee fusion to the affected joint.

  • 10.

    Patients with a history of infection in the index knee.

  • 11.

    Incarcerated patients.

All participants in the parent study (n = 111) were evaluated against supplemental inclusion/exclusion criteria that were defined to identify optimally functioning TKA implants. Two patients were lost to follow-up before this evaluation.

Supplemental inclusion criteria

  • 1.

    Patients enrolled in the parent study who underwent a primary TKA and do not have evidence of implant failure.

  • 2.

    Patients able to comply with all study requirements, including stress radiographs.

  • 3.

    Patients with an ROM within 0° to 5° of extension and a minimum of 120° of flexion.

Supplemental exclusion criterion

  • 1.

    Patients with hyperextension of the index knee.

Of the remaining participants in the parent study (n = 109), 8 patients did not satisfy the supplemental inclusion/exclusion criteria due to hyperextension (n = 1), ≥ 5° of flexion (n = 5), or < 120° of extension (n = 2) of the implanted knee. Eligible patients (n = 101) were contacted from each implant group for enrollment and scheduling for the present study in an order determined by a random number generator. Selection was blind to an individual’s clinical and functional attributes within the pool of eligible participants. Twenty patients declined to participate, and 16 patients were lost to follow-up between their assessment for eligibility and enrollment. Sixty five patients enrolled. A Consolidated Standards of Reporting Trials-style flowchart was adapted to demonstrate the selection of participants from the randomized parent study (Fig. 1).

Figure 1.

Figure 1

Participant selection flowchart.

Each participant from the parent study was evaluated at a single additional visit between 5 and 8 years postoperatively. A provider measured ROM; KS Scores, LEAS, and FJS outcome surveys were completed; stress radiographs were performed; and mechanical (hip-knee-ankle) and anatomic tibiofemoral alignment measurements were obtained from the 1-year postoperative long-axis radiographs acquired as part of the parent study.

Stress radiograph technique

Following a protocol described by Sur et al., a Telos Stress Device (Telos Medical USA, Austin, TX) applied a 150-Newton force anteriorly and posteriorly to the knee in approximately 45° and 90° of flexion to obtain stress radiographs in the lateral plane [19]. The Telos device was selected for its accuracy and reproducibility for quantitative stress measurements of the knee [[21], [22], [23], [24], [25], [26]]. Subsequently, anterior and posterior displacement measurements were made on stress radiographs by tracing lines along the posterior margin of the tibial component and the posterior edge of the femoral component, which were perpendicular to the tibial baseplate. A standard 25.4-millimeter (1-inch) calibration marker was used to determine the true size of the anatomy in each X-ray by defining the magnification factor. All measurements were adjusted for the magnification factor.

Operative technique

The senior author performed all TKAs without computer assistance using an unrestricted, caliper-verified, kinematic alignment surgical technique with medial parapatellar arthrotomy and intramedullary femoral and extramedullary tibial guides [27]. In all cases, a tourniquet was inflated to 300 mmHg prior to skin incision and deflated after component cementation. All posterior cruciate ligaments (PCLs) were sacrificed, all patellae were resurfaced, and all devices were cemented. The distal femur was resurfaced with measured resection to recreate the native, prearthritic joint line. The posterior slope and varus alignment of the proximal tibia were assessed intraoperatively and matched for each patient. As necessary, adjustments were made to account for cartilage and bone loss. Ligament releases beyond the creation of a medial soft-tissue sleeve during the initial exposure were not performed.

Power analysis

An a priori power analysis was performed to estimate the minimum sample size for an independent 2-sample t-test to compare the mean laxity measurements between 2 groups. The PS group was projected to have a mean mid-flexion laxity of 5 mm with a standard deviation of 2 mm, and the CS group was projected to have a mean mid-flexion laxity of 10 mm with a standard deviation of 5 mm [19]. The effect size (d) was calculated to be 1.3 based on these projections. A 2-tailed test with a significance level (α) of 0.05 and a desired power (1-β) of 80% was performed. This analysis indicated that 11 or more participants per group were necessary to detect the differences in mid-flexion laxity between the 2 groups with adequate statistical power.

Data analyses

Microsoft Excel (Mac version 16.77) and an electronic data capture and analysis software platform (Ortho Research Master, Spokane Joint Replacement Center, Inc., Spokane, WA, USA) were used to compare groups using the student’s t-test. The level of significance was defined as P ≤ .05.

Results

Of the 111 eligible patients recruited from the parent study, 65 participants enrolled, including 28 who received a PS implant and 37 who received a CS implant. Within these groups, 19 women and 18 men received a CS implant, while 15 women and 13 men received a PS implant. There were no statistically significant differences in mean age or body mass index between the groups (Table 1); however, mean years postoperative differed (P < .02). The CS and PS groups averaged 85.8 (range 75-90) and 77.3 (range 70-85) months postoperative, respectively, from their index primary TKA (Table 2).

Table 1.

Mean demographic data.

PS
CS
P values
Participants Age BMI Participants Age BMI Age BMI
All (n = 28) 64.5 (49-79) 32.17 All (n = 37) 63.3 (51-78) 30.34 .53 .09
Men (n = 13) 63.6 (49-79) 31.40 Men (n = 18) 64.3 (51-78) 29.19 .84 .13
Women (n = 15) 65.3 (53-74) 32.83 Women (n = 19) 62.3 (53-75) 31.43 .15 .36

BMI, body mass index; CS, condylar-stabilized; PS, posterior-stabilized.

Age in years (mean and range), body mass index = kg/M2, and statistical significance determined by Student’s t-test.

Table 2.

Mean months postoperative from index arthroplasty.

Measurement PS
CS
P value
All Men Women All Men Women All Men Women
Months postoperative 77.3 78.2 76.5 85.8 83.6 87.4 .02 .03 .02
Standard deviation 11 10 12 12 11 13
Range 70-85 70-84 71-85 75-90 75-88 76-90

CS, condylar-stabilized; PS, posterior-stabilized.

Statistical significance (in bold) determined by Student’s t-test.

The CS group exhibited a higher mean total anterior-posterior displacement in 45° of flexion than the PS group (12.44 mm vs 7.33 mm, P ≤ .0001, Table 3, Fig. 2). In 90° of flexion, the CS group also demonstrated significantly greater total anterior-posterior displacement, almost 3 times the displacement of the PS group (9.74 mm vs 3.54 mm, P ≤ .0001, Table 3, Fig. 2). Randomly chosen representative radiographs illustrating anterior and posterior displacement of the CS and PS implants in 45° of flexion are shown in Figures 3a and b and 4a and b.

Table 3.

Mean laxity (mm) via radiographic stability analysis.

Measurement
PS
CS
P values
Total laxity All Men Women All Men Women All Men Women
45° 7.33 6.50 8.17 12.44 10.78 14.11 <.0001 <.004 <.005
90° 3.54 2.18 4.69 9.74 9.61 9.87 <.0001 <.0002 <.005

CS, condylar-stabilized; PS, posterior-stabilized.

Statistical significance (in bold) determined by Student’s t-test.

Figure 2.

Figure 2

Patients with CS implants exhibited greater mean total anteroposterior (AP) displacement (mm) in 45° and 90° of flexion. Statistical significance was determined by a Student’s t-test. CS, condylar-stabilized.

Figure 3.

Figure 3

Condylar-stabilized (CS) implant (a) anterior and (b) posterior displacement at 45° flexion.

Figure 4.

Figure 4

Posterior-stabilized implant (PS) (a) anterior and (b) posterior displacement at 45° flexion.

The KS Pain/Motion scores (Table 4), LEAS, and FJS (Table 5) were equivalent; however, the KS Function and Total Composite scores varied between women in the PS group vs the CS group. For women, the KS Function (94.67 vs 81.84, P < .02, Fig. 5) and Total Composite (192.80 vs 176.95, P < .01, Fig. 6) scores were superior in the PS group compared to the CS group.

Table 4.

Mean knee society scores.

Measurement PS
CS
P values
All Men Women All Men Women All Men Women
Pain/motion 97.39 96.54 98.13 95.30 95.50 95.11 .25 .71 .22
 Standard deviation 6 8.5 2.6 8.1 7.1 9.1
Function 90.00 84.62 94.67 87.70 93.89 81.84 .60 .17 .02
 Standard deviation 19.6 25.9 10.6 15.2 9.8 17.2
Total score 187.39 181.15 192.8 183.00 189.39 176.95 .39 .33 <.01
 Standard deviation 23.6 32.3 10.4 17.1 12.8 18.8

CS, condylar-stabilized; PS, posterior-stabilized.

Statistical significance (in bold) determined by Student’s t-test.

Table 5.

Mean LEAS and FJS.

Instrument PS
CS
P value
All Men Women All Men Women All Men Women
LEAS 10.64 10.54 10.73 10.51 11.44 9.63 .86 .45 .19
 Standard deviation 3.3 4.1 2.4 2.5 2.4 2.3
FJS 62.57 61.22 63.75 68.64 72.11 65.35 .41 .32 .88
 Standard deviation 32.3 34.3 31.7 27 25.7 28.5

CS, condylar-stabilized; FJS, Forgotten Joint Score; LEAS, Lower Extremity Activity Scale; PS, posterior-stabilized.

Statistical significance determined by Student’s t-test.

Figure 5.

Figure 5

Women with PS implants achieved better mean Knee Society Function Scores than women with CS implants. Statistical significance was determined by a Student’s t-test. CS, condylar-stabilized; PS, posterior-stabilized.

Figure 6.

Figure 6

Patients with PS implants achieved better mean total Knee Society Scores than women with CS implants. Statistical significance was determined by a Student’s t-test. CS, condylar-stabilized; PS, posterior-stabilized.

Neither ROM nor hip-knee-ankle and anatomic tibiofemoral angular measurements significantly differed between the PS and CS groups (Table 6).

Table 6.

Mean results of ROM and radiographic knee angulation.

Measurement CS
PS
P value
All Men Women All Men Women All Men Women
Extension° 0.18 0.38 0.00 0.24 0.22 0.26 .80 .70 .38
 Standard deviation 0.9 1.4 0 1 0.9 1.2
Flexion° 127.46 128.77 126.33 126.19 125.39 126.95 .39 .12 .77
 Standard deviation 5.6 6.8 4.3 6.1 4.8 7.2
ATF angle° 5.80 6.18 5.50 6.06 6.86 5.55 .67 .50 .95
 Standard deviation 1.9 1.9 2 1.9 2.39 1.6
Alignmenta 181.2 179.12 181.77 179.62 179.11 180.36 .56 .66 .43
 Standard deviation 2.0 1.9 2.2 1.8 1.7 1.9

ATF, anatomic tibiofemoral; CS, condylar-stabilized; PS, posterior-stabilized; ROM, range of motion.

a

Mechanical hip-knee-ankle alignment. Statistical significance determined by Student’s t-test.

Discussion

The term mid-flexion instability emerged from a cadaver study to describe a distinct phenomenon of increased varus-valgus excursion between 30° and 60° flexion yet stability at 90° flexion and full extension [28]. Instability accounts for up to 20% of TKA revisions, emphasizing the importance of understanding the various types of instability and their associated risk factors [29]. Consequently, mid-flexion instability continues to be defined and examined in clinical trials and computational models, with some using it to reference laxity in the sagittal/AP plane [4,7,9,16,[30], [31], [32], [33], [34]]. Some of these studies, including the present study, explore the susceptibility of various implant designs to mid-flexion laxity [15,34].

Our findings support our primary hypothesis that the studied CS and PS implants are associated with different sagittal/AP stability. Consistent with 90° laxity data presented by Sur et al., PS knees in this study provided superior AP stability in 45° and 90° of flexion compared to CS knees [19]. This difference in stability did not correspond to appreciable differences in patient outcomes, with the exception that the KS Function and Total Composite Scores were better for women in the PS implant subgroup (94.67 vs 81.84, P < .02), which exceeds the minimal clinically important difference of 6.1-6.4 for this score [35]. Therefore, the secondary hypothesis that participants with greater AP knee stability would have improved clinical outcomes compared to those with less AP stability was only partially supported.

Unlike Sur et al.’s previous findings that all patients with PS implants exhibit less mean laxity at 90° and a greater mean combined KS score than patients with CS implants, the inverse relationship between sagittal laxity and clinical outcomes only applied to women in the present study. Several limitations may explain these results. First, the PROMs used in this study may not distinguish subtle differences in otherwise well-functioning knees due to a ceiling effect. With only 65 study participants, our ability to discern small differences in outcome scores and other clinical parameters may have been limited and underpowered. An analysis of a larger sample population is warranted to explore the significant inverse relationship between laxity and clinical outcomes observed in women.

The difference in mean postoperative time between the PS and CS groups in the present study is another limitation. Although we would not expect a 10-month postoperative time difference—7 years and 3 months (CS) vs 6 years and 5 months (PS)—to correspond to clinically significant disparities, PROMs and sagittal stability might change over this period [36,37]. Additionally, the present study quantified the knee’s sagittal stability under static loading conditions, assuming our measurements approximated the anterior-posterior displacement patients experience during daily activities; however, dynamic loading may subject the knee to unique forces and recruit additional soft tissue structures that lead to different results and conclusions.

Beyond these limitations, the magnitude of AP laxity observed in the present study suggests that these single-radius CS and PS implants are effectively anterior cruciate ligament-deficient and PCL-deficient in mid-flexion. If this were true, the overall clinical outcome scores may not have significantly differed between the 2 implant groups despite a significant difference in laxity, because these scores were adversely impacted by substantial mid-flexion laxity. In other words, participants in both implant groups may have possessed a degree of sagittal laxity that adversely impacted the outcome parameters. Support for this conclusion comes from a report comparing the mid-flexion AP stability of PS vs medial-pivot (MP) implants using the same radiographic stress testing protocol as the present study [15]. The MP implant design features a distinct geometry from the PS and CS implants and is produced by a different manufacturer. The lesser AP laxity of 3.6 mm observed in the MP group in 45° of flexion in the report contrasts with the 12.44 mm and 7.33 mm of laxity observed in the CS and PS groups, respectively, in the present study. Based on Longo et al.’s findings that ≥ 7 mm of anterior-posterior displacement is an independent risk factor for mid-flexion instability in 30° of flexion, patients with CS and PS implants may have a higher risk of mid-flexion instability than patients with MP implants because both groups’ mean mid-flexion laxity exceeds this threshold [3]. Furthermore, these sagittal translation measurements observed in varying post-TKA patient groups exceed the expected sagittal translation of a healthy, native knee. The total sagittal translation of the medial tibio-femoral compartment of a healthy knee is less than a few millimeters, while translation of the lateral compartment can reach 20 mm [[38], [39], [40]].

The magnitude of mid-flexion instability observed in the present study may be attributed to various causes. For instance, the CS implant in the present study is only moderately congruent—not a true “deep-dish” or ultra-congruent design—so it may not provide sufficient stability in the sagittal plane, especially in the absence of the PCL. If the PCL had been retained in the CS implant group, reduced posterior translation of the CS implant may have been observed, leading to decreased laxity and potentially different results. Similarly, the PS implant in the present study may not be equivalent to other PS designs. Differences in tibiofemoral conformity and the degree of flexion required to engage the cam may impact PS implant stability.

Alternatively, the sagittal laxity observed in the present study might be attributed to the surgical technique. Although experts disagree, some evidence suggests that increasing the posterior slope in patients receiving a PS implant and releasing the PCL can both cause instability [[41], [42], [43]]. However, the posterior slope angle was anatomically matched, and the PCL was released in both implant groups in the present study, so we do not anticipate these factors confounded our results. Furthermore, the sagittal laxity cannot be attributed to the kinematic alignment approach. Previous research demonstrates that a kinematic alignment surgical technique identical to the procedure used in the present study leads to substantially less mid-flexion laxity in knees implanted with ball-in-socket designs [15].

That said, we acknowledge that the kinematic alignment surgical approach used in the present study may not be directly comparable to traditional mechanical alignment or gap-balancing approaches. Kinematic alignment rarely requires ligament releases. Bone cuts can typically achieve knee balancing unless a patient presents with significant preoperative flexion and valgus contractures. Thus, our findings cannot extend to other surgical approaches where ligament releases are routinely performed because these differences may affect the stability and kinematics of the knee [44,45].

Conclusions

Mid-flexion sagittal laxity was observed in both PS and CS single-radius implant groups. Based on the limited numbers available, our primary hypothesis that mid-flexion laxity would differ between these implant groups was supported, and our secondary hypothesis that the implant group with less sagittal/AP laxity would have improved outcome scores was partially supported. We encourage the continued exploration of the relationships among implant design, mid-flexion stability, and clinical outcomes.

Funding

Stryker (Mahwah, NJ, USA) provided financial support for this research.

Conflicts of interest

David F. Scott received royalties from Innomed; is a paid consultant for Microport and Medacta; received research support from Microport, Medacta, and Stryker as a Principal Investigator; and is the reviewer of Journal of Arthroplasty, Knee, Journal of Knee Surgery, and Journal of Bone & Mineral. All other authors declare no potential conflicts of interest.

CRediT authorship contribution statement

Lauren K. Holbrook: Writing – review & editing, Writing – original draft, Visualization, Validation, Methodology, Investigation, Formal analysis, Data curation, Conceptualization. Emma N. Horton: Writing – review & editing. David F. Scott: Writing – review & editing, Supervision, Software, Resources, Project administration, Methodology, Investigation, Funding acquisition, Formal analysis, Data curation, Conceptualization.

Appendix A. Supplementary data

Conflict of Interest Statement for Scott
mmc1.docx (27KB, docx)
Conflict of Interest Statement for Horton
mmc2.docx (18.1KB, docx)
Conflict of Interest Statement for

Holbrook

mmc3.docx (18.1KB, docx)

References

  • 1.Incavo S.J., Schmid S., Sreenivas K., Ismaily S., Noble P.C. Total knee arthroplasty using anatomic alignment can produce mid-flexion laxity. Clin Biomech (Bristol) 2013;28:429–435. doi: 10.1016/j.clinbiomech.2013.02.012. [DOI] [PubMed] [Google Scholar]
  • 2.Kappel A., Laursen M., Nielsen P.T., Odgaard A. Relationship between outcome scores and knee laxity following total knee arthroplasty: a systematic review. Acta Orthop. 2019;90:46–52. doi: 10.1080/17453674.2018.1554400. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Longo U.G., Candela V., Pirato F., Hirschmann M.T., Becker R., Denaro V. Midflexion instability in total knee arthroplasty: a systematic review. Knee Surg Sports Traumatol Arthrosc. 2021;29:370–380. doi: 10.1007/s00167-020-05909-6. [DOI] [PubMed] [Google Scholar]
  • 4.Luyckx T., Vandenneucker H., Ing L.S., Vereecke E., Ing A.V., Victor J. Raising the joint line in TKA is associated with mid-flexion laxity: a study in cadaver knees. Clin Orthop Relat Res. 2018;476:601–611. doi: 10.1007/s11999.0000000000000067. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Matsumoto K., Ogawa H., Yoshioka H., Akiyama H. Postoperative anteroposterior laxity influences subjective outcome after total knee arthroplasty. J Arthroplasty. 2017;32:1845–1849. doi: 10.1016/j.arth.2016.12.043. [DOI] [PubMed] [Google Scholar]
  • 6.Mochizuki T., Tanifuji O., Sato T., Hijikata H., Koga H., Watanabe S., et al. Association between anteroposterior laxity in mid-range flexion and subjective healing of instability after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2017;25:3543–3548. doi: 10.1007/s00167-016-4375-8. [DOI] [PubMed] [Google Scholar]
  • 7.Nagle M., Glynn A. Midflexion instability in primary total knee arthroplasty. Knee Surg. 2020;33:459–465. doi: 10.1055/s-0039-1678537. [DOI] [PubMed] [Google Scholar]
  • 8.Nowakowski A.M., Majewski M., Müller-Gerbl M., Valderrabano V. Measurement of knee joint gaps without bone resection: "Physiologic" extension and flexion gaps in total knee arthroplasty are asymmetric and unequal and anterior and posterior cruciate ligament resections produce different gap changes. J Orthop Res. 2012;30:522–527. doi: 10.1002/jor.21564. [DOI] [PubMed] [Google Scholar]
  • 9.Tsubosaka M., Ishida K., Kodato K., Shibanuma N., Hayashi S., Kurosaka M., et al. Mid-flexion stability in the anteroposterior plane is achieved with a medial congruent insert in cruciate-retaining total knee arthroplasty for varus osteoarthritis. Knee Surg Sports Traumatol Arthrosc. 2021;29:467–473. doi: 10.1007/s00167-020-05927-4. [DOI] [PubMed] [Google Scholar]
  • 10.Pagnano M.W., Hanssen A.D., Lewallen D.G., Stuart M.J. Flexion instability after primary posterior cruciate retaining total knee arthroplasty. Clin Orthop Relat Res. 1998:39–46. doi: 10.1097/00003086-199811000-00008. [DOI] [PubMed] [Google Scholar]
  • 11.Schwab J.H., Haidukewych G.J., Hanssen A.D., Jacofsky D.J., Pagnano M.W. Flexion instability without dislocation after posterior stabilized total knees. Clin Orthop Relat Res. 2005;440:96–100. doi: 10.1097/00003086-200511000-00018. [DOI] [PubMed] [Google Scholar]
  • 12.Vince K. Mid-flexion instability after total knee arthroplasty: woolly thinking or a real concern? Bone Jt J. 2016;98-b(1 Suppl A):84–88. doi: 10.1302/0301-620X.98B1.36445. [DOI] [PubMed] [Google Scholar]
  • 13.Banks S.A., Hodge W.A. 2003 Hap Paul Award paper of the International Society for Technology in Arthroplasty: design and activity dependence of kinematics in fixed and mobile-bearing knee arthroplasties. J Arthroplasty. 2004;19:809–816. doi: 10.1016/j.arth.2004.04.011. [DOI] [PubMed] [Google Scholar]
  • 14.Wang H., Simpson K.J., Ferrara M.S., Chamnongkich S., Kinsey T., Mahoney O.M. Biomechanical differences exhibited during sit-to-stand between total knee arthroplasty designs of varying radii. J Arthroplasty. 2006;21:1193–1199. doi: 10.1016/j.arth.2006.02.172. [DOI] [PubMed] [Google Scholar]
  • 15.Scott D.F., Hellie A.A. Mid-Flexion, anteroposterior stability of total knee replacement implanted with kinematic alignment: a randomized, quantitative radiographic laxity study with posterior-stabilized and medial-stabilized implants. J Bone Joint Surg Am. 2022;105:9–19. doi: 10.2106/JBJS.2122.00549. [DOI] [PubMed] [Google Scholar]
  • 16.Mehta N., Burnett R.A., Kahlenberg C.A., Miller R., Chalmers B., Cross M.B. Mid-Flexion instability after total knee arthroplasty: diagnosis, implant design, and outcomes. Orthopedics. 2023;46:e13–e19. doi: 10.3928/01477447-20220719-01. [DOI] [PubMed] [Google Scholar]
  • 17.Angerame M.R., Holst D.C., Jennings J.M., Komistek R.D., Dennis D.A. Total knee arthroplasty kinematics. J Arthroplasty. 2019;34:2502–2510. doi: 10.1016/j.arth.2019.05.037. [DOI] [PubMed] [Google Scholar]
  • 18.Scott D.F., Smith R.R. A prospective, randomized comparison of posterior stabilized versus cruciate-substituting total knee arthroplasty: a preliminary report with minimum 2-year results. J Arthroplasty. 2014;29(9 Suppl):179–181. doi: 10.1016/j.arth.2014.03.050. [DOI] [PubMed] [Google Scholar]
  • 19.Sur Y.J., Koh I.J., Park S.W., Kim H.J., In Y. Condylar-stabilizing tibial inserts do not restore anteroposterior stability after total knee arthroplasty. J Arthroplasty. 2015;30:587–591. doi: 10.1016/j.arth.2014.11.018. [DOI] [PubMed] [Google Scholar]
  • 20.Scott D.F. Prospective randomized comparison of posterior-stabilized versus condylar-stabilized total knee arthroplasty: final report of a five-year study. J Arthroplasty. 2018;33:1384–1388. doi: 10.1016/j.arth.2017.11.037. [DOI] [PubMed] [Google Scholar]
  • 21.Ryu D.J., Kwon K.B., Jung E.Y., Lee S.S., Kim J.H., Jang M.C., et al. Clinically reliable knee flexion angle measured on stress radiography for quantifying posterior instability in posterior cruciate ligament injury. Orthop J Sports Med. 2021;9 doi: 10.1177/2325967121989252. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Kappel A., Mortensen J.F., Nielsen P.T., Odgaard A., Laursen M. Reliability of stress radiography in the assessment of coronal laxity following total knee arthroplasty. Knee. 2020;27:221–228. doi: 10.1016/j.knee.2019.09.013. [DOI] [PubMed] [Google Scholar]
  • 23.Panisset J.C., Ntagiopoulos P.G., Saggin P.R., Dejour D. A comparison of Telos™ stress radiography versus Rolimeter™ in the diagnosis of different patterns of anterior cruciate ligament tears. Orthop Traumatol Surg Res. 2012;98:751–758. doi: 10.1016/j.otsr.2012.07.003. [DOI] [PubMed] [Google Scholar]
  • 24.Garavaglia G., Lubbeke A., Dubois-Ferrière V., Suva D., Fritschy D., Menetrey J. Accuracy of stress radiography techniques in grading isolated and combined posterior knee injuries: a cadaveric study. Am J Sports Med. 2007;35:2051–2056. doi: 10.1177/0363546507306466. [DOI] [PubMed] [Google Scholar]
  • 25.Jung T.M., Reinhardt C., Scheffler S.U., Weiler A. Stress radiography to measure posterior cruciate ligament insufficiency: a comparison of five different techniques. Knee Surg Sports Traumatol Arthrosc. 2006;14:1116–1121. doi: 10.1007/s00167-006-0137-3. [DOI] [PubMed] [Google Scholar]
  • 26.Schulz M.S., Russe K., Lampakis G., Strobel M.J. Reliability of stress radiography for evaluation of posterior knee laxity. Am J Sports Med. 2005;33:502–506. doi: 10.1177/0363546504269723. [DOI] [PubMed] [Google Scholar]
  • 27.Scott D.F., Howell S.M. Kinematic alignment possible with manual instrumentation, medially stabilized implantsOrthopedics Today. 2019. https://www.healio.com/news/orthopedics/20190211/kinematic-alignment-possible-with-manual-instrumentation-medially-stabilized-implants
  • 28.Martin J.W., Whiteside L.A. The influence of joint line position on knee stability after condylar knee arthroplasty. Clin Orthop Relat Res. 1990:146–156. [PubMed] [Google Scholar]
  • 29.Sharkey P.F., Lichstein P.M., Shen C., Tokarski A.T., Parvizi J. Why are total knee arthroplasties failing today—has anything changed after 10 years? J Arthroplasty. 2014;29:1774–1778. doi: 10.1016/j.arth.2013.07.024. [DOI] [PubMed] [Google Scholar]
  • 30.Vajapey S.P., Pettit R.J., Li M., Chen A.F., Spitzer A.I., Glassman A.H. Risk factors for mid-flexion instability after total knee arthroplasty: a systematic review. J Arthroplasty. 2020;35:3046–3054. doi: 10.1016/j.arth.2020.05.026. [DOI] [PubMed] [Google Scholar]
  • 31.Ramappa M. Midflexion instability in primary total knee replacement: a review. SICOT J. 2015;1:24. doi: 10.1051/sicotj/2015020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Del Gaizo D.J., Della Valle C.J. Instability in primary total knee arthroplasty. Orthopedics. 2011;34:e519–e521. doi: 10.3928/01477447-20110714-46. [DOI] [PubMed] [Google Scholar]
  • 33.Evangelista P.J., Laster S.K., Lenz N.M., Sheth N.P., Schwarzkopf R. A computer model of mid-flexion instability in a balanced total knee arthroplasty. J Arthroplasty. 2018;33:S265–S269. doi: 10.1016/j.arth.2018.02.021. [DOI] [PubMed] [Google Scholar]
  • 34.Hino K., Ishimaru M., Iseki Y., Watanabe S., Onishi Y., Miura H. Mid-flexion laxity is greater after posterior-stabilised total knee replacement than with cruciate-retaining procedures: a computer navigation study. Bone Jt J. 2013;95-b:493–497. doi: 10.1302/0301-620X.95B4.30664. [DOI] [PubMed] [Google Scholar]
  • 35.Lee W.C., Kwan Y.H., Chong H.C., Yeo S.J. The minimal clinically important difference for Knee Society Clinical Rating System after total knee arthroplasty for primary osteoarthritis. Knee Surg Sports Traumatol Arthrosc. 2017;25:3354–3359. doi: 10.1007/s00167-016-4208-9. [DOI] [PubMed] [Google Scholar]
  • 36.Benjamin J., Johnson R., Porter S. Knee scores change with length of follow-up after total knee arthroplasty. J Arthroplasty. 2003;18:867–871. doi: 10.1016/s0883-5403(03)00264-x. [DOI] [PubMed] [Google Scholar]
  • 37.Carlson V.R., Post Z.D., Orozco F.R., Davis D.M., Lutz R.W., Ong A.C. When does the knee feel normal again: a cross-sectional study assessing the Forgotten Joint Score in patients after total knee arthroplasty. J Arthroplasty. 2018;33:700–703. doi: 10.1016/j.arth.2017.09.063. [DOI] [PubMed] [Google Scholar]
  • 38.Pinskerova V., Vavrik P. In: Personalized hip and knee joint replacement. Rivière C., Vendittoli P.-A., editors. Springer International Publishing; Cham: 2020. Knee anatomy and biomechanics and its relevance to knee replacement; pp. 159–168. [PubMed] [Google Scholar]
  • 39.Iwaki H., Pinskerova V., Freeman M.A.R. Tibiofemoral movement 1: the shapes and relative movements of the femur and tibia in the unloaded cadaver knee. J Bone Joint Surg Br. 2000;82-B:1189–1195. doi: 10.1302/0301-620x.82b8.10717. [DOI] [PubMed] [Google Scholar]
  • 40.Brantigan O.C., Voshell A.F. The mechanics of the ligaments and menisci of the knee joint. J Bone Joint Surg Am. 1941;23 [PubMed] [Google Scholar]
  • 41.Okamoto S., Mizu-uchi H., Okazaki K., Hamai S., Nakahara H., Iwamoto Y. Effect of tibial posterior slope on knee kinematics, quadriceps force, and patellofemoral contact force after posterior-stabilized total knee arthroplasty. J Arthroplasty. 2015;30:1439–1443. doi: 10.1016/j.arth.2015.02.042. [DOI] [PubMed] [Google Scholar]
  • 42.Mihalko W.M., Krackow K.A. Posterior cruciate ligament effects on the flexion space in total knee arthroplasty. Clin Orthop Relat Res. 1999:243–250. doi: 10.1097/00003086-199903000-00029. [DOI] [PubMed] [Google Scholar]
  • 43.Matthews J., Chong A., McQueen D., O'Guinn J., Wooley P. Flexion-extension gap in cruciate-retaining versus posterior-stabilized total knee arthroplasty: a cadaveric study. J Orthop Res. 2014;32:627–632. doi: 10.1002/jor.22587. [DOI] [PubMed] [Google Scholar]
  • 44.Ishii Y., Noguchi H., Takeda M., Sato J., Toyabe S. Anteroposterior translation does not correlate with knee flexion after total knee arthroplasty. Clin Orthop Relat Res. 2014;472:704–709. doi: 10.1007/s11999-013-3274-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Wautier D., Thienpont E. Changes in anteroposterior stability and proprioception after different types of knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2017;25:1792–1800. doi: 10.1007/s00167-016-4038-9. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Conflict of Interest Statement for Scott
mmc1.docx (27KB, docx)
Conflict of Interest Statement for Horton
mmc2.docx (18.1KB, docx)
Conflict of Interest Statement for

Holbrook

mmc3.docx (18.1KB, docx)

Articles from Arthroplasty Today are provided here courtesy of Elsevier

RESOURCES