Establishing Quality Control Procedures for Large-Scale Plasma Proteomics Analyses

Khiry L Patterson 1, Albert B Arul 2, Min Ji Choi 3, Nekesa C Oliver 4, Marsalas D Whitaker 5, Angel C Bodrick 6, Julia B Libby 7, Shania Hansen 8, Logan Dumitrescu 9, Katherine A Gifford 10, Angela L Jefferson 11, Timothy J Hohman 12, Renã A S Robinson 13
PMCID: PMC12204224  NIHMSID: NIHMS2091340  PMID: 37163770

Abstract

Proteomics research has been transformed due to high-throughput liquid chromatography (LC-MS/MS) tandem mass spectrometry instruments combined with highly sophisticated automated sample preparation and multiplexing workflows. However, scaling proteomics experiments to large sample cohorts (hundreds to thousands) requires thoughtful quality control (QC) protocols. Robust QC protocols can help with reproducibility, quantitative accuracy, and provide opportunities for more decisive troubleshooting. Our laboratory conducted a plasma proteomics study of a cohort of N = 335 patient samples using tandem mass tag (TMTpro) 16-plex batches. Over the course of a 10-month data acquisition period for this cohort we collected 271 pooled QC LC-MS/MS result files obtained from MS/MS analysis of a patient-derived pooled plasma sample, representative of the entire cohort population. This sample was tagged with TMTzero or TMTpro reagents and used to inform the daily performance of the LC-MS/MS instruments and to allow within and across sample batch normalization. Analytical variability of a number of instrumental and data analysis metrics including protein and peptide identifications, peptide spectral matches (PSMs), number of obtained MS/MS spectra, average peptide abundance, percent of peptides with a Δ m/z between ±0.003 Da, percent of MS/MS spectra obtained at the maximum injection time, and the retention time of selected tracking peptides were evaluated to help inform the design of a robust LC-MS/MS QC workflow for use in future cohort studies. This study also led to general tips for using selected metrics to inform real-time troubleshooting of LC-MS/MS performance issues with daily QC checks.

Keywords: liquid chromatography, mass spectrometry, plasma, proteins, proteomics, quality control

Graphical Abstract

graphic file with name nihms-2091340-f0001.jpg

INTRODUCTION

Liquid chromatography tandem mass spectrometry (LC-MS/MS)-based proteomics continues to be an integral and relevant approach for discoveries in biomedical research that include cancer,1 sepsis,2,3 autoimmune disorders,4,5 and Alzheimer’s disease.6–9 As an analytical platform LC-MS/MS-based proteomics can provide absolute and relative protein and peptide abundances. Additionally, this platform can tackle proteome complexity and provide high sample throughput with well-developed sample preparation techniques, chromate-graphic separations, and MS/MS data acquisition strategies.10–16 However, the value and reproducibility of the analytical results from MS-based proteomics is broadly dependent upon the quality of the data obtained.

Quality control (QC) approaches in MS-based proteomics have been reviewed and focus on solutions for improved reproducibility,17 specifically during sample preparation and LC-MS/MS analyses.18,19 These reviews suggest that in order to manage variability in proteomics workflows, an efficient application of robust protocols, automation, and validated standard operating procedures (SOPs) is critical. However, there is no consensus on the ideal QC methodology to employ for the routine monitoring of LC and MS/MS experiments. For example, multiple QC sample options exist ranging from digested single protein samples derived from bovine serum albumin to more complex HeLa cell samples.20–23 The former can be used for quick QC checks while the latter may mimic more the complexity of study samples. However, the selection of any QC sample type should be informative of the QC metrics regarding the performance of the sample and instrument workflow. Arguably, when considering a robust QC protocol for analysis of clinical studies, pooled patient-derived samples may provide a more direct sample matrix that cannot be achieved with plasma standards or other complex digests. However, the experimental study design and availability of patient samples may be limiting factors.

In addition to various QC sample types (e.g., single protein digests,22 whole-cell lysates24), multiple QC software tools have also been developed (e.g., MSstatsQC,25 iMonDB,26 QC-ART,27 PTXQC28) to help assess LC-MS/MS performance by computing and plotting a number of advanced QC metrics. However, deciding which QC software tool to use and onboarding of these tools can be difficult. For example, users may find a specific metric or plot challenging to interpret in the context of what corrective action is needed or the software layout is hard to navigate.17 Furthermore, the most sophisticated QC strategy will not be appropriate across all samples or projects without the need for laboratory-specific modification.

Our goal was to develop a robust QC approach that could be applied to a LC-MS/MS-based proteomics workflow of plasma samples. Specifically, we were interested in introducing measures into the data acquisition period that would ensure optimal LC-MS/MS data were obtained across batches of TMT-labeled samples in clinical cohort studies. We initiated QC measures that primarily relied on protein and peptide counts for a cohort of 335 patient samples. However, from an initial proteomics analysis it was clear more stringent QC measures would be necessary to inform real-time adjustments to LC-MS/MS instrumental parameters necessary to maintain optimal performance. Thus, we established additional QC metrics (i.e., peptide retention time, peptide spectral matches (PSMs), MS/MS spectra, peptide abundance (x¯), %m/z, and %ITmax) and evaluated their ability to establish normal operating limits for each metric in future cohort studies. Here, we present the analyses of these metrics from 271 QC injections and discuss the generation of a LC-MS/MS QC workflow for longitudinal monitoring across large-scale proteomics studies.

MATERIALS AND METHODS

Sample Selection.

Patient-derived human plasma samples (N = 335) collected from cognitively normal (N = 168, age 73 ± 8 years, 41% female) and mild cognitively impaired (N = 167, age 72 ± 7 years, 41% female) participants were obtained from the Vanderbilt Memory and Aging Project (VMAP), a case-control longitudinal study that examines vascular health and brain aging.1 Written informed consent was obtained from participants, and secondary analyses of all data was carried out in accordance with Vanderbilt University and the Vanderbilt University Medical Center Institutional Review Board (IRB)-approved protocols.

Plasma Depletion.

A 1:4 plasma:buffer dilution was generated for each 40 μL crude patient plasma sample using in-house Buffer A containing 1.71 mM NaCl, 3.38 μM KCl, 1.25 mM Na2HPO4, and 2.25 μM KH2PO4 in Optima LC-MS water (ThermoFisher Scientific, Waltham). The dilution was centrifuged at 16,000g for 1 min at 4 °C through a 0.22 μM filter (MilliporeSigma, Burlington). The sample flow-through was injected onto a Multiple Affinity Removal System (MARS) Human 14 (4.6 × 100 mm, Agilent, Santa Clara) column using a Waters Alliance e2695 Separation module (Milford) high-performance liquid chromatography (HPLC) system and the unbound fractions were collected into an Eppendorf tube. The unbound fraction was concentrated using a 10 kDa molecular weight cutoff concentrator filter (MilliporeSigma, Burlington) for 4 min at 14,000g and 4 °C. The concentrate was collected and the concentration determined by bicinchoninic acid assay (ThermoFisher Scientific, Waltham). Human plasma standard (MilliporeSigma, Burlington) was used as a quality control (QCstd) sample to monitor for optimal daily HPLC performance during the depletion protocol.

Automated Protein Digestion.

Using a 96-well plate design on a Biomek i7 Automated Workstation (Beckman Coulter, Brea), each depleted sample (100 μg) was randomized into 23 batches with each consisting of 15 participant samples and one QCstd sample per batch to monitor the digestion protocol. Each depleted patient and QCstd sample was prepared in 1 M ammonium bicarbonate (Honeywell, Charlotte) buffer and the volume normalized with LC-MS grade water. Proteins were reduced by 200 mM dithiothreitol (DTT) (ThermoFisher Scientific, Waltham) for 45 min at 55 °C followed by alkylation in 200 mM iodoacetamide (IAM, 98%) (ThermoFisher Scientific, Waltham) for a 30 min incubation period in the dark at 25 °C. In-solution digestion of proteins was performed using trypsin/Lys-C (Promega, Madison) with a 1:50 enzyme:substrate ratio for 14 h at 37 °C. The digested samples were acidified with 5% formic acid (ThermoFisher Scientific, Waltham) to a pH < 3 and desalted with 0.1% formic acid using BioPureSPE C18 96-well plates (The Nest Group, Ipswich) with an automated robotic liquid handler (Biomek i7) equipped with a Positive Pressure X-Well SPE Extractor (Ultimaration, Rostock). Samples were eluted with 60:40 acetonitrile (ThermoFisher Scientific, Waltham):H2O with 0.1% formic acid and dried with centrifuged evaporation.

Sample Pooling and Tagging.

Each sample was reconstituted in 100 mM tetraethylammonium bromide (TEAB, pH 8.5) (MilliporeSigma, Burlington). A pooled-plasma stock solution (Spool) was prepared by combining 5 μg of peptide from each reconstituted patient-derived sample. Automated TMTpro 16-plex labeling using a Biomek i7 was performed following manufacturer’s instruction with peptides from patient samples (25 μg/sample) and included Spool peptides (25 μg). Specifically, each 16-plex TMTpro batch contained patient samples (N = 15) and Spool (N = 1). The Spool sample after TMTpro labeling is referred to as QCbatch. There were 23 16-plex batches in total. The QCbatch channel was generated to allow within batch and across batch normalization. TMTpro reporter ion channel assignments were randomized among patient and Spool samples. Each batch was desalted, dried down, and separated into 24 fractions using high-pH (pH = 10) reversed-phase fractionation on a Zorbax Extend300 C-18 analytical column (4.6 mm i.d. × 12.5 mm, 5 μm) (Agilent, Santa Clara) on a Waters e2696 HPLC system. The fractions were dried by centrifugal evaporation and reconstituted to 0.25 μg/μL in 0.1% formic acid.

Additionally, Spool was used to generate a QC sample (QCpool) for daily LC-MS/MS performance monitoring. In order to evaluate reporter ion fragmentation, aliquots of Spool were labeled with TMTzero (ThermoFisher Scientific, Waltham). After labeling, tagged peptides were desalted on a 10 mg Oasis HLB cartridge (Waters Corp., Milford). The eluent was dried down and reconstituted to a concentration of 0.25 μg/μL in 0.1% formic acid generating the QCpool sample used for daily QC injections.

LC-MS/MS Analysis.

Peptides were analyzed using an Ultimate 3000 nano ultrahigh-performance liquid chromatography system (ThermoFisher Scientific, Waltham) coupled with an Orbitrap Fusion Lumos (ThermoFisher Scientific, Waltham) mass spectrometer in positive nanospray ionization mode. Injections were loaded onto an Acclaim PepMap C18, 3 μm packed trap column (75 μm i.d. × 2 cm) (ThermoFisher Scientific, Waltham) at 2 μL/min using loading solvent (water with 2% acetonitrile and 0.1% formic acid), and peptides separated on an in-house XBridge BEH C18, 2.5 μm packed analytical column (100 μm i.d. × 28 cm) (Waters Corporation, Milford) at 0.3 μL/min using solvent A (water with 0.1% formic acid) and solvent B (acetonitrile with 0.1% formic acid). QCpool and samples were separated with the following LC gradient: 0–14 min, 2% B; 14–17 min, 2–7% B; 17–100 min, 7–16% B; 100–155 min, 16–25% B; 155–160 min, 25–85% B; 160–168 min, 85% B; 168–170 min, 85–4% B; and 170–180 min, 4% B. Peptides were ionized with a positive voltage of 2.5 kV and ion transfer tube temperature of 300 °C. Full MS scans were acquired using Orbitrap isolation over m/z = 400–1600, at a resolution of 120,000 with microscans set to 1. The automatic gain control (AGC) target was set to 4.0 × 105 and the maximum injection time was set to 100 ms. Monoisotopic peak determination was set to Peptide including charge states from 2 to 6. Dynamic exclusion was enabled after a single charge state per precursor for a duration of 20 s and the intensity threshold set to 5.0 × 104 ions. Data dependent acquisition was used to acquire MS/MS spectra with the top 15 scans using a quadrupole isolation window of 0.7 m/z and Orbitrap detection with a resolution of 50,000. Peptide fragmentation was performed using high-energy C-trap dissociation with a normalized collision energy of 35% and normalized AGC target set to 250%.

LC-MS/MS Monitoring.

Approximately 1100 analytical plasma sample LC-MS/MS injections were performed over the course of 10 months conducted by two different operators (Op), hereafter referred to as OpA and OpB. Longitudinal monitoring of the LC-MS/MS system was performed by first queuing a daily blank (0.1% formic acid) at a flow rate of 0.3 μL/min with the following 1 h LC gradient: 0–10 min, 10% B; 10–41 min; 10–30% B, 41–45 min; 30–60% B, 45–46 min; 60–90% B, 46–49 min; 90% B, 49–50 min; 90–10% B, and 50–60 min; 10% B. The initial blank injection was followed by a QCpool (3 h gradient) and blank injection sequence prior to performing six sample injections (three fractions analyzed in duplicate each with a 3 h gradient) (Figure 1). Additional QC injections were performed after each maintenance event including: analytical and trap column change, LC solvent refresh, MS calibration, and MS front-end cleaning.

Figure 1.

Figure 1.

Semiautomated plasma proteomics workflow highlighting incorporation of quality controls (QCs). A cohort of plasma samples (N = 335) was depleted offline and the unbound fraction was concentrated, quantified, digested, and desalted using an automated robotic liquid handler (Biomek i7). Equimolar aliquots of each sample were pooled together creating an Spool sample used to generate both a within batch QC sample (QCbatch) and daily instrument performance QC sample (QCpool). QCpool samples were generated from TMTzero labeled Spool aliquots. Patient samples (N = 15) and Spool aliquot (N = 1) were tagged with TMTpro–16plex reagents in a randomized fashion. This resulted in 23 batches total. The QCbatch channel was used for within batch and across batch normalization. TMTpro batches were fractionated into 24 fractions by high-pH reversed-phase fractionation and subjected to liquid chromatography tandem mass spectrometry (LC-MS/MS) analysis. The total LC-MS/MS acquisition time was 10 months and was conducted by two different operators (OpA and OpB). QCpool samples were not subjected to high-pH reversed-phase fractionation and injected daily onto the LC-MS/MS system. The daily injection order was (1) blank (LC-MS/MS grade water with 0.1% FA), (2) QCpool, (3) blank, (4) fraction i in duplicate, (5) fraction j in duplicate, and (6) fraction k in duplicate, where i, j, and k = 1–23, randomized. The injection order was repeated daily through competition of all 24 fractions in duplicate per batch. A total of 271 QC search result files (OpA = 112, OpB = 159) were used to evaluate overall LC-MS/MS performance. Created with BioRender.com.

Database Searching.

The resulting QCpool raw files (N = 278) were analyzed using Proteome Discoverer software (version 2.5) and searched against the Uniprot human reviewed protein database (04/02/2021, 79740 sequences). The database was searched using the following Sequest HT parameters: a maximum of two missed cleavage sites, peptide length range of 6–144; precursor mass tolerance 10 ppm; fragment mass tolerance 0.02 Da; dynamic modifications of methionine oxidation (15.995 Da), lysine TMT (224.152 Da), and N-terminal TMT (224.152 Da). Peptides and proteins were filtered using the following parameters: peptide confidence at least high with a minimum peptide length of 6; protein filtering using a minimum number of one peptide sequence counting only rank one and top scored proteins. Proteins were grouped by applying the strict parsimony principle and ID distributions only used unique peptides based on protein groups. Protein and peptide abundances were determined using the precursor ions quantifier node.

Data Processing.

Seven QCpool raw files acquired from OpB were filtered out due to either an aborted LC-MS/MS method, incorrect LC method, or injection failure resulting in a total of 271 QCpool search result files (112 and 159 from OpA and B, respectively). We selected eight result file derived metrics (Table 1) to evaluate instrument performance across both OpA and B: protein count, peptide count, PSMs, number of MS/MS spectra, protein and peptide abundance, %m/z, and %ITmax. “Identification-centric” metrics included, identified protein count, identified peptide count, and the number of PSMs. Additionally, the number of collected MS/MS spectra and the average identified protein and peptide abundances were collected per injection across both operators. Peptide abundances generated by the Proteome Discoverer software were calculated using the precursor ion quantifier node, which quantifies peptide abundance by precursor ion m/z intensity. Identified peptide abundances associated with identified proteins are used to quantify protein abundance. Peptide and protein abundance hereafter will refer to the average log10 transformed abundances of all identified and quantified peptides and proteins, respectively, per injection. The “instrument specific” metric, %m/z, was derived from the mass accuracy specifications (±3 ppm) of the Lumos Fusion, where the percentage of peptides identified with an m/z value within ±0.003 Da of the theoretical m/z value (Δm/z value) is recorded relative to the total number of identified peptides. The method specific, %ITmax, metric was derived from the percentage of peptides identified at the user defined maximum MS/MS injection time relative to the total number of peptides identified at the average MS/MS injection time.

Table 1.

LC-MS/MS Quality Control Metrics

QC metric description
protein count number of proteins identified
peptide count number of peptides identified
PSMs number of peptide spectral matches
MS/MS spectra number of acquired MS/MS spectra
peptide abundance (x¯) average of all summed peptide abundances
%m/z percentage of peptides with a Δm/z between ±0.003 Da
%ITmax percentage of MS/MS spectra obtained at the maximum user set injection time
Qx the x-th equidistant retention time quadranta describing peptide elution
t R retention time, minutes
a

The maximum and minimum observed retention times are used to estimate the peptide elution range. The range is divided into four equal sections denoted as retention time quadrants.

Statistical Analysis.

Analytical values for selected QC metrics tR, protein count, peptide count, PSMs, MS/MS spectra, peptide abundance (x¯), %m/z, and %ITmax were recorded longitudinally and represent the LC-MS/MS analysis performance over the 10 month period. The average, standard deviation, and coefficient of variation (%CV) were calculated per metric and used to assess experimental variability. The variability associated with this data set was evaluated by principal component analysis (PCA) and used to identify statistical outliers that required operator intervention. PCA was executed using the GraphPad Prism version 9.3.1 for Windows (San Diego) software.

RESULTS

An overview of the plasma proteomics pipeline used here (Figure 1) details aspects of the sample preparation for patient and QC samples (which are the focus of this analysis). The raw LC-MS/MS data of 271 QC injections that occurred daily over the course of a 10-month proteomics study of 335 patient samples were assessed. The duration of this experiment necessitated measures to ensure optimal data collection for each sample across 23 sample batches. To this end, a QCpool sample was created that contained equimolar amounts of the 335 patient samples that were tagged with TMTzero for daily monitoring. Analysis of the QCpool samples was conducted across a series of analytical metrics (i.e., peptide retention time, peptide spectral matches (PSMs), MS/MS spectra, peptide abundance (x¯), %m/z, and %ITmax) (Table 1).

Peptide Tracking.

The data were searched for identified peptides observed through completion of LC-MS/MS data acquisition hereafter referred to as “tracking peptides”. Three consecutive QCpool result files were filtered to generate a master list of tracking peptides based on the following criteria: the peptide must be TMT modified, identified at high confidence with an Xcorr value ≥2, possess no more than one missed cleavage, be unique to a single identified protein, and have a Δm/z between ±0.003 Da. This generated a list of 2,733 tracking peptides. This list of peptides was then grouped into four subsets based on their average retention time (tR) along the elution gradient (i.e., the time during active peptide elution, here observed between 30–170 min (Figure 2A). Each subset was defined by dividing the elution gradient into four equidistant quadrants (Q, henceforth referred to as Q1, Q2, Q3, and Q4) from which one tracking peptide each was selected for tR (min) monitoring (Figure 2B). Here, we use proteolytic peptides RYIETDPANR, IVLGQEQDSYGGK, TEVIPPLIENR, and SQDILLSVENTVIYR observed in our QCpool sample to characterize and evaluate the elution profile of the LC gradient.29

Figure 2.

Figure 2.

(A) An example extracted chromatogram of base peak ions from QCpool injection #8. The chromatogram is divided into quadrants Q1 (30–65 min), Q2 (65–100 min), Q3 (100–135 min), Q4 (135–170 min). (B) A scatter plot of the observed retention time (tR, min) for the quadrant Q1 peptide RYIETDPANR, Q2 peptide IVLGQEQDSYGGK, Q3 peptide TEVIPPLIENR, and Q4 peptide SQDILLSVENTVIYR. (C) Box and whisker plots of the standardized tR for each tracking peptide for OpA is shaded gray and OpB is white. The standardized tR was calculated by dividing (x¯−xi) by the corresponding tR standard deviation across operators per tracked peptide, where x¯ is the average tR across both operators and xi is the observed tR. (D) A bar graph of the percentage of data sets acquired from OpA (N = 112) and OpB (N = 159) where each individual tracking peptide was detected in the full MS scan. The total number of data sets each individual tracking peptide was detected for OpsA and B, respectively, were 111 and 158 (RYIETDPANR), 112 and 158 (IVLGQEQDSYGGK), 111 and 158 (TEVIPPLIENR), and 107 and 129 (SQDILLSVENTVIYR). Data collected under OpA and OpB are indicated by the gray and white bars, respectively.

Over the course of LC-MS/MS monitoring, OpA trained OpB on daily operation of data acquisition and LC-MS/MS maintenance. At QCpool injection #114 OpB assumed operation of the LC-MS/MS experiments. Across operators the observed tR (%CV) for each tracked peptide (Table 2) was less than 20%.

Table 2.

Coefficient of Variation (%CV) for Retention Time (tR) of Tracking Peptides

overalla operator Aa operator Ba
sequence t R N t R N t R N
RYIETDPANR 46.04 (17) 269b 42.20 (6) 112 48.78 (18) 157
IVLGQEQDSYGGK 90.46 (15) 270 86.01 (8) 113 93.67 (17) 157
TEVIPPLIENR 109.31 (11) 269 105.17 (2) 112 112.26 (14) 157
SQDILLSVENTVIYR 143.56 (2) 236 140.99 (1) 108 145.73 (1) 128
a

Each tR value represents the x¯, (%CV), and number of injections.

b

The number of injections varies as each tracking peptide was not identified across all injections. The total number of injections evaluated was 271.

Figure 2B shows a cluster of QCpool injections (#146–173) observed with peptide tR values right shifted into adjacent quadrants and include peptide RYIETDPANR (Q1), IVLG-QEQDSYGGK (Q2), and TEVIPPLIENR (Q3). Consequently, the Q4 peptide SQDILLSVENTVIYR was not observed, representing a loss of data due to retention times that shifted beyond the programmed acquisition window. Shifted retention times are likely due to inconsistent flow rates resulting from drifting flow sensors, unstable pressure transducers, and/or inconsistent solvent composition.

Standardized tR values (Figure 2C) were calculated for each operator by eq 1,

xstd=xi−x¯/Sx (1)

where xi is the raw value, x¯ the mean, and Sx the standard deviation of tR. Retention time values observed for OpA generally had a tighter data distribution than OpB (i.e., %CV ranging from 1–8% and 1–18% for OpA and OpB, respectively) (Table 2). Several procedural steps including solvent and analytical column preparation may account in part to the variation observed between OpA and OpB.

To ensure the four tracked peptides were reasonable to use, we also determined how often each tracking peptide was detected across all QC injections (N = 271). Tracking peptides were observed 98–100% across all injections in Q1–Q3; however, this was dramatically reduced to 80% in Q4 (Figure 2D). We believe this was likely due to right shifted retention times in injections #146–173, impacting the most hydrophobic peptides.

QC Monitoring Analysis.

Identification rates in terms of protein/peptide counts and PSMs are often used as primary indicators of proteomics data quality;18 however, we suspected these metrics alone do not fully represent contributors to experimental variability. These metrics are also not sophisticated enough to help troubleshoot specifics of LC-MS/MS instrumentation and encourage timely QC intervention. Therefore, prior to implementation of QC metrics we longitudinally tracked protein count, peptide count, PSMs, number of MS/MS spectra, average protein and peptide abundances, %m/z, and % ITmax across all QC injections. Our goal was to determine the platform baseline and establish normal operating limits for future daily QC monitoring (Figure 3). Normal operating limits were established for each metric based on the standard error distribution of the mean value for a given metric. Data points that fell outside of 2σ signaled operator intervention was necessary before proceeding further with sample injections.

Figure 3.

Figure 3.

Scatter plots of the selected metrics across QCpool injections (N = 271). These metrics included (A) identified protein count, (B) identified peptide count, (C) protein abundance (x¯), (D) peptide abundance (x¯), (E) peptide spectral match (PSMs) count, (F) the percentage of identified peptides with a Δm/z between ±0.003 Da (%m/z), (G) MS/MS spectra count, and (H) the percentage of peptides identified at the maximum MS/MS injection time (%ITmax). Operator B began acquisition at QC injection #114 indicated by a gray vertical line. The average value across both operators for each metric, ±1σ, and ±2σ are indicated by the solid blue, dashed black, and dashed red lines, respectively. For the metric %m/z, the maximum value limit indicated by the solid red line was less than +1σ and +2σ.

The average (i.e., x¯±σ) observed protein (Figure 3A) and peptide counts (Figure 3C) and PSMs (Figure 3E) across operators were 366 ± 33, 4340 ± 434, and 7760 ± 1862, respectively. Protein count dropped at the point of transition from OpA to OpB (i.e., QC injection #114). Interestingly, no change in average peptide count was observed at the transition point. PSMs and the number of MS/MS spectra exhibited an increase in value starting at QC injection #116 coinciding with a change in the user defined maximum injection time from 80 to 86 ms. We suspect the increase in maximum MS/MS injection time allowed more time for the AGC target to be reached resulting in a high number of obtained MS/MS spectra and potential PSMs. A cluster of QC injections (#146–173) dropped in peptide count coinciding with right shifted tR values (Figure 2D) under OpB. In general, the %CV associated with OpB is higher when compared to OpA for all QC metrics (Table 3).

Table 3.

Coefficient of Variation (%CV) for Selected Quality Control Metrics

QC metric overalla operator Aa operator Ba
number of injections 271 112 159
identified proteins 366 (9) 388 (5) 349 (9)
identified peptides 4340 (10) 4550 (6) 4190 (12)
protein abundance, x¯ (log10) 7 (2) 7 (2) 7 (3)
peptide abundance, x¯ (log10) 7 (3) 7 (3) 7 (3)
PSMs 7760 (24) 5990 (10) 9030 (14)
MS/MS spectra 53645 (8) 51188 (6) 55402 (7)
%m/z 99 (2) 99 (1) 99 (2)
%ITmax 84 (9) 86 (7) 82 (10)
a

Values are presented as x¯, mean (%CV).

Quantification accuracy of precursor m/z is correlated with signal-to-noise levels.30 Several factors can increase experimental noise such as chemical contaminants, electrical interference, dirty ion optics, and inadequate gas pressures or purities, causing inconsistencies in observed signal between QC injections. Insufficient precursor signal to meet user defined intensity thresholds may also result in a decrease in triggered MS/MS events. We monitored the average protein and peptide abundances (Figures 3B and 3D, respectively) associated with the QCpool in each sample batch and the number of MS/MS spectra triggered per QC injection. Across operators, the average protein and peptide abundance, and MS/MS spectra were 7.37 ± 0.15, 7.03 ± 0.21, and 53645 ± 4292, respectively. Due to the large number of proteins and peptides used to determine average abundance values, it is also helpful to examine the range of abundances for a given QC injection (Supplemental Figure S1). Overall, the %CV of the protein and peptide average abundances and MS/MS spectra across operators was observed at 2, 3, and 8%, respectively, indicating high reproducibility across both operators.

The metrics %m/z (Figure 3F) and %ITmax (Figure 3H) have maxima at 100% indicating a majority of the total number of identified peptides per injection had a Δm/z ± 0.003 Da and a majority of the total number of MS/MS spectra were collected at the user defined maximum injection time, respectively. Across operators the average %m/z and %ITmax were 99 ± 2 and 84 ± 9, respectively.

Mass accuracy is subject to significant shifts for a given MS system and should be evaluated to minimize negative influences including temperature and humidity fluctuations, particularly for ion traps where mass accuracy is more sensitive to the ion population.31 In addition, laser desorption and electrospray ionization Orbitrap instruments may suffer performance and mass accuracy loss due to power cycling events.32 The overall population of modified peptide ions present in the QCpool ranging in abundance from 105–109 ions was used to evaluate mass accuracy shifts (%m/z, Figure 3F). A decline in %m/z was observed from QC injections #219–227. Prior to these QC injections, the mass spectrometer was shut down and serviced (i.e., turbo pump oil change, ion optics and quadrupole cleaning, system bake-out, and calibration), such that the instrument should have easily met the normal operating limits for this metric. That the instrument did not, suggests the electronics may have not properly reinitialized when the instrument was rebooted.

Drops in ion current from poor spray stability33 and problems with sample load34 may be detected through monitoring the observed injection times associated with the collected MS/MS spectra. The effective ion injection time was monitored by tracking %ITmax (Figure 3H). The average %ITmax across operators was 84 ± 8. If most QC injections had a %ITmax at 100%, this would signify the MS/MS scan settings are too short or that inadequate m/z signal is being transmitted causing the injection time threshold to be triggered before the AGC target is met. The overall %CV for %ITmax across operators was 9% indicating high reproducibility, despite the high variability from one QC injection sample to the next. Interestingly, QC injections #265–270 belong to a group of QC injections (i.e., QC #258–271) where the QCpool sample had been stored at 4 °C for several weeks before use. The decrease in %ITmax for these QC injections may be a result sample degradation or the adsorption of more hydrophobic peptides to the vial surface during the storage time,35–38 resulting in less signal.

Multivariate Analysis of Tracked QC Metrics.

When a single metric deviates outside of normal operating limits it indicates an underlying instrument issue needs to be addressed. However, LC-MS/MS systems consist of multiple interconnected experimental stages (i.e., sample loading, separation, spray injection, ion transmission, ion trapping, fragmentation, and detection) that are difficult to analytically monitor when only a single metric is considered. Tracking multiple metrics consistent with the various stages of LC-MS/MS experiments allows for a more robust view of instrument performance. Furthermore, assessing experimental variability highlights the usefulness of selected QC metrics and distinguishes QC injections outside of normal operating limits.

PCA was performed using standardized data from seven QC metrics (i.e., protein count, peptide count, peptide abundance (x¯), PSMs, MS/MS spectra, %m/z, and %ITmax) collected across 271 injections to evaluate the variability of the collective QC data set (Table 4). The experimental variation observed across QCpool replicates in this data set is spread across selected QC metrics. Therefore, multimetric QC tracking is essential to capturing experimental variability that would otherwise be ignored when performing QC monitoring.

Table 4.

Variance Contributions Determined by Principal Component Analysisa (PCA) of Selected QC Metrics

principal componentsb PC1 PC2 PC3 PC4 PC5 PC6 PC7
proportion of variance (%)c 36.27 26.86 17.05 11.40 3.58 2.71 2.13
a

Principal component analysis was performed using data from seven QC metrics including protein count, peptide count, PSMs, MS/MS spectra, % m/z, and %ITmax, collected over 271 QC injections.

b

All principal components (PC) were selected.

c

The contribution of variance explained by each PC sums to explain 100% of the variance in this data set.

PCA was performed for protein count paired with the remaining six metrics (Figure 4), where outliers fall outside of 95% confidence. Across each PCA plot OpB had significantly more QC outliers than OpA. This is consistent with OpB having larger %CV than OpA for each metric. Singular data clusters are observed within the confidence region for protein count with peptide abundance (x¯), %m/z, MS/MS spectra, and %ITmax (Figures 4B, 4D, 4E, 4F, respectively). However, peptide count (Figure 4A) and PSMs (Figure 4C) have two distinct data clusters. The smaller cluster of data points observed for peptide count (Figure 4A) coincides with QC injections performed after a mass spectrometer calibration. The two clusters observed for PSMs are separated based on injection times of 80 and 86 ms.

Figure 4.

Figure 4.

Principal component analysis (PCA) plots of selected quality control metrics. Each plot was generated by performing PCA analysis using the variable pair, protein count and (A) peptide count, (B) peptide abundance (x¯), (C) peptide spectral matches (PSMs), (D) the percentage of identified peptides with a Δm/z between ±0.003 Da (%m/z), (E) the number of MS/MS spectra obtained, and (F) the percentage of peptides identified at the maximum MS/MS injection time (%ITmax). QC data entries from OpA and OpB are colored blue and orange, respectively. Confidence ellipses (95%) are represented by the black dashed line.

One outlier under OpA (i.e., QC #31) and three outliers under OpB (i.e., QC #212, 213, and 262) were observed for every PCA plot. These QC injections are also measures that fell outside of the normal operating limits for protein count alone (Figure 3A). This demonstrates the strong ability for protein count to identify metrics outside of normal operating limits. Additional outliers are observed for peptide count, average peptide abundance (x¯), %m/z, MS/MS spectra, and %ITmax (Figures 4A, 4B, and 4D–F, respectively). This reiterates the importance of including multiple metrics when evaluating LC-MS/MS performance over time. No additional outliers (except for QC# 31, 212, 213, and 262) were observed for PSMs. Overall, the PCA analyses highlight the highly reproducible nature and quality of the project data set.

DISCUSSION

The purpose of this study was to use a real project data set to help establish the baseline of our plasma proteomics workflow. This was deemed necessary to ensure optimal LC-MS/MS data were collected over the 10-month project acquisition window. However, it was clear that QC metrics had to be established including normal operating limits for those metrics to inform data quality and allow for more real-time troubleshooting and instrument optimization. Below, we briefly describe the utility of each metric, how to globally incorporate these QC metrics in a multimetric QC evaluation scheme, and provide recommendations for implementing these metrics in a daily LC-MS/MS QC workflow.

QC Metrics.

Monitoring LC performance is achieved through use of selected tracking peptides. Solvent and column composition and flow rate can alter the elution of selected peptides, generating substantial shifts that may cause both strongly hydrophilic and hydrophobic peptides to be unobserved. Additionally, selected peptides that elute in the middle of the elution profile may be strongly left or right shifted, respectively. However, the missingness of tR data may not always be directly related to LC issues nor impact each tracked peptide. It is important that the peptides selected for tracking are observed throughout the duration of the proteomics study. Therefore, in addition to applying the filtering criteria discussed previously to generate the potential tracked peptides list (N = 2,733), we selected peptides that (1) were observed in three consecutive QCpool injections, (2) were derived from traditional tryptic cleavage sites, and (3) generated intense m/z signal (≥1.0 × 106) in order to minimize missing tR measures.39 This resulted in 2,463 peptides representing a robust population of tracking peptides from which we sorted peptides into four 35 min retention quadrants based on their reported tR. We randomly selected one peptide per quadrant resulting in four selected QC tracking peptides, each originating from a unique protein. However, a single missing tR value should require that another QC injection be performed and analyzed before resuming sample analysis.

Protein and peptide counts that deviate in a QC injection from normal operating limits originate from sample preparation, sample load, or LC and MS instrument issues.17 Specifically, the baseline observed for peptide count can be affected by insufficient extraction or digestion, poor sample injection, and instrument instability.40 Additionally, the success of peptide identification informs the number of observed proteins; therefore these metrics generally correlate with each other. Nevertheless, isolation of any one potential source of error is not possible when evaluating protein and peptide count singularly. However, protein and peptide count do represent the most relevant metrics for proteomics researchers prior to quantitative and biological interpretation of the data.

The PSMs metric is interconnected to protein/peptide count and MS/MS spectra. The number of MS/MS spectra obtained from LC-MS/MS analysis can impact the observed number of PSMs. Expectedly, QC injections with PSMs that fall outside of normal operating limits correlated with protein and peptide counts in the same injection. However, the number of MS/MS spectra collected is less sensitive to potential changes in sample load/integrity compared to protein and peptide counts and PSMs. Protein and peptide counts and PSMs are generally redundant QC metrics. However, a substantial decrease in the number of MS/MS spectra collected pinpoints instrument performance concerns.

Peptide abundance (x¯) as a metric informs of precursor ion intensity. During this project, cleaning the ion source components and the quadrupole rods and/or performing a mass spectrometer calibration returned this metric to within normal operating limits. Protein abundance if used as a metric would likely be redundant with peptide abundance (x¯).

We observed >99% of peptides across all QC injections have a %m/z metric that falls within normal operating limits. Performing a mass calibration was generally sufficient intervention to return the %m/z metric within normal operating limits. However, following a prolonged power down of the mass spectrometer (i.e., scheduled source cleanings or primary maintenance) our %m/z metric was outside of normal range. A second instrument reboot, bake-out, system equilibration, and calibration restored this metric. Real-time monitoring of this metric could be implemented using a lock mass.31,32

Unstable ion signal and poor ion detection can lead to long ion trap injection times and a decreased number of MS/MS spectra.41 In part, the baseline for %ITmax is influenced by the spray stability and AGC settings.42 Therefore, when establishing the baseline of this metric it is essential to ensure the entire LC-MS/MS system is stable.

It is important to note that some sources of experimental variability during LC-MS/MS data acquisition are not directly gleaned from injection of QCpool sample including but not limited to analytical column length, daily analytical and trap column pressures before and after sample injections, collision gas pressures, spray stability prior to QC and sample injections, and the pH of autosampler wash vials. Additionally, daily QC monitoring of LC-MS/MS performance should be in addition to prior QC efforts during sample preparation tracking depletion, digestion, and labeling efficiency. These points represent opportunities for researchers to design a robust list of QC metrics spanning the entire proteomics workflow. We have also collected data from QC monitoring of sample preparation steps for the patient samples (N = 335) represented in this work (data not shown). We believe the record and results from every QC test performed throughout the proteomics pipeline adds to the integrity of our cohort study analysis.

Implementation of QC Metrics.

Establishing the characteristic baseline of each QC metric must be completed prior to the start of any LC-MS/MS proteomics study. Therefore, a potentially overlooked consideration is the integrity of the QC sample intended for daily monitoring. We briefly described in the Materials and Methods section the different QC samples generated and their intended use. One of the utilities of using a patient-derived pooled sample for QC monitoring is that the sample closely represents the complexity of actual study samples and can help minimize missingness of QC channels for normalization. However, an obvious limitation when planning QC efforts for very large cohorts and ongoing studies whereby all patient samples would require preparation steps if they are combined after depletion, digestion, or tagging steps. For cases, where there is no limited automation to help with sample preparation, the QCpool would not be available until those steps are complete. Additionally, depending on the quantity of sample needed the generation of a QCpool may not be possible. We however recommend the use of a QCpool sample when possible and encourage the use of other suitable QC samples.

Additionally, efforts to minimize operator induced variability (i.e., stringent column and solvent preparation protocols, stockpiling QC sample to last the duration of the LC-MS/MS analysis, and scheduling routine LC and MS calibrations) must be in place before starting sample analysis. Note sample cleanliness may alter the frequency of needed calibrations and source/quadrupole cleanings for example on an Orbitrap Fusion Lumos mass spectrometer. We recommend determining the maximum amount of sample mass that can be injected before instrument cleaning is necessary (e.g., approximately 1 mg for the system and sample complexity that generated this data set).

Normal operating limits for each QC metric should be determined by analyzing three or more test QC injections on a recently cleaned and calibrated system. Experimental variability is determined for each metric in terms of %CV by longitudinally reporting QCpool LC-MS/MS results. We demonstrated a %CV < 10% for each metric is achievable through establishment of QC protocols and adequate operator training. Therefore, each QC metric must be within normal operating limits in order to move forward with data acquisition of study samples.

Quality Control Workflow.

A general LC-MS/MS QC workflow optimized for a 24 h data acquisition period (Figure 5) is provided. Generous laboratory notebook documentation detailing instrument operators, injected QCpool aliquot identity, instrument cleanings and issues, solvent replenishments, and column changes is critical. Prior to sample analysis a blank (0.1% formic acid), QCpool, and blank injection sequence is executed. Database search of the QCpool raw file is performed and each selected QC metric is assessed to determine if the metric passes or fails. An example QC monitoring workflow used to guide systematic identification of common sources of error when tR or protein and peptide counts outliers are observed is also provided (Supplemental Figure S2). This workflow can help users with quickly troubleshooting LC-MS/MS instruments and restoring optimal instrument operation. All instances of intervention are followed by an additional QCpool and blank injection test to verify the normal limits of each metric are restored. Note that performance thresholds and protocols including pressure fluctuations and recommended LC maintenance must be determined for each LC-MS/MS system in use. We recommend taking detailed notes on all actions performed by operators/technicians and the outcomes, providing essential information toward the creation and implementation of QC monitoring workflows. N number of sample injections and replicates are queued only when all QC metrics are observed within normal operating limits, followed by the next blank, QCpool, blank injection sequence. Ideally, this workflow would be best implemented with a friendly user GPU interface, automated in real-time for faster feedback, such as the recently developed PaSER 2022 systems (Bruker Daltonik GmbH, Bremen, Germany).

Figure 5.

Figure 5.

Quality control workflow. An illustration of a 24 h LC-MS/MS workflow optimized for the following quality control sequence performed prior to queuing any sample injection: (1) blank, (2) QCpool, (3) blank. The QCpool raw file is then database searched and the liquid chromatography and mass spectrometry metrics extracted and applied to the QC assessment. N number of sample injections including additional number of replicates are queued only after a successful navigation through the pass/fail QC assessment workflow. Created with BioRender.com.

CONCLUSIONS

Managing the experimental variability of high-throughput plasma proteomics LC-MS/MS analysis can be executed using simple QC metrics and the daily use of a standard QC sample. Though the selected metrics evaluated only represent a fraction of possible QC metrics available, we have demonstrated the capabilities of each metric to promote reproducibility and inform of irregular instrument performance that requires operator intervention. This analysis was done in the context of conducting large-scale bottom-up proteomics analyses of patient plasma samples that require months of instrument time. QC monitoring should be adjusted across workflows and further advanced to real-time and automated schemes. Application of consistent, well-planned, and well-documented QC measures ultimately increases the value and confidence of large-scale LC-MS/MS proteomics data sets.

Supplementary Material

Supplemental Information

Supporting Information

The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/jasms.3c00050.

Supplemental Figure S1: Average protein and peptide abundances across quality control (QC) injections; Supplemental Figure S2: A recommended quality control (QC) monitoring workflow; Supplemental Table S1: Tracking peptide retention times across quality control (QC) injections; Supplemental Table S2: Analytical metrics across quality control (QC) injections (PDF)

ACKNOWLEDGMENTS

This work was supported using resources from Vanderbilt University Institutional Start-up funds, National Institutes of Health (R01AG064950, RASR), Vanderbilt University Trans-institutional Program Vanderbilt Research Award to the Vanderbilt Memory and Alzheimer’s Center, Alzheimer’s Association (IIRG-08-88733, Jefferson), and the National Institute of Aging Grants: (R01-AG034962, Jefferson), (R01-AG0563534, Jefferson), and (R01-AG062826, Gifford). Study samples were obtained from the Vanderbilt Memory and Aging Project (VMAP). Participant data were collected by Vanderbilt Memory and Alzheimer’s Center Investigators at Vanderbilt University Medical Center.

Footnotes

The authors declare no competing financial interest.

Complete contact information is available at: https://pubs.acs.org/10.1021/jasms.3c00050

Contributor Information

Khiry L. Patterson, Department of Chemistry, Vanderbilt University, Nashville, Tennessee 37235, United States

Albert B. Arul, Department of Chemistry, Vanderbilt University, Nashville, Tennessee 37235, United States

Min Ji Choi, Department of Chemistry, Vanderbilt University, Nashville, Tennessee 37235, United States.

Nekesa C. Oliver, Department of Chemistry, Vanderbilt University, Nashville, Tennessee 37235, United States

Marsalas D. Whitaker, Department of Chemistry, Vanderbilt University, Nashville, Tennessee 37235, United States

Angel C. Bodrick, Department of Chemistry, Vanderbilt University, Nashville, Tennessee 37235, United States; Department of Biochemistry, Cancer Biology, Neuroscience, and Pharmacology, Meharry Medical College, Nashville, Tennessee 37208, United States

Julia B. Libby, Department of Neurology, Vanderbilt University Medical Center, Nashville, Tennessee 37232, United States

Shania Hansen, Vanderbilt Memory and Alzheimer’s Center, Vanderbilt University Medical Center, Nashville, Tennessee 37212, United States.

Logan Dumitrescu, Department of Neurology, Vanderbilt University Medical Center, Nashville, Tennessee 37232, United States.

Katherine A. Gifford, Department of Neurology, Vanderbilt University Medical Center, Nashville, Tennessee 37232, United States; Vanderbilt Memory and Alzheimer’s Center, Vanderbilt University Medical Center, Nashville, Tennessee 37212, United States

Angela L. Jefferson, Department of Neurology, Vanderbilt University Medical Center, Nashville, Tennessee 37232, United States; Vanderbilt Brain Institute, Vanderbilt University Medical Center, Nashville, Tennessee 37232, United States

Timothy J. Hohman, Vanderbilt Memory and Alzheimer’s Center, Vanderbilt University Medical Center, Nashville, Tennessee 37212, United States; Vanderbilt Brain Institute, Vanderbilt University Medical Center, Nashville, Tennessee 37232, United States

Renã A. S. Robinson, Department of Chemistry, Vanderbilt University, Nashville, Tennessee 37235, United States; Vanderbilt Institute of Chemical Biology, Vanderbilt University, Nashville, Tennessee 37232, United States; Vanderbilt Brain Institute, Vanderbilt University Medical Center, Nashville, Tennessee 37232, United States;

Data Availability Statement

The data underlying this study are available in the published article and the Supporting Information.

REFERENCES

  • (1).Enroth S; Berggrund M; Lycke M; Broberg J; Lundberg M; Assarsson E; Olovsson M; Stalberg K; Sundfeldt K; Gyllensten U High throughput proteomics identifies a high-accuracy 11 plasma protein biomarker signature for ovarian cancer. Commun. Biol 2019, 2, 221. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (2).Dubois C; Payen D; Simon S; Junot C; Fenaille F; Morel N; Becher F Top-Down and Bottom-Up Proteomics of Circulating S100A8/S100A9 in Plasma of Septic Shock Patients. J. Proteome Res 2020, 19 (2), 914–925. [DOI] [PubMed] [Google Scholar]
  • (3).Sharma NK; Ferreira BL; Tashima AK; Brunialti MKC; Torquato RJS; Bafi A; Assuncao M; Azevedo LCP; Salomao R Lipid metabolism impairment in patients with sepsis secondary to hospital acquired pneumonia, a proteomic analysis. Clin. Proteomics 2019, 16, 29. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (4).Solimani F; Didona D; Li J; Bao L; Patel PM; Gasparini G; Kridin K; Cozzani E; Hertl M; Amber KT Characterizing the proteome of bullous pemphigoid blister fluid utilizing tandem mass tag labeling coupled with LC-MS/MS. Arch Dermatol Res. 2022, 314 (9), 921–928. [DOI] [PubMed] [Google Scholar]
  • (5).Sperk M; van Domselaar R; Rodriguez JE; Mikaeloff F; Sa Vinhas B; Saccon E; Sonnerborg A; Singh K; Gupta S; Vegvari A; Neogi U Utility of Proteomics in Emerging and Re-Emerging Infectious Diseases Caused by RNA Viruses. J. Proteome Res 2020, 19 (11), 4259–4274. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (6).Chen M; Xia W Proteomic Profiling of Plasma and Brain Tissue from Alzheimer’s Disease Patients Reveals Candidate Network of Plasma Biomarkers. J. Alzheimers Dis 2020, 76 (1), 349–368. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (7).Khan MJ; Chung NA; Hansen S; Dumitrescu L; Hohman TJ; Kamboh MI; Lopez OL; Robinson RAS Targeted Lipidomics To Measure Phospholipids and Sphingomyelins in Plasma: A Pilot Study To Understand the Impact of Race/Ethnicity in Alzheimer’s Disease. Anal. Chem 2022, 94 (10), 4165–4174. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (8).Khan MJ; Desaire H; Lopez OL; Ilyas Kamboh M; Robinson RAS Dataset of why inclusion matters for Alzheimer’s disease biomarker discovery in plasma. Data Brief 2021, 35, 106923. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (9).Muenchhoff J; Poljak A; Thalamuthu A; Gupta VB; Chatterjee P; Raftery M; Masters CL; Morris JC; Bateman RJ; Fagan AM; Martins RN; Sachdev PS Changes in the plasma proteome at asymptomatic and symptomatic stages of autosomal dominant Alzheimer’s disease. Sci. Rep 2016, 6, 29078. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (10).Aebersold R; Mann M Mass-spectrometric exploration of proteome structure and function. Nature 2016, 537 (7620), 347–55. [DOI] [PubMed] [Google Scholar]
  • (11).Arul AB; Robinson RAS Sample Multiplexing Strategies in Quantitative Proteomics. Anal. Chem 2019, 91 (1), 178–189. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (12).Cravatt BF; Simon GM; Yates JR 3rd The biological impact of mass-spectrometry-based proteomics. Nature 2007, 450 (7172), 991–1000. [DOI] [PubMed] [Google Scholar]
  • (13).Domon B; Aebersold R Mass spectrometry and protein analysis. Science 2006, 312 (5771), 212–7. [DOI] [PubMed] [Google Scholar]
  • (14).Frost DC; Rust CJ; Robinson RAS; Li L Increased N,N-Dimethyl Leucine Isobaric Tag Multiplexing by a Combined Precursor Isotopic Labeling and Isobaric Tagging Approach. Anal. Chem 2018, 90 (18), 10664–10669. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (15).Wu Z; Shen Y; Zhang X TAG-TMTpro, a Hyperplexing Quantitative Approach for High-Throughput Proteomic Studies. Anal. Chem 2022, 94 (37), 12565–12569. [DOI] [PubMed] [Google Scholar]
  • (16).Yates JR 3rd Mass spectral analysis in proteomics. Annu. Rev. Biophys. Biomol. Struct 2004, 33, 297–316. [DOI] [PubMed] [Google Scholar]
  • (17).Walzer M; Vizcaino JA Review of Issues and Solutions to Data Analysis Reproducibility and Data Quality in Clinical Proteomics. Methods Mol. Biol 2020, 2051, 345–371. [DOI] [PubMed] [Google Scholar]
  • (18).Bittremieux W; Tabb DL; Impens F; Staes A; Timmerman E; Martens L; Laukens K Quality control in mass spectrometry-based proteomics. Mass Spectrom Rev. 2018, 37 (5), 697–711. [DOI] [PubMed] [Google Scholar]
  • (19).Tabb DL Quality assessment for clinical proteomics. Clin Biochem 2013, 46 (6), 411–20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (20).Bereman MS; Beri J; Sharma V; Nathe C; Eckels J; MacLean B; MacCoss MJ An Automated Pipeline to Monitor System Performance in Liquid Chromatography-Tandem Mass Spectrometry Proteomic Experiments. J. Proteome Res 2016, 15 (12), 4763–4769. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (21).Chiva C; Olivella R; Borras E; Espadas G; Pastor O; Sole A; Sabido E QCloud: A cloud-based quality control system for mass spectrometry-based proteomics laboratories. PLoS One 2018, 13 (1), No. e0189209. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (22).Kocher T; Pichler P; Swart R; Mechtler K Quality control in LC-MS/MS. Proteomics 2011, 11 (6), 1026–30. [DOI] [PubMed] [Google Scholar]
  • (23).Pichler P; Mazanek M; Dusberger F; Weilnbock L; Huber CG; Stingl C; Luider TM; Straube WL; Kocher T; Mechtler K SIMPATIQCO: a server-based software suite which facilitates monitoring the time course of LC-MS performance metrics on Orbitrap instruments. J. Proteome Res 2012, 11 (11), 5540–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (24).Beasley-Green A; Bunk D; Rudnick P; Kilpatrick L; Phinney K A proteomics performance standard to support measurement quality in proteomics. Proteomics 2012, 12 (7), 923–31. [DOI] [PubMed] [Google Scholar]
  • (25).Dogu E; Taheri SM; Olivella R; Marty F; Lienert I; Reiter L; Sabido E; Vitek O MSstatsQC 2.0: R/Bioconductor Package for Statistical Quality Control of Mass Spectrometry-Based Proteomics Experiments. J. Proteome Res 2019, 18 (2), 678–686. [DOI] [PubMed] [Google Scholar]
  • (26).Bittremieux W; Willems H; Kelchtermans P; Martens L; Laukens K; Valkenborg D iMonDB: Mass Spectrometry Quality Control through Instrument Monitoring. J. Proteome Res 2015, 14 (5), 2360–6. [DOI] [PubMed] [Google Scholar]
  • (27).Stanfill BA; Nakayasu ES; Bramer LM; Thompson AM; Ansong CK; Clauss TR; Gritsenko MA; Monroe ME; Moore RJ; Orton DJ; Piehowski PD; Schepmoes AA; Smith RD; Webb-Robertson BM; Metz TO Quality Control Analysis in Real-time (QC-ART): A Tool for Real-time Quality Control Assessment of Mass Spectrometry-based Proteomics Data. Mol. Cell Proteomics 2018, 17 (9), 1824–1836. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (28).Bielow C; Mastrobuoni G; Kempa S Proteomics Quality Control: Quality Control Software for MaxQuant Results. J. Proteome Res 2016, 15 (3), 777–87. [DOI] [PubMed] [Google Scholar]
  • (29).Burkhart JM; Premsler T; Sickmann A Quality control of nano-LC-MS systems using stable isotope-coded peptides. Proteomics 2011, 11 (6), 1049–57. [DOI] [PubMed] [Google Scholar]
  • (30).Bakalarski CE; Elias JE; Villen J; Haas W; Gerber SA; Everley PA; Gygi SP The impact of peptide abundance and dynamic range on stable-isotope-based quantitative proteomic analyses. J. Proteome Res 2008, 7 (11), 4756–65. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (31).Zhou M; Gucinski AC; Boyne MT 2nd Performance metrics for evaluating system suitability in liquid chromatography-Mass spectrometry peptide mass mapping of protein therapeutics and monoclonal antibodies. MAbs 2015, 7 (6), 1104–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (32).Perry RH; Cooks RG; Noll RJ Orbitrap mass spectrometry: instrumentation, ion motion and applications. Mass Spectrom Rev. 2008, 27 (6), 661–99. [DOI] [PubMed] [Google Scholar]
  • (33).Rudnick PA; Clauser KR; Kilpatrick LE; Tchekhovskoi DV; Neta P; Blonder N; Billheimer DD; Blackman RK; Bunk DM; Cardasis HL; Ham AJ; Jaffe JD; Kinsinger CR; Mesri M; Neubert TA; Schilling B; Tabb DL; Tegeler TJ; Vega-Montoto L; Variyath AM; Wang M; Wang P; Whiteaker JR; Zimmerman LJ; Carr SA; Fisher SJ; Gibson BW; Paulovich AG; Regnier FE; Rodriguez H; Spiegelman C; Tempst P; Liebler DC; Stein SE Performance metrics for liquid chromatography-tandem mass spectrometry systems in proteomics analyses. Mol. Cell Proteomics 2010, 9 (2), 225–41. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (34).Kalli A; Hess S Effect of mass spectrometric parameters on peptide and protein identification rates for shotgun proteomic experiments on an LTQ-orbitrap mass analyzer. Proteomics 2012, 12 (1), 21–31. [DOI] [PubMed] [Google Scholar]
  • (35).Kawashima Y; Takahashi N; Satoh M; Saito T; Kado S; Nomura F; Matsumoto H; Kodera Y Enhanced recovery of lyophilized peptides in shotgun proteomics by using an LC-ESI-MS compatible surfactant. Proteomics 2013, 13 (5), 751–5. [DOI] [PubMed] [Google Scholar]
  • (36).Kraut A; Marcellin M; Adrait A; Kuhn L; Louwagie M; Kieffer-Jaquinod S; Lebert D; Masselon CD; Dupuis A; Bruley C; Jaquinod M; Garin J; Gallagher-Gambarelli M Peptide storage: are you getting the best return on your investment? Defining optimal storage conditions for proteomics samples. J. Proteome Res 2009, 8 (7), 3778–85. [DOI] [PubMed] [Google Scholar]
  • (37).Magdeldin S; Moresco JJ; Yamamoto T; Yates JR 3rd Off-Line Multidimensional Liquid Chromatography and Auto Sampling Result in Sample Loss in LC/LC-MS/MS. J. Proteome Res 2014, 13 (8), 3826–36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (38).Stejskal K; Potesil D; Zdrahal Z Suppression of peptide sample losses in autosampler vials. J. Proteome Res 2013, 12 (6), 3057–62. [DOI] [PubMed] [Google Scholar]
  • (39).Tabb DL; Vega-Montoto L; Rudnick PA; Variyath AM; Ham AJ; Bunk DM; Kilpatrick LE; Billheimer DD; Blackman RK; Cardasis HL; Carr SA; Clauser KR; Jaffe JD; Kowalski KA; Neubert TA; Regnier FE; Schilling B; Tegeler TJ; Wang M; Wang P; Whiteaker JR; Zimmerman LJ; Fisher SJ; Gibson BW; Kinsinger CR; Mesri M; Rodriguez H; Stein SE; Tempst P; Paulovich AG; Liebler DC; Spiegelman C Repeatability and reproducibility in proteomic identifications by liquid chromatography-tandem mass spectrometry. J. Proteome Res 2010, 9 (2), 761–76. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (40).Piehowski PD; Petyuk VA; Orton DJ; Xie F; Moore RJ; Ramirez-Restrepo M; Engel A; Lieberman AP; Albin RL; Camp DG; Smith RD; Myers AJ Sources of technical variability in quantitative LC-MS proteomics: human brain tissue sample analysis. J. Proteome Res 2013, 12 (5), 2128–37. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (41).Kalli A; Smith GT; Sweredoski MJ; Hess S Evaluation and optimization of mass spectrometric settings during data-dependent acquisition mode: focus on LTQ-Orbitrap mass analyzers. J. Proteome Res 2013, 12 (7), 3071–86. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (42).Andrews GL; Dean RA; Hawkridge AM; Muddiman DC Improving proteome coverage on a LTQ-Orbitrap using design of experiments. J. Am. Soc. Mass Spectrom 2011, 22 (4), 773–83. [DOI] [PMC free article] [PubMed] [Google Scholar]