Introduction
On August 15, 2025, we will celebrate the 75th anniversary of the National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK) at the National Institutes of Health (NIH). This symposium is a collection of six manuscripts1-6 highlighting the development and clinical translation of the artificial pancreas (AP)—a landmark achievement in the treatment of diabetes, to a significant extent due to the critical support of NIDDK and the Institute’s Special Diabetes Program—a statutory funding program that began in 1998, focusing on basic, translational, and clinical research for the prevention and cure of type 1 diabetes (T1D). Over two decades, the AP grew from a disputed idea in 2005, to the standard of care for T1D today, improving the lives of those who struggle daily with this condition. The symposium is also part of a series of publications on a variety of relevant topics highlighting NIDDK-supported research contributions as the Institute celebrates its 75th anniversary. All manuscripts are written by those who had prominent contributions to the science, engineering, and clinical testing of the AP technology. Here is their story:
On a cold and cloudy December day in 2005, physicians, scientists, and engineers working on diabetes technologies gathered in the Lister Hill Auditorium at the National Institutes of Health in Bethesda, Maryland. This was the first workshop, “Obstacles and Opportunities on the Road to Artificial Pancreas: Closing the Loop,” organized by NIDDK and the Juvenile Diabetes Research Foundation (JDRF, now Breakthrough T1D) in collaboration with the Food and Drug Administration (FDA). In the previous 30 years, the treatment of T1D had been a technology-rich clinical endeavor with increasingly significant engineering components. In the 1970s, the first trials of continuous subcutaneous insulin delivery (CSII) using insulin pumps were reported,7,8 and by 2005, CSII has established its presence in the treatment of T1D. Research was also underway in the 1990s on designing automation algorithms for insulin delivery, anticipating the development of suitably accurate sensors and pumps, laying the groundwork for what was to become the core algorithm in modern commercial devices—model predictive control (MPC). 9 The progress of an AP suitable for outpatient was enabled by the introduction of continuous glucose monitoring (CGM).10,11 In 2004-2005, first attempts to automate glucose control using CGM and CSII were made.12,13 Thus, the focus of this first NIH/JDRF workshop was on the viability of an AP system using subcutaneous glucose sensing via CGM and CSII. The room was divided on this subject, with opinions ranging from “impossible” to “not too difficult”—and, as expected, the truth lied between these two extremes. Nevertheless, the workshop achieved its objective: a movement was initiated toward solving the challenges of automated insulin delivery (AID), and in 2006, a Request for Applications was issued initiating the JDRF Artificial Pancreas Project. The second FDA-NIH-JDRF workshop, “Towards an Artificial Pancreas,” was held in July 2008, followed by solicitations issued by the NIDDK to support research developing new AP technologies.
This collection of papers1-6 commemorates the academic and industrial research and development supported by NIDDK, Breakthrough T1D, and the Helmsley Charitable Trust, which drove the creation and validation for clinical use of several viable AP systems. A Supplement on Development & Future of AID Systems, published recently in Diabetes Technology and Therapeutics, detailed the private sector developments during this same period of time, highlighting the pioneering contributions of a number of companies to the AP field. 14
The Following 10 Years
A number of studies documented the benefits of CGM15-17 and outlined its future as a component of an AP system.18-21 The long way toward the contemporary interoperable user-friendly closed-loop AP began in 2007 with the introduction of a universal research platform—the APS—developed at the University of California, Santa Barbara (UCSB). 22 The APS enabled automated communication between a variety of CGM devices, insulin pumps, and control algorithms, and powered the first inpatient clinical trials.23,24 In April 2007, a major milestone was achieved with the APS enabling the first closed-loop human subject trial in the United States at UCSB. 25 In 2008, a computer simulator of the human metabolic system in T1D developed at the universities in Padova, Italy and Virginia, USA, and was accepted by the FDA as a substitute to animal trials in the testing of AID algorithms—this unprecedented event accelerated the AP development by years. 26 Outpatient studies took an AP system to a camp setting, where children with T1D were observed overnight, comparing AP to sensor-augmented pump therapy. 27 This same system was then taken to patients’ homes, achieving similar results. 28 However, these first inpatient and outpatient studies used a system that was based on laptop computers, still cumbersome and unsuitable for routine outpatient use.
In 2009 JAMA wrote “Artificial pancreas may soon be a reality,” 29 and by 2010, the Artificial Pancreas became a global research topic. In May 2012, a Diabetes Outlook published in Nature 30 highlighted the AP, and then Science featured the same topic. 31 These Nature 30 and Science 31 reviews featured a photo of a smart phone presenting a dual traffic-light display—the face of the first portable AP platform—the Diabetes Assistant developed at the University of Virginia in 2011. Diabetes Assistant was built using an Android smart phone as a computational hub, and included graphical user interface designed for the patient 32 and Web-based remote monitoring. 33 Diabetes Assistant was used in a number of clinical trials at ten centers in the United States and Europe.34-42
In 2014, NIDDK issued a program announcement titled “Advanced Clinical Trials to test Artificial Pancreas Device Systems in Type 1 Diabetes (UC4).” Increasing research activity was directed toward designing innovative control algorithms that could integrate CGMs and pumps, drawing on insights and expertise from the process dynamics and control field. 43 Studies supported by this initiative resulted in significant clinical translation—the roots of several contemporary AP systems can be traced to this NIDDK support. In July 2016, Diabetes Care published a symposium 44 dedicated to outpatient AP studies, 45 including trials at patient’s homes lasting a month or more37,38,45 and studies in young children. 39 The AP systems were pilot-tested, and were found effective, in pregnant women with T1D,46,47 and in people with type 2 diabetes on insulin therapy. 48 In 2016, a pivotal trial of the first commercial hybrid AID system was completed—the Medtronic 670G—which automatically modulates basal rates but not insulin boluses 49 —and the system was cleared by the FDA for clinical use. Between 2016 and 2018, the National Library of Medicine included more than 150 publications per year related to the engineering development and clinical testing of various closed-loop systems—a trend that confirmed the growing prominence of AP. Systematic reviews and meta-analyses were published by various journals, confirming the efficacy of closed-loop control for the insulin treatment of diabetes.50-53 The AP was coming of age.
State of the Art Today
A dream in 2005, closed-loop control of diabetes, also known as AID, or the AP, is now the gold-standard treatment for T1D, and is expanding to people with type 2 diabetes on insulin treatment. Several commercial systems are now available in the United States, including: Medtronic’s MiniMed 670G, 770G, and 780G, the latter adjusting automatically basal rate and administering correction boluses; t:slim X2 and Mobi with Control-IQ Technology by Tandem Diabetes Care, which automates both insulin basal rate and correction boluses; OmniPod 5 hybrid closed-loop system, Insulet Corporation; the iLet Bionic Pancreas system by Beta Bionics, CamAPS FX (CamDiab Ltd., Cambridge, UK)—a mobile app that is the first to be approved for use by pregnant women with T1D, and the twiist Automated Insulin Delivery (AID) system by Sequel. Other AP systems are available overseas, but are not cleared by the FDA in the United States, for example, Diabeloop DBLG1 (Diabeloop France) and Inreda AP—a bi-hormonal fully automated closed-loop system (Inreda Diabetic B.V., the Netherlands). The development and testing of most of these systems, for example, MiniMed 780G, Control-IQ, iLet, Cam APS FX, was supported by the NIDDK UC4 grants noted above, while NIDDK R01, R21, and DP3 grants supported developments at UCSB that led to the predictive algorithm powering the OmniPod 5. The major difference between AP configurations is the control algorithm, which directs insulin delivery based on CGM readings and other factors, such as insulin on board. For example, the MiniMed systems use proportional integral derivative algorithms, while most others use controllers based on the MPC algorithm, originally developed for the chemical process industry in the 1960s. 43 MiniMed 780G added a previously introduced fuzzy logic algorithm 27 responsible for the administration of correction boluses. Except for Inreda AP, which injects insulin and glucagon, all closed-loop systems inject insulin only, with the iLet potentially capable of injecting glucagon as well.
Multiple well respected peer reviewed medical journals embraced the AP topic. In the span of four years the New England Journal of Medicine, published seven clinical trials,54-60 indicating that AP technologies have firmly transitioned to the clinical practice of T1D. These papers featured several AP systems noted above: Control-IQ,54,55,57 the iLet, 58 Cam APX,59,60 and an open-source control algorithm. 56 The AP technologies made successful strides into insulin-using type 2 diabetes as well.61-63 Large volumes of papers reported outcomes of real-world AP use by thousands of people for extended periods of time,64-73 for example, 780G,64-67 Control-IQ,68-71 and OmniPod 5.72,73 Figure 1 presents the number of publications in PubMed per year resulting from a search on “artificial pancreas” or “closed loop control of diabetes.”
Figure 1.
Number of publications in PubMed per year resulting from a search on “artificial pancreas” or “closed loop control of diabetes.”
Clusters of publications addressed important subtopics of the general AP theme, such as use of AID in pregnancy complicated by T1D74-78; ultra-fast insulin analogues attempting to minimize postprandial glucose excursions with hybrid AP systems, which resulted in mixed success79-82; use of AID in various type of exercise83-86; and the efficacy of AP in people with history of severe hypoglycemia and hypoglycemia unawareness.87-89 International forums crafted consensus statements on the clinical use of closed-loop control.90-92
Overall, while detailed comparisons of results across studies are not statistically plausible due to differences in sensors used, experimental setup, and lack of randomization, it became evident that (1) all contemporary AP systems perform significantly better than both multiple daily injections or sensor-augmented insulin pump therapy, and (2) the first-generation AP systems achieve, on average, strikingly similar results, for example, time in the target range of 70 to 180 mg/dL (TIR) of ~70-75 percent and rate of hypoglycemia (below 70 mg/dL) in the order of 2% to 3%. The “ceiling effect” of TIR is an important observation pointing at the limitations of the current hybrid closed-loop devices: both the physiology of the subcutaneous route of insulin delivery and factors stemming from the behavior of AP users present challenges that need to (and will be) addressed in the near future.
The Next Frontier
While predictions are often inaccurate, in this section, we will attempt to forecast what is coming next to the AP field, based on current major research trends. In 2023, the Fifth Artificial Pancreas Workshop: Enabling Full Automation, Access, and Adoption was held at the NIH Campus in Bethesda, Maryland. 93 The Workshop outlined several major trends in the development of AP systems: (1) data management, integration, and cybersecurity; (2) adaptation of AID through the lifespan; (3) approaches to fully automated closed-loop control, and (4) novel artificial intelligence (AI) strategies to develop smarter, personalized, diabetes management systems. 93 Overall, the next frontier of AP development is abundantly clear: the quest is for a fully automated system that would not require input from its user to maintain optimal glucose control. This is in contrast to virtually all contemporary AID systems, which are “hybrid,” in that the user is expected to announce meals and initiate meal boluses or anticipate physical activity and attenuate insulin delivery in advance. While the most advanced hybrid AID systems have consistently shown improvements in overnight control, all studies to date point out that achieving optimal control during the day is still a challenge. This is a direct result of behavioral disturbances typically occurring during the day, for example, meals or exercise that contribute to unexpected glycemic fluctuations. Moreover, the current speed of insulin action, even with modern rapid-acting analogues, is still too slow to enable adequate AID response—it is simply too late to automatically inject insulin when a CGM indicates a postprandial glucose excursion or to discontinue insulin delivery when exercise is detected. This motivates the development of a fully automated closed-loop (FCL) system that does not require meal or exercise announcements. Interestingly, one of the first published closed-loop trials in the United States employed an FCL system, which showed promise, albeit using small (30 g carbohydrate) unannounced meals. 24 While the task is clear, the routes to achieving FCL are different and can be classified as follows: (1) ultra-fast insulin attempting to compensate for both the time difference between glucose absorption during meals and the onset of insulin action, and the attenuation of insulin action during exercise; (2) dual insulin plus glucagon control typically increasing insulin doses around meals and counteracting potential insulin-induced hypoglycemia with automated glucagon injection; (3) use of adjuvant hormones, for example, amylin, GLP-1 receptor agonists, SGLT-2 inhibitors, which are known to mitigate postprandial meal excursions, or additional signals, such as motion sensing or heart rate, reacting fast to physical activity; and (4) new control algorithms trying to anticipate the timing of the next meal or exercise based on historical data for each user. The latter is a route to adaptive personalized AP systems, which are certainly on the list of upcoming developments. Currently, some algorithms are “trainable” adapting control parameters based on the experience of their users.94,95 In one study using adaptation of the algorithm’s “aggressiveness” during periods of uncertainty, a remarkable 88% TIR was achieved, which remains one of the highest recorded in a closed-loop study. 95
We can, however, anticipate a brighter future for truly personal FCL using the vast amounts of data available now. Looking back at the beginning of the AP project 20 years ago, data were sparse. The CGM was in infancy and data sets including both glucose readings and insulin delivery were generally unavailable, with one notable exception: elaborate studies of the human metabolic system, typically involving tracers containing labeled glucose, have been done, allowing the creation of comprehensive mathematical models of insulin-glucose dynamics.96-98 Consequently, to date, most AP control algorithms are model-based using models of the metabolic system, simplified enough to be tractable in real time, to calculate control actions. 43 With the accumulation of large data sets and the increasing ability of contemporary CGM/CSII systems to track an individual across long periods of time, “model-free” data-driven controllers are emerging based on techniques that fall in the domain of data science, such as machine learning (ML) and AI. Both ML and AI can remedy certain shortcomings of the current AID systems, as elaborated by a recent review of data science methods available to diabetes research and clinical practice. 99 These methods range from detection of glycemic patterns and prediction of high-risk events, to the first AID system based entirely on ML and a deep neural network, which opened the field for a new class of purely data-driven control algorithms.100,101 By design, ML and AI are capable of real-time learning, adaptation, and problem-solving. While CGM data are still relatively sparse (one year of five-minute sampled CGM = approximately 100 000 points) compared with the volume of data required to train typical Large Language Models (hundreds of GBs to TBs), as more data are accumulated across AP users and time, we can expect that future AID developments will gradually transition from traditional equation (model)-driven control algorithms to smarter, personalized, trainable, fully automated ML, and generative AI implementations. This transition has already begun 101 and the movement toward the AI AP of the near future is likely inevitable.
About This Symposium
This symposium is a collection of six manuscripts highlighting the development and clinical translation of the AP idea,1-6 assembled in celebration of two decades of NIDDK and the Special Diabetes Program support to this technology. All manuscripts are written by scholars, engineers, and physicians, who had prominent contributions to the development and clinical translation of the AP. The collection includes a paper on the hardware components of the AP—CGM and insulin pumps, 1 as well as a paper on the metabolic models, in silico pre-clinical trials, and algorithms that integrated the components in what is now known as AP system. 2 The clinical testing and translation of the AP is reflected by two papers—one on early studies 3 and another on pivotal registration trials, which earned FDA clearance to the systems currently in use. 4 Finally, two papers discuss the challenges and opportunities to AP research 5 and the practical considerations for widespread AP use in the treatment of diabetes. 6 As pointed out in the beginning, the focus of this collection is more on academic than industry developments, and in particular highlights the contributions of NIDDK and organizations such as Breakthrough T1D to the creation of the AP.
Acknowledgments
As stated in the manuscript, the purpose of this symposium is to celebrate NIDDK’s 75th anniversary, to recognize the scientific input from NIDDK and its partners, and to acknowledge NIDDK, its partners and the Special Diabetes Program overseen by NIDDK, for their sustained long-term financial support toward the development and translational success of the AP project. The genesis for this symposium is credited to Drs. Guillermo Arreaza-Rubin of NIDDK. The authors thank Dr Thomas Eggerman and Neal Green (NIDDK) for the support in organizing this complex endeavor. This manuscript was not prepared under the auspices of the NIDDK and does not represent analyses or conclusions of the NIH or the NIDDK. FJD would like to acknowledge the stimulating and fruitful multi-decade collaborative endeavors with Eyal Dassau and Howard Zisser. BPK would like to acknowledge his long-standing collaborators Sue Brown and Marc Breton, as well as his outstanding colleagues at the UVA Center for Diabetes Technology.
Footnotes
Author Contributions: BPK proposed wrote the first draft of the manuscript. FJD revised the manuscript and added additional material.
The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: FJD reports funding over the time course described in this editorial from the Showalter Foundation, Roche Diagnostics Corporation, the Juvenile Diabetes Research Foundation, the Helmsley Charitable Trust, and the National Institutes of Health (NIDDK). FJD also reports royalty payments from patents in a portfolio that have been licensed to Insulet, Roche, and Dexcom. BPK reports research grants handled by the University of Virginia from the National Institutes of Health, Novo Nordisk, Dexcom, and Tandem Diabetes Care. In addition, BPK has a number of patents with royalties paid to Dexcom and Novo Nordisk.
Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.
ORCID iDs: Francis J. Doyle
https://orcid.org/0000-0002-3293-9114
Boris P. Kovatchev
https://orcid.org/0000-0003-0495-3901
References
- 1. Bergenstal RM, Heller A, Marc D. Breton, et al. Evolution of the Artificial Pancreas: Components and Integration - CGMs, Insulin, and AP Systems. J Diabetes Sci Technol. 2025; 19(4):883-894. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Cinar A, Basu A, Bequette W, et al. Metabolic Models, In Silico Trials, and Algorithms. J Diabetes Sci Technol. 2025; 19(4):895-907. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Zisser H, Hovorka R, Basu A, et al. Early stages of Automatic Insulin Delivery. J Diabetes Sci Technol. 2025; 19(4):908-923. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Akturk HK, Brown SA, Beck RW, et al. Registration and Real-Life Studies on Automated Insulin Delivery Systems. J Diabetes Sci Technol. 2025; 19(4):924-936. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Jacobs PG, Levy CJ, Brown SA, et al. Research Gaps, Challenges and Opportunities in Automated Insulin Delivery Systems. J Diabetes Sci Technol. 2025; 19(4):937-949. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Messer LH, Forlenza GP, Gonder-Frederick L, et al. Practical Considerations and Implementation of Automated Insulin Delivery Systems. J Diabetes Sci Technol. 2025; 19(4):950-957. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Pickup JC, Keen H, Parsons JA, Alberti KG. Continuous subcutaneous insulin infusion: an approach to achieving normoglycaemia. Br Med J. 1978;1:204-207. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Tamborlane WV, Sherwin RC, Genel M, Felig P. Reduction to normal of plasma glucose in juvenile diabetes by subcutaneous administration of insulin with a portable infusion pump. N Engl J Med. 1979;300:573-578. [DOI] [PubMed] [Google Scholar]
- 9. Parker RS, Doyle FJ, 3rd, Peppas NA. A model-based algorithm for blood glucose control in type I diabetic patients. IEEE Trans Biomed Eng. 1999;46(2):148-157. doi: 10.1109/10.740877. [DOI] [PubMed] [Google Scholar]
- 10. Mastrototaro JJ. The MiniMed continuous glucose monitoring system. Diabetes Technol Ther. 2000;2:S13-S18. [DOI] [PubMed] [Google Scholar]
- 11. Feldman B, Brazg R, Schwartz S, Weinstein R. A continuous glucose sensor based on wired enzyme technology—results from a 3-day trial in patients with type 1 diabetes. Diabetes Technol Ther. 2003;5(5):769-779. [DOI] [PubMed] [Google Scholar]
- 12. Hovorka R, Chassin LJ, Wilinska ME, et al. Closing the loop: the ADICOL experience. Diabetes Technol Ther. 2004;6(3):307-318. [DOI] [PubMed] [Google Scholar]
- 13. Steil GM, Rebrin K, Darwin C, Hariri F, Saad MF. Feasibility of automating insulin delivery for the treatment of Type 1 diabetes. Diabetes. 2006;55(12):3344-3350. [DOI] [PubMed] [Google Scholar]
- 14. Garg SK, McVean JJ. Development and future of Automated Insulin Delivery (AID) systems. Diabetes Technol Ther. 2024;26(suppl. 3):1-6. doi: 10.1089/dia.2023.0467. [DOI] [PubMed] [Google Scholar]
- 15. Deiss D, Bolinder J, Riveline JP, et al. Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring. Diabetes Care. 2006;29(12):2730-2732. [DOI] [PubMed] [Google Scholar]
- 16. Garg S, Zisser H, Schwartz S, et al. Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor. Diabetes Care. 2006;29(1):44-50. [DOI] [PubMed] [Google Scholar]
- 17. The Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group. Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med. 2008;359:1464-1476. [DOI] [PubMed] [Google Scholar]
- 18. Klonoff DC. Continuous glucose monitoring: roadmap for 21st century diabetes therapy. Diabetes Care. 2005;28(5):1231-1239. [DOI] [PubMed] [Google Scholar]
- 19. Hovorka R. Continuous glucose monitoring and closed-loop systems. Diabet Med. 2006;23:1-12. [DOI] [PubMed] [Google Scholar]
- 20. Klonoff DC. The artificial pancreas: how sweet engineering will solve bitter problems. J Diabetes Sci Technol. 2007;1(1):72-81. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Buckingham B, Cobry E, Clinton P, et al. Preventing hypoglycemia using predictive alarm algorithms and insulin pump suspension. Diabetes Technol Ther. 2009;11(2):93-97. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Dassau E, Zisser H, Palerm C, Buckingham BA, Jovanovic L, Doyle FJ, 3rd. Modular artificial beta-cell system: a prototype for clinical research. J Diabetes Sci Technol. 2008;2(5):863-872. doi: 10.1177/193229680800200518. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Breton M, Farret A, Bruttomesso D, et al. Fully integrated artificial pancreas in type 1 diabetes: modular closed-loop glucose control maintains near normoglycemia. Diabetes. 2012;61(9):2230-2237. doi: 10.2337/db11-1445. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Dassau E, Zisser H, Harvey RA, et al. Clinical evaluation of a personalized artificial pancreas. Diabetes Care. 2013;36(4):801-809. doi: 10.2337/dc12-0948. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Zisser H, Palerm C, Finan D, et al. Clinical insights from preliminary testing of an artificial β-cell control algorithm—modeling and tuning. In: Klonoff DC. Summary of Seventh Annual Diabetes Technology Meeting October 25-27, 2007 San Francisco, California. J Diabetes Sci Technol. 2008;2:A187. [PMC free article] [PubMed] [Google Scholar]
- 26. Cobelli C, Kovatchev B. Developing the UVA/Padova type 1 diabetes simulator: modeling, validation, refinements, and utility. J Diabetes Sci Technol. 2023;17(6):1493-1505. doi: 10.1177/19322968231195081. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Phillip M, Battelino T, Atlas E, et al. Nocturnal glucose control with an artificial pancreas at a diabetes camp. N Engl J Med. 2013;368:824-833. [DOI] [PubMed] [Google Scholar]
- 28. Nimri R, Danne T, Kordonouri O, et al. The “glucositter” overnight automated closed loop system for type 1 diabetes: a randomized crossover trial. Pediatr Diabetes. 2013;14:159-167. [DOI] [PubMed] [Google Scholar]
- 29. Friedrich MJ. Artificial pancreas may soon be a reality. JAMA. 2009;301:1525-1527. [DOI] [PubMed] [Google Scholar]
- 30. Dolgin E. Managed by machine. Nature. 2012;485:S6-S8. [DOI] [PubMed] [Google Scholar]
- 31. Clery D. A pancreas in a box. Science. 2014;343:133-115. [DOI] [PubMed] [Google Scholar]
- 32. Keith Hynes P, Guerlain S, Mize LB, et al. DiAs user interface: a patient-centric interface for mobile artificial pancreas systems. J Diabetes Sci Technol. 2013;7:1416-1426. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Place J, Robert A, Ben Brahim N, et al. DiAs web monitoring: a real-time remote monitoring system designed for artificial pancreas outpatient trials. J Diabetes Sci Technol. 2013;7:1427-1435. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Cobelli C, Renard E, Kovatchev BP, et al. Pilot studies of wearable artificial pancreas in type 1 diabetes. Diabetes Care. 2012;35:e65-e67. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Kovatchev BP, Renard E, Cobelli C, et al. Feasibility of outpatient fully integrated closed-loop control: first studies of wearable artificial pancreas. Diabetes Care. 2013;36(7):1851-1858. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Kovatchev BP, Renard E, Cobelli C, et al. Safety of outpatient closed-loop control: first randomized crossover trials of a wearable artificial pancreas. Diabetes Care. 2014;37(7):1789-1796. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Anderson SM, Raghinaru D, Pinsker JE, et al. Multinational home use of closed-loop control is safe and effective. Diabetes Care. 2016;39(7):1143-1150. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Renard E, Farret A, Kropff J, et al. AP@home Consortium. Day and night closed loop glucose control in patients with type 1 diabetes under free-living conditions: comparison of a single-arm, 1-month experience to results of a previously reported feasibility study of evening and night at home. Diabetes Care. 2016;39:1151-1160. [DOI] [PubMed] [Google Scholar]
- 39. Del Favero S, Boscari F, Messori M, et al. Randomized summer camp crossover trial in 5- to 9-year-old children: outpatient wearable artificial pancreas is feasible and safe. Diabetes Care. 2016;39(7):1180-1185. [DOI] [PubMed] [Google Scholar]
- 40. Kovatchev B, Cheng P, Anderson SM, et al. Feasibility of long-term closed-loop control: a multicenter 6-month trial of 24/7 automated insulin delivery. Diabetes Technol Ther. 2017;19(1):18-24. [DOI] [PubMed] [Google Scholar]
- 41. Anderson SM, Buckingham BA, Breton MD, et al. Hybrid closed-loop control is safe and effective for people with type 1 diabetes who are at moderate to high risk for hypoglycemia. Diabetes Technol Ther. 2019;21(6):356-363. doi: 10.1089/dia.2019.0018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Breton MD, Cherñavvsky DR, Forlenza GP, et al. Closed loop control during intense prolonged outdoor exercise in adolescents with type 1 diabetes: the artificial pancreas ski study. Diabetes Care. 2017;40(12):1644-1650. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Doyle FJ, III, Huyett LM, Lee JB, Zisser HC, Dassau E. Closed-loop artificial pancreas systems: engineering the algorithms. Diabetes Care. 2014;37(5):1191-1197. doi: 10.2337/dc13-2108. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Kovatchev B, Tamborlane WV, Cefalu WT, Cobelli C. The artificial pancreas in 2016: a digital treatment ecosystem for diabetes. Diabetes Care. 2016;39(7):1123-1126. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Tauschmann M, Allen JM, Wilinska ME, et al. Day-and-night hybrid closed-loop insulin delivery in adolescents with type 1 diabetes: a free-living, randomized clinical trial. Diabetes Care. 2016;39(7):1168-1174. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46. Stewart ZA, Wilinska ME, Hartnell S, et al. Closed-loop insulin delivery during pregnancy in women with type 1 diabetes. N Engl J Med. 2016;375:644-654. [DOI] [PubMed] [Google Scholar]
- 47. Kaur RJ, Smith BH, Ozaslan B, et al. Hypoglycemia in prospective multicenter study of pregnancies with pre-existing type 1 diabetes on sensor-augmented pump therapy: the LOIS-P study. Diabetes Technol Ther. 2022;24(8):544-555. doi: 10.1089/dia.2021.0479. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48. Thabit H, Hartnell S, Allen JM, et al. Closed-loop insulin delivery in inpatients with type 2 diabetes: a randomised, parallel-group trial. Lancet Diabetes Endocrinol. 2017;5(2):117-124. [DOI] [PubMed] [Google Scholar]
- 49. Bergenstal RM, Garg S, Weinzimer SA, et al. Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes. JAMA. 2016;316:1407-1408. [DOI] [PubMed] [Google Scholar]
- 50. Weisman A, Bai JW, Cardinez M, Kramer CK, Perkins BA. Effect of artificial pancreas systems on glycaemic control in patients with type 1 diabetes: a systematic review and meta-analysis of outpatient randomised controlled trials. Lancet Diabetes Endocrinol. 2017;5(7):501-512. [DOI] [PubMed] [Google Scholar]
- 51. Bekiari E, Kitsios K, Thabit H, et al. Artificial pancreas treatment for outpatients with type 1 diabetes: systematic review and meta-analysis. BMJ. 2018;361:k1310. doi: 10.1136/bmj.k1310. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52. Boughton CK, Hovorka R. New closed-loop insulin systems. Diabetologia. 2021;64(5):1007-1015. doi:10.1007/s00125-021-05391-w [Erratum in: Diabetologia. 2021;64(6):1455. doi:10.1007/s00125-021-05443-1]. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53. Moon SJ, Jung I, Park CY. Current advances of artificial pancreas systems: a comprehensive review of the clinical evidence. Diabetes Metab J. 2021;45(6):813-839. doi: 10.4093/dmj.2021.0177. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54. Brown SA, Kovatchev BP, Raghinaru D, et al. ; iDCL Trial Research Group. 6-month randomized multicenter trial of closed-loop control in type 1 diabetes. N Engl J Med. 2019;381:1707-1717. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. Breton MD, Kanapka LG, Beck RW, et al. ; iDCL Trial Research Group. A randomized trial of closed-loop control in children with type 1 diabetes. N Engl J Med. 2020;383:836-845. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56. Burnside MJ, Lewis DM, Crocket HR, et al. Open-source automated insulin delivery in type 1 diabetes. N Engl J Med. 2022;387(10):869-881. doi: 10.1056/NEJMoa2203913. [DOI] [PubMed] [Google Scholar]
- 57. Wadwa RP, Reed ZW, Buckingham BA, et al. Trial of hybrid closed-loop control in young children with type 1 diabetes. N Engl J Med. 2023;388(11):991-1001. doi: 10.1056/NEJMoa2210834. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58. Bionic Pancreas Research Group; Russell SJ, Beck RW, Damiano ER, et al. Multicenter, randomized trial of a bionic pancreas in type 1 diabetes. N Engl J Med. 2022;387(13):1161-1172. doi: 10.1056/NEJMoa2205225. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59. Boughton CK, Allen JM, Ware J, et al. Closed-loop therapy and preservation of C-peptide secretion in type 1 diabetes. N Engl J Med. 2022;387(10):882-893. doi: 10.1056/NEJMoa2203496. [DOI] [PubMed] [Google Scholar]
- 60. Lee TTM, Collett C, Bergford S, et al. Automated insulin delivery in women with pregnancy complicated by type 1 diabetes. N Engl J Med. 2023;389(17):1566-1578. doi: 10.1056/NEJMoa2303911. [DOI] [PubMed] [Google Scholar]
- 61. Amer BE, Yaqout YE, Abozaid AM, Afifi E, Aboelkhier MM. Does fully closed-loop automated insulin delivery improve glycaemic control in patients with type 2 diabetes? A meta-analysis of randomized controlled trials. Diabet Med. 2024;41(1):e15196. doi: 10.1111/dme.15196. [DOI] [PubMed] [Google Scholar]
- 62. Bauza C, Kanapka LG, Greene E, Lal RA, Arbiter B, Beck RW. Use of the community-derived open-source automated insulin delivery loop system in type 2 diabetes. Diabetes Technol Ther. 2024;26(7):494-497. doi: 10.1089/dia.2023.0569. [DOI] [PubMed] [Google Scholar]
- 63. Fabris C, Kovatchev B. Real-life use of automated insulin delivery in individuals with type 2 diabetes. J Diabetes Sci Technol. 2024:19322968241274786. doi: 10.1177/19322968241274786. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64. Silva JD, Lepore G, Battelino T, et al. Real-world performance of the MiniMed™ 780G system: first report of outcomes from 4120 users. Diabetes Technol Ther. 2022;24(2):113-119. doi: 10.1089/dia.2021.0203. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65. Gruber N, Wittenberg A, Brener A, et al. Real-life achievements of MiniMed 780G advanced closed-loop system in youth with type 1 diabetes: AWeSoMe study group multicenter prospective trial. Diabetes Technol Ther. 2024;26(11):869-880. doi: 10.1089/dia.2024.0148. [DOI] [PubMed] [Google Scholar]
- 66. Halim B, Abraham MB, Manos G, et al. Advances in automated insulin delivery with the medtronic 780G: the Australian experience. Diabetes Technol Ther. 2024;26(3):190-197. doi: 10.1089/dia.2023.0348. [DOI] [PubMed] [Google Scholar]
- 67. Choudhary P, Arrieta A, van den Heuvel T, Castañeda J, Smaniotto V, Cohen O. Celebrating the data from 100,000 real-world users of the MiniMed™ 780G system in Europe, Middle East, and Africa collected over 3 years: from data to clinical evidence. Diabetes Technol Ther. 2024;26(suppl. 3):32-37. doi: 10.1089/dia.2023.0433. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68. Breton MD, Kovatchev BP. One year real-world use of the control-IQ advanced hybrid closed-loop technology. Diabetes Technol Ther. 2021;23(9):601-608. doi: 10.1089/dia.2021.0097. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69. Kovatchev BP, Singh H, Mueller L, Gonder-Frederick LA. Biobehavioral changes following transition to automated insulin delivery: a large real-life database analysis. Diabetes Care. 2022;45(11):2636-2643. doi: 10.2337/dc22-1217. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70. Wang XS, Dunlop AD, McKeen JA, Feig DS, Donovan LE. Real-world use of Control-IQ™ technology automated insulin delivery in pregnancy: a case series with qualitative interviews. Diabet Med. 2023;40(6):e15086. doi: 10.1111/dme.15086. [DOI] [PubMed] [Google Scholar]
- 71. Graham R, Mueller L, Manning M, et al. Real-world use of control-IQ technology is associated with a lower rate of severe hypoglycemia and diabetic ketoacidosis than historical data: results of the Control-IQ Observational (CLIO) prospective study. Diabetes Technol Ther. 2024;26(1):24-32. doi: 10.1089/dia.2023.0341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72. Marks BE, Meighan S, Zehra A, et al. Real-world glycemic outcomes with early omnipod 5 use in youth with type 1 diabetes. Diabetes Technol Ther. 2023;25(11):782-789. doi: 10.1089/dia.2023.0337. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73. Forlenza GP, DeSalvo DJ, Aleppo G, et al. Real-world evidence of Omnipod® 5 automated insulin delivery system use in 69,902 people with type 1 diabetes. Diabetes Technol Ther. 2024;26(8):514-525. doi: 10.1089/dia.2023.0578. [DOI] [PubMed] [Google Scholar]
- 74. Lawton J, Kimbell B, Closs M, et al. Listening to women: experiences of using closed-loop in type 1 diabetes pregnancy. Diabetes Technol Ther. 2023;25(12):845-855. doi: 10.1089/dia.2023.0323. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75. Benhalima K, Beunen K, Van Wilder N, et al. Comparing advanced hybrid closed loop therapy and standard insulin therapy in pregnant women with type 1 diabetes (CRISTAL): a parallel-group, open-label, randomised controlled trial. Lancet Diabetes Endocrinol. 2024;12(6):390-403. doi: 10.1016/S2213-8587(24)00089-5. [DOI] [PubMed] [Google Scholar]
- 76. Levy CJ, Kudva YC, Ozaslan B, et al. At-home use of a pregnancy-specific zone-MPC closed-loop system for pregnancies complicated by type 1 diabetes: a single-arm, observational multicenter study. Diabetes Care. 2023;46(7):1425-1431. doi: 10.2337/dc23-0173. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77. Guibert C, Amoura L, Rakotoarisoa L, et al. MiniMedTM 780G advanced hybrid closed-loop system study in pregnant women with type 1 diabetes. Diabetes Technol Ther. 2023;25(12):893-901. doi: 10.1089/dia.2023.0267. [DOI] [PubMed] [Google Scholar]
- 78. Quirós C, Herrera Arranz MT, Amigó J, et al. Real-world evidence of off-label use of commercially automated insulin delivery systems compared to multiple daily insulin injections in pregnancies complicated by type 1 diabetes. Diabetes Technol Ther. 2024;26(8):596-606. doi: 10.1089/dia.2023.0594. [DOI] [PubMed] [Google Scholar]
- 79. Dovc K, Bergford S, Fröhlich-Reiterer E, et al. A comparison of faster insulin aspart with standard insulin aspart using hybrid automated insulin delivery system in active children and adolescents with type 1 diabetes: a randomized double-blind crossover trial. Diabetes Technol Ther. 2023;25(9):612-621. doi: 10.1089/dia.2023.0178. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80. Haliloglu B, Boughton CK, Lakshman R, et al. Postprandial glucose excursions with ultra-rapid insulin analogs in hybrid closed-loop therapy for adults with type 1 diabetes. Diabetes Technol Ther. 2024;26(7):449-456. doi: 10.1089/dia.2023.0509. [DOI] [PubMed] [Google Scholar]
- 81. Levy CJ, Bailey R, Laffel LM, et al. Multicenter evaluation of ultra-rapid Lispro insulin with control-IQ technology in adults, adolescents, and children with type 1 diabetes. Diabetes Technol Ther. 2024;26(9):652-660. doi: 10.1089/dia.2024.0048. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82. Nwokolo M, Lakshman R, Hartnell S, et al. CamAPS FX hybrid closed-loop with ultra-rapid lispro compared with standard lispro in adults with type 1 diabetes: a double-blind, randomized, crossover study. Diabetes Technol Ther. 2023;25(12):856-863. doi: 10.1089/dia.2023.0262. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83. McCarthy OM, Christensen MB, Kristensen KB, et al. Automated insulin delivery around exercise in adults with type 1 diabetes: a pilot randomized controlled study. Diabetes Technol Ther. 2023;25(7):476-484. doi: 10.1089/dia.2023.0009. [DOI] [PubMed] [Google Scholar]
- 84. Mameli C, Rigamonti A, Felappi B, et al. Performance of tandem control IQ during outdoor physical activity in children and adolescents with type 1 diabetes. Diabetes Technol Ther. 2024;26(2):112-118. doi: 10.1089/dia.2023.0324. [DOI] [PubMed] [Google Scholar]
- 85. Askari MR, Ahmadasas M, Shahidehpour A, et al. Multivariable automated insulin delivery system for handling planned and spontaneous physical activities. J Diabetes Sci Technol. 2023;17(6):1456-1469. doi: 10.1177/19322968231204884. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86. Hughes MS, Kingman RS, Hsu L, Lal RA, Buckingham BA, Zaharieva DP. Swimming with the omnipod 5 automated insulin delivery system: connectivity in the water. Diabetes Care. 2023;46(8):e148-e150. doi: 10.2337/dc23-0470. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87. Renard E, Joubert M, Villard O, et al. Safety and efficacy of sustained automated insulin delivery compared with sensor and pump therapy in adults with type 1 diabetes at high risk for hypoglycemia: a randomized controlled trial. Diabetes Care. 2023;46(12):2180-2187. doi: 10.2337/dc23-0685. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88. Malone SK, Matus AM, Flatt AJ, et al. Prolonged use of an automated insulin delivery system improves sleep in long-standing type 1 diabetes complicated by impaired awareness of hypoglycemia. J Diabetes Sci Technol. 2023;18:1416-1423. doi: 10.1177/19322968231182406. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89. Lee MH, Gooley J, Obeyesekere V, et al. Hybrid closed loop in adults with type 1 diabetes and severely impaired hypoglycemia awareness. J Diabetes Sci Technol. 2024:19322968241245627. doi: 10.1177/19322968241245627. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90. Sherr JL, Heinemann L, Fleming GA, et al. Automated insulin delivery: benefits, challenges, and recommendations. a consensus report of the joint diabetes technology working group of the European Association for the Study of Diabetes and the American Diabetes Association. Diabetes Care. 2022;45(12):3058-3074. doi: 10.2337/dci22-0018. [DOI] [PubMed] [Google Scholar]
- 91. Sherr JL, Heinemann L, Fleming GA, et al. Automated insulin delivery: benefits, challenges, and recommendations. A consensus report of the joint diabetes technology working group of the European Association for the Study of Diabetes and the American Diabetes Association. Diabetologia. 2023;66(1):3-22. doi: 10.1007/s00125-022-05744-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92. Phillip M, Nimri R, Bergenstal RM, et al. Consensus recommendations for the use of automated insulin delivery technologies in clinical practice. Endocr Rev. 2023;44(2):254-280. doi: 10.1210/endrev/bnac022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93. Aaron RE, Tian T, Yeung AM, et al. NIH fifth artificial pancreas workshop 2023: meeting report: the fifth artificial pancreas workshop: enabling fully automation, access, and adoption. J Diabetes Sci Technol. 2024;18(1):215-239. doi: 10.1177/19322968231201829. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94. Shi D, Dassau E, Doyle FJ, III. Multivariate learning framework for long-term adaptation in the artificial pancreas. Bioeng Transl Med. 2018;4(1):61-74. doi: 10.1002/btm2.10119. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95. Pinsker JE, Laguna Sanz AJ, Lee JB, et al. Evaluation of an artificial pancreas with enhanced model predictive control and a glucose prediction trust index with unannounced exercise. Diabetes Technol Ther. 2018;20(7):455-464. doi: 10.1089/dia.2018.0031. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96. Basu R, Di Camillo B, Toffolo G, et al. Use of a novel triple-tracer approach to assess postprandial glucose metabolism. Am J Physiol Endocrinol Metab. 2003;284(1):E55-E69. doi: 10.1152/ajpendo.00190.2001. [DOI] [PubMed] [Google Scholar]
- 97. Dalla Man C, Rizza RA, Cobelli C. Meal simulation model of the glucose-insulin system. IEEE Trans Biomed Eng. 2007;54(10):1740-1749. doi: 10.1109/TBME.2007.893506. [DOI] [PubMed] [Google Scholar]
- 98. Cobelli C, Man CD, Sparacino G, Magni L, De Nicolao G, Kovatchev BP. Diabetes: models, signals, and control. IEEE Rev Biomed Eng. 2009;2:54-96. doi: 10.1109/RBME.2009.2036073. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99. Jacobs PG, Herrero P, Facchinetti A, et al. Artificial intelligence and machine learning for improving glycemic control in diabetes: best practices, pitfalls, and opportunities. IEEE Rev Biomed Eng. 2024;17:19-41. doi: 10.1109/RBME.2023.3331297. [DOI] [PubMed] [Google Scholar]
- 100. Castillo A, Villa-Tamayo MF, Pryor E, et al. Deep neural network architectures for an embedded MPC Implementation: application to an automated insulin delivery system. 22nd IFAC World Congress, 2023. [Google Scholar]
- 101. Kovatchev B, Castillo A, Pryor E, et al. Neural-net artificial pancreas: a randomized crossover trial of a first-in-class automated insulin delivery algorithm. Diabetes Technol Ther. 2024;26(6):375-382. doi: 10.1089/dia.2023.0469. [DOI] [PMC free article] [PubMed] [Google Scholar]