Metabolic Models, in Silico Trials, and Algorithms*

Ali Cinar 1,, Ananda Basu 2, B Wayne Bequette 3, Marc D Breton 4, Bruce Buckingham 5, Eda Cengiz 6, Claudio Cobelli 7, Eyal Dassau 8, Francis J Doyle III 9, Chiara Fabris 4, Andrea Facchinetti 7, Irl Hirsch 10, Roman Hovorka 11,12, Peter G Jacobs 13,14, Boris P Kovatchev 4, Chiara Dalla Man 7, Laurie Quinn 15, Jay Skyler 16
PMCID: PMC12213549  PMID: 40590459

Abstract

Artificial pancreas (AP) systems, also called automated insulin delivery systems, have improved the time in range of glucose levels, reduced the daily burden of the user for glucose regulation, and improved their quality of life. Several commercially available AP systems operate in hybrid closed-loop mode that requires manual information from the user for meals and exercise. This article summarizes the progress on mathematical models of glucose-insulin dynamics, continuous glucose monitoring systems, and insulin pumps that form the building blocks of AP systems, the shift from animal studies to in silico clinical trials that accelerated the rate of progress in AP technologies and the efforts for developing the next-generation AP systems, and the fully automated AP that eliminates manual inputs and mitigates the effects of disturbances to glucose homeostasis—meals, physical activities, acute stress, and variations in sleep characteristics. A section is devoted to discuss the unique glycemic management challenges faced by women with diabetes across the lifespan (menstrual cycle, menopause, pregnancy) and summarize progress made to reduce their impact on glycemic management.

Keywords: mathematical models, artificial pancreas, simulators for in silico clinical trials, glucose control algorithms, digital twins, women with diabetes


This article is part of a collection of publications to commemorate the 75th Anniversary of the NIDDK.

The chapter focuses on the development of metabolic models, in silico trials, and control algorithms. The mission of NIDDK is to conduct and support basic, preclinical, and clinical research and to disseminate science-based information on diabetes and other metabolic diseases to improve peoples’ health. In this article, we are celebrating the 75 years of progress enabled by NIDDK to improve treatments and quality of life for people living with type 1 diabetes, focusing on the building blocks that led to the development of the artificial pancreas (AP) system. The commercialization of AP systems was only possible through substantial support from the NIDDK.

Automated insulin delivery (AID) has transformed care for people with T1D, enabling glycosylated hemoglobin (HbA1c) reductions of 0.2% to 0.5% compared with open-loop therapy.1-4 Yet, studies indicate that a majority of people using AID still are not achieving the target of 70% in range (70-180 mg/dL) 5 as recommended by the American Diabetes Association. Multiple possible reasons may cause this, including carbohydrate misestimations, 6 forgetting to dose for meals or dosing after the meal event has occurred, high carbohydrate diets, sedentary, highly active lifestyles, or variations in lifestyles, limited information used by AID systems especially for fully automated AID systems, and shortcomings in AID’s control system logic. This chapter is highlighting the progress made in the journey of AID system development and future directions of progress.

The Early Years: The Building Blocks for Automated Insulin Delivery Systems—Models, Insulin Pumps, and First Continuous Glucose Monitors

Diabetes is a condition known for more than 3000 years. Historical progress of medical knowledge and hypotheses leading to qualitative models 7 provide a strong foundation for recent progress in advanced treatments such as AP (also called AID systems for insulin-only AP systems). Since the discovery of insulin more than a century ago, scientists have tirelessly strived to advance its administration to improve blood glucose concentration (BGC) control in people with diabetes mellitus. The current AP systems represent the culmination of inspired collaboration between physician-scientists, control engineers, and skilled mathematicians over the last 50 years.

The first prototypes of AP systems for closed-loop control (CLC) of BGC, involved recurring intravenous measurements of BGC coupled with intravenous insulin delivery,8-12 resulting in prototypical commercial devices such as the Biostator and the Nikkiso STG-22 systems.13,14 However, these devices, although clearly impractical for daily outpatient use, demonstrated the possibility of closed-loop glucose control.

Advances in technologies to capture information about BGC and insulin pumps were complemented by the development of effective CLC strategies. The emergence of commercially available minimally invasive continuous glucose monitoring (CGM) systems by MiniMed (Medtronic), revolutionized this field that saw the forerunners of subcutaneous-subcutaneous CLC systems for effective glucose control in feasibility studies.15-17

The next set of landmark studies18,19 established the feasibility and efficacy of subcutaneous continuous insulin delivery via insulin pumps in type 1 diabetes (T1D) and paved the way for further research in intravenous glucose sensing coupled with intraperitoneal insulin delivery.20,21 The early real-time CGM devices (Medtronic, Dexcom, Abbott) had high levels of inaccuracies (mean absolute relative differences—MARDs of 15%-21%), 22 especially in the first 24 hours and also in the hypoglycemia ranges. These devices required multiple daily calibrations with fingerstick glucose measurements and experienced glitches with drug interference, most notably, acetaminophen (Tylenol).23,24 In silico simulations determined that a sensor MARD of ≤10% would permit safe, non-adjunctive use of CGM data for insulin dosing. 25 Continuous research and development in the CGM field resulted in factory-calibrated CGMs with MARD <10% that can be used for 10 or 14 days and improved reliability of glucose data from CGM to the CLC systems.

Early insulin pumps included the “Autosyringe” of Dean Kamen, the Mill-Hill pump of John Pickup, and Minimed of Al Mann. The “Backpack” of Arnold Kadish in 1963, provided dual-hormone (insulin/glucagon) delivery. Two pumps: Minimed in the United States and Disetronic (later Ypsomed) in Europe offered pumps with tubes for a number of years. Mann Foundation and Medtronic proposed pumps in peritoneal cavity. A patch pump, OmniPod, was introduced by Insulet. 26

Since 1970s, there has been interest in developing quantitative models of glucose-insulin dynamics. 27 These models were generally detailed physiological models focused on predicting a population average from clinical trial data. They enabled the early work (Figure 1) on the development of model predictive control (MPC) algorithms for AID systems.28,29 Bergman and Cobelli reported the minimal model, with two quasi-linear differential equations: one representing insulin kinetics in plasma and a second representing the effects of insulin and glucose itself on restoration of the glucose after perturbation by intravenous injection. 30

Figure 1.

Figure 1.

Milestones for AP development, progress, and funding initiatives.

A milestone achievement was the development of the University of Virginia-University of Padova (UVa-Padova) T1D Simulator,31-34 accepted by the U.S. Food and Drug Administration (FDA) in 2008, that was underpinned by extensive dynamic profiling of carbohydrate metabolism in over 200 adults at Mayo Clinic. 35 This enabled rapid in silico testing of CLC algorithms’ performance on virtual patients with different metabolic characteristics, eliminating the need for preclinical animal trials of CLC systems.

The critical aspects of progress in AP systems include iterative refinements of the essential components of the AP system—CGM, controller platform often on a smartphone and the insulin pump—and their connectivity and communication with each other. Early systems were larger and multicomponent (eg, Dexcom Sevenplus CGM; DiAs platform; 36 Medtronic or Tandem t-slim pump) with suboptimal performance and frequent connectivity issues.37,38 Significant progress continues to be made in all these aspects. Recent advances in this century including AP use in pregnancy,39,40 during menstrual cycle41,42 and inpatient management 43 will be highlighted in this commemorative venture.

None of the research and progress in the AP arena could have been possible without the unprecedented commitment of the researchers and equally importantly, funding support from various agencies—National Institutes of Health, Juvenile Diabetes Research Foundation, European Commission, Helmsley Charitable Trust and Industry partners—whose combined forces made safe and effective CLC insulin delivery an everyday reality in the United States, Europe, and expanding to other parts of the world.

Animals Replaced by In Silico Trials: The University of Virginia-University of Padova Simulator of the Human Metabolic System in Type 1 Diabetes

Given the large observed inter-individual variability, an average model cannot describe the spectrum of individual responses to diabetes treatment. A cohort of in silico “subjects” that spans the inter-individual variability observed in the general population of people with diabetes would enable a more realistic assessment of the performance of a proposed treatment.

Data from 204 individuals without diabetes 35 who underwent a triple tracer meal protocol provided plasma glucose and insulin concentrations, and model-independent estimates of fundamental fluxes of the glucose system enabled the development of a large-scale glucose-insulin model. The model had 18 differential equations with 42 parameters (33 free and 9 derived from steady-state constraints). Estimates of the model parameters of the 204 subjects enabled the generation of any number of virtual subjects by using the joint multivariate probability distribution of the model parameters.

As the AP development research accelerated with the JDRF initiated Artificial Pancreas Project in 2006 and significant investment by the National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK), the demonstration of the safety and feasibility of AP systems in animals, before any testing in humans slowed the progress. A simulator of type 1 diabetes mellitus (T1DM) having virtual patients with realistic inter-subject variability could enable fast and extensive testing of AP control algorithms and be more representative than small-size animal trials. Researchers at the UVa-Padova collaborated to modify the model by substituting exogenous insulin delivery in place of endogenous insulin secretion subsystem and addressed inter-person variability, yielding a T1DM simulation model with 13 differential equations and 35 parameters (26 free and 9 derived from steady-state constraints) (Figure 2). Model validity was tested using T1DM data sets including adults, adolescents, and children, resulting in the UVA-Padova simulator with 100 adult, 100 adolescent, and 100 children virtual subjects. The simulator included CGM and insulin pump models, enabling testing of any meal and insulin delivery scenario in silico. The simulator was accepted by the FDA as a substitute to animal trials for the preclinical testing of control strategies in AP studies in January 2008.

Figure 2.

Figure 2.

Scheme of the glucose metabolism model in the UVa-Padova T1DM simulator (45).

The UVa-Padova simulator focused on the effects of meals on glucose dynamics. As new clinical data were collected from many clinical studies, its features were enhanced over the years 33 and simulation period increased. 34 Physiological inputs, based on carefully conducted physiology studies in T1D44-47 enabled the incorporation of factors such as circadian changes to insulin sensitivity and dawn phenomenon into the simulator.

Several research-grade simulators have been developed with capabilities to simulate the effects of physical activities and generate additional signals such as energy expenditure from the model as inputs to control algorithms48-50 and dual-hormone AIDs. 51

Designing the Algorithms: Major Problems and Innovative Solutions

The heart of the AP system is the control algorithm that links the input—subcutaneous interstitial glucose measurements—with the output—subcutaneous insulin infusion rates—via the insulin pump. The CLC algorithms need to account for several physiological delays involved in glucose transport from the vascular to the interstitial compartment; insulin absorption and transport from the subcutaneous to the venous compartment; arterial insulin delivery to end organs such as liver, muscle, and fat tissue from the venous compartment; and insulin action on responsive tissues viz, muscle and liver. These algorithms also need to accommodate rapid and dynamic changes to the metabolic state that occur frequently in free-living—meals and physical activity. Early efforts in CLC algorithm design used proportional-integral-derivative (PID) control that relied on the difference between the CGM-reported GC and the target GC. It was supplemented by estimators that used mathematical models describing the dynamics of glucose metabolism.52,53 Over time, powerful GC control algorithms based on MPC (Figure 3) and adaptive control were developed.38,15,54-61 This generation of AID systems was called hybrid CLC because of their dependency on user-provided meal and exercise information.

Figure 3.

Figure 3.

Consensus CGM outcomes for baseline and 12 months Control-IQ use broken down by age groups. 54

Models used in simulators and in CLC algorithms have different tasks and characteristics. Simulator models must have a high level of accuracy. Physiology-based compartment models have been used for this purpose.15,62-65 Their parameters were fitted to data from individual subjects, to create a pool of values for each model parameter. These sets of values enabled the development of statistical distributions for each parameter for use in generating many virtual patients for the simulators. Recently, deep neural networks (DNNs) were also considered to develop models, but they require significantly more data. Physiology-based DNNs can merge both approaches. Models used in AP systems are more compact for computational efficiency. Both physiology-based models and data-driven models are used to build the state-space models that form the core of MPC algorithms.

Two physiology-based models have been used in developing most MPC systems: Hovorka model 15 and UVa-Padova model.63,64,66 In the early stages of MPC algorithm development, fixed-parameter linear state-space models were used. Parameter adaptation was introduced to improve CLC performance. In most algorithms, adaptation was made infrequently, mostly once a day. Data-driven models based on time series model structure provide more compact state-space models and permit higher frequencies of adaptation.67-69 The rationale for frequent adaptation (every 5 minutes when new CGM data are provided) is to adjust the model to rapidly changing metabolic dynamics caused by food consumption or physical activities. To reduce the computational burden and frequent adjustments of insulin infusion rates, zone MPC algorithms have been proposed.70,71

Large disturbances to glucose homeostasis can be caused by food consumption, physical activities, acute psychological stress, irregularities in sleep patterns, and metabolic variations in the female body caused by factors such as pregnancy and menstrual cycle. Recent work focused on the detection of meals and physical activities, the estimation of their characteristics, and forecasting their potential effects on GC in the quest for fully automated AP systems.

A major challenge to the performance of all control algorithms is the delays caused by inferring BGC from subcutaneous CG and the delay of the insulin infused to subcutaneous tissue to affect the glucose metabolism. These delays can be accommodated under manual control by providing insulin boluses before consumption of a meal or adjustment of target glucose values and basal insulin levels before starting an exercise. However, an AP does not know about these future activities. The hybrid CLC system required such manual inputs at the expense of increasing the daily burden on the user. Eliminating the need for manual user inputs to achieve fully automated AP systems continues to be an active research and development area.

Several approaches have been proposed to eliminate manual meal information entry:

  • (1) Determining meals times and characteristics from historical data to predict the time and dose of insulin boluses.72,73 Large amounts of individual historical data are needed to improve accuracy, and precautions must be added to modify bolusing if the user’s pattern changes on a specific day;

  • (2) Extending the historical data-based approach to capture the various patterns of behavior of an individual and use these patterns to select the current behavior pattern of this individual. Use this information to provide a sequence of miniboluses as evidence is developing over time on the occurrence of a meal.74,75

  • (3) Detecting the occurrence of a meal from current CGM data, the estimation of meal characteristics from recent CGM data and suggesting insulin miniboluses based on dynamic estimation of meal characteristics.76,77

  • (4) Leveraging AI by developing Bayesian networks to estimate current and future meals from historical and current data, and adjusting insulin infusions accordingly. 78

  • (5) Leveraging accelerometer data from a smartwatch, eating and drinking can be detected based on hand gestures in real-time.79,80

A combination of these techniques can detect meals as soon as they begin, estimate carbohydrate content based on CGM data in real-time, affirm by comparing to individual behavior patterns.

Some approaches for automating the detection of physical activities and adjustment of insulin-dosing/target glucose levels relied on strategies similar to meal detection from historical data.81,82 Others approaches used information such as heart rate or energy expenditure variations collected in real-time.37,71,83-87 Relying exclusively on heart rate signals poses a challenge since psychological stress can also increase heart rate yet will have opposite effects on BGC.88,89 With the recent increase in attention to anxieties and stress, it is foreseeable to expect additional modules in AP algorithms to detect, discriminate, and quantify stress.89,90

The control algorithms have many additional features and constraints for reducing the possibility of aggressive control actions that may lead to hypoglycemia or hyperglycemia, and for minimizing the daily total insulin dose. They are developed as a modular architecture to enable optimization for improved efficiency60,91 and incorporate of algorithms for fault detection,92,93 diagnosis, and mitigation of their effects. Pressure-induced sensor attenuation, 94 imputation of missing sensor data, 95 pump and insulin delivery faults,96,97 and errors in control logic have been addressed.

A recent initiative is to replace the MPC that requires real-time optimization at each sampling time with a neural network trained to mimic the MPC, eliminating the time and resource consumption by the repetitive optimization computations. 98

Dual-hormone AP systems that automate glucagon infusions have been an attractive design that would reduce the potential of hypoglycemia episodes.99-104 The lack of long-term-stable glucagon, user concerns to carry two pumps, having two cannulas for insulin and glucagon, and potential interactions between insulin and glucagon controllers are being addressed by various research teams and companies 105 and clinical trials are being conducted to assess the performance of dual-hormone AP systems.106,107

Digital Twins Technology to Optimize and Personalize Closed-Loop Therapy

A major reason why many people with T1D may fail to achieve clinically recommended time-in-range targets is the lack of personalization of AID control algorithms. Historically, they have been designed to perform well for an entire population of people with T1D, rather than being tailored to individual needs. As a result, an AID control algorithm may be effective for one person but not for others. One way to address personalization is the use of recursively updated models in MPC algorithms.87,108 Digital twins (DTs) offer an alternative by enabling simulations with a digital clone of an individual to determine the appropriate insulin therapy for that individual.

A digital twin (DT) is a mathematical representation of a physical system. They have been utilized in automation, 109 manufacturing, 110 and aerospace applications 111 to optimize processes through simulations, in personalized medicine112,113 and in diabetes and obesity management.114,115 In diabetes technology, a DT refers to a metabolic mathematical model of a single individual that describes how glucose dynamics are impacted by carbohydrates, other nutrients, exercise, insulin, and other hormones such as glucagon and pramlintide. The models can be represented by differential equations15,33,116 or neural networks.114,117

The CGM, insulin, and nutrition data can be used to identify each person’s DT, estimating the parameters of the metabolic model. Figure 4 shows real-world CGM, insulin, carbohydrate, and exercise data collected from the T1D Exercise Initiative 118 that are fit to several DTs. 116 Once the closest DT is identified, it can then be used to compare the person’s current insulin-dosing therapy with new therapies by replaying various alternative therapies based on the person’s individual nutrition, exercise, and insulin-dosing choices. Digital twins have been successfully used to identify optimal changes to insulin-dosing strategies by replaying insulin-dosing settings such carbohydrate ratios,119,120 making suggestions on changes to insulin dosing or carbohydrate intake around exercise to avoid hypoglycemia, 121 or to develop new algorithms for decision support system 122 outcomes based on nutrition.

Figure 4.

Figure 4.

Example of a DT fit during aerobic exercise (shaded green area on the right). Top panel is CGM. Blue trace is real-world data from a T1Dexi participant and hypoglycemia occurred post-exercise. Predictions of DT candidates are light gray–dashed lines. The black-dashed line indicates the best DT most closely matching real-world CGM. The bottom trace shows heart rate changes during and after the exercise event, and carbohydrate intake shown as red X’s with the amount of carbs in grams (124).

Finally, DTs allow patients to experiment with their own data using interactive computer simulations and thereby get insights into what would happen if they change their parameters of their AID system. 123 This latter application was first tested in a pilot clinical study, refined, and deployed in a recently completed larger 6-month randomized clinical study (NCT05610111).

A challenge of the DTs is the necessity to update their parameters as the individuals’ metabolism and lifestyle may change over time. The current DTs are reliable for short periods, but their parameters need to be optimized periodically for longer-duration simulations.

As the DT technology matures, DTs may be used to personalize settings within AIDs to optimize the dosing strategy for each individual. We anticipate that advancements in DT approaches, enhanced by AI and driven by progress in cloud and edge computing, will enable DTs to operate efficiently on devices with limited computational power such as insulin pumps and smartphones.

Extension of Models to Women-Specific Characteristics (Menstrual Cycle, Menopause, Pregnancy)

Women with diabetes (WD) face unique glycemic management challenges across the lifespan. Physiological events such as menstrual cycles, pregnancy, and menopause drastically impact insulin requirements and can exacerbate disease burden, worsen glycemic variability, and increase the risk of complications. Optimizing insulin-dosing algorithms (IDAs) for WD requires addressing sex-specific physiological and lifestyle factors that impact glucose regulation.

The menstrual cycle influences insulin’s glucodynamic action with reduced glucose-lowering action during luteal phase (LP) compared to follicular phase (FP). Data derived from insulin pharmacokinetic and pharmacodynamic studies allowed to quantify the variability in insulin glucodynamic action across menstrual cycle phases in women with T1D. When introduced in a simulator of glucose metabolism in T1D, this variability pattern led to increased time spent >180 and >250 mg/dL during LP vs FP, suggesting that current AID may benefit from informing the IDA with menstrual cycle-related insulin action changes.41,42 Building on this preliminary in silico evidence, there is ongoing work to design and test innovative IDA to optimize insulin treatment across the menstrual cycle in women with T1D.

The WD have an increased risk of pregnancy complications, including pregnancy loss, pre-eclampsia and large-for-gestational-age neonates. Intensification of insulin treatment to achieve glycemic targets is required starting from the pregestational period. There is emerging data regarding the use of machine learning methods for early identification of women with a high risk of developing gestational diabetes. Adverse pregnancy outcomes despite contemporary treatment methods are still inexcusably high for WD. Incremental improvements as small as a 5-unit increase in TIR at 12 weeks reduce the risk of preeclampsia and large for gestational age infants by ~50%. 124 Currently, there is only one AID system that is approved for use in pregnancy, the CamAPS FX (CamDiab Ltd, Cambridge, UK), with a treat-to-target adaptive MPC algorithm. The CamAPS FX use in pregnant women with T1D resulted in a significant increase in time in range (TIR) of 10.5%. 125 The Longitudinal Observation of Insulin Use and Glucose Sensor Metrics in Pregnant Women with Type 1 Diabetes Using Continuous Glucose Monitors and Insulin Pumps (LOIS-P) consortium’s zone-MPC algorithm customized for use in pregnancy, allowing for lower glycemic targets during the day (80-110 mg/dL) and during the night (80-100 mg/dL), yielded favorable glycemic outcomes in a small-scale study. 126 There are evolving research efforts to characterize insulin sensitivity fluctuations across pregnancy trimesters to inform metabolic models and in silico studies for more advanced AID.

Adverse metabolic changes and reduced insulin sensitivity associated with menopause might exacerbate dysglycemia in WD. Data regarding the effect of menopause on insulin action and AID are scarce. Conducting studies to determine the effect of menopause on insulin action would be key to model menopause-associated metabolic changes and customize AID treatment strategies for WD during menopause.

Metabolic models, in silico trials, and advanced IDA have been revolutionizing diabetes care for people with diabetes. These tools must be designed with sex-specific considerations and validated extensively. Most of the current AID systems are often trained on generalized data sets. Expanding data sets to include diverse female populations and tailoring models to help optimize insulin dosing based on sex-specific metabolic characteristics represent a crucial step toward achieving personalized diabetes care.

Conclusions

Many of the advances in technology contributing to the transformation of diabetes care have been supported by the NIDDK over the last 75 years. This chapter illustrated the advances in technology, the development of mathematical models, in silico trial systems, and novel control algorithms. It also highlighted current research activities that will enhance the performance of the AP systems (AID and dual hormone), serve the needs of various groups of people with diabetes that have specific challenges. The trend toward eliminating manual information provided by users, accommodation of the effects of meals, physical activities, stress, variations in sleep, and seasonal changes will improve the TIR of glucose levels and the quality of life of people with diabetes. The NIDDK will continue to play a major role in contributing to and enabling these multifaceted advances.

Footnotes

Abbreviations: ADA, American Diabetes Association; AID, automatic insulin delivery system; AP, artificial pancreas; BCG, blood glucose concentration; CGM, continuous glucose monitoring; CLC, closed-loop control; GC, glucose concentration; FP, follicular phase; FDA, U.S. Food and Drug Administration; HbA1c, glycosylated hemoglobin (%); HCL, hybrid closed-loop system; IDA, insulin-dosing algorithm; LP, luteal phase; MARD, mean absolute relative difference; MPC, model predictive control; PID, proportional-integral-derivative control; T1D, type 1 diabetes; T1DM, type 1 diabetes mellitus; WD, women with diabetes.

The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: AB received research funding from NIDDK and Medtronic. MDB reports research grants handled by the University of Virginia from the National Institutes of Health, Novo Nordisk, and Dexcom, and nonfinancial research support from Tandem Diabetes Care, outside of the submitted study. In addition, MDB has patents with royalties licensed to Dexcom, Sanofi, Tandem Diabetes Care, and Novo Nordisk. MDB also reports consulting/speakership activities with Dexcom, Tandem Diabetes Care, Roche, Portal Insulin LLC, BoydSense, and Vertex. BB has received research support from NIDDK, NIH, Breakthrough T1D, Helmsley Charitable Trust, Medtronic Diabetes, Tandem, Insulet, Bigfoot, Dexcom, Abbott, Convatec, and Lilly. Advisory Boards for Medtronic, Tandem, Animas, Sanofi, Roche, Novo Nordisk, Lilly, Convatec, BD, Profusa, and Glysense. AC reports research grants handled by Illinois Institute of Technology from the National Institutes of Health (NIDDK), National Science Foundation, Walder Foundation and Breakthrough T1D (JDRF); research device support from Dexcom; and honoraria from Eli Lilly, Medtronic and American Diabetes Association. EC is scientific advisory board member of Novo Nordisk, Eli Lilly, MannKind, Arecor, Portal Insulin, Tandem, and Ypsomed. FJD reports funding over the time course described in this paper from the Showalter Foundation, Roche Diagnostics Corporation, the Juvenile Diabetes Research Foundation, the Helmsley Charitable Trust, and the National Institutes of Health (NIDDK). FJD also reports royalty payments from patents in a portfolio that have been licensed to Insulet, Roche, and Dexcom. PGJ receives research support from Dexcom and Eli Lilly and is a co-founder and shareholder in Pacific Diabetes Technologies. ED has received personal fees from Roche and Eli Lilly and Company; holds patents on artificial pancreas technology; and has received product support from Insulet Corporation, Tandem Diabetes Care, Roche, and Dexcom, Inc. The work presented in this article was performed as part of his academic appointment and is independent of his employment with Eli Lilly and Company. CF reports receiving research support from Novo Nordisk, Dexcom, and Tandem Diabetes Care handled by the University of Virginia, and patent royalties from Novo Nordisk and Dexcom handled by the University of Virginia’s Licensing and Ventures Group. IBH reports Research funding: NIDDK, Tandem, Dexcom, and MannKind and consulting with Abbott, Roche, and Hagar. PGJ receives research support from Dexcom and Eli Lilly and is a co-founder and shareholder in Pacific Diabetes Technologies. BPK reports research grants handled by the University of Virginia from the National Institutes of Health, Novo Nordisk, Dexcom, and Tandem Diabetes Care. In addition, BPK has patents with royalties licensed to Dexcom and Novo Nordisk. CC, CDM, AF, and LQ have no disclosures relevant to this work. JSS has been an advisor to 4Immune, AbbVie, Abvance, ActoBiotics, Adocia, Aerami/Dance Biopharma, AiTA, Applied Therapuetics, Arecor, AstraZeneca, Avotres, Bayer, Biomea Fusion, COUR Pharmaceuticals, Dexcom, Eli Lilly, Kriya Therapeutics, Levicure, Novo Nordisk, Oramed, Provention Bio, PolTREG, Quell Therapeutics, RegCell, Remedy Plan Inc., Respond Health, Sanofi, Shoreline Biosciences, Signos, Vertex, vTv Therapeutics, and WiNK. He is a member of the Board of Directors of Applied Therapeutics and of SAB Biotherapeutics. He is Chair of the Strategic Advisory Board of the EU EDENT1FI consortium. JSS has equity in: 4Immune, Abvance, AiTA, Applied Therapeutics, Avotres, Dexcom, Immunomolecular Therapeutics, Oramed, SAB Biotherapeutics, Signos, vTv Therapeutics, and WiNK.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

References

  • 1. Bergenstal RM, Garg S, Weinzimer SA, et al. Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes. JAMA. 2016;316(13):1407-1408. [DOI] [PubMed] [Google Scholar]
  • 2. Boughton CK, Hartnell S, Thabit H, et al. Hybrid closed-loop glucose control compared with sensor augmented pump therapy in older adults with type 1 diabetes: an open-label multicentre, multinational, randomised, crossover study. Lancet Healthy Longev. 2022;3(3):e135-e142. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Brown SA, Forlenza GP, Bode BW, et al. Multicenter trial of a tubeless, on-body automated insulin delivery system with customizable glycemic targets in pediatric and adult participants with type 1 diabetes. Diabetes Care. 2021;44(7):1630-1640. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Brown SA, Kovatchev BP, Raghinaru D, et al. Six-month randomized, multicenter trial of closed-loop control in type 1 diabetes. N Engl J Med. 2019;381(18):1707-1717. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Choudhary P, Kolassa R, Keuthage W. Advanced hybrid closed loop study in an adult population with type 1 diabetes (adapt): a randomized controlled study. Lancet Diabetes Endocrinol. 2022;10(10):720-731. [DOI] [PubMed] [Google Scholar]
  • 6. Gillingham MB, Li Z, Beck RW, et al. Assessing mealtime macronutrient content: patient perceptions versus expert analyses via a novel phone app. Diabetes Technol Ther. 2021;23(2):85-94. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Poretsky L. Principles of Diabetes Mellitus. Cham: Springer; 2017:XXIX,1066. [Google Scholar]
  • 8. AH K. Automation control of blood sugar. I. A servomechanism for glucose monitoring and control. Am J Med Electron. 1964;3:82-86. [PubMed] [Google Scholar]
  • 9. Albisser A, Leibel B, Ewart T, et al. An artificial endocrine pancreas. Diabetes. 1974;23(5):389-396. [DOI] [PubMed] [Google Scholar]
  • 10. Kraegen EW, Campbell LV, Chia YO, Meler H, Lazarus L. Control of blood glucose in diabetics using an artificial pancreas. Aust N Z J Med. 1977;7(3):280-286. [DOI] [PubMed] [Google Scholar]
  • 11. Pfeiffer EF, Thum C, Clemens AH. The artificial beta cell-a continuous control of blood sugar by external regulation of insulin infusion (glucose controlled insulin infusion system). Horm Metab Res. 1974;6(5):339-342. [DOI] [PubMed] [Google Scholar]
  • 12. Yamasaki Y, Shichiri M, Kawamori R, et al. Counterregulatory system in an artificial endocrine pancreas glucose infusion algorithm. Artif Organs. 1979;3(3):265-270. [DOI] [PubMed] [Google Scholar]
  • 13. Parra CA. Transepithelial elimination in necrobiosis lipoidica. Br J Dermatol. 1977;96(1):83-86. [DOI] [PubMed] [Google Scholar]
  • 14. Yatabe T, Yamazaki R, Kitagawa H, et al. The evaluation of the ability of closed-loop glycemic control device to maintain the blood glucose concentration in intensive care unit patients. Crit Care Med. 2011;39(3):575-578. [DOI] [PubMed] [Google Scholar]
  • 15. Hovorka R, Canonico V, Chassin LJ, et al. Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes. Physiol Meas. 2004;25(4):905-920. [DOI] [PubMed] [Google Scholar]
  • 16. Steil GM, Rebrin K, Darwin C, Hariri F, Saad MF. Feasibility of automating insulin delivery for the treatment of type 1 diabetes. Diabetes. 2006;55(12):3344-3350. [DOI] [PubMed] [Google Scholar]
  • 17. Weinzimer SA, Steil GM, Swan KL, Dziura J, Kurtz N, Tamborlane WV. Fully automated closed-loop insulin delivery versus semiautomated hybrid control in pediatric patients with type 1 diabetes using an artificial pancreas. Diabetes Care. 2008;31(5):934-939. [DOI] [PubMed] [Google Scholar]
  • 18. Pickup J, Keen H, Parsons J, Alberti K. Continuous subcutaneous insulin infusion: an approach to achieving normoglycaemia. Br Med J. 1978;1(6107):204-207. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Tamborlane WV, Sherwin RS, Genel M, Felig P. Reduction to normal of plasma glucose in juvenile diabetes by subcutaneous administration of insulin with a portable infusion pump. N Engl J Med. 1979;300(11):573-578. [DOI] [PubMed] [Google Scholar]
  • 20. LeBlanc H, Chauvet D, Lombrail P, Robert JJ. Glycemic control with closed-loop intraperitoneal insulin in type I diabetes. Diabetes Care. 1986;9(2):124-128. [DOI] [PubMed] [Google Scholar]
  • 21. Renard E, Place J, Cantwell M, Chevassus H, Palerm CC. Closed-loop insulin delivery using a subcutaneous glucose sensor and intraperitoneal insulin delivery: feasibility study testing a new model for the artificial pancreas. Diabetes Care. 2010;33(1):121-127. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Kovatchev B, Anderson S, Heinemann L, Clarke W. Comparison of the numerical and clinical accuracy of four continuous glucose monitors. Diabetes Care. 2008;31(6):1160-1164. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Basu A, Veettil S, Dyer R, Peyser T, Basu R. Direct evidence of acetaminophen interference with subcutaneous glucose sensing in humans: a pilot study. Diabetes Technol Ther. 2016;18(suppl 2):S243-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Cobelli C, Renard E, Kovatchev B. Artificial pancreas: past, present, future. Diabetes. 2011;60(11):2672-2682. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Kovatchev BP, Patek SD, Ortiz EA, Breton MD. Assessing sensor accuracy for non-adjunct use of continuous glucose monitoring. Diabetes Technol Ther. 2015;17(3):177-186. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Zisser H, Breton M, Dassau E, et al. Novel methodology to determine the accuracy of the OmniPod insulin pump: a key component of the artificial pancreas system. J Diabetes Sci Technol. 2011;5(6):1509-1518. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Viceconti M, Cobelli C, Haddad T, et al. In silico assessment of biomedical products: the conundrum of rare but not so rare events in two case studies. Proc Inst Mech Eng H. 2017;231(5):455-466. [DOI] [PubMed] [Google Scholar]
  • 28. Parker RS, Doyle FJ, Peppas NA. A model-based algorithm for blood glucose control in type I diabetic patients. IEEE Trans Biomed Eng. 2002;46(2):148-157. [DOI] [PubMed] [Google Scholar]
  • 29. Sorensen J. A Physiologic Model of Glucose Metabolism in Man and its Use to Design and Assess Improved Insulin Therapies for Diabetes [PhD dissertation]. Cambridge, MA: Department of Chemical Engineering, Massachusetts Institute of Technology; 1985. [Google Scholar]
  • 30. Bergman RN. Origins and history of the minimal model of glucose regulation. Front Endocrinol. 2021;11:583016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Cobelli C, Kovatchev B. Developing the UVA/Padova type 1 diabetes simulator: modeling, validation, refinements, and utility. J Diabetes Sci Technol. 2023;17(6):1493-1505. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Kovatchev BP, Breton M, Dalla Man C, Cobelli C. In silico preclinical trials: a proof of concept in closed-loop control of type 1 diabetes. Los Angeles, CA: SAGE Publications Sage CA; 2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Dalla Man C, Micheletto F, Lv D, Breton M, Kovatchev B, Cobelli C. The UVA/PADOVA type 1 diabetes simulator: new features. J Diabetes Sci Technol. 2014;8(1):26-34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Visentin R, Campos-Náñez E, Schiavon M, et al. The UVA/Padova type 1 diabetes simulator goes from single meal to single day. J Diabetes Sci Technol. 2018;12(2):273-281. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Basu R, Dalla Man C, Campioni M, et al. Effects of age and sex on postprandial glucose metabolism: differences in glucose turnover, insulin secretion, insulin action, and hepatic insulin extraction. Diabetes. 2006;55(7):2001-2014. [DOI] [PubMed] [Google Scholar]
  • 36. Place J, Robert A, Ben Brahim N, et al. DiAs web monitoring: a real-time remote monitoring system designed for artificial pancreas outpatient trials. J Diabetes Sci Technol. 2013;7(6):1427-1435. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37. Breton MD, Cherñavvsky DR, Forlenza GP, et al. Closed-loop control during intense prolonged outdoor exercise in adolescents with type 1 diabetes: the artificial pancreas ski study. Diabetes Care. 2017;40(12):1644-1650. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Kovatchev B, Cheng P, Anderson SM, et al. Feasibility of long-term closed-loop control: a multicenter 6-month trial of 24/7 automated insulin delivery. Diabetes Technol Ther. 2017;19(1):18-24. [DOI] [PubMed] [Google Scholar]
  • 39. Levy CJ, Kudva YC, Ozaslan B, et al. At-home use of a pregnancy-specific zone-MPC closed-loop system for pregnancies complicated by type 1 diabetes: a single-arm, observational multicenter study. Diabetes Care. 2023;46(7):1425-1431. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Stewart ZA, Wilinska ME, Hartnell S, et al. Closed-loop insulin delivery during pregnancy in women with type 1 diabetes. N Engl J Med. 2016;375(7):644-654. [DOI] [PubMed] [Google Scholar]
  • 41. Diaz C JL, Cengiz E, Breton MD, Fabris C. Modeling the variability of insulin sensitivity during the menstrual cycle in women with type 1 diabetes to adjust open-loop insulin therapy. Annu Int Conf IEEE Eng Med Biol Soc. 2021;2021:1543-1546. [DOI] [PubMed] [Google Scholar]
  • 42. Diaz C JL, Fabris C, Breton MD, Cengiz E. Insulin replacement across the menstrual cycle in women with type 1 diabetes: an in silico assessment of the need for ad hoc technology. Diabetes Technol Ther. 2022;24(11):832-841. [DOI] [PubMed] [Google Scholar]
  • 43. Boughton CK, Bally L, Martignoni F, et al. Fully closed-loop insulin delivery in inpatients receiving nutritional support: a two-centre, open-label, randomised controlled trial. Lancet Diabetes Endocrinol. 2019;7(5):368-377. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Hinshaw L, Dalla Man C, Nandy DK, et al. Diurnal pattern of insulin action in type 1 diabetes: implications for a closed-loop system. Diabetes. 2013;62(7):2223-2229. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Kudva YC, Carter RE, Cobelli C, Basu R, Basu A. Closed-loop artificial pancreas systems: physiological input to enhance next-generation devices. Diabetes Care. 2014;37(5):1184-1190. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Mallad A, Hinshaw L, Dalla Man C, et al. Nocturnal glucose metabolism in type 1 diabetes: a study comparing single versus dual tracer approaches. Diabetes Technol Ther. 2015;17(8):587-595. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Visentin R, Dalla Man C, Kudva A, Basu C, Cobelli AC. Circadian variability of insulin sensitivity: physiological input for in silico artificial pancreas. Diabetes Technol Ther. 2015;17(1):1-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Hobbs N, Samadi S, Rashid M, et al. A physical activity-intensity driven glycemic model for type 1 diabetes. Comput Methods Programs Biomed. 2022;226:107153. [DOI] [PubMed] [Google Scholar]
  • 49. Siket M, Rashid MM, Cinar A. py-mgipsim: an open-source python library for simulating type 1 diabetes with diverse meals and physical activities. J Diabetes Sci Technol. 2025;19(3):855-856. doi: 10.1177/19322968251328664. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50. Rashid M, Samadi S, Sevil M, et al. Simulation software for assessment of nonlinear and adaptive multivariable control algorithms: glucose—insulin dynamics in type 1 diabetes. Comput Chem Eng. 2019;130:106565. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51. Smaoui MR, Rabasa-Lhoret R, Haidar A. Development platform for artificial pancreas algorithms. PLoS ONE. 2020;15(12):e0243139. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52. Pinsker JE, Lee JB, Dassau E, et al. Randomized crossover comparison of personalized MPC and PID control algorithms for the artificial pancreas. Diabetes Care. 2016;39(7):1135-1142. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53. Steil GM. Algorithms for a closed-loop artificial pancreas: the case for proportional-integral-derivative control. J Diabetes Sci Technol. 2013;7(6):1621-1631. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54. Breton MD, Kovatchev BP. One year real-world use of the control-IQ advanced hybrid closed-loop technology. Diabetes Technol Ther. 2021;23(9):601-608. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55. Cameron F, Niemeyer G, Wilson DM, et al. Inpatient trial of an artificial pancreas based on multiple model probabilistic predictive control with repeated large unannounced meals. Diabetes Technol Ther. 2014;16(11):728-734. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56. Cinar A. Automated insulin delivery algorithms. Diabetes Spectr. 2019;32(3):209-214. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57. Forlenza GP, Cameron FM, Ly TT, et al. Fully closed-loop multiple model probabilistic predictive controller artificial pancreas performance in adolescents and adults in a supervised hotel setting. Diabetes Technol Ther. 2018;20(5):335-343. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58. Isganaitis E, Raghinaru D, Ambler-Osborn L, et al. Closed-loop insulin therapy improves glycemic control in adolescents and young adults: outcomes from the international diabetes closed-loop trial. Diabetes Technol Ther. 2021;23(5):342-349. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59. Kudva YC, Laffel LM, Brown SA, et al. Patient-reported outcomes in a randomized trial of closed-loop control: the pivotal international diabetes closed-loop trial. Diabetes Technol Ther. 2021;23(10):673-683. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60. Patek SD, Magni L, Dassau E, et al. Modular closed-loop control of diabetes. IEEE Trans Biomed Eng. 2012;59(11):2986-2999. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61. Resalat N, Hilts W, Youssef JE, Tyler N, Castle JR, Jacobs PG. Adaptive control of an artificial pancreas using model identification, adaptive postprandial insulin delivery, and heart rate and accelerometry as control inputs. J Diabetes Sci Technol. 2019;13(6):1044-1053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62. Ambrus JL, Ambrus JL, Jr, Chadha KC. Interferon inhibitors in lupus erythematosus. N Engl J Med. 1988;319(9):582-583. [DOI] [PubMed] [Google Scholar]
  • 63. Dalla Man C, Raimondo DM, Rizza RA, Cobelli C. GIM, simulation software of meal glucose-insulin model. J Diabetes Sci Technol. 2007;1(3):323-330. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64. Dalla Man C, Rizza RA, Cobelli C. Meal simulation model of the glucose-insulin system. IEEE Trans Biomed Eng. 2007;54(10):1740-1749. [DOI] [PubMed] [Google Scholar]
  • 65. Hovorka R, Shojaee-Moradie F, Carroll PV, et al. Partitioning glucose distribution/transport, disposal, and endogenous production during IVGTT. Am J Physiol Endocrinol Metab. 2002;282(5):E992-1007. [DOI] [PubMed] [Google Scholar]
  • 66. Visentin R, Dalla Man C, Kovatchev B, Cobelli C. The university of Virginia/Padova type 1 diabetes simulator matches the glucose traces of a clinical trial. Diabetes Technol Ther. 2014;16(7):428-434. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67. El-Khatib FH, Russell SJ, Magyar KL, et al. Autonomous and continuous adaptation of a bihormonal bionic pancreas in adults and adolescents with type 1 diabetes. J Clin Endocrinol Metab. 2014;99(5):1701-1711. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68. Hajizadeh I, Rashid M, Samadi S, et al. Adaptive and personalized plasma insulin concentration estimation for artificial pancreas systems. J Diabetes Sci Technol. 2018;12(3):639-649. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69. Turksoy K, Bayrak ES, Quinn L, Littlejohn E, Cinar A. Multivariable adaptive closed-loop control of an artificial pancreas without meal and activity announcement. Diabetes Technol Ther. 2013;15(5):386-400. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70. Forlenza GP, Deshpande S, Ly TT, et al. Application of zone model predictive control artificial pancreas during extended use of infusion set and sensor: a randomized crossover-controlled home-use trial. Diabetes Care. 2017;40(8):1096-1102. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71. Huyett LM, Ly TT, Forlenza GP, et al. Outpatient closed-loop control with unannounced moderate exercise in adolescents using zone model predictive control. Diabetes Technol Ther. 2017;19(6):331-339. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 72. Cameron FM, Ly TT, Buckingham BA, et al. Closed-loop control without meal announcement in type 1 diabetes. Diabetes Technol Ther. 2017;19(9):527-532. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73. Lee H, Buckingham BA, Wilson DM, Bequette BW. A closed-loop artificial pancreas using model predictive control and a sliding meal size estimator. J Diabetes Sci Technol. 2009;3(5):1082-1090. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 74. Corbett JP, Breton MD, Patek SD. A multiple hypothesis approach to estimating meal times in individuals with type 1 diabetes. J Diabetes Sci Technol. 2021;15(1):141-146. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 75. Garcia-Tirado J, Colmegna P, Villard O, et al. Assessment of meal anticipation for improving fully automated insulin delivery in adults with type 1 diabetes. Diabetes Care. 2023;46(9):1652-1658. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76. Samadi S, Rashid M, Turksoy K, et al. Automatic detection and estimation of unannounced meals for multivariable artificial pancreas system. Diabetes Technol Ther. 2018;20(3):235-246. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 77. Samadi S, Turksoy K, Hajizadeh I, Feng J, Sevil M, Cinar A. Meal detection and carbohydrate estimation using continuous glucose sensor data. IEEE J Biomed Health Inform. 2017;21(3):619-627. [DOI] [PubMed] [Google Scholar]
  • 78. Ahmadasas MS, Rashid M, Cinar A, Bilgic M. Stochastic model predictive control of blood glucose levels using probabilistic meal anticipation. Paper presented at the 1st IFAC Workshop on Engineering Diabetes Technologies (EDT 2025); May 2025: Valencia, Spain. [Google Scholar]
  • 79. Corbett JP, Hsu L, Brown SA, et al. Smartwatch gesture-based meal reminders improve glycaemic control. Diabetes Obes Metab. 2022;24(8):1667-1670. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80. Grosman B, Roy A, Lintereur L, et al. A peek under the hood: explaining the MiniMed™ 780G algorithm with meal detection technology. Diabetes Technol Ther. 2024;26(S3):17-23. [DOI] [PubMed] [Google Scholar]
  • 81. Garcia-Tirado J, Brown SA, Laichuthai N, et al. Anticipation of historical exercise patterns by a novel artificial pancreas system reduces hypoglycemia during and after moderate-intensity physical activity in people with type 1 diabetes. Diabetes Technol Ther. 2021;23(4):277-285. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 82. Garcia-Tirado J, Corbett JP, Boiroux D, Jørgensen JB, Breton MD. Closed-loop control with unannounced exercise for adults with type 1 diabetes using the ensemble model predictive control. J Process Control. 2019;80:202-210. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 83. Askari MR, Ahmadasas M, Shahidehpour A, et al. Multivariable automated insulin delivery system for handling planned and spontaneous physical activities. J Diabetes Sci Technol. 2023;17(6):1456-1469. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84. Breton MD, Brown SA, Karvetski CH, et al. Adding heart rate signal to a control-to-range artificial pancreas system improves the protection against hypoglycemia during exercise in type 1 diabetes. Diabetes Technol Ther. 2014;16(8):506-511. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 85. Garcia-Tirado J, Colmegna P, Corbett JP, Ozaslan B, Breton MD. In silico analysis of an exercise-safe artificial pancreas with multistage model predictive control and insulin safety system. J Diabetes Sci Technol. 2019;13(6):1054-1064. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 86. Jacobs PG, Resalat N, El Youssef J, et al. Incorporating an exercise detection, grading, and hormone dosing algorithm into the artificial pancreas using accelerometry and heart rate. J Diabetes Sci Technol. 2015;9(6):1175-1184. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 87. Turksoy K, Hajizadeh I, Hobbs N, et al. Multivariable artificial pancreas for various exercise types and Intensities. Diabetes Technol Ther. 2018;20(10):662-671. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 88. Hobbs N, Brandt R, Maghsoudipour S, et al. Observational study of glycemic impact of anticipatory and early-race athletic competition stress in type 1 diabetes. Front Clin Diabetes Healthc. 2022;3:816316. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 89. Sevil M, Rashid M, Hajizadeh I, Park M, Quinn L, Cinar A. Physical activity and psychological stress detection and assessment of their effects on glucose concentration predictions in diabetes management. IEEE Trans Biomed Eng. 2021;68(7):2251-2260. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90. Sevil M, Rashid M, Hajizadeh I, et al. Discrimination of simultaneous psychological and physical stressors using wristband biosignals. Comput Methods Programs Biomed. 2021;199:105898. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 91. Breton M, Farret A, Bruttomesso D, et al. Fully integrated artificial pancreas in type 1 diabetes: modular closed-loop glucose control maintains near normoglycemia. Diabetes. 2012;61(9):2230-2237. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 92. Bequette BW. Fault detection and safety in closed-loop artificial pancreas systems. J Diabetes Sci Technol. 2014;8(6):1204-1214. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 93. Feng J, Hajizadeh I, Yu X, et al. Multi-level supervision and modification of artificial pancreas control system. Comput Chem Eng. 2018;112:57-69. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 94. Baysal N, Cameron F, Buckingham BA, et al. A novel method to detect pressure-induced sensor attenuations (PISA) in an artificial pancreas. J Diabetes Sci Technol. 2014;8(6):1091-1096. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 95. Feng J, Turksoy K, Samadi S, Hajizadeh I, Littlejohn E, Cinar A. Hybrid online sensor error detection and functional redundancy for systems with time-varying parameters. J Process Control. 2017;60:115-127. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 96. Cescon M, DeSalvo DJ, Ly TT, et al. Early detection of infusion set failure during insulin pump therapy in type 1 diabetes. J Diabetes Sci Technol. 2016;10(6):1268-1276. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 97. Howsmon DP, Baysal N, Buckingham BA, et al. Real-time detection of infusion site failures in a closed-loop artificial pancreas. J Diabetes Sci Technol. 2018;12(3):599-607. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 98. Kovatchev B, Castillo A, Pryor E, et al. Neural-net artificial pancreas: a randomized crossover trial of a first-in-class automated insulin delivery algorithm. Diabetes Technol Ther. 2024;26(6):375-382. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 99. Canadian Paediatric Society Annual Meeting. Vancouver, British Columbia, June 26-30, 1993. Clin Invest Med. 1993;16(3 suppl):A1-52. [PubMed] [Google Scholar]
  • 100. El-Khatib FH, Russell SJ, Nathan DM, Sutherlin RG, Damiano ER. A bihormonal closed-loop artificial pancreas for type 1 diabetes. Sci Transl Med. 2010;2(27):27ra27. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 101. Haidar A, Legault L, Matteau-Pelletier L, et al. Outpatient overnight glucose control with dual-hormone artificial pancreas, single-hormone artificial pancreas, or conventional insulin pump therapy in children and adolescents with type 1 diabetes: an open-label, randomised controlled trial. Lancet Diabetes Endocrinol. 2015;3(8):595-604. [DOI] [PubMed] [Google Scholar]
  • 102. Jacobs PG, El Youssef J, Reddy R, et al. Randomized trial of a dual-hormone artificial pancreas with dosing adjustment during exercise compared with no adjustment and sensor-augmented pump therapy. Diabetes Obes Metab. 2016;18(11):1110-1119. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 103. Lindkvist EB, Laugesen C, Reenberg AT, et al. Performance of a dual-hormone closed-loop system versus insulin-only closed-loop system in adolescents with type 1 diabetes. A single-blind, randomized, controlled, crossover trial. Front Endocrinol (Lausanne). 2023;14:1073388. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 104. Rayannavar A, Mitteer LM, Balliro CA, et al. The bihormonal bionic pancreas improves glycemic control in individuals with hyperinsulinism and postpancreatectomy diabetes: a pilot study. Diabetes Care. 2021;44(11):2582-2585. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 105. Infante M, Baidal DA, Rickels MR, et al. Dual-hormone artificial pancreas for management of type 1 diabetes: Recent progress and future directions. Artif Organs. 2021;45(9):968-986. doi:10.1111/aor.14023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 106. El-Khatib FH, Balliro C, Hillard MA, et al. Home use of a bihormonal bionic pancreas versus insulin pump therapy in adults with type 1 diabetes: a multicentre randomised crossover trial. Lancet. 2017;389(10067):369-380. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 107. Russell SJ, Beck RW, Damiano ER, et al. Multicenter, randomized trial of a bionic pancreas in type 1 diabetes. N Engl J Med. 2022;387(13):1161-1172. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 108. Hajizadeh I, Rashid M, Cinar A. Plasma-insulin-cognizant adaptive model predictive control for artificial pancreas systems. J Process Control. 2019;77:97-113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 109. Attaran M, Celik BG. Digital twin: benefits, use cases, challenges, and opportunities. Decis Anal J. 2023;6:100165. [Google Scholar]
  • 110. Cimino C, Negri E, Fumagalli L. Review of digital twin applications in manufacturing. Comput Ind. 2019;113:103130. [Google Scholar]
  • 111. Glaessgen EH, Stargel D. The digital twin paradigm for future NASA and U.S. air force vehicles. Paper presented at the 53rd AIAA/ASME/ASCE/AHS/ASC Structures, Structural Dynamics and Materials Conference; April 2012; Honolulu, Hawaii. [Google Scholar]
  • 112. Kamel Boulos MN, Zhang P. Digital twins: from personalised medicine to precision public health. J Pers Med. 2021;11(8):745. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 113. Sahal R, Alsamhi SH, Brown KN. Personal digital twin: a close look into the present and a step towards the future of personalised healthcare industry. Sensors (Basel). 2022;22(15):5918. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 114. Mosquera-Lopez C, Jacobs PG. Digital twins and artificial intelligence in metabolic disease research. Trends Endocrinol Metab. 2024;35(6):549-557. [DOI] [PubMed] [Google Scholar]
  • 115. Cappon G, Facchinetti A. Digital twins in type 1 diabetes: a systematic review. J Diabetes Sci Technol. 2024. doi: 10.1177/19322968241262112. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 116. Resalat N, El Youssef J, Tyler N, Castle J, Jacobs PG. A statistical virtual patient population for the glucoregulatory system in type 1 diabetes with integrated exercise model. PLoS ONE. 2019;14(7):e0217301-17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 117. Lv Z, Xie S. Artificial intelligence in the digital twins: state of the art, challenges, and future research topics. DigitalTwin. 2021;1:12. [Google Scholar]
  • 118. Riddell M, Li Z, Gal R, et al. The type 1 diabetes exercise initiative (T1DEXI): examining the acute glycemic effects of different types of structured exercise sessions in type 1 diabetes in a real-world settting. Diabetes Care. 2023;46(4):704-713. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 119. Cappon G, Vettoretti M, Sparacino G, Favero SD, Facchinetti A. ReplayBG: a digital twin-based methodology to identify a personalized model from type 1 diabetes data and simulate glucose concentrations to assess alternative therapies. IEEE Trans Biomed Eng. 2023;70(11):3227-3238. [DOI] [PubMed] [Google Scholar]
  • 120. Patek SD, Lv D, Ortiz EA, et al. Empirical representation of blood glucose variability in a compartmental model. Prediction Methods for Blood Glucose Concentration. Cham: Springer; 2016. [Google Scholar]
  • 121. Young G, Dodier R, Youssef JE, et al. Design and in silico evaluation of an exercise decision support system using digital twin models. J Diabetes Sci Technol. 2024;18(2):324-334. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 122. Pellizzari E, Prendin F, Cappon G, Sparacino G, Facchinetti A. drCORRECT: an algorithm for the preventive administration of postprandial corrective insulin boluses in type 1 diabetes management. J Diabetes Sci Technol. 2023;19(3):711-721. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 123. Colmegna P, McFadden R, Fabris C, et al. Adaptive biobehavioral control: a pilot analysis of human-machine coadaptation in type 1 diabetes. Diabetes Technol Ther. 2024;26(9):644-651. [DOI] [PubMed] [Google Scholar]
  • 124. Sobhani NC, Goemans S, Nguyen A, et al. Continuous glucose monitoring in pregnancies with type 1 diabetes: small increases in time-in-range improve maternal and perinatal outcomes. Am J Obstet Gynecol. 2024;231(4):467.e1-467. [DOI] [PubMed] [Google Scholar]
  • 125. Lee TTM, Collett C, Bergford S, et al. Automated insulin delivery in women with pregnancy complicated by type 1 diabetes. N Engl J Med. 2023;389(17):1566-1578. [DOI] [PubMed] [Google Scholar]
  • 126. Ozaslan B, Deshpande S, Doyle FJ, 3rd, Dassau E. Zone-MPC automated insulin delivery algorithm tuned for pregnancy complicated by type 1 diabetes. Front Endocrinol (Lausanne). 2021;12:768639. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society

RESOURCES