Skip to main content
. 2025 Jul 1;15:21521. doi: 10.1038/s41598-025-05780-w

Table 5.

Matrix effect test results in plasma and whole blood.

Plasma
Analyte Lot ID LQC HQC
Nominal Conc. (nM) n Mean Calculated Conc. (nM) Accuracy (%) CV (%) Nominal Conc. (nM) n Mean Calculated Conc. (nM) Accuracy (%) CV (%)
PPIX 1 3.00 3 2.86 95.3 5.4 1500 3 1600 106.7 2.4
2 3 2.79 92.9 10.9 3 1570 104.7 1.1
3 3 2.93 97.7 4.6 3 1490 99.4 2.7
4 3 2.68 89.2 4.8 3 1550 103.2 1.5
5 3 2.92 97.3 5.1 3 1500 100.1 0.8
6 3 2.76 92.1 10.0 3 1530 102.2 2.0
Whole blood
Analyte Lot ID Endogenous Conc. (nM) LQC HQC
Nominal Conc. (nM) n Mean Calculated Conc. (nM) Accuracy (%) CV (%) Nominal Conc. (nM) n Mean Calculated Conc. (nM) Accuracy (%) CV (%)
PPIX 1 26.7 56.7 3 51.8 91.3 0.59 1530 3 1530 100 0.76
2 77.7 108 3 99.2 92.1 1.9 1580 3 1540 97.5 3.6
3 46.0 76.0 3 74.1 97.5 1.5 1550 3 1440 92.9 2.0
4 247 277 3 275 99.3 1.6 1750 3 1660 94.9 0.29
5 67.6 97.6 3 98.3 101 2.4 1570 3 1520 96.8 0.95
6 84.8 115 3 109 95.0 0.57 1580 3 1460 92.4 1.7