Table 1.

Recommended imaging parameters for the seven-color program

Scan unit manufacturer and model Leica STELLARIS 8;
Leica SP8
Details
Acquisition Model stage Using the stage model
Sequential Scan Between stacks, Sequential setting provided in Figure 2
xy Format 512∗512 for 20X;
1024∗1024 for 40X
Speed 400 Hz
Scan directionality bidirectional
Zoom factor 1
Line average 2 for 20X; 3 for 40X
Pinhole 1 A.U.
z-stack z-stack size 1.5 μm
Stage Filed size According to experimental requirements
Tile Scan Stitching Overlap x: 10%, y: 10%
Accessories laser 405 nm; WLL2 (440 nm–790 nm)
Objective 20X (NA 0.75, Air); 40X (NA 1.3, oil)
Detector Power HyD S