Abstract
Background
This study aims to evaluate the stress distribution in peri-implant bone, implant surfaces, and within the framework material under vertical and oblique loads, considering various angulations and abutment types of bendable one-piece implants in the All-on-4 concept.
Methods
A three-dimensional model of an edentulous mandible was constructed. Six different configurations were modeled according to prosthesis connection type and inclination of the posterior implants based on the All-on-4 protocol. Four one-piece implants were placed in each model and a cobalt-chromium frameworks with 14 mm cantilever lengths were designed. Under vertical and oblique loading conditions, maximum principal stress and minimum principal stress values were obtained for cortical and trabecular bone and von Mises stress values were calculated for implants and frameworks.
Results
Implant angulation, connection type, and loading conditions affected stress distribution in the peri-implant bone, implants, and frameworks. Higher stress values were observed in posterior implants with increased angulation. Screw-retained models demonstrated greater stress values compared to cement-retained ones under both loading conditions.
Conclusions
For optimal stress distribution in peri-implant bone and frameworks, cement-retained prosthesis may be preferred, especially in cases with increased implant angulation in one-piece bendable systems. Additionally, careful consideration of posterior implant angulation and loading conditions is crucial to minimize the excessive stress and to enhance long-term stability.
Clinical trial number
Not applicable.
Keywords: Bendable implant, One-piece implant, Finite element analysis, Stress distribution, All-on-4, Angulated implant
Background
The primary objective of contemporary dentistry is to address the functional, aesthetic, and psychological needs of those who have experienced tooth loss for a variety of reasons [1]. Dental implants have emerged as a crucial component of prosthodontic treatment for patients with partial or complete edentulism. It is essential to comprehend the distribution of stress that arises during the phases of mastication and functioning, as the occlusal forces exerted on the prosthesis are transmitted to the adjacent bone. Excessive stress loading can result in the deterioration of osseointegration and subsequent failure of the implant [2].
Unfavorable implant inclination can lead to biomechanical problems, causing harmful stresses on the implants, resulting in typical failures such as screw loosening, abutment fracture, and porcelain veneer chipping [3, 4]. In order to prevent clinical complications arising from the improper placement of implants in two-piece implant systems, a number of dental implant manufacturers offer pre-angled abutments [5]. In some cases, dental implants may be intentionally tilted due to factors such as reduced bone volume, maxillary sinus pneumatization, or posterior bone resorption in the mandible after posterior tooth extraction, making implant placement challenging [6]. Recent research suggests that posterior dental implant tilting up to 45 degrees can be a viable option to bone grafts and sinus augmentation, allowing for less involvement with the maxillary sinuses [7–9].
The All-on-4 design eliminates the need for complex surgical interventions in patients who are entirely missing teeth and have atrophic jaws, using four implants: two axial anterior and two tilted posteriors. These implants are carefully positioned to pass through the sinus floor and make the most of the patient’s remaining bone [10]. By increasing the anteroposterior spread, the All-on-4 concept reduces the distal cantilever length and improves cortical anchorage by permitting the implantation of longer dental implants that are optimally positioned to provide prosthetic support [11]. These clinical situations might be handled with pre-angled abutments in two-piece implant systems. On the other hand, clinicians may face difficulties when using single-piece dental implant systems in the aforementioned cases especially when mounting the angled abutments into the most distal implants.
Single-piece implants offer several advantages, including the elimination of gaps between the implant body and the abutment, preventing loosening of abutments or prosthetic screws, which are common in two-piece systems [12, 13]. Additionally, they prevent epithelial damage and bone resorption while saving time by eliminating the need to remove or adjust gingival formers and abutments during prosthetic procedures [14]. The macro design of the implant neck also reduces the risk of peri-implantitis [15], and the surgical and prosthetic process is faster compared to two-piece systems [16, 17].
In single-piece implants requiring angulation, only solution can be bending the abutments if the system allows the clinicians. However, to the best of authors’ knowledge, no biomechanical studies have yet evaluated the stress distribution in screw-retained and cement-retained restorations using bendable single-piece implant systems. This study aims to use finite element analysis (FEM) to assess stress distribution in screw-retained and cement-retained bendable single-piece implant systems subjected to vertical and oblique forces, utilizing four implant-supported fixed prosthesis. The null hypotheses of this study were that the connection type and the angulation of the posterior implant would have no effect on the von Mises stress distribution in implants, peri-implant bone, or the framework.
Methods
Model construction and ethical approval
A three-dimensional (3D) model of a completely edentulous mandible was developed using high-resolution computed tomography (CT) data obtained from the publicly available Visible Human Project, which provides anatomically detailed datasets suitable for biomechanical simulations. The study protocol was approved by the Istanbul Yeni Yuzyil University Science and Health Sciences Research Ethics Committee (No. 2022/05-860).
Mandibular geometry acquisition and processing
CT images were reconstructed with a slice thickness of 0.1 mm and imported into 3D Slicer (Kitware, Clifton Park, NY, USA) in Digital Imaging and Communications in Medicine (DICOM) format. This open-source software was employed for image segmentation and 3D model generation. Semi-automatic segmentation was performed using appropriate Hounsfield Unit (HU) thresholds to isolate the mandibular bone, followed by manual refinement and region-growing techniques to ensure anatomical accuracy, especially near cortical-trabecular boundaries and critical landmarks such as the mental foramen and mandibular canal.
The 3D surface model was generated using the marching cubes algorithm and smoothed with Laplacian filters to eliminate voxel artifacts while preserving essential morphological features related to stress transmission.
Further refinement was conducted in ANSYS SpaceClaim (Ansys Inc., Canonsburg, PA, USA). A 2-mm cortical bone shell was created by offsetting the outer surface of the mandible, and the trabecular bone was modeled by referencing the internal geometry. Mucosal thickness was defined based on the gingival height of the abutments. Implants were placed unicortically, with all regions except the neck embedded in the trabecular bone. All components were spatially aligned and positioned in the appropriate coordinates in 3D space within ANSYS SpaceClaim, ensuring an anatomically faithful representation for subsequent biomechanical analyses.
Prosthetic modeling and experimental design
The mandible and prosthesis models were combined with computational models of the implants and prosthetic components to represent an implant-supported mandibular fixed complete prosthesis. Six different test models based on the All-on-4 protocol (Table 1) were created according to the retention type of the prosthesis and the inclination of the posterior implants. Four one-piece implants, based on the dimensions of Mode Provo implants (Mode Implant / Mode Medikal San. Tic. Ltd. Şti., İstanbul, Türkiye), were placed in each model, and 3D computer-aided design (CAD) models were obtained from the manufacturer.
Table 1.
Study models with different posterior implant inclinations and prosthesis retention types
| Model | Implant location and inclination | Retention type |
|---|---|---|
| Model 1 |
Anterior: Lateral incisor region, parallel Posterior: 2nd premolar region, 17° distal inclination |
Screw-retained |
| Model 2 |
Anterior: Lateral incisor region, parallel Posterior: 2nd premolar region, 17° distal inclination |
Cement-retained |
| Model 3 |
Anterior: Lateral incisor region, parallel Posterior: 2nd premolar region, 30° distal inclination |
Screw-retained |
| Model 4 |
Anterior: Lateral incisor region, parallel Posterior: 2nd premolar region, 30° distal inclination |
Cement-retained |
| Model 5 |
Anterior: Lateral incisor region, parallel Posterior: 2nd premolar region, 45° distal inclination |
Screw-retained |
| Model 6 |
Anterior: Lateral incisor region, parallel Posterior: 2nd premolar region, 45° distal inclination |
Cement-retained |
In the screw-retained test groups (models 1, 3, and 5), data from the Mode Provo-S implants were used (Fig. 1a), while in the cement-retained test groups (models 2, 4, and 6), data from the Mode Provo-C implants were modeled (Fig. 1b). In all models, the two anterior implants (3.5 mm in diameter and 12 mm in length) were positioned vertically at the lateral incisor sites. The posterior one-piece implants (3.5 mm in diameter and 15 mm in length) were modeled as bent at the neck, with inclinations of 17° (models 1 and 2), 30° (models 3 and 4), 45° (models 5 and 6) and placed in the second premolar regions (Fig. 2).
Fig. 1.
(A) Representation of the screw-retained bendable one-piece implant model, including the implant, prosthetic framework, and screw-retention components. (B) Representation of the cement-retained bendable one-piece implant model, illustrating the implant, prosthetic framework, and cement layer used for retention
Fig. 2.
The inclination of posterior implants varies across models as 17°, 30°, and 45°
In cement-retained groups, a 20-µm polycarboxylate cement layer was modeled at the superior region of the implant, conforming to the abutment surface. Cobalt-chromium frameworks incorporating a 14-mm distal cantilever were designed for all models. All parts were meshed and assembled in ANSYS SpaceClaim, and mesh compatibility was ensured in ANSYS Workbench (Table 2).
Table 2.
Quantitative model information
| Materials | Elastic Modulus (MPa) | Poisson Coefficient (ν) |
|---|---|---|
| Cortical Bone | 13,700 | 0.3 |
| Trabecular Bone | 1370 | 0.3 |
| Feldspathic Porcelain | 82,800 | 0.35 |
| Cobalt-Chrome (CoCr) | 218,000 | 0.33 |
| Titanium | 110,000 | 0.3 |
| Cement | 5110 | 0.35 |
Mesh generation and convergence validation
A volumetric finite element mesh was constructed using 4-node tetrahedral (TET4) elements to accurately represent the complex anatomical geometry of the mandible and implant-supported prosthetic system. Element sizes ranged from 0.1 to 0.5 mm for cortical and trabecular bone, and from 0.1 to 0.25 mm for implants, abutments, and metallic superstructures. Local mesh refinement was applied in regions with high stress gradients, particularly around the implant–abutment interfaces. On average, the meshed models contained approximately 836,596 nodes and 3,479,223 elements.
To ensure solution accuracy without excessive computational demand, a mesh convergence analysis was conducted on a representative submodel. Key outcome measures included the maximum von Mises stress in implants, principal stresses in peri-implant bone, and reaction forces at boundary nodes. The mesh was refined iteratively, and the relative error between successive simulations was calculated using the formula:
![]() |
Refinement was continued until the relative error dropped below the predefined threshold of 2–3%, confirming mesh convergence and numerical stability (Tables 3 and 4). Mesh quality was further assessed using geometric criteria such as skewness (> 80°) and minimum edge length (> 0.001 mm); any elements that failed to meet these criteria were corrected accordingly. The final mesh characteristics for each model are summarized in Table 5.
Table 3.
Implant convergence test results
| Mesh Size (mm) | Estimated Stress (MPa) | Relative Error (%) |
|---|---|---|
| 0.5 | 68.96 | - |
| 0.3 | 75.06 | 8.84 |
| 0.2 | 78.63 | 4.75 |
| 0.1 | 80.94 | 2.93 |
Table 4.
Bone (around implant) convergence test results
| Mesh Size (mm) | Estimated Stress (MPa) | Relative Error (%) |
|---|---|---|
| 0.5 | 6.56 | - |
| 0.3 | 7.34 | 11.89 |
| 0.2 | 7.75 | 5.57 |
| 0.1 | 7.97 | 2.80 |
Table 5.
Mesh characteristics of the finite element models
| Total # of Nodes | Total # of Elements | |
|---|---|---|
| Model 1 | 836,596 | 3,479,223 |
| Model 2 | 872,903 | 3,561,709 |
| Model 3 | 818,812 | 3,406,392 |
| Model 4 | 855,404 | 3,490,379 |
| Model 5 | 868,215 | 3,618,125 |
| Model 6 | 904,817 | 3,702,547 |
Implant geometry and discretization
Implant geometries were obtained directly from the manufacturer as CAD files to ensure dimensional accuracy. These models were imported into ANSYS Workbench and discretized using the same meshing strategy. Mesh refinement was applied in curved and high-stress areas to accurately capture load distribution behavior.
Model count and simulation overview
Six finite element models were developed, and two different linear static analyses were performed on each model under distinct loading scenarios, resulting in a total of 12 simulations. Simulations were conducted using LS-DYNA under quasi-static assumptions.
Loading and boundary conditions
Two loading scenarios were defined to simulate functional masticatory forces:
Oblique loading: A 150 N force at a 75° angle in the linguo-buccal direction was applied to the buccal cusps of the 1st premolar, 2nd premolar, and 1st molar teeth (Fig. 3a).
Fig. 3.
Boundary conditions applied in the finite element analysis. (A) Oblique loading scenario, (B) Vertical loading scenario
Vertical loading: A vertical 150 N force was applied to the same regions (Fig. 3b).
To avoid stress singularities, loads were distributed over peripheral nodes within the contact zones. Boundary conditions were assigned to appropriately constrain mandibular movement during load application.
Results
Stresses in peri-implant bone
The Pmax and Pmin stress distributions in the cortical bone are depicted in Figs. 4 and 5. According to the analyses, implant angulation, implant-framework connection type, and loading conditions were found to be effective on the stress distributions in the peri-implant bone. Both Pmax and Pmin stresses in anterior and posterior peri-implant bone increased with the increase in the angulation degree of posterior implant. In the anterior implant region, vertical forces produced slightly lower Pmax and Pmin values in comparison to oblique forces, whereas vertical loading composed higher stress values in the posterior implant region. Stress values were lower in cement-retained models (2, 4, and 6) compared to their screw-retained counterparts (1, 3, and 5).
Fig. 4.
Pmax maps in the cortical bone under oblique (top row) and vertical (bottom row) loading conditions. The images are unprocessed except for merging multiple simulation outputs for layout clarity. No changes to brightness, contrast, or color have been applied. The color scales and data remain as originally generated from the finite element analysis
Fig. 5.
Pmin maps in the cortical bone under oblique (top row) and vertical (bottom row) loading conditions. The images are unprocessed except for merging multiple simulation outputs for layout clarity. No changes to brightness, contrast, or color have been applied. The color scales and data remain as originally generated from the finite element analysis
Maximum principal stresses
Regardless of the loading conditions, the highest Pmax values were observed in the anterior and posterior implant regions of Model 5 (45° angulation, screw-retained); while the lowest values were observed in both regions of Model 2 (17° angulation, cement-retained). In all models, highest Pmax values were concentrated around the mesial and buccal aspects of the implant necks.
The highest Pmax values determined under oblique and vertical loadings were in the anterior implant region of Model 5 with the values of 14.483 MPa and 13.432 MPa, respectively. The lowest Pmax values, recorded in the posterior implant region of Model 2 were 0.243 MPa under oblique loading and 0.278 MPa under vertical loading (Fig. 6).
Fig. 6.
Distribution of Pmax values in the cortical bone under oblique and vertical loading conditions for all models
Minimum principal stresses
The highest Pmin values were identified in posterior implant region of Model 5 whereas the lowest values were found in Model 2, regardless of the loading direction. According to stress maps, the highest compressive stresses were determined on the distal aspects of the cortical bone around the posterior implants.
Among the models analyzed, the posterior implant region of Model 5 demonstrated the highest Pmin stresses, with values of − 75.093 MPa under oblique loading and − 79.256 MPa under vertical loading. The lowest Pmin stresses were identified in Model 2 as − 32.598 MPa and − 35.919 MPa during oblique and vertical loadings, respectively (Fig. 7).
Fig. 7.
Distribution of Pmin values in the cortical bone under oblique and vertical loading conditions for all models
Stresses in implants
The angulation of the posterior implants and the connection type influenced the von Mises stresses detected on the implants. Regardless of the loading conditions, higher stress values were determined in posterior implants in comparison to anterior implants. Von Mises stresses increased with the increase of the implant angulation (Fig. 8). Screw-retained models exhibited higher stresses than cemented-connection models in both anterior and posterior implants. Color-coded stress maps revealed that von Mises stresses were concentrated in the neck regions of anterior and posterior implants. In posterior implants, highest stresses were observed in the mesial aspect of the implant neck closer to the abutment while in anterior implants stresses were higher in buccal aspect of the implant neck near the crestal region (Figs. 9 and 10).
Fig. 8.
Comparison of von Mises stress values in the implants under oblique and vertical loading
Fig. 9.
Color-coded stress maps indicating von Mises stress distributions in the implant structures under vertical loading. The images are unprocessed except for merging multiple simulation outputs for layout clarity. No changes to brightness, contrast, or color have been applied. The color scales and data remain as originally generated from the finite element analysis
Fig. 10.
Color-coded stress maps indicating von Mises stress distributions in the implant structures under oblique loading. The images are unprocessed except for merging multiple simulation outputs for layout clarity. No changes to brightness, contrast, or color have been applied. The color scales and data remain as originally generated from the finite element analysis
In anterior implants, oblique forces generated higher stresses while in posterior implants higher stress values were observed under vertical forces. Among the models, for both anterior and posterior implants, the highest Von Mises stress values were determined in Model 5 under both loading conditions. For anterior implants, the highest stress value was recorded as 43.657 MPa under oblique loading and for posterior implants the highest value was detected as 108.568 MPa under vertical loading in Model 5. The lowest von Mises stresses were identified in anterior and posterior implants of Model 2 with the values of 23.396 MPa and 72.794 MPa, respectively (Fig. 8).
Stresses in framework
Connection type significantly affected the stress values determined in frameworks. Under both oblique and vertical loading conditions, the von Mises stresses observed in the frameworks of screw-retained models were more than 3.5 times higher than those in cement-retained models. The increase in the angulation of posterior implants resulted in a slight decrease in the stress values determined in the framework. The highest stresses were found in Model 1 (17° angulation, screw-retained) with values of 75.297 MPa and 74.827 MPa, under oblique and vertical forces, respectively (Fig. 11). Color maps indicated that stress was concentrated around the posterior implant regions in all models. More uniform stress distribution was observed in cement-retained models.
Fig. 11.
Distribution of von Mises stress values in the prosthetic framework under oblique and vertical loading
Discussion
This study evaluated the effects of different implant inclinations and implant-abutment connections on the biomechanical behavior of bendable single-piece implants and stress distribution across the peri-implant bone and prosthetic framework. The null hypotheses of the study was rejected since the findings of the finite element analysis revealed that both connection type and the implant angulation had significant effects on stress distribution.
All-on-4 treatment concept has been established as a reliable option in the treatment of edentulous arches with fixed restorations according to the long-term clinical outcomes [18–21]. Angulated implants allow the placement of longer implants in the posterior region, which increases the contact area between bone and implant and also improves implant stability [22]. Furthermore, tilting the posterior implant may eliminate the necessity for grafting procedures and reduce the cantilever length [23].
To fabricate full arch prosthetic restorations, the angulation of the tilted implants can be adjusted using various solutions, such as angled multi-unit abutments, angulated screw channels, or built-in angle corrected implants [3, 22]. These approaches endeavor to optimize prosthetic design while addressing the clinical limitations. However, it has been reported that the use of angulated abutments might result in technical complications, especially screw loosening [3]. In single-piece bendable Provo implants, the correction of the implant angulation is achieved without the need for additional components reducing the risk of potential mechanical failures. Single-piece implants have advantages over two-piece implant systems including minimizing micromotion and consequently reducing inflammatory responses in the surrounding tissues [24]. Furthermore, finite element analyses have reported that single-piece implants exhibit lower stress concentrations [25, 26] and allow for more uniform stress distribution along the implant body in All-on-4 concept [27] compared to two-piece implants. However, studies on single-piece bendable remain limited. Therefore, the current study aimed to investigate the stress distribution in implants, peri-implant bone, and framework when single-piece bendable implants are connected to the framework using either screw or cement.
In the present study, the cantilever length and implant lengths were standardized in all models, regardless of the angulation of the posterior implants to eliminate variables related to the lever arm effect and ensure that the detected differences in stress distributions could be attributed only to the changes in implant angulation and the bending of implant neck. The results of the present study revealed that angulation of the posterior implant significantly affected the stress distribution in peri-implant bone. Increased angulation resulted in higher Pmax and Pmin values indicating that higher implant inclinations compose greater stress concentration in cortical bone that is consistent with previous studies [28, 29] in which cantilever lengths were standardized among the groups. This finding may be attributed to increased bending moments in relation to higher implant angulations since the forces distributed unevenly across the implant and led to localized stress concentrations in the cortical bone.
In cases of total edentulism with multiple implants, screw-retained restorations are usually indicated due to their predictable retrievability, which facilitates hygiene maintenance and possible repairs [30]. Additionally, the elimination of excess cement, which has been reported to be a risk factor for biological complications such as peri-implant diseases, is achieved with screw-retained restorations [31]. On the other hand, cement-retained restorations can achieve a passive fit through the cement layer gap especially when the implants are not placed ideally [32]. Screw-retained and cement-retained restorations exhibit different biomechanical behaviors in full-arch implant supported prosthesis which may affect the stress distribution across the implant, peri-implant bone, and prosthetic components. The results of this study showed that von Mises stresses in screw-retained frameworks were consistently higher than those in cement-retained frameworks under both loading conditions. In vertical loading, screw-retained models exhibited stress values ranging between 71.97 MPa and 74.83 MPa, whereas cement-retained models displayed significantly lower stress values, ranging between 18.97 MPa and 20.05 MPa. Similarly, under oblique loading, von Mises stresses in screw-retained frameworks were between 72.44 MPa and 75.29 MPa, compared to 19.28 MPa to 20.29 MPa in cement-retained frameworks. In screw-retained restorations, the rigid connection between the implant and prosthetic framework often leads to localized stress concentrations, particularly at the implant-abutment interface and peri-implant cortical bone, which may increase the risk of mechanical complications such as screw loosening or implant fatigue. In contrast, cement-retained restorations introduce a degree of elasticity at the connection interface, facilitating a more uniform distribution of occlusal forces. This reduces stress peaks in the peri-implant bone, particularly in the cortical region, potentially lowering the risk of bone resorption [33]. The results of this study are consistent with previous research, suggesting that screw-retained restorations revealed higher stress concentrations in the implants and framework [33–35].
Excessive occlusal loads applied to dental implants can negatively affect the long-term success of implant-supported restorations [36]. When applying FEM to dental implants, it is important to consider not only axial loads and horizontal forces (moment-causing loads) but also a combined load (oblique occlusal force) because the latter represents more realistic occlusal directions and, for a given force, will cause the highest localized stress in cortical bone [37]. The enduring success of implant-supported prostheses heavily hinges on the occlusal loads they are subjected [36, 38].
To simulate average posterior bite forces, previous studies applied oblique and axial loads of 100–150 N on dental implants, particularly those inclined at 30–45 degrees [39–43]. In accordance with these studies, in this study, vertical and oblique (linguo-buccal direction) forces of 150 N were applied to the 1st and 2nd premolars and the 1st molar. Previous research [44–47] has demonstrated that oblique loads induce higher stress concentrations on implants, peri-implant bone, and prosthetic components, significantly altering stress distribution patterns. Consistent with these findings, the present study revealed that oblique loading resulted in higher von Mises stress values in the prosthetic framework and in anterior implants placed vertically. Additionally, Pmax and Pmin values in the cortical and trabecular bone of the anterior region were higher under oblique loading than under vertical loading.
Akça & Iplikcioglu [48] reported that exceeding the von Mises stress value of 550 MPa, which corresponds to the yield strength of a titanium implant, could potentially lead to implant failure. In the current study, von Mises stress values over 550 MPa were not detected on the substructure material and implants under oblique and vertical loads. The highest value obtained in the study was 108.568 MPa observed in posterior implants placed at a 45-degree angle under vertical load.
The reliability of the findings was enhanced through careful standardization during the finite element modeling process. Anatomically accurate mandibular geometry was obtained from CT data, and segmentation of cortical and trabecular bone was performed using appropriate Hounsfield unit thresholds. All materials were defined as homogeneous, isotropic, and linearly elastic. Each model was constructed with a fine tetrahedral mesh exceeding 3 million elements, and bonded contacts were used to simulate fixed interfaces without relative motion. Implant dimensions and cantilever lengths were standardized across all models, allowing stress differences to be attributed solely to implant angulation and connection type. These modeling decisions improve the validity of the observed outcomes, particularly in the comparison of cement- and screw-retained frameworks under increased posterior angulation.
This study has several limitations. First, although FEM provides valuable insights into biomechanical behavior, it cannot fully replicate the complex nature of masticatory forces, which vary in magnitude, direction, and duration in vivo. Moreover, the cantilever length was standardized at 14 mm across all models, which, while allowing for controlled comparisons, may not fully represent the varying clinical scenarios seen in full-arch rehabilitations. Only a single prosthetic material was evaluated in this study, whereas different materials could influence stress distribution and overall biomechanical performance. Further studies evaluating different prosthetic materials, dynamic loading conditions, and long-term clinical outcomes are needed.
Conclusions
Within the limitations the following conclusions can be drawn:
Increasing the angulation of posterior implants results in higher stress values in the cortical and trabecular bone, as well as in anterior and posterior implants, but these values are in the safe range and far below the yield strength of the implants which may cause failure.
Cement-retained frameworks exhibit significantly lower stress values compared to screw-retained frameworks under both oblique and vertical loads.
Higher stress values occur in posterior implants compared to anterior implants, particularly under vertical loading, whereas anterior implants experience greater stress under oblique forces.
Acknowledgements
The authors thank MODE Medikal for their help in providing the data for the various implant types.
Abbreviations
- FEM
Finite element method
- 3D
Three-dimensional
- CT
Computed tomography
- DICOM
Digital imaging and communications in medicine
- CoCr
Cobalt-Chrome
- CAD
Computer-aided design
- HU
Hounsfield Unit
Author contributions
EBO; conceptualization, data curation, formal analysis, investigation, methodology, visualization, writing – original draft, writing – review and editing. GD; writing – original draft, methodology, formal analysis, visualization, investigation. AS; investigation, methodology, writing – original draft. SCD; data curation, investigation, methodology, writing – review and editing, project administration. SKY; data curation, formal analysis, visualization, writing – review and editing, supervision, conceptualization. OG; formal analysis, methodology, project administration, supervision, writing – review and editing, conceptualization.All authors read and approved the final manuscript.
Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Data availability
The datasets used and/or analyzed during the current study are available from the corresponding author upon reasonable request.
Declarations
Ethical approval
The study protocol was approved by the Istanbul Yeni Yuzyil University Science and Health Sciences Research Ethics Committee (No. 2022/05-860).
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Misch CE. Dental implant prosthetics. Elsevier Health Sciences. 2014.
- 2.Anusavice KJ, Shen C, Rawls HR. Phillips’ science of dental materials. Elsevier Health Sciences. 2012.
- 3.Hotinski E, Dudley J. Abutment screw loosening in angulation-correcting implants: an in vitro study. J Prosthet Dent. 2019;121(1):151–5. 10.1016/j.prosdent.2018.03.005 [DOI] [PubMed]
- 4.Poletto-Neto V, Tretto PHW, Zen BM, Bacchi A, Dos Santos MBF. Influence of implant inclination and prosthetic abutment type on the biomechanics of implant-supported fixed partial dentures. J Oral Implantol. 2019;45(5):343–50. 10.1563/aaid-joi-D-18-00305. [DOI] [PubMed] [Google Scholar]
- 5.Cavallaro J Jr, Greenstein G. Angled implant abutments: a practical application of available knowledge. J Am Dent Assoc. 2011;142(2):150–8. 10.14219/jada.archive.2011.0057. [DOI] [PubMed] [Google Scholar]
- 6.Mumcu E, Dayan SC, Genceli E, Geckili O. Comparison of four-implant-retained overdentures and implant-supported fixed prostheses using the All-on-4 concept in the maxilla in terms of patient satisfaction, quality of life, and marginal bone loss: a 2-year retrospective study. Quintessence Int. 2020;51(5):388–96. 10.3290/j.qi.a44368. [DOI] [PubMed] [Google Scholar]
- 7.Ayali A, Altagar M, Ozan O, Kurtulmus-Yilmaz S. Biomechanical comparison of the All-on-4, M-4, and V-4 techniques in an atrophic maxilla: A 3D finite element analysis. Comput Biol Med. 2020;123:103880. 10.1016/j.compbiomed.2020.103880. [DOI] [PubMed] [Google Scholar]
- 8.Mehta SP, Sutariya PV, Pathan MR, Upadhyay HH, Patel SR, Kantharia NDG. Clinical success between Tilted and axial implants in edentulous maxilla: A systematic review and meta-analysis. J Indian Prosthodont Soc. 2021;21(3):217–28. 10.4103/jips.jips_79_21. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Yang Q, Guan X, Wang B, Zhang D, Bai J, Zhang X, Zhou Y. Implant survival rate and marginal bone loss with the all-on-4 immediate-loading strategy: A clinical retrospective study with 1 to 4 years of follow-up. J Prosthet Dent. 2023;130(6):849–57. 10.1016/j.prosdent.2021.12.020. [DOI] [PubMed] [Google Scholar]
- 10.Dayan SC, Geckili O. The influence of framework material on stress distribution in maxillary complete-arch fixed prostheses supported by four dental implants: a three-dimensional finite element analysis. Comput Methods Biomech Biomed Engin. 2021;24(14):1606–17. 10.1080/10255842.2021.1903450. [DOI] [PubMed] [Google Scholar]
- 11.Patzelt SB, Bahat O, Reynolds MA, Strub JR. The all-on‐four treatment concept: a systematic review. Clin Implant Dent Relat Res. 2014;16(6):836–55. 10.1111/cid.12068. [DOI] [PubMed] [Google Scholar]
- 12.Broggini N, McManus LM, Hermann JS, Medina RU, Oates TW, Schenk RK, Cochran DL. Persistent acute inflammation at the implant-abutment interface. J Dent Res. 2003;82(3):232–7. 10.1177/154405910308200316. [DOI] [PubMed] [Google Scholar]
- 13.Elakkiya S, Ramesh AS, Prabhu K. Systematic analysis on the efficacy of bone enhancement methods used for success in dental implants. J Indian Prosthodont Soc. 2017;17(3):219–25. 10.4103/jips.jips_19_17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Finne K, Rompen E, Toljanic J. Clinical evaluation of a prospective multicenter study on 1-piece implants. Part 1: marginal bone level evaluation after 1 year of follow-up. Int J Oral Maxillofac Implants. 2007;22(2):226–34. [PubMed] [Google Scholar]
- 15.Iglhaut G, Schwarz F, Winter RR, Mihatovic I, Stimmelmayr M, Schliephake H. Epithelial attachment and downgrowth on dental implant abutments—a comprehensive review. J Esthet Restor Dent. 2014;26(5):324–31. 10.1111/jerd.12097. [DOI] [PubMed] [Google Scholar]
- 16.Agbaje JO, Diederich H. Minimal invasive concept for the rehabilitation of edentulous jaw with one-piece implants. Int J Case Rep Short Reviews. 2018;4:028–30. [Google Scholar]
- 17.Bollen CM, Papaioanno W, Van Eldere J, Schepers E, Quirynen M, Van Steenberghe D. The influence of abutment surface roughness on plaque accumulation and peri-implant mucositis. Clin Oral Implants Res. 1996;7(3):201–11. 10.1034/j.1600-0501.1996.070302.x. [DOI] [PubMed] [Google Scholar]
- 18.Maló P, de Araújo Nobre M, Lopes A, Ferro A, Botto J. The All-on‐4 treatment concept for the rehabilitation of the completely edentulous mandible: A longitudinal study with 10 to 18 years of follow‐up. Clin Implant Dent Relat Res. 2019;21(4):565–77. 10.1111/cid.12769. [DOI] [PubMed] [Google Scholar]
- 19.Pera P, Menini M, Pesce P, Bevilacqua M, Pera F, Tealdo T. Immediate versus delayed loading of dental implants supporting fixed full-arch maxillary prostheses: a 10-year follow-up report. Int J Prosthodont. 2019;32:27–31. 10.11607/ijp.5804. [DOI] [PubMed] [Google Scholar]
- 20.Grandi T, Signorini L. Rehabilitation of the completely edentulous mandible by all-on-four treatment concept: A retrospective cohort study with up to 10 years follow-up. Med (Kaunas). 2021;58(1):10. 10.3390/medicina58010010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Uesugi T, Shimoo Y, Munakata M, Sato D, Yamaguchi K, Fujimaki M, Malo P. The All-on-four concept for fixed full-arch rehabilitation of the edentulous maxilla and mandible: a longitudinal study in Japanese patients with 3–17-year follow-up and analysis of risk factors for survival rate. Int J Implant Dent. 2023;9(1):43. 10.1186/s40729-023-00511-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Ibrahim CRM, Sameh A, Askar O. A finite element analysis study on different angle correction designs for inclined implants in All-On-Four protocol. BMC Oral Health. 2024;24(1):331. 10.1186/s12903-024-04091-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Wang Q, Zhang ZZ, Bai SZ, Zhang SF. Biomechanical analysis of stress around the Tilted implants with different cantilever lengths in all-on-4 concept. BMC Oral Health. 2022;22(1):469. 10.1186/s12903-022-02520-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Prithviraj DR, Gupta V, Muley N, Sandhu P. One-piece implants: placement timing, surgical technique, loading protocol, and marginal bone loss. J Prosthodontics: Implant Esthetic Reconstr Dentistry. 2013;22(3):237–44. 10.1111/j.1532-849X.2012.00928.x. [DOI] [PubMed] [Google Scholar]
- 25.Cavit Çehreli M, Akça K, İplikçioğlu H. Force transmission of one-and two‐piece morse‐taper oral implants: a nonlinear finite element analysis. Clin Oral Implants Res. 2004;15(4):481–9. 10.1111/j.1600-0501.2004.01025.x. [DOI] [PubMed] [Google Scholar]
- 26.Moustafa SM, Elattar MS, Ahmed TM. Comparison of stresses transmitted to one-piece and two-piece narrow-diameter implants in mandibular over dentures (a finite element stress analysis). Alexandria Dent J. 2018;43(2):51–6. [Google Scholar]
- 27.Mohamed AMA, Askar MG, El Homossany MEMB. Stresses induced by one-piece and two-piece dental implants in All-on-4® implant supported prosthesis under simulated lateral occlusal loading: nonlinear finite element analysis study. BMC Oral Health. 2022;22(1):196. 10.1186/s12903-022-02228-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Sannino G. All-on-4 concept: a 3-dimensional finite element analysis. J Oral Implantol. 2015;41(2):163–71. 10.1563/AAID-JOI-D-12-00312. [DOI] [PubMed] [Google Scholar]
- 29.Tribst JPM, Campanelli de Morais D, Melo de Matos JD, Lopes GDRS, Dal Piva AMDO, Souto Borges AL, Ausiello P. Influence of framework material and posterior implant angulation in full-arch all-on-4 implant-supported prosthesis stress concentration. Dent J (Basel). 2022;10(1):12. 10.3390/dj10010012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Shadid R, Sadaqa N. A comparison between screw-and cement-retained implant prostheses. A literature review. J Oral Implantol. 2012;38(3):298–307. 10.1563/AAID-JOI-D-10-00146. [DOI] [PubMed] [Google Scholar]
- 31.Staubli N, Walter C, Schmidt JC, Weiger R, Zitzmann NU. Excess cement and the risk of peri-implant disease–a systematic review. Clin Oral Implants Res. 2017;28(10):1278–90. 10.1111/clr.12954. [DOI] [PubMed] [Google Scholar]
- 32.Aglietta M, Siciliano VI, Zwahlen M, Brägger U, Pjetursson BE, Lang NP, Salvi GE. A systematic review of the survival and complication rates of implant supported fixed dental prostheses with cantilever extensions after an observation period of at least 5 years. Clin Oral Implants Res. 2009;20(5):441–51. 10.1111/j.1600-0501.2009.01706.x. [DOI] [PubMed] [Google Scholar]
- 33.Silva GC, Cornacchia TM, de Magalhães CS, Bueno AC, Moreira AN. Biomechanical evaluation of screw-and cement-retained implant-supported prostheses: a nonlinear finite element analysis. J Prosthet Dent. 2014;112(6):1479–88. 10.1016/j.prosdent.2014.06.010. [DOI] [PubMed] [Google Scholar]
- 34.Pellizzer EP, Tonella BP, Ferraço R, Falcón-Antenucci RM, de Carvalho PSP, Alves-Rezende MCR. Photoelastic stress analysis in screwed and cemented implant-supported dentures with external hexagon implants. J Craniofac Surg. 2010;21(4):1110–3. 10.1097/SCS.0b013e3181e1b46e. [DOI] [PubMed] [Google Scholar]
- 35.de Aguiar FA Jr, Tiossi R, Macedo AP, de Mattos MDGC, Ribeiro RF, Rodrigues RCS. Photoelastic analysis of stresses transmitted by universal cast to long abutment on implant-supported single restorations under static occlusal loads. J Craniofac Surg. 2012;23(7):77–81. 10.1097/SCS.0b013e3182597c09. [DOI] [PubMed] [Google Scholar]
- 36.Bal BT, Çağlar A, Aydın C, Yılmaz H, Bankoğlu M, Eser A. Finite element analysis of stress distribution with splinted and nonsplinted maxillary anterior fixed prostheses supported by zirconia or titanium implants. Int J Oral Maxillofac Implants. 2013;28(1):27–38. 10.11607/jomi.2442. [DOI] [PubMed] [Google Scholar]
- 37.Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: a review of the literature. J Prosthet Dent. 2001;85(6):585–98. 10.1067/mpr.2001.115251. [DOI] [PubMed] [Google Scholar]
- 38.Torcato LB, Pellizzer EP, Verri FR, Falcón-Antenucci RM, Júnior JFS, de Faria Almeida DA. Influence of parafunctional loading and prosthetic connection on stress distribution: a 3D finite element analysis. J Prosthet Dent. 2015;114(5):644–51. 10.1016/j.prosdent.2015.04.018. [DOI] [PubMed] [Google Scholar]
- 39.Faverani LP, Barão VAR, Ramalho-Ferreira G, Delben JA, Ferreira MB, Júnior IRG, Assunção WG. The influence of bone quality on the Biomechanical behavior of full-arch implant-supported fixed prostheses. Mater Sci Engineering: C. 2014;37:164–70. 10.1016/j.msec.2014.01.013. [DOI] [PubMed] [Google Scholar]
- 40.Ozdemir Doğan D, Polat NT, Polat S, Şeker E, Gül EB. Evaluation of All-on‐Four concept and alternative designs with 3D finite element analysis method. Clin Implant Dent Relat Res. 2014;16(4):501–10. 10.1111/cid.12024. [DOI] [PubMed] [Google Scholar]
- 41.Almeida EO, Rocha EP, Júnior ACF, Anchieta RB, Poveda R, Gupta N, Coelho PG. Tilted and short implants supporting fixed prosthesis in an atrophic maxilla: a 3D-FEA Biomechanical evaluation. Clin Implant Dent Relat Res. 2015;17:332–42. 10.1111/cid.12129. [DOI] [PubMed] [Google Scholar]
- 42.Bhering CLB, Mesquita MF, Kemmoku DT, Noritomi PY, Consani RLX, Barão VAR. Comparison between all-on-four and all-on-six treatment concepts and framework material on stress distribution in atrophic maxilla: A prototyping guided 3D-FEA study. Mater Sci Eng C Mater Biol Appl. 2016;69:715–25. 10.1016/j.msec.2016.07.059. [DOI] [PubMed] [Google Scholar]
- 43.Ozan O, Kurtulmus-Yilmaz S. Biomechanical comparison of different implant inclinations and cantilever lengths in All-on-4 treatment concept by Three-Dimensional finite element analysis. Int J Oral Maxillofac Implants. 2018;33(1):64–71. 10.11607/jomi.6201. [DOI] [PubMed] [Google Scholar]
- 44.Chang HS, Chen YC, Hsieh YD, Hsu ML. Stress distribution of two commercial dental implant systems: A three-dimensional finite element analysis. J Dent Sci. 2013;8(3):261–71. 10.1016/j.jds.2012.04.006. [Google Scholar]
- 45.Takahashi JMFK, Dayrell AC, Consani RLX, de Arruda Nóbilo MA, Henriques GEP, Mesquita MF. Stress evaluation of implant-abutment connections under different loading conditions: a 3D finite element study. J Oral Implantol. 2015;41(2):133–7. 10.1563/AAID-JOI-D-11-00205. [DOI] [PubMed] [Google Scholar]
- 46.Kaleli N, Sarac D, Külünk S, Öztürk Ö. Effect of different restorative crown and customized abutment materials on stress distribution in single implants and peripheral bone: A three-dimensional finite element analysis study. J Prosthet Dent. 2018;119(3):437–45. 10.1016/j.prosdent.2017.03.008. [DOI] [PubMed] [Google Scholar]
- 47.Lee JH, Jang HY, Lee SY. Finite element analysis of dental implants with zirconia crown restorations: conventional cement-retained vs. cementless screw-retained. Mater (Basel). 2021;14(10):2666. 10.3390/ma14102666. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Akça K, Iplikçioğlu H. Finite element stress analysis of the effect of short implant usage in place of cantilever extensions in mandibular posterior edentulism. J Oral Rehabil. 2002;29(4):350–6. 10.1046/j.1365-2842.2002.00872.x. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets used and/or analyzed during the current study are available from the corresponding author upon reasonable request.