Abstract

The loss of a limb or paralysis resulting from spinal cord injury has devastating consequences on quality of life. One approach to restoring lost sensory and motor abilities in amputees and patients with tetraplegia is to supply them with implants that provide a direct interface with the CNS. Such brain–machine interfaces might enable a patient to exert voluntary control over a prosthetic or robotic limb or over the electrically induced contractions of paralysed muscles. A parallel interface could convey sensory information about the consequences of these movements back to the patient. Recent developments in the algorithms that decode motor intention from neuronal activity and in approaches to convey sensory feedback by electrically stimulating neurons, using biomimetic and adaptation-based approaches, have shown the promise of invasive interfaces with sensorimotor cortices, although substantial challenges remain.


A study published nearly 15 years ago described how rats learned to operate, using signals recorded from two 16-electrode arrays implanted in the primary motor cortex (M1) and thalamus, a simple, single-degree of freedom (DOF) robotic arm to retrieve a water reward1. Since this pioneering study, the capabilities of brain–machine interfaces (BMIs) have been considerably extended2–8. Advances in microelectrode, electronics and computing technologies have made it possible to construct electronic interfaces with the brain that enable us to read out and ‘decode’ hundreds of neural signals from motor areas of the brain in order to control increasingly sophisticated robotic arms. Recently, an individual with tetraplegia was able to operate seven of the DOFs of a complex anthropomorphic arm and hand using inputs from two 96-channel intracortical microelectrode arrays implanted in M1 (REF. 9). In this Review, we explore the advances in intracortical BMIs over the 15 years that have elapsed since the first experiments by discussing studies in rodents, monkeys and humans that have led to the tremendous increase in function.

Current BMIs have a number of limitations. First among them is the fact that, with few exceptions, they involve only efferent neural interfaces (BOX 1; FIG. 1); that is, they decode movement-related information from measured brain activity. Virtually all BMIs are guided only by visual feedback, lacking the somatosensory feedback that is a critical component of normal motor behaviour10. The absence of somatosensory feedback may contribute substantially to the relatively poor quality of BMI-controlled movements and to the difficulty with which they are produced. Afferent interfaces are being developed to restore a sense of touch and limb movement to patients through trains of electrical stimulation delivered to electrodes in the somatosensory cortex11–20 (FIG. 1), but they are still much less advanced than their efferent counterparts. Ultimately, a BMI would pass information in both directions to restore both motor and somatosensory function.

Box 1 |. Decoding movement plans from the brain.

Computing a neural ‘decoder’

Neurons emit action potentials, or ‘spikes,’ which are discrete events whose rate and timing carry information. The process of ‘neural decoding’ typically involves developing a set of (usually linear) functions that convert strings of action potentials, recorded on many electrodes, into a small number of continuously varying signals (usually two or three). These functions can then be used to control the movement of a cursor or robotic limb given any recorded pattern of neuronal activity. Alternatively, a range of machine-learning methods (for example, linear discriminant analysis, support vector machines or Bayesian classifiers) can be used to classify neural recordings into discrete control states that might be used to identify one of several potential targets or to determine whether the hand is open or closed.

A linear ‘decoder’ is essentially a weighted average across the recent firing rate history of the recorded neurons, with different control variables determined by different linear combinations of the inputs. Decoders have also been based on local field potentials38–40,83,133–138, which are continuous signals that are thought to represent the summed synaptic potentials from many hundreds or thousands of neurons near the electrode tip139. Particularly with small numbers of electrodes, these potentials seem to carry nearly as much information as action potentials and have the added benefit that they may be more stable over long periods of time37,136.

Constructing a decoder is a matter of finding the weighting coefficients that optimally transform spike trains into, for example, the intended trajectory of the subject’s hand. A simple approach is to record 5–10 minute samples of neural activity from the limb representation in the primary motor cortex (M1) together with limb movement data and to apply a least squares regression calculation between the neural signals and limb kinematics. A similar process can be used in the case of a paralysed patient if neural activity is measured while the patient imagines making a series of movements that correspond to designated trajectories. Although linear decoders are common and effective, many non-linear decoder models have also been developed140–144.

Given the weighting coefficients that constitute the decoder, one can record neural activity, use the decoder to compute the weighted averages, and generate predictions of the subject’s intended movements. In an able-bodied subject, the accuracy of these predictions can be directly compared to the subject’s actual movements in later, offline analysis. Alternatively, the predictions can be generated in real time and used to control the movement of a cursor or a robotic limb. In this case, the subject becomes part of a feedback loop and is able to correct errors and to adapt to any inaccuracies in the intended limb movement (BOX 2).

Kinematics versus kinetics

Nearly all brain–machine interfaces (BMIs) are designed to predict the position or velocity of hand movements; that is, their ‘kinematics’. However, M1 discharge also encodes the ‘kinetics’ (or ‘dynamics’) of movement and can even be driven under isometric conditions in which there is no movement68,70,145,146. The degree to which M1 encodes movement kinetics or kinematics (or ‘muscles versus movements’) has been the subject of debate for many decades147,148. Now, in the context of BMIs, the discussion has taken on a practical dimension: if M1 responses carry kinetic information, then a kinematic decoder that is designed to predict the movement trajectory of an unloaded limb may behave unpredictably if the user attempts to pick up a heavy object. If the user plans a movement that includes greater muscle activation to compensate for the greater mass, a purely kinematic decoder might result in the equivalent of the error that results when picking up an empty jug of milk that appears to be full. Likewise, lifting the jug with the arm extending across the body requires very different muscle activation than if the arm is to the side of the body. This type of postural dependence leads to different M1 discharge for similar actions149–151. A BMI must therefore be able to work across the full range of possible postures and accommodate both anticipated and unexpected external loads without the need for repeated recalibration. Relatively little work has been devoted to this requirement (although see REFS 80,152–154), but this need to generalize is likely to be of increasing importance as the performance of BMIs improves. There are different considerations when using neural signals from premotor or parietal cortices4,155–158, areas that are thought to represent higher-order movement planning and may be more closely related to movement kinematics.

Figure 1 |. Idealized bidirectional brain–machine interface for prosthetic control.

Figure 1 |

Neural signals from motor-related areas of the brain — for example, the primary motor cortex (M1) — that encode the intended movement (motor intent) are decoded and used to control the movement of the prosthetic limb. Sensors on the prosthetic limb convey information about movements of the limb and any objects with which it comes into contact. The output of these sensors is converted into patterns of electrical stimulation (stimulus pulses), which are delivered to sensory areas of the brain — for example, the primary somatosensory cortex (S1) — via chronically implanted arrays of electrodes.

Second, these efferent interfaces are used almost exclusively to control the ‘kinematics’ of reaching. Much less experimental effort has been devoted to developing efferent decoders that control movement ‘kinetics’: that is, the forces and muscle activity underlying movement (for exceptions, see REFS 2,21–23) (BOX 1). Furthermore, much progress remains to be made towards the control of hand use, object grasping and manipulation. The large area of M1 devoted to controlling the complex mechanics of hand use thus remains largely untapped.

Last, most efferent BMIs rely on decoders that attempt to make use of the natural mapping between brain activity and limb movement24–26 — a process known as biomimicry. Some afferent BMIs use an analogous strategy by attempting to produce naturalistic patterns of neuronal activation in the somatosensory cortex through electrical stimulation13–15,19. However, perfect biomimetic mappings cannot be achieved for either type of interface with the current state of science and technology. An alternative approach is to adopt a systematic but not necessarily biomimetic mapping between neuronal activation and limb state, which the subject must then learn through adaptation, sometimes in combination with simultaneous adaptation of the decoder algorithms12,16,17,27–31 (BOX 2). Ultimately, both biomimicry and adaptation are likely to be important principles in BMI development.

Box 2 |. Adapting the decoder as well as the user.

Adaptation of the user’s efforts while he or she is learning to use a brain–machine interface (BMI) is an important factor in achieving successful control of a prosthetic limb. As discussed in the main text, improvements in BMI performance have been shown to correlate with changes in neuronal discharge that may reflect this learning28,31,159. Adaptation of the decoder algorithms may be equally important. Indeed, under realistic clinical conditions, decoders cannot be based on actual movement, and patients may not tolerate extended periods of data collection for decoder training8,9. As a result, initial decoder estimates may not be very accurate. Decoders may also need to adapt in the face of neuron turnover: that is, the gradual loss of particular neurons and their replacement with others160. If the new neurons carry different information, the characteristics of the old decoder will no longer be appropriate.

Gradient descent algorithms are commonly used to adapt linear filter decoders on the basis of errors in their predictions. The fundamental challenge is to infer the user’s intended movement as an antecedent to computing movement error. If there are explicit targets, the decoder could assume that the subject intended to reach the correct target directly and could then compute the corresponding error5,8,161. This approach has yielded accurate movements at more than twice the speed of fixed decoders, approaching that of the native limb5. When there are no explicit targets, an alternative approach may involve detecting inflections in the predicted trajectory that are indicative of attempted corrective movements162. Determining an optimal learning rate for the algorithm adaptation is a crucial factor that requires online, real-time testing. A rapid update rate that works well offline may lead to poor real-time performance, becoming unstable when coupled with the user’s own adaptation163. The effect might be akin to trying to perfect a golf swing but having to use a different club on each attempt.

Adapting a muscle activity (electromyographic) decoder that might be used for a paralysed patient2,21,81 is less straightforward, as there is no well-defined motor error in this case. One might infer a muscle activation error indirectly, through a position error measured at the end point, but the complex, redundant dynamics of the musculoskeletal system present a formidable computational challenge. One possibility to overcome these problems may be to use a reinforcement learning algorithm that does not require explicit error information. The basic feasibility of such an approach has been demonstrated in a pair of studies using an actor–critic architecture164,165, in which the ‘actor’ — that is, movement generation — was controlled by the primary motor cortex discharge, and the ‘critic’ was determined by either task success or a proxy: namely, the activity recorded from the nucleus accumbens.

Owing to space restrictions, we focus in this Review on approaches that involve implantation of intracortical electrode arrays rather than on less-invasive approaches that use epidural, subdural or scalp recordings. For a discussion of the relative merits of invasive and non-invasive approaches, see REF. 32. However, although intracortical approaches offer great potential for the restoration of sensorimotor function, they face considerable challenges (BOX 3). Perhaps the most crucial challenge is that current intracortical electrode technologies do not have the lifespan required for a clinically viable efferent interface33. The quality of signals obtained with less-invasive approaches tends to be maintained over longer periods of time given the larger, lower-impedance electrodes that sample the aggregate activity from many thousands of neurons, but they have poorer spatial resolution and yield fewer independent signals34–36. The recording longevity of intracortical BMIs might be increased by recording local field potentials (rather than single-neuron activity)37–39 as well as by recording multi-unit rather than single-unit activity40,41. The lifespan of existing electrode technologies for stimulation is likely to be significantly longer than that for recording, in the sense that stimulation is less affected by many of the factors that cause loss of recording quality (BOX 3). A further major challenge is that BMIs should eventually be available to the patient in his or her home for 24 hours a day rather than only in the laboratory — a goal that is closer to reality for the fully non-invasive approaches42.

Box 3 |. Reliability, safety and usability of neural interfaces.

Reliability

An important concern is whether implanted electrode arrays remain both effective and sufficiently robust for decades in the brain, especially considering the risks associated with their surgical implantation. Unfortunately, the current state of the art in chronically implanted electrode arrays has not yet achieved such long-term clinical viability166. First, neural signals recorded from individual electrodes exhibit variation over days, and even on any given day. These variations can have a substantial impact on the performance of decoders167. There is also a progressive decay of neuronal signals over months or tens of months that has been attributed to various biological reactions to the implanted arrays, including the death of neurons around the electrode tips, inflammation, disruption in the blood–brain barrier and gliosis166,168,169. Some evidence suggests that chronic intracortical microstimulation (ICMS) at clinically relevant levels can accelerate neuronal loss170, but it is unclear whether this loss has any functional consequences171,172. The electrodes themselves may also degrade over time, as shown by a progressive decay in their impedance169,173. The choice of electrode material is particularly important for the longevity of electrode arrays in stimulation applications174. Finally, there have been some incidences of outright failure of electrode arrays, most often due to connector issues33,175.

The next generation of interfaces must be fabricated with materials that are more mechanically and biologically compatible with the neural tissue so as to minimize the tissue reaction and increase the longevity of the measured signals168,176,177. Existing electrodes are at least one million times stiffer than the surrounding brain tissue — a mechanical mismatch that leads to chronic shear forces and micro-motion, both of which are thought to play a significant part in the inflammatory response to implantation178,179. To address the impedance mismatch, more compliant electrodes, and methods to insert them, are currently being developed180–184. Surface chemistry is also thought to have an important role in biological compatibility by modulating the immune response. The combined goals of maintaining or increasing electrical conductivity while reducing glial formation and promoting neurofilament growth present considerable challenges for the future development of implantable electrode arrays185–191. Carbon nanotubes are an interesting option that may provide the means to address both the mechanical and surface chemistry issues192.

Safety and usability

Another important consideration is whether the implants pose a risk to the patient beyond that associated with the surgical procedure. Although the implants themselves cause some tissue damage, they seem to have little effect on sensorimotor function and seldom cause serious infections. The relative safety of the increasingly common and clinically beneficial deep brain stimulation (DBS)193 — a much more invasive procedure — suggests that neuronal damage will not be a substantial obstacle to intracortical neural interfaces. However, unlike existing DBS implants, the use of transcranial and percutaneous components in current BMI implants are a cause for concern given their potential to cause infection as well as for cosmetic reasons. These concerns have spurred the development of wireless implants194–196, with the extracranial component to be implanted in the chest or another more surgically accessible and less conspicuous region of the body. With implantation comes the challenge of providing adequate power to the devices as well as reducing power dissipation and preventing heating of the device, particularly for components in close proximity to brain tissue197.

A related challenge is the transition of brain–machine interfaces from experiments in the laboratory to the home. Indeed, most existing devices are run on desktop computers and require constant monitoring by the experimenters. Many wires run from the implant to the computer and from the computer to the prosthesis, and the whole system is unwieldy and delicate. As implants become more reliable and the decoding and encoding algorithms improve, the importance of miniaturization and usability will come to the fore.

Controlling movement through an efferent BMI

Control of reach.

Movements controlled by BMIs progressed within a short period from the one-DOF robotic arm used by rats to collect water droplets1 to two-dimensional24 and three-dimensional26 control of cursor movement by monkeys. Each of these three seminal studies used a decoder that was based on data recorded during natural movement to attempt to mimic the existing mapping from M1 to movement (BOX 1). In an amputee or paralysed patient, who cannot generate actual movement, the neuronal discharge can be measured while he or she observes and attempts to mimic the movement of a robotic arm as it makes a series of preprogrammed movements8,9. Together with the cursor trajectories, this neuronal activity is then used to compute an ‘observation-based’ decoder43,44. Indeed, experiments with monkeys have shown that many M1 neurons discharge when a monkey is observing (and presumably attempting to imitate) hand movements represented by the moving cursor in a manner that is at least qualitatively similar to the way they discharge when the monkey is making the actual limb movements45. This observation-based discharge may in part reflect the activity of ‘mirror neurons’ — neurons located predominantly in the ventral premotor cortex (PMv; area F5) and the inferior parietal lobule that discharge both during particular movements and during observation of those movements performed by other individuals46. However, neurons with mirror-like properties account for less than 15% of neurons in the premotor cortex47, and presumably even less in M1, so it is unlikely that they make a major contribution to the decoding.

Observation-based decoders are increasingly being used in experiments with non-paralysed monkeys in an effort to mimic the conditions that would be encountered by a paralysed patient. However, the practice may not be ideal. Abolishing limb movement better mimics the clinical condition in that it eliminates (or at least holds constant) the substantial sensory input that M1 receives from the moving limb48,49; however, for a non-paralysed monkey to imagine moving without actually moving involves either a dramatic change in its motor intent or a downstream mechanism to gate off the cortical output. As there is little evidence that such gating occurs (although see REF. 50), a non-paralysed monkey that imagines — but suppresses — limb movements may not be a very good model of a paralysed patient who is actually attempting to move. Indeed, behaviour-related changes in afferent input or motor intent may explain some of the changes in neural discharge (described below) that occur when a non-paralysed subject transitions from actual limb movement to brain control of a robotic arm26,28.

Arguably the best clinical demonstration of BMI control of a robotic limb achieved so far was by a 52-year-old individual with tetraplegia resulting from spinocerebellar degeneration9 (FIG. 2a–c). The patient was implanted with two microelectrode arrays in the motor cortex with which she controlled an anthropomorphic robotic arm51 (using the hand pictured in FIG. 2d). The decoders were initially recomputed daily in a two-step sequence. Neural recordings were made while the patient observed a series of limb movements, which were used to compute an observation-based decoder. She then used this initial decoder to move the robotic arm herself, thus generating the data necessary to compute a more accurate movement-based decoder. She achieved three-dimensional control of hand position on the second day of training, progressing to four, and finally seven DOFs (six in the arm and one for grasp) after 3–4 months. Early in the learning process, the movement trajectory of the prosthetic limb was constrained by the decoder computer to a simple, accurate path. This computer assistance was adjusted as the subject learned to control the limb and as task complexity increased, so as to maintain a reasonable success rate. It was eventually eliminated altogether (red portion of the trace in FIG. 2c). The combination of a simple, linear biomimetic decoder and such guided adaptation may have been key to the success of this endeavour.

Figure 2 |. Seven-DOF control of a prosthetic limb for reaching and grasping.

Figure 2 |

a | A patient with tetraplegia had two electrode arrays (Blackrock Microsystems, Salt Lake City, Utah, USA) implanted into the primary motor cortex. b | The patient controlled the position and orientation of the modular prosthetic limb in order to grasp objects. c | The patient’s performance on grasping tasks improved over time, even as the task complexity increased and the degree of computer assistance decreased. The grey trace shows chance level given the level of assistance. d–f | State-of-the-art anthropomorphic prosthetic hands. The modular prosthetic limb (part d) developed by DARPA (Defense Advanced Research Projects Agency, USA) and the Johns Hopkins University Applied Physics Laboratory (Laurel, Maryland, USA) has 19 degrees of freedom (DOFs) in the hand alone (one less than in the actual hand) and over 100 sensors in the arm and hand. The DLR Hand II (part e) from the Institute of Robotics and Mechatronics of the German Aerospace Center (Wessling, Germany) has 15 active DOFs and a wide range of position, contact and torque sensors. The i-limb ultra revolution (part f) from Touch Bionics (Mansfield, Massachusetts, USA) features a powered rotating thumb that moves automatically into position. All five digits move independently, bending at the joints to fit around the shape of whatever object is grasped. It can also be automatically configured into any of 24 grip patterns. Images in parts a and b courtesy of University of Pittsburgh Medical Center, USA. Part c is based on data from REF. 9. Image in part d courtesy of the Johns Hopkins University Applied Physics Laboratory, USA. Image in part e courtesy of DLR, Institute of Robotics and Mechatronics, Germany. Image in part f courtesy of Advanced Arm Dynamics, USA.

Control of grasp.

Prosthetic hands have become increasingly anthropomorphic, with several being nearly as versatile as their human counterparts (FIG. 2d–f). Despite these advances in technology and in our understanding of the representation of grasping within M1 (REFS 52–55), no BMI user has achieved more than one DOF of control in opening and closing a robotic hand3,9 or in controlling simulated grasp force of a virtual object2. For example, the BrainGate experiments8 used a linear discriminant classifier to determine whether the neural activity in M1 represented a hand-open or hand-closed state. This enabled paralysed human patients to grasp objects safely using a robotic hand (including the hand pictured in FIG. 2e).

Although online control of hand movement remains quite limited, offline decoding of hand kinematics is more advanced and points the way towards future improvements in BMI control of prosthetic hands. Both the type of grip (that is, power or precision) and wrist orientation have been classified offline using signals from the anterior intraparietal cortex and PMv56. Similarly, six different object-specific hand conformations57, and even individual finger movements58, have been classified using PMv and M1 signals, respectively. In addition to such state classifiers, continuous decoders have been used offline to predict various high-dimensional, grasp-related kinematic signals. In one study, estimates of hand position and 18 wrist and finger joint angles were computed based on recordings from M1, the dorsal premotor cortex (PMd) and PMv59. In another study, M1 activity was used to predict 25 joint angles from the arm and hand as monkeys grasped different objects60. Here, the authors noted that >95% of the variance of these 25 joint angles could be captured by 10 principal components. In fact, just two or three principal components (similar conceptually to DOFs) are sufficient to capture the variation in hand conformation that is needed to grasp a large variety of objects61, so achieving this level of control with an online BMI is a realistic near-term goal. Whether this kind of higher-level control might be better accomplished using signals recorded from premotor areas56,57,59 than from M1 remains to be established.

The dexterous manipulation of objects requires considerably more DOFs than does simple grasping62,63. To extend control from two or three DOFs to two or three DOFs in each digit — and thus to confer the ability to manipulate objects — will be a considerable challenge for the next phase of BMI development. In fact, simply evaluating the success of BMI-mediated grasping may prove to be a challenge: the complexity of object manipulation means that it is more difficult to evaluate grasping performance than reaching performance. The establishment of a common set of metrics that can be applied across laboratories to assess reaching performance of a BMI and, ultimately, grasping experiments would be an important contribution to the field64,65.

Control of limb impedance and end point forces.

When we reach for and grasp objects, our arm and hand follow a trajectory through space that can be described in terms of the position of the arm and hand and their velocity at any point in time. This description of motion is referred to as kinematics. But we control much more than the kinematics of a movement (BOX 1). For example, we can exert precise amounts of force on delicate objects or large amounts of force to manipulate heavy tools, and we can stiffen the entire limb by co-contracting our muscles when greater stability is required (this is typically referred to as ‘impedance control’ (REF. 66)). These force-related characteristics are referred to as the kinetics of movement or as movement dynamics. The muscle forces that cause movement and regulate stiffness are richly represented in M1 (REFS 67–72) and can be predicted from M1 recordings nearly as well as or, in some cases, better than can movement kinematics2,73–75 (FIG. 3). Some evidence suggests that a subpopulation of neurons in M1 is specialized to regulate limb impedance through muscle co-contraction76.

Figure 3 |. Offline electromyography-based predictions during flexion and extension isometric wrist torque.

Figure 3 |

Most brain–machine interfaces (BMIs) use signals from the primary motor cortex (M1) to make predictions about limb kinematics, typically the position or the velocity of the hand. Force- and muscle-related information is also richly represented in M1 and offers a different approach to movement control with a BMI. In this example, a monkey was trained to place its hand in an isometric force transducer and to exert wrist force in either the flexion (flx.) or extension (ext.) direction. The monkey learned to control a cursor, the movement of which was determined by the forces exerted on the transducer, to place it in a series of targets displayed on a video screen. The resulting force signal is shown by the white trace. At the same time, recordings were made from 77 single- and multi-unit neural recordings in the hand area of M1. The firing rate of each unit, normalized to its maximum rate within this period, is indicated by a colour code, with each line corresponding to a different unit. Note the bursts of unit activity that slightly precede each torque pulse. Electromyography (EMG) signals from the four wrist muscles (red traces) were recorded simultaneously and correlated well with the exerted force. The investigators computed the decoder using data collected in this task and used these data to make EMG predictions for each of the four wrist muscles (blue traces)81. Real-time predictions of muscle activity have been used to control electrical stimulation of muscles aimed at restoring voluntary arm movement in patients with tetraplegia21,84. Similar predictions might also be used to control the movement of a robotic arm by incorporating an appropriate musculoskeletal model80. Such an approach would enable one to control the impedance of the limb during movement as well as the motion itself. The figure is reproduced, with permission, from REF. 75 © (2010) Academic Press, Elsevier.

The tight coupling between M1 activity and muscle activity or force generation was exploited in a pair of studies that used M1 recordings to predict both the kinematics of reaching and the joint torques necessary to cause the limb to move. When monkeys used these hybrid kinematic–torque decoders to control simulated limbs that had dynamics like those of the natural limb (that is, stiffness, damping and mass)22,23, they achieved movement trajectories that were straighter, smoother and arguably more naturalistic than those generated by pure kinematic decoders. By contrast, pure torque decoders used in the same studies produced movements that were abnormally ballistic and difficult for the monkeys to control22,23.

The difficulty in controlling movement using a pure torque decoder of neural signals may be due in part to the inability to control the impedance of the limb77–79. For example, the monkeys would have been unable to stiffen the limb at the end of movement to stabilize the hand at the desired target. In principle, the necessary impedance control could be accomplished with a muscle-based BMI. To test this possibility, another research group used a neural network to reproduce limb motion and joint torque indirectly: they used the network to generate the inputs to a musculoskeletal model of the arm while training the network only on the resultant torque errors80. Although this system was only tested offline, it reproduced torque during reaching across different force-field-loading conditions more accurately than did a model of torque that did not incorporate musculoskeletal mechanics. Importantly, the system also accurately captured the impedance of the limb under the different conditions. Interestingly, the inputs to the musculoskeletal model produced by the neural network generated patterns of modulation that resembled a monkey’s actual electromyographic (EMG) signals, despite the fact that the network was not explicitly trained to do so80.

Several groups have begun to explore the possibility of using M1 recordings to control muscle activity directly — a strategy that could be used to restore voluntary control of limb movements and impedance after spinal cord injury81–83. Real-time EMG predictions such as those shown in FIG. 3 have been used to control the intensity of electrical stimulation delivered to paralysed muscles, causing them to contract — a process referred to as ‘functional electrical stimulation’ (FES). An FES-based BMI has restored voluntary grasp to monkeys whose forearm flexor muscles were temporarily paralysed as a result of a peripheral nerve block21,84. In another study, monkeys were trained to control the activity of individual neurons in order to produce specified levels of isometric wrist torque through FES of both flexor and extensor muscles during nerve block85.

Results from these studies suggest that BMIs that incorporate some form of movement kinetics (for example, torque prediction combined with impedance control or muscle activity predictions) are feasible and might provide the means for patients to interact more naturally with the environment than would be possible with a purely kinematic BMI.

The role of adaptation.

Plastic, adaptive changes in the brain underlie both normal learning and recovery after injury. Adaptation is perhaps most obvious within the motor system when we practise new skills, but it also occurs in the somatosensory system both during skill acquisition86 and following injury87. For example, when humans make reaching movements against a force field88 or with rotated visual feedback89,90, initial errors in reaching direction become smaller and eventually disappear with repeated exposure to the perturbation over tens of minutes. These adaptive processes also cause changes in the subjects’ perception of their hand position91,92 and in somatosensory-evoked potentials93. Similar motor and sensory adaptive processes will almost certainly be engaged as patients learn to control a prosthetic limb.

As mentioned above, two contrasting and complementary strategies have been used in BMI design. One approach — biomimicry — aims to maintain, as much as possible, the mapping between natural patterns of neuronal activity and limb state. The goal is to maximize performance and minimize the amount of learning required of the user. The other approach — typically referred to as ‘adaptation’ — uses systematic, but not necessarily biomimetic, mappings between neural activity and limb state, which the patient must learn. The rationale behind the adaptation approach is that, given the remarkable ability of the brain to adapt and the relatively crude interfaces afforded by the current state of technology, there may be little to gain in attempting to mimic the normal mapping.

Most motor BMIs are designed to be biomimetic. In able-bodied monkeys, the decoder can be computed from the activity of neurons recorded during actual limb movement2,21,24,26,28,44 (BOX 1); in paralysed individuals, it is often computed based on neural discharge during observation of movement43–45. However, even with a biomimetic decoder, the discharge properties of M1 neurons progressively change as an able-bodied monkey switches from making normal reaching movements and starts practising with the BMI94. Typically, the tuning curves of neurons that are used for the BMI become sharper and they often shift towards a new preferred direction. If a monkey is tested in this manner over a series of experimental sessions, the tuning curves of neurons that are tracked across sessions tend to shift by the same amount each time2,26,28. The changes in neuronal discharge may reflect changes in the motor intent, as though the monkey is learning a new motor behaviour. In fact, the adaptation may be necessitated by the use of a kinematic decoder (BOX 1). Normally, M1 controls movement though the activation of muscles and the generation of forces. During kinematic brain control, the normal dynamics of movement are eliminated, as though the limb has suddenly lost all of its mass. The monkey may need to adapt to this change by learning a new motor strategy. As intact monkeys become better at BMI control, they generally alter their actual limb movements or cease making them altogether. Consequently, in addition to the altered motor intent, the altered discharge noted above may result from changes in afferent input to M1 as the actual limb movements change or no longer occur.

These adaptive processes have been further investigated in experiments in which the BMI decoder was intentionally perturbed. In one example, monkeys were trained to use a biomimetic decoder to move a cursor before and after all inputs to the decoder were randomly swapped (so that the decoder was no longer biomimetic)27. Under these conditions, the monkeys’ performance initially declined and then gradually returned, over several sessions, to that achieved with the initial biomimetic decoder. Changes in success rate correlated with the progressive changes in neuronal tuning curves27.

In another set of experiments, half of the tuning curves used for control were intentionally rotated by 90 degrees about a common axis28, so that any attempted movement towards a given target, which was displayed in a three-dimensional virtual-reality environment, resulted in cursor movement in the wrong direction, much like the classic visual rotation experiments89,90. Monkeys learned to compensate for this perturbation within a single session, although more slowly and less completely than in natural (non-BMI) adaptation experiments28,31. Most neurons changed their discharge pattern in a manner that suggests that the monkeys simply learned to aim at a different target28. However, there were indications of more specific learning as well. Although the tuning curves of all of the recorded neurons tended to rotate, those of the ‘perturbed’ neurons rotated more than the others, perhaps to compensate directly for the perturbation. Furthermore, the overall discharge of the perturbed neurons was gradually reduced so that they contributed less to the output. Consistent with this result, another study found that the neurons used for BMI-based control of cursor movement underwent larger changes upon transition from actual hand control to brain control than did neurons whose responses were recorded but not used in the decoder29. Thus, in both of these experiments, the monkeys apparently learned to modulate the activity of individual neurons selectively in order to improve the BMI control.

These changes happen automatically, probably through normal motor learning mechanisms. A different approach might be to train the animal to control the activation of individual neurons directly. Monkeys can learn to adapt the discharge of neurons in M1 if they are provided with rewards based on a desired discharge rate30,95. Interestingly, monkeys could learn to control the activation of individual motor cortical neurons regardless of any prior relation the neurons may have had to movement. In fact, the monkeys even learned to modulate neurons recorded in the somatosensory cortex96. When signals were recorded from pairs of neurons but rewards were given based on the discharge of only one of the two neurons, the firing rate of the second, unrewarded cell, was generally uncorrelated with that of the rewarded neuron, suggesting that the monkey did not use a strategy involving indiscriminate activation of many neurons. In fact, monkeys can even learn to control two neurons independently if they are differentially rewarded30.

The first implementation of an intracortical interface in a human patient used essentially this approach. Two locked-in patients with terminal amyotrophic lateral sclerosis learned, with at least some success, to control the movement of a cursor that provided a communication interface97. More recently, single-neuron conditioning was used to implement the one-DOF FES-based BMI and to generate the isometric contractions described above85. The extent to which this approach can be scaled up to include larger numbers of neurons and more complex movements remains to be determined.

Monkeys, and presumably humans, can thus learn new mappings between the activity of neurons and movement — a process that may be analogous to normal motor learning. Remarkably, these results indicate that some learning occurs even at the level of individual neurons during BMI control. Nonetheless, one might assume that a biomimetic decoder that closely approximates the normal mapping from brain to movement would yield faster learning and perhaps better ultimate performance than a non-biomimetic mapping, but this remains an important question for future research.

Conveying somatosensory feedback with a BMI

Movements are normally guided by somatosensory as well as visual feedback. The absence of somatosensory feedback in current BMIs undoubtedly limits the quality of the movements that are produced by those BMIs. Below, we discuss the importance of touch and proprioception in motor control, and describe approaches to restore somatosensation through intracortical microstimulation (ICMS).

The importance of somatosensation.

Loss of somatosensation — including touch and proprioception — has a devastating effect on motor abilities. Signals from mechanoreceptive afferents in our skin convey information about the location of contact98,99 and about the forces exerted on the skin when we grasp an object100–102. Cutaneous afferents also signal when our grip on an object is slipping103. Without these afferents, we would routinely crush or drop objects we are grasping. As might be expected, cutaneous anaesthesia of the hand makes it virtually impossible to manipulate small objects104. The importance of proprioception in motor control has been brought to light most compellingly in lesion and clinical studies. Lesioning or inactivation of the primary somatosensory cortex (S1) in monkeys results in uncoordinated finger movements, and coarse and exaggerated grasp behaviours, the latter presumably as an adaptive response to compensate for attenuated sensation105. Loss of proprioception largely eliminates the ability to plan the dynamics of limb movements106. In addition to its importance in motor control, somatosensation plays an important part in embodiment and in our affective lives. Indeed, the sensory experience of our limbs’ position and movements in space, and of events impinging upon our skin, confers to the limbs an embodiment, so that they are experienced as a part of ‘me’ (REF. 107). Furthermore, touch is essential for communicating emotions and for sexual behaviour108.

Given the importance of somatosensation, the functionality of cortical BMIs may not justify the invasiveness of the required surgery109 until they provide for tactile and proprioceptive feedback — a prospect that has spurred the development of the first bidirectional BMI, in which stimulation of the somatosensory cortex guided monkeys’ use of an efferent interface20.

Elicitation of somatosensory percepts through cortical stimulation.

In the 1930s and 1940s, Wilder Penfield110 provided the first examples of somatosensory percepts elicited through electrical stimulation. He induced tactile sensations by applying electrical stimulation to the exposed brains of patients undergoing neurosurgical procedures. These sensations were ‘felt’ at body regions whose location was systematically related to the site of stimulation on the cortical surface111. Much later, in a seminal experiment using ICMS in monkeys, Romo and colleagues112 elicited percepts of cutaneous flutter at frequencies corresponding to the ICMS pulse frequency, showing that the temporal structure of ICMS pulse trains shapes, in systematic ways, the evoked sensations. More recent experiments have characterized how the detectability of ICMS depends on stimulation parameters in rodents113,114 and primates14,16,20,115. These studies paved the way for attempts to elicit a broader range of sensations through ICMS in S1, with the hope to one day restore touch and proprioception to individuals who have lost these senses.

One approach to restoring somatosensation through a brain interface is to attempt to elicit percepts through stimulation that approximates the sensory experiences that arise during everyday behaviour, much like the biomimetic approach to a motor BMI. This involves restoring different aspects of touch and proprioception in ways on the basis of our understanding of the relevant natural somatosensory coding. In one example of such a biomimetic mapping in rodents, ICMS was delivered to a single location in the barrel cortex whenever the rat’s corresponding whisker crossed a particular spatial location (although it touched nothing). Rats were able to locate virtual objects using this artificial sensation in a manner that mimicked the sensory experience the rats would have had when whisking a physical target13,116. A similar example of the biomimetic approach, in primates, is the use of somatotopy to convey information about contact location in the hand. ICMS in S1 evokes percepts that can be highly localized, down to a single palmar whorl or fingertip14 (FIG. 4a). This somatotopic organization can be exploited to signal which regions of a prosthetic limb have contacted an object: activation of sensors on the prosthesis could be used to drive activation of S1 neurons with corresponding receptive fields.

Figure 4 |. Conveying information about contact location and pressure using a biomimetic approach.

Figure 4 |

a | Monkeys were mechanically touched on the hand twice in a row and were trained to judge whether the second indentation was to the left or right (lateral or medial) of the first one (left panel). (Each blue dot corresponds to the location of a mechanical indentation.) On a subset of trials (hybrid trials), one of the indentations was replaced with an intracortical microstimulation (ICMS) pulse train delivered to an electrode with a receptive field centred at the site of the indentation (red shaded area). Mean performance on mechanical and hybrid trials is shown in the right panel. Although animals performed better on mechanical trials than on hybrid trials, performance in the latter was considerably (and significantly) above chance (0.5). In other words, ICMS through individual electrodes produces percepts that are localized to a patch of skin corresponding to their receptive field. b | Monkeys were trained to detect indentations of the skin or ICMS pulse trains delivered to the primary somatosensory cortex (S1); they were also trained to discriminate the intensity of these mechanical or electrical stimuli using the same behavioural paradigm. The graph on the left shows the mapping between pressure magnitude (that is, the depth of the skin indentation) and the ICMS pulse amplitude that yielded percepts of equal sensory magnitude (for example, that were equally detectable or equally discriminable from an equivalently detectable standard stimulus). The graph on the right shows that animals performed equally well on a force discrimination task when the tactile stimuli were delivered to the actual finger (blue dots and trace) as when they were delivered to a prosthetic finger (red dots and trace, with each trace corresponding to a different electrode). Data in right panel of part a from REF. 14. Part b is modified from REF. 14.

One might ask how a ‘somatotopically appropriate’ mapping between sensors on the prosthetic hand and stimulating electrodes in S1 could be achieved in an amputee or tetraplegic patient. In these patients, although the limb is missing or no longer connected to the brain, the part of the brain that responded to movement or tactile stimulation of the limb is still intact. In fact, amputees often experience sensations projected to a so-called ‘phantom limb’; that is, the sensations are felt to be located where the missing limb used to be117. One approach to achieve a sensor–electrode mapping might consist of delivering ICMS through individual electrodes while the patient reports where on the phantom or deafferented limb he or she ‘feels’ the evoked sensations.

In addition to location information, a biomimetic approach can be used to convey information about the pressure exerted on objects by the prosthesis by converting the output of force sensors into ICMS pulse trains that evoke percepts of appropriate magnitude118 (FIG. 4b). Furthermore, information about different stimulus properties (for example, surface texture or direction of limb movement) might be conveyed by stimulating separate sets of neurons in S1. For example, some S1 neurons signal the direction in which a tactile stimulus moves across the skin119, whereas others signal the orientation of an edge indented into the skin120 and still others signal the position and motion of the limb121–123. Thus, targeted stimulation of specific S1 neurons could in principle elicit percepts of cutaneous motion, of shape or of limb movement19. In the clinic, one could rely on patient reports to identify the quality associated with individual electrodes.

It is currently impossible to activate large numbers of neurons individually and independently. Consequently, to evoke a specific perceptual quality requires that neurons that are activated together by a given ICMS train all have similar receptive field properties, so that they convey a consistent message. Even low-current ICMS involves the synchronous activation of populations of neurons, some of which may be a large distance away from the electrode124. These more distant neurons may exhibit very different response dynamics than those near the electrode tip125. Considering this, the effort to produce biomimetic patterns of neuronal activation might seem to be doomed to failure. Fortunately, there is some evidence of functional clustering within the tactile126 and proprioceptive18 representations in S1. Nonetheless, it remains unclear whether these functional clusters are segregated enough to enable electrical activation of a sufficiently large and homogenous group of neurons to elicit well-defined sensations of particular qualities (but see REF. 19 for an example of a sensation of limb movement induced using this biomimetic strategy). Biophysical models might provide some insight into the effects of stimulation on neuronal activity and thereby help to refine the stimulation regimes required to produce verisimilar patterns of neuronal activation. Stimulation across large numbers of electrodes, each with very low currents, may be one approach115. Optogenetic stimulation may enable more selective activation of neurons than is possible with electrical stimulation127,128. However, technical advances alone cannot get around the fact that neurons are embedded in networks, and our understanding of network dynamics and of their relationship with perception and behaviour is still primitive. Thus, although biomimicry is an intuitive way to convey sensory feedback, it faces considerable technical and scientific challenges. Sensory BMIs, like motor BMIs, will also benefit from the brain’s ability to adapt; in this case, to learn to compensate for limitations in the verisimilitude or intuitiveness of the artificial percepts.

Adaptation in sensory BMIs.

The ‘adaptation’ approach requires systematic mappings between sensors on a prosthetic limb and patterns of electrical stimulation applied to the brain, but without the constraint that the stimulation should necessarily lead to naturalistic patterns of neural activity. The patient must learn to make associations between the percepts elicited by different patterns of stimulation and the corresponding sensory events (and resulting sensor activations). The idea is that, as long as different states of sensor activation evoke discriminable percepts, the patient may be able to learn the mapping. Animals have been trained to discriminate small numbers of fixed temporal11,20,112 and spatiotemporal16 patterns of electrical stimulation applied to the somatosensory cortex (FIG. 5).

Figure 5 |. Bidirectional interface based on learned associations.

Figure 5 |

The monkey explored virtual space with a virtual hand (or avatar) to find which of three visually identical targets had the right ‘texture’. The avatar was controlled by signals recorded in the primary motor cortex (M1); tactile sensations of texture were elicited by stimulating the primary somatosensory cortex (S1) with patterned pulse trains. When the avatar hovered over two of the three visual targets, the monkey received one of two pulse trains of intracortical microstimulation (ICMS). The third target did not trigger ICMS. The animal was successful if it correctly identified the target associated with the rewarded ICMS train. Over time, the animal learned to identify the rewarded target. This study demonstrates that animals can learn to discriminate between different patterns of sensory stimulation applied to S1 and combined, for the first time, motor decoding and ICMS feedback. The figure is modified, with permission, from REF. 20 © (2011) Macmillan Publishers Ltd. All rights reserved.

In other experiments, continuously modulated stimulation was used to guide ongoing movement. In one study, the strength of ICMS in the whisker area of the somatosensory cortex in rats was modulated according to the distance of the rat’s head to an infrared light located above a water port; the closer the rat’s head was to the port, the stronger the ICMS. The rats learned to locate the water port even though the somatosensory areas of their brains were being activated in an entirely novel manner compared to the patterns normally elicited during whisking129. In another experiment, a monkey learned to guide the movement of a cursor on a screen on the basis of the strength of somatosensory ICMS delivered across eight electrodes. The strength of ICMS was modulated according to an error vector that pointed from cursor to target. The stimulation frequency on each of the eight electrodes was determined by the projection of the error vector onto one of eight evenly spaced unit vectors17. The monkey learned the mapping over a period of several months, during which the ICMS was paired with visual signals carrying the same target-related information. In both of these studies, animals could perform the tasks despite the fact that the information provided by the stimulation was not information that is normally encoded in these brain areas.

Importantly, the limb can adopt a huge number of (proprioceptive) states, and the number of different combinations of cutaneous events that can impinge upon the limb is virtually infinite. The main question as to the viability of the adaptation approach is whether the adult brain can learn a mapping between such a complex space of possible states and corresponding patterns of ICMS. Nevertheless, even the most ardent biomimicry enthusiast must acknowledge that artificial percepts will be of limited verisimilitude and that patients will have to rely at least in part on learning and plasticity to master the use of a sensory prosthesis.

A compelling example of the synergy between biomimicry and adaptation is provided by a study involving targeted reinnervation. In this study, amputees underwent surgery to reimplant nerves that had once innervated a severed limb into remaining skin and muscle sites. When this newly innervated skin was touched, these individuals experienced tactile sensations that were projected to the phantom limb130. Two of these patients underwent an experiment in which the newly innervated skin was mechanically stimulated each time the corresponding ‘receptive field’ on the prosthetic limb was touched. The repeated pairing caused the percepts to move from being experienced on the phantom limb to being experienced on the prosthetic limb, as determined by self-reports and physiological measures of embodiment131. Such a migration might also be observed when spatiotemporally appropriate ICMS is delivered during object manipulation with the prosthetic limb132.

As is the case for motor BMIs, biomimetic feedback will probably be more intuitive to use initially and require less learning to master, which may have a substantial impact on how readily sensory BMIs will be embraced by patients. However, the ability to elicit naturalistic patterns of neuronal activation is extremely limited given the current state of science and technology; therefore, learning and adaptation will have important roles in the interpretation of artificial sensory feedback. Studies within the biomimicry framework will tell us just how verisimilar artificial percepts can be, and studies within the adaptation-based framework will tell us how far learning can take us.

Summary

The quality of motor BMIs has progressed dramatically since a single-DOF robot was first controlled from the brain, but BMIs still remain largely restricted to executing the kinematics of reaching under visual guidance. The major engineering challenges of the next generation of BMIs will include the incorporation of somatosensory feedback through an afferent interface, direct control of limb impedance and of forces applied on the environment (BOX 1), and the ability to manipulate objects with anthropomorphic prosthetic hands. Understanding and exploiting the interaction between biomimetic designs and both user and decoder adaptation (BOX 2) will be essential for this development effort.

BMI technology must also move out of the laboratory and into the clinic. Percutaneous connectors must be eliminated to address safety, cosmetic and convenience concerns. The overall reliability of BMIs must be improved, especially the limited life of existing microelectrode arrays (BOX 3). The accessibility of BMIs also needs improvement, so that they can be used 24 hours a day without requiring continuous technical support. The challenges that face the field are substantial, but the opportunities afforded by BMI technology are unprecedented. Acceptance of this technology into the clinical arena is likely to hinge on the success of these developments in the next 5 to 10 years.

Acknowledgements

The authors gratefully thank J. Yau, H. Saal, A. Suminski and K. Otto for their comments on a previous version of the manuscript. The authors also thank G. Tabot for designing figure 1. S.J.B.is supported by US Defense Advanced Research Projects Agency (DARPA) contract #N66001-10-C-4056, US National Science Foundation (NSF) grant IOS-1150209 and US National Institutes of Health (NIH) grant 082865. L.E.M. is supported by grants from the US NIH (NS053603, NS048845) and the US NSF (0932263), with additional funding from the Chicago Community Trust through the Searle Program for Neurological Restoration.

Glossary

Degree of freedom

(DOF). The number of signals required to control a device. The DOF is determined approximately by the number of parameters that defines its configuration.

Decoders

A set of (often linear) coefficients used to transform a large number of signals recorded from the brain into a small number of control signals. A decoder might also be used simply to classify the brain signals into two or more clusters that could be used to control the state of a limb.

Offline analysis

A test of decoder performance, typically using signals previously recorded from an able-bodied subject, enabling comparison of the decoder’s ‘predictions’ with the actual movement-related signals.

Impedance

In electricity, the opposition to alternating current by an electric circuit. In limb movement, a measure of how much the limb resists motion when subjected to a force.

Redundant

A limb having more degrees of freedom (for example, muscles or joint rotations) than are minimally necessary to position and orient its end point. Redundancy conveys flexibility but also requires more complex control algorithms.

Actor–critic

A reinforcement learning approach that consists of having an ‘actor’ perform an action based on the state of the system and a ‘critic’ evaluate the consequences of that action. The probability of performing that action given the state is then modified based on the consequences.

Online control

Actual predictions made with a decoder in real-time, allowing the user to control a robotic limb or the motion of a cursor.

Ballistic

A preprogrammed movement that is sufficiently rapid that it cannot be modified by online sensory feedback.

Proprioception

The sense of the relative position and motion of parts of the body (particularly limbs) and of the effort deployed in movement.

Flutter

Low-frequency (~5–50 Hz) oscillations.

Verisimilitude

In the context of sensory brain–machine interfaces, the similarity to naturally occurring percepts.

Percutaneous

Literally, ‘by way of the skin’. In this context, an interface that penetrates the skin in order to convey signals to and from the nervous system.

Footnotes

Competing interests statement

The authors declare no competing interests.

References

  • 1.Chapin JK, Moxon KA, Markowitz RS & Nicolelis MA Real-time control of a robot arm using simultaneously recorded neurons in the motor cortex. Nature Neurosci. 2, 664–670 (1999). [DOI] [PubMed] [Google Scholar]
  • 2.Carmena JM et al. Learning to control a brain–machine interface for reaching and grasping by primates. PLoS Biol. 1, e42 (2003). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Velliste M, Perel S, Spalding MC, Whitford AS & Schwartz AB Cortical control of a prosthetic arm for self-feeding. Nature 453, 1098–1001 (2008). [DOI] [PubMed] [Google Scholar]
  • 4.Santhanam G, Ryu SI, Yu BM, Afshar A & Shenoy KV A high-performance brain–computer interface. Nature 442, 195–198 (2006). [DOI] [PubMed] [Google Scholar]
  • 5.Gilja V et al. A high-performance neural prosthesis enabled by control algorithm design. Nature Neurosci. 15, 1752–1757 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Mulliken GH, Musallam S & Andersen RA Decoding trajectories from posterior parietal cortex ensembles. J. Neurosci 28, 12913–12926 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Li Z, O’Doherty JE, Lebedev MA & Nicolelis MA Adaptive decoding for brain–machine interfaces through Bayesian parameter updates. Neural Comput. 23, 3162–3204 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Hochberg LR et al. Reach and grasp by people with tetraplegia using a neurally controlled robotic arm. Nature 485, 372–375 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Collinger JL et al. High-performance neuroprosthetic control by an individual with tetraplegia. Lancet 381, 557–564 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]; This study represents the current state of the art of a tetraplegic human controlling a prosthetic limb with many DOFs. Optimal performance was obtained through a two-stage biomimetic decoder and guided, progressive user adaptation.
  • 10.Johansson RS & Flanagan JR Coding and use of tactile signals from the fingertips in object manipulation tasks. Nature Rev. Neurosci 10, 345–359 (2009). [DOI] [PubMed] [Google Scholar]; This paper provides an excellent review of how tactile signals are important for manipulating objects.
  • 11.London BM, Jordan LR, Jackson CR & Miller LE Electrical stimulation of the proprioceptive cortex (area 3a) used to instruct a behaving monkey. IEEE Trans. Neural Syst. Rehabil. Eng 16, 32–36 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.O’Doherty JE, Lebedev MA, Li Z & Nicolelis MA Virtual active touch using randomly patterned intracortical microstimulation. IEEE Trans. Neural Syst. Rehabil. Eng 20, 85–93 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Venkatraman S & Carmena JM Active sensing of target location encoded by cortical microstimulation. IEEE Trans. Neural Syst. Rehabil. Eng 19, 317–3241 (2011). [DOI] [PubMed] [Google Scholar]
  • 14.Tabot GA et al. Restoring the sense of touch with a prosthetic hand through a brain interface. Proc. Natl Acad. Sci. USA 110, 18279–18284 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]; This study describes approaches to convey information about contact location, force and timing through a biomimetic strategy of ICMS applied to the somatosensory cortex.
  • 15.Romo R, Hernandez A, Zainos A, Brody CD & Lemus L Sensing without touching: psychophysical performance based on cortical microstimulation. Neuron 26, 273–278 (2000). [DOI] [PubMed] [Google Scholar]
  • 16.Fitzsimmons NA, Drake W, Hanson TL, Lebedev MA & Nicolelis MA Primate reaching cued by multichannel spatiotemporal cortical microstimulation. J. Neurosci 27, 5593–5602 (2007). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Dadarlat MC, O’Doherty JE & Sabes PN Multisensory integration of vision and intracortical microstimulation for sensory substitution and augmentation. Soc. Neurosci. Abstr 792.12 (2012). [Google Scholar]
  • 18.Weber DJ et al. Limb-state information encoded by peripheral and central somatosensory neurons: Implications for an afferent interface. IEEE Trans. Neural Syst. Rehabil. Eng 19, 501–513 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Tomlinson T, Ruiz Torres R & Miller LE Multi-electrode stimulation in somatosensory area 2 induces a natural sensation of limb movement. Soc. Neurosci. Abstr 835.03 (2013). [Google Scholar]
  • 20.O’Doherty JE et al. Active tactile exploration using a brain-machine-brain interface. Nature 479, 228–231 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]; This study was the first to describe a bidirectional BMI in which a cursor was controlled by signals from the motor cortex while stimulation was delivered to the somatosensory cortex to signal the consequences of those movements. The monkey had to learn the mapping of the afferent interface.
  • 21.Ethier C, Oby ER, Bauman MJ & Miller LE Restoration of grasp following paralysis through brain-controlled stimulation of muscles. Nature 485, 368–371 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]; Monkeys were able to grasp and move objects despite temporary paralysis of forearm flexor muscles in this study, which used a biomimetic decoder to evoke electrically induced muscle contractions controlled in real-time by predictions of muscle activity.
  • 22.Suminski AJ, Willett FR, Fagg AH, Bodenhamer M & Hatsopoulos NG Continuous decoding of intended movements with a hybrid kinetic and kinematic brain machine interface. Conf. Proc. IEEE Eng. Med. Biol. Soc 2011, 5802–5806 (2011). [DOI] [PubMed] [Google Scholar]
  • 23.Chhatbar PY & Francis JT Towards a naturalistic brain–machine interface: hybrid torque and position control allows generalization to novel dynamics. PLoS ONE 8, e52286 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Serruya MD, Hatsopoulos NG, Paninski L, Fellows MR & Donoghue JP Instant neural control of a movement signal. Nature 416, 141–142 (2002). [DOI] [PubMed] [Google Scholar]; This classic study was the first to achieve continuous control of two-dimensional cursor movement using intracortical recordings of neural activity.
  • 25.Wessberg J et al. Real-time prediction of hand trajectory by ensembles of cortical neurons in primates. Nature 408, 361–365 (2000). [DOI] [PubMed] [Google Scholar]
  • 26.Taylor DM, Tillery SI & Schwartz AB Direct cortical control of 3D neuroprosthetic devices. Science 296, 1829–1832 (2002). [DOI] [PubMed] [Google Scholar]
  • 27.Ganguly K & Carmena JM Emergence of a stable cortical map for neuroprosthetic control. PLoS Biol. 7, e1000153 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]; Given several days of practice, monkeys became as proficient in using a decoder with scrambled inputs as they had been in its original, biomimetic state. They could readily switch between the two decoders. Changes in neural tuning tracked the behavioural improvement.
  • 28.Jarosiewicz B et al. Functional network reorganization during learning in a brain–computer interface paradigm. Proc. Natl Acad. Sci. USA 105, 19486–19491 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]; This study emulates the classic ‘visual rotation’ studies by systematically (rather than randomly) remapping the decoder output. It reveals a combination of global and local adaptive changes in neural tuning.
  • 29.Ganguly K, Dimitrov DF, Wallis JD & Carmena JM Reversible large-scale modification of cortical networks during neuroprosthetic control. Nature Neurosci. 14, 662–667 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Fetz EE & Baker MA Operantly conditioned patterns on precentral unit activity and correlated responses in adjacent cells and contralateral muscles. J. Neurophysiol 36, 179–204 (1973). [DOI] [PubMed] [Google Scholar]; This paper extends earlier work on the conditioning of single neurons in M1 by showing that monkeys can learn to control the discharge of different neurons independently, even that of adjacent pairs of neurons.
  • 31.Chase SM, Kass RE & Schwartz AB Behavioral and neural correlates of visuomotor adaptation observed through a brain–computer interface in primary motor cortex. J. Neurophysiol 108, 624–644 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Millán JR & Carmena JM Invasive or noninvasive: understanding brain–machine interface technology. IEEE Eng. Med. Biol. Mag 29, 16–19 (2010). [DOI] [PubMed] [Google Scholar]
  • 33.Barrese JC et al. Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates. J. Neural Eng 10, 066014 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Hwang EJ & Andersen RA The utility of multichannel local field potentials for brain–machine interfaces. J. Neural Eng 10, 046005 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Schalk G et al. Two-dimensional movement control using electrocorticographic signals in humans. J. Neural Eng 5, 75–84 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Wolpaw JR & McFarland DJ Control of a two-dimensional movement signal by a noninvasive brain–computer interface in humans. Proc. Natl Acad. Sci. USA 101, 17849–17854 (2004). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Flint RD, Wright ZA, Scheid MR & Slutzky MW Long term, stable brain machine interface performance using local field potentials and multiunit spikes. J. Neural Eng 10, 056005 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Mehring C et al. Inference of hand movements from local field potentials in monkey motor cortex. Nature Neurosci. 6, 1253–1254 (2003). [DOI] [PubMed] [Google Scholar]
  • 39.Bansal AK, Vargas-Irwin CE, Truccolo W & Donoghue JP Relationships among low-frequency local field potentials, spiking activity, and three-dimensional reach and grasp kinematics in primary motor and ventral premotor cortices. J. Neurophysiol 105, 1603–1619 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Stark E & Abeles M Predicting movement from multiunit activity. J. Neurosci 27, 8387–8394 (2007). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Chestek CA et al. Long-term stability of neural prosthetic control signals from silicon cortical arrays in rhesus macaque motor cortex. J. Neural Eng 8, 045005 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Sellers E, McFarland D, Vaughan T & Wolpaw J in Brain–Computer Interfaces: Revolutionizing Human–Computer Interaction (eds Graimann B, Allison B & Pfurtscheller G) 97–111 (Springer, 2010). [Google Scholar]
  • 43.Wahnoun R, He J & Helms Tillery SI Selection and parameterization of cortical neurons for neuroprosthetic control. J. Neural Eng 3, 162–171 (2006). [DOI] [PubMed] [Google Scholar]
  • 44.Hochberg LR et al. Neuronal ensemble control of prosthetic devices by a human with tetraplegia. Nature 442, 164–171 (2006). [DOI] [PubMed] [Google Scholar]; This study was the first in which a paralysed human patient achieved continuous, two-dimensional control of a cursor through an intracortical brain interface.
  • 45.Tkach D, Reimer J & Hatsopoulos NG Congruent activity during action and action observation in motor cortex. J. Neurosci 27, 13241–13250 (2007). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Rizzolatti G & Craighero L The mirror-neuron system. Annu. Rev. Neurosci 27, 169–192 (2004). [DOI] [PubMed] [Google Scholar]
  • 47.di Pellegrino G, Fadiga L, Fogassi L, Gallese V & Rizzolatti G Understanding motor events: a neurophysiological study. Exp. Brain Res 91, 176–180 (1992). [DOI] [PubMed] [Google Scholar]
  • 48.Cheney PD & Fetz EE Corticomotoneuronal cells contribute to long-latency stretch reflexes in the rhesus monkey. J. Physiol 349, 249–272 (1984). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Asanuma H Functional role of sensory inputs to the motor cortex. Prog. Neurobiol 16, 241–262 (1981). [DOI] [PubMed] [Google Scholar]
  • 50.Fetz EE, Perlmutter SI, Prut Y, Seki K & Votaw S Roles of primate spinal interneurons in preparation and execution of voluntary hand movement. Brain Res. Rev 40, 53–65 (2002). [DOI] [PubMed] [Google Scholar]
  • 51.Johannes MS et al. An overview of the developmental process for the modular prosthetic limb. Johns Hopkins Apl. Tech. Digest 30, 207–216 (2011). [Google Scholar]
  • 52.Landgren S, Phillips CG & Porter R Cortical fields of origin of the monosynaptic pyramidal pathways to some alpha motoneurones of the baboon’s hand and forearm. J. Physiol 161, 112–125 (1962). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Schieber MH & Hibbard LS How somatotopic is the motor cortex hand area? Science 261, 489–492 (1993). [DOI] [PubMed] [Google Scholar]
  • 54.Maier MA, Bennett KMB, Hepp-Reymond MC & Lemon RN Contribution of the monkey corticomotoneuronal system to the control of force in precision grip. J. Neurophysiol 69, 772–785 (1993). [DOI] [PubMed] [Google Scholar]
  • 55.Hendrix CM, Mason CR & Ebner TJ Signaling of grasp dimension and grasp force in dorsal premotor cortex and primary motor cortex neurons during reach to grasp in the monkey. J. Neurophysiol 102, 132–145 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Townsend BR, Subasi E & Scherberger H Grasp movement decoding from premotor and parietal cortex. J. Neurosci 31, 14386–14398 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Carpaneto J et al. Continuous decoding of grasping tasks for a prospective implantable cortical neuroprosthesis. J. Neuroeng. Rehabil 9, 84 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Egan J, Baker J, House PA & Greger B Decoding dexterous finger movements in a neural prosthesis model approaching real-world conditions. IEEE Trans. Neural Syst. Rehabil. Eng 20, 836–844 (2012). [DOI] [PubMed] [Google Scholar]
  • 59.Aggarwal V, Mollazadeh M, Davidson AG, Schieber MH & Thakor NV State-based decoding of hand and finger kinematics using neuronal ensemble and LFP activity during dexterous reach-to-grasp movements. J. Neurophysiol 109, 3067–3081 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Vargas-Irwin CE et al. Decoding complete reach and grasp actions from local primary motor cortex populations. J. Neurosci 30, 9659–9669 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Santello M, Flanders M & Soechting JF Postural hand synergies for tool use. J. Neurosci 18, 10105–10115 (1998). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Thakur PH, Bastian AJ & Hsiao SS Multidigit movement synergies of the human hand in an unconstrained haptic exploration task. J. Neurosci 28, 1271–1281 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Todorov E & Ghahramani Z Analysis of the synergies underlying complex hand manipulation. Conf. Proc. IEEE Eng. Med. Biol. Soc 6, 4637–4640 (2004). [DOI] [PubMed] [Google Scholar]
  • 64.Thompson D, Blain-Moraes S & Huggins J Performance assessment in brain-computer interface-based augmentative and alternative communication. BioMed Eng. Online 12, 43 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Yuan P et al. A study of the existing problems of estimating the information transfer rate in online brain–computer interfaces. J. Neural Eng 10, 026014 (2013). [DOI] [PubMed] [Google Scholar]
  • 66.Hogan N Active control of mechanical impedance by coactivation of antagonist muscles. IEEE Trans. Automat. Control 29, 681–690 (1984). [Google Scholar]
  • 67.Evarts EV Relation of pyramidal tract activity to force exerted during voluntary movement. J. Neurophysiol 31, 14–27 (1968). [DOI] [PubMed] [Google Scholar]
  • 68.Smith AM, Hepp-Reymond MC & Wyss UR Relation of activity in precentral cortical neurons to force and rate of force change during isometric contractions of finger muscles. Exp. Brain Res 23, 315–332 (1975). [DOI] [PubMed] [Google Scholar]
  • 69.Kalaska JF, Cohen DAD, Hyde ML & Prud’homme M A comparison of movement direction-related versus load direction-related activity in primate motor cortex, using a two-dimensional reaching task. J. Neurosci 9, 2080–2102 (1989). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Boline J & Ashe J On the relations between single cell activity in the motor cortex and the direction and magnitude of three-dimensional dynamic isometric force. Exp. Brain Res 167, 148–159 (2005). [DOI] [PubMed] [Google Scholar]
  • 71.Cheney PD & Fetz EE Functional classes of primate corticomotorneuronal cells and their relation to active force. J. Neurophysiol 44, 773–791 (1980). [DOI] [PubMed] [Google Scholar]
  • 72.Hepp-Reymond MC, Wyss UR & Anner R Neuronal coding of static force in the primate motor cortex. J. Physiol. Paris 74, 287–291 (1978). [PubMed] [Google Scholar]
  • 73.Humphrey DR, Schmidt EM & Thompson WD Predicting measures of motor performance from multiple cortical spike trains. Science 170, 758–761 (1970). [DOI] [PubMed] [Google Scholar]; This classic study was the first to make real-time predictions of movement-related parameters using single-unit recordings from multiple neurons in the motor cortex.
  • 74.Fagg AH, Ojakangas GW, Miller LE & Hatsopoulos NG Kinetic trajectory decoding using motor cortical ensembles. IEEE Trans. Neural Syst. Rehabil. Eng 17, 487–496 (2009). [DOI] [PubMed] [Google Scholar]
  • 75.Oby ER et al. in Statistical Signal Processing for Neuroscience and Neurotechnology (ed. O’Weiss KG) 369–406 (Academic Press, Elsevier, 2010). [Google Scholar]
  • 76.Humphrey DR & Reed DJ in Motor Control Mechanisms in Health and Disease (ed. Desmedt JE) 347–372 (Raven, 1983). [Google Scholar]
  • 77.Burdet E, Osu R, Franklin DW, Milner TE & Kawato M The central nervous system stabilizes unstable dynamics by learning optimal impedance. Nature 414, 446–449 (2001). [DOI] [PubMed] [Google Scholar]
  • 78.Franklin D, Burdet E, Osu R, Kawato M & Milner T Functional significance of stiffness in adaptation of multijoint arm movements to stable and unstable dynamics. Exp. Brain Res 151, 145–157 (2003). [DOI] [PubMed] [Google Scholar]
  • 79.Gribble PL, Mullin LI, Cothros N & Mattar A Role of cocontraction in arm movement accuracy. J. Neurophysiol 89, 2396–2405 (2003). [DOI] [PubMed] [Google Scholar]
  • 80.Kim HK et al. The muscle activation method: an approach to impedance control of brain-machine interfaces through a musculoskeletal model of the arm. IEEE Trans. Biomed. Eng 54, 1520–1529 (2007). [DOI] [PubMed] [Google Scholar]
  • 81.Pohlmeyer EA, Solla SA, Perreault EJ & Miller LE Prediction of upper limb muscle activity from motor cortical discharge during reaching. J. Neural Eng 4, 369–379 (2007). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 82.Santucci DM, Kralik JD, Lebedev MA & Nicolelis MA Frontal and parietal cortical ensembles predict single-trial muscle activity during reaching movements in primates. Eur. J. Neurosci 22, 1529–1540 (2005). [DOI] [PubMed] [Google Scholar]
  • 83.Flint RD, Ethier C, Oby ER, Miller LE & Slutzky MW Local field potentials allow accurate decoding of muscle activity. J. Neurophysiol 108, 18–24 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84.Pohlmeyer EA et al. Toward the restoration of hand use to a paralyzed monkey: brain-controlled functional electrical stimulation of forearm muscles. PLoS ONE 4, e5924 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 85.Moritz CT, Perlmutter SI & Fetz EE Direct control of paralysed muscles by cortical neurons. Nature 456, 639–642 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 86.Xerri C, Merzenich MM, Jenkins W & Santucci S Representational plasticity in cortical area 3b paralleling tactual-motor skill acquisition in adult monkeys. Cereb. Cortex 9, 264–276 (1999). [DOI] [PubMed] [Google Scholar]
  • 87.Qi HX, Chen LM & Kaas JH Reorganization of somatosensory cortical areas 3b and 1 after unilateral section of dorsal columns of the spinal cord in squirrel monkeys. J. Neurosci 31, 13662–13675 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 88.Shadmehr R & Mussa-Ivaldi FA Adaptive representation of dynamics during learning of a motor task. J. Neurosci 14, 3208–3224 (1994). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 89.Krakauer JW, Pine ZM, Ghilardi MF & Ghez C Learning of visuomotor transformations for vectorial planning of reaching trajectories. J. Neurosci 20, 8916–8924 (2000). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90.Cunningham HA Aiming error under transformed spatial mappings suggests a structure for visual-motor maps. J. Exp. Psychol. Hum. Percept. Perform 15, 493–506 (1989). [DOI] [PubMed] [Google Scholar]
  • 91.Ostry DJ, Darainy M, Mattar AA, Wong J & Gribble PL Somatosensory plasticity and motor learning. J. Neurosci 30, 5384–5393 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 92.Cressman EK & Henriques DYP Sensory recalibration of hand position following visuomotor adaptation. J. Neurophysiol 102, 3505–3518 (2009). [DOI] [PubMed] [Google Scholar]
  • 93.Nasir SM, Darainy M & Ostry DJ Sensorimotor adaptation changes the neural coding of somatosensory stimuli. J. Neurophysiol 109, 2077–2085 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 94.Lebedev MA et al. Cortical ensemble adaptation to represent velocity of an artificial actuator controlled by a brain-machine interface. J. Neurosci 25, 4681–4693 (2005). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 95.Fetz EE & Finocchio DV Operant conditioning of specific patterns of neural and muscular activity. Science 174, 431–435 (1971). [DOI] [PubMed] [Google Scholar]
  • 96.Moritz CT & Fetz EE Volitional control of single cortical neurons in a brain-machine interface. J. Neural Eng 8, 025017 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 97.Kennedy PR, Bakay RA, Moore MM, Adams K & Goldwaithe J Direct control of a computer from the human central nervous system. IEEE Trans. Rehabil. Eng 8, 198–202 (2000). [DOI] [PubMed] [Google Scholar]
  • 98.Ochoa J & Torebjork E Sensations evoked by intraneural microstimulation of single mechanoreceptor units innervating the human hand. J. Physiol 342, 633–654 (1983). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 99.Wheat HE, Goodwin AW & Browning AS Tactile resolution: peripheral neural mechanisms underlying the human capacity to determine positions of objects contacting the fingerpad. J. Neurosci 15, 5582–5595 (1995). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 100.Knibestol M Stimulus-response functions of slowly adapting mechanoreceptors in the human glabrous skin area. J. Physiol 245, 63–80 (1975). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 101.Macefield VG, Hager-Ross C & Johansson RS Control of grip force during restraint of an object held between finger and thumb: responses of cutaneous afferents from the digits. Exp. Brain Res 108, 155–171 (1996). [DOI] [PubMed] [Google Scholar]
  • 102.Goodwin AW & Wheat HE Sensory signals in neural populations underlying tactile perception and manipulation. Annu. Rev. Neurosci 27, 53–77 (2004). [DOI] [PubMed] [Google Scholar]
  • 103.Johansson RS & Westling G Roles of glabrous skin receptors and sensorimotor memory in automatic control of precision grip when lifting rougher or more slippery objects. Exp. Brain Res 56, 550–564 (1984). [DOI] [PubMed] [Google Scholar]
  • 104.Monzée J, Lamarre Y & Smith AM The effects of digital anesthesia on force control using a precision grip. J. Neurophysiol 89, 672–683 (2003). [DOI] [PubMed] [Google Scholar]
  • 105.Brochier T, Boudreau MJ, Paré M & Smith AM The effects of muscimol inactivation of small regions of motor and somatosensory cortex on independent finger movements and force control in the precision grip. Exp. Brain Res 128, 31–40 (1999). [DOI] [PubMed] [Google Scholar]
  • 106.Sainburg RL, Ghilardi MF, Poizner H & Ghez C Control of limb dynamics in normal subjects and patients without proprioception. J. Neurophysiol 73, 820–835 (1995). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 107.Botvinick M & Cohen J Rubber hands “feel” touch that eyes see. Nature 391, 756 (1998). [DOI] [PubMed] [Google Scholar]
  • 108.Dunbar RI The social role of touch in humans and primates: behavioural function and neurobiological mechanisms. Neurosci. Biobehav. Rev 34, 260–268 (2010). [DOI] [PubMed] [Google Scholar]
  • 109.Blabe C et al. Assessing the brain–machine interface priiorities from the perspective of spinal cord injury participants. Soc. Neurosci. Abstr 584.14 (2012). [Google Scholar]
  • 110.Penfield W & Boldrey E Somatic motor and sensory representation in the cerebral cortex of man as studied by electrical stimulation. Brain 60, 389–443 (1937). [Google Scholar]
  • 111.Rasmussen T & Penfield W The human sensorimotor cortex as studied by electrical stimulation. Fed. Proc 6, 184 (1947). [PubMed] [Google Scholar]
  • 112.Romo R, Hernandez A, Zainos A & Salinas E Somatosensory discrimination based on cortical microstimulation. Nature 392, 387–390 (1998). [DOI] [PubMed] [Google Scholar]; In this landmark study, monkeys were trained to discriminate the frequencies of sequential mechanical vibrations applied to the fingertip. They were then able to discriminate the frequency of ICMS pulse trains delivered to S1 when substituted for one or both of the mechanical stimuli.
  • 113.Fridman G, Blair H, Blaisdell A & Judy J Perceived intensity of somatosensory cortical electrical stimulation. Exp. Brain Res 203, 499–515 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 114.Semprini M, Bennicelli L & Vato A A parametric study of intracortical microstimulation in behaving rats for the development of artificial sensory channels. Conf. Proc. IEEE Eng. Med. Biol. Soc 2012, 799–802 (2012). [DOI] [PubMed] [Google Scholar]
  • 115.Zaaimi B, Ruiz-Torres R, Solla SA & Miller LE Multi-electrode stimulation in somatosensory cortex increases probability of detection. J. Neural Engineer 10, 056013 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 116.O’Connor DH et al. Neural coding during active somatosensation revealed using illusory touch. Nature Neurosci. 16, 958–965 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 117.Ramachandran VS & Hirstein W The perception of phantom limbs. The D. O. Hebb lecture. Brain 121, 1603–1630 (1998). [DOI] [PubMed] [Google Scholar]; This paper provides an overview of the phenomenon of the phantom limb, which in turn sheds light on how somatosensory representations evolve after deafferentation.
  • 118.Berg JA et al. Behavioral demonstration of a somatosensory neuroprosthesis. IEEE Trans. Neural Syst. Rehabil. Eng 21, 500–507 (2013). [DOI] [PubMed] [Google Scholar]
  • 119.Pei YC, Hsiao SS, Craig JC & Bensmaia SJ Shape invariant coding of motion direction in somatosensory cortex. PLoS Biol. 8, e1000305 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 120.Bensmaia SJ, Denchev PV, Dammann JF, Craig JC & Hsiao SS The representation of stimulus orientation in the early stages of somatosensory processing. J. Neurosci 28, 776–786 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 121.Costanzo RM & Gardner EP Multiple-joint neurons in somatosensory cortex of awake monkeys. Brain Res. 214, 321–333 (1981). [DOI] [PubMed] [Google Scholar]
  • 122.Prud’homme MJL & Kalaska JF Proprioceptive activity in primate primary somatosensory cortex during active arm reaching movements. J. Neurophysiol 72, 2280–2301 (1994). [DOI] [PubMed] [Google Scholar]
  • 123.London BM & Miller LE Responses of somatosensory area 2 neurons to actively and passively generated limb movements. J. Neurophysiol 109, 1505–1513 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 124.Histed MH, Bonin V & Reid RC Direct activation of sparse, distributed populations of cortical neurons by electrical microstimulation. Neuron 63, 508–522 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 125.Logothetis NK et al. The effects of electrical microstimulation on cortical signal propagation. Nature Neurosci. 13, 1283–1291 (2010). [DOI] [PubMed] [Google Scholar]
  • 126.Mountcastle VB & Powell TP Central nervous mechanisms subserving position sense and kinesthesis. Bull. Johns Hopkins Hosp 105, 173–200 (1959). [PubMed] [Google Scholar]
  • 127.Diester I et al. An optogenetic toolbox designed for primates. Nature Neurosci. 14, 387–397 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 128.Cardin JA et al. Targeted optogenetic stimulation and recording of neurons in vivo using cell-type-specific expression of Channelrhodopsin-2. Nature Protoc. 5, 247–254 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 129.Thomson EE, Carra R & Nicolelis MA Perceiving invisible light through a somatosensory cortical prosthesis. Nature Commun. 4, 1482 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 130.Marasco PD, Schultz AE & Kuiken TA Sensory capacity of reinnervated skin after redirection of amputated upper limb nerves to the chest. Brain 132, 1441–1448 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 131.Marasco PD, Kim K, Colgate JE, Peshkin MA & Kuiken TA Robotic touch shifts perception of embodiment to a prosthesis in targeted reinnervation amputees. Brain 134, 747–758 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 132.Shokur S et al. Expanding the primate body schema in sensorimotor cortex by virtual touches of an avatar. Proc. Natl Acad. Sci. USA 110, 15121–15126 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 133.Ince NF et al. High accuracy decoding of movement target direction in non-human primates based on common spatial patterns of local field potentials. PLoS ONE 5, e14384 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 134.Rickert J et al. Encoding of movement direction in different frequency ranges of motor cortical local field potentials. J. Neurosci 25, 8815–8824 (2005). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 135.Andersen RA, Musallam S & Pesaran B Selecting the signals for a brain-machine interface. Curr. Opin. Neurobiol 14, 720–726 (2004). [DOI] [PubMed] [Google Scholar]
  • 136.Markowitz DA, Wong YT, Gray CM & Pesaran B Optimizing the decoding of movement goals from local field potentials in macaque cortex. J. Neurosci 31, 18412–18422 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 137.Pesaran B, Pezaris JS, Sahani M, Mitra PP & Andersen RA Temporal structure in neuronal activity during working memory in macaque parietal cortex. Nature Neurosci. 5, 805–811 (2002). [DOI] [PubMed] [Google Scholar]
  • 138.Dangi S, Orsborn AL, Moorman HG & Carmena JM Design and analysis of closed-loop decoder adaptation algorithms for brain–machine interfaces. Neural Comput. 25, 1693–1731 (2013). [DOI] [PubMed] [Google Scholar]
  • 139.Buzsáki G, Anastassiou CA & Koch C The origin of extracellular fields and currents — EEG, ECoG, LFP and spikes. Nature Rev. Neurosci 13, 407–420 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 140.Kim SP et al. Divide-and-conquer approach for brain machine interfaces: nonlinear mixture of competitive linear models. Neural Netw. 16, 865–871 (2003). [DOI] [PubMed] [Google Scholar]
  • 141.Kim SP et al. A comparison of optimal MIMO linear and nonlinear models for brain-machine interfaces. J. Neural Eng 3, 145–161 (2006). [DOI] [PubMed] [Google Scholar]
  • 142.Shoham S et al. Statistical encoding model for a primary motor cortical brain-machine interface. IEEE Trans. Biomed. Eng 52, 1312–1322 (2005). [DOI] [PubMed] [Google Scholar]
  • 143.Yu BM et al. Mixture of trajectory models for neural decoding of goal-directed movements. J. Neurophysiol 97, 3763–3780 (2007). [DOI] [PubMed] [Google Scholar]
  • 144.Yu B, Cunningham J, Shenoy K & Sahani M in Neural Information Processing; Lecture Notes in Computer Science Vol. 4984 (eds Ishikawa M, Doya K, Miyamoto H & Yamakawa T) 586–595 (Springer, 2008). [Google Scholar]
  • 145.Palmer SS & Fetz EE Discharge properties of prima te forearm motor units during isometric muscle activity. J. Neurophysiol 54, 1178–1193 (1985). [DOI] [PubMed] [Google Scholar]
  • 146.Sergio LE, Hamel-Paquet C & Kalaska JF Motor cortex neural correlates of output kinematics and kinetics during isometric-force and arm-reaching tasks. J. Neurophysiol 94, 2353–2378 (2005). [DOI] [PubMed] [Google Scholar]
  • 147.Fetz EE Are movement parameters recognizably coded in activity of single neurons? Behav. Brain Sci 15, 679–690 (1992). [Google Scholar]
  • 148.Phillips CG Laying the ghost of ‘muscles versus movements’. Can. J. Neurol. Sci 2, 209–218 (1975). [DOI] [PubMed] [Google Scholar]
  • 149.Scott SH & Kalaska JF Reaching movements with similar hand paths but different arm orientations. I. Activity of individual cells in motor cortex. J. Neurophysiol 77, 826–852 (1997). [DOI] [PubMed] [Google Scholar]
  • 150.Caminiti R, Johnson PB & Urbano A Making arm movements within different parts of space: dynamic aspects in the primate motor cortex. J. Neurosci 10, 2039–2058 (1990). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 151.Kakei S, Hoffman DS & Strick PL Muscle and movement representations in the primary motor cortex. Science 285, 2136–2139 (1999). [DOI] [PubMed] [Google Scholar]
  • 152.Oby ER, Ethier C & Miller LE Movement representation in the primary motor cortex and its contribution to generalizable EMG predictions. J. Neurophysiol 109, 666–678 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 153.Cherian A, Krucoff MO & Miller LE Motor cortical prediction of EMG: evidence that a kinetic brain-machine interface may be robust across altered movement dynamics. J. Neurophysiol 106, 564–575 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 154.Gupta R & Ashe J Lack of adaptation to random conflicting force fields of variable magnitude. J. Neurophysiol 97, 738–745 (2007). [DOI] [PubMed] [Google Scholar]
  • 155.Hatsopoulos N, Joshi J & O’Leary JG Decoding continuous and discrete motor behaviors using motor and premotor cortical ensembles. J. Neurophysiol 92, 1165–1174 (2004). [DOI] [PubMed] [Google Scholar]
  • 156.Hauschild M, Mulliken GH, Fineman I, Loeb GE & Andersen RA Cognitive signals for brain–machine interfaces in posterior parietal cortex include continuous 3D trajectory commands. Proc. Natl Acad. Sci. USA 109, 17075–17080 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 157.Pesaran B, Nelson MJ & Andersen RA Dorsal premotor neurons encode the relative position of the hand, eye, and goal during reach planning. Neuron 51, 125–134 (2006). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 158.Musallam S, Corneil BD, Greger B, Scherberger H & Andersen RA Cognitive control signals for neural prosthetics. Science 305, 258–262 (2004). [DOI] [PubMed] [Google Scholar]
  • 159.Ganguly K & Carmena JM Neural correlates of skill acquisition with a cortical brain-machine interface. J. Motor Behav 42, 355–360 (2010). [DOI] [PubMed] [Google Scholar]
  • 160.Heliot R, Venkatraman S & Carmena JM Decoder remapping to counteract neuron loss in brain-machine interfaces. Conf. Proc. IEEE Eng. Med. Biol. Soc 2010, 1670–1673 (2010). [DOI] [PubMed] [Google Scholar]
  • 161.Orsborn AL, Dangi S, Moorman HG & Carmena JM Closed-loop decoder adaptation on intermediate time-scales facilitates rapid BMI performance improvements independent of decoder initialization conditions. IEEE Trans. Neural Syst. Rehabil. Eng 20, 468–477 (2012). [DOI] [PubMed] [Google Scholar]
  • 162.Fishbach A, Roy S, Bastianen C, Miller L & Houk J Kinematic properties of on-line error corrections in the monkey. Exp. Brain Res 164, 442–457 (2005). [DOI] [PubMed] [Google Scholar]
  • 163.Danziger Z, Fishbach A & Mussa-Ivaldi FA Learning algorithms for human–machine interfaces. IEEE Trans. Biomed. Eng 56, 1502–1511 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 164.Mahmoudi B & Sanchez JCA Symbiotic brain-machine interface through value-based decision making. PLoS ONE 6, e14760 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 165.Mahmoudi B, Pohlmeyer EA, Prins NW, Geng S & Sanchez JC Towards autonomous neuroprosthetic control using Hebbian reinforcement learning. J. Neural Eng 10, 066005 (2013). [DOI] [PubMed] [Google Scholar]
  • 166.Judy JW Neural interfaces for upper-limb prosthesis control: opportunities to improve long-term reliability. IEEE Pulse 3, 57–60 (2012). [DOI] [PubMed] [Google Scholar]
  • 167.Perge JA et al. Intra-day signal instabilities affect decoding performance in an intracortical neural interface system. J. Neural Eng 10, 036004 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 168.Polikov VS, Tresco PA & Reichert WM Response of brain tissue to chronically implanted neural electrodes. J. Neurosci. Methods 148, 1–18 (2005). [DOI] [PubMed] [Google Scholar]
  • 169.Prasad A et al. Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants. J. Neural Eng 9, 056015 (2012). [DOI] [PubMed] [Google Scholar]
  • 170.McCreery D, Pikov V & Troyk P Neuronal loss due to prolonged controlled-current stimulation with chronically implanted microelectrodes in the cat cerebral cortex. J. Neural Eng 7, 036005 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 171.Parker RA, Davis TS, House PA, Normann RA & Greger B The functional consequences of chronic, physiologically effective intracortical microstimulation. Prog. Brain Res 194, 145–165 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 172.Chen KH et al. The effect of chronic intracortical microstimulation on the electrode–tissue interface. J. Neural Eng 11, 026004 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 173.Kane S et al. Electrical performance of penetrating microelectrodes chronically implanted in cat cortex. IEEE Trans. Biomed. Eng 60, 2153–2160 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 174.Cogan SF Neural stimulation and recording electrodes. Annu. Rev. Biomed. Eng 10, 275–309 (2008). [DOI] [PubMed] [Google Scholar]
  • 175.Simeral JD et al. Some preliminary longitudinal findings from five trial participants using the BrainGate neural interface system. Soc. Neurosci. Abstr 142.04 (2011). [Google Scholar]
  • 176.Grill WM, Norman SE & Bellamkonda RV Implanted neural interfaces: biochallenges and engineered solutions. Annu. Rev. Biomed. Eng 11, 1–24 (2009). [DOI] [PubMed] [Google Scholar]
  • 177.Reichert WM Indwelling Neural Implants: Strategies for Contending with the In Vivo Environment (CRC, 2010). [PubMed] [Google Scholar]
  • 178.Sridharan A, Rajan SD & Muthuswamy J Long-term changes in the material properties of brain tissue at the implant–tissue interface. J. Neural Eng 10, 066001 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 179.Harris JP et al. In vivo deployment of mechanically adaptive nanocomposites for intracortical microelectrodes. J. Neural Eng 8, 046010 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 180.Ware T, Simon D, Rennaker RL & Voit W Smart polymers for neural interfaces. Polymer Rev. 53, 108–129 (2013). [Google Scholar]
  • 181.Capadona JR, Tyler DJ, Zorman CA, Rowan SJ & Weder C Mechanically adaptive nanocomposites for neural interfacing. MRS Bull. 37, 581–589 (2012). [Google Scholar]
  • 182.Lind G, Linsmeier CE, Thelin J & Schouenborg J Gelatine-embedded electrodes—a novel biocompatible vehicle allowing implantation of highly flexible microelectrodes. J. Neural Engineer 7, 046005 (2010). [DOI] [PubMed] [Google Scholar]
  • 183.Khodagholy D et al. Highly conformable conducting polymer electrodes for in vivo recordings. Adv. Mater 23, H268–H272 (2011). [DOI] [PubMed] [Google Scholar]
  • 184.Kim T, Branner A, Gulati T & Giszter SF Braided multi-electrode probes: mechanical compliance characteristics and recordings from spinal cords. J. Neural Eng 10, 045001 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 185.Rousche PJ et al. Flexible polyimide-based intracortical electrode arrays with bioactive capability. IEEE Trans. Biomed. Eng 48, 361–371 (2001). [DOI] [PubMed] [Google Scholar]
  • 186.Moxon KA et al. Nanostructured surface modification of ceramic-based microelectrodes to enhance biocompatibility for a direct brain-machine interface. IEEE Trans. Biomed. Eng 51, 881–889 (2004). [DOI] [PubMed] [Google Scholar]
  • 187.Kim D-H & Martin DC Sustained release of dexamethasone from hydrophilic matrices using PLGA nanoparticles for neural drug delivery. Biomaterials 27, 3031–3037 (2006). [DOI] [PubMed] [Google Scholar]
  • 188.Cho Y & Ben Borgens R Electrically controlled release of the nerve growth factor from a collagen-carbon nanotube composite for supporting neuronal growth. J. Mater. Chem. B 1, 4166–4170 (2013). [DOI] [PubMed] [Google Scholar]
  • 189.Keefer EW, Botterman BR, Romero MI, Rossi AF & Gross GW Carbon nanotube coating improves neuronal recordings. Nature Nanotechnol. 3, 434–439 (2008). [DOI] [PubMed] [Google Scholar]
  • 190.Kotov NA et al. Nanomaterials for Neural Interfaces. Adv. Mater 21, 3970–4004 (2009). [Google Scholar]
  • 191.Stauffer WR & Cui XT Polypyrrole doped with 2 peptide sequences from laminin. Biomaterials 27, 2405–2413 (2006). [DOI] [PubMed] [Google Scholar]
  • 192.Malarkey E & Parpura V in Brain Edema XIV Vol. 106 Acta Neurochirurgica Supplementum (eds Czernicki Z et al. ) 337–341 (Springer, 2010). [Google Scholar]
  • 193.Kenney C et al. Short-term and long-term safety of deep brain stimulation in the treatment of movement disorders. J. Neurosurg 106, 621–625 (2007). [DOI] [PubMed] [Google Scholar]
  • 194.Rizk M et al. A fully implantable 96-channel neural data acquisition system. J. Neural Eng 6, 026002 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 195.Zhang F, Aghagolzadeh M & Oweiss KA Fully implantable, programmable and multimodal neuroprocessor for wireless, cortically controlled brain-machine interface applications. J. Signal Process. Syst 69, 351–361 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 196.Borton DA, Yin M, Aceros J & Nurmikko A An implantable wireless neural interface for recording cortical circuit dynamics in moving primates. J. Neural Eng 10, 026010 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 197.Harrison RR The design of integrated circuits to observe brain activity. Proc. IEEE 96, 1203–1216 (2008). [Google Scholar]

RESOURCES