Table 5.
Relative deviations of neutron beam parameters at the BSA collimator exit compared to original ones after replacing mg, F, pb, and Li neutron reaction cross-sections in ENDF/B VIII.1 library with corresponding data from ENDF/B VII.1 library.
| Neutron libraries |
(%) |
(%) |
(%) |
(%) |
(%) |
|
|---|---|---|---|---|---|---|
| ENDF/B VIII.1 | ENDF/B VII.1(F) | −7.19 | 0.23 | 2.17 | −21.10 | 0.81 |
| ENDF/B VII.1(Mg) | −0.01 | 0.00 | 0.13 | 0.01 | 0.03 | |
| ENDF/B VII.1(Pb) | −0.80 | −0.03 | 0.08 | 0.79 | −2.41 | |
| ENDF/B VII.1(Li) | −0.01 | 0.00 | 0.03 | 0.01 | 0.01 | |
| ENDF/B VII.1(F + Mg) | −7.89 | 0.19 | 1.93 | −20.36 | −1.82 | |
| ENDF/B VII.1 | −7.87 | 0.19 | 2.38 | −20.28 | −1.55 | |