Skip to main content
. 2005 Aug;43(8):3829–3834. doi: 10.1128/JCM.43.8.3829-3834.2005

TABLE 2.

Average manipulation time of VITEK 2 and Phoenix (calculated for 39 batches, each with seven isolates)

Workflow VITEK 2
Phoenix
Step(s) Time (s) Step(s) Time (s)
Preprocessing Organize bench with supplies 19 Organize bench with supplies 19
Label purity plates 15 Label purity plates 16
Label tubes with salinea 44 Label ID brotha 56
Dispense saline 12
Inoculum adjustment Make suspension and check with DensiChek or vortex 230 Inoculate ID broth and check with crystal spec or vortex 462
Panel preparation Inoculate purity plate 47 Inoculate purity plate 67
Open package 63
Drop indicator in AST broth 67
Transfer 25 μl from ID to AST broth 137
Open package and label panels 107
Place ID and AST broth in inoculation station and inoculate panels with ID and AST broth 96
Close the panels 32
Put panels in transport caddy 17
System loading Enter accession no., read barcode on SCSb, put card on SCS rack 138 Enter accession number and put panels in Phoenix 113
Put carrier in VITEK2 9
Postprocessing Put purity plates in incubator 46 Put purity plates in incubator 46
Remove panels from instrument and throw away 13 Remove panels from instrument and throw away 17
Total per batch 636 1,252
a

Labeling was done for the study but is optional for routine procedures.

b

SCS, smart carrier station.