Abstract
Introduction:
Over 52% of people who use a lower-limb prosthesis fall once every year, but fall risk is still not effectively screened. Few studies have identified clinical outcome measures that could help screen fall risk. Gait asymmetries between the intact and prosthetic limbs could also potentially help identify fall risk, based on findings in people with stroke. However, no studies have examined the relationship between gait asymmetries and falls in people who use a lower-limb prosthesis. Therefore, we sought to determine if any gait asymmetry parameters could significantly distinguish 12-month retrospective fallers from non-fallers.
Methods:
People were recruited from private practice and the Amputee Coalition National Conference. Participants completed a 12-month retrospective fall history and seven clinical outcome measures, as well as level-ground walking at self-selected pace with wearable sensors that collected kinematic (sagittal plane: hip, knee, and ankle range of motion) and kinetic (peak braking and propulsion ground reaction force) data.
Results:
Twenty-two individuals who use a unilateral prosthesis participated (age 57.6 ± 4.2 years; 15 transtibial, 7 transfemoral). No gait asymmetry parameters significantly distinguished 12-month retrospective fallers from non-fallers. However, the Four Square Step Test did significantly distinguish fallers from non-fallers (p=0.040, Hedge’s g=−0.739, area under the curve= 0.725, CI= 50–95%, cut-off time= 13.14 seconds), irrespective of level of prosthesis use. No parameters significantly distinguished fallers from non-fallers by level of prosthesis use (transtibial, transfemoral).
Conclusions:
The Four Square Step Test cut-off time may be useful to distinguish fallers from non-fallers in unilateral lower-limb prosthesis users.
Keywords: artificial limbs, accidental falls, lower extremity amputees, biomechanics, outcome measures
1. Introduction
Individuals who use unilateral lower-limb prostheses have higher fall risks compared to the general population.1 Fall risks are defined as the likelihood of whether or not a fall will occur.1 Specifically, community-dwelling adults 18+ years of age with lower-limb amputation are estimated to have similar or higher fall risks compared to older adults 65+ years of age without amputation.1 Further, over 52% of individuals who use unilateral lower-limb prostheses fall per year.2 Falls in this population have been associated with injury,1–3 decreased mobility,4 and reduced quality of life.5 Despite this knowledge, people in this population are still not effectively screened for fall risk in clinical practice.6–10
A recent study found the Four Square Step Test (FSST), Timed Up and Go (TUG), and Amputee Mobility Predictor with Prosthesis (AMPPRO) could significantly distinguish fallers from non-fallers who use unilateral lower-limb prostheses based on a 12-month retrospective recall of fall history.11 Further, retrospective fall history has recently been shown to be useful in predicting future falls in this population,10 and growing pressures to justify prosthetic componentry for reimbursement might necessitate increased outcome measure use. Thus, simple and quick clinical outcome measures could help clinicians identify fall risk in clinical practice.6–8,10–11
Gait parameters could also be effective at distinguishing fallers from non-fallers in people who use a unilateral lower-limb prosthesis, but no studies have investigated this relationship.8 In the general population of older adults and stroke survivors, recent studies have suggested gait asymmetry, specifically at peak ankle flexion/extension during swing and push off phases while walking on level ground, could be used to distinguish fallers from non-fallers.12–14 Since people who use a unilateral lower-limb prosthesis can have similar, and even increased amounts of sagittal plane gait asymmetry and unilateral mechanics due to prosthesis use,14 lower-limb asymmetry parameters might also help distinguish fallers from non-fallers in people who use lower-limb prostheses. Additionally, people who use a transfemoral prosthesis have demonstrated both a higher risk of falling post-rehabilitation and greater gait asymmetry than those who use a transtibial prosthesis,15,16 indicating that gait asymmetry could be a useful parameter to distinguish fall risks between people who use a transfemoral and transtibial prosthesis.
Therefore, this study aimed to determine if any gait asymmetry parameters or clinical outcome measures could distinguish fallers from non-fallers, based on 12-month retrospective fall history. It was hypothesized that walking asymmetry data would: 1) be able to significantly distinguish fallers from non-fallers, and 2) be more useful at significantly distinguishing fallers from non-fallers than clinical outcome measures, defined as having greater levels of significance with higher sensitivity and specificity. Further, it was hypothesized that the above would be true irrespective of level of prosthesis use and for both transtibial and transfemoral prosthesis users separately.
2. Methods
2.1. Participants:
The North Texas Regional Institutional Review Board (#2020-048), approved this study. Individuals who use unilateral transtibial or transfemoral prostheses were recruited from prosthetics clinics in the [redacted] metroplex and the Amputee Coalition National Conference. Individuals were included if they were: between 18 and 95 years of age, walking with a prosthesis for at least one year, and able to walk independently for at least five minutes or 100 yards. Individuals were excluded if they had: pain, open wounds or discomfort on their lower limbs or trunk on the day of testing, limb difference on any other limbs (e.g. bilateral lower-limb loss, an upper-limb loss), major musculoskeletal injury or surgery in the last year, or other comorbidities that would make standing, turning, or walking unsafe.
2.2. Equipment:
Seven inertial measurement unit (IMU) sensors (iSen, STT Systems, San Sebastian, Spain) were used to collect and store lower-body IMU data.14 Studies have provided evidence of reliability and validity for IMUs relative to motion capture equipment in this population.14,17 IMUs were secured to each participant by the same researcher (first author, [initials redacted]) using recommendations from STT Systems for the lower-body model: one on the sacrum, one on each thigh, one on each shank, and one on each foot, according to Figure 1. IMUs were secured to body segments with elastic straps, except the foot IMUs which were secured directly to the subject’s shoe with velcro. Additionally, the thigh IMUs on transfemoral prosthetic limbs were secured directly to the prosthesis using self-adherent wrap (Coban 1580 Series, 3M™, Saint Paul, MN) to prevent sliding. For transfemoral participants, the IMU magnetometers were disabled during data collection to prevent magnetic interference from the microprocessor knees.
Figure 1:

Anterior (panels A and C) and posterior (panels B and D) views of inertial measurement unit and pressure insole equipment placed on a representative participant who uses a transtibial prosthesis.
Pressure insoles (Loadsol, Novel, Saint Paul, Minnesota, USA), with evidence of reliability and validity in prosthesis users, were also used to collect ground reaction force data.18,19 Each insole contained three pressure sensors on the medial, lateral, and heel aspects of the foot. Each participant was assessed for the proper size of pressure insoles, which were in standard US sizes. Pressure insoles were less than 3.4mm thick, worn bilaterally, and the same researcher (first author, [initials redacted]) placed the insole between the participant’s foot and their shoe. All data were collected at a sampling rate of 100Hz.
2.3. Data Collection:
Data was collected during a single in-person visit by the first author [initials redacted] at either the University of [redacted] or the Amputee Coalition National Conference. After participants provided informed consent, they completed demographic information and prosthetic componentry information. Then, they completed a 12-month retrospective fall history, self-report clinical outcome measure questionnaires, and functional clinical outcome measures. Participants were then outfitted with IMUs and pressure insoles to complete several walking trials.
Research Electronic Data Capture (REDCap) software hosted at the University of [redacted] was used to collect and manage all data (Supplemental Document 1).20 Data was collected by the first author [initials redacted] with an iPad (Apple, Cupertino, California, USA).
Prosthetic componentry, residual limb length, and K-levels were determined by a certified Prosthetist and Orthotist (MGF). K-levels were determined based on the Amputee Mobility Predictor with Prosthesis (AMPPRO)21 assessment and the prosthetic componentry worn at the test session.
Retrospective fall histories were collected to define faller and non-faller groupings. Participants were asked if they had experienced a fall in the previous 12 months, and if so, how many they had experienced. Falls were defined as “a loss of balance where your body landed on the ground or floor” (page 10).22 Participants were further asked to describe their most memorable fall (if multiple), such as, if they received medical treatment due to the fall, changed their activities or ways in which they interact with their environment, and had a change in confidence during daily activities of living. Additionally, participants were asked if they experienced any near-falls, defined as “a loss of balance where you caught yourself or recovered without landing on the ground or floor” (page 10).22
Several self-report clinical outcome measure questionnaires were completed as potential identifiers of fall risk: the Socket Comfort Score (SCS),23 Patient Reported Outcomes Measurement Information System-29 v2.0 profile (PROMIS-29),24 the Prosthetic Limb Users Survey of Mobility (PLUS-M) 12-Item Short Form,25 and the Activities-Specific Balance Confidence Scale (ABC).26 These self-report measures were selected because they have been shown to have acceptable psychometric properties, such as test-retest reliability, construct validity, and measurement error in individuals who use lower-limb prostheses.24
Additionally, functional clinical outcome measures were also completed as potential identifiers of fall risk including the AMPPRO, Timed Up and Go (TUG) task,27 and the Four Square Step Test (FSST).28
To capture gait asymmetry data, participants walked across 6 meters of level ground at their self-selected habitual walking speed for five trials. Participants completed walking data in comparable ways (same walkway, private area) on a tile floor at the University of [redacted] and a low carpeted floor at the Amputee Coalition National Conference.
2.4. Data analysis:
IMU data were processed within the proprietary iSen software provided by STT Systems to define lower-limb joint angles, and ground reaction force data were processed within proprietary Loadsol software.
For each participant, three of their five walking trials were selected for analysis based on visual observation that the joint angles and iSen model were appropriate (e.g. both legs were shown to be forward-facing and moved forwards in the sagittal plane during the walking trial). Then, the three middle strides from each trial for both the prosthetic and intact limbs were extracted to minimize effects from the participant accelerating or decelerating, and gait events were used to determine single and double-limb support. Therefore, a total of nine strides (three walking trials × three middle strides) were used for each limb (prosthetic, intact) for analysis. For IMU-derived data, strides were extracted based on the automatic detection of initial contact and toe-off events identified within iSen IMU software. This was previously validated against motion capture detection for accuracy.14 For Loadsol data, steps were extracted based on evidence of the first non-zero point (initial contact) and last non-zero point (toe-off).29 Loadsol and IMU events were compared and consistently within 2–3 frames (0.02–0.03s) of each other.14
Peak flexion and extension values during each stride at the hip, knee, and ankle were extracted for both prosthetic and intact limbs. Range of motion was calculated as the difference between peak flexion and extension for each stride and limb. Peak braking and propulsive force were also extracted for both limbs. The equation below was used to calculate asymmetry between intact (I) and prosthetic (P) limbs for both kinematic and kinetic measures.14,16 Asymmetry values were calculated for each stride, then averaged separately for each limb (prosthetic, intact).
| Eq. 1 |
For our primary analysis, participants were stratified into two groups, fallers and non-fallers, irrespective of level of prosthesis use, based on prior studies.6,7 Participants were also grouped as fallers and non-fallers by level of prosthesis use to examine potential differences by transtibial or transfemoral prosthesis use. A faller was defined as any participant who reported experiencing 1 or more falls over the past 12 months based on retrospective recall, while a non-faller reported experiencing no falls over the past 12 months. The Shapiro-Wilks test was used to determine normality of the data. Significance between fallers and non-fallers was assessed using Mann-Whitney U tests with a significance level set at p ≤ 0.05. For each clinical outcome measure or gait parameter that could significantly distinguish fallers from non-fallers, cut-off times were determined using area under the curve and receiver operating characteristic (AUC/ROC) analysis. For AUC analysis, a clinical threshold of 0.800 or greater is typically considered to show clinical acceptability.7 Hedge’s g was also calculated to demonstrate effect sizes, where a threshold of 0.8 or greater indicates a large effect size. Effect sizes provide an indication of the generalizability of the results, given a larger sample size. A larger effect size value indicates a higher likelihood that the parameter would be statistically significant if the sample size in this study were larger. Positive effect size values represent lower scores, slower times, and less asymmetry in fallers, compared to non-fallers.
3. Results
3.1. Irrespective of Prosthesis Use
Table 1 presents demographic characteristics, clinical outcome measure scores, and gait data stratified by fall status irrespective of the level of prosthesis use (n = 22; fallers n = 10, non-fallers n = 12). The only demographic characteristic that was statistically significantly different between fallers and non-fallers, as expected, was the average number of falls participants experienced (0 for non-fallers; 3.6 for fallers; p=0.012). For clinical outcome measures, only the FSST significantly distinguished fallers from non-fallers (p=0.040, g=−0.739), with an AUC of 0.725, optimal cut-off time of ≥13.14s, sensitivity of 80%, and specificity of 75% (confidence interval 50% – 95%). For gait parameters, none were statistically significantly different between groups. While not statistically significant, Hedge’s g effect size was medium for the Global Pain scale (g=−0.781) and the TUG task (g=−0.562)), as well as the following gait parameters: (walking speed (g=0.590), ankle dorsiflexion asymmetry (g=−0.663), ankle range of motion asymmetry (g=−0.620), peak propulsive asymmetry(g=−0.616)).
Table 1:
Demographic characteristics, clinical outcome measure scores, and gait data stratified by non-fallers and fallers irrespective of level of prosthesis use.
| Demographics | Non-Fallers (n=12) | Fallers (n=10) | p-values | Hedge’s g |
|---|---|---|---|---|
| Age (yrs) | 57.5 ± 15.0 | 56.8 ± 13.8 | 0.461 | 0.048 |
| Stature (cm) | 175.9 ± 8.8 | 173.0 ± 10.2 | 0.286 | 0.307 |
| Body Mass (kg) | 90.0 ± 25.4 | 92.1 ± 21.6 | 0.461 | −0.088 |
| Sex | 7 M / 5 F | 8 M / 2 F | 0.144 | - |
| Race | 10 White/1 Prefer not to answer | 10 White | 0.915 | - |
| Ethnicity | 1 Hispanic or Latine/10 Not Hispanic or Latine | 0 Hispanic or Latine/10 Not Hispanic or Latine | 0.915 | - |
| TSA (yrs) | 19.0 ± 17.9 | 14.0 ± 14.3 | 0.265 | 0.305 |
| Etiology | 6 Tr / 2 Vasc / 3 Can / 1 Cong | 4 Tr / 3 Vasc / 2 Can / 1 Cong | 0.403 | - |
| Residual Limb length (cm) | 18.2 ± 4.0 | 17.3 ± 5.0 | 0.322 | 0.201 |
| Level of Prosthesis Use | 8 TT / 4 TF | 7 TT / 3 TF | 0.435 | - |
| K-Level | 9 K3 / 3 K2 | 8 K3 / 2 K2 | 0.497 | - |
| Time using current prosthesis (prosthesis used during study) (yrs) | 4.0 ± 2.6 | 3.3 ± 2.6 | 0.251 | 0.269 |
| Last prosthetic adjustment (months) | 3.1 ± 3.6 | 1.0 ± 0.5 | 0.185 | 0.780 |
| # Falls | 0.0 ± 0.0 | 3.6 ± 4.2 | 0.012 * | −1.278 |
| # Near-Falls | 2.2 ± 1.1 | 8.2 ± 15.8 | 0.249 | −0.564 |
| Clinical Outcomes | Non-Fallers (n=12) | Fallers (n=10) | p-values | Hedge’s g |
| SCS | 8.7 ± 1.1 | 7.9 ± 2.3 | 0.268 | 0.427 |
| PROMIS-Physical T-score | 47.5 ± 6.7 | 45.0 ± 5.7 | 0.160 | 0.397 |
| PROMIS-Anxiety T-score | 44.8 ± 5.9 | 44.7 ± 5.9 | 0.470 | 0.006 |
| PROMIS-Depression T-score | 43.5 ± 4.6 | 43.6 ± 5.9 | 0.411 | −0.013 |
| PROMIS-Fatigue T-score | 46.8 ± 6.5 | 47.0 ± 8.1 | 0.286 | −0.036 |
| PROMIS-Sleep T-score | 45.4 ± 8.3 | 46.4 ± 7.3 | 0.371 | −0.131 |
| PROMIS-Social T-score | 55.0 ± 7.6 | 55.0 ± 6.3 | 0.500 | 0.002 |
| PROMIS-Pain Interference T-score | 47.8 ± 6.9 | 47.8 ± 7.0 | 0.472 | 0.006 |
| PROMIS-Global Pain | 43.3 ± 1.2 | 58.8 ± 2.5 | 0.089 | −0.781 |
| ABC | 3.4 ± 0.2 | 3.8 ± 0.6 | 0.234 | −0.225 |
| PLUS-M 12 T-score | 52.6 ± 4.6 | 50.2 ± 8.1 | 0.334 | 0.358 |
| AMPPRO | 41.3 ± 2.3 | 40.4 ± 3.2 | 0.303 | −.298 |
| TUG (s) | 9.7 ± 1.8 | 11.1 ± 3.0 | 0.090 | −0.562 |
| FSST (s) | 12.7 ± 5.4 | 14.8 ± 3.4 | 0.040 * | −0.739 |
| Gait Parameters | Non-Fallers (n=12) | Fallers (n=10) | p-values | Hedge’s g |
| Walking Speed (m/s) | 0.65 ± 0.23 | 0.52 ± 0.14 | 0.142 | 0.590 |
| SLS Asymmetry (%) | 31.32 ± 20.20 | 30.17 ± 12.20 | 0.500 | 0.063 |
| Hip Flexion Asymmetry (%) | 27.94 ± 19.21 | 31.71 ± 16.66 | 0.209 | −0.200 |
| Hip Ext. Asymmetry (%) | 43.91 ± 18.69 | 39.77 ± 18.57 | 0.291 | 0.214 |
| Hip ROM Asymmetry (%) | 27.33 ± 14.18 | 32.50 ± 15.41 | 0.337 | −0.337 |
| Knee Flexion Asymmetry (%) | 30.69 ± 17.40 | 30.69 ± 9.86 | 0.291 | 0.000 |
| Knee Ext. Asymmetry (%) | 41.47 ± 20.12 | 46.82 ± 13.95 | 0.114 | −0.292 |
| Knee ROM Asymmetry (%) | 29.27 ± 15.12 | 30.13 ± 19.08 | 0.462 | −0.048 |
| Ankle Dorsi. Asymmetry (%) | 33.91 ± 16.35 | 40.41 ± 15.47 | 0.085 | −0.633 |
| Ankle Plantar. Asymmetry (%) | 38.76 ± 18.10 | 39.02 ± 18.48 | 0.436 | −0.014 |
| Ankle ROM Asymmetry (%) | 29.14 ± 15.32 | 40.02 ± 18.59 | 0.070 | −0.620 |
| Peak Braking Asymmetry (%) | 24.04 ± 11.95 | 28.38 ± 20.69 | 0.500 | −0.252 |
| Peak Propulsive Asymmetry (%) | 15.33 ± 11.20 | 24.57 ± 17.44 | 0.158 | −0.616 |
Mean ± standard deviation data for 22 participants. Chi-Squared or Mann-Whitney tests were used as appropriate to determine significance. Bold p-value with asterisk (*) indicates significance. Hedge’s g =absolute values of 0.2, 0.5, and 0.8 indicate small, moderate, and large effect sizes, respectively. Bold Hedge’s g values indicates a moderate to large effect size. K-levels were determined based on AMPPRO scores and prosthetic componentry. Abbreviations: TSA= time since amputation, M= male, F= female, Tr= trauma, Vasc= vascular, Can= cancer, Cong= congenital, TT= transtibial, TF= transfemoral, SCS= Socket Comfort Score (scale of 0 to 10; higher score indicates greater comfort), PROMIS= Patient Reported Outcomes Measurement Information System (T-score scale with a mean of 50 and standard deviation of 10; higher score indicates more of the construct being measured), ABC= Activities Specific Balance Confidence Scale (scale of 0 to 5; higher score indicates greater perceived balance confidence), PLUS-M= Prosthetic Limb Users Survey of Mobility (T-score scale with a mean of 50 and standard deviation of 10; higher score indicates greater perceived mobility), AMPPRO= Amputee Mobility Predictor with Prosthesis (scale of 0 to 42; higher score indicates greater potential for functional mobility), TUG= Timed Up and Go (faster times indicate greater functional mobility), FSST= Four Square Step Test (faster times indicate greater functional mobility), SLS= Single Limb Support, Ext.= extension, Dorsi.= dorsiflexion, Plantar.= plantarflexion.
3.2. Transtibial Fallers v Non-fallers, Transfemoral Fallers v Non-fallers
Table 2 presents demographic characteristics, clinical outcome measure scores, and gait data stratified by fall status for transtibial prosthesis users (n = 15; fallers n = 7, non-fallers n = 8), and transfemoral prosthesis users (n = 7; fallers n = 3, non-fallers n = 4).
Table 2:
Demographic characteristics, clinical outcome measure scores, and gait data stratified by non-fallers and fallers for transtibial and transfemoral prosthesis users.
| Transtibial (n=15) | Transfemoral (n=7) | |||||||
|---|---|---|---|---|---|---|---|---|
| Demographics | Non-Fallers (n= 8) | Fallers (n= 7) | p-values | Hedge’s g | Non-Fallers (n= 4) | Fallers (n= 3) | p-values | Hedge’s g |
| Age (yrs) | 61.6 ± 12.9 | 62.6 ± 7.8 | 0.500 | −0.092 | 49.4 ± 17.5 | 43.1 ± 16.5 | 0.315 | 0.368 |
| Stature (cm) | 176.2 ± 7.7 | 172.0 ± 9.7 | 0.232 | 0.484 | 175.3 ± 12.1 | 175.2 ± 13.2 | 0.429 | 0.008 |
| Body Mass (kg) | 97.2 ± 24.3 | 97.8 ± 22.6 | 0.390 | −0.025 | 75.5 ± 23.8 | 78.8 ± 13.6 | 0.315 | −0.162 |
| Sex | 4 M / 4 F | 5 M / 2 F | 0.268 | - | 3 M / 1 F | 3 M / 0 F | 0.315 | - |
| Race | 8 White | 7 White | 0.478 | - | 3 White/1 Prefer not to answer | 3 White/0 Prefer not to answer | 0.315 | - |
| Ethnicity | 0 Hispanic or Latine/8 Not Hispanic or Latine | 0 Hispanic or Latine/7 Not Hispanic or Latine | 0.478 | - | 1 Hispanic or Latine/3 Not Hispanic or Latine | 0 Hispanic or Latine/3 Not Hispanic or Latine | 0.315 | - |
| TSA (yrs) | 22.0 ± 20.9 | 11.2 ± 14.3 | 0.141 | 0.595 | 13.0 ± 9.4 | 20.3 ± 14.7 | 0.200 | −0.618 |
| Etiology | 4 Tr / 2 Vasc / 1 Can / 1 Cong | 3 Tr / 3 Vasc / 1 Can / 0 Cong | 0.478 | - | 2 Tr / 0 Vasc / 3 Can / 0 Cong | 1 Tr / 0 Vasc / 1 Can / 1 Cong | 0.315 | - |
| Residual Limb length (cm) | 19.6 ± 3.6 | 18.0 ± 5.2 | 0.232 | 0.363 | 15.3 ± 3.5 | 15.7 ± 5.0 | 0.500 | −0.096 |
| K-Level | 5 K3 / 3 K2 | 5 K3 / 2 K2 | 0.390 | - | 4 K3 / 0 K2 | 3 K3 / 0 K2 | 0.500 | - |
| Time using current prosthesis (prosthesis used during study) (yrs) | 3.9 ± 2.4 | 4.0 ± 2.7 | 0.478 | −0.039 | 4.0 ± 3.5 | 1.7 ± 1.2 | 0.200 | 0.817 |
| Time since last prosthetic adjustment (months) | 2.9 ± 2.5 | 1.1 ± 0.6 | 0.141 | 0.958 | 3.6 ± 5.6 | 0.8 ± 0.3 | 0.129 | 0.645 |
| Clinical Outcomes | Non-Fallers (n= 8) | Fallers (n= 7) | p-values | Hedge’s g | Non-Fallers (n= 4) | Fallers (n= 3) | p-values | Hedge’s g |
| SCS | 8.8 ± 1.0 | 8.9 ± 0.9 | 0.434 | −0.103 | 8.5 ± 1.3 | 5.7 ± 3.2 | 0.115 | 0.051 |
| PROMIS-Physical T-score | 47.0 ± 7.3 | 46.4 ± 5.7 | 0.478 | 0.083 | 48.6 ± 6.0 | 41.6 ± 4.6 | 0.057 | 0.079 |
| PROMIS-Anxiety T-score | 47.0 ± 6.1 | 43.0 ± 4.7 | 0.116 | 0.687 | 40.3 ± 1.3 | 48.8 ± 7.5 | 0.115 | −0.498 |
| PROMIS-Depression T-score | 44.8 ± 5.3 | 41.0 ± 4.2 | 0.116 | 0.913 | 41.0 ± 6.9 | 49.6 ± 9.0 | 0.115 | 0.274 |
| PROMIS-Fatigue T-score | 46.9 ± 3.3 | 44.2 ± 7.8 | 0.390 | 0.426 | 46.6 ± 11.3 | 53.6 ± 4.5 | 0.200 | −0.645 |
| PROMIS-Sleep T-score | 46.4 ± 7.7 | 44.4 ± 6.8 | 0.390 | 0.263 | 43.4 ± 10.4 | 51.4 ± 7.0 | 0.200 | −0.735 |
| PROMIS-Social T-score | 54.8 ± 8.5 | 57.7 ± 5.0 | 0.232 | −0.384 | 55.5 ± 6.6 | 48.7 ± 4.0 | 0.115 | 0.990 |
| PROMIS-Pain Interference T-score | 49.4 ± 7.1 | 46.2 ± 5.9 | 0.199 | 0.461 | 44.6 ± 6.1 | 51.4 ± 9.4 | 0.200 | −0.757 |
| PROMIS-Global Pain | 43.3 ± 1.3 | 40.4 ± 2.8 | 0.347 | −0.566 | 48.2 ± 0.8 | 58.8 ± 0.5 | 0.115 | −0.189 |
| ABC | 2.8 ± 0.9 | 3.1 ± 0.7 | 0.347 | 0.048 | 3.7 ± 0.5 | 3.4 ± 0.2 | 0.200 | −0.359 |
| PLUS-M 12 T-score | 52.1 ± 5.2 | 53.0 ± 5.2 | 0.390 | −0.158 | 53.5 ± 1.5 | 43.7 ± 11.0 | 0.225 | 0.109 |
| AMPPRO | 40.9 ± 2.7 | 39.9 ± 3.8 | 0.390 | 0.294 | 42.0 ± 0.8 | 41.7 ± 0.6 | 0.315 | 0.384 |
| TUG (s) | 9.7 ± 2.0 | 10.8 ± 3.2 | 0.199 | −0.392 | 9.7 ± 1.4 | 11.8 ± 3.0 | 0.200 | −0.828 |
| FSST (s) | 13.1 ± 6.5 | 14.1 ± 3.7 | 0.168 | −0.175 | 11.9 ± 2.4 | 16.5 ± 2.7 | 0.057 | −0.520 |
| Gait Parameters | Non-Fallers (n= 8) | Fallers (n= 7) | p-values | Hedge’s g | Non-Fallers (n= 4) | Fallers (n= 3) | p-values | Hedge’s g |
| Walking Speed (m/s) | 0.67 ± 0.19 | 0.51 ± 0.16 | 0.060 | 0.864 | 0.60 ± 0.32 | 0.57 ± 0.10 | 0.315 | 0.089 |
| SLS Asymmetry (%) | 34.0 ± 22.1 | 28.28 ± 13.0 | 0.434 | 0.294 | 25.79 ± 17.38 | 34.58 ± 11.36 | 0.500 | −0.484 |
| Hip Flexion Asymmetry (%) | 26.02 ± 18.58 | 29.87 ± 12.34 | 0.232 | −0.226 | 31.78 ± 22.78 | 35.98 ± 27.44 | 0.500 | −0.143 |
| Hip Ext. Asymmetry (%) | 39.74 ± 18.05 | 43.11 ± 19.61 | 0.434 | −0.169 | 52.26 ± 19.52 | 31.99 ± 16.41 | 0.115 | −0.929 |
| Hip ROM Asymmetry (%) | 25.80 ± 16.98 | 36.95 ± 14.35 | 0.232 | −0.663 | 30.41 ± 6.76 | 22.12 ± 14.82 | 0.200 | −0.648 |
| Knee Flexion Asymmetry (%) | 31.99 ± 18.61 | 33.76 ± 7.37 | 0.232 | −0.114 | 28.08 ± 16.99 | 23.53 ± 12.83 | 0.500 | 0.248 |
| Knee Ext. Asymmetry (%) | 41.63 ± 15.33 | 47.37 ± 14.41 | 0.168 | −0.362 | 41.17 ± 30.58 | 45.53 ± 15.79 | 0.315 | −0.143 |
| Knee ROM Asymmetry (%) | 32.01 ± 14.94 | 36.78 ± 18.91 | 0.268 | −0.266 | 23.81 ± 16.09 | 14.62 ± 7.07 | 0.315 | 0.583 |
| Ankle Dorsi. Asymmetry (%) | 35.00 ± 18.86 | 43.92 ± 18.30 | 0.141 | −0.451 | 25.74 ± 8.58 | 38.87 ± 6.75 | 0.057 | −0.397 |
| Ankle Plantar. Asymmetry (%) | 34.88 ± 15.49 | 39.27 ± 20.79 | 0.500 | −0.228 | 46.53 ± 22.83 | 38.44 ± 15.44 | 0.315 | 0.337 |
| Ankle ROM Asymmetry (%) | 28.98 ± 15.45 | 41.71 ± 15.36 | 0.061 | −0.778 | 29.46 ± 17.44 | 36.08 ± 28.55 | 0.315 | −0.247 |
| Peak Braking Asymmetry (%) | 27.59 ± 9.60 | 33.19 ± 25.52 | 0.331 | −0.277 | 18.71 ± 14.56 | 20.36 ± 6.45 | 0.429 | −0.116 |
| Peak Propulsive Asymmetry (%) | 14.94 ± 12.91 | 26.56 ± 19.50 | 0.215 | −0.657 | 15.92 ± 9.89 | 21.25 ± 16.66 | 0.500 | −0.334 |
Mean ± standard deviation data for 22 participants, stratified by non-faller v faller for transtibial and transfemoral users. Chi-Squared or Mann-Whitney tests were used as appropriate to determine significance. Bold p-value with asterisk (*) indicates significance. Hedge’s g =absolute values of 0.2, 0.5, and 0.8 indicate small, moderate, and large effect sizes, respectively. Bold Hedge’s g values indicate a moderate to large effect size. K-levels were determined based on AMPPRO scores and prosthetic componentry. Abbreviations: TSA= time since amputation, M= male, F= female, Tr= trauma, Vasc= vascular, Can= cancer, Cong= congenital, TT= transtibial, TF= transfemoral, SCS= Socket Comfort Score (scale of 0 to 10; higher score indicates greater comfort), PROMIS= Patient Reported Outcomes Measurement Information System (T-score scale with a mean of 50 and standard deviation of 10; higher score indicates more of the construct being measured), ABC= Activities Specific Balance Confidence Scale (scale of 0 to 5; higher score indicates greater perceived balance confidence), PLUS-M= Prosthetic Limb Users Survey of Mobility (T-score scale with a mean of 50 and standard deviation of 10; higher score indicates greater perceived mobility), AMPPRO= Amputee Mobility Predictor with Prosthesis (scale of 0 to 42; higher score indicates greater potential for functional mobility), TUG= Timed Up and Go (faster times indicate greater functional mobility), FSST= Four Square Step Test (faster times indicate greater functional mobility), SLS= Single Limb Support, Ext.= extension, Dorsi.= dorsiflexion, Plantar.= plantarflexion.
For transtibial prosthesis users, no parameters showed statistically significant differences between fallers and non-fallers. While no parameters were statistically significant, Hedge’s g effect size was large for the PROMIS-Depression (p=0.116, g=0.993) and walking speed (p=0.060, g=0.864). Several parameters also showed a medium effect size: time since amputation (p=0.141, g=0.595), PROMIS-Anxiety (p=0.116, g=0.687), PROMIS-Global Pain (p=0.347, g=−0.566), Hip ROM Asymmetry (p=0.232, g=−0.663), Ankle ROM Asymmetry (p=0.061, g=−0.778), and Peak Propulsive Asymmetry (p=0.215, g=−0.657).
For transfemoral prosthesis users, no parameters showed statistically significant differences between fallers and non-fallers. While no parameters were statistically significant, Hedge’s g effect size was large for time using current prosthesis (p=0.200, g=0.817), PROMIS-Social (p=0.115, g=0.990), the TUG task (p=0.200, g=−0.828), and hip extension asymmetry (p=0.115, g=−0.929). Several parameters also showed a medium effect size: time since amputation (p=0.200, g=−0.618), time since last prosthetic adjustment (p=0.129, g=0.645), PROMIS-Fatigue (p=0.200, g=−0.645), PROMIS-Sleep (p=0.200, g=−0.735), PROMIS-Pain Interference (p=0.200, g=−0.757), and Hip ROM Asymmetry (p=0.200, g=−0.648).
4. Discussion
This study sought to examine if gait asymmetry in people who use a lower-limb prosthesis could significantly distinguish 12-month retrospective fallers from non-fallers. While our hypothesis was not supported, one clinical outcome measure parameter did significantly distinguish non-fallers from fallers (FSST, cut-off time ≥13.14s, medium effect size of g=0.739), and only irrespective of level of prosthesis use, which aligns with prior work.6–8 Additionally, no parameters distinguished fallers from non-fallers by level of prosthesis use (transtibial, transfemoral). While our sample size was small (n=22), the mean age and PLUS-M scores (indication of self-perceived mobility) of participants in this study are in line with the overall population of prosthesis users,30,31 which strengthens the generalizability of the clinical outcome measure and gait parameter results. However, more research is needed to determine if our findings are replicable with a larger sample size of lower-limb prosthesis users, or if a larger subset of transtibial or transfemoral users would yield unique parameters to distinguish fallers from non-fallers by level of prosthesis use.
Findings from this study indicate that the FSST was the most useful parameter at distinguishing fallers from non-fallers, irrespective of level of prosthesis use, with a caveat of showing a medium effect size (g=0.739). This finding aligns with prior work,6,7 even with our sample including a greater proportion of individuals who are older, with K2 level of mobility, or with a dysvascular etiology which better reflects the overall population of lower-limb prosthesis users. The FSST cut off time (≥13.14s) in this study was several seconds slower than a cut-off time determined in a prior study (≥8.49s) suggesting that cut-off times should be reflective of the individual demographics.7 Therefore, if replicable in a larger sample, the FSST cut-off time could provide a practical assessment to help clinicians routinely distinguish fallers from non-fallers in clinical practice.
No gait asymmetry parameters were able to significantly distinguish fallers from non-fallers either irrespective of prosthesis use or by level of prosthesis use, which did not support our hypothesis. While not statistically significant, several medium effect sizes were found, which may be useful to include in future studies. Additionally, participants, regardless of faller status or level of prosthesis use tended to have a larger range with greater gait asymmetry compared to a prior review of summarized values across over forty studies, likely due to the more reflective sample population utilized in this study (e.g., K2 mobility).16 Despite our study sample showing greater asymmetry than prior studies, no gait measures were able to distinguish fallers from non-fallers based on retrospective recall of fall history, indicating gait asymmetry may not be useful in distinguishing fall risk. However, given the large effect size for transfemoral fallers vs non-fallers in hip extension asymmetry (more asymmetry in non-fallers, g=−0.929, p=0.115) (Table 2) and walking speed for transtibial fallers vs non-fallers (slower in fallers, g=0.864, p=0.060) (Table 2), further research is needed with larger sample sizes by prosthesis level to make this determination.
Besides the FSST, no other clinical outcome measure scores could significantly distinguish fallers from non-fallers (either irrespective of level of prosthesis use or by level of prosthesis use). While one study found that the TUG test could significantly distinguish retrospective fallers from non-fallers irrespective of level of prosthesis use,6 the TUG was not significant in this study nor did it show a large effect size (p=0.090, g=−0.562) (Table 1). This may be due to differences in how participants were grouped (prior study distinguished recurrent fallers (people who fell multiple times) from non-fallers), or our small sample size. Additionally, irrespective of level of prosthesis use, the PROMIS-Global Pain score was the closest clinical outcome measure to reaching significance (p=0.089) and showed a medium effect size (g=−0.781) (Table 1), with fallers reporting more pain. This relationship between global pain and fall risk could be related to the number of medications used, as a prior study found that the use of greater than two prescription medications increased fall risk in this population.15 Specifically, tricyclic antidepressants, gabapentin, and opioids are commonly prescribed for pain management in this population32 and have each been shown to increase fall risk.33,34 Further, time since last prosthetic appointment showed large effect sizes (irrespective of prosthesis use, fallers showed less time than non-fallers), which suggests a potential need for targeted fall risk training after prosthetic adjustments.
Further, when considering level of prosthesis use, transtibial prosthesis users showed a large effect size in PROMIS-Depression (lower score in fallers, g=0.913) (Table 2), while transfemoral prosthesis users showed a large effect size in PROMIS-Social (lower scores in fallers, g=0.990) (Table 2). This finding aligns with previous research in older adults in the general population, which has demonstrated that higher PROMIS-Depression scores and lower PROMIS-Social scores are associated with increased risk of falls.35,36 Demographic factors with medium to large effect sizes may also provide insight into fall risks. Specifically, transtibial fallers showed a shorter time since amputation (11.2 years on average) compared to non-fallers (22.0 years on average), highlighting the need for fall risk training for people who use a transtibial prosthesis within approximately the first decade of their amputation. Additionally, transfemoral fallers showed a shorter time using their current prosthesis (1.7 years on average) than non-fallers (4.0 years on average), which demonstrates the potential benefit of additional fall risk training when people who use a transfemoral prosthesis receive new componentry or a new prosthetic socket.
A limitation with collecting retrospective fall history is that gait changes can occur after a fall, thus, gait data may not be fully reflective of prior fall status. For example, 7 out of 10 prosthesis users in this study, irrespective of level of prosthesis use, reported a change in behavior or activity after a fall had occurred (e.g., changed how you do certain activities, become more careful or cautious). Prospective falls, as an alternative to retrospective, could further confirm the findings in this study and potentially identify additional measures of fall risk that were not captured in this study due to the inherent limitation of retrospective fall history. Further, our a priori analysis indicated a sample size of 40 participants was needed to adequately power this study.6,7,10 Despite our limited sample size, results were still comparable to that of prior studies6,7 and provide a useful starting point for screening in clinical practice and future research. However, our results should still be interpreted with caution, given that our study was underpowered.
5. Conclusions
The FSST was most effective at distinguishing 12-month retrospective fallers from non-fallers, with a cut-off time of 13.14 seconds, irrespective of level of prosthesis use. This cut-off time may help distinguish fallers from non-fallers in people who use unilateral lower-limb prostheses. While gait asymmetry was not able to statistically significantly distinguish fallers from non-fallers, hip extension asymmetry and walking speed showed large effect sizes, which suggests that these measures could be important to examine by level of prosthesis use with a larger sample size.
Supplementary Material
Acknowledgements:
The authors would like to thank Caitlyn Finnerty, Shawn Kennedy, and Tasha Buxton for assisting with data collection. Thanks also to Cody Logenbaugh at Baker Orthotics and Prosthetics and Mark Ashford at Hanger Clinic for their help with participant recruitment. This project is taken in part from a dissertation MGF submitted to the University of North Texas Health Science Center in partial fulfillment of the requirements for the degree of Doctor of Philosophy.
Funding Source:
This project was supported by an American Orthotics and Prosthetics Association research award administered by the Center for Orthotic and Prosthetic Learning and Outcomes/Evidence-Based Practice. MGF was supported by the National Institutes of Health/National Institute on Aging (T32 AG020494) and the Institute for Healthy Aging.
Footnotes
Declaration of Interest: None
References
- 1.Wong C, Chihuri S, Li G. Risk of fall-related injury in people with lower limb amputations: A prospective cohort study. J Rehabil Med. 2016; 48: 80–85. [DOI] [PubMed] [Google Scholar]
- 2.Kim J, Major MJ, Hafner B, et al. Frequency and Circumstances of Falls Reported by Ambulatory Unilateral Lower Limb Prosthesis Users: A Secondary Analysis. Journal of Injury, Function and Rehabilitation. 2019; 11: 344–353. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Chihuri S, Wong CK. Factors associated with the likelihood of fall-related injury among people with lower limb loss. Inj Epidemiol. 2018; 5: 42. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Miller WC, Deathe AB, Speechley M, et al. The influence of falling, fear of falling, and balance confidence on prosthetic mobility and social activity among individuals with a lower extremity amputation. Archives of Physical Medicine and Rehabilitation. 2001; 82: 1238–1244. [DOI] [PubMed] [Google Scholar]
- 5.Pezzin LE, Dillingham TR, MacKenzie EJ. Rehabilitation and the long-term outcomes of persons with trauma-related amputations. Archives of Physical Medicine and Rehabilitation. 2000; 81: 292–300. [DOI] [PubMed] [Google Scholar]
- 6.Hakim RM, Frey CM, Spadoni KE, et al. Identifying Fallers Using Clinical Balance Measures in Community-Dwelling Adults with Lower Extremity Amputation: A Cross-Sectional Study. J Dev Phys Disabil. 2018; 30: 677–688. [Google Scholar]
- 7.Sawers A, Hafner BJ. Using Clinical Balance Tests to Assess Fall Risk among Established Unilateral Lower Limb Prosthesis Users: Cutoff Scores and Associated Validity Indices. PM&R. 2020; 12: 16–25. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Finco MG, Sumien N, Moudy SC. Clinical evaluation of fall risk in older adults who use lower-limb prostheses: A scoping review. JAGS. 2023; [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Clemens S, Doerger C, Lee SP. Current and Emerging Trends in the Management of Fall Risk in People with Lower Limb Amputation. Curr Geriatr Rep. 2020;9(3):134–141. doi: 10.1007/s13670-020-00328-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Tobaigy M, Hafner BJ, Sawers A. Recalled Number of Falls in the Past Year—Combined With Perceived Mobility—Predicts the Incidence of Future Falls in Unilateral Lower Limb Prosthesis Users. Physical Therapy. 2022; 102: pzab267. [DOI] [PubMed] [Google Scholar]
- 11.Lusardi MM, Fritz S, Middleton A, et al. Determining Risk of Falls in Community Dwelling Older Adults: A Systematic Review and Meta-analysis Using Posttest Probability. Journal of Geriatric Physical Therapy. 2017; 40: 1–36. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Gillain S, Boutaayamou M, Schwartz C, et al. Gait symmetry in the dual task condition as a predictor of future falls among independent older adults: a 2-year longitudinal study. Aging Clin Exp Res. 2019; 31: 1057–1067. [DOI] [PubMed] [Google Scholar]
- 13.Liang JN, Ho K-Y, Lee Y-J, et al. Slow Walking in Individuals with Chronic Post-Stroke Hemiparesis: Speed Mediated Effects of Gait Kinetics and Ankle Kinematics. Brain Sci. 2021; 11: 365. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.[redacted]
- 15.Hunter SW, Batchelor F, Hill KD, Hill AM, Mackintosh S, Payne M. Risk Factors for Falls in People With a Lower Limb Amputation: A Systematic Review. PM R. 2017;9(2):170–180.e1. doi: 10.1016/j.pmrj.2016.07.531. [DOI] [PubMed] [Google Scholar]
- 16.Finco MG, Sarah Moudy, Rita Patterson. Normalized Kinematic Walking Symmetry Data for Individuals Who Use Lower-Limb Prostheses: Considerations for Clinical Practice and Future Research. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Rattanakoch J, Samala M, Limroongreungrat W, Guerra G, Tharawadeepimuk K, Nanbancha A, Niamsang W, Kerdsomnuek P, Suwanmana S. Validity and Reliability of Inertial Measurement Unit (IMU)-Derived 3D Joint Kinematics in Persons Wearing Transtibial Prosthesis. Sensors. 2023; 23(3):1738. 10.3390/s23031738. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Loiret I, Villa C, Dauriac B, et al. Are wearable insoles a validated tool for quantifying transfemoral amputee gait asymmetry? Prosthetics & Orthotics International. 2019; 43: 492–499. [DOI] [PubMed] [Google Scholar]
- 19.Timmermans C, Cutti AG, van Donkersgoed H, Roerdink M. Gaitography on lower-limb amputees: Repeatability and between-methods agreement. Prosthet Orthot Int. 2019;43(1):71–79. doi: 10.1177/0309364618791618 [DOI] [PubMed] [Google Scholar]
- 20.Harris PA, Taylor R, Minor BL, et al. The REDCap consortium: Building an international community of software platform partners. Journal of Biomedical Informatics. 2019; 95: 103208. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Gailey RS, Roach KE, Applegate EB, et al. The Amputee Mobility Predictor: An instrument to assess determinants of the lower-limb amputee’s ability to ambulate. Archives of Physical Medicine and Rehabilitation. 2002; 83: 613–627. [DOI] [PubMed] [Google Scholar]
- 22.Sawers A, McDonald CL, Hafner BJ. A systematic review of outcome measures used to assess standing balance or posture in people with lower limb amputation. Prosthet Orthot Int. 2020; 44: 391–402. [Google Scholar]
- 23.Hanspal R, Fisher K, Nieveen R. Prosthetic socket fit comfort score. Disabil Rehabil. 2003;25:1278–1280. [DOI] [PubMed] [Google Scholar]
- 24.Hafner BJ, Morgan SJ, CPO, et al. Psychometric evaluation of self-report outcome measures for prosthetic applications. J Rehabil Res Dev. 2016;53:797–812. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Amtmann D, Bamer AM, Kim J, et al. A comparison of computerized adaptive testing and fixed-length short forms for the Prosthetic Limb Users Survey of Mobility (PLUS-MTM). Prosthet Orthot Int. 2018;42:476–482. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Miller WC, Speechley M, Deathe AB. Balance confidence among people with lower-limb amputations. Phys Ther. 2002;82:856–865. [PubMed] [Google Scholar]
- 27.Schoppen T, Boonstra A, Groothoff JW, et al. The timed “up and go” test: Reliability and validity in persons with unilateral lower limb amputation. Arch Phys Med Rehabil. 1999;80:825–828. [DOI] [PubMed] [Google Scholar]
- 28.Langford Z The Four Square Step Test. J Physiother. 2015;61:162. [DOI] [PubMed] [Google Scholar]
- 29.Robinson RO, Herzog W, Nigg BM. Use of force platform variables to quantify the effects of chiropractic manipulation on gait symmetry. J Manip Physiol Ther. 1987;10:172–176. [PubMed] [Google Scholar]
- 30.Ziegler-Graham K, MacKenzie EJ, Ephraim PL, Travison TG, Brookmeyer R. Estimating the prevalence of limb loss in the United States: 2005 to 2050. Arch Phys Med Rehabil. 2008;89(3):422–429. doi: 10.1016/j.apmr.2007.11.005 [DOI] [PubMed] [Google Scholar]
- 31.Fylstra BL, England DL, Stevens PM, Campbell JH, Wurdeman SR. Creating Adjusted Scores Targeting mobiLity Empowerment (CASTLE 1): determination of normative mobility scores after lower limb amputation for each year of adulthood. Disabil Rehabil. Published online 2023. doi: 10.1080/09638288.2023.2208376 [DOI] [PubMed] [Google Scholar]
- 32.Modest JM, Raducha JE, Testa EJ, Eberson CP. Management of post-amputation pain. R I Med J. 2020;103(4):19–22. Published 2020 May 1. [PubMed] [Google Scholar]
- 33.Rentsch CT, Morford KL, Fiellin DA, Bryant KJ, Justice AC, Tate JP. Safety of gabapentin prescribed for any indication in a large clinical cohort of 571,718 US veterans with and without alcohol use disorder. Alcohol Clin Exp Res. 2020;44(9):1807–1815. doi: 10.1111/acer.14408 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Seppala LJ, Wermelink AMAT, de Vries M, et al. Fall-risk-increasing drugs: a systematic review and meta-analysis: II. Psychotropics. J Am Med Dir Assoc. 2018;19(4):371.e11–371.e17. doi: 10.1016/j.jamda.2017.12.098 [DOI] [PubMed] [Google Scholar]
- 35.Gambaro E, Gramaglia C, Azzolina D, Campani D, Molin AD, Zeppegno P. The complex associations between late-life depression, fear of falling and risk of falls: a systematic review and meta-analysis. Ageing Res Rev. 2022;73:101532. doi: 10.1016/j.arr.2021.101532 [DOI] [PubMed] [Google Scholar]
- 36.Dashner J, Espín-Tello SM, Chen SW, et al. Influence of falls, fall-related injuries, and fear of falling on social participation in people aging with long-term physical disability: a cross-sectional study. Disabil Rehabil. Published online December 18, 2023. doi: 10.1080/09638288.2023.2293990 [DOI] [PMC free article] [PubMed] [Google Scholar]