Table 3.
Overall Changes in Parameters and Changes in Parameters Stratified by Lead Age
| Overall Changes in Parameters (n=95) | p-value | ||
|---|---|---|---|
| Sensing | Pre-LVAD | 10.70±5.15 | <0.01 |
| Post-LVAD | 9.09±5.87 | ||
| Impedance | Pre-LVAD | 461.69±120.40 | 0.857 |
| Post-LVAD | 463.36±151.60 | ||
| Capture Threshold | Pre-LVAD | 1.00±0.57 | 0.017 |
| Post-LVAD | 1.18±0.81 | ||
| Pulse width | Pre-LVAD | 0.46±0.14 | 0.206 |
| Post-LVAD | 0.49±0.19 | ||
| High Voltage Impedance | Pre-LVAD | 53.21±15.28 | <0.01 |
| Post-LVAD | 46.2±17.25 | ||
| Changes in parameters stratified for RV lead age less or equal to than 2 years (n=33) | |||
| Sensing | Pre-LVAD | 10.89±4.87 | 0.537 |
| Post-LVAD | 10.43±6.74 | ||
| Impedance | Pre-LVAD | 458.27±79.84 | 0.333 |
| Post-LVAD | 445.24±100.87 | ||
| Capture Threshold | Pre-LVAD | 0.96±0.67 | 0.405 |
| Post-LVAD | 1.03±0.83 | ||
| Pulse width | Pre-LVAD | 0.46±0.12 | 0.211 |
| Post-LVAD | 0.48±0.17 | ||
| High Voltage Impedance | Pre-LVAD | 56.82±15.41 | <0.01 |
| Post-LVAD | 50.85±16.73 | ||
| Changes in parameters stratified for RV lead age greater than 2 years (n=62) | |||
| Sensing | Pre-LVAD | 10.60±5.33 | <0.01 |
| Post-LVAD | 8.38±5.27 | ||
| Impedance | Pre-LVAD | 463.52±137.78 | 0.438 |
| Post-LVAD | 474.0±172.64 | ||
| Capture Threshold | Pre-LVAD | 1.03±0.51 | 0.026 |
| Post-LVAD | 1.25±0.78 | ||
| Pulse width | Pre-LVAD | 0.46±0.15 | 0.374 |
| Post-LVAD | 0.49±0.19 | ||
| High Voltage Impedance | Pre-LVAD | 51.29±14.98 | <0.01 |
| Post-LVAD | 43.73±17.14 | ||