Skip to main content
. 2025 Aug 19;15:30457. doi: 10.1038/s41598-025-14552-5

Table 3.

Linearity, limit of detection, and limit of quantification of Rb/Cs.

Elemental Standard solution range(ug·L−1) Calibration curves R2 LDQsolvent (ug·L−1) LDQ matrix (ug·L−1)
Rb 5-400 y=0.0072x-0.013 0.9998 0.030 0.039
Cs 5-400 y=0.0030x-0.0050 0.9999 0.003 0.005