Table 1.
Comparison of the consensus and the study protocol in BP measurement.
| Consensus protocol | Study protocol | |
|---|---|---|
| Validation sample size | At least 85 subjects. In case an independent general population of 85 subject studies has been completed successfully, a compromise was agreed to accept a minimum of 35 special population subjects. |
35 subjects in the acute ischemic stroke population. |
| Cuff size-stratified subgroups | A minimum of 22 subjects per cuff, which means, that 4 cuffs could be evaluated in an 88-subject study. | Not applicable in the watch device. |
| Reference blood pressure measurement and validation procedure | Same arm sequential measurement. Four reference BP measurements. follow, alternated by 3 test device measurements (R1–T1–R2–T2–R3–T3–R4). Measurements will be performed with at least 60-second intervals. |
Same arm sequential measurement with 2 arms performed simultaneously. Divided into 2 arms, first reference BP measurements and device calibration, followed by 2 reference BP measured simultaneously with 2 test devices. Measurement performed with 60-second intervals. |
| Validation criteria | BP difference meets both criteria 1 and 2. Presented with Bland–Altman scatterplots. |
BP difference meets both criteria 1 and 2. Presented with Bland–Altman scatterplots. |