Table 1.

Intra-run and inter-run accuracy and precision for TAK-981 in human K2EDTA plasma.

Run
Number
LLOQ 0.100 ng/mL LQC 0.300 ng/mL MQC 30.0 ng/mL HQC 75.0 ng/mL
1 0.0938 0.309 32.0 79.6
  0.0972 0.325 31.9 77.6
  0.0951 0.309 30.9 75.3
  0.104 0.317 32.9 77.5
  0.0958 0.288 30.8 78.8
  0.0991 0.307 32.2 77.6
Intra-run Mean 0.0975 0.309 31.8 77.7
Intra-run S.D. 0.00367 0.0124 0.804 1.46
Intra-run %CV 3.8 4.0 2.5 1.9
Intra-run %RE −2.5 3.0 6.0 3.6
n 6 6 6 6
2 0.0940 0.342 28.9 76.2
  0.0929 0.256 30.2 73.3
  0.108 0.287 30.1 72.7
  0.0854 0.291 29.5 74.8
  0.102 0.289 30.0 71.1
  0.0890 0.311 29.5 77.2
Intra-run Mean 0.0952 0.296 29.7 74.2
Intra-run S.D. 0.00838 0.0286 0.494 2.28
Intra-run %CV 8.8 9.7 1.7 3.1
Intra-run %RE −4.8 −1.3 −1.0 −1.1
n 6 6 6 6
3 0.109 0.323 31.1 76.7
  0.107 0.346* 31.6 74.1
  0.113 0.325 33.4 76.9
  0.109 0.320 32.8 76.6
  0.112 0.318 31.2 74.8
  0.105 0.334 31.1 75.3
Intra-run Mean 0.109 0.328 31.9 75.7
Intra-run S.D. 0.00299 0.0106 0.991 1.16
Intra-run %CV 2.7 3.2 3.1 1.5
Intra-run %RE 9.0 9.3 6.3 0.9
n 6 6 6 6
Inter-run Mean 0.101 0.311 31.1 75.9
Inter-run S.D. 0.00817 0.0223 1.27 2.18
Inter-run %CV 8.1 7.2 4.1 2.9
Inter-run %RE 1.0 3.7 3.7 1.2
n 18 18 18 18

Note: *: Deviation from nominal > 15%; value included in statistics.

%CV: percent coefficient of variation; %RE: percent relative error; HQC: high-quality control; LLOQ: lower limit of quantitation; LQC: low-quality control; MQC: medium-quality control; n: number; S.D.: standard deviation.