Skip to main content
Ophthalmology and Therapy logoLink to Ophthalmology and Therapy
. 2025 Jul 24;14(9):2313–2322. doi: 10.1007/s40123-025-01212-0

Navigating the Lens Labyrinth: A Practical Approach to Intraocular Lens Classification and Selection

Joaquín Fernández 1, Filomena Ribeiro 2,3, H Burkhard Dick 4, Carlos Rocha-de-Lossada 5,6,7, Manuel Rodríguez-Vallejo 1,✉
PMCID: PMC12370590  PMID: 40705184

Abstract

The expanding array of intraocular lenses (IOLs) designed to increase the depth-of-field (DOFi) presents cataract and refractive surgeons with considerable complexity when selecting the most appropriate IOL to meet patient expectations. This practical review underscores the importance of using the functional classification framework, given the large number of different available models and clinical studies. By adopting a classification system grounded in clinical outcomes, surgeons can systematically choose IOLs on the basis of the probability of achieving spectacle independence and optimal visual acuity. Beyond functional classification, the review emphasizes the critical importance of considering additional factors such as contrast sensitivity and dysphotopsia risk to tailor IOL selection to individual visual needs. This layered approach helps optimize patient-centered outcomes and postoperative satisfaction.

Keywords: Intraocular lens, Functional classification, Spectacle independence, Visual acuity, Biometric factors

Key Summary Points

Patients primarily desire spectacle independence with intraocular lenses (IOLs) that increase the depth-of-field; therefore, surgeons must thoroughly evaluate patient-specific daily visual tasks and assess the patient’s willingness to accept potential visual compromises in pursuit of this goal.
The evidence-based functional classification enables surgeons to make structured, outcome-driven decisions by selecting an IOL from a category according to the probability of achieving specific visual results.
Although functional classification guides IOL selection on the basis of spectacle independence and visual acuity, it may overlook subtle yet important differences between lenses. Surgeons should also assess contrast sensitivity and dysphotopsia risk to better align with patient expectations.
A thorough assessment of pupil size, corneal aberrations, astigmatism, and comorbidities is essential because they significantly influence the likelihood of achieving the expected visual outcomes.

Introduction

Intraocular lenses (IOLs) that extend the depth-of-field (DOFi) are used in cataract and refractive lens exchange (RLE) surgeries to increase the likelihood of postoperative spectacle independence [1–3]. Excluding conventional monofocal IOLs, surgeons currently have access to a global portfolio of more than 100 IOL models from various manufacturers [4]. Therefore, it is reasonable to assume that many of these models deliver similar clinical outcomes, while it is understandable that marketing departments strive to differentiate their products from the competition.

However, the sheer volume of available models, combined with an ever-growing body of literature characterized by heterogeneity in study design and outcome reporting often leads to confusion among surgeons aiming to make informed choices [5, 6]. While several approaches to classifying IOLs have been proposed, many are shaped by regional practices, clinical traditions, or theoretical perspectives of their authors [4, 7, 8]. This complexity highlights the importance of an evidence-based approach that not only identifies the most probable clinical outcomes for each IOL category but also quantifies the certainty with which those outcomes can be expected. Against this background, the purpose of this practical review is to provide cataract and refractive surgeons with an evidence-based framework, developed through consensus by the European Society of Cataract and Refractive Surgeons (ESCRS) Functional Vision Working Group [2] (Fig. 1), for the selection of IOLs that increase DOFi. This article is based on previously conducted studies and does not contain any new studies with human participants or animals performed by any of the authors.

Fig. 1.

Fig. 1

Average defocus curves for each evidence-based functional classification category. Shaded areas indicate the cutoff regions for PARTIAL-Depth-of-Field (DOFi), shown with a green gradient background and dotted colored lines, and for FULL-DOFi, shown with a yellow background and solid-colored lines. The FULL-DOFi category exhibits a progressive decrease in the difference between intermediate and near defocus steps from steep to continuous categories

By synthesizing current clinical evidence and integrating it with functional classification systems, this review aims to facilitate a patient-centered approach to IOL selection, informing about the likelihood of achieving postoperative spectacle independence and satisfactory visual acuity. Moreover, it seeks to bridge the gap between clinical outcomes and patient-reported expectations, offering practical guidance on managing visual performance variables such as contrast sensitivity and dysphotopsia.

Managing Expectations

Preoperative selection of an IOL that increases the DOFi should begin with a thorough patient interview. The primary objective of this interview is to align the patient’s visual needs with the potential outcomes of the procedure [9]. Therefore, the discussion should focus on accurately identifying the balance between the desired postoperative outcomes and the level of personal investment, whether financial or in terms of potential compromises in vision, that they are prepared to accept in pursuit of those outcomes [10–13]. Importantly, this conversation must also address the permanent nature of IOL surgery, particularly as simultaneous vision IOLs remain the only currently available surgical solution that offers a lifelong correction at several distances. However, RLE patients should be informed of key caveats, including the increased risks of retinal complications in certain populations and the potential technical challenges associated with IOL exchange if expectations are not achieved [14].

The primary expectation of candidates for IOLs that increase the DOFi is the ability to perform daily tasks without the need for spectacles [15]. Therefore, to achieve functional vision, surgeons must identify the specific tasks the patient commonly performs and select the procedure that will provide the level of visual performance required to complete these tasks without spectacles [16]. Fortunately, current evidence provides the necessary information to guide surgeons in selecting the IOL most likely to help achieve this goal.

Evidence-Based IOL Selection

The ESCRS, through its Functional Vision Working Group of experts, supported the classification of IOLs purely on the basis of clinical outcomes [1, 2]. This new classification was developed through an analysis of historical evidence, allowing for the formation of IOL categories based on the similarity of monocular distance-corrected visual acuity defocus curves reported in published studies that included this endpoint. This analysis resulted in two main categories: partial DOFi and full DOFi. Each of these categories is further divided into three subcategories. The partial DOFi group includes narrowed, enhanced, and extended DOFi, while the full DOFi group comprises continuous, smooth, and steep DOFi [2]. Although the specific defocus curve shapes and DOFi cutoff values can be found in the original references [1, 2], the key point for readers of this practical review is that this classification supports evidence-based selection of IOLs. In other words, it enables surgeons to choose lenses according to the probability of achieving a specific visual outcome, as informed by a synthesis of current clinical evidence.

From Functional Classification to Spectacle Independence

As described in the previous sections, spectacle independence is the primary clinical outcome to consider when selecting IOLs that increase the DOFi, and six general categories of IOLs are currently available to support this goal. However, a discrepancy exists between outcome measures. Spectacle independence is a patient-reported outcome, typically derived from psychometric assessments or questionnaires [15], whereas the visual acuity defocus curve is a clinically measured outcome, based on psychophysical testing conducted in a clinical setting [17].

Clinical outcomes in IOL studies are often correlated, enabling estimation of the probability of achieving spectacle independence at far, intermediate, and near distances for each functional IOL category [15]. Table 1 presents the pooled estimates from the meta-analysis with a random-effects model using the DerSimonian and Laird estimator of all available studies in IOLEvidence database in April 2025 (QvisionAcademy, https://apple.co/3NQ6Feu), including the mean probability of spectacle independence, its 95% confidence interval (95% CI), and 95% prediction interval (95% PI). The IOLEvidence database is publicly available through the IOLEvidence app and is periodically updated to include new reports containing the primary endpoints defined in the referenced standards [5]. Studies involving populations with comorbidities (e.g., ocular disease or prior laser refractive surgery), target modifications such as micro-monovision, or corneal astigmatism requiring toric IOLs or additional corneal incisions for astigmatism correction were excluded.

Table 1.

Probability of spectacle independence reported in studies evaluating intraocular lenses by functional category

Category N Intermediate N Near
Pooled mean 95% CI 95% PI Pooled mean 95% CI 95% PI
Partial DOFi narrowed 17 0.52 0.44–0.60 0.23−0.80 18 0.12 0.07−0.19 0.01−0.56
Partial DOFi enhanced 12 0.77 0.66−0.85 0.33−0.96 19 0.30 0.21−0.42 0.05−0.79
Partial DOFi extended 27 0.90 0.85−0.93 0.61−0.98 47 0.65 0.56−0.73 0.13−0.96
Full DOFi continuous 5 0.96 0.92−0.98 – 8 0.91 0.81−0.96 0.47−0.99
Full DOFi smooth 64 0.95 0.93−0.96 0.74−0.99 87 0.89 0.87−0.91 0.67−0.97
Full DOFi steep 13 0.85 0.74−0.92 0.38−0.98 23 0.81 0.73−0.87 0.42−0.96

Pooled estimates from meta-analysis with a random-effects model using the DerSimonian and Laird estimator

CI confidence interval, PI prediction interval, N number of studies in the group

“–” Not enough studies to calculate PI

To understand the practical value of this information, consider a patient who is a professional driver, requiring acceptable intermediate vision without spectacles but not concerned about near vision. Suppose we are evaluating IOLs from the narrowed category (conventional monofocals) [2]. The pooled mean probability of achieving intermediate spectacle independence in this IOL category is 0.52 (Table 1), indicating that, on average, patients implanted with IOLs from this category achieve spectacle independence in 52% of cases. However, while 0.52 represents the best point estimate, individual samples of studies may yield slightly different results. The 95% confidence interval (95% CI) for this estimate ranges from 0.44 to 0.60. This means that if meta-analysis is repeated using different but comparable sets of studies, 95% of the resulting pooled means would fall within this interval. The 95% CI thus provides an important measure of the precision of the pooled estimate, helping clinicians understand the expected range of average performance for IOLs within this category.

In contrast, the 95% prediction interval (95% PI) for the same IOL category ranges from 0.23 to 0.80. This interval provides an estimate of the range within which the intermediate spectacle independence rate of a new, future study, assuming it is similar to those included in the meta-analysis, would be expected to fall. 95% PI accounts for both within-study variability and between-study heterogeneity, including differences in study populations, specific narrow category IOL models, and methodologies. Thus, if a clinician selects an IOL from the narrow category and consults a single relevant study, it is reasonable to expect that the observed spectacle independence rate would fall between 0.23 and 0.80 with 95% confidence.

From a clinical perspective, both the pooled mean and its 95% CI are important for estimating the average expected effect across all comparable studies. However, the 95% PI provides crucial insight into the potential variability in outcomes observed in individual clinical settings, lenses, or studies. In this case, while the pooled mean spectacle independence rate of 0.52 suggests a generally moderate outcome for narrow IOLs, the wide PI highlights considerable uncertainty when predicting the result for any single IOL or individual study from this category.

As an alternative, IOLs from the enhanced category demonstrate a higher pooled mean spectacle independence rate of 0.77 (Table 1), although the 95% PI remains wide, ranging from 0.33 to 0.96, reflecting substantial variability. In contrast, the extended category presents a pooled mean of 0.90, accompanied by a narrower 95% PI of 0.61–0.98, indicating not only a superior average outcome but also greater consistency across IOLs and studies from this category. From an evidence-based perspective, to achieve spectacle independence at intermediate distances, surgeons should preferentially select an IOL from the extended category, as it offers a high probability of spectacle independence (approximately 0.90) with a high level of certainty that the outcome from a particular IOL study of this category would fall within the range of 0.61–0.98. However, as will be described later, there are other factors not included in the functional classification that must also be considered.

From Functional Classification to Visual Acuity

Spectacle independence can be considered an absolute expectation, meaning that postoperatively, this outcome may or may not be achieved, thereby fulfilling or failing to fulfill the patient’s expectations. Conversely, visual performance is regarded as a relative expectation, as the preoperative state serves as a baseline or reference point for comparison [18]. For instance, if a patient with cataracts achieves a visual acuity of 0.1 logMAR at distance and 0.2 logMAR at near with spectacles and subjectively deems these visual acuities sufficient for performing daily near tasks, these values can be considered baseline targets to maintain postoperatively. Similarly, if a patient with a clear crystalline lens achieves 0 logMAR at distance and 0.3 logMAR at near with multifocal contact lenses, this performance can also serve as a baseline reference. Furthermore, for a patient achieving −0.1 logMAR at distance and 0 logMAR at near with spectacles but willing to accept a compromise in visual acuity to achieve spectacle independence, it is possible to simulate reduced vision defocusing with trial frame, contact lenses, or simulators [19], by presenting a standardize chart with progressively lower levels of acuity at the distance of interest, thereby exploring the degree of visual acuity the patient is willing to sacrifice [5]. Thus, in cases of relative expectations, the surgeon gains preoperative insight into the level of visual acuity that the patient considers functional. Later, looking at the evidence, the visual acuity provided by the different functional categories can be explored in the same way as in the case of spectacle independence (Table 2).

Table 2.

Uncorrected binocular visual acuity reported in studies evaluating intraocular lenses by functional category

Category N Intermediate (66 cm) N Near (40 cm)
Pooled mean 95% CI 95% PI Pooled mean 95% CI 95% PI
Partial DOFi narrowed 27 0.23 0.19−0.27 0.02−0.44 23 0.39 0.34−0.44 0.16−0.63
Partial DOFi enhanced 14 0.14 0.09−0.19 −0.07−0.35 16 0.31 0.27−0.35 0.14−0.49
Partial DOFi extended 21 0.07 0.05−0.10 −0.03−0.18 40 0.18 0.15−0.21 −0.01−0.38
Full DOFi continuous 6 0.00 −0.03−0.02 −0.09−0.08 12 0.04 0.03−0.05 −0.01−0.08
Full DOFi smooth 19 0.07 0.04−0.10 −0.05−0.19 63 0.06 0.05−0.07 −0.03−0.14
Full DOFi steep 2* 0.11 0.09−0.14 – 18 0.09 0.05−0.13 −0.08−0.26

Pooled estimates from meta-analysis with a random-effects model using the DerSimonian and Laird estimator

CI confidence interval, PI prediction interval, N number of studies in the group

*Not enough studies reported uncorrected binocular visual acuity to allow reliable pooling for the full DOFi steep category

“–” Not enough studies to calculate PI

Beyond Visual Acuity in Managing Expectations

Previously, the process of conducting an evidence-based selection of an IOL on the basis of a specific level of functional classification was described, depending on the patient’s spectacle dependence and visual acuity at various distances. However, certain endpoints exhibit subtle variations that may not be captured within the evidence-based functional classification, as the bias between categories can be substantially greater than the differences in mean outcomes [1]. Therefore, the selection of an IOL is fundamentally a top–down process, wherein a general category is chosen according to the desired levels of spectacle independence and visual acuity. Subsequently, a deeper level of specialization is required to select a specific IOL model, considering other critical visual and patient reported outcomes important for patients. We therefore recommend reviewing studies of IOLs selected from a given category and examining additional outcomes in greater depth, particularly those derived from validated questionnaires. These include instruments recommended for assessing spectacle independence (Patient-Reported Spectacle Independence Questionnaire (PRSIQ)) [15], visual symptoms (Assessment of IntraOcular Lens Implant Symptoms (AIOLIS)) [12], night driving difficulties (Vision and Night Driving Questionnaire (VND-Q)) [20], and the ability to perform daily activities (Visual Function Index‑14 (VF-14)) [21], among others. It is essential to understand the specific purpose of each questionnaire and the characteristics of the population used for its validation, to ensure accurate interpretation of the results. Furthermore, single-item questions that report the proportion of subjects achieving a specific level on a 5-point Likert scale can also provide valuable insights, as outlined in existing standards [5]. In addition to validated questionnaires, the structured assessment of patient needs can be further enhanced by incorporating emerging digital tools and technologies. These include online educational tools offered by industry that allow patients to better understand trade-offs; advanced wearable systems that objectively capture real-world visual behavior; and visual simulators that enable patients to experience the anticipated visual performance with simultaneous vision IOLs [19, 22].

To reestablish far-distance contrast sensitivity and mitigate complaints of dysphotopsia in patients undergoing implantation with IOLs that increase the DOFi, it is crucial to objectively assess the level of crystalline lens sclerosis or cataract [18]. Using objective cataract grading metrics such as the Ocular Scatter Index (OSI) [23], Dysfunctional Lens Index (DLI) [24], and Pentacam Nucleus Staging (PNS) can help predict the postoperative restoration of contrast sensitivity [25]. Specifically, cutoff values for OSI > 1.25, DLI < 7.67, and PNS > 1, combined with patient age > 62 years, are important thresholds, indicating a higher probability of the patient to have a better contrast sensitivity (≥ 0.8 logCS at high spatial frequency) at far distance with a multifocal intraocular lens than the preoperative contrast sensitivity affected by crystalline lens sclerosis [26]. In addition, positive dysphotopsia, which includes bothersome visual artifacts around bright lights, might also be prevented by these cataract-grading methods. An OSI < 2.8, DLI > 7.6, or PNS < 2 might suggest increased susceptibility to bothersome photic phenomena (dysphotopsia) postoperatively if the patient already complains preoperatively, since there is not an optical reason before surgery that explains the increase in photic phenomena size [27]. However, these cutoff points for positive dysphotopsia were derived on the basis of correlations with a single 5-point Likert-scale question owing to the absence of a validated, dysphotopsia-specific questionnaire comprising multiple items that capture varying levels of severity and that have been validated using Rasch analysis [28].

Consequently, evaluating both contrast sensitivity and dysphotopsia risk using these cataract grading indices can guide surgeons in selecting appropriate candidates for IOL implantation, managing patient expectations and potentially reducing postoperative visual complaints.

Conclusions

As described above, evidence-based IOL selection from functional classification is a top–down process and serves as an evidence-based tool from which the surgeon can enter the adoption of new IOL technologies with prior knowledge of the range of visual acuity and spectacle independence outcomes to be expected with such IOLs. While clinical endpoints remain the gold standard for guiding patient-centered decision-making, it is important to acknowledge that optical bench testing plays a complementary role in the preliminary assessment of early prototypes and design performance [2].

Once a general functional category has been selected, a more detailed analysis of the individual evidence for each specific IOL model is required. In this stage, the surgeon must assess various biometric and clinical factors [29–31], such as pupil size [32–36], axes of the eye [37], corneal aberrations [38–40], corneal astigmatism [41–43], intraocular lens positioning [44, 45], posterior capsule opacification [46, 47], and the presence of ocular comorbidities [48, 49], among others, that may influence the probability of achieving the expected outcomes. Careful consideration of these individual variables is essential to refine the IOL choice and to maximize the likelihood of meeting patient expectations in clinical practice.

Author Contributions

All authors contributed to the study’s conception and design. Material preparation, data collection, and analysis were carried out by Joaquín Fernández and Manuel Rodríguez-Vallejo. Joaquín Fernández and Manuel Rodríguez-Vallejo drafted the initial manuscript, while Filomena Ribeiro, Burkhard Dick, and Carlos Rocha-de-Lossada reviewed and refined subsequent versions. All authors read and approved the final manuscript.

Funding

No funding or sponsorship was received for this study or publication of this article.

Data Availability

Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.

Declarations

Conflicts of Interest

Carlos Rocha-de-Lossada is an Editorial Board member of Ophthalmology and Therapy. Carlos Rocha-de-Lossada was not involved in the selection of peer reviewers for the manuscript nor any of the subsequent editorial decisions. Joaquín Fernández, Filomena Ribeiro, H. Burkhard Dick, and Manuel Rodríguez-Vallejo have no competing interests to declare.

Ethical Approval

This article is based on previously conducted studies and does not contain any new studies with human participants or animals performed by any of the authors.

References

  • 1.Fernández J, Ribeiro F, Rocha-de-Lossada C, Rodríguez-Vallejo M. Functional classification of intraocular lenses based on defocus curves: a scoping review and cluster analysis. J Refract Surg. 2024;40:e108–16. [DOI] [PubMed] [Google Scholar]
  • 2.Ribeiro F, Dick HB, Kohnen T, Findl O, Nuijts R, Cochener B, et al. Evidence-based functional classification of simultaneous vision intraocular lenses: seeking a global consensus by the ESCRS Functional Vision Working Group. J Cataract Refract Surg. 2024;50:794–8. [DOI] [PubMed] [Google Scholar]
  • 3.Riaz KM, Wendelstein JA, Koch DD. Depth of field or depth of focus? J Cataract Refract Surg. 2024;50:1291–2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Rampat R, Gatinel D. Multifocal and extended depth-of-focus intraocular lenses in 2020. Ophthalmology. 2021;128:e164–85. [DOI] [PubMed] [Google Scholar]
  • 5.Fernández J, Ribeiro FJ, Rodríguez-Vallejo M, Dupps WJ, Werner L, Srinivasan S, et al. Standard for collecting and reporting outcomes of IOL–based refractive surgery: update for enhanced monofocal, EDOF, and multifocal IOLs. J Cataract Refract Surg. 2022;48:1235–41. [DOI] [PubMed] [Google Scholar]
  • 6.Fernández J, Rocha-De-Lossada C, Rodríguez-Vallejo M. Recommendations for writing clinical research manuscripts: from monofocal to multifocal intraocular lenses. Int J Environ Res Public Heal. 2022;19:17036. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Kanclerz P, Toto F, Grzybowski A, Alio JL. Extended depth-of-field intraocular lenses: an update. Asia-Pac J Ophthalmol (Phila, Pa). 2020;9:194–202. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Daya SM. A discussion on modern IOL terminology. 2021. Available from: https://crstodayeurope.com/articles/feb-2021/lets-speak-lenspeak/
  • 9.Wu Y-H, Yu D, Goldstein JE, Kwon M, Gobeille M, Watson E, et al. Estimating visual acuity without a visual acuity chart. Transl Vis Sci Technol. 2024;13:20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Sharma G, Chiva-Razavi S, Viriato D, Naujoks C, Patalano F, Bentley S, et al. Patient-reported outcome measures in presbyopia: a literature review. Bmj Open Ophthalmol. 2020;5: e000453. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Lan Q, Liu Y, Xu F, Li M, Li Y, Yang T, et al. Cost-effectiveness of presbyopia correction among seven strategies of bilateral cataract surgery based on a prospective single-blind two-center trial in China. Ophthalmol Ther. 2022;11:2067–82. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Hays RD, MacRae S, Holladay J, Tarver ME, Lum F, Stark W, et al. Development of a patient-reported outcome measure to assess symptoms associated with cataract surgery and intraocular lens implants. Ophthalmology. 2023;130:715–25. [DOI] [PubMed] [Google Scholar]
  • 13.Yeu E, Cuozzo S. Matching the Patient to the intraocular lens preoperative considerations to optimize surgical outcomes. Ophthalmology. 2021;128:e132–41. [DOI] [PubMed] [Google Scholar]
  • 14.Rodríguez-Calvo-de-Mora M, Rocha-de-Lossada C, Rodríguez-Vallejo M, Cruz DZL, Pérez JF. Retinal detachment after refractive lens exchange: a narrative review. Arch Soc Española Oftalmol (Engl Ed). 2023;98:507–20. [DOI] [PubMed] [Google Scholar]
  • 15.Morlock R, Wirth RJ, Tally SR, Garufis C, Heichel CWD. Patient-Reported Spectacle Independence Questionnaire (PRSIQ): development and validation. Am J Ophthalmol. 2017;178:101–14. [DOI] [PubMed] [Google Scholar]
  • 16.Atkinson MJ, Tally S, Heichel CW, Kozak I, Leich J, Levack A. A qualitative investigation of visual tasks with which to assess distance-specific visual function. Qual Life Res. 2013;22:437–53. [DOI] [PubMed] [Google Scholar]
  • 17.Rodríguez-Vallejo M, Burguera N, Rocha-de-Lossada C, Aramberri J, Fernández J. Refraction and defocus curves in eyes with monofocal and multifocal intraocular lenses. J Optom. 2023;16:236–43. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Fernández J, Rodríguez-Vallejo M, Martínez J, Tauste A, Piñero DP. From presbyopia to cataracts: a critical review on dysfunctional lens syndrome. J Ophthalmol. 2018;2018:1–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Fernández J, Burguera N, Maldonado-Carmona C, Ginel J, Sáez-Martín Á, Rodríguez-Vallejo M. Simulations of multifocal vision in patients with previous monofocal intraocular lens implantation. J Refract Surg. 2023;39:831–9. [DOI] [PubMed] [Google Scholar]
  • 20.Kimlin JA, Black AA, Wood JM. Older drivers’ self-reported vision-related night-driving difficulties and night-driving performance. Acta Ophthalmol. 2020;98:e513–9. [DOI] [PubMed] [Google Scholar]
  • 21.Friedman DS, Tielsch JM, Vitale S, Bass EB, Schein OD, Steinberg EP. VF-14 item specific responses in patients undergoing first eye cataract surgery: can the length of the VF-14 be reduced? Br J Ophthalmol. 2002;86:885. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Sarkar S, Khuu SK, Vera-Diaz FA, Kang P. Validation of real-time viewing distance and ambient illumination measurement with two wearable sensors. Ophthalmic Physiol Opt. 2025;45:820–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Artal P, Benito A, Pérez GM, Alcón E, Casas ÁD, Pujol J, et al. An objective scatter index based on double-pass retinal images of a point source to classify cataracts. PLoS One. 2011;6: e16823. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Li Z, Yu L, Chen D, Chang P, Wang D, Zhao Y, et al. Dysfunctional lens index serves as a novel surgery decision-maker for age-related nuclear cataracts. Curr Eye Res. 2019;44:733–8. [DOI] [PubMed] [Google Scholar]
  • 25.Pan A-P, Wang Q-M, Huang F, Huang J-H, Bao F-J, Yu A-Y. Correlation among lens opacities classification system III grading, Visual Function Index-14, Pentacam nucleus staging, and objective scatter index for cataract assessment. Am J Ophthalmol. 2015;159:241-247.e2. [DOI] [PubMed] [Google Scholar]
  • 26.Fernández J, Burguera N, Rocha-de-Lossada C, Rachwani-Anil R, Rodríguez-Vallejo M. Objective cataract grading methods and expected contrast sensitivity reestablishment with multifocal intraocular lenses. Int Ophthalmol. 2023;43:2825–32 [DOI] [PubMed] [Google Scholar]
  • 27.Fernández J, Burguera N, Rocha-de-Lossada C, Zamorano-Martín F, Rodríguez-Vallejo M. Screening of positive dysphotopsia before multifocal intraocular lens implantation. Indian J Ophthalmol. 2024;72:S211–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Fernández J, Rodríguez-Vallejo M, Martínez J, Burguera N, Piñero DP. What we have learnt from 30 years living with positive dysphotopsia after intraocular lens implantation?: a review. Expert Rev Ophthalmol. 2021;16:1–10. [Google Scholar]
  • 29.Fernández J, Rodríguez-Vallejo M, Martínez J, Tauste A, Piñero DP. Biometric factors associated with the visual performance of a high addition multifocal intraocular lens. Curr Eye Res. 2018;43:998–1005. [DOI] [PubMed] [Google Scholar]
  • 30.Takabatake R, Takahashi M. Preoperative factors affecting visual acuity following the implantation of diffractive multifocal intraocular lenses. J Refract Surg. 2021;37:674–9. [DOI] [PubMed] [Google Scholar]
  • 31.Liu Y, Wei D, Bai T, Luo J, Wood J, Vashisht A, et al. Using machine learning to predict post-operative depth of focus after cataract surgery with implantation of Tecnis Symfony. Eur J Ophthalmol. 2021;112067212199177. [DOI] [PubMed]
  • 32.Fernández J, Rodríguez-Vallejo M, Martínez J, Burguera N, Piñero DP. Pupil dependence assessment with multifocal intraocular lenses through visual acuity and contrast sensitivity defocus curves. Eur J Ophthalmol. 2020;112067212094020. [DOI] [PubMed]
  • 33.Ang RET, Khoramnia R, Uy HS, Hoffmann P, Anello RD, Zalewski D, et al. Comparison of visual acuity outcomes of enhanced monofocal versus standard monofocal intraocular lenses from a randomized, multicenter, active-controlled trial. J Refract Surg. 2025;41:e300–9. [DOI] [PubMed] [Google Scholar]
  • 34.León-Ibáñez L, Piñero DP. Analysis of the visual function with the continuous transitional focus intraocular lens Precizon Presbyopic. Eur J Ophthalmol. 2024;11206721241293830. [DOI] [PubMed]
  • 35.Marín JM, Hervella L, Villegas E, Robles C, Alcón E, Yago I, et al. Visual performance at all distances and patient satisfaction with a new aspheric inverted meniscus intraocular lens. J Refract Surg. 2023;39:582–8. [DOI] [PubMed] [Google Scholar]
  • 36.Auffarth GU, Gerl M, Tsai L, Janakiraman DP, Jackson B, Alarcon A, et al. Clinical evaluation of a new monofocal IOL with enhanced intermediate function in patients with cataract. J Cataract Refr Surg. 2021;47:184–91. [DOI] [PubMed] [Google Scholar]
  • 37.Fernández J, Ribeiro F, Burguera N, Burguera-Giménez N, Rodríguez-Vallejo M. Replacing theories with evidence around the axes of the eye in intraocular lens selection: a review for the clinician. Ophthalmol Ther. 2025. [DOI] [PMC free article] [PubMed]
  • 38.Fernández J, Burguera N, Calvo-de-Mora MR, Rocha-de-Lossada C, Rodríguez-Vallejo M. Influence of corneal higher order aberrations on visual performance with a multifocal intraocular lens. J Refract Surg. 2025;41:e291–9. [DOI] [PubMed] [Google Scholar]
  • 39.Danzinger V, Schartmüller D, Lisy M, Schranz M, Abela-Formanek C, Menapace R, et al. Prospective fellow-eye comparison of a spherical monofocal and an enhanced monofocal intraocular lens in intermediate and near vision. J Refract Surg. 2025;41:e382–90. [DOI] [PubMed] [Google Scholar]
  • 40.Savini G, Galzignato A, Coutinho CP, Huang J, Wu Y, Barboni P, et al. Influence of preoperative variables on the 3-month functional outcomes of the Vivity extended depth-of-focus intraocular lens: a prospective case series. Eye Vis. 2025;12:8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Rocha-de-Lossada C, Rodríguez-Vallejo M, Rodríguez-Calvo-de-Mora M, Ribeiro FJ, Fernández J. Managing low corneal astigmatism in patients with presbyopia correcting intraocular lenses: a narrative review. BMC Ophthalmol. 2023;23:254. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Pérez-Sanz L, Vega F, Azor-Morón JA, Cuellar F, Millán MS, Garzón N. Tolerance to residual astigmatism of an isofocal intraocular lens. Graefe’s Arch Clin Exp Ophthalmol. 2023;1–12. [DOI] [PubMed]
  • 43.Zheng H, Tan Q, Zhou S, Luo W, Ortega-Usobiaga J, Wang L, et al. The tolerance of refractive errors of extended depth of focus intraocular lens in patients with previous corneal refractive surgery. Int Ophthalmol. 2023;1–9. [DOI] [PubMed]
  • 44.Fernández J, Burguera N, Rocha-de-Lossada C, Rachwani-Anil R, Rodríguez-Vallejo M. Influence of a multifocal intraocular lens centration and eye angles on light distortion and ocular scatter index. Graefe’s Arch Clin Exp Ophthalmol. 2023;261:2291–9. [DOI] [PubMed] [Google Scholar]
  • 45.Zhang Y, Zhang J, Liang C, Xu Y, Wu J, Wu J, et al. Effect of crystalline lens decentration and tilt on visual performance in eyes implanted with bifocal or extended depth of focus intraocular lenses. BMC Ophthalmol. 2025;25:58. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Fernández J, García-Montesinos J, Martínez J, Piñero DP, Rodríguez-Vallejo M. Posterior capsular opacification evaluation through contrast sensitivity defocus curves with two multifocal intraocular lenses of similar material. Graefe’s Archive Clin Exp Ophthalmol. 2021;1–8. [DOI] [PubMed]
  • 47.Chang H, Fang Q, Liu X. Influencing factors and prediction model construction of posterior capsular opacification after intraocular lens implantation treated with Nd: YAG laser. BMC Ophthalmol. 2025;25:188. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Fernández J, Ribeiro F, Burguera N, Rodríguez-Calvo-de-Mora M, Rodríguez-Vallejo M. Visual and patient-reported outcomes of an enhanced versus monofocal intraocular lenses in cataract surgery: a systematic review and meta-analysis. Eye. 2025;1–16. [DOI] [PMC free article] [PubMed]
  • 49.Álvarez-García MT, Fuente-García C, Muñoz-Puyol C, Piñero DP. Clinical outcomes with extended depth of focus intraocular lenses in cases in which multifocal lenses are not primarily recommended. J Ophthalmol. 2023;2023:8814627. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.


Articles from Ophthalmology and Therapy are provided here courtesy of Springer

RESOURCES