. 2025 Aug 22;15(8):e095640. doi: 10.1136/bmjopen-2024-095640

Table 3. Comparative diagnostic accuracy of intra-individual robust coefficient of variation for peripheral blood neutrophil myeloperoxidase expression between a single-use flow cytometric lyophilised reagent tube (BD Lyotube Stain 468) and its laboratory-developed liquid reagent counterpart (n=100)*.

BD Lyotube Stain 468 Laboratory-developed liquid reagent test
Intra-individual RCV, median (range), %
 Confirmed suspicion of MDS (n=37) 34.9 (25.0–69.4) 34.5 (25.9–60.3)
 Unconfirmed suspicion of MDS (n=63) 29.0 (23.8–44.6) 29.8 (23.9–48.5)
RCV≥30%
 True positive, n 33 (…) 35 (…)
 False positive, n 23 (…) 29 (…)
 False negative, n 4 (…) 2 (…)
 True negative, n 40 (…) 34 (…)
 Sensitivity, % (95% CI) 89 (75 to 97) 95 (82 to 99)
 Specificity, % (95% CI) 63 (50 to 75) 54 (41 to 67)
 PPV, % (95% CI) 59 (45 to 72) 55 (42 to 67)
 NPV, % (95% CI) 91 (78 to 97) 94 (81 to 99)
 LR+ (95% CI) 2.44 (1.73 to 3.45) 2.05 (1.56 to 2.71)
 LR− (95% CI) 0.17 (0.07 to 0.44) 0.10 (0.03 to 0.39)
*

The analytical sample consisted of 37 confirmed and 63 unconfirmed suspicions of myelodysplastic neoplasm, after excluding three patients with uninterpretable bone marrow cytomorphology.

LR+, likelihood ratio of a positive result; LR−, likelihood ratio of a negative result; MDS, myelodysplastic neoplasm; NPV, negative predictive value; PPV, positive predictive value; RCV, robust coefficient of variation.