Table 1.
Operating conditions of HPLC-ICP-MS
| ICP-MS operating parameters | HPLC operating parameters | ||
|---|---|---|---|
| RF power | 1600 W | Mobile phase composition | 2 mM 1-octanesulfonic acid sodium salt, 2 mM malonic acid, 4 mM tetramethylammonium hydroxide solution |
| Plasma gas flow | 16.0 L/min | Mobile phase pH | 4.13 (± 0.02) |
| Auxiliary (makeup) gas flow | 1.2 L/min | Mobile phase flow rate | 1 mL/min |
| Nebulizer (carrier) gas flow | 0.98 L/min | Injection volume | 20 µL |
| Nebulizer type | Meinhard | Degasser | On |
| Monitored ion | m/z 91 (75As 16O+) | Column temperature | Ambient |
| Acquisition time | 240 s (4 min) | ||