Skip to main content
. Author manuscript; available in PMC: 2026 Jun 2.
Published in final edited form as: Radiat Phys Chem Oxf Engl 1993. 2025 Jun 2;237:113013. doi: 10.1016/j.radphyschem.2025.113013

Table 2.

Mean organ dose calculated by MC-GPU for the incomplete patient anatomy and the extended version of the patient for the 8-year-old obese male abdominal CT scan.

Calculated Dose (mGy)**
Incomplete CT Extended CT
Spleen 23.7 24.1
Kidneys * 21.2 21.4
Liver 24.7 25.4
Stomach 23.5 23.9
Pancreas 20.8 21.1
Lungs * 20.2 13.4
Esophagus 17.0 8.5
Heart 16.8 16.2
Colon 22.4 22.5
**

Monte Carlo statistical error for all organs of interest (<1%).

*

Refers to left and right combined.