Abstract
Purpose
This study aims to evaluate rotational stability and patient-reported outcomes associated with Tecnis Eyhance toric II “monofocal plus” intraocular lens (IOL), particularly in light of a modified haptic design.
Methods
A prospective, interventional study was conducted with patients undergoing routine cataract surgery and Tecnis Eyhance toric II IOL implantation. The inclusion criteria were topographic corneal astigmatism of 0.5–4.0D. Exclusion criteria included previous corneal or ocular surgeries, mesopic pupils >5.5mm, wear of rigid gas-permeable contact lenses, and ophthalmic pathologies that might limit vision outcomes. Toric stability was assessed on day 1, week 1, and month 3 using slit-lamp exams and toriCAM app. Quality of Vision (QoV) and Near Activity Visual Questionnaires (NAVQ) before the surgery and 3 months post-operatively.
Results
The study analyzed 98 eyes from 65 patients. Mean absolute IOL rotation was 1.73±1.64° (range: 0–7°) from day 1 to week 1, with 95% rotating ≤5°. From week 1 to month 3, mean rotation was 1.13±1.53° (range: 0–8°), with 98% rotating ≤5°.
Conclusion
The data from this study suggest that the Tecnis Eyhance toric II IOL exhibits good postoperative rotational stability and positive patient-reported visual outcomes.
Keywords: rotational stability, visual acuity, subjective results, toric IOL
Introduction
Cataracts and corneal astigmatism are prevalent conditions that significantly affect visual acuity and quality of life. Approximately 30% of individuals presenting for cataract surgery have corneal astigmatism of 0.75 diopters (D) or more, which can compound the visual impairment caused by cataracts.1 In the past, traditional cataract surgery with standard intraocular lenses (IOLs) would not address corneal astigmatism, often necessitating additional refractive procedures or the use of corrective eyewear postoperatively.2
Toric IOLs have solidified their place as the primary method of addressing corneal astigmatism at the time of cataract surgery.3 However, optimal correction is dependent on the rotational stability within the eye as rotational misalignment can diminish or alter the intended astigmatic correction, potentially resulting in suboptimal visual results for patients.4 Depending on the implanted toric power, every degree of rotation may reduce cylindrical correction by as much as 3.3%, leading to residual astigmatism and reduced visual acuity. Ten or more degrees of misalignment may necessitate surgical repositioning.5–7 Rotation is presumed to occur as a result of the instability of the IOL within the pre-fibrotic capsular bag.4 Other contributing factors may include cyclotorsion of the eye and its effect on planning or alignment, imprecise positioning of the IOL, IOL haptic design, IOL material, capsulorhexis size, and capsular bag shrinkage.6 In addition to reduced uncorrected acuity, patients with rotated toric IOLs may experience visual disturbances such as distorted vision, ghosting or monocular diplopia.8 Misalignment can also cause higher-order aberrations.5 Ultimately, these problems can result in decreased patient satisfaction and a possible need for re-operation. The goal of the modern toric IOL design is to minimize these risks and enhance the stability and predictability of the refractive outcomes.
The Tecnis Eyhance Toric II “monofocal plus” IOL aims to combine the benefits of toric correction with an extended range of intermediate vision via a graduated power increase in the central optical zone of the lens. The design aims to provide improved intermediate vision while maintaining high-quality distance vision. A recent design modification was introduced with haptic edges left frosted rather than polished to increase capsular bag friction and improve rotational stability. Post-market analysis is required to validate the effect of this change.4,9
This study assesses the rotational stability and visual outcomes of the Tecnis Eyhance Toric II IOL in a real-world clinical environment.
Methods
Study Design and Objectives
This prospective observational study assessed the rotational stability and subjective visual quality outcomes of the Tecnis Eyhance Toric II Extended Depth-of-Focus IOL in patients undergoing cataract surgery by 4 surgeons (GM, HO, SW, KV). The study received ethics approval from an independent review board, Veritas. Consent was obtained from all participants after providing them with detailed information about the study. This study adheres to the Declaration of Helsinki. This research was supported through an investigator-initiated study grant from Johnson & Johnson Vision. The primary outcome was rotational stability of the IOL during the first three postoperative months. Secondary outcomes included subjective visual quality outcomes before surgery and 3 months postoperatively.
Participants
Individuals who were 21 years of age or older, scheduled for routine cataract surgery, and had preoperative corneal astigmatism of 0.5 to 4.0 diopters were included. The exclusion criteria included prior corneal or intraocular surgery, mesopic pupil sizes exceeding 5.5 mm, recent use of rigid gas-permeable contact lenses, pre-existing macular pathology, intraocular hypertension and poorly controlled diabetes were excluded from the study.
The IOL
The Tecnis Toric EDof IOL is a foldable, UV light-absorbing, hydrophobic acrylic one-piece IOL with a C-loop haptic. It is available in spherical powers of +5.00D to +34.00D in 0.5D increments, with cylinder power from 1.00D to 8.00D. The lens is preloaded in the Tecnis Simplicity Delivery System injector with frosted haptics, as discussed, intended to increase rotational stability.10 The power of the IOL was calculated using the toric IOL calculator provided by Johnson & Johnson Vision online. The target refraction aimed for emmetropia, set between 0 and −0.5 D. The selection was based on the smallest predicted residual cylinder along the correct axis to minimize the risk of overcorrection and subsequent axis inversion.
Surgical Procedure
All participants underwent standard phacoemulsification cataract surgery through a 2.4 mm wound without the use of a femtosecond laser. The Verion Image Guided System (Alcon) was used to capture preoperative anatomic landmarks, and images were transferred to the Verion Digital Marker. Under local anesthesia, a temporal clear cornea incision was used in each case, with surgeons adjusting for their own surgically induced astigmatism (SIA). An anterior capsulorhexis of 5 mm was created, and the IOL was placed in the capsular bag according to the manufacturer’s guidelines and rotated to the predetermined axis using the Verion system’s intraoperative digital toric marking system.
Outcome Measures
The rotational stability of the IOL after surgery was measured by comparing the alignment of the IOL’s at one day, one week, and three months post-operatively, using the toriCAM smartphone application. Patients were examined at the slit lamp with the head adjusted to be in vertical alignment and fixated on a distance target with the fellow eye. With the pupil dilated to expose paired IOL toric markings, a narrow-slit beam was created and rotated to intersect both markings. The orientation of this beam in degrees was then recorded by overlaying it onto the ToriCAM application alignment screen. The patient-reported visual quality was assessed using the Quality of Vision (QoV) questionnaire and the Near Activity Visual Questionnaire (NAVQ) three months postoperatively. The QoV questionnaire was used to evaluate the frequency, severity, and degree of bother associated with visual disturbances such as glare, halos, starbursts, hazy vision, blurred vision, distortion, double/multiple images, fluctuation in vision, focusing difficulties, and difficulties in judging distance or depth perception. Each symptom was evaluated based on frequency, severity, and bothersomeness, with raw data scaled from 0 to 100, where 0 indicates no symptoms and better visual quality.
The NAVQ assessed functional outcomes related to near activities, including reading, using electronic devices, and other close-up tasks, by asking patients to rate their difficulty with each activity. With 11 questions rated from 0–3, 0 indicating no difficulty and 3 representing extreme difficulty. The questionnaire is scored from 0–33, with a lower score indicating better visual quality.
Statistical Analysis
Participant demographics and baseline characteristics were summarized using descriptive statistics. To evaluate the statistical significance of the change of absolute rotation over time, we employed the Kruskal–Wallis Test. Mean, interquartile range, the range between the first and third quartile (IQR), and outliers were calculated.
Results
Patient Demographics and Baseline Characteristics
This study included 98 eyes from 65 patients, 53 OD (54%), 45 OS (46%). The mean age was 68.47 ± 9.28 years. There were 36 male and 29 female patients (Table 1). No capsular tension rings were used.
Table 1.
Patient Characteristics
| Demographics | |
|---|---|
| Eyes | 98 |
| Patients | 65 |
| Age (Mean ± SD) | 68.47 ± 9.28 |
| Gender (Male/Female) | 36/29 |
| Laterality (Right/Left) | 53/45 |
Rotational Stability
In Figure 1A, the Target IOL Axis to Day 1 shows a mean change of 2.99±3.49 degrees, with a median of 2.0 degrees. Several outliers were observed, with values of 14, 11, four occurrences of 9, and three occurrences of 15 degrees. In Figure 1B, there was a mean change of 1.73±1.64 degrees and a median change of 1.0 degrees from Day 1 to Week 1. The 1st quartile (Q1) was 0.75 degrees, the 3rd quartile (Q3) was 3.0 degrees, and the IQR was 2.25 degrees with one outlier at 7 degrees. 95% of eyes had an IOL rotation of ≤5 degrees. From Week 1 to Month 3, the mean change was 1.13±1.52 degrees and a median change of 1.0 degrees was found. Q1 was 0.0 degrees, Q3 was 2.0 degrees, and the IQR was 2.0 degrees with two outliers at 6 and 8 degrees. 98% of eyes had an IOL rotation of ≤5 degrees. From Day 1 to Month 3, the total lens rotation was 1.90 ± 2.03 degrees. All time intervals indicated no statistical significance (p-value >0.05).
Figure 1.
Boxplot analysis- (A) Difference in IOL Axis between Target and Day 1, (B) Change in absolute IOL axis rotation for two intervals: Day 1 to Week 1 and Week 1 to Month 3. The box plot illustrates the mean, median, IQR, and outliers (1.5xIQR+Q3). The whiskers extend to the remainder of the distribution, with outliers depicted separately.
The IOL rotated clockwise in 30 eyes (31%, 2 ± 1.51°), counterclockwise in 42 eyes (43%, −2 ± 1.57°) and did not rotate in 26 eyes (26%) between Week 1 and Day 1. Between Month 3 and Week 1, the IOL rotated clockwise in 30 eyes (31%, 1.6 ± 0.86°), counterclockwise in 23 eyes (23%, −3 ± 1.96°) and did not rotate in 45 eyes (46%) (Table 2).
Table 2.
Absolute Rotation Values and Distribution of Rotation Direction
| Observation Period | Absolute IOL Rotation in Degree | ≤5° (%) | P value | Absolute Number of IOLs Rotating Clockwise: Counter-Clockwise: No Change | |
|---|---|---|---|---|---|
| Median (Range) | Mean±SD | Distribution | |||
| Target IOL Axis to Day 1 | 2.0 (0–15) | 2.99±3.49 | 84 | >0.05 | 48:26:24 |
| Day 1–Week 1 | 1.0 (0–7) | 1.2±0.9 | 95 | >0.05 | 30:42:26 |
| 1 week–Month 3 | 1.0 (0–8) | 1.1±0.8 | 98 | >0.05 | 30:23:45 |
Patient Reported Outcomes
Patients completed the QoV questionnaire and the NAVQ at the three-month follow-up (Tables 3 and 4, respectively). The QoV questionnaire results showed positive results for all categories.
Table 3.
Quality of Vision Questionnaire Responses
| Subscale | Postoperative Mean ± SD |
|---|---|
| Frequency | 40.2 ± 10.1* |
| Severity | 38.6 ± 11.5* |
| Bothersome | 39.5 ± 12.6* |
Note: (*) indicates a significant value p<0.05.
Table 4.
Near Activity Visual Questionnaire Responses
| Question | Postoperative Mean ± SD |
|---|---|
| 1. Reading small print (eg, newspaper) | 0.8 ± 0.4 |
| 2. Reading large print (eg, books) | 0.7 ± 0.3 |
| 3. Reading very large print (eg, signs) | 0.6 ± 0.2 |
| 4. Seeing objects close up (eg, needlework) | 0.9 ± 0.5 |
| 5. Seeing objects in the distance (eg, driving) | 0.5 ± 0.3 |
| 6. Recognizing faces | 0.4 ± 0.2 |
| 7. Using a computer or digital screen | 0.6 ± 0.3 |
| 8. Seeing at night | 1.0 ± 0.4 |
| 9. Seeing in dim light | 0.9 ± 0.3 |
| 10. Performing near tasks (eg, cooking) | 0.7 ± 0.3 |
| 11. General satisfaction with near vision | 0.8 ± 0.4 |
| Overall Score | 6.95 ± 1.5 |
Discussion
This study evaluated the rotational stability and patient-reported outcomes following cataract extraction and insertion of the Tecnis Eyhance Toric II IOL in 65 patients (98 eyes). The mean absolute rotation of the IOL was 1.73 ± 1.64° from day 1 to week 1 and 1.13 ± 1.53° from week 1 to month 3, with 95% and 98% of eyes demonstrating rotations of ≤5° in these intervals, respectively. Additionally, patient-reported outcomes showed significant improvements in visual quality and near-activity performance.
Rotational Stability
A determining factor in the success of toric IOLs is rotational stability, as misalignment from the intended axis reduces the effectiveness of the lens in correcting astigmatism. This study showed consistent rotational stability, with an observed rotation of 1.90 ± 2.03 degrees from Day 1 to Month 3, comparable to the results reported by Zeilinger et al for the Tecnis Toric II IOL who reported rotational stability as 1.34 ± 1.46 degrees.11
Various factors influence the rotational stability of a toric IOL, including IOL interaction with the capsular bag, IOL material, and haptic design. Studies suggest that the capsular bag fibrosis typically stabilizes within 3 months.12 Studies have found that achieving a strong adhesion between the IOL and the capsular bag helps to minimize rotation.13,14 In a paper comparing the rotational stability of toric IOLs, unpolished surfaces demonstrated more rotational stability compared to polished surfaces of IOLs with made with identical design and material, likely due to the higher coefficient of friction between the surface of the lens and the capsule.15 Our findings revealed no significant changes in IOL rotation from day 1 to month 3, which is in line with the typical stabilization period observed with capsular bag fibrosis. Adhesion to the capsular bag is heavily determined by the material of the IOL. The lens and haptics are made of soft, foldable, UV-light absorbing, hydrophobic acrylic material, which has been found to have the highest levels of adhesion when compared to other materials, such as PMMA and silicone.15–17
Furthermore, research has highlighted that extracellular matrix proteins such as fibronectin, type IV collagen and vitronectin play a role in adhesion and these proteins are seen in higher levels in acrylics compared to other materials. These factors likely contributed to the rotational stability of the Tecnis Toric II IOL. Takaku et al (2021) observed a significant decrease in axis misalignment and uncorrected astigmatism with the Tecnis Toric II featuring frosted haptics compared to the previous model, Tecnis Toric I, with smooth haptics.17 They reported that the mean axis misalignment significantly decreased from 3.8 ± 2.7° with the Tecnis Toric I to 1.9 ± 1.3° with the Tecnis Toric II (p=0.04).17 Osawa et al (2022) highlighted that modified haptics can significantly enhance surgical outcomes and rotational stability, reinforcing our findings.15,18,19
Connell et al (2025) compared the rotational stability of the Eyhance Tecnis Toric II with the Clareon Toric IOL. Stability between the two lenses was comparable. Additionally, as found in this study, Connell et al did not find any significant rotation in the Tecnis Toric II between day 1 and month 6.20
Similarly, Jung et al (2024) found minimal rotation at month 3, with a mean misalignment of 0.68 ± 4.17°, and average absolute misalignment was 3.41 ± 2.44°.21
Patient Reported Outcomes
This study, using the QoV and NAVQ questionnaires, demonstrated that the Tecnis Eyhance toric IOL provides significant improvements in both quality of vision and near vision function. To our knowledge, this is the first study to report patient-reported outcomes alongside rotational stability. These findings indicate a substantial reduction in visual disturbances such as glare, halos, and distortion, leading to higher patient satisfaction.11
Similarly, the NAVQ scores underscore the effectiveness of the Tecnis Eyhance toric IOL in enhancing near vision for daily tasks such as reading and using digital devices.22 It was observed that patients with rotational instability exceeding 5 degrees experienced compromised patient-reported outcomes. Auffarth et al (2020) reported that the enhanced monofocal IOL improved intermediate vision performance while maintaining distance vision and managing photic phenomena, based on Catquest-SF9 results.23 Similarly, Visser et al found in a randomized clinical trial that toric IOLs resulted in higher rates of spectacle independence for distance vision compared to aspherical IOLs in patients with cataracts and corneal astigmatism, benefiting daily life activities.24 Moreover, Chang et al noted that 92.5% of surgeons were satisfied with the Tecnis Toric II IOL, reflecting its overall clinical success.25
Study Limitations and Conclusion
Possible limitations to our study include the unmeasured effects of tilt. The optical performance of toric IOLs can be marginally impacted by tilt, potentially leading to post-operative astigmatic error. Incorporating tilt measurements in future research would improve the assessment of IOL positioning, resulting in a more thorough analysis. Furthermore, expanding the study to encompass a larger and more diverse patient population and a singular surgeon with the same SIA could yield more comprehensive data on the performance of the Eyhance Toric II IOL across various demographics and ocular conditions.
Overall, our analysis indicates that the Eyhance Toric II IOL exhibits strong performance in terms of rotational stability and patient-reported outcomes. The design of this device appears to enhance its capability to maintain proper alignment post-surgery and improve patient-reported outcomes, making it a valuable addition to the existing IOL options for cataract surgery.
Disclosure
H. O’Donnell reports consulting agreement with Alcon and Bausch + Lomb, outside the submitted work. G. Moloney is a consultant for Kowa Ind. manufacturers of Glanatec (ripasudil) and has received a research grant from Johnson and Johnson. The remaining authors have no conflicts of interest to disclose for this work.
References
- 1.Núñez MX, Henriquez MA, Escaf LJ, et al. Consensus on the management of astigmatism in cataract surgery. Clin Ophthalmol. 2019;13:311. doi: 10.2147/OPTH.S178277 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Hamel C, Bailey S. Intraocular lenses for cataract surgery. Intraocular Lenses Cataract Surg CADTH Heal Technol Rev. 2023. Available from: https://www.ncbi.nlm.nih.gov/books/NBK599976/. Accessed November 18, 2024. [PubMed]
- 3.Lin X, Ma D, Yang J. Insights into the rotational stability of toric intraocular lens implantation: diagnostic approaches, influencing factors and intervention strategies. Front Med. 2024;11:1349496. doi: 10.3389/FMED.2024.1349496 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Singh A, Kapoor G, Baranwal VK, Kalra N. Rotational stability of toric intraocular lenses. Med J Armed Forces India. 2022;78(1). doi: 10.1016/J.MJAFI.2020.03.014 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Oshika T, Inamura M, Inoue Y, et al. Incidence and outcomes of repositioning surgery to correct misalignment of toric intraocular lenses. Ophthalmology. 2018;125(1):31–35. doi: 10.1016/J.OPHTHA.2017.07.004 [DOI] [PubMed] [Google Scholar]
- 6.Park RB, Aref AA. Astigmatism management in modern cataract surgery. Vision. 2024;8(1):9. doi: 10.3390/VISION8010009 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Kim CY, Chung SH, Kim TI, Cho YJ, Yoon G, Seo KY. Comparison of higher-order aberration and contrast sensitivity in monofocal and multifocal intraocular lenses. Yonsei Med J. 2007;48(4):627. doi: 10.3349/YMJ.2007.48.4.627 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Buratto L, Brint SF, Boccuzzi D. Toric Intraocular Lenses. Cataract Surg Intraocular Lenses. 2023;39–52. doi: 10.1201/9781003522874-8 [DOI] [Google Scholar]
- 9.Negishi K, Masui S, Ayaki M, Torii H, Yotsukura E, Nishi Y. Clinical results and factors affecting visual function in eyes implanted with an enhanced monofocal intraocular lens. Clin Ophthalmol. 2023;17:3965–3973. doi: 10.2147/OPTH.S438599 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Tecnis Eyhance Toric II IOL - Johnson & Johnson Vision. De Ceunynck Ophthalmology. 2024. [Google Scholar]
- 11.Zeilinger J, Hienert J, Ruiss M, Pilwachs C, Findl O. Rotational stability of a new toric intraocular lens with an advanced optical profile. J Cataract Refract Surg. 2023;49(6):584–588. doi: 10.1097/J.JCRS.0000000000001158 [DOI] [PubMed] [Google Scholar]
- 12.Chen L, Zhong Y, Yao K, Fu Q. Effect of intraocular lens material and haptic design on anterior capsule contraction after cataract surgery: a systematic review and meta-analysis. Graefes Arch Clin Exp Ophthalmol. 2024;262(5):1421–1432. doi: 10.1007/S00417-023-06230-X [DOI] [PubMed] [Google Scholar]
- 13.Vukich JA, Ang RE, Straker BJK, et al. Evaluation of intraocular lens rotational stability in a multicenter clinical trial. Clin Ophthalmol. 2021;15:3001–3016. doi: 10.2147/OPTH.S309214 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Felipe A, Artigas JM, Díez-Ajenjo A, García-Domene C, Alcocer P. Residual astigmatism produced by toric intraocular lens rotation. J Cataract Refract Surg. 2011;37(10):1895–1901. doi: 10.1016/J.JCRS.2011.04.036 [DOI] [PubMed] [Google Scholar]
- 15.Vandekerckhove K. Rotational stability of monofocal and trifocal intraocular toric lenses with identical design and material but different surface treatment. J Refract Surg. 2018;34(2):84–91. doi: 10.3928/1081597X-20171211-01 [DOI] [PubMed] [Google Scholar]
- 16.Kramer BA, Hardten DR, Berdahl JP. Rotation characteristics of three toric monofocal intraocular lenses. Clin Ophthalmol. 2020;14:4379. doi: 10.2147/OPTH.S285818 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Takaku R, Nakano S, Iida M, Oshika T. Influence of frosted haptics on rotational stability of toric intraocular lenses. Sci Reports 2021 111. 2021;11(1):1–7. doi: 10.1038/s41598-021-94293-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Osawa R, Sano M, Yuguchi T, Kaiya T, Oshika T. Effects of modified haptics on surgical outcomes and rotational stability of toric intraocular lens implantation. J Refract Surg. 2022;38(10):648–653. doi: 10.3928/1081597X-20220715-02 [DOI] [PubMed] [Google Scholar]
- 19.Pastor-Pascual F, Orts-Vila P, Tañá-Sanz P, Tañá-Sanz S, Ruiz-Mesa R, Tañá-Rivero P. Non-diffractive, toric, extended depth-of-focus intraocular lenses in eyes with low corneal astigmatism. Eye Vis. 2024;11(1):1–11. doi: 10.1186/S40662-024-00380-7/FIGURES/5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Connell B, Lahood B. Comparing rotational stability of clareon and eyhance toric IOLs in a real-world setting. Clin Ophthalmol. 2025;19:141–149. doi: 10.2147/OPTH.S503337 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Jung H, Jun I, Lee HK, Seo KY, Kim TI. Real-world outcomes of cataract surgeries using a new type of enhanced monofocal toric intraocular lens. Sci Rep. 2024;14(1). doi: 10.1038/S41598-024-65746-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Sims J, Sloesen B, Bentley S, et al. Psychometric evaluation of the near activity visual questionnaire presbyopia (NAVQ-P) and additional patient-reported outcome items. J Patient-Reported Outcomes. 2024;8(1). doi: 10.1186/S41687-024-00717-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Auffarth GU, Gerl M, Tsai L, et al. Clinical evaluation of a new monofocal IOL with enhanced intermediate function in patients with cataract. J Cataract Refract Surg. 2021;47(2):184–191. doi: 10.1097/J.JCRS.0000000000000399 [DOI] [PubMed] [Google Scholar]
- 24.Visser N, Bauer NJC, Nuijts RMMA. Toric intraocular lenses: historical overview, patient selection, IOL calculation, surgical techniques, clinical outcomes, and complications. J Cataract Refract Surg. 2013;39(4):624–637. doi: 10.1016/J.JCRS.2013.02.020 [DOI] [PubMed] [Google Scholar]
- 25.Chang DH, Hu J, Miller KM, Vilupuru S, Zhao W. Post-market evaluation of rotational stability and visual performance of a new toric intraocular lens with frosted haptics. Clin Ophthalmol. 2022;16:4055. doi: 10.2147/OPTH.S389304 [DOI] [PMC free article] [PubMed] [Google Scholar]