. 2025 Aug 12;13(9):e03219-24. doi: 10.1128/spectrum.03219-24

Performance of novel digital real-time PCR for detection of SARS-CoV-2, respiratory syncytial viruses, and influenza viruses in Ghana

Michael Owusu 1,2,3,, Bernard Nkrumah 4, Godfred Acheampong 1, Stephen Opoku Afriyie 1, Ebenezer Kojo Addae 1, Richard Larbi 3, Richard Owusu Ansah 3, Chrysantus Kubio 5, Farouk Saeed 5, Nana Kwame Ayisi-Boateng 6, Eric Darko 6, Jacob Amonoo-Neizer 7, Abena Gyawu Owusu-Ansah 8, Frederick Ayensu 8, Peter Kojo Brenya 8, Veronica Bannor 9, Pawan Angra 10, Danielle Thompson Barradas 10
Editor: Rosemary C She11
PMCID: PMC12403756  PMID: 40792500

ABSTRACT

Digital PCR (dPCR) systems offer high sensitivity and reproducibility without requiring external control standards. However, their performance against real-time reverse transcription-PCR (rRT-PCR) for detecting respiratory viruses remains unexplored in Ghana. We therefore evaluated the performance of a novel dPCR, Lab-On-An-Array (LOAA), for the detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), respiratory syncytial virus (RSV), and influenza viruses type A (Flu A) and B (Flu B). A cross-sectional hospital-based study was conducted between August 2022 and January 2023 in Ghana’s Ashanti and Savannah Regions. Oropharyngeal swabs from 356 participants with a median age of 19 years, presenting with suspected respiratory illness, were tested using LOAA and rRT-PCR. Viral RNA was extracted using a Qiagen Viral Mini Kit (Qiagen Diagnostics GmbH, Germany). LOAA and rRT-PCR tests were performed using Genoplexor COVID-19/Flu/RSV Detection Kit (Optolane Technologies Inc, South Korea) and FluoroType SARS-CoV-2/Flu/RSV kits (Hain Lifescience GmbH, Germany), respectively. LOAA’s performance metrics were assessed using rRT-PCR as the gold standard. Overall positivity rates were 29.78% and 30.90% for LOAA and rRT-PCR, respectively. Compared to rRT-PCR, LOAA’s sensitivity was 87.76% for RSV, 91.30% for SARS-CoV-2, 86.21% for Flu B, and 88.89% for Flu A. Positive predictive value was the highest for RSV (97.73%) and lowest for Flu A (61.54%); negative predictive values were ≥98.00% for all respiratory viruses. LOAA recorded an “almost perfect” agreement (κ ≥0.88) with rRT-PCR for RSV, SARS-CoV-2, and Flu B and good agreement for Flu A (κ = 0.72). LOAA is sensitive in detecting SARS-CoV-2, RSV, and Flu B infections; however, minor improvements for Flu A are required.

IMPORTANCE

This study presents the potential of a digital PCR as a highly sensitive and reproducible tool for detecting respiratory viruses in Ghana, where robust diagnostic methods are essential for managing public health challenges. By evaluating the novel Lab-On-An-Array (LOAA) system, we provide its critical operational performance against the gold-standard rRT-PCR for detecting severe acute respiratory syndrome coronavirus 2, respiratory syncytial virus, and influenza viruses. Our findings show that LOAA demonstrates excellent agreement with rRT-PCR for most viruses, offering a promising alternative for respiratory virus surveillance and diagnosis. This research is particularly significant for resource-limited settings, as it supports the adoption of advanced molecular diagnostics to improve early detection and response to respiratory infections. Minor refinements for specific viruses, such as influenza A, could further enhance its utility in clinical and epidemiological applications.

KEYWORDS: respiratory viruses, SARS-CoV-2, RSV, influenza, Lab On An Array, RT-PCR, Ghana

INTRODUCTION

Respiratory viruses include a wide range of viruses that infect cells of the respiratory tract and are associated with significant morbidity and mortality worldwide. Respiratory viruses are responsible for about 90% of upper respiratory tract infections and 30% of lower respiratory tract infections (1). Common human respiratory viruses include influenza viruses, respiratory syncytial virus, adenoviruses, rhinoviruses, and coronaviruses. Respiratory viruses belong to diverse families and differ in their mode of transmission, seasonality of circulation, and disease severity (2). Since 2019, severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has emerged as an important respiratory virus accounting for over 776 million confirmed cases and 7.1 million deaths worldwide as of 16 July 2024 (3).

To contain and minimize the spread of infection, prompt and accurate diagnosis is of utmost importance to detect, isolate, treat, and trace infected individuals. Clinical diagnosis of SARS-CoV-2 infection is complicated by nonspecific symptoms such as fever, chills, cough, headache, shortness of breath, and sore throat that may be caused by other respiratory viruses or bacteria (4). The similarities in symptoms pose a challenge to clinicians; hence, the need for laboratory diagnosis to differentiate between infections caused by respiratory viruses, bacteria, and fungi.

Conventional viral diagnostic methods, such as viral culture, immunofluorescence assays, and serological techniques, are time-consuming, labor intensive, and more prone to cross-reactivity relative to emerging methods such as real-time reverse transcription PCR (rRT-PCR). Rapid diagnostic tests, such as lateral flow assays, are relatively inexpensive and easy to use; however, they are often of limited value in diagnostics due to poor sensitivity and specificity (5). Currently, the World Health Organization has established rRT-PCR, a nucleic acid amplification test, as the gold standard to detect SARS-CoV-2 (6). rRT-PCR makes use of target specific primers and probes to accurately distinguish between viral pathogens, thus conferring higher sensitivity and specificity to these types of assays. In addition, rRT-PCR is capable of high throughput screening and can provide an estimation of viral load by the use of cycle threshold (Ct) values with an appropriate standard or calibration curves, although it is important to avoid sole reliance on Ct values when estimating viral load, given that these values can be affected by other factors including inherent inter-run variability (7).

To compensate for the limitations of rRT-PCR for diagnosis, digital PCR (dPCR) assays are becoming increasingly popular for use in both research and diagnostic laboratory settings (8, 9). The dPCR method is similar to rRT-PCR in that specific primers and probes bind to complementary segments in the target organism genome (10). However, dPCR differs from rRT-PCR in that each reaction is divided into thousands of individual partitions with some partitions containing the target molecule while others are lacking. After thermal cycling, each partition is then analyzed for the presence or absence of a fluorescent signal. The ratio of partitions with a fluorescent signal to the total number of partitions is analyzed using Poisson distribution to calculate the copy number of the target molecule. As such, dPCR assays do not require standard curves for quantification and provide precise and comparatively more reproducible results than rRT-PCR (9). Unlike rRT-PCR, partitioning in dPCR ensures that thousands of reactions occur simultaneously, thus increasing tolerance to inhibitors and sensitivity to detect low copy numbers of target molecules (11, 12).

In this study, we evaluated the clinical performance of a Lab-On-An-Array (LOAA) digital real-time PCR analyzer (Optolane Technologies Inc, South Korea), a novel platform for the multiplex detection of four respiratory viruses—SARS-CoV-2, RSV, Flu A, and Flu B—using rRT-PCR as the gold standard. To date, the evaluation of LOAA has been limited to singleplex, small-scale laboratory-based studies using cultured cells (8) and already confirmed samples for analysis (10). In a previous study, LOAA perfectly matched rRT-PCR (Allplex 2019-nCoV Kit, Seegene Inc, Korea) for singleplex detection of COVID-19 (10). LOAA was reported to be 10-fold more sensitive than rRT-PCR; however, the results of this study may not be generalizable due to sampling bias, small sample size, and poor study design adopted for the study (10).

The current study is the first to evaluate the performance of LOAA using prospective clinical samples collected from the field in Ghana. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), area under curve (AUC), kappa, and accuracy of LOAA for differential diagnosis of SARS-CoV-2, RSV, Flu A, and Flu B using rRT-PCR as the gold standard were assessed.

MATERIALS AND METHODS

Study area

The study was conducted at seven health facilities located within the Ashanti and Savannah Regions of Ghana. In the Ashanti Region, two health care facilities—KNUST Hospital and HopeXchange Medical Center (Fig. 1) and in Savannah Region, five health care facilities—Bole, Sawla-Tuna-Kalba, West Gonja, Central Gonja, and East Gonja District Hospitals (Fig. 2) were selected for the study.

Fig 1.

The map depicts the Greater Kumasi area road network with towns and two hospitals. Inset maps localize Ashanti Region within Ghana and Ghana within Africa. Major highways and minor roads connect facilities across western and eastern zones.

Map of the Ashanti Region showing the study sites. The map was produced using ESRI ArcGIS 10.8, with base layers sourced from https://www.openstreetmap.org. Regional boundary shapefiles were obtained from https://gadm.org/download_country_v4.html and used in accordance with their licensing terms.

Fig 2.

The map depicts the Savannah Region with district hospitals in Sawla Tuna Kalba, Bole, West Gonja, Central Gonja, and East Gonja. Insets locate the region within Ghana and Africa.

Map showing the location of the study sites in the Savannah Region of Ghana. The map was produced using ESRI ArcGIS 10.8, with base layers sourced from https://www.openstreetmap.org. Regional boundary shapefiles were obtained from https://gadm.org/download_country_v4.html and used in accordance with their licensing terms.

These two regions were purposively selected for the study based on anecdotal knowledge of respiratory viral diseases in the health facilities located in these regions. Located in the southern part of Ghana, the Ashanti Region doubles as the second largest region in terms of population size and the region with the second highest number of confirmed COVID-19 cases (13). About 61% of its inhabitants live in urban areas (14). The Savannah Region occupies the northern part of Ghana and is the largest region in terms of land mass. Notwithstanding, the region is the second least populated with about 70% of its inhabitants living in rural areas (14). As of 7 April 2024, the Savannah Region had the least number of COVID-19 cases in Ghana (13).

Study design and study population

This was a cross-sectional hospital-based study conducted between August 2022 and January 2023. The inclusion criterion for enrollment of participants in the study was a clinical presentation with a suspected case of respiratory tract infection as defined by two or more of the following symptoms: fever, cough, headache, loss of smell, shortness of breath, nausea, vomiting, general weakness, loss of appetite, muscle pain, and diarrhea. Participants of any age or sex were eligible for enrollment in the study. The aim of the study was explained to all participants in their local language, and their consent was obtained before enrollment. Written informed consent was obtained from participants who were 18 years of age and above, while parental/guardian consent and child assent were obtained for those below 18 years of age. Study participants who could not provide informed consent were excluded from the study.

Sample size determination

The sample size was pre-calculated as the minimum number of samples required to attain 95% sensitivity and specificity assuming a 5% margin of error at a 95% confidence level (α  =  0.05, power 1 – β  =  0.20) using previously described formulas by Hajian-Tilaki (15). To calculate the final sample size, a previous prevalence of 43.2% was used (16). After substituting in the variables, a minimum sample size of 299 was required for the study.

Collection of clinical specimens

Oropharyngeal specimens were collected from the participants by trained laboratory personnel using flocked sterile swabs. Swabs were taken at the posterior oropharynx area without touching the tongue, as previously described (17). Afterward, the swabs were kept in well-labeled tubes and transported in 1 mL viral medium on ice to the Genomic and Infectious Disease Laboratory at Kwame Nkrumah University of Science and Technology, Kumasi, Ghana, for further testing.

Laboratory procedures

Experimental controls

Prior to testing clinical specimens, experimental controls were obtained from Kumasi Centre for Collaborative Research in Tropical Medicine, Kumasi, Ghana, to validate assay protocols for the detection of respiratory viruses by LOAA and rRT-PCR. The controls included archived SARS-CoV-2, RSV, Flu A (H1N1, H3N2), and Flu B samples confirmed positive by rRT-PCR.

Extraction of viral RNA

Nucleic acid extraction was performed at ambient temperature for the control and clinical samples in a class II biosafety cabinet. Viral RNA was extracted using QIAamp Viral RNA Mini Kit (Qiagen Diagnostics GmbH, Germany) following the manufacturer’s instructions. Viral RNA was eluted into 60 µL buffer AVE and stored at −20°C prior to molecular diagnosis.

Real-time RT-PCR

Real-time RT-PCR amplification was performed on the CFX96 Touch Real-Time PCR Detection System (Bio-Rad Inc, USA) for the detection of four respiratory viruses (Flu A, Flu B, SARS-CoV-2, and RSV). The multiplex FluoroType SARS-CoV-2/Flu/RSV Ver 1.0 Kit (Hain Lifescience GmbH, Germany) was used to qualitatively detect respiratory viruses following target specific protocols and primers according to the manufacturer’s instructions. The kit targets the RNA-dependent RNA polymerase (RdRP) and nucleocapsid (N) genes for SARS-CoV-2, the matrix protein 1 (M1) sequence for influenza A, the hemagglutinin sequence for influenza B, and the nucleoprotein (N) sequence for RSV. Real-time RT-PCR was carried out in a 20 µL reaction volume consisting of 8 µL master mix and 12 µL of RNA template. Respective volumes of reagents were adjusted based on final concentration for each primer and probe as specified in the manufacturer’s instruction. The cycling conditions consisted of initial activation at 50°C for 5 min, final activation at 95°C for 20 seconds, denaturation at 95°C for 3 seconds, and annealing/extension at 60°C for 26 seconds. After 50 cycles, the reactions were incubated at 4°C for 1 min. Samples with Ct values of 40 and below were interpreted as positive, and those greater than 40 or not detected were interpreted as negative according to a modified manufacturer’s instructions.

LOAA

The LOAA dPCR has a compact size, measuring 24 cm in length, 20 cm in width, and 25 cm in height. The Genoplexor COVID-19/Flu/RSV Detection Kit (Optolane Technologies Inc, South Korea) with a dynamic range of 5–7 log was used for real-time qualitative detection of SARS-CoV-2 (RdRp/E gene), Flu A (M gene), Flu B (NP gene), and RSV (M gene) RNA using LOAA Analyzer (model: On-Point). A 60 µL reaction mixture consisting of 30 µL of extracted RNA and 30 µL Onestep dRT-PCR (2×) mixture was prepared and loaded onto a chip provided by the manufacturer. The chip was placed in a cartridge and then carefully slotted into the LOAA Analyzer for real-time detection (Fig. 3). The cycling conditions included 52°C of reverse transcription, then denaturation at 97°C, followed by binding of the probe to the specific sequence at 60°C. The reaction mixture was allowed to complete 45 thermal cycles. Samples with Ct values ≤38 were interpreted as positive, while those with Ct values above 38 or not detected were determined to be negative as specified by the manufacturer.

Fig 3.

Diagram depicts LOAA instrument with height 25 cm, depth 24 cm, and width 20 cm beside LOAA Postman. Two circular PCR cartridges, Dr. PCR Cartridge and Dr. PCR R9 Cartridge, are placed in front.

LOAA Analyzer.

Quality control

To ensure reliable diagnosis, internal controls provided by the manufacturers of the kits used in the study, together with the experimental controls obtained, were tested on both LOAA and rRT-PCR prior to testing actual clinical samples. Internal controls were included alongside every batch of clinical samples tested by LOAA and rRT-PCR to validate the results. Biosafety techniques and standard operating procedures were strictly adhered to during processing. Testing was carried out by trained laboratory professionals in a clean working environment.

Data analysis

Data were entered into Microsoft Excel 2019 (Microsoft Inc, USA) and exported into STATA 14 (College Station, TX, USA) and GraphPad Prism 9.0 (San Diego, CA, USA) for analysis. Incomplete, missing, and duplicate data sets were excluded from the final analysis (Fig. 4). Age and Ct value data were presented as median and interquartile range (IQR), whereas categorical data were presented as proportions. The “diagtest” syntax in STATA was used to calculate sensitivity, specificity, PPV, and NPV. Student’s t-test was calculated to compare measurements by LOAA and rRT-PCR. P-values <0.05 were considered statistically significant. The AUC for the diagnostic performance of LOAA was obtained from ROC curves, assuming rRT-PCR as a reference point. The kappa value of inter-rater agreement was used to assess the strength of agreement between LOAA and rRT-PCR. Kappa values were interpreted following McHugh’s guidelines; values <0.20 = poor agreement, 0.21–0.40 = fair, 0.41–0.60 = moderate, 0.61–0.80 = good/substantial agreement, and 0.81–1.00 = very good/perfect agreement (18). Spearman’s correlation coefficient, r, was used to express the strength of agreement between measurements by LOAA and rRT-PCR. Correlation coefficient, r, was interpreted as a “perfect,” “very strong,” “moderate,” “fair,” “poor,” or “no” correlation when r = 1.00, 0.80–0.90, 0.60–0.70, 0.30–0.50, 0.10–0.20, or 0.00, respectively (19).

Fig 4.

Flowchart depicts analysis of 356 participants tested by LOAA and rRT-PCR. LOAA detects 22 SARS-CoV-2, 44 RSV, 13 FLU A, and 27 FLU B. rRT-PCR detects 23 SARS-CoV-2, 49 RSV, 9 FLU A, and 29 FLU B. Seven records were excluded due to missing or duplication.

Flowchart describing participant recruitment and tests performed.

RESULTS

Socio-demographic characteristics of the study population

A total of 356 patients were recruited for the study—298 (83.70%) from the Ashanti region and 58 (16.30%) from the Savannah region. All the viruses investigated in the study were detected in each region, with 67.79% (n = 202) and 24.14% (n = 14) positivity rates in Ashanti and Savannah, respectively. The age of participants ranged from 2 weeks to 83 years, with a median age of 19 years (IQR = 2 weeks–30 years). Detailed socio-demographic characteristics of the study population have been summarized in Table 1.

TABLE 1.

Characteristics of the study population

Characteristic Ashanti (n = 298)
No. (%)
Savannah (n = 58)
No. (%)
Total (n = 356)
No. (%)
Gender
 Male 121 (40.60) 20 (34.48) 141 (39.61)
 Female 177 (59.40) 38 (65.52) 215 (60.39)
Age group (years)
 ≤10 124 (41.61) 0 (0.00) 124 (34.83)
 11–20 42 (14.09) 11 (18.97) 53 (14.88)
 21–30 73 (24.50) 21 (36.21) 94 (26.40)
 31–40 30 (10.06) 7 (12.07) 37 (10.39)
 41–50 15 (5.03) 8 (13.79) 23 (6.46)
 51–60 5 (1.67) 6 (10.34) 11 (3.09)
 61–70 6 (2.01) 3 (5.17) 9 (2.53)
 >70 3 (1.00) 2 (3.44) 5 (1.40)
Education
 No formal education 53 (17.79) 8 (13.79) 61 (17.13)
 Primary 91 (30.54) 27 (46.55) 118 (33.15)
 Secondary 41 (13.76) 12 (20.69) 53 (14.89)
 Tertiary 91 (30.54) 0 (0.00) 91 (25.56)
aNo response 22 (7.38) 11 (18.97) 33 (9.27)
a

No response: This individuals did not provided information on their educational status.

Experimental controls

A total of 12 experimental controls were obtained for testing all the viruses under study by LOAA and rRT-PCR prior to testing of clinical samples. We used eight positive controls (two per virus) and four negative controls. However, for all experimental controls tested, LOAA qualitatively matched rRT-PCR for the detection of the respiratory viruses (data not shown).

Prevalence of respiratory viruses by LOAA and rRT-PCR

Of 356 clinical specimens tested, rRT-PCR detected 110 (30.90%) respiratory viruses while LOAA detected 106 (29.78%). For both rRT-PCR and LOAA, RSV was the predominant respiratory virus, followed by Flu B, SARS-CoV-2, and Flu A. The difference in the detection of respiratory viruses by LOAA and rRT-PCR was not statistically significant (Table 2).

TABLE 2.

Detection of respiratory viruses by LOAA and rRT-PCRa

Respiratory virus rRT-PCR
(n = 356)
LOAA
(n = 356)
P-value
SARS-CoV-2, n (%) 23 (6.46) 22 (6.18) >0.99
RSV, n (%) 49 (13.76) 44 (12.36) 0.66
Flu A, n (%) 9 (2.53) 13 (3.65) 0.52
Flu B, n (%) 29 (8.15) 27 (7.58) 0.89
Total, n (%) 110 (30.90) 106 (29.78) 0.81
a

Fischer’s exact test was used to compute the P-value.

Clinical performance of LOAA vs rRT-PCR

Using rRT-PCR as a reference, the sensitivity (Sn) of LOAA to detect respiratory viruses ranged between 86.21% (Flu B) and 91.30% (SARS-CoV-2). LOAA showed high specificity (Sp) (>98.00%) for all respiratory viruses tested in the study. LOAA was most sensitive in detecting SARS-CoV-2 (Sn: 91.30%, Sp: 99.70%), while the lowest sensitivity was recorded for Flu B (Sn: 86.21%, Sp: 99.38%). Sensitivity for RSV (Sn: 87.76%, Sp: 99.67%) and Flu A (Sn: 88.89%, Sp: 98.54%) was slightly higher, though specificity remained consistently high across all viruses. For all respiratory viruses, LOAA showed “very good/almost perfect agreement” (κ = 0.88–0.93) with rRT-PCR except for Flu A (κ = 0.72), which showed “good/substantial agreement” (Table 3). Overall, LOAA produced 9 false positives and 13 false negatives as compared to the gold standard rRT-PCR assay for all respiratory viruses tested.

TABLE 3.

Clinical performance of LOAA vs rRT-PCR

Respiratory virus LOAA rRT-PCR Sensitivity
(95% CI)
Specificity
(95% CI)
PPV
(95% CI)
NPV
(95% CI)
AUC Accuracy
%
Kappa
Result Positive Negative Total
SARS-CoV-2 Positive 21 1 22 91.30 (73.20–98.45) 99.70 (98.32–99.98) 95.45 (78.20–99.77) 99.40 (97.84–99.89) 0.75 99.16 0.93
Negative 2 332 334
Total 23 333 356
RSV Positive 43 1 44 87.76 (75.76–94.27) 99.67 (98.18–99.98) 97.73 (88.19–99.88) 98.08 (95.87–99.12) 0.59 98.03 0.91
Negative 6 306 312
Total 49 307 356
Flu A Positive 8 5 12 88.89 (56.50–99.43) 98.54 (96.62–99.37) 61.54 (35.52–82.29) 99.70 (98.34–99.98) 0.78 98.31 0.72
Negative 1 342 343
Total 9 347 356
Flu B Positive 25 2 27 86.21 (69.44–94.50) 99.38 (97.76–99.89) 92.59 (76.63–98.63) 98.77 (96.87–99.52) 0.70 98.31 0.88
Negative 4 325 329
Total 29 327 356

Ct values of respiratory viruses by LOAA and rRT-PCR

The Ct values of respiratory viruses detected by rRT-PCR and LOAA ranged from 17.56 to 39.97 and 14.12 to 34.14, respectively. The median Ct value, 29.79 (IQR: 25.61, 32.89), detected by rRT-PCR was significantly higher (P < 0.0001) than that of LOAA, 27.25 (IQR: 24.45, 29.62) (Fig. 5).

Fig 5.

Dot plot compares Ct values of SARS-CoV-2, RSV, FLU A, and FLU B detected by LOAA and rRT-PCR. LOAA values for SARS-CoV-2, RSV, and FLU B are significantly lower. Include LOAA cutoff value, and no significant difference was found for FLU A.

Comparing Ct values of individual respiratory viruses detected and assay sensitivity of rRT-PCR and LOAA. The red dotted line indicates the Ct cutoff value of LOAA. Pearson’s correlation was used to compute the P-value.

Correlation of LOAA and rRT-PCR Ct values for detection of respiratory viruses

Spearman’s correlation coefficient, r, was calculated to determine the strength of the agreement between Ct values of LOAA and rRT-PCR. All individual respiratory viruses showed significant differences in Ct values for LOAA and rRT-PCR except for Flu A, which had comparable Ct values for LOAA and rRT-PCR (Fig. 5). For the detection of respiratory viruses, a “very strong” positive correlation, rs = 0.86 (95% CI: 0.79, 0.90), was observed between the combined Ct values by LOAA and rRT-PCR as shown in Fig. 6. For the detection of respiratory viruses by either method only, an unpaired t-test was used to compare the mean Ct values, which was found to be significant (Fig. 7).

Fig 6.

Scatterplot depicts LOAA versus rRT-PCR values with positive correlation. Most values cluster between 20 and 35. Trend line slopes upward with correlation coefficient of 0.86.

Correlation of Ct values detected by LOAA and rRT-PCR.

Fig 7.

Dot plot depicts Ct values for LOAA-only and rRT-PCR-only groups with statistical comparison. Median and spread are higher in the rRT-PCR-only group. The p-value is less than 0.01.

Ct of infections detected by rRT-PCR and LOAA only.

Comparison of respiratory viral coinfections detected by LOAA and rRT-PCR

Coinfections were identified in the study, with rRT-PCR detecting four cases and LOAA identifying three cases. Both methods detected one case of SARS-CoV-2/RSV coinfection. LOAA uniquely detected SARS-CoV-2/Flu B and Flu A/Flu B, which were missed by rRT-PCR, while rRT-PCR identified three cases of RSV/Flu B that were not detected by LOAA (Table 4).

TABLE 4.

Respiratory viral coinfection detected by LOAA and rRT-PCR

Coinfection rRT-PCR,
No.
LOAA,
No.
SARS-CoV-2/RSV 1 1
SARS-CoV-2/Flu B 0 1
Flu A/Flu B 0 1
RSV/Flu B 3 0
Total 4 3

Discordant results

A total of 22 discordant test results were recorded in the study. Out of this number, 13/22 (59.09%) infections were exclusively detected by rRT-PCR, while 9/22 (40.90%) were detected by LOAA only.

DISCUSSION

Prompt and accurate detection of infectious pathogens, particularly epidemic/pandemic-prone viruses, is critical for successful disease management and control. During the height of the COVID-19 pandemic, manufacturers of biomedical diagnostic devices were faced with the task of developing highly sensitive and rapid tests to support testing and tracing of suspected infected individuals. Gold-standard rRT-PCR, despite its sensitivity, is time-consuming and requires skilled labor (20, 21), whereas most rapid diagnostic tests generally lacked the needed sensitivity and specificity for active case detection (20, 22). It was during this period that Optolane Technologies Inc (Seongnam, South Korea) launched LOAA, a third-generation PCR and a novel compact-sized digital PCR equipment capable of diagnostic applications.

LOAA is a separation-type dPCR instrument that makes use of semiconductor chip-based microelectromechanical system technology to significantly reduce the size of the equipment. LOAA is advantageous over other currently commercialized platforms owing to its compact size, rapid turnaround time, precise measurement, and potential for improved clinical utility (9, 23). Early versions of LOAA were designed to solely detect SARS-CoV-2 (Dr. PCR 20K COVID-19 Detection Kit) and have been reported to show absolute sensitivity and perfect diagnostic agreement with commercially available rRT-PCR-based tests performed on Bio-Rad’s CFX96 (10) and QX200 droplet dPCR platforms (8). Recently, LOAA has been equipped for multiplex detection of common respiratory viruses by the aid of the Genoplexor COVID-19/RSV/Flu Detection Kit (Optolane Technologies Inc, South Korea). In this study, we assessed the clinical performance of LOAA for multiplex detection of SARS-CoV-2, Flu A, Flu B, and RSV using a well-established rRT-PCR assay as the reference standard.

Compared to rRT-PCR, LOAA showed sensitivities of 91.30%, 87.76%, 88.89%, and 86.21% for the detection of SARS-CoV-2, RSV, Flu A, and Flu B, respectively. The sensitivity of LOAA to detect SARS-CoV-2 (91.30%) in the present study was lower than in previous studies in Korea (8, 10, 24). This observation may be explained by three main factors: (i) the current study reports clinical sensitivity based on field evaluations, whereas those cited were laboratory-based studies using already confirmed archived samples, (ii) sensitivity in the current study was estimated from multiplex reaction, while the previous studies were obtained from singleplex assays, and (iii) the present study focuses on oropharyngeal specimens, while the earlier studies included a variety of specimen types in their analyses. The variability in field and laboratory conditions, differences in study design, sample size, and specimen type among the study participants in the current and previous studies may have accounted for the differences in LOAA sensitivity to detect SARS-CoV-2 (25, 26). There are conflicting reports regarding the sensitivity of multiplex and singleplex assays to detect respiratory viruses by nucleic acid amplification tests. While several studies (2730) showed comparable detection by both assays, other authors (3134) reported slightly higher sensitivity of singleplex assays than multiplex assays, particularly at the detection of low levels of viremia. The relatively low sensitivity of multiplex assays has been attributed to the competition among different targets for primers and other reagents, which may lead to low amplification efficiency (35, 36). Notwithstanding, multiplex assays are ideally more preferred in the laboratory due to their ability to simultaneously detect multiple infections, thus reducing turnaround time, preserving staff resources, and conserving expensive reagents (30). Despite the difference in sensitivity, the specificity and kappa of LOAA to detect SARS-CoV-2 were similar to previous studies (8, 10, 24).

For both tests, RSV was the predominant respiratory virus detected in the study. This finding was not particularly surprising since the majority of the study participants were aged 10 years and below (Table 1). It has been established that RSV is a common cause of acute respiratory infection among children under 5 years of age (37). Few multiplex dPCR assays have been developed to detect RSV and other respiratory viruses, with very limited data about their comparative field evaluations (38, 39). In this study, LOAA recorded the third highest sensitivity for the detection of RSV (87.76%). This sensitivity was higher than those reported by other digital PCR systems (nCounter; RSV-A 74.30%, RSV-B 77.60%) evaluated in Brazil (40). In addition, LOAA reported marginally higher specificity for RSV compared to RSV-A and RSV-B in the Brazilian study (99.67% vs 98.40%, 99.67% vs 97.80%) and accuracy (98.03% vs 95.20%, 98.03% vs 95.30%), respectively. In contrast to the few multiplex dPCR assays for respiratory viruses, there are several multiplex RT-PCR assays developed to target about 2–18 individual respiratory viruses (30, 4144). Compared to multiplex RT-PCR assays, LOAA sensitivity to detect RSV was lower than reported in earlier studies (33, 41, 45), which reported absolute sensitivity or positive percent agreement with varying reference standards. This may be due to the low Ct threshold of LOAA, hence, its inability to pick up low-density infections. Again, this may be attributed to different primer designs, differences in viral load, optimization, or efficiency of extraction/amplification process between both tests.

From the study, LOAA recorded low sensitivity for the detection of Flu B (86.21%) and Flu A (88.89%). This sensitivity was higher than that of Qiagen ResPlex II, a multiplex RT-PCR assay for the detection of 17 respiratory viruses in a study held in China (32), but was lower than reported in earlier studies (33, 45). The difference in sensitivities may be attributed to differences in patient populations and, most importantly, variability in the reference standards employed in each study (46). Studies by Mark GC et al. revealed a higher sensitivity of ResPlex II when compared to viral culture as a standard but lower sensitivity when compared to singleplex PCR. Despite the relatively low sensitivity, LOAA showed high specificity (98.54%–99.38%) and overall accuracy (98.31%) for the detection of Flu A and B. Together with a PPV/NPV of 92.59%/98.77% and kappa value indicative of “almost perfect agreement” with rRT-PCR, the LOAA remains a viable choice to detect Flu B infections. For Flu A, however, the low PPV (61.54%) but high NPV (99.70%) and kappa value indicative of “good/substantial agreement” suggest LOAA is best used to rule out Flu A infections rather than detecting them.

From the study, LOAA reported a lower percentage of false positive (n = 9/22; 40.90%) and a higher rate of false negative (n = 13/22; 59.09%) results when compared to rRT-PCR. False positive dPCR results have been attributed to poor assay design and off-target amplification stemming from cross-contamination during experimental setup (47). False negatives, on the other hand, may be attributed to the relatively low detection limit of LOAA, hence, its inability to detect low-density infections. False negative dPCR results at low viral load have been reported in other studies (48, 49).

Overall, our results suggest LOAA is a sensitive and highly accurate platform when compared to the gold standard, rRT-PCR, for the detection of SARS-CoV-2, RSV, and Flu B, with minor improvements needed for Flu A. The ease of operation and portable compact size of LOAA validate the novel platform as an efficient tool for swift diagnostic outcomes, catering to the need for timely clinical decision-making. The LOAA device necessitates a nucleic acid extraction step before testing. Resolving this requirement in the future will enhance the platform’s efficiency, potentially resulting in a significantly improved functionality as a point-of-care device or bedside test device. In the current study, only a few cases of viral coinfection (LOAA = 3/356, rRT-PCR = 4/356) were reported (Table 4). This was lower than previous studies in China (39) where 6/98 viral coinfections were detected by a multiplex digital droplet PCR assay and in Ecuador (50) where 19/158 viral coinfections were detected by multiplex RT-qPCR assays. It is plausible that the variation in the study population, period of the study, and geographical area may have accounted for differences in coinfections in both studies. Our results showed that LOAA perfectly matched rRT-PCR to detect SARS-CoV-2/RSV. It outperformed rRT-PCR in detecting SARS-CoV-2/Flu B and Flu A/Flu B coinfections but failed to detect all three RSV/Flu B coinfections detected by rRT-PCR. Viral coinfections have been implicated in altered viral pathogenicity, disrupted host response, and complicated diagnosis (51). Studies have shown an increase in false negative test results due to masking of target viral RNA by undesired pathogens in coinfections (52). Further studies are recommended to investigate the efficacy of LOAA to detect coinfections—both viral and bacterial-viral—and how it influences its accuracy.

Conclusion

The emergence of novel highly sensitive digital PCR tests represents a significant breakthrough in overcoming the challenges associated with conventional PCR testing. Our results validate the Genoplexor COVID-19/RSV/Flu Detection Kit performed on LOAA as a highly sensitive tool for the detection of SARS-CoV-2, RSV, and influenza type B (Flu B) with minor improvements required for type A (Flu A). With the omission of the nucleic acid extraction step, LOAA will be more efficient for the detection of SARS-CoV-2 and other respiratory viruses. It goes without saying that the sustainability of this method and its practicality for surveillance testing further outlines its potential for widespread use in monitoring and controlling respiratory virus outbreaks.

Limitations of the study

One limitation of the study is the potential impact of using oropharyngeal swabs in our study which could have implications on the results. This difference in sampling methods could potentially lead to an overestimation of prevalence, consequently affecting the accuracy of the sample size calculation.

In addition, while LOAA qualitatively matched rRT-PCR for all experimental controls, the small sample size may not provide sufficient assurance for pathogen detection across diverse viral lineages. This limitation constrains the robustness of the validation, and further testing with a broader range of controls is necessary to ensure broader applicability of the results.

Moreover, though the current requirement for standard nucleic acid extraction of LOAA may limit immediate point-of-care use, it ensures high sensitivity and specificity; ongoing efforts to integrate simplified extraction methods are expected to enhance the kit’s accessibility and utility in diverse settings.

Again, the absence of sequencing or other molecular methods to resolve discrepancies between LOAA and rRT-PCR results limits definitive interpretation of diagnostic accuracy.

Finally, the two assays, LOAA and rRT-PCR, use different viral targets, which impact the limits of detection and Ct values. These differences may introduce challenges in making meaningful comparisons between the assays, as they may reflect assay-specific sensitivities rather than true performance equivalency. Future work should aim to harmonize or adjust for these assay variations to improve the validity of cross-comparison.

ACKNOWLEDGMENTS

This study was funded by the U.S. Center for Disease Control and the Korea International Cooperation Agency under award number NU2HGH000041-02-00. The findings and conclusions in this report are those of the author(s) and do not necessarily represent the official position of the U.S. Centers for Disease Control and Prevention. The use of trade names and commercial sources is for identification purposes only and does not imply endorsement by the U.S. Department of Health and Human Services.

M.O., B.N., DTB., and G.A. contributed to the conceptualization and methodology of the study. G.A., S.O.A., E.K.A., R.L., R.O.A., C.K., F.S., N.K.A-.B., E.D., J.A-.N., A.G.O-.A., F.A., P.K.B., and V.B. were involved in data collection. Data analysis and visualization were led by M.O., G.A., S.O.A., and E.K.A. Supervision was provided by M.O., B.N., P.A., and D.T.B. The manuscript was drafted and revised primarily by M.O., B.N., DTB., and G.A., with input and critical revisions from all authors. All authors reviewed and approved the final manuscript. The findings and conclusions in this report are those of the author(s) and do not necessarily represent the official position of the U.S. Centers for Disease Control and Prevention. The use of trade names and commercial sources is for identification purposes only and does not imply endorsement by the U.S. Department of Health and Human Services.

The authors are grateful to all staff of the various health facilities for their support and assistance during data collection. We also extend our appreciation to the study participants for their time and willingness to contribute to this research.

The findings and conclusions in this report are those of the author(s) and do not necessarily represent the official position of the U.S. Centers for Disease Control and Prevention.

The use of trade names and commercial sources is for identification purposes only and does not imply endorsement by the U.S. Department of Health and Human Services.

Contributor Information

Michael Owusu, Email: michaelowusu80@gmail.com.

Rosemary C. She, City of Hope Department of Pathology, Duarte, California, USA

ETHICAL APPROVAL

The objectives and procedures of the study were reviewed and approved by the Ethics Review Board of the Ghana Health Service (Reference number: GHS-ERC:007/05/22) and U.S. Centers for Disease Control and Prevention (Accession number: CGH-GOSB-5/24/22-368df). All participants provided written informed consent before their involvement in the study. Patients were de-identified using unique codes to ensure confidentiality. The link to the ethics documents is available online (10.6084/m9.figshare.29606108)

DATA AVAILABILITY

The datasetsdata sets supporting the conclusions of this article are included within the article and available online (10.6084/m9.figshare.29606108).

REFERENCES

  • 1. Korsman S, Zyl G, Nutt L, Andersson M, Preiser W. 2012. Respiratory Viruses Virology:108–109. doi: 10.1016/B978-0-443-07367-0.00047-1 [DOI] [Google Scholar]
  • 2. Leung NHL. 2021. Transmissibility and transmission of respiratory viruses. Nat Rev Microbiol 19:528–545. doi: 10.1038/s41579-021-00535-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. World Health Organization . 2023. WHO (Coronavirus) COVID-19 Dashboard. Available from: https://covid19.who.int
  • 4. Zhang N, Wang L, Deng X, Liang R, Su M, He C, Hu L, Su Y, Ren J, Yu F, Du L, Jiang S. 2020. Recent advances in the detection of respiratory virus infection in humans. J Med Virol 92:408–417. doi: 10.1002/jmv.25674 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Burrell C, Howard C, Murphy F. 2017. Laboratory diagnosis of virus diseases. Fenner and White’s Medical Virology 2017:135–154. doi: 10.1016/B978-0-12-375156-0.00010-2 [DOI] [Google Scholar]
  • 6. 2021. Recommendations for national SARS-CoV-2 testing strategies and diagnostic capacities: interim guidance . World Health Organization. [Google Scholar]
  • 7. Bustin SA, Nolan T. 2020. RT-qPCR Testing of SARS-CoV-2: A Primer. Int J Mol Sci 21:3004. doi: 10.3390/ijms21083004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Lee S-S, Park J-H, Bae Y-K. 2021. Comparison of two digital PCR methods for EGFR DNA and SARS-CoV-2 RNA quantification. Clinica Chimica Acta 521:9–18. doi: 10.1016/j.cca.2021.06.016 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Lee H, Lee C-J, Kim DH, Cho C-S, Shin W, Han K. 2021. High-accuracy quantitative principle of a new compact digital PCR equipment: Lab On An Array. Genomics Inform 19:e34. doi: 10.5808/gi.21035 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Lee C-J, Shin W, Mun S, Yu M, Choi Y-B, Kim DH, Han K. 2021. Diagnostic evaluation of qRT-PCR-based kit and dPCR-based kit for COVID-19. Genes Genom 43:1277–1288. doi: 10.1007/s13258-021-01162-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Nyaruaba R, Mwaliko C, Dobnik D, Neužil P, Amoth P, Mwau M, Yu J, Yang H, Wei H. 2022. Digital PCR Applications in the SARS-CoV-2/COVID-19 Era: a Roadmap for Future Outbreaks. Clin Microbiol Rev 35:e0016821. doi: 10.1128/cmr.00168-21 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Dong L, Zhou J, Niu C, Wang Q, Pan Y, Sheng S, Wang X, Zhang Y, Yang J, Liu M, Zhao Y, Zhang X, Zhu T, Peng T, Xie J, Gao Y, Wang D, Dai X, Fang X. 2021. Highly accurate and sensitive diagnostic detection of SARS-CoV-2 by digital PCR. Talanta 224:121726. doi: 10.1016/j.talanta.2020.121726 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Ghana Health Service . 2023. Ghana Health Service. COVID-19 Situation Dashboard. Available from: https://www.ghs.gov.gh/covid19/dashboardm.php
  • 14. Ghana Statistical Service Population of Regions and District. 2021
  • 15. Hajian-Tilaki K. 2014. Sample size estimation in diagnostic test studies of biomedical informatics. J Biomed Inform 48:193–204. doi: 10.1016/j.jbi.2014.02.013 [DOI] [PubMed] [Google Scholar]
  • 16. Annan A, Ebach F, Corman VM, Krumkamp R, Adu-Sarkodie Y, Eis-Hübinger AM, Kruppa T, Simon A, May J, Evans J, Panning M, Drosten C, Drexler JF. 2016. Similar virus spectra and seasonality in paediatric patients with acute respiratory disease, Ghana and Germany. Clin Microbiol Infect 22:340–346. doi: 10.1016/j.cmi.2015.11.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Akowuah E, Acheampong G, Ayisi-Boateng NK, Amaniampong A, Agyapong FO, Senyo Kamasah J, Agyei G, Owusu DO, Nkrumah B, Mutocheluh M, Sylverken AA, Owusu M. 2022. Comparable Detection of SARS-CoV-2 in sputum and oropharyngeal swab samples of suspected COVID-19 Patients. COVID 2:858–866. doi: 10.3390/covid2070062 [DOI] [Google Scholar]
  • 18. McHugh ML. 2012. Interrater reliability: the kappa statistic. Biochem Med (Zagreb) 22:276–282. https://hrcak.srce.hr/89395 [PMC free article] [PubMed] [Google Scholar]
  • 19. Chan YH. 2003. Biostatistics 104: correlational analysis. Singapore Med J 44:614–619. [PubMed] [Google Scholar]
  • 20. Gupta N, Augustine S, Narayan T, O’Riordan A, Das A, Kumar D, Luong JHT, Malhotra BD. 2021. Point-of-Care PCR Assays for COVID-19 Detection. Biosensors (Basel) 11:141. doi: 10.3390/bios11050141 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Wang C, Liu M, Wang Z, Li S, Deng Y, He N. 2021. Point-of-care diagnostics for infectious diseases: From methods to devices. Nano Today 37:101092. doi: 10.1016/j.nantod.2021.101092 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Ranadive N, Kunene S, Darteh S, Ntshalintshali N, Nhlabathi N, Dlamini N, Chitundu S, Saini M, Murphy M, Soble A, Schwartz A, Greenhouse B, Hsiang MS. 2017. Limitations of rapid diagnostic testing in patients with suspected malaria: a diagnostic accuracy evaluation from swaziland, a low-endemicity country aiming for malaria elimination. Clin Infect Dis 64:1221–1227. doi: 10.1093/cid/cix131 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Khan MS, Tariq MO, Nawaz M, Ahmed J. 2021. MEMS sensors for diagnostics and treatment in the fight against COVID-19 and other pandemics. IEEE Access 9:61123–61149. doi: 10.1109/ACCESS.2021.3073958 [DOI] [Google Scholar]
  • 24. Shin W, Lee CJ, Lee YM, Choi YB, Mun S, Han K. 2022. Rapid identification of SARS-CoV-2 in the point-of-care using digital PCR-based Dr. PCR Di20K COVID-19 Detection Kit without viral RNA extraction. Genes Genomics 44:617–628. doi: 10.1007/s13258-022-01242-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Bwire GM, Majigo MV, Njiro BJ, Mawazo A. 2021. Detection profile of SARS-CoV-2 using RT-PCR in different types of clinical specimens: a systematic review and meta-analysis. J Med Virol 93:719–725. doi: 10.1002/jmv.26349 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Kaufman HW, Niles JK, Nash DB. 2021. Disparities in SARS-CoV-2 positivity rates: associations with race and ethnicity. Popul Health Manag 24:20–26. doi: 10.1089/pop.2020.0163 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Kim S, Eom K-W, Cho C-R, Um TH. 2016. Comparison of the Ability of Multiplex and Singleplex PCR to Detect Human Respiratory Viruses. Lab Med Online 6:240. doi: 10.3343/lmo.2016.6.4.240 [DOI] [Google Scholar]
  • 28. Laamiri N, Aouini R, Marnissi B, Ghram A, Hmila I. 2018. A multiplex real-time RT-PCR for simultaneous detection of four most common avian respiratory viruses. Virology (Auckl) 515:29–37. doi: 10.1016/j.virol.2017.11.021 [DOI] [PubMed] [Google Scholar]
  • 29. Sahajpal NS, Mondal AK, Ananth S, Njau A, Jones K, Ahluwalia P, Oza E, Ross TM, Kota V, Kothandaraman A, Fulzele S, Hegde M, Chaubey A, Rojiani AM, Kolhe R. 2022. Clinical validation of a multiplex PCR-based detection assay using saliva or nasopharyngeal samples for SARS-Cov-2, influenza A and B. Sci Rep 12:3480. doi: 10.1038/s41598-022-07152-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Shu B, Kirby MK, Davis WG, Warnes C, Liddell J, Liu J, Wu K-H, Hassell N, Benitez AJ, Wilson MM, Keller MW, Rambo-Martin BL, Camara Y, Winter J, Kondor RJ, Zhou B, Spies S, Rose LE, Winchell JM, Limbago BM, Wentworth DE, Barnes JR. 2021. Multiplex real-time reverse transcription PCR for Influenza A Virus, Influenza B virus, and severe acute respiratory syndrome coronavirus 2. Emerg Infect Dis 27:1821–1830. doi: 10.3201/eid2707.210462 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Parker J, Fowler N, Walmsley ML, Schmidt T, Scharrer J, Kowaleski J, Grimes T, Hoyos S, Chen J. 2015. Analytical sensitivity comparison between singleplex real-time PCR and a multiplex PCR platform for detecting respiratory viruses. PLoS One 10:e0143164. doi: 10.1371/journal.pone.0143164 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Deng J, Ma Z, Huang W, Li C, Wang H, Zheng Y, Zhou R, Tang Y-W. 2013. Respiratory virus multiplex RT-PCR assay sensitivities and influence factors in hospitalized children with lower respiratory tract infections. Virol Sin 28:97–102. doi: 10.1007/s12250-013-3312-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Bierbaum S, Forster J, Berner R, Rücker G, Rohde G, Neumann-Haefelin D, Panning M, CAPNETZ study group . 2014. Detection of respiratory viruses using a multiplex real-time PCR assay in Germany, 2009/10. Arch Virol 159:669–676. doi: 10.1007/s00705-013-1876-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Onwuchekwa C, Moreo LM, Menon S, Machado B, Curcio D, Kalina W, Atwell JE, Gessner BD, Siapka M, Agarwal N, Rubbrecht M, Nair H, Rozenbaum M, Aponte-Torres Z, Vroling H, Begier E. 2023. Underascertainment of respiratory syncytial virus infection in adults due to diagnostic testing limitations: a systematic literature review and meta-analysis. J Infect Dis 228:173–184. doi: 10.1093/infdis/jiad012 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Elnifro EM, Ashshi AM, Cooper RJ, Klapper PE. 2000. Multiplex PCR: optimization and application in diagnostic virology. Clin Microbiol Rev 13:559–570. doi: 10.1128/CMR.13.4.559 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36. Perkel J. 2012. Overcoming the challenges of multiplex. Available from: http://www.biocompare.com/Editorial-Articles/117895-Multiplex-PCR
  • 37. Kafintu-Kwashie AA, Nii-Trebi NI, Obodai E, Neizer M, Adiku TK, Odoom JK. 2022. Molecular epidemiological surveillance of viral agents of acute lower respiratory tract infections in children in Accra, Ghana. BMC Pediatr 22:364. doi: 10.1186/s12887-022-03419-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Hijano DR, Maron G, Hayden RT. 2018. Respiratory viral infections in patients with cancer or undergoing hematopoietic cell transplant. Front Microbiol 9:3097. doi: 10.3389/fmicb.2018.03097 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Leong NKC, Gu H, Ng DYM, Chang LDJ, Krishnan P, Cheng SSM, Peiris M, Poon LLM. 2023. Development of multiplex RT-ddPCR assays for detection of SARS-CoV-2 and other common respiratory virus infections. Influenza Other Respir Viruses 17:e13084. doi: 10.1111/irv.13084 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Bouzas ML, Oliveira JR, Queiroz A, Fukutani KF, Barral A, Rector A, Wollants E, Keyaerts E, Van der Gucht W, Van Ranst M, Beuselinck K, de Oliveira CI, Van Weyenbergh J, Nascimento-Carvalho CM. 2018. Diagnostic accuracy of digital RNA quantification versus real-time PCR for the detection of respiratory syncytial virus in nasopharyngeal aspirates from children with acute respiratory infection. J Clin Virol 106:34–40. doi: 10.1016/j.jcv.2018.07.003 [DOI] [PubMed] [Google Scholar]
  • 41. Kim TY, Bae GE, Kim J-Y, Kang M, Jang J-H, Huh HJ, Chung DR, Lee NY. 2022. Evaluation of the Kaira COVID-19/Flu/RSV Detection Kit for detection of SARS-CoV-2, influenza A/B, and respiratory syncytial virus: A comparative study with the PowerChek SARS-CoV-2, influenza A&B, RSV Multiplex Real-time PCR Kit. PLoS One 17:e0278530. doi: 10.1371/journal.pone.0278530 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Neopane P, Nypaver J, Shrestha R, Beqaj S. 2022. Performance Evaluation of TaqMan SARS-CoV-2, Flu A/B, RSV RT-PCR Multiplex Assay for the Detection of Respiratory Viruses. Infect Drug Resist 15:5411–5423. doi: 10.2147/IDR.S373748 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. Anderson TP, Werno AM, Barratt K, Mahagamasekera P, Murdoch DR, Jennings LC. 2013. Comparison of four multiplex PCR assays for the detection of viral pathogens in respiratory specimens. J Virol Methods 191:118–121. doi: 10.1016/j.jviromet.2013.04.005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Milucky J, Pondo T, Gregory CJ, Iuliano D, Chaves SS, McCracken J, Mansour A, Zhang Y, Aleem MA, Wolff B, et al. 2020. The epidemiology and estimated etiology of pathogens detected from the upper respiratory tract of adults with severe acute respiratory infections in multiple countries, 2014-2015. PLoS One 15:e0240309. doi: 10.1371/journal.pone.0240309 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Li W, Wang X, Cui W, Yuan L, Hu X. 2022. Clinical Evaluation of a Multiplex PCR Assay for Simultaneous Detection of 18 Respiratory Pathogens in Patients with Acute Respiratory Infections. Pathogens 12:21. doi: 10.3390/pathogens12010021 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Mak GC, Cheng PKC, Lim W. 2011. Evaluation of Qiagen Resplex II for the detection of pandemic influenza A (H1N1) 2009 and influenza A (H3N2) virus. J Clin Virol 51:88–89. doi: 10.1016/j.jcv.2011.02.006 [DOI] [PubMed] [Google Scholar]
  • 47. Quan PL, Sauzade M, Brouzes E. 2018. dPCR: A Technology Review. Sensors (Basel) 18:1271. doi: 10.3390/s18041271 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Nixon G, Garson JA, Grant P, Nastouli E, Foy CA, Huggett JF. 2014. Comparative study of sensitivity, linearity, and resistance to inhibition of digital and nondigital polymerase chain reaction and loop mediated isothermal amplification assays for quantification of human cytomegalovirus. Anal Chem 86:4387–4394. doi: 10.1021/ac500208w [DOI] [PubMed] [Google Scholar]
  • 49. Hayden RT, Gu Z, Ingersoll J, Abdul-Ali D, Shi L, Pounds S, Caliendo AM. 2013. Comparison of droplet digital PCR to real-time PCR for quantitative detection of cytomegalovirus. J Clin Microbiol 51:540–546. doi: 10.1128/JCM.02620-12 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50. Morales-Jadán D, Muslin C, Viteri-Dávila C, Coronel B, Castro-Rodríguez B, Vallejo-Janeta AP, Henríquez-Trujillo AR, Garcia-Bereguiain MA, Rivera-Olivero IA. 2023. Coinfection of SARS-CoV-2 with other respiratory pathogens in outpatients from Ecuador. Front Public Health 11:1264632. doi: 10.3389/fpubh.2023.1264632 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51. Du Y, Wang C, Zhang Y. 2022. Viral Coinfections. Viruses 14:2645. doi: 10.3390/v14122645 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52. Sabzevari AG, Nikbakht M. 2022. Considerations about Reducing False-Negative PCR Test for COVID-19. Int J MCH AIDS 11:e571. doi: 10.21106/ijma.571 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The datasetsdata sets supporting the conclusions of this article are included within the article and available online (10.6084/m9.figshare.29606108).


Articles from Microbiology Spectrum are provided here courtesy of American Society for Microbiology (ASM)