TABLE 1.

Details of parameters checked or reported by ImageCompliance V2.0.

Data DICOM Tag Reported for Reference value Tolerance Warning tier
Patient ID 0010,0020 CT & MR NA NA Report only
Patient name 0010,0010 CT & MR NA NA Report only
Study date 0008,0020 CT & MR NA NA Report only
Modality 0008,0060 CT & MR NA NA Report only
Body region 0018,0015 & 0008,1030 CT & MR NA NA Report only
Slice thickness 0018,0050 CT & MR

1.0 mm for stereotactic scans;

Protocol‐specific for non‐stereotactic scans, typically 2.0–2.5 mm

Fail
kVp 0018,0060 CT Protocol‐specific, typically 120 = Fail
Data collection diameter 0018,0090 CT Protocol‐specific, typically 500 mm Fail
Reconstruction diameter 0018,1100 CT Protocol‐specific, typically 700 mm Fail
Gantry tilt 0018,1120 CT 0 = Fail
Tube current 0018,1151 CT Scanner model‐specific, determined from historical data Within 95% confidence interval Physics review a
Filter type 0018,1160 CT Protocol‐specific = Fail
Focal spot 0018,1190 CT Protocol‐specific = Fail
Convolution Kernel 0018,1210 CT Protocol‐specific = Fail
Patient position 0018,5100 CT Protocol‐specific, typically Headfirst Supine = Warning
CTDIvol 0018,9345 CT Scanner model‐specific, determined from historical data Within 95% Confidence Interval Physics review
Exposure time 0018,1150 CT NA NA Report only
Protocol name 0018,1030 CT NA NA Report only
2D distortion correction 0008,0008 MR Yes = Fail
Manufacturer 0008,0070 MR Protocol‐specific = Fail
Model 0008,1090 MR Protocol‐specific = Fail
Serial number 0018,1000 MR Protocol‐specific = Fail
Series description 0008,103E MR Protocol‐specific = Fail b
3D distortion Correction 0008,0008 & 0008,103E MR Yes = Fail b or Physics Review c
Sequence weighting 0008,103E MR Protocol‐specific = Fail b
Scanning sequence 0018,0020 MR Protocol‐specific = Fail b
Voxel dimension 0028,0030 MR Protocol‐specific Fail b
FOV size 0028,0030, 0028,0011 & 0028,0010 MR NA NA Report only
Phase encode direction 0018,9034 MR NA NA Report only
Pixel bandwidth 0018,0095 MR NA NA Report only
a

Check of tube current is only applicable to GE scanners, as they do not report CTDIvol.

b

For cranial stereotactic treatment only.

c

For non‐cranial stereotactic treatment only.