Updated Guidance for Communicating PFAS Identification Confidence with Ion Mobility Spectrometry

Anna K Boatman 1, Jessie R Chappel 2, Kaylie I Kirkwood-Donelson 3, Jonathon F Fleming 4, David M Reif 5, Emma L Schymanski 6, Julia E Rager 7, Erin S Baker 8
PMCID: PMC12424050  NIHMSID: NIHMS2109076  PMID: 40808630

Abstract

Over the past decade, global contamination from per- and polyfluoroalkyl substances (PFAS) has become apparent due to their detection in countless matrices worldwide, from consumer products to human blood to drinking water. As researchers implement nontargeted analyses (NTA) to more fully understand the PFAS present in the environment and human bodies, clear guidance is needed for consistent and objective reporting of the identified molecules. Confidence levels for small molecules analyzed and identified with high-resolution mass spectrometry (HRMS) have existed since 2014; however, unification of currently used levels and improved guidance for their application is needed due to inconsistencies in reporting and continuing innovations in analytical methods. Here, we (i) investigate current practices for confidence level reporting of PFAS identified with liquid chromatography (LC), gas chromatography (GC), and/or ion mobility spectrometry (IMS) coupled with HRMS and (ii) propose a simple, unified confidence level guidance that incorporates both PFAS-specific attributes and IMS collision cross section (CCS) values.

Keywords: ion mobility spectrometry (IMS), nontargeted analysis (NTA), high-resolution mass spectrometry (HRMS), identification confidence, per- and polyfluoroalkyl substances (PFAS)

Graphical Abstract

graphic file with name nihms-2109076-f0001.jpg

INTRODUCTION

Per- and polyfluoroalkyl substances (PFAS) are a large and growing class of anthropogenic fluorinated chemicals,13 with estimates ranging from 4,700 chemicals on the widely cited National Institute of Standards and Technology (NIST) PFAS Suspect List4 to over 7 million PFAS in PubChem.5 As PFAS production, use, and monitoring expands, these numbers are likely to increase. To date, researchers have typically focused on a limited number of PFAS due to the low availability of analytical standards, but the increased use of nontargeted analyses (NTA) with high resolution mass spectrometry (HRMS) has enabled broader monitoring and novel chemical discovery.13 Similarly, the use of analytical techniques beyond gas or liquid chromatography (GC or LC) coupled to HRMS is expanding. For example, numerous studies cite ion mobility spectrometry (IMS) as a promising technology for PFAS research due to its ability to separate PFAS from biomolecules,1,2,6,7 and its adoption is already growing.

When reporting NTA results, it is critical to assign a confidence level to each identified molecule.8 Confidence levels provide context for chemical identifications and are important for transparent reporting of results,9 interpretation of exposure and toxicity data,10 and appropriate response in the event of an environmental release of unknown chemicals.11 Schymanski et al.8 first defined confidence levels for small molecule identifications made using HRMS in 2014 and others have subsequently adapted these levels. For example, Celma et al.12 incorporated CCS values into the 2014 levels, and Charbonnet et al.3 defined a PFAS-specific system, adding additional sublevels to the 2014 levels using PFAS-specific attributes. However, manual level assignments following any of these schemes are often subjective and prone to unintended user bias (often in the form of overestimation), because lower confidence identifications are often perceived as “worse” than higher confidence ones. Additionally, little guidance is available for what to do when some parameters do not meet the defined requirements.1214 Even our own extensive experience with PFAS NTA using LC-IMS-HRMS has frequently generated situations where confidence levels are not defined by any existing guidance.15,16 Further, there are vast and growing numbers of PFAS candidates, and a rapidly increasing number of researchers performing PFAS NTA on different analytical platforms including GC-HRMS, LC-HRMS, LC-IMS-HRMS,1721 strengthening the need for simple and objective confidence level assignment to support PFAS NTA. Here, we investigate the current reporting of PFAS NTA identifications, address identified gaps by clarifying and unifying existing guidance, and introduce a simplified checklist to help standardize confidence level assignments for PFAS identified using LC-IMS-HRMS. The updated confidence level guidance proposed here expands on efforts by the NTA community to standardize and improve study reporting quality.9,22

MATERIALS AND METHODS

To investigate current practices for confidence level reporting of PFAS identified in NTA using HRMS, information was aggregated across our team’s extensive experience in this field as well as a dedicated literature review examining recently published papers worldwide. Our team’s experience includes the analysis of hundreds of analytical PFAS standards and numerous environmental matrices (water, serum, plasma, whole blood, tissue, etc.).7,15,16,2325 For the literature review, a search was performed in PubMed (RRID: SCR_004846) on 30 September 2024 using the following search query: ((PFAS[Title/Abstract]) OR (polyfluor*[Title/Abstract]) OR (perfluor*[Title/Abstract])) AND ((nontarget*[Title/Abstract]) OR (nontarget*[Title/Abstract]) OR (untarget*-[Title/Abstract])). A total of 426 publications from as early as 1997 were identified through this query, with the 94 studies published in 2023 selected for manual review to assess research trends in more detail. Inclusion in the in-depth analysis required articles specifically use HRMS for PFAS analysis, yielding 38 studies in total from 2023.2663 Information from these articles was then summarized, including platform type, chemical identification reporting approach, matrix analyzed, number of PFAS identified, and confidence level scheme cited (see Table S1). This information was then used to help create the guidance criteria for assigning confidence levels to PFAS identified using LC-IMS-HRMS by understanding gaps in literature and practices which need to be considered in the guidance.

RESULTS AND DISCUSSION

Assessing Current Practices for Confidence Levels in PFAS NTA.

Our literature search identified 38 papers published in 2023 using HRMS for PFAS NTA worldwide (Table S1). The most common platforms were LC-Orbitrap and LC-QTOF (see Table S2 for full breakdown of instruments and experiment types). One study used Fourier transform ion cyclotron resonance (FTICR) and no chromatography, and one study did not specify the type of HRMS instrument used. This illustrates the wide range of platforms available for PFAS NTA and the lack of transparency in reporting. Some studies analyzed PFAS exclusively, while others evaluated a broader range of pollutants that also included PFAS. The types of experiments conducted were diverse, including environmental monitoring (searching for unknown molecules in samples collected from subjects or nature), remediation studies (evaluating PFAS breakdown via intentional degradation technology treatment), industrial analysis (investigating composition of consumer products), and more. A variety of matrices were analyzed, including water (most common), blood, soil, air and analytical standards. Studies identified a wide range of PFAS (0 to 102 PFAS per study; Figure S1), and interestingly, more than a quarter of these studies (12) did not report confidence levels (Figure 1). Among the 26 studies that reported confidence levels, the 2014 levels were the most cited for PFAS identifications. Additionally, we observed a wide distribution of confidence level assignments, with Level 2 being the most frequently assigned (Figure 2). Previous guidance by Schymanski et al.8 and Celma et al.12 stated that Level 2b (probable structure by diagnostic evidence) identifications are usually rare;8,12 however, we found that was not the case with PFAS NTA. In fact, 8 of the 10 studies from the literature search that used sublevels for Level 2 reported Level 2b and/or 2c identifications (range: 5 to 87 PFAS at Level 2b/c PFAS.26,33,36,38,39,55,56,62 Guidance from Charbonnet et al.3 defines Level 2b [probable by diagnostic fragmentation evidence] and 2c [probable by diagnostic homologue evidence] as equivalent confidence). An in-depth evaluation of the methods showed that the confidence level guidelines were applied inconsistently (three such examples are shown in Figure S2). Examples include reported mass errors exceeding author-defined thresholds (which qualifies as an incorrect identification and should not be assigned a confidence level at all) or assigned levels that contradict established guidance. For example, in some cases Level 2b was assigned to structures with explicitly noted potential isomers. However, one exact structure is required for Level 2b, and the possibility of isomers points to Level 3 (tentative candidates per Schymanski et al. or Celma et al. guidance,8,12 or positional isomer candidates per Charbonnet et al. guidance).3 These inconsistencies complicate cross-study comparisons even across identical analytical platforms and call into question the accuracy of reported confidence levels. The Study Reporting Tool (SRT), developed by the Best Practices for Non-Targeted Analysis (BP4NTA) working group to improve reporting quality in NTA studies, includes a prompt for users to report confidence levels and to attach data supporting each identification.9 Of the 38 studies in the literature evaluation, only 2 of them utilized the SRT.26,38 Both of the studies that used the SRT reported confidence levels; however, both of them contained inconsistencies with their cited scheme. These inconsistencies not only hinder the interpretation of published data but also impede comparison of IMS to non-IMS results, and underscore the need for more objective, standardized confidence level assignments. Clearly, there is an opportunity to unify existing confidence level frameworks in a way that enables clear, simple and more comparable reporting as IMS is increasingly adopted in PFAS NTA, and the updates we propose here are complementary with other tools such as the SRT.

Figure 1.

Figure 1.

Confidence level scheme cited in the 38 papers from 2023. 2014 (Small Molecules): cited Schymanski et al.8 scale for small molecules. 2022 (PFAS): cited Charbonnet et al.3 scale for PFAS. None: no confidence levels were reported. Other: cited different confidence levels that do not use a 1–5 scale.

Figure 2.

Figure 2.

Distribution of confidence levels among the 26 papers that identified >0 PFAS and cited Schymanski or Charbonnet scales. A) Percent of PFAS identified at each level by study. In several instances not all PFAS identified were assigned a confidence level, e.g. Study 9 reported 93 PFAS identified, assigned Level 2 to 28 of them, and did not assign confidence levels to the remainder. B) Breakdown of the 10 papers that assigned sublevels to Level 2 probable structures. Nine papers did not assign sublevels, and 8 papers did not report any Level 2 detections. Level 2 identifications are shown with thicker outlines and pattern fill (dots for 2a, diagonal stripes for 2b and 2c). The Charbonnet scale defines 2b and 2c as equivalent confidence, so these levels are combined to facilitate comparison to the Schymanski scale.

To understand potential improvement areas for confidence levels, the 38 NTA studies were evaluated further. A lack of universal standards for acceptable NTA performance3,13,14,64 has been noted elsewhere and was evident in the literature evaluation. For instance, mass tolerances were sometimes not defined or set as high as 20 ppm for HRMS data. Automation is one potential solution for these inconsistencies, where confidence levels would be assigned objectively based on predefined criteria for CCS values and other dimensions. For automation to function, discrete criteria must be provided. Existing scales have attempted to define criteria for mass, CCS, and RT accuracy. However, given the wide variety of analytical techniques and modeling tools already used and being developed for PFAS NTA, universally suitable acceptance tolerances cannot yet be defined. For example, in other literature beyond the 38 studies evaluated here, experimental CCS values can be highly reproducible (with reported reproducibility of ≤0.30% in some cases),65 while predicted CCS values are more variable, with one study noting machine learning tools with errors up to 8%.24 All methods have utility depending on the experimental context, and defining an explicit tolerance to apply universally would be detrimental. Thus, we faced a challenge in unifying objective tolerance criteria. We propose a potential way forward of using an instrument- and method-based set of recommended tolerances, as discussed further below.

Addressing Gaps in Existing Confidence Level Assignments.

To address the current challenges in communicating confidence for PFAS identified using LC-IMS-HRMS our first step was to unify and simplify the existing confidence level guidance for small molecules,8 IMS,12 and PFAS.3 An overarching observation of our literature search was an inconsistent application of current guidance, leading us to conclude that simplification and clarification of existing guidance was necessary. The updated requirements are summarized in a checklist66 (Figure 3), described in detail in the following section, and provided as an extended printable document in the Supplemental Checklist. One key change is the removal of sublevels. Existing scales distinguish between a probable structure based on a library match (Level 2a) and one based on diagnostic evidence if no library entry exists (Level 2b). Charbonnet et al. also defined a Level 2c, defined as equivalent to Level 2b for probable structures elucidated using PFAS homologous series. However, in our experience analyzing PFAS using LC-IMS-HRMS, we have detected PFAS that exist in spectral libraries but not CCS libraries. We have also detected PFAS that match our in-house CCS library, fall exactly on CCS vs m/z and RT vs m/z trendlines for a homologous series, and do not have observable or diagnostic mobility-aligned fragments. Both scenarios would not qualify as a Level 2a under current guidance because a CCS or spectral library match was absent; however, more evidence is available than required for a 2b/c identification. Additionally, IMS libraries are growing, but most spectral libraries do not include CCS values and the vast majority of PFAS in PubChem do not yet have experimental CCS values. Of the 8,099 unique CIDs in PubChem with experimental CCS values, only 208 are PFAS according to the OECD definition implemented in PubChem5 (12 Nov. 2024). As spectral and CCS libraries grow and as analytical instrumentation advances, the gray area between a library-based identification and a diagnostic evidence-based identification will likely blur further. Therefore, to align with our goal of reducing ambiguity, we are removing the sublevel distinction in the updated confidence levels. Other key changes from current guidance are described below and summarized in Table S3.

Figure 3.

Figure 3.

Requirement checklist for PFAS identified using LC- or GC-IMS-HRMS. All listed criteria must be met to assign each confidence level. An extended printable version of the checklist is available in the Supporting Information and online (see Data Availability Statement).

UPDATED PFAS LC- OR GC-IMS-HRMS CONFIDENCE LEVELS

The proposed LC- or GC-IMS-HRMS confidence levels for PFAS unify existing guidance from three sources: small molecules confidence levels introduced in 2014,8 levels incorporating CCS values from 2020,12 and levels specific to PFAS defined in 2022.3 The updated levels for PFAS identified using IMS range from 1 to 5 and have no sublevels. These levels also require that acceptable tolerances for all measured dimensions in each experiment must be defined based on the capabilities of their analytical platform. To avoid selecting confidence levels based on desired results, these tolerances should be defined in advance, with rationale clearly described in the experimental methods, and widely accepted as appropriate tolerances for the analytical platform in question. Platform-specific, experiment-appropriate error tolerances will reduce ambiguity and enable uniform confidence level assignment as technology advances. Table 1 (available as a printable worksheet in Table S4) provides suggested tolerances for the Agilent 6560 IMS-QTOF mass spectrometer based on our own experience with PFAS NTA and reference material analysis, widely accepted platform tolerances,6770 and typical workflows used. Future work (beyond the scope of the current manuscript) includes engaging with other IMS PFAS community members to define recommended reporting and tolerances for other instruments and workflows, for broader applicability.

Table 1.

Example Tolerance Definitions for Each Measured Dimensiona

Parameter Match Type Example Tolerance Definition
Exact m/z All Measured m/z is ±10 ppm of the exact m/z; same adducts observed as with the reference standard (where applicable)
Isotopic pattern All Isotopic pattern visually matches the formula distribution within 10%, with exceptions made for low-abundance features (<5000 peak area) with no M+1 or M+2 detected
CCS Reference standard Measured CCS is ±0.2% of the reference standard value
Library value Measured CCS is ±2.0% of external DTIMS library value
Predicted value Measured CCS value is ±5.0% of expected value based on homologous series trend lines
RT Reference standard Measured RT is ±0.5 min of in-house library values (to account for matrix effects), and must elute in same order
Predicted value Measured RT increases with m/z within homologous series
Mobility-aligned fragments Reference standard MS/MS match cosine similarity >0.9 for mobility-separated fragments
Library value MS/MS match cosine similarity >0.8 for mobility-separated fragments on similar instrument; 0.7 for mobility-separated fragments on other HRMS instrumentation
a

These values are recommended default settings for an Agilent 6560 LC-IMS-HRMS platform and have been used for analysis of PFAS in a variety of environmental matrices, in addition to homologous series trendlines for CCS prediction.

Broadly, at Level 1, the identified PFAS is confirmed by matching experimentally validated values from a reference material in all dimensions. At Level 2, sufficient evidence is present that exactly one structure is probable for the identified PFAS, with all other structural possibilities ruled out. At Level 3, at least one candidate structure must fulfill all available evidence, and multiple candidate structures are acceptable. At Level 4, a structure cannot be elucidated but an unequivocal molecular formula is determined, while at Level 5 the feature cannot be identified beyond being a molecule of interest. Definitions and additional requirements are detailed below.

Definitions.

  • Match: The measured value for a feature is within the user-defined acceptable error tolerance of the comparison value (i.e., the reference standard, library, or predicted value)

  • Diagnostic evidence: Experimental observations that corroborate compound identifications. Examples for PFAS could include homologues, compound synthesis information, MS/MS fragments, ionization behavior, parent-transformation product relationships, etc.8

  • PFAS homologues: Structurally similar compounds differing by one or more repeating structural unit e.g. CF2, CH2CF2, or CF2O71

  • Library: A reputable repository containing validated mass spectra and/or CCS values

  • Prediction: A calculated spectrum or value obtained from a computational tool or extrapolated from experimental data. Example for PFAS include CCS and RT and values derived from homologous series trendlines, subclass-aligned fragments, and more3,24,71

  • Support: The comparison value aligns with the measured value for a feature and may be outside of the user-defined acceptable error (only applicable at Level 3 and lower).

Using CCS Values in Confidence Level Assignment.

IMS is highly useful for separating PFAS from biomolecules, resolving isomers, and generating CCS values which can be plotted along homologous series trendlines.6,7,21 Several different types of IMS separations exist.17,72 Two common methods are drift tube IMS (DTIMS) and traveling wave IMS (TWIMS), both of which have been used to measure CCS values for molecules from a variety of classes, including PFAS.73 CCS values are highly reproducible when using the same IMS platform;20,65,74 however, cross-platform comparisons (e.g., comparing CCS values obtained using DTIMS to those obtained from TWIMS) tend to have higher error.75 For example, Dodds et al. detailed discrepancies between a DTIMS platform and a TWIMS platform, noting CCS measurement differences of around 1% on average between the two platforms, and described a protocol for comparing CCS values across platforms.73 Critically, therefore, different tolerances may be required when using library values from a different IMS platform and should be reported transparently. Additionally, some types of IMS separations, such as field asymmetric IMS (FAIMS), do not generate CCS values.17,72 These platforms add an additional separation dimension and may be considered additional diagnostic evidence using our updated guidance; however, this separation is not equivalent to a CCS value.

IMS performance is commonly assessed by evaluating separation efficiency (typically by measuring the resolving power, Rp, calculated using CCS/ΔCCS) and reproducibility (typically with relative standard deviation, or % RSD). Currently, DTIMS platforms can reasonably achieve Rp up to 60 and certain TWIMS platforms can achieve up to 300 without data processing.73 Analysts should carefully consider the differences in CCS values for their features of interest and ensure their platform can separate them when defining tolerances. For example, Feature X has a measured CCS of 200.1 Å2 using a DTIMS platform with a Rp of 50. Feature X matches two library entries with identical m/z and RT values: Isomer A (library CCS value of 199.7 Å2) and Isomer B (library CCS value of 200.0 Å2). Separating Isomer A and B requires a much higher Rp (>660) than the DTIMS can reasonably achieve, and it would not be appropriate to annotate Feature X as Isomer B based on having a closer CCS value. Without other evidence, such as diagnostic fragmentation spectra, both Isomer A and Isomer B should both be reported as tentative candidates for Feature X, at a Level 3 confidence as described subsequently, potentially nominating Isomer B as the “most likely” candidate as the CCS value is closer. Certain data processing tools, such as high resolution demultiplexing, may promote higher resolution separations on these platforms and should also be considered when defining tolerances.76,77

Confirmed Structure (Level 1).

All Level 1 identifications must meet the minimum requirements of exact m/z, isotopic envelope, CCS, and RT (in the case of LC) or retention index (RI; in the case of GC) match to a reference standard (Figure 3). Reference values from a matching analytical standard must be collected on the same analytical platform as the unknown. These spectra may come from separate injections of the analytical standards or from spiked internal native or mass labeled standards, either analyzed at the same time as the samples of interest, or analyzed previously in-house using identical analytical methods.8,12 The high reproducibility of intra-laboratory CCS values and the ability of IMS to separate PFAS and matrix interferences makes it possible for analysts to use in-house libraries containing RT values, CCS values, and mass spectra without having to run new reference standards alongside experimental samples within the same analytical run. The number of possible Level 1 identifications is thus limited by the number of standards available to the analyst. Existing scales require MS2 spectra for all detected features; however, fragmentation spectra are not always obtainable for low abundance features, which is particularly problematic for environmental analysis of complex biological matrices. Given the high reproducibility of CCS values, the ability of IMS to separate PFAS from interfering biomolecules, and the increased confidence afforded by adding a CCS dimension, mobility-aligned fragments are not required for Level 1 PFAS identifications if diagnostic evidence is available such as the presence of other homologues in the series. Analysts should, however, consider all empirical data available and if anything does not align with the identification, the identification should be downgraded.

Probable Structure (Level 2).

At Level 2, all evidence in all available dimensions must point to exactly one structure and eliminate all others from contention.8 This is the highest confidence level possible when standards cannot be obtained for the molecule in question. All Level 2 identifications must match either a library value or a predicted value (if a library value is not available) for exact m/z, isotopic envelope, CCS, and RT (for LC) or RI (for GC). If the measured value does not match the library or predicted value within the predefined, experiment-appropriate tolerance, it is not sufficient to justify a Level 2 identification; however, it may support a Level 3. The library value must come from a reputable spectral or CCS library or literature appropriate to the experiment.78,79 Predicted values may come from a variety of sources, for example CCS vs m/z and RT vs m/z trendlines for homologous series, and the error of the prediction must be sufficiently narrow to eliminate other candidate structures.15,24 Fragmentation spectra should be used to determine a structure when available; however, it is possible that a combination of MS1 data, CCS values, RT/RI, and diagnostic evidence (such as the presence of homologues at other masses) may be sufficient to justify a Level 2 assignment for some probable structures when no other possibility satisfies all available evidence and when available fragmentation information does not contradict the proposed structure.

Tentative Candidates (Level 3).

With Level 3 identifications, at least one candidate structure is supported by all available evidence.3,8,12 As with Level 2, Level 3 requires either a library or predicted value for exact m/z, isotopic envelope, CCS, and RT or RI for all candidate structures. However, at Level 3, multiple tentative candidate structures are permitted, i.e., the requirement for exactly one structure only is relaxed. Fragmentation spectra are typically required to determine a structure and should be used when available; however, as with Level 2, diagnostic evidence (see Definitions) may be sufficient to propose a tentative candidate structure in the absence of fragmentation data. Additionally, no contradictory fragmentation information that negates the identification should be present. When multiple candidates are proposed, it is recommended to report a “most likely” candidate if possible, although we emphasize that the most likely candidate is not equivalent to a Level 2 probable structure. In the earlier example with Feature X (CCS 200.1 Å2), Isomer B (CCS 200.0 Å2) could be reported as the most likely candidate based on closest library match, with Isomer A (CCS 199.7 Å2) also included as a tentative candidate. Various other scoring terms or lines of evidence could be used for selecting a most likely candidate depending on the software approaches used or information available, such as experimental or in silico fragment matching, experimental or predicted CCS values, literature/patent scores, retention information, etc.80 Level 3 is also an appropriate assignment when a reference standard or library was used, but the calculated errors are outside of the given tolerance criteria; in this case the measured values do not match the comparison values but may still be used to support a tentative candidate structure. While Metz et al.80 recently proposed a “probability” approach to rank cases with multiple “n” candidates, this is not yet integrated in this level system while we await community feedback on their proposal.

Unequivocal Molecular Formula (Level 4).

The definition of a Level 4 identification is an unequivocal formula for the detected feature.3,8,12 The exact m/z and isotopic pattern must match exactly one possible molecular formula. Molecular formula determination typically requires an isotopic envelope and rich fragmentation information, but these are not fixed requirements. The highest scoring formula of a list of possible formulas does not qualify as an unequivocal formula without additional diagnostic evidence, as these are often wrong, especially with the presence of fluorine in the candidate formulas – a fact that is unavoidable in PFAS NTA.

Feature of Interest (Level 5).

Level 5 is the catchall for any feature in the data set that cannot be annotated with an unequivocal formula or candidate structure but is still of potential interest to the experimental question.8,12 This does not mean that every remaining feature is annotated with Level 5. Often in environmental NTA there will be thousands of features that do not even qualify for Level 5; for example, matrix lipids when evaluating PFAS, or statistically insignificant features when comparing between two groups of samples (e.g., statistical comparisons of features detected in upstream vs downstream industrial effluent). The requirements for a Level 5 identification include a measured monoisotopic m/z, a measured CCS, and at least one piece of evidence that supports the feature being a PFAS of interest. In the move toward automation, “of interest” must be explicitly defined. The additional evidence could be database driven, such as an exact mass match to a suspect list PFAS (equivalent to a Level 5a per the 2022 PFAS levels3) with insufficient data to bump up the identification to a Level 4. The evidence could also be experimental (equivalent to a Level 5b per the 2022 PFAS levels3), such as a negative mass defect, feature presence in fluorinated space when plotting its CCS vs m/z, presence of CF2 homologues, presence of CF2-containing fragments, or any number of other justifications.

Discussion and Examples of PFAS Identified with IMS.

In typical LC- and GC-HRMS workflows, chromatographic separation and confirmation with diagnostic MS2 fragments is required for compounds with very close m/z values. Figure 4A shows two PFAS, Hydro-EVE and 6:2 FTS, that have m/z values within 10 ppm and require LC separation on most platforms. However, their CCS values are distinct enough that IMS fully resolves these two molecules.7 An evaluation of our in-house library25 containing experimentally measured RT and CCS values from authenticated standards for 175 PFAS revealed only 3 pairs of PFAS with the potential to overlap with another PFAS in the library when LC, IMS, and HRMS dimensions are all considered with the tolerances shown in Table 1, not including branched isomers (e.g., linear PFOA separates from branched PFOA, but individual branched isomers do not always separate from each other).7 Figure 4B shows one of these pairs, 8:2 FTS and 8:2 FTPA, that may not fully separate using MS1, RT, and CCS depending on the resolving power of the IMS system (remaining pairs shown in Figure S3).

Figure 4.

Figure 4.

Two examples of PFAS measured using LC-IMS-HRMS. The vertical error bars show ±2.0% CCS, and the horizontal error bars show ±10 ppm (for m/z) and ±0.1 min (for RT). A) The exact m/z of deprotonated Hydro-EVE and 6:2 FTS are quite close (426.9651 and 426.9679, respectively, a difference of 6.57 ppm). However, their CCS values (151.2 and 168.1 Å2) diverge enough (11.1%) to fully resolve these two molecules even if they were to coelute. B) On the other hand, 8:2 FTS and 8:2 FTPA have a larger difference in m/z (526.9615 and 526.9705, 17 ppm) that are resolved by a tolerance of ±5 ppm but have a chance of overlapping with a tolerance of ±10 ppm. Because the CCS values are also close and these molecules coelute in this method, MS2 is recommended to confirm which structures are present in the sample. However, depending on the LC gradient and the resolving power of the IMS platform used, MS2 may not be necessary.

Existing confidence level guidance does not currently consider subjective parameters like geographical knowledge and whether certain PFAS may be coming from manufacturing products in the area. However, such context is often important to the PFAS identification workflow and may sometimes factor into confidence level assignments when assigning tentative Level 3 structures, selecting a most likely candidate at Level 3, bumping a tentative Level 3 candidate to a probable Level 2 structure, or confirming a Level 1 identification in the absence of observable fragments. Our updated guidance permits the analyst to use such knowledge as diagnostic evidence so long as it is reported transparently. For example, if a researcher is studying PFAS in water upstream and downstream of a fluorochemical manufacturer and they have existing knowledge that only one of the candidates is produced by that manufacturer, the user could change the confidence from Level 3 to Level 2. All minimum requirements for mass spectra, CCS, and RT must still be met in this case. Thus, in the example illustrated in Figure 4, confidence levels would be assigned as follows, assuming a mass tolerance of ±10 ppm and a CCS tolerance of ±0.5%:

  • 6:2 FTS and Hydro-EVE: Level 1

  • 8:2 FTS and 8:2 FTPA: Level 1 (with diagnostic or fragmentation evidence), Level 3 otherwise (report feature as “8:2 FTS or 8:2 FTPA”).

PERSPECTIVE

Effective NTA workflows are essential for monitoring and discovering PFAS and other emerging compounds, and transparent communication of results is critical. Multidimensional techniques like LC-IMS-HRMS are invaluable for analyzing complex environmental and biological samples; however, current methods suffer from inconsistent data quality and confidence reporting. Here, we adapted existing confidence level frameworks to PFAS identified using LC- and GC-IMS-HRMS to improve data quality and reporting of identification confidence in this rapidly growing area of research. Our research experiences among study coauthors combined with a dedicated literature search revealed the inconsistent use of existing confidence level guidance for PFAS across a wide range of NTA platforms. To address this gap, we unified and simplified existing confidence level guidance by creating a minimum requirements checklist for PFAS identified at each level. We additionally proposed improvements including user-defined, instrument-specific tolerances for quality control and updated requirements for CCS and MS2 fragmentation across all levels. Recommended tolerances for the platform best known to the authors are given in Table 1, and future work will involve consultation with community members to establish other sets of recommended tolerances for other platforms, to ensure broader applicability. These updates provide clear methods for analysts to communicate confidence of PFAS identifications and increase the transparency of NTA results. While this guidance is currently intended for PFAS identified using LC- or GC-IMS-HRMS, researchers using non-IMS platforms focused on molecules other than PFAS may also benefit from these updates. This guidance should be updated as new analytical technologies appear and predictive tools improve, ultimately growing toward automation. To support this, we are developing a web-based application (Small Molecule Identification Scoring Made Easy, or “SMISE”) for automated confidence level assignment as well as additional functionality to aid NTA researchers in proposing and eliminating candidate structures. Release of this tool will enable automated and objective confidence level assignment and unified reporting of evidence used in NTA PFAS identifications.

Supplementary Material

SI 1
SI 2

The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acs.est.5c01354.

Supplemental tables (including the 38 papers included in the literature analysis, breakdown of experiment types, summary of proposed changes to existing guidance, and worksheet for defining tolerances) and figures (including number of PFAS identified in NTA studies from 2023, examples of inconsistent confidence levels, and LC-IMS-HRMS library evaluation of overlapping PFAS) (PDF) Supplemental Checklist: A printable version of the checklist shown in Figure 3, with definitions added, to serve as a standalone reference (PDF)

ACKNOWLEDGMENTS

Funding for this work was made possible through grants from the NIH National Institute of Environmental Health Sciences (P42 ES027704, P42 ES031007, and T32 ES007126), a cooperative agreement with the Environmental Protection Agency (STAR RD 84003201), and support was additionally provided through the Institute for Environmental Health Solutions at the University of North Carolina Gillings School of Global Public Health. E.L.S. acknowledges funding support from the Luxembourg National Research Fund (FNR) for project A18/BM/12341006.

Footnotes

Complete contact information is available at: https://pubs.acs.org/10.1021/acs.est.5c01354

The authors declare no competing financial interest.

Contributor Information

Anna K. Boatman, Department of Chemistry, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina 27514, United States

Jessie R. Chappel, Department of Environmental Sciences and Engineering, Gillings School of Global Public Health and The Institute for Environmental Health Solutions, Gillings School of Global Public Health, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina 27514, United States; Curriculum in Toxicology and Environmental Medicine, School of Medicine, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina 27514, United States

Kaylie I. Kirkwood-Donelson, Immunity, Inflammation, and Disease Laboratory, National Institute of Environmental Health Sciences, Durham, North Carolina 27709, United States

Jonathon F. Fleming, Predictive Toxicology Branch, Division of Translational Toxicology, National Institute of Environmental Health Sciences, Durham, North Carolina 27713, United States

David M. Reif, Predictive Toxicology Branch, Division of Translational Toxicology, National Institute of Environmental Health Sciences, Durham, North Carolina 27713, United States

Emma L. Schymanski, Luxembourg Centre for Systems Biomedicine (LCSB), University of Luxembourg, L-4367 Belvaux, Luxembourg

Julia E. Rager, Department of Environmental Sciences and Engineering, Gillings School of Global Public Health and The Institute for Environmental Health Solutions, Gillings School of Global Public Health, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina 27514, United States; Curriculum in Toxicology and Environmental Medicine, School of Medicine, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina 27514, United States

Erin S. Baker, Department of Chemistry, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina 27514, United States

Data Availability Statement

The Baker lab PFAS LC-IMS-HRMS library is freely available on Zenodo (DOI: 10.5281/zenodo.14341320). A standalone checklist is also available on Zenodo (DOI: 10.5281/zenodo.15441451).

REFERENCES

  • (1).Perera D; Scott W; Smolinski R; Mukhopadhyay L; McDonough CA Techniques to characterize PFAS burden in biological samples: Recent insights and remaining challenges. Trends in Environmental Analytical Chemistry 2024, 41, No. e00224. [Google Scholar]
  • (2).Strynar M; McCord J; Newton S; Washington J; Barzen-Hanson K; Trier X; Liu Y; Dimzon IK; Bugsel B; Zwiener C; Munoz G Practical application guide for the discovery of novel PFAS in environmental samples using high resolution mass spectrometry. Journal of Exposure Science & Environmental Epidemiology 2023, 33 (4), 575–588. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (3).Charbonnet JA; McDonough CA; Xiao F; Schwichtenberg T; Cao D; Kaserzon S; Thomas KV; Dewapriya P; Place BJ; Schymanski EL; Field JA; Helbling DE; Higgins CP Communicating Confidence of Per- and Polyfluoroalkyl Substance Identification via High-Resolution Mass Spectrometry. Environmental Science & Technology Letters 2022, 9 (6), 473–481. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (4).PFAS Suspect List. National Institute of Standards and Technology Public Data Repository. 2023. https://data.nist.gov/od/id/mds2-2387 (accessed 2024-10-22).
  • (5).Schymanski EL; Zhang J; Thiessen PA; Chirsir P; Kondic T; Bolton EE Per- and Polyfluoroalkyl Substances (PFAS) in PubChem: 7 Million and Growing. Environ. Sci. Technol 2023, 57 (44), 16918–16928. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (6).Díaz-Galiano FJ; Murcia-Morales M; Monteau F; Le Bizec B; Dervilly G Collision cross-section as a universal molecular descriptor in the analysis of PFAS and use of ion mobility spectrum filtering for improved analytical sensitivities. Anal. Chim. Acta 2023, 1251, No. 341026. [DOI] [PubMed] [Google Scholar]
  • (7).Dodds JN; Hopkins ZR; Knappe DRU; Baker ES Rapid Characterization of Per- and Polyfluoroalkyl Substances (PFAS) by Ion Mobility Spectrometry–Mass Spectrometry (IMS-MS). Anal. Chem 2020, 92 (6), 4427–4435. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (8).Schymanski EL; Jeon J; Gulde R; Fenner K; Ruff M; Singer HP; Hollender J Identifying Small Molecules via High Resolution Mass Spectrometry: Communicating Confidence. Environ. Sci. Technol 2014, 48 (4), 2097–2098. [DOI] [PubMed] [Google Scholar]
  • (9).Peter KT; Phillips AL; Knolhoff AM; Gardinali PR; Manzano CA; Miller KE; Pristner M; Sabourin L; Sumarah MW; Warth B; Sobus JR Nontargeted Analysis Study Reporting Tool: A Framework to Improve Research Transparency and Reproducibility. Anal. Chem 2021, 93 (41), 13870–13879. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (10).Schrimpe-Rutledge AC; Codreanu SG; Sherrod SD; McLean JA Untargeted Metabolomics StrategiesChallenges and Emerging Directions. J. Am. Soc. Mass Spectrom 2016, 27 (12), 1897–1905. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (11).Phillips AL; Williams AJ; Sobus JR; Ulrich EM; Gundersen J; Langlois-Miller C; Newton SR A framework for utilizing high-resolution mass spectrometry and nontargeted analysis in rapid response and emergency situations. Environmental toxicology and chemistry 2022, 41 (5), 1117–1130. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (12).Celma A; Sancho JV; Schymanski EL; Fabregat-Safont D; Ibáñez M; Goshawk J; Barknowitz G; Hernández F; Bijlsma L Improving Target and Suspect Screening High-Resolution Mass Spectrometry Workflows in Environmental Analysis by Ion Mobility Separation. Environ. Sci. Technol 2020, 54 (23), 15120–15131. [DOI] [PubMed] [Google Scholar]
  • (13).Schulze B; Jeon Y; Kaserzon S; Heffernan AL; Dewapriya P; O’Brien J; Gomez Ramos MJ; Ghorbani Gorji S; Mueller JF; Thomas KV; Samanipour S An assessment of quality assurance/quality control efforts in high resolution mass spectrometry non-target workflows for analysis of environmental samples. TrAC Trends in Analytical Chemistry 2020, 133, No. 116063. [Google Scholar]
  • (14).Fisher CM; Peter KT; Newton SR; Schaub AJ; Sobus JR Approaches for assessing performance of high-resolution mass spectrometry–based non-targeted analysis methods. Anal. Bioanal. Chem 2022, 414 (22), 6455–6471. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (15).Kirkwood-Donelson KI; Dodds JN; Schnetzer A; Hall N; Baker ES Uncovering per- and polyfluoroalkyl substances (PFAS) with nontargeted ion mobility spectrometry–mass spectrometry analyses. Science Advances 2023, 9 (43), No. eadj7048. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (16).Boatman AK; Chappel JR; Polera ME; Dodds JN; Belcher SM; Baker ES Assessing Per- and Polyfluoroalkyl Substances in Fish Fillet Using Non-Targeted Analyses. Environ. Sci. Technol 2024, 58 (32), 14486–14495. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (17).Dodds JN; Baker ES Ion mobility spectrometry: fundamental concepts, instrumentation, applications, and the road ahead. J. Am. Soc. Mass Spectrom 2019, 30 (11), 2185–2195. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (18).Zheng X; Aly NA; Zhou Y; Dupuis KT; Bilbao A; Paurus VL; Orton DJ; Wilson R; Payne SH; Smith RD; et al. A structural examination and collision cross section database for over 500 metabolites and xenobiotics using drift tube ion mobility spectrometry. Chemical science 2017, 8 (11), 7724–7736. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (19).Picache JA; Rose BS; Balinski A; Leaptrot KL; Sherrod SD; May JC; McLean JA Collision cross section compendium to annotate and predict multi-omic compound identities. Chemical Science 2019, 10 (4), 983–993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (20).Izquierdo-Sandoval D; Fabregat-Safont D; Lacalle-Bergeron L; Sancho JV; Hernández F; Portoles T Benefits of Ion Mobility Separation in GC-APCI-HRMS Screening: From the Construction of a CCS Library to the Application to Real-World Samples. Anal. Chem 2022, 94 (25), 9040–9047. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (21).Celma A; Alygizakis N; Belova L; Bijlsma L; Fabregat-Safont D; Menger F; Gil-Solsona R Ion mobility separation coupled to high-resolution mass spectrometry in environmental analysis – Current state and future potential. Trends in Environmental Analytical Chemistry 2024, 43, No. e00239. [Google Scholar]
  • (22).Phillips AL; Peter KT; Sobus JR; Fisher CM; Manzano CA; McEachran AD; Williams AJ; Knolhoff AM; Ulrich EM Standardizing non-targeted analysis reporting to advance exposure science and environmental epidemiology. Journal of Exposure Science & Environmental Epidemiology 2023, 33 (4), 501–504. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (23).Dodds JN; Kirkwood-Donelson KI; Boatman AK; Knappe DR; Hall NS; Schnetzer A; Baker ES Evaluating Solid Phase Adsorption Toxin Tracking (SPATT) for passive monitoring of per-and polyfluoroalkyl substances (PFAS) with Ion Mobility Spectrometry-Mass Spectrometry (IMS-MS). Science of The Total Environment 2024, 947, No. 174574. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (24).Foster M; Rainey M; Watson C; Dodds JN; Kirkwood KI; Fernández FM; Baker ES Uncovering PFAS and Other Xenobiotics in the Dark Metabolome Using Ion Mobility Spectrometry, Mass Defect Analysis, and Machine Learning. Environ. Sci. Technol 2022, 56 (12), 9133–9143. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (25).Joseph K; Boatman A; Dodds J; Kirkwood Donelson K; Ryan J; Zhang J; Thiessen P; Bolton E; Valdiviezo A; Sapozhnikova Y; Rusyn I; Schymanski E; Baker E Dataset for “Multidimensional library for the improved identification of per- and polyfluoroalkyl substances (PFAS)”. Zenodo. 2024. DOI: 10.5281/zenodo.14341320 (accessed 2024-12-09). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (26).Bowers BB; Lou Z; Xu J; De Silva AO; Xu X; Lowry GV; Sullivan RC Nontarget analysis and fluorine atom balances of transformation products from UV/sulfite degradation of perfluoroalkyl contaminants. Environ. Sci. Process Impacts 2023, 25 (3), 472–483. [DOI] [PubMed] [Google Scholar]
  • (27).Carlsson H; Sreenivasan AP; Erngren I; Larsson A; Kultima K Combining the targeted and untargeted screening of environmental contaminants reveals associations between PFAS exposure and vitamin D metabolism in human plasma. Environ. Sci. Process Impacts 2023, 25 (6), 1116–1130. [DOI] [PubMed] [Google Scholar]
  • (28).Casey JS; Jackson SR; Ryan J; Newton SR The use of gas chromatography - high resolution mass spectrometry for suspect screening and non-targeted analysis of per- and polyfluoroalkyl substances. J. Chromatogr A 2023, 1693, No. 463884. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (29).Cheng X; Liu L; Ge Y; Weber R; Huang J Target and non-target analysis of per- and polyfluoroalkyl substances in representative chrome mist suppressants on the Chinese market. Chemosphere 2023, 337, No. 139419. [DOI] [PubMed] [Google Scholar]
  • (30).Chibwe L; De Silva AO; Spencer C; Teixera CF; Williamson M; Wang X; Muir DCG Target and Nontarget Screening of Organic Chemicals and Metals in Recycled Plastic Materials. Environ. Sci. Technol 2023, 57 (8), 3380–3390. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (31).Dewapriya P; Nilsson S; Ghorbani Gorji S; O’Brien JW; Braunig J; Gomez Ramos MJ; Donaldson E; Samanipour S; Martin JW; Mueller JF; Kaserzon SL; Thomas KV Novel Per- and Polyfluoroalkyl Substances Discovered in Cattle Exposed to AFFF-Impacted Groundwater. Environ. Sci. Technol 2023, 57 (36), 13635–13645. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (32).Gobindlal K; Zujovic Z; Jaine J; Weber CC; Sperry J Solvent-Free, Ambient Temperature and Pressure Destruction of Perfluorosulfonic Acids under Mechanochemical Conditions: Degradation Intermediates and Fluorine Fate. Environ. Sci. Technol 2023, 57 (1), 277–285. [DOI] [PubMed] [Google Scholar]
  • (33).Gonda N; Choyke S; Schaefer C; Higgins CP; Voelker B Hydroxyl Radical Transformations of Perfluoroalkyl Acid (PFAA) Precursors in Aqueous Film Forming Foams (AFFFs). Environ. Sci. Technol 2023, 57 (21), 8053–8064. [DOI] [PubMed] [Google Scholar]
  • (34).Gutierrez-Martin D; Gil-Solsona R; Saaltink MW; Rodellas V; Lopez-Serna R; Folch A; Carrera J; Gago-Ferrero P Chemicals of emerging concern in coastal aquifers: Assessment along the land-ocean interface. J. Hazard Mater 2023, 448, No. 130876. [DOI] [PubMed] [Google Scholar]
  • (35).Hao S; Reardon PN; Choi YJ; Zhang C; Sanchez JM; Higgins CP; Strathmann TJ Hydrothermal Alkaline Treatment (HALT) of Foam Fractionation Concentrate Derived from PFAS-Contaminated Groundwater. Environ. Sci. Technol 2023, 57 (44), 17154–17165. [DOI] [PubMed] [Google Scholar]
  • (36).Hu J; Lyu Y; Chen H; Cai L; Li J; Cao X; Sun W Integration of target, suspect, and nontarget screening with risk modeling for per- and polyfluoroalkyl substances prioritization in surface waters. Water Res. 2023, 233, No. 119735. [DOI] [PubMed] [Google Scholar]
  • (37).Jacob P; Helbling DE Exploring the Evolution of Organofluorine-Containing Compounds during Simulated Photo-lithography Experiments. Environ. Sci. Technol 2023, 57 (34), 12819–12828. [DOI] [PubMed] [Google Scholar]
  • (38).Kim Y; Pike KA; Gray R; Sprankle JW; Faust JA; Edmiston PL Non-targeted identification and semi-quantitation of emerging per- and polyfluoroalkyl substances (PFAS) in US rainwater. Environ. Sci. Process Impacts 2023, 25 (11), 1771–1787. [DOI] [PubMed] [Google Scholar]
  • (39).Kirkwood-Donelson KI; Dodds JN; Schnetzer A; Hall N; Baker ES Uncovering per- and polyfluoroalkyl substances (PFAS) with nontargeted ion mobility spectrometry-mass spectrometry analyses. Sci. Adv 2023, 9 (43), No. eadj7048. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (40).Koelmel JP; Lin EZ; Parry E; Stelben P; Rennie EE; Godri Pollitt KJ Novel perfluoroalkyl substances (PFAS) discovered in whole blood using automated non-targeted analysis of dried blood spots. Sci. Total Environ 2023, 883, No. 163579. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (41).Li L; Gao R; Wang X; Deng Y; Sun H; Sun H; Zhang B; Yu N; Gu C; Pan B; Yu H; Wei S SWATH-F: A Novel Nontarget Strategy Based on the SWATH-MS Deconvolution Method Assisting in Annotating PFAS Homologues in Multisample Studies. Anal. Chem 2023, 95 (39), 14551–14557. [DOI] [PubMed] [Google Scholar]
  • (42).Liu J; Zhao Z; Li J; Hua X; Zhang B; Tang C; An X; Lin T Emerging and legacy perfluoroalkyl and polyfluoroalkyl substances (PFAS) in surface water around three international airports in China. Chemosphere 2023, 344, No. 140360. [DOI] [PubMed] [Google Scholar]
  • (43).Liu Y; Lu MY; Bao J; Shao LX; Yu WJ; Hu XM; Zhao X Periodically reversing electrocoagulation technique for efficient removal of short-chain perfluoroalkyl substances from contaminated groundwater around a fluorochemical facility. Chemosphere 2023, 334, No. 138953. [DOI] [PubMed] [Google Scholar]
  • (44).Manz KE; Dodson RE; Liu Y; Scheidl L; Burks S; Dunn F; Gairola R; Lee NF; Walker ED; Pennell KD; Braun JM Effects of Corsi-Rosenthal boxes on indoor air contaminants: non-targeted analysis using high resolution mass spectrometry. J. Expo Sci. Environ. Epidemiol 2023, 33 (4), 537–547. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (45).Mok S; Lee S; Choi Y; Jeon J; Kim YH; Moon HB Target and non-target analyses of neutral per- and polyfluoroalkyl substances from fluorochemical industries using GC-MS/MS and GC-TOF: Insights on their environmental fate. Environ. Int 2023, 182, No. 108311. [DOI] [PubMed] [Google Scholar]
  • (46).Munoz G; Liu M; Vo Duy S; Liu J; Sauve S Target and nontarget screening of PFAS in drinking water for a large-scale survey of urban and rural communities in Quebec, Canada. Water Res. 2023, 233, No. 119750. [DOI] [PubMed] [Google Scholar]
  • (47).Oro-Nolla B; Dulsat-Masvidal M; Bertolero A; Lopez-Antia A; Lacorte S Target and untargeted screening of perfluoroalkyl substances in biota using liquid chromatography coupled to quadrupole time of flight mass spectrometry. J. Chromatogr A 2023, 1701, No. 464066. [DOI] [PubMed] [Google Scholar]
  • (48).Qiao B; Song D; Fang B; Yu H; Li X; Zhao L; Yao Y; Zhu L; Chen H; Sun H Nontarget Screening and Fate of Emerging Per- and Polyfluoroalkyl Substances in Wastewater Treatment Plants in Tianjin, China. Environ. Sci. Technol 2023, 57 (48), 20127–20137. [DOI] [PubMed] [Google Scholar]
  • (49).Renyer A; Ravindra K; Wetmore BA; Ford JL; DeVito M; Hughes MF; Wehmas LC; MacMillan DK Dose Response, Dosimetric, and Metabolic Evaluations of Replacement PFAS Perfluoro-(2,5,8-trimethyl-3,6,9-trioxadodecanoic) Acid (HFPO-TeA). Toxics 2023, 11 (12), 951. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (50).Sanchez-Resino E; Marques M; Gutierrez-Martin D; Restrepo-Montes E; Martinez MA; Salas-Huetos A; Babio N; Salas-Salvado J; Gil-Solsona R; Gago-Ferrero P Exploring the Occurrence of Organic Contaminants in Human Semen through an Innovative LC-HRMS-Based Methodology Suitable for Target and Nontarget Analysis. Environ. Sci. Technol 2023, 57 (48), 19236–19252. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (51).Schwartz-Narbonne H; Xia C; Shalin A; Whitehead HD; Yang D; Peaslee GF; Wang Z; Wu Y; Peng H; Blum A; Venier M; Diamond ML Per- and Polyfluoroalkyl Substances in Canadian Fast Food Packaging. Environ. Sci. Technol. Lett 2023, 10 (4), 343–349. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (52).Sdougkou K; Xie H; Papazian S; Bonnefille B; Bergdahl IA; Martin JW Phospholipid Removal for Enhanced Chemical Exposomics in Human Plasma. Environ. Sci. Technol 2023, 57 (28), 10173–10184. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (53).Siegel HG; Nason SL; Warren JL; Prunas O; Deziel NC; Saiers JE Investigation of Sources of Fluorinated Compounds in Private Water Supplies in an Oil and Gas-Producing Region of Northern West Virginia. Environ. Sci. Technol 2023, 57 (45), 17452–17464. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (54).Steeves KL; Bissram MJ; Kleywegt S; Stevens D; Dorman FL; Simpson AJ; Simpson MJ; Cahill LS; Jobst KJ Nontargeted screening reveals fluorotelomer ethoxylates in indoor dust and industrial wastewater. Environ. Int 2023, 171, No. 107634. [DOI] [PubMed] [Google Scholar]
  • (55).Tang C; Zheng R; Zhu Y; Liang Y; Liang Y; Liang S; Xu J; Zeng YH; Luo XJ; Lin H; Huang Q; Mai BX Nontarget Analysis and Comprehensive Characterization of Iodinated Polyfluoroalkyl Acids in Wastewater and River Water by LC-HRMS with Cascade Precursor-Ion Exclusions and Algorithmic Approach. Environ. Sci. Technol 2023, 57 (44), 17099–17109. [DOI] [PubMed] [Google Scholar]
  • (56).Tang C; Zhu Y; Liang Y; Zeng YH; Peng X; Mai BX; Xu J; Huang Q; Lin H First Discovery of Iodinated Polyfluoroalkyl Acids by Nontarget Mass-Spectrometric Analysis and Iodine-Specific Screening Algorithm. Environ. Sci. Technol 2023, 57, 1378. [DOI] [PubMed] [Google Scholar]
  • (57).Tisler S; Savvidou P; Jorgensen MB; Castro M; Christensen JH Supercritical Fluid Chromatography Coupled to High-Resolution Mass Spectrometry Reveals Persistent Mobile Organic Compounds with Unknown Toxicity in Wastewater Effluents. Environ. Sci. Technol 2023, 57 (25), 9287–9297. [DOI] [PubMed] [Google Scholar]
  • (58).van Gerwen M; Colicino E; Guan H; Dolios G; Nadkarni GN; Vermeulen RCH; Wolff MS; Arora M; Genden EM; Petrick LM Per- and polyfluoroalkyl substances (PFAS) exposure and thyroid cancer risk. EBioMedicine 2023, 97, No. 104831. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (59).Wang Y; Ji Y; Tishchenko V; Huang Q Removing per- and polyfluoroalkyl substances (PFAS) in water by foam fractionation. Chemosphere 2023, 311, No. 137004. [DOI] [PubMed] [Google Scholar]
  • (60).Wickersham LC; Mattila JM; Krug JD; Jackson SR; Wallace MAG; Shields EP; Halliday H; Li EY; Liberatore HK; Farrior SM; Preston W; Ryan JV; Lee CW; Linak WP Characterization of PFAS air emissions from thermal application of fluoropolymer dispersions on fabrics. J. Air Waste Manag Assoc 2023, 73 (7), 533–552. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (61).Zhao M; Yao Y; Dong X; Baqar M; Fang B; Chen H; Sun H Nontarget Identification of Novel Per- and Polyfluoroalkyl Substances (PFAS) in Soils from an Oil Refinery in Southwestern China: A Combined Approach with TOP Assay. Environ. Sci. Technol 2023, 57 (48), 20194–20205. [DOI] [PubMed] [Google Scholar]
  • (62).Zweigle J; Bugsel B; Rohler K; Haluska AA; Zwiener C PFAS-Contaminated Soil Site in Germany: Nontarget Screening before and after Direct TOP Assay by Kendrick Mass Defect and FindPFDeltaS. Environ. Sci. Technol 2023, 57 (16), 6647–6655. [DOI] [PubMed] [Google Scholar]
  • (63).Zweigle J; Capitain C; Simon F; Roesch P; Bugsel B; Zwiener C Non-extractable PFAS in functional textiles - characterization by complementary methods: oxidation, hydrolysis, and fluorine sum parameters. Environ. Sci. Process Impacts 2023, 25 (8), 1298–1310. [DOI] [PubMed] [Google Scholar]
  • (64).BP4NTA. QA/QC Metrics. 2024. https://nontargetedanalysis.org/reference-content/results/qa-qc-metrics/ (accessed 2024-10-04).
  • (65).Stow SM; Causon TJ; Zheng X; Kurulugama RT; Mairinger T; May JC; Rennie EE; Baker ES; Smith RD; McLean JA; Hann S; Fjeldsted JC An Interlaboratory Evaluation of Drift Tube Ion Mobility–Mass Spectrometry Collision Cross Section Measurements. Anal. Chem 2017, 89 (17), 9048–9055. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (66).Boatman A; Chappel J; Kirkwood-Donelson K; Fleming J; Reif D; Schymanski E; Rager J; Baker E LC- or GC-IMS-HRMS Confidence Level Requirement Checklist for PFAS v2. Zenodo. 2025. DOI: 10.5281/zenodo.15441451. [DOI] [Google Scholar]
  • (67).Hollender J; Schymanski EL; Ahrens L; Alygizakis N; Béen F; Bijlsma L; Brunner AM; Celma A; Fildier A; Fu Q; Gago-Ferrero P; Gil-Solsona R; Haglund P; Hansen M; Kaserzon S; Kruve A; Lamoree M; Margoum C; Meijer J; Merel S; Rauert C; Rostkowski P; Samanipour S; Schulze B; Schulze T; Singh RR; Slobodnik J; Steininger-Mairinger T; Thomaidis NS; Togola A; Vorkamp K; Vulliet E; Zhu L; Krauss M NORMAN guidance on suspect and non-target screening in environmental monitoring. Environmental Sciences Europe 2023, 35 (1), 75. [Google Scholar]
  • (68).Oberacher H; Reinstadler V; Kreidl M; Stravs MA; Hollender J; Schymanski EL Annotating Nontargeted LC-HRMS/MS Data with Two Complementary Tandem Mass Spectral Libraries. Metabolites 2019, 9 (1), 3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (69).Oberacher H; Sasse M; Antignac J-P; Guitton Y; Debrauwer L; Jamin EL; Schulze T; Krauss M; Covaci A; Caballero-Casero N; Rousseau K; Damont A; Fenaille F; Lamoree M; Schymanski EL A European proposal for quality control and quality assurance of tandem mass spectral libraries. Environmental Sciences Europe 2020, 32 (1), 43. [Google Scholar]
  • (70).Harwood TV; Treen DGC; Wang M; de Jong W; Northen TR; Bowen BP BLINK enables ultrafast tandem mass spectrometry cosine similarity scoring. Sci. Rep 2023, 13 (1), No. 13462. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (71).Koelmel JP; Paige MK; Aristizabal-Henao JJ; Robey NM; Nason SL; Stelben PJ; Li Y; Kroeger NM; Napolitano MP; Savvaides T; Vasiliou V; Rostkowski P; Garrett TJ; Lin E; Deigl C; Jobst K; Townsend TG; Godri Pollitt KJ; Bowden JA Toward Comprehensive Per- and Polyfluoroalkyl Substances Annotation Using FluoroMatch Software and Intelligent High-Resolution Tandem Mass Spectrometry Acquisition. Anal. Chem 2020, 92 (16), 11186–11194. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (72).Kirkwood KI; Odenkirk MT; Baker ES Ion Mobility Spectrometry. In Mass Spectrometry for Lipidomics: Methods and Applications; Wiley-VCH GmbH, 2023. [Google Scholar]
  • (73).Dodds JN; Ford LC; Ryan JP; Solosky AM; Rusyn I; Baker ES Evaluating Ion Mobility Data Acquisition, Calibration, and Processing for Small Molecules: A Cross-Platform Assessment of Drift Tube and Traveling Wave Methodologies. J. Am. Soc. Mass Spectrom 2025, 36, 1456. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (74).Vasilopoulou CG; Sulek K; Brunner A-D; Meitei NS; Schweiger-Hufnagel U; Meyer SW; Barsch A; Mann M; Meier F Trapped ion mobility spectrometry and PASEF enable in-depth lipidomics from minimal sample amounts. Nat. Commun 2020, 11 (1), 331. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (75).George AC; Schmitz I; Rouvière F; Alves S; Colsch B; Heinisch S; Afonso C; Fenaille F; Loutelier-Bourhis C Interplat-form comparison between three ion mobility techniques for human plasma lipid collision cross sections. Anal. Chim. Acta 2024, 1304, No. 342535. [DOI] [PubMed] [Google Scholar]
  • (76).Bilbao A; Gibbons BC; Stow SM; Kyle JE; Bloodsworth KJ; Payne SH; Smith RD; Ibrahim YM; Baker ES; Fjeldsted JC A Preprocessing Tool for Enhanced Ion Mobility–Mass Spectrometry-Based Omics Workflows. J. Proteome Res 2022, 21 (3), 798–807. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (77).Butler KE; Baker ES A High-Throughput Ion Mobility Spectrometry–Mass Spectrometry Screening Method for Opioid Profiling. J. Am. Soc. Mass Spectrom 2022, 33 (10), 1904–1913. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (78).BP4NTA. Data Processing and Analysis. 2024. https://nontargetedanalysis.org/reference-content/methods/data-processing-and-analysis/ (accessed 2025-01-20).
  • (79).Elapavalore A; Kondić T; Singh RR; Shoemaker BA; Thiessen PA; Zhang J; Bolton EE; Schymanski EL Adding open spectral data to MassBank and PubChem using open source tools to support non-targeted exposomics of mixtures. Environmental Science: Processes & Impacts 2023, 25 (11), 1788–1801. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (80).Metz TO; Chang CH; Gautam V; Anjum A; Tian S; Wang F; Colby SM; Nunez JR; Blumer MR; Edison AS; Fiehn O; Jones DP; Li S; Morgan ET; Patti GJ; Ross DH; Shapiro MR; Williams AJ; Wishart DS Introducing “Identification Probability” for Automated and Transferable Assessment of Metabolite Identification Confidence in Metabolomics and Related Studies. Anal. Chem 2025, 97 (1), 1–11. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

SI 1
SI 2

Data Availability Statement

The Baker lab PFAS LC-IMS-HRMS library is freely available on Zenodo (DOI: 10.5281/zenodo.14341320). A standalone checklist is also available on Zenodo (DOI: 10.5281/zenodo.15441451).

RESOURCES