Skip to main content
PLOS One logoLink to PLOS One
. 2025 Sep 11;20(9):e0331969. doi: 10.1371/journal.pone.0331969

Point-of-care creatinine-based eGFR (StatSensor) in detecting kidney dysfunction (KD) among people living with HIV in Tanzania

Daniel Msilanga 1,2,*, Anthony Muiru 3, Elizabeth Msangi 1,2, Jacqueline Shoo 1,2, Jonathan Mngumi 1,2, Ewaldo Komba 1,2, Emmanuel Balandya 1, Paschal Ruggajo 4, Rajendra Bhimma 5, Kathleen Liu 3
Editor: Mogamat-Yazied Chothia6
PMCID: PMC12425315  PMID: 40934196

Abstract

Introduction

Kidney Dysfunction (KD) is prevalent among people living with HIV (PLHIV) in low- and middle-income countries (LMICs), but routine screening is limited due to inadequate laboratory infrastructure. The StatSensor® Point-of-Care (POC) Creatinine Test offers a rapid, cost-effective alternative for early KD detection, though its accuracy in PLHIV remains uncertain.

Methods

We conducted a diagnostic accuracy cross-sectional study at Temeke Regional Referral Hospital (TRRH) HIV Clinic from January to March 2025 among PLHIV aged ≥18 years. Kidney dysfunction (KD) was defined as an estimated glomerular filtration rate (eGFR) <60 mL/min/1.73 m² using the CKD-EPI 2021 equation. We compared StatSensor point-of-care eGFR results with eGFR derived from serum creatinine measured by the Jaffe method. Diagnostic performance metrics including sensitivity, specificity, predictive values, and receiver operating characteristic (ROC) curves were reported.

Results

Among 358 participants, the median age was 48 years, with 66.2% female and 15.6% having KD (eGFR < 60 mL/min/1.73m²). The StatSensor demonstrated 92.9% sensitivity, 94.7% specificity, and 94.4% overall diagnostic accuracy compared to the Jaffe method. The ROC curve (AUC = 0.938) indicated strong test performance, showing substantial agreement with a kappa value of 0.805. Bland-Altman analysis revealed a negative bias of 4.36 mL/min/1.73 m² with limits of agreement from −19.68 to 28.40 and a strong correlation (R² = 0.813) between the two methods.

Conclusion

The StatSensor POC Creatinine test demonstrated high diagnostic accuracy and strong agreement with the standard Jaffe method, indicating its potential as a reliable screening tool for kidney dysfunction in PLHIV in resource-limited settings.

Introduction

Kidney dysfunction (KD) is a significant concern among people living with HIV (PLHIV) in low-middle-income countries (LMICs), with prevalence rates ranging from 10% to 35% [1–3]. Despite guideline recommendations for routine screening, implementation remains limited, particularly in primary healthcare settings, where nearly 70% of people seek care. It is estimated that only one in three patients receives recommended creatinine testing (2). This is largely due to inadequate laboratory infrastructure [4–6]As a result, many patients are not diagnosed in the early stages of KD, and patients are typically identified only when they present with symptoms of end stage kidney failure (ESKD), often requiring hemodialysis [7–9]. There is an urgent need for a feasible, and accessible test, to improve early detection and management of KD in PLHIV [9].

The StatSensor® POC Creatinine test (by Nova Biomedical Waltham, MA, USA), is a handheld point-of-care test designed to overcome the limitations of traditional laboratory creatinine testing [10,11]. This user-friendly and device requires minimal training and provides rapid results using capillary/venous blood from a finger stick, enabling early identification of KD [11]. Several studies on the StatSensor Creatinine test have shown variable findings, including sensitivity and specificity above 90% and a positive mean eGFR bias compared to laboratory-based creatinine measurements by Jaffe method [10,11–14]. While this test presents a promising option for KD screening, its performance variability across different populations warrants further evaluation, and its utility as a screening tool for KD in PLHIV in resource-limited settings remains unexplored [9].

To address the gap in KD screening among PLHIV in Tanzania and similar resource-limited settings, this study aimed to evaluate the diagnostic accuracy of the StatSensor Creatinine test in detecting KD among PLHIV receiving HIV care. By comparing its performance with, conventional laboratory-based serum creatinine test, this study aimed to provide critical insights into its potential as a reliable and cost-effective screening tool in settings with limited laboratory capacity.

Methods

Ethical approval

The study was reviewed and approved by the National Institute for Medical Research (NIMR) under reference number NIMR/HQ/R.8a/Vol.IX/4695 and permission to conduct the study was given by Temeke Regional Referral Hospital. The study objectives, methods, procedures, and potential risks were thoroughly explained to participants before obtaining their consent. All participants provided written informed consent before enrollment and granted permission before undergoing any study procedures.

Study design and settings

We conducted a diagnostic accuracy cross-sectional study at Temeke Regional Referral Hospital (TRRH) HIV Clinic from 5th January to 30th March 2025. TRRH is one of the largest HIV care centers in the region, with 24,000–30,000 study visits per year, and a catchment area of 1,346,674 resident of Tanzania.

Recruitment procedure

We included people living with HIV (PLHIV) aged 18 years or older who were receiving care at TRRH during the study period. Participants were selected daily using Probability Proportional to Size (PPS) sampling based on the clinic’s appointment register to ensure representation across varying clinic volumes. Each eligible patient was assigned a unique study ID, and recruitment continued daily until the target sample size was reached.

Kidney dysfunction was defined as eGFR < 60 mL/min/1.73m², calculated using the CKD-EPI 2021 equation (2,17). All eligible patients were informed about the study, and those who provided written informed consent were enrolled. Data were collected using a structured, interviewer-administered electronic questionnaire, which included socio-demographic factors, clinical history, and selected laboratory results.

Blood samples were collected by trained phlebotomists using standard aseptic techniques via venipuncture. A total of 5 mL of whole blood was drawn from each participant. A drop of whole blood was immediately analyzed at the bedside using the Nova StatSensor Xpress Creatinine Point-of-Care Meter (Nova Biomedical, Waltham, MA, USA). The device employs multi-enzyme amperometric biosensor technology to measure creatinine levels in whole blood, with a coefficient of variation (CV) ranging from 4.4% to 5.8% at clinical decision levels [10]. Calibration was factory-encoded in the reagent strip, and trained technicians performed all measurements according to the manufacturer’s instructions. The technicians were blinded to the results of the reference method.

The remaining blood sample was placed in a non-heparinized serum-separating tube (SST), securely sealed to minimize air exposure, and stored in a cool box. Samples were transported to the central laboratory within one hour of collection and centrifuged at 3000G. Serum creatinine was measured using the Jaffe method on the Cobas Integra 400 Plus analyzer (Roche Diagnostics), which is traceable to isotope dilution mass spectrometry (IDMS). The assay demonstrated coefficients of variation (CVs) of 2.2% at 88 µmol/L and 1.8% at 482 µmol/L, based on internal quality control data. The laboratory participates in an External Quality Assurance (EQA) program coordinated by, and accredited to, the National Accreditation Board for Testing and Calibration Laboratories (NABL) and the College of American Pathologists (CAP).

Statistical analysis

All analyses were conducted using STATA 17. Descriptive statistics included measures of central tendency (mean, median) and dispersion (standard deviation, quartiles, and interquartile range). StatSensor creatinine values were compared with laboratory-based Jaffe creatinine measurements using ordinary least squares linear regression and Bland–Altman analysis. The regression analysis included estimation of the slope, intercept, and coefficient of determination (R²), with the laboratory Jaffe method serving as the reference. In the Bland–Altman analysis, the mean bias and 95% limits of agreement were calculated, and comparisons were assessed against the allowable total error (TEa) margin of ±15%, as recommended by Clinical Laboratory Standards Institute (CLSI). The diagnostic performance of the StatSensor was further evaluated through sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV), calculated using two-by-two contingency tables with the reference method as the gold standard. Receiver Operating Characteristic (ROC) curve analysis was performed to determine overall diagnostic accuracy, with the Area Under the Curve (AUC) reported. Likelihood ratios and diagnostic odds ratios were also calculated. Pearson’s correlation coefficient was used to assess linear relationships between continuous creatinine values, and simple linear regression was used to evaluate the association between StatSensor-derived eGFR and laboratory-measured creatinine. A p-value <0.05 was considered statistically significant.

Results

A total of 358 participants were included in the study. The median age was 48 years (IQR: 39–54), with 56.9% of participants under the age of 50, and 66.2% were female. The median duration of HIV was 132 months (IQR: 84–180), with most participants on a tenofovir disoproxil fumarate (TDF)-based regimen (91.1%). A total of 15.6% had reduced eGFR (<60 mL/min/1.73m²), and the median estimated eGFR using the Jaffe method was 87.0 mL/min/1.73m² (IQR: 68.8–104.0), and 80.0 mL/min/1.73m² (IQR: 65.0, 101.4) by Statsensor method (Table 1).

Table 1. Socio-demographic and clinical characteristics of the study participants, n = 358.

Variable Frequency (n) Percent (%)
Age group (years)
 < 50 204 56.9
 ≥ 50 154 43.1
Median age in years (IQR) 48 (39, 54)
Gender
 Male 121 33.8
 Female 237 66.2
Median duration of HIV in months (IQR) 132 (84, 180)
Type of ART
 TDF Based Regimen 326 91.1
 Non TDF Based Regimen 32 8.9
Median creatinine (Jaffe method) (IQR) µmol/L 80.0 68.0 to 94.75
Median creatinine (Statsensor method) (IQR) µmol/L 85.5 (71.0 to 101.0)
Estimated eGFR (Jaffe Method)
 < 60 56 15.6
 ≥ 60 302 84.4
Median standard eGFR (Jaffe method) (IQR) 87.0 (68.8, 104.0)
Median and IQR eGFR (Statsensor) 80.0 (65.0, 101.4)

TDF-Tenofovir Disoproxil Fumarate, eGFR-Estimated glomerular filtration rate, ART-Antiretroviral therapy

Among the 358 participants, the Stat-Sensor correctly identified 52 out of 56 participants with low eGFR, while 4 were misclassified as normal. Additionally, 286 out of 302 participants were correctly classified as having normal eGFR, whereas 16 were misclassified as having low eGFR (Table 2)

Table 2. Diagnostic Comparison of StatSensor eGFR versus standard creatinine eGFR.

Standard creatinine estimated GFR
StatSensor estimated GFR Low (< 60) Normal (≥ 60) Total
 Low (< 60) 52 16 68
 Normal (≥ 60) 4 286 290
Total 56 302 358

The StatSensor demonstrated high diagnostic performance, with a sensitivity of 92.9% (95% CI: 82.7–98.0), specificity of 94.7% (95% CI: 91.5–96.9), and overall accuracy of 94.4% (95% CI: 91.5–96.6) (Table 3).

Table 3. Diagnostic performance of StatSensor for detecting renal dysfunction.

95% Confidence interval
Estimate Lower CL Upper CL
Sensitivity (%) 92.9 82.7 98.0
Specificity (%) 94.7 91.5 96.9
Diagnostic accuracy (%) 94.4 91.5 96.6
Positive predictive value (%) 76.5 66.7 84.0
Negative predictive value (%) 98.6 96.5 99.5
Positive diagnostic likelihood ratio 17.5 10.8 28.4
Negative diagnostic likelihood ratio 0.08 0.03 0.19
Diagnostic odds ratio 232.4 74.7 722.8

The area under the ROC curve (AUC) is 0.938 with standard error of 0.0213, 95% CI (0.896–0.979) p-value <0.001 (Fig 1). The agreement between Stat-Sensor eGFR and Jaffe method based eGFR showed a Kappa value of 0.805 (SE: 0.042, p < 0.001) (Table 4).

Fig 1. ROC curve of the StatSensor in detecting renal dysfunction.

Fig 1

Table 4. Measure of agreement (Kappa) between StatSensor and creatinine by jaffe method.

Symmetric measures
Value Asymptotic standard errora Approximate Tb Approximate significance
Measure of Agreement Kappa .805 .042 15.342 < 0.001
N of Valid Cases 358

a. Not assuming the null hypothesis.

b. Using the asymptotic standard error assuming the null hypothesis.

A scatter plot with linear regression demonstrated a strong positive correlation between the two methods, with a slope of 0.903 and an R² of 0.813. (Fig 2). The Bland–Altman analysis showed that StatSensor-derived eGFR had a negative bias of 4.36 mL/min/1.73 m² (95% CI: 3.09–5.64) compared to the reference laboratory Jaffe kinetic creatinine-based eGFR. The 95% limits of agreement ranged from –19.68 to 28.40 mL/min/1.73 m², and the majority of data points fell within the ± 15% TEA margin, supporting acceptable agreement between the two methods (Fig 3)

Fig 2. Linear regression plot of StatSensor eGFR versus Jaffe eGFR with ±15% total allowable error limits.

Fig 2

Fig 3. Bland–Altman plot comparing StatSensor-derived eGFR to laboratory Jaffe method eGFR (mL/min/1.73m²).

Fig 3

Discussion

In this cross-section study of individuals receiving HIV care in Tanzania, we demonstrated the accuracy of the StatSensor compared to the conventional serum creatinine method. The StatSensor showed high accuracy in detecting KD, with strong sensitivity, specificity, and diagnostic agreement compared to standard creatinine testing using Jaffe method, with a slight negative bias in eGFR making it effective in distinguishing individuals with and without KD.

This test has the potential to identify 9 out of 10 people with KD correctly. However, 3 out of 10 positive cases may be misclassified, highlighting the need for confirmatory testing for those classified with KD to ensure accurate diagnosis and avoid unnecessary clinical interventions. Our findings align with studies by Chandrasekar et al., and Dally et al., which assessed the accuracy of the StatSensor and reported sensitivity ranging from 85% to 97% and specificity ranging from 90% to 96% and 95%, respectively [11–13]. These results are notably higher than those reported by Nataatmadja et al., likely due to the smaller sample size and fewer events in their study. These values make it an excellent initial screening tool, particularly in primary healthcare settings with limited laboratory infrastructure [15]. By enabling early detection and timely intervention, the StatSensor has the potential to improve clinical outcomes for at-risk patients, especially in resource-limited settings where access to conventional laboratory testing is restricted.

A negative bias of 4.36 mL/min/1.73 m² in eGFR was observed, indicating a under estimation of kidney function when using the StatSensor. This finding is consistent with the study by Kosack et al., which also reported a negative mean eGFR bias when comparing the StatSensor eGFR to laboratory-based eGFR derived from enzymatically measured creatinine [10]. Interestingly, the bias appeared to vary across the eGFR range. At lower eGFR values (<60 mL/min/1.73 m²), there was a mild positive bias, contributing to the misclassification of a few normal cases as abnormal. Conversely, at higher eGFR values, the StatSensor tended to underestimate eGFR, consistent with the overall negative bias, leading to some misclassification of normal as reduced kidney function. These patterns underscore the need to interpret point-of-care results cautiously, especially near diagnostic thresholds. While a mild under-estimation of eGFR is unlikely to cause significant under-diagnosis of KD or delay interventions for patients at risk of CKD progression, it may still be relevant for the misclassification of CKD stages, particularly near diagnostic cutoffs. However, R² alone (0.813) does not fully demonstrate method agreement, as it reflects correlation but not accuracy. To address this, we evaluated the agreement using Bland–Altman analysis and Total Allowable Error (TEa) limits. Over 95% of StatSensor-derived eGFR results fell within the CLSI-recommended ±15% TEa margin, indicating clinically acceptable agreement. Despite the mild bias, the StatSensor demonstrated strong diagnostic performance, similar to previous studies reinforcing its reliability [11,16]. Its ability to provide rapid, point-of-care results makes it particularly valuable for early detection in resource-limited settings where laboratory-based creatinine testing is not always available.

The World Health Organization (WHO) has outlined essential criteria for ideal POC diagnostics in resource-limited settings, summarized by the acronym ASSURED, Affordable, Sensitive, Specific, User-friendly, Rapid and Robust, Equipment-free, and Deliverable to end users [12]. The StatSensor Xpress Creatinine test aligns well with most of these principles, it is portable, easy to operate with minimal training, does not require refrigeration or ancillary laboratory infrastructure, and produces rapid results [11,10].. These diagnostic characteristics, along with its simplicity, make it suitable for use even by non-clinical personnel trained in device handling, thus expanding its applicability in decentralized and primary care settings where access to laboratory services is limited.

Routine and frequent kidney function testing is core part of recommended HIV care. Screening facilitates early detection of KD and enables timely interventions to prevent or slow the progression of disease as well as adjustment of potentially nephrotoxic medication. The StatSensor Xpress Creatinine Meter offers a promising alternative to traditional laboratory-based screening methods, particularly in resource-limited settings where access to conventional testing is limited [9]. Additionally, the device’s portability and minimal infrastructure requirements make it particularly suitable for decentralized healthcare settings, ensuring that KD screening is more accessible to people living with HIV in low-resource environments [9,11].

A major strength of this study is its pragmatic design, conducted within a busy HIV clinic setting in Tanzania, reflecting the real-world feasibility and performance of the StatSensor POC test. The study employed rigorous diagnostic accuracy methods, including blinding and standardized comparisons with the Jaffe method. However, several limitations should be noted. First, the study did not use measured glomerular filtration rate (mGFR), the gold standard for kidney function assessment, which may have introduced classification bias. Second, although the CKD-EPI 2021 equation is widely used, its accuracy in sub-Saharan African populations has been questioned, potentially affecting the precision of eGFR estimates in this cohort. Third, while the StatSensor performed well overall, confirmatory testing may still be warranted near diagnostic thresholds to avoid misclassification. Finally, the study did not evaluate agreement across individual KDIGO eGFR stages, which may have provided additional insights into stage-specific classification discrepancies.

Conclusion

The StatSensor Xpress Creatinine Meter demonstrated high diagnostic accuracy and a strong correlation with conventional testing, making it a valuable screening tool for KD among PLHIV. Integrating StatSensor into routine HIV care programs could improve patient outcomes by enabling clinicians to adjust ART, implement reno-protective measures, and avoid nephrotoxic drugs. To ensure its effective implementation, further cost-effectiveness analyses are needed to evaluate the feasibility of scaling up its use in national HIV programs.

Supporting information

S1 Checklist. A completed STROBE checklist.

(DOC)

pone.0331969.s001.doc (79KB, doc)
S2 Data. Dataset of hemodialysis patients.

(XLSX)

pone.0331969.s002.xlsx (52.3KB, xlsx)

Data Availability

All relevant data are within the paper and its Supporting Information files.

Funding Statement

This project was funded by the Fogarty International Center of the National Institutes of Health (NIH) Award Number D43TW009343 and the University of California Global Health Institute (UCGHI). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1.Mwemezi O, Ruggajo P, Mngumi J, Furia FF. Renal dysfunction among hiv-infected patients on antiretroviral therapy in Dar es Salaam, Tanzania: a cross-sectional study. Int J Nephrol. 2020;2020:8378947. doi: 10.1155/2020/8378947 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Ottaru TA, Kwesigabo GP, Butt Z, Caputo M, Chillo P, Siril H, et al. Prevalence and risk factors for renal insufficiency among adults living with HIV in Tanzania: results from a cross-sectional study in 2020-2021. Healthcare (Basel). 2024;12(6):657. doi: 10.3390/healthcare12060657 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Coetzee L, Bogler L, De Neve J-W, Geldsetzer P, Vollmer S, Author C. HIV, antiretroviral therapy and non-communicable diseases in sub-Saharan Africa: empirical evidence from 44 countries over the period 2000 to 2016. J Int AIDS Soc. 2019. doi: 10.1002/jia2.25364 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Crowley T, Stellenberg EL. Integrating HIV care and treatment into primary healthcare: are clinics equipped?. Afr J Prim Health Care Fam Med. 2014;6(1):E1-7. doi: 10.4102/phcfm.v6i1.616 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Rugakingira AA, Yondu E, Thobias JM, Dionis I, Kamata CC, Kilonzi M, et al. Opportunities and challenges for the integration of managing non-communicable diseases within HIV care and treatment services in Tanzania. HIV Res Clin Pract. 2024;25(1):2398869. doi: 10.1080/25787489.2024.2398869 [DOI] [PubMed] [Google Scholar]
  • 6.Swanepoel CR, Atta MG, D’Agati VD, Estrella MM, Fogo AB, Naicker S, et al. Kidney disease in the setting of HIV infection: conclusions from a kidney disease: improving global outcomes (KDIGO) controversies conference. Kidney International. Elsevier B.V.; 2018. 545–59. doi: 10.1016/j.kint.2017.11.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Manaye GA, Abateneh DD, Niguse W. Chronic kidney disease and associated factors among HIV/AIDS patients on HAART in Ethiopia. HIV AIDS (Auckl). 2020;12:591–9. doi: 10.2147/HIV.S276356 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Menez S, Hanouneh M, McMahon BA, Fine DM, Atta MG. Pharmacotherapy and treatment options for HIV-associated nephropathy. Expert Opin Pharmacother. 2018;19(1):39–48. doi: 10.1080/14656566.2017.1416099 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Msilanga D, Muiru A, Balandya E, Liu K. Point of care creatinine testing for early detection of renal dysfunction in Tanzanian HIV patients: a study protocol. BMC Nephrol. 2024;25(1):419. doi: 10.1186/s12882-024-03840-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Kosack CS, de Kieviet W, Bayrak K, Milovic A, Page AL. Evaluation of the Nova StatSensor® Xpress(TM) creatinine point-of-care handheld analyzer. PLoS One. 2015;10(4):e0122433. doi: 10.1371/journal.pone.0122433 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Nataatmadja M, Fung AWS, Jacobson B, Ferera J, Bernstein E, Komenda P, et al. Performance of StatSensor point-of-care device for measuring creatinine in patients with chronic kidney disease and postkidney transplantation. Can J Kidney Health Dis. 2020;7:2054358120970716. doi: 10.1177/2054358120970716 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Dally M, Amador JJ, Butler-Dawson J, Lopez-Pilarte D, Gero A, Krisher L, et al. Point-of-care testing in chronic kidney disease of non-traditional origin: considerations for clinical, epidemiological, and health surveillance research and practice. Ann Glob Health. 2023;89(1):7. doi: 10.5334/aogh.3884 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Ilieva K, Landgraf K, Boxhammer R, Steidl S, Flesher D. Wcn25-744 felzartamab selectively and potently targets cd38 antibody secreting cells from patients with immune-mediated kidney diseases.
  • 14.Snaith B, Harris MA, Shinkins B, Jordaan M, Messenger M, Lewington A. Point-of-care creatinine testing for kidney function measurement prior to contrast-enhanced diagnostic imaging: evaluation of the performance of three systems for clinical utility. Clin Chem Lab Med. 2018;56(8):1269–76. doi: 10.1515/cclm-2018-0128 [DOI] [PubMed] [Google Scholar]
  • 15.Maxim LD, Niebo R, Utell MJ. Screening tests: a review with examples. Inhal Toxicol. 2014;26(13):811–28. doi: 10.3109/08958378.2014.955932 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Korpi-Steiner NL, Williamson EE, Karon BS. Comparison of three whole blood creatinine methods for estimation of glomerular filtration rate before radiographic contrast administration. Am J Clin Pathol. 2009;132(6):920–6. doi: 10.1309/AJCPTE5FEY0VCGOZ [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Mogamat-Yazied Chothia

23 Jun 2025

Dear Dr. Msilanga,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Aug 07 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org . When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols .

We look forward to receiving your revised manuscript.

Kind regards,

Mogamat-Yazied Chothia, MBChB, FCP(SA), MMed, PhD

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please include a complete copy of PLOS’ questionnaire on inclusivity in global research in your revised manuscript. Our policy for research in this area aims to improve transparency in the reporting of research performed outside of researchers’ own country or community. The policy applies to researchers who have travelled to a different country to conduct research, research with Indigenous populations or their lands, and research on cultural artefacts. The questionnaire can also be requested at the journal’s discretion for any other submissions, even if these conditions are not met.  Please find more information on the policy and a link to download a blank copy of the questionnaire here: https://journals.plos.org/plosone/s/best-practices-in-research-reporting. Please upload a completed version of your questionnaire as Supporting Information when you resubmit your manuscript.

3. Thank you for stating the following financial disclosure:

“This project was funded by the Fogarty International Center of the National Institutes of Health (NIH) Award Number D43TW009343 and the University of California Global Health Institute (UCGHI).”

Please state what role the funders took in the study. If the funders had no role, please state: "The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."

If this statement is not correct you must amend it as needed.

Please include this amended Role of Funder statement in your cover letter; we will change the online submission form on your behalf.

4. Please include a separate caption for each figure in your manuscript.

5. We note you have included a table to which you do not refer in the text of your manuscript. Please ensure that you refer to Table 5 in your text; if accepted, production will need this reference to link the reader to the Table.

6. Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information.

7. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

Reviewer #1: Partly

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #1: No

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #1: Yes

Reviewer #2: Yes

**********

Reviewer #1: Thank you to the authors for performing this study. There is a need to confirm performance of point of care tests available. Please address the following concerns.

1. Grammatical error:

a. title should rather read: Point of care creatinine demonstrates high accuracy in detecting kidney dysfunction among people living with HIV in Tanzania

b. Line 58: the sentence does not appear to be complete – a word has been left out after “accessible”?

c. Line 61: “and” should be removed

2. Under the Methods section, the following should be provided:

a. Instrument information required:

i. Methodology in use in the Nova StatSensor Xpress Creatinine Point-of-care meter

ii. Coefficient of variation data at the medical decision limits for the Stat Sensor creatinine measurement

iii. Automated instrument name and model, as well as the correct methodology used for creatinine measurement in the laboratory should be mentioned. The authors mention that the “enzymatic Jaffe method” was used. The method can be either enzymatic or Jaffe (these are two separate methods).

iv. It is also important to mention whether the method is isotope dilution mass spectrometry traceable.

v. Coefficient of variation data at the medical decision limits for the automated creatinine method should also be mentioned.

b. Tube information

i. – “non-heparinized tube” was mentioned. If possible, the authors should be more specific e.g. SST tubes were used

ii. Sample handling information should be summarised e.g. samples were transported to the laboratory and centrifuged at 3000G within 2 hours.

3. Statistical analysis and results

a. As a creatinine analyser is being evaluated, the creatinine data should also be described.

i. In the form of a method comparison using ordinary linear regression and/or difference plots to demonstrate the difference in creatinine results between the methods. Allowable error limits should be included (these can be obtained from EFLM, Westgard or RCPA websites). Remember to include full name of creatinine methods on each axis, as well as correct units used.

b. The eGFR comparison difference plot: the units have not been included along the X or Y-axes. For the X-axis, the laboratory creatinine based eGFR should be used (and not the average eGFR), as this method is being considered the gold-standard method. It is also more useful to demonstrate the allowable error limits on the graphs instead of the 95% distribution limits.

c. eGFR method comparison should ideally be presented before the difference plot graph. Here, as well, please add units on both X and Y axes, as well as the allowable error limits on the graph. The laboratory-based creatinine method should be named on the graph (including the instrument name).

d. Graph legends need to be more descriptive of the findings.

e. Under the results section:

i. Table 1 should omit information not relevant to the study e.g. marital status, level of education and employment status, so that relevant data is easier to read

ii. Lines 123 and 124 are already in table 1 and therefore should be omitted in the paragraph section.

iii. Table 2 data: To be able to get a better idea of the performance of the eGFR results, it would be useful to also compare the KDIGO stages between the two methods in a separate table and then assess % agreement.

iv. Table 3 should be omitted if also being described in lines 135 – 141.

v. In line 153, the positive bias of 4.36 is mentioned. Please add units. This data is more appropriately assessed using total allowable error limits (rather than the limits of agreement alone) – refer to the eGFR difference plot comment.

f. Try to present data in a logical order:

i. Descriptive data

ii. Method comparison and diff plot for creatinine findings

iii. Method comparison and diff plot for eGFR findings

iv. Agreement data (incl kappa), ROC and diagnostic accuracy findings

4. Discussion:

a. Line 173: be clear when discussing the positive bias, whether referring to creatinine or eGFR.

b. Line 186: include the units for eGFR

c. Line 188: This is contradictory to what has been said earlier in the paper. A positive bias for creatinine measurement will mean a negative bias for eGFR measurement.

d. Line 193: Using the R2 value is not enough to confirm good performance between methods. An R2 of 0.813 is not considered a very good correlation. It would be better to demonstrate that the majority of results fall within total allowable error limits (see point above regarding demonstrating the allowable error limits on the graphs)

e. Lines 193 and 218: The authors use the “Jaffe” and “enzymatic” methods interchangeably. This is not correct. Please confirm the exact methodology in use in the laboratory.

This study shows valuable information. Looking forward to seeing the revised version.

Reviewer #2: The Article will contribute to a better understanding of POCT in a resource limited setting

Throughout the article the authors have used the method "enzymatic Jaffe method" for creatinine. There are 2 main methods for creatinine. it is the Jaffe kinetic method or the Enzymatic creatinine method. Please confirm the correct method

Under Methods. Please include the laboratory platform, The CVs of the method used in the laboratory and please add if the lab is engaged in an External Quality assurance programme

Under Results. Please also include the difference plot (Bland Altman) for the 2 creatinine methods. Please ensure that you recognize the hierarchy of methods and label the plot appropriately

The Bland Altman plot for the eGFR needs labelling

Under Discussion: line 173 compared to standard creatinine testing, with a positive bias....

Line 186.... a positive bias of 4.36 eGFR

Please elaborate as to how a positive bias of creatinine is associated with a higher eGFR? Please review

**********

what does this mean? ). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy

Reviewer #1: No

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/ . PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org

PLoS One. 2025 Sep 11;20(9):e0331969. doi: 10.1371/journal.pone.0331969.r002

Author response to Decision Letter 1


24 Jun 2025

We confirm that all manuscript and figure files meet PLOS ONE’s formatting and naming requirements.

The PLOS questionnaire on inclusivity in global research has been attached in the supplementary materials submitted with the manuscript.

The statement "The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript" has been added to the manuscript.

Separate captions for each figure have now been included in the manuscript.

The required information has been incorporated into the manuscript.

The reference list has been thoroughly reviewed and updated to ensure completeness and accuracy.

Decision Letter 1

Mogamat-Yazied Chothia

22 Jul 2025

Dear Dr. Msilanga,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Sep 05 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org . When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols .

We look forward to receiving your revised manuscript.

Kind regards,

Mogamat-Yazied Chothia, MBChB, FCP(SA), MMed, PhD

Academic Editor

PLOS ONE

Journal Requirements:

1. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise. 

2. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

Reviewer #1: (No Response)

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions??>

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #1: Yes

Reviewer #2: Yes

**********

Reviewer #1: Thank you again for performing this useful study and for addressing the concerns raised. Just a few more concerns, before it can be published:

1. Including the creatinine data would add value to this publication (at least the central data for each method, but even better, would be to also include the linear regression equation (y=0.903x+3.82; r2=0.813) with/without graphs). This can be included in the RESULTS section, just after table 1.

2. Other changes required before publication:

a) Legend for Table 3 (line 183/184) still refers to "laboratory enzymatic eGFR". Should be "Jaffe"

b) Line 154, please add the median (IQR) data for the point of care eGFR results either in Table 1 or in the paragraph describing Table 2 data.

c) eGFR graphs:

* Bland-Altman graph: The TAE error limits have been plotted incorrectly. The TAE is a percentage, whereas

the results are plotted in ml/min/1.73m2. See attached example.

* If possible, please add the TAE limits to the ordinary linear regression graph as well. See attached example.

d) Under the RESULTS and DISCUSSION sections:

* Line 174 – 176:The linear regression relationship demonstration should be described as the equation

(y=0.903x+3.82). In the next line, it is incorrectly stated that a positive bias of 4.36 ml/min/1.73m2 is

present – it appears to be a negative bias (which correlates with the linear regression equation).

* Lines 189: eGFR shows small negative bias.

* Line 202: A negative bias was noted, indicating an underestimation of kidney function, and therefore the

overestimation of low kidney function vs normal function (seen in table 1).

The rest of the discussion and conclusions should be adjusted around this small negative bias and over-calling

of decreased kidney function.

e) Consider changing the title to state "Point-of-care creatinine based eGFR (StatSensor) in detecting kidney dysfunction (KD) among people living with HIV in Tanzania"

f) Adjust abstract according to above changes. In the Results section: An example would be: The StatSensor eGFR correlated with Roche Jaffe method-derived eGFR (y=0.903x+3.82; r2=0.813), with a mean negative bias of 4.36 ml/min/1.73m2.

Reviewer #2: I'm happy with the reviewed changes. Just one note. THE bland Altman plot clearly when truncated shows a positive bias for eGFR at lower ends upto 60 ml/min.This explains the 4 normal missclassfication,however at higher eGfrs there is a negative bias explaining the 16 cases of miss classification. This explanation must be highlighted in the results discussion.

**********

what does this mean? ). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy

Reviewer #1: No

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/ . PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org

PLoS One. 2025 Sep 11;20(9):e0331969. doi: 10.1371/journal.pone.0331969.r004

Author response to Decision Letter 2


8 Aug 2025

We sincerely thank the reviewers for their insightful and constructive comments, which significantly helped improve the clarity and rigor of our manuscript.

We have carefully addressed all concerns, including clarifying the bias observed in the Bland–Altman plot, correcting terminologies, and revising the figures and abstract accordingly.

We believe these revisions have strengthened the manuscript and enhanced its value for readers interested in point-of-care kidney function assessment.

Attachment

Submitted filename: Reviewers comments.docx

pone.0331969.s005.docx (16.3KB, docx)

Decision Letter 2

Mogamat-Yazied Chothia

24 Aug 2025

Point-of-care creatinine-based eGFR (StatSensor) in detecting kidney dysfunction (KD) among people living with HIV in Tanzania

PONE-D-25-28343R2

Dear Dr. Daniel Msilanga,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager®  and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support .

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Mogamat-Yazied Chothia, MBChB, FCP(SA), MMed, PhD

Academic Editor

PLOS ONE

Acceptance letter

Mogamat-Yazied Chothia

PONE-D-25-28343R2

PLOS ONE

Dear Dr. Msilanga,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Prof. Mogamat-Yazied Chothia

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Checklist. A completed STROBE checklist.

    (DOC)

    pone.0331969.s001.doc (79KB, doc)
    S2 Data. Dataset of hemodialysis patients.

    (XLSX)

    pone.0331969.s002.xlsx (52.3KB, xlsx)
    Attachment

    Submitted filename: Reviewers comments.docx

    pone.0331969.s005.docx (16.3KB, docx)

    Data Availability Statement

    All relevant data are within the paper and its Supporting Information files.


    Articles from PLOS One are provided here courtesy of PLOS

    RESOURCES