Skip to main content
Insights into Imaging logoLink to Insights into Imaging
. 2025 Sep 18;16:198. doi: 10.1186/s13244-025-02081-8

Technical requirements and optimization strategies for home-based teleradiology workstations: a review article

Mustafa S Alhasan 1,2,, Ayman S Alhasan 1,3
PMCID: PMC12446176  PMID: 40965740

Abstract

Abstract

Teleradiology has advanced from an occasional modality to a cornerstone of modern radiology practice, with the COVID-19 pandemic catalyzing widespread adoption of home-based reading environments. This review synthesizes current literature and expert recommendations on hardware and software optimization for effective home-based teleradiology implementation. Available data indicate 65% of institutions established home workstations during the pandemic, with 74% transitioning routine daytime shifts to internal teleradiology. We reviewed key components of successful remote reading environments, including diagnostic display specifications, environmental controls, ergonomic considerations, computational infrastructure, and network architecture. Evidence suggests that properly configured remote workstations maintain diagnostic performance equivalent to hospital reading rooms while enhancing radiologist satisfaction and productivity. We found that 65% of radiologists reported reduced stress levels when working from home, and 96% observed similar or improved report turnaround times. Essential technical specifications include display luminance standards, ambient lighting levels between 25 and 75 lux, acoustic conditions below 40 decibels, and temperature control within 20–24 °C. Computational requirements include a minimum sustained bandwidth of 50–100 Mbps. We review multi-layered security architectures and workflow integration strategies supporting distributed reading environments. Our review concludes that properly implemented home-based teleradiology is a viable practice model extending specialized expertise across geographic boundaries while promoting radiologist well-being. However, knowledge gaps remain in technical standardization, regulatory harmonization, and long-term clinical outcomes, underscoring the need for further research to support confident, data-driven teleradiology implementation.

Critical relevance statement

This review critically evaluates the technical, ergonomic, and operational requirements for home-based teleradiology, offering evidence-based recommendations that address current practice gaps and support the development of sustainable, high-performance remote reading environments in modern clinical radiology.

Key Points

  • Home teleradiology maintains diagnostic quality while improving radiologist well-being; 65% report reduced stress and 96% show similar or improved report turnaround times.

  • Optimal implementation requires medical-grade displays, a controlled environment (25–75 lux lighting), 50–100 Mbps bandwidth, and robust security measures.

  • Standardization varies across jurisdictions; some countries have protocols, but gaps persist in cross-border teleradiology and long-term outcomes assessment.

Graphical Abstract

graphic file with name 13244_2025_2081_Figa_HTML.jpg

Keywords: Teleradiology, Telemedicine, Radiology practice, Computational radiology

Introduction

Teleradiology has evolved from addressing after-hours coverage gaps to becoming a cornerstone of modern radiology practice, with 85.6% of U.S. radiologists reporting engagement in remote reading by 2019 [1].

The COVID-19 pandemic significantly accelerated the adoption of teleradiology, transforming it from a primarily after-hours or supplementary service into a mainstream radiology practice model (Fig. 1). In response to social distancing measures, healthcare institutions rapidly implemented home workstations to ensure continuity of diagnostic services while safeguarding staff [2]. This abrupt shift exposed both the strengths and limitations of existing teleradiology infrastructure. Academic medical centers and private practices alike faced technical, regulatory, and operational challenges in establishing home-based reading environments capable of maintaining diagnostic quality while promoting radiologist well-being. Survey data reflect this paradigm shift: approximately 65% of institutions deployed home workstations, and 74% transitioned routine daytime shifts to internal teleradiology during the pandemic. Notably, over half of radiologists continued remote work practices even after the easing of acute pandemic restrictions, indicating a sustained change in radiology practice patterns [3].

Fig. 1.

Fig. 1

Teleradiology workflow architecture. Complete workflow showing: Imaging Facility (green) with modalities, PACS, and on-site reading; Teleradiology Infrastructure (blue) providing secure connectivity, worklist management, and reporting platforms; Home Reading Environment (orange) with diagnostic displays and workstation hardware; and Quality Assurance framework (red) with peer review and performance monitoring. Arrows indicate workflow direction and feedback loops

The sustainability of the shift toward hybrid and home-based practice models hinges on the effective optimization of remote workstations. Unlike the controlled conditions of hospital reading rooms, home-based setups present unique challenges related to hardware specifications, software configurations, network reliability, ergonomic design, and quality assurance protocols. In response, professional organizations have issued guidelines tailored specifically to home-based teleradiology. Notably, the American College of Radiology (ACR), American Association of Physicists in Medicine (AAPM), and Society for Imaging Informatics in Medicine (SIIM) jointly released the Electronic Practice Standard in 2022, outlining requirements for display systems, computing hardware, and security infrastructure appropriate for remote diagnostic interpretation [4]. Similarly, the Royal College of Radiologists (RCR) and the Turkish Society of Radiology (TSR) have published technical and operational recommendations to support high-quality home-based radiology reporting [5, 6].

The widespread adoption of remote reading has yielded substantial data on both technical performance and radiologist experience. Empirical evidence indicates that well-configured home workstations can maintain diagnostic accuracy while enhancing radiologist satisfaction and productivity. Approximately 65% of radiologists report reduced stress levels when working from home, and 96% note similar or improved report turnaround times [7]. Nonetheless, notable challenges persist, particularly in the areas of trainee education, team cohesion, and quality assurance in distributed reading environments. The educational impact is especially concerning, with 52% of trainees reporting diminished opportunities for case review and real-time feedback during remote instruction [8]. In this review article, we aim to synthesize current literature and expert perspectives to provide evidence-based guidance for optimizing hardware and software in home-based teleradiology workstations.

Display technology and image interpretation environment

Display technology comparison

The diagnostic display is the critical interface between radiological data and the interpreting radiologist, directly influencing detection sensitivity and diagnostic confidence. Medical-grade monitors, specifically engineered for radiologic interpretation, differ markedly from consumer-grade displays in several key dimensions. One of the most important distinctions is compliance with the DICOM Part 14 Grayscale Standard Display Function (GSDF), which ensures accurate grayscale rendering and consistent contrast perception across varying luminance levels [5, 912]. Medical-grade displays also maintain strict luminance uniformity, typically with less than 30% variation across the screen, compared to potential variations exceeding 50% in consumer monitors. Additionally, medical displays are designed for long-term calibration stability, with premium models retaining accuracy over thousands of operating hours, whereas consumer displays may exhibit significant drift after just a few hundred hours. These technical differences account for the substantial price gap: medical-grade displays generally range from $5000 to $15,000 per unit, compared to $500–$2000 for high-end consumer monitors. This cost differential underscores why professional guidelines consistently discourage the use of general-purpose displays for primary diagnosis in teleradiology settings [13, 14].

Liquid crystal displays (LCD) with light-emitting diode (LED) backlighting remain the standard for diagnostic imaging, offering proven performance and established quality assurance protocols. [1517]. While some medical display manufacturers have begun incorporating OLED and mini-LED into diagnostic-grade monitors, professional organizations have adopted a cautious stance, awaiting further validation studies to confirm their equivalence, or superiority, to standard LCD systems for primary diagnostic use (Table 1).

Table 1.

Essential components of a home-based teleradiology workstation

Component category General radiography/CT/MRI Mammography Nuclear medicine/fusion imaging Minimum specifications (all modalities)
Display
 Resolution ≥ 3 MP (2048 × 1536) ≥ 5 MP (2560 × 2048) ≥ 3 MP with color calibration As specified by modality
 Luminance ≥ 350 cd/m² ≥ 420 cd/m² ≥ 350 cd/m² Stable during operational lifetime
 Contrast ratio ≥ 350:1 ≥ 450:1 ≥ 350:1 Measurable using AAPM TG18 test patterns
 Pixel pitch ≤ 0.21 mm ≤ 0.17 mm ≤ 0.21 mm Uniform across display surface
 Calibration DICOM GSDF DICOM GSDF DICOM GSDF + color (sRGB) Initial and quarterly verification
 Configuration

Minimum total display area: 6–10 MP effective viewing

Options: (a) Dual 3–5 MP displays, OR (b) Single large display (≥ 30”, ≥ 6MP)

Selection based on workflow, space, and user preference

Same options, with ≥ 5 MP primary for mammography Same options + color-capable display for fusion imaging Plus 1 administrative display for all configurations
Computing hardware
 Processor Multi-core CPU (≥ 8 cores, ≥ 3.0 GHz) Same Same plus specialized processing Intel i7/i9 or AMD Ryzen 7/9 equivalent
 Memory ≥ 32 GB RAM ≥ 32 GB RAM ≥ 64 GB RAM DDR4-3200 or faster
 Storage Primary: NVMe SSD ≥ 1 TB Secondary: ≥ 2 TB for local caching Same Same plus expanded cache ≥ 4 TB for local caching

≥ 3000 MB/s read speed

Encrypted at rest

 Graphics Professional GPU, ≥ 8 GB VRAM Same Professional GPU, ≥ 16 GB VRAM NVIDIA RTX/Quadro or AMD Radeon Pro
Network
 Download bandwidth ≥ 100 Mbps sustained ≥ 150 Mbps sustained ≥ 200 Mbps sustained Business-class service with SLA
 Upload bandwidth ≥ 20 Mbps sustained ≥ 30 Mbps sustained ≥ 50 Mbps sustained Verified through regular testing
 Latency ≤ 50 ms to primary systems Same Same ≤ 15 ms jitter
 Reliability ≥ 99.9% uptime Same Same Redundant connectivity recommended
 Security VPN with AES-256 encryption Same Same Split-tunnel configuration
Environment
 Ambient lighting 25–75 lux at display surface 20–40 lux 25–75 lux with controlled reflections Indirect, adjustable intensity
 Sound level ≤ 40 dB ambient noise Same Same Sound isolation from household
 Temperature 20–24 °C (68–75 °F) Same Same Stable during reading sessions
 Ergonomics

ANSI/BIFMA compliant chair

Monitor height at/below eye level

Same Same

Height-adjustable desk

Adjustable monitor arms

Power protection
 UPS ≥ 15 min runtime at full load Same Same Pure sine wave output
 Surge protection ≥ 1500 joule rating Same Same Equipment insurance
Peripherals
 Input devices

Ergonomic keyboard and mouse

Optional: programmable keypad

Same Same plus specialized input

Programmable function keys

Customizable macros

 Dictation

Noise-canceling microphone

Optional: foot pedal

Same Same

≥ 16-bit/48 kHz audio quality

Programmable control

Display configuration should prioritize total effective viewing area and workflow efficiency. Both dual-monitor and single large monitor configurations demonstrate equivalent diagnostic performance when meeting resolution and calibration requirements (ACR Technical Standard). Selection factors include available space, workflow patterns, radiologist preference, and institutional standardization needs

MP megapixel, cd/m² Candela per square meter, DICOM GSDF Digital Imaging and Communications in Medicine Grayscale Standard Display Function, SLA service level agreement, UPS uninterruptible power supply, AAPM American Association of Physicists in Medicine, NVMe non-volatile memory express, SSD solid state drive, VRAM video random access memory, AES advanced encryption standard, ANSI/BIFMA American National Standards Institute/Business and Institutional Furniture Manufacturer’s Association, VPN virtual private network, GPU graphics processing unit, CPU central processing unit, RAM random access memory, dB decibel, TB terabyte, GB gigabyte, GHz gigahertz, Mbps megabits per second, ms milliseconds

Luminance standards and modality-specific requirements

Luminance specifications for diagnostic displays vary based on imaging modality, reflecting the distinct visual demands of different types of radiologic interpretation. For general radiography, computed tomography (CT), and magnetic resonance imaging (MRI), displays are typically required to maintain a calibrated luminance of at least 350 cd/m², with a minimum black level below 1 cd/m² and a contrast ratio exceeding 350:1. Mammographic interpretation necessitates more stringent performance criteria due to the need for detecting microcalcifications and subtle architectural distortions. Current guidelines recommend a minimum luminance of 420 cd/m² and a display resolution of at least 5 megapixels for mammography [18]. Additionally, nuclear medicine and certain advanced visualization applications benefit from color-capable displays with accurate color rendering, provided they also meet luminance and grayscale standards appropriate for diagnostic interpretation [19, 20].

In home-based teleradiology, modality-specific imaging requirements must be directly reflected in equipment selection and validation protocols. The Royal College of Radiologists (RCR) recommends that remote workstations replicate the display specifications used in on-site environments for equivalent imaging modalities. This guidance emphasizes that home-reporting monitors should not only match hospital-based systems in screen size or appearance, but also in technical parameters such as resolution, luminance, contrast ratio, and calibration status. Similarly, the Turkish Society of Radiology (TSR) advises that home workstations be equipped with displays offering a resolution of at least three megapixels for general radiologic interpretation, and 5 megapixels or higher for mammography, thereby maintaining diagnostic consistency with institutional standards [5, 6].

Ambient environment control

The interpretation environment surrounding diagnostic displays significantly influences visual perception and diagnostic accuracy. Among environmental variables, ambient lighting is paramount; excessive or improperly directed light can degrade contrast perception through screen glare and pupillary constriction. Professional guidelines recommend maintaining ambient lighting between 25–75 lux at the display surface, substantially lower than typical residential lighting levels, which often exceed 300 lux. This necessitates targeted workspace modifications for home-based teleradiology, such as room-darkening window treatments, indirect lighting fixtures, and matte-finished surfaces to minimize reflective artifacts [2123].

In addition to lighting, acoustic conditions also affect interpretive performance. Background noise levels below 40 decibels are advised to reduce cognitive distraction and promote sustained focus, particularly during complex case evaluations. Achieving this in a home setting may require sound insulation, dedicated quiet spaces isolated from household activity, or the use of active noise-canceling technologies. Temperature regulation further contributes to optimal cognitive function, with evidence suggesting peak mental performance in environments maintained between 20 and 24 °C (68–75 °F) [24, 25]. Collectively, these environmental elements form an interpretation ecosystem that can either support or undermine diagnostic accuracy, regardless of the quality of the display hardware itself.

Quality assurance and calibration requirements

Diagnostic displays require quality assurance protocols to maintain consistent performance in distributed teleradiology environments (Table 2). Daily automated calibration using internal photometers ensures DICOM GSDF compliance, while weekly manual verification with external photometers validates luminance stability and contrast ratios. Monthly assessments should evaluate luminance uniformity, with deviations not exceeding 30% across the display surface.

Table 2.

Teleradiology quality assurance framework

QA component Requirements Measurement method Implementation frequency Remote-specific considerations
Technical quality assurance
 Display calibration DICOM GSDF compliance; Luminance 350–450 cd/m²; Contrast ratio ≥ 350:1 AAPM TG18 test patterns; Luminance uniformity measurement Initial validation plus quarterly recalibration Remote calibration verification; Home environment lighting assessment; Remote monitoring software
 Network performance Sustained bandwidth meeting modality requirements; Latency < 50 ms; Uptime > 99.9% Active monitoring of throughput, latency, packet loss; Periodic stress testing Continuous monitoring with weekly reporting ISP service level guarantees; Redundant connection availability; Automated alert systems
 Workstation validation Standard enterprise configuration; Performance benchmarking; Security compliance Benchmark suite execution; Configuration compliance scanning Initial certification plus monthly validation Remote management capabilities; Auto-remediation tools; Secure administration channel
Interpretive quality assurance
 Peer review process Random case selection (5–10% of studies); Blinded review methodology; Standardized scoring system RADPEER or equivalent methodology; Subspecialty-appropriate case matching Minimum 5–10 cases per radiologist monthly Digital peer review platform; Cross-location case distribution; Anonymous reviewer identification
 Diagnostic accuracy Reference standard comparison; Surgical/pathological correlation; Follow-up analysis Sensitivity/specificity calculation; Positive predictive value assessment Quarterly analysis of critical findings; Annual comprehensive review Remote access to clinical outcome data; Integrated feedback mechanisms
 Critical result communication Direct communication protocols; Closed-loop documentation; Escalation pathways Time to notification measurement; Documentation completeness audit 100% of critical findings tracking; Monthly compliance audit Remote communication tools; Digital acknowledgment tracking; Multi-channel notification options
Operational quality assurance
 Turnaround time monitoring Modality-specific targets; Priority-based metrics Time-stamped workflow tracking; Modality-specific TAT calculation Continuous monitoring with daily reporting Remote productivity dashboards; Comparative performance metrics; Environmental factor assessment
 Workload distribution Volume-based assignments; Complexity factor adjustment; Subspecialty alignment RVU tracking; Complexity-adjusted productivity measurement Daily workload assessment; Weekly balance review Remote worklist management tools; AI-assisted complexity scoring; Dynamic assignment algorithms
 System availability 99.9% uptime target; Scheduled maintenance windows; Redundancy verification End-to-end availability measurement; Component-specific uptime calculation Continuous monitoring with real-time alerting Remote system monitoring; Distributed availability tracking; Disconnected operation capabilities
Remote-specific quality assurance
 Home environment assessment Private workspace; Lighting control (25–75 lux); Acoustic management (< 40 dB) Standardized assessment checklist; Photographic documentation Initial implementation plus annual recertification Remote workspace certification; Environmental factor verification; Alternative location requirements
 Remote radiologist well-being Virtual engagement strategies; Burnout prevention protocols Standardized well-being surveys; Productivity pattern analysis Quarterly well-being assessment Virtual community development; Isolation mitigation protocols; Structured connection opportunities
 Communication effectiveness Virtual communication platforms; Structured interaction protocols Communication frequency measurement; Satisfaction survey data Monthly assessment Multi-channel communication tools; Virtual presence technologies; Backup communication pathways

AAPM American Association of Physicists in Medicine, AI artificial intelligence, DICOM Digital Imaging and Communications in Medicine, GSDF grayscale standard display function, ISP internet service provider, RADPEER radiology peer review system, RVU relative value unit, TAT turnaround time, TG task group

Acceptance testing must establish baseline performance characteristics, including maximum luminance output (≥ 350 cd/m² for general radiology, ≥ 420 cd/m² for mammography), contrast ratios, and spatial resolution compliance with manufacturer specifications. Previous evidence has demonstrated significant performance variations between monitor technologies, focusing on the importance of standardized QA protocols tailored to specific display characteristics [26]. Environmental monitoring should document ambient lighting levels (25–75 lux) and viewing conditions to ensure the best diagnostic performance.

Regulatory compliance requires extensive and detailed documentation of calibration activities, performance measurements, and corrective actions. For teleradiology applications, remote QA execution may utilize video documentation and electronic monitoring systems to satisfy accreditation requirements while maintaining the flexibility of distributed reading environments.

Ergonomics and human factors engineering

Musculoskeletal disorder prevention

Radiologists are at considerable risk for developing work-related musculoskeletal disorders (MSDs) due to prolonged static postures, repetitive motions, and suboptimal workstation configurations. Studies report that up to 60% of radiologists experience symptoms of musculoskeletal discomfort, most commonly affecting the neck, lower back, and upper extremities [27]. Effective prevention begins with ergonomically designed seating [2830].

Workstation layout is another critical factor. Height-adjustable desks that accommodate both sitting and standing postures have been shown to reduce static spinal loading. Evidence supports alternating between sitting and standing every 30–45 min to maximize musculoskeletal benefit without impairing workflow efficiency or diagnostic accuracy [31, 32]. Monitor positioning has a direct impact on cervical strain. Optimal setup involves placing the top edge of the primary display at or just below eye level and maintaining a viewing distance of approximately 20–40 inches from the user [33]. When using multiple displays, monitors should be arranged to minimize head rotation, typically through curved or angled configurations that preserve consistent viewing distances across all screens (Fig. 2).

Fig. 2.

Fig. 2

Optimized home reading environment illustration

Visual fatigue mitigation strategies

One widely supported intervention is the 20-20-20 rule: looking at an object 20 feet away for 20 s every 20 min. This technique has been shown to alleviate accommodation fatigue and can be effectively integrated into the workflow through scheduled break reminders [34, 35].

High-performance computing infrastructure

Computational requirements analysis

The computing hardware supporting teleradiology workstations must provide consistent, high-performance capabilities across a broad range of imaging modalities and advanced visualization tasks. Hardware requirements vary by modality, with basic radiography and limited CT interpretation demanding relatively modest resources, whereas applications involving multiplanar reconstruction, volumetric rendering, and fusion imaging require substantially greater processing power [36].

Graphical processing units acceleration benefits

Graphics processing units (GPUs) play an increasingly critical role in radiologic interpretation, particularly in the context of advanced visualization and artificial intelligence (AI) applications. Many modern PACS viewers leverage GPU acceleration to enhance performance for resource-intensive tasks such as three-dimensional rendering, maximum intensity projections, and real-time volumetric manipulation. Performance benchmarking has shown that GPU-accelerated rendering can improve frame rates by 30–600% compared to CPU-only processing, with the greatest benefits observed in complex operations like cinematic rendering and volume visualization [37].

System redundancy engineering

Redundancy engineering is essential to mitigate potential points of failure, starting with power supply reliability. Uninterruptible power supply (UPS) systems capable of supporting the entire workstation for at least 15–30 min are considered a minimum standard for teleradiology, enabling graceful shutdown during outages and protecting against power fluctuations commonly encountered in residential settings [38].

Connectivity and network architecture

Bandwidth requirements

Network connectivity is a critical component of teleradiology infrastructure, serving as the primary link between remote radiologists and institutional imaging systems. Bandwidth requirements are influenced by study complexity, workflow intensity, and daily interpretative volume. Download bandwidth directly affects image retrieval speed, with current guidelines recommending a minimum sustained rate of 50–100 Mbps for general teleradiology applications, and over 200 Mbps for high-volume or subspecialty workflows involving large datasets, such as multiphase CT, breast tomosynthesis, or dynamic cardiac MRI [39].

Upload bandwidth demands have also increased, driven by the need for robust bidirectional communication. Sustained upload speeds of 20–50 Mbps are recommended to support smooth video conferencing, screen sharing during interdisciplinary consultations, and timely transmission of finalized reports. Importantly, these figures represent minimum operational thresholds rather than ideal targets. High-performing teleradiology environments often exceed these benchmarks to ensure consistent performance during peak usage periods and to accommodate the growing size and complexity of imaging datasets [3941].

Latency management strategies

While bandwidth metrics quantify data transfer capacity, latency measurements are equally critical for teleradiology performance, particularly in tasks requiring real-time interactivity. Latency, commonly measured as round-trip time between the radiologist’s workstation and institutional systems, directly affects responsiveness during operations such as window/level adjustments, image scrolling, and region-of-interest manipulation. For optimal performance, round-trip latency should ideally remain below 50 milliseconds to ensure fluid and responsive image interaction [42].

Latency management often involves network route optimization, which aims to minimize physical distance and the number of network hops between endpoints. Techniques such as Border Gateway Protocol (BGP) routing controls, traffic engineering, and strategic peering arrangements can reduce latency by 20–40% compared to default internet paths. However, implementing such optimizations typically requires enterprise-level network management capabilities, which may not be feasible for individual practitioners operating from home environments [43].

Network security architectures

Security requirements for teleradiology connections must strike a careful balance between robust protection of patient data and operational efficiency to support timely clinical access. Modern security architectures employ a layered, defense-in-depth approach, leveraging multiple technologies rather than relying on a single point of protection. Virtual Private Network (VPN) solutions, including internet protocol security (IPsec), secure sockets layer/transport layer security, and datagram transport layer security (DTLS) protocols, create encrypted tunnels between teleradiology workstations and institutional systems, safeguarding data in transit across public or unsecured networks [4446].

Authentication frameworks have evolved significantly, moving beyond basic password-based systems to more advanced multi-factor authentication (MFA) models. These frameworks incorporate combinations of knowledge factors (e.g., passwords, PINs), possession factors (e.g., smart cards, hardware tokens, registered mobile devices), and biometric factors (e.g., fingerprint scans, facial recognition, behavioral biometrics). The goal is to enhance security without impairing usability, acknowledging that overly burdensome authentication processes may encourage workarounds that ultimately compromise system integrity [4751].

Software optimization

Distributed architecture models

The technological architecture underlying teleradiology has evolved significantly, from basic remote viewing of centralized image repositories to advanced distributed systems engineered for enhanced performance, scalability, and reliability across diverse environments. Traditional thick-client architectures, in which teleradiologists operate full-featured workstation software connected to centralized PACS databases, continue to provide strong rendering performance and support for advanced visualization. However, these benefits come with increased complexity in terms of deployment, configuration, and ongoing maintenance.

More recently, cloud-native PACS architectures have emerged as a transformative development in distributed imaging platforms. These systems leverage global cloud infrastructure to offer improved accessibility, scalability, and resilience, often outperforming traditional on-premises solutions. Built on microservices architectures, they break down monolithic PACS functionality into modular components that can be independently deployed and dynamically scaled based on workload demands. For teleradiology, this architecture presents distinct advantages, including geographically distributed processing to reduce latency, elastic scaling during peak workloads, and built-in redundancy via multi-region deployments [52, 53].

Prefetching algorithms

Intelligent prefetching mechanisms play a crucial role in enhancing teleradiology performance by proactively retrieving relevant prior studies and related examinations before radiologists explicitly request them. This approach effectively masks network latency by aligning data retrieval with workflow timing. Rule-based prefetching systems apply static criteria, such as anatomical region, imaging modality, and temporal proximity, to identify pertinent comparison studies. These are typically fetched automatically when a new case is added to the radiologist’s worklist, provided they meet predefined relevance thresholds [54].

In teleradiology implementations, the prefetching strategy significantly influences both interpretation efficiency and overall network usage. Aggressive prefetching maximizes study availability and minimizes delays during image interpretation but may result in the retrieval of many studies that are ultimately unused, leading to unnecessary bandwidth consumption. Conversely, conservative prefetching reduces nonessential data transfers but may introduce workflow delays when radiologists request prior studies not previously retrieved. Optimal prefetching strategies must therefore balance performance gains with bandwidth efficiency, particularly in distributed or bandwidth-limited environments [54, 55].

Modern PACS architectures increasingly leverage real-time streaming technologies that have largely superseded traditional prefetching approaches. Progressive JPEG 2000 and adaptive streaming protocols enable dynamic image delivery based on bandwidth availability and user interaction patterns. These streaming implementations provide immediate image access while minimizing bandwidth waste, particularly beneficial for teleradiology, where network resources may be limited. However, hybrid approaches combining intelligent prefetching of high-priority studies with streaming capabilities for on-demand access represent the current state-of-the-art for optimized teleradiology performance [5659].

Voice recognition technology comparison

Modern speech recognition systems achieve over 95% accuracy for radiological dictation [6062]. Cloud-based solutions offer superior accuracy but introduce 200 ms to 500 ms latency compared to local processing [6368].

Structured reporting implementation strategies

In teleradiology applications, structured reporting offers distinct advantages, including improved consistency across distributed interpreters, streamlined quality monitoring, and reduced dictation burden, which may help mitigate speech recognition challenges in variable acoustic environments [69]. Effective implementations prioritize template harmonization across all reading locations to ensure uniform report structure and terminology, regardless of where the interpretation is performed. At the same time, systems must allow for location-specific modifications to accommodate the unique preferences or clinical requirements of particular facilities or service lines [70].

Security and compliance framework

Multi-layered security architecture

Effective teleradiology security necessitates coordinated protection across multiple technical layers, system components, and potential attack vectors, rather than reliance on isolated security measures. Modern security architectures adopt a defense-in-depth strategy, emphasizing redundant safeguards to ensure that the compromise of any single control does not lead to total system exposure. At the foundation of this architecture is physical security, which includes controls that restrict unauthorized access to teleradiology workstations and supporting infrastructure, forming the baseline upon which higher-level security measures are built [71, 72].

System security combines workstation hardening (restricted privileges, endpoint protection) with application-level controls including secure authentication and comprehensive threat protection (Table 3).

Table 3.

Overview of teleradiology security framework

Security domain Implementation requirements Regulatory requirements Risk mitigation strategies Validation methods Implementation priority Implementation timeline Responsible parties Monitoring frequency Cost impact Breach impact
Physical security: workstation environment Dedicated lockable space; Access restrictions; Privacy screens (≥ 60° angle); Environmental monitoring HIPAA §164.310(a)(1); GDPR Art. 32; ISO 27001 A.11 Room access logs; Clear screen policy; Authorized personnel documentation; Clean desk protocols Physical security audit (quarterly); Documented ownership; Photographic setup verification Required Implementation Phase 1 (0–30 days) Radiologist; IT Security Quarterly validation Low High
Physical security: media handling Encrypted media (AES-256); Secure disposal (DoD 5220.22-M); Asset tracking; No unencrypted PHI HIPAA §164.310(d)(1); HITECH §13402; ISO 27001 A.8.3 Media inventory management; Destruction certification; DLP scanning; Media handling policy Destruction verification; Quarterly compliance check; DLP reports Required Implementation Phase 1 (0–30 days) IT Operations; Security Team; Radiologist Monthly scanning; Quarterly audit Low-Medium High
Network security: perimeter protection Enterprise-grade NGFW; IDS/IPS with medical protocols; Network segmentation; Dedicated teleradiology VLAN HIPAA §164.312(e)(1); NIST SP 800-41; ISO 27001 A.13.1 Boundary protection devices; Default-deny rules; Regular rule review; Protocol filtering Vulnerability scanning (monthly); Penetration testing (annual); Configuration review Critical Implementation Phase 1 (0–30 days) Network Security Team; IT Operations Continuous monitoring; Monthly reporting Medium-High Severe
Network security: data transmission TLS 1.3 or IPsec with PFS; AES-256 minimum; Certificate authentication; Automated certificate management HIPAA §164.312(e)(1); FIPS 140-2/3; NIST SP 800-52r2 Protocol downgrade prevention; Certificate monitoring; Cryptographic assessment; Key management Encryption validation; TLS configuration audit; Packet capture analysis Critical Implementation Phase 1 (0–30 days) Network Security Team; Security Architecture Weekly certificate monitoring; Quarterly configuration review Medium Severe
Network security: remote access Split-tunnel VPN (clinical traffic); MFA (NIST AAL2+); Context-aware policies; Detailed connection logging HIPAA §164.312(a)(2)(iv); NIST SP 800-63B; ISO 27001 A.6.2 Connection time limits; Geo-restrictions; Device posture checking; Access alerting VPN configuration audit; Auth log review; Connection pattern analysis Critical Implementation Phase 1 (0–30 days) Identity Team; Network Security Daily log review; Monthly VPN audit Medium-High Severe
System security: endpoint protection EDR/XDR with behavioral analysis; Application allowlisting; Vulnerability management; Privileged access controls HIPAA §164.308(a)(5)(ii)(B); NIST SP 800-53 SI-7; ISO 27001 A.12.2 BYOD prohibition; Centralized management; Admin restriction; Security baselines Bi-weekly vulnerability scanning; Configuration audits; Privilege testing Required Implementation Phase 2 (31–60 days) Endpoint Security Team; IT Operations Daily EDR monitoring; Bi-weekly scans High High
System security: patch management Critical patches < 14 days; Monthly maintenance; Pre-deployment testing; Automated compliance HIPAA §164.308(a)(5)(ii)(B); NIST SP 800-40r4; ISO 27001 A.12.6 Vulnerability prioritization; Exception process; Legacy isolation; Compensating controls Patch compliance reports; Remediation tracking; Exception documentation Required Implementation Phase 2 (31–60 days) IT Operations; Security Team Weekly patch status; Monthly compliance review Medium High
System security: device hardening DISA STIG/CIS benchmarks; Disabled services; Disk encryption; Local firewall HIPAA §164.308(a)(5)(ii)(D); NIST SP 800-70; ISO 27001 A.14.1.3 Custom security baselines; System snapshots; Boot protection; FIPS 140-2 encryption Configuration scans; Baseline audits; Benchmark scoring Required Implementation Phase 2 (31–60 days) System Administrators; Security Engineers Quarterly baseline verification Medium High
Application security: authentication Federated identity (SAML/OIDC); Role-based access; 15–30 min timeouts; Unique identification HIPAA §164.312(a)(1); NIST SP 800-63 C; ISO 27001 A.9.2 Password complexity; Login monitoring; Account lockout; Privileged monitoring Auth log analysis; Credential storage tests; Session review Critical Implementation Phase 1 (0–30 days) Identity Team; Application Security Daily monitoring; Monthly audit Medium-High Severe
Application security: PACS/RIS Current application versions; Vendor security certification; API validation; Third-party assessment HIPAA §164.308(b)(1); FDA Cybersecurity Guidance; ISO 27001 A.14.2 Vendor security questionnaire; Pen-testing approval; Compensating controls Vendor assessment; Vulnerability scanning; Configuration review Critical Implementation Phase 2 (31–60 days) Radiology IT; Vendor Management Quarterly security review High Severe
Application security: mobile devices Container technology; Remote wipe; Camera/microphone controls; Restricted storage HIPAA §164.310(b); NIST SP 800-124; ISO 27001 A.6.2.1 App whitelisting; Device encryption; Mobile DLP; Screen capture prevention Mobile security audit; Container isolation tests; Data leakage testing Recommended Implementation Phase 3 (61–90 days) Mobile Security Team; End User Computing Monthly compliance check Medium Medium-High
Data security: classification PHI identification/tagging; Automated discovery; Minimal necessary principle; Data flow mapping HIPAA §164.502(b); GDPR Art. 5(1)(c); ISO 27001 A.8.2 DLP rules; Data minimization; Privacy by design; De-identification when possible Classification audit; Flow mapping; DLP testing; Accuracy validation Required Implementation Phase 2 (31–60 days) Data Security Team; Privacy Officer Quarterly assessment Medium High
Data security: protection Database encryption (TDE/field); Key management; Data masking; DLP for all states HIPAA §164.312(a)(2)(iv); GDPR Art. 32(1)(a); ISO 27001 A.10.1 Key rotation; Secure escrow; Hardware security modules; Duty separation Encryption audit; Key management review; Protection testing Critical Implementation Phase 2 (31–60 days) Data Security; Encryption Team Monthly key validation; Quarterly audit High Severe
Data security: retention Automated post-report purging; Study-specific policies; Deletion verification; Exception workflow HIPAA §164.310(d)(2)(i); GDPR Art. 5(1)(e); ISO 27001 A.18.1.3 Schedule enforcement; Cryptographic erasure; Media sanitization; Storage monitoring Deletion verification; Compliance review; Storage scanning Required Implementation Phase 3 (61–90 days) Records Management; Legal; IT Storage Weekly purge validation; Monthly compliance Medium High
Operational security: audit logging Centralized SIEM; Tamper-evident logs; ≥ 1 year retention; Access/activity/system logs HIPAA §164.312(b); NIST SP 800-92; ISO 27001 A.12.4 Correlation analysis; Automated alerting; Integrity validation; Access monitoring Coverage verification; Integrity testing; Alert effectiveness Required Implementation Phase 2 (31–60 days) Security Operations; Compliance Daily log review; Monthly integrity check High High
Operational security: monitoring 24/7 security monitoring; Behavioral anomaly detection; IOC monitoring; Activity tracking HIPAA §164.308(a)(1)(ii)(D); NIST SP 800-137; ISO 27001 A.12.7 Behavior profiling; Threshold alerts; Anomaly detection; Real-time monitoring Detection capability tests; False positive analysis; Response timing Recommended Implementation Phase 3 (61–90 days) Security Operations Center; CIRT Continuous monitoring; Weekly review High Medium-High
Operational security: incident response IR playbooks; Escalation procedures; Breach notification process; Annual exercises HIPAA §164.308(a)(6); HITECH §13402; ISO 27001 A.16.1 Incident classification; Evidence preservation; Forensic capability; Communication Quarterly tabletop exercises; Annual simulation; Post-incident reviews Required Implementation Phase 2 (31–60 days) Incident Response Team; Legal Quarterly validation exercises Medium-High High
User security: awareness Initial/annual training; Phishing simulations; Role-specific training; Policy acknowledgment HIPAA §164.308(a)(5); GDPR Art. 32(4); ISO 27001 A.7.2.2 Targeted role training; Microlearning; Security champions; Knowledge assessment Completion audits; Assessment scores; Simulation metrics Required Implementation Phase 3 (61–90 days) Security Awareness; HR Monthly phishing tests; Quarterly training Low-Medium Medium-High
User security: remote work policies Acceptable use policy; Home network requirements; Work/personal separation; Environmental standards HIPAA §164.310(b); NIST SP 800-46; ISO 27001 A.6.2.2 Policy acknowledgment; Remote certification; Environment verification; Agreements Policy compliance audit; Environment checks; Usage monitoring Required Implementation Phase 3 (61–90 days) HR; Security Policy Team; Teleradiology Director Annual policy refresh; Quarterly checks Low Medium-High
Vendor security: assessment Vendor security questionnaires; Contract security requirements; Risk rating methodology; Right-to-audit clauses HIPAA §164.308(b); HITECH §13401; ISO 27001 A.15.1 Security SLAs; Risk-based assessment; Security incident reporting; Key personnel vetting Initial assessment; Annual review; Security incident analysis Required Implementation Phase 2 (31–60 days) Vendor Management; Procurement; Security Initial then annual reassessment Medium High
Compliance management: documentation Security policies; Technical standards; Procedures; Guidelines tailored for teleradiology HIPAA §164.316; GDPR Art. 24; ISO 27001 A.5.1 Document management system; Version control; Approval workflows; Accessibility Document currency audit; Implementation verification; Effectiveness review Required Implementation Phase 1 (0–30 days) Compliance Team; Security Policy Annual review; Quarterly spot checks Low-Medium Medium-High
Risk management: assessment Asset inventory; Risk analysis methodology; Threat modeling; Vulnerability management program HIPAA §164.308(a)(1)(ii)(A); NIST SP 800-30; ISO 27001 A.8.1 Risk register; Mitigation planning; Residual risk acceptance; Executive reporting Quarterly risk review; Annual comprehensive assessment; Gap analysis Required Implementation Phase 1 (0–30 days) Risk Management; Security Leadership Quarterly risk review; Daily vulnerability tracking Medium High
Business continuity: teleradiology DR planning; Backup interpretation capabilities; Alternative workstation access; Communication planning HIPAA §164.308(a)(7); NIST SP 800-34; ISO 27001 A.17 Redundant connectivity; Alternative reading locations; Cloud-based backup PACS; Prioritization Annual DR testing; Backup capability verification; RTO/RPO validation Required Implementation Phase 3 (61–90 days) Business Continuity; Radiology Operations Annual full test; Quarterly component tests High Severe

AAL2+ Authentication Assurance Level 2 or higher, AES advanced encryption standard, API application programming interface, BYOD bring your own device, CIS Center for Internet Security, CIRT Computer Incident Response Team, DISA STIG Defense Information Systems Agency Security Technical Implementation Guide, DLP data loss prevention, DR disaster recovery, EDR/XDR endpoint/extended detection and response, FIPS Federal Information Processing Standards, GDPR General Data Protection Regulation, HIPAA Health Insurance Portability and Accountability Act, HITECH Health Information Technology for Economic and Clinical Health Act, IDS/IPS intrusion detection/prevention system, IOC indicators of compromise, IPsec Internet Protocol Security, IR incident response, ISO International Organization for Standardization, MFA multi-factor authentication, NGFW next-generation firewall, NIST National Institute of Standards and Technology, OIDC OpenID Connect, PFS perfect forward secrecy, PHI protected health information, RIS Radiology Information System, RPO/RTO recovery point objective/recovery time objective, SAML security assertion markup language, SIEM Security information and event management, SLA service level agreement, TDE transparent data encryption, TLS transport layer security, VLAN virtual local area network

Patient data protection strategies

Protecting patient information in teleradiology workflows requires robust, structured, and context-specific strategies that extend beyond generic security controls to address the unique properties of medical imaging data [73]. One foundational principle is data minimization, which restricts the transmission of personal health information to only those elements essential for diagnostic interpretation. Advanced implementations often employ partial anonymization, preserving clinically relevant identifiers (e.g., age, gender, scan date) while stripping extraneous personal details, thereby reducing the privacy impact in the event of a security breach [74].

Practical implementation guide

Successful teleradiology implementation requires structured decision-making regarding scope, ranging from limited emergency coverage to comprehensive enterprise deployment, with phased approaches recommended (Table 4).

Table 4.

Teleradiology implementation models: strategic considerations and relative cost framework

Implementation component Academic medical center Community hospital network Subspecialty practice Private teleradiology service Key considerations Risk factors
Technical infrastructure
 Display systems High investment priority Medical-grade displays essential Multiple modality support Medium-High investment Standard diagnostic quality Dual-modality optimization Highest investment priority Subspecialty-specific requirements Premium calibration needs Medium investment Standardized enterprise solution Volume-based procurement Cannot compromise on diagnostic quality. Medical-grade displays required for primary interpretation. Modality-specific requirements must be met. Technology obsolescence without proper planning. Inadequate specifications reducing diagnostic confidence.
 Computing hardware High performance requirements Multi-tasking capabilities Advanced visualization support Medium performance requirements Standard clinical workflows Reliable operation priority High performance requirements Specialized processing needs Advanced analytics support Medium performance requirements Standardized configurations Proven reliability focus Hardware must meet minimum performance specifications for expected study volumes. Future-proofing considerations important. Hardware failure without redundancy. Insufficient performance affecting workflow efficiency.
 Network infrastructure Enterprise-grade connectivity Redundancy essential Institutional integration Business-class service Multiple site coordination Reliability priority Specialized optimization High-bandwidth requirements Low-latency critical Standardized deployment Proven connectivity models Service level agreements Bandwidth and latency directly impact user experience. Redundancy critical for clinical operations. Connectivity failures disrupting patient care. Inadequate bandwidth causing workflow bottlenecks.
 Security framework Comprehensive enterprise security Multi-layer protection Institutional compliance Standard healthcare security Proven security models Regulatory compliance Specialized security needs Data protection priority Audit trail requirements Enterprise security framework Standardized implementation Comprehensive monitoring Must meet all applicable regulatory requirements. Security cannot be compromised. Regulatory penalties from inadequate security. Data breaches affecting institutional reputation.
Operational considerations
 Quality assurance Comprehensive academic protocols Peer review integration Teaching requirements Standard clinical protocols Quality monitoring systems Performance tracking Subspecialty-focused protocols Specialized metrics Expert review processes Standardized quality framework Automated monitoring systems Performance benchmarking QA standards must match on-site interpretation quality. Continuous monitoring essential. Quality compromise affecting patient outcomes. Inconsistent performance across locations.
 Training and support Advanced continuing education Research integration Trainee supervision Standard continuing education Clinical workflow focus User support priority Subspecialty development Advanced techniques Expert consultation Standardized training programs Efficient support models User competency focus Training must address both technical and clinical components. Ongoing support critical. Knowledge gaps causing operational inefficiency. Inadequate support affecting user adoption.
 Workflow integration Complex academic workflows Research coordination Teaching integration Standard clinical workflows Multi-site coordination Efficiency optimization Specialized workflows Expert consultation Integration complexity Optimized production workflows Standardized processes Efficiency maximization Workflow integration affects user acceptance and efficiency. Change management critical. Workflow disruption reducing productivity. Poor integration affecting user satisfaction.
Implementation models
 Scope definition Hybrid academic/clinical model Research support requirements Educational integration Clinical service focus Multi-site coverage Emergency support Subspecialty expertise distribution Expert consultation model Geographic expansion Production optimization Volume-based efficiency Service scalability Scope must align with institutional goals and capabilities. Phased approach often optimal. Scope creep affecting project success. Unrealistic expectations causing implementation failure.
 Timeline expectations 24–36 months comprehensive implementation Complex requirements Extended validation 18–24 months standard implementation Proven approaches Focused deployment 24–30 months specialized implementation Custom requirements Expert validation 12–18 months standardized implementation Proven methodologies Efficient deployment Realistic timelines critical for success. Adequate testing and validation time required. Rushed implementation compromising quality. Extended timelines affecting institutional support.
 Success metrics Academic productivity measures Teaching effectiveness Research integration Quality improvements Clinical efficiency measures Service quality metrics User satisfaction Cost effectiveness Subspecialty expertise metrics Expert consultation quality Geographic reach Clinical outcomes Production volume metrics Service efficiency Quality consistency Cost optimization Metrics must align with implementation goals. Regular monitoring and adjustment needed. Inappropriate metrics misguiding decisions. Lack of measurement preventing optimization.
Strategic benefits
 Operational flexibility Enhanced scheduling flexibility Geographic coverage expansion Improved work-life balance Staff retention benefits Multi-site service optimization After-hours coverage Emergency response Geographic expansion Subspecialty expertise distribution Expert consultation availability Improved service reach Enhanced recruitment Service scalability Production optimization Geographic flexibility Cost efficiency Benefits realization requires proper implementation and management. Quantification varies by system. Benefit overestimation affecting ROI. Implementation quality determining actual benefits.
 Competitive advantage Academic excellence maintenance Recruitment enhancement Research capabilities Service differentiation Service territory expansion Competitive positioning Market leadership Improved efficiency Subspecialty expertise positioning Market differentiation Expert service delivery Competitive recruitment Service efficiency optimization Market expansion Cost competitiveness Production scalability Competitive advantages require sustained execution and quality maintenance. Competitive threats from poor execution. Market changes affecting advantage sustainability.
Implementation scale
 Small scale (1–5 users) Limited economies of scope Pilot program potential Research focus Limited optimization potential Emergency coverage focus Specialized focus advantage Niche service development Generally not viable Fixed cost challenges Requires careful scope management to achieve positive outcomes. Often serves as proof of concept. Insufficient scale for full benefit realization. High per-user costs.
 Medium scale (6–15 users) Balanced approach Operational efficiency Educational integration Optimized clinical model Multi-site integration Specialty service expansion Expert network development Viable service model Standard deployment Ideal scale for many practice environments. Balanced approach with significant benefit potential. Change management challenges without proper planning. Complexity requiring structured approach.
 Large scale (16+ users) Enterprise optimization Multi-specialty coordination Transformative potential Multi-site integration comprehensive deployment Multi-specialty coordination Complex service networks Scalable enterprise model Maximum efficiency Maximum benefit realization through comprehensive deployment. Requires enterprise project management. Implementation complexity requiring phased approach. Coordination challenges across multiple sites.

Regional considerations: Cost relationships and implementation considerations based on developed healthcare markets. Significant regional variations expected in absolute costs, regulatory requirements, reimbursement models, and productivity benefits. All operational considerations listed per workstation unless otherwise specified. Institutional costs vary significantly based on scale and integration requirements

QA quality assurance, ROI return on investment, SLA service level agreement

Successful teleradiology deployment requires comprehensive checklists covering technical infrastructure (workstation specifications, connectivity validation, security implementation, environmental assessment) and administrative readiness (credential verification, billing integration, scheduling planning, support infrastructure) to ensure operational readiness.

Conclusions

Literature review and real-world implementation experience demonstrate that properly implemented home-based teleradiology achieves diagnostic performance equivalent to on-site interpretation while enhancing radiologist satisfaction and extending subspecialty expertise across geographic boundaries. Technical evidence supports high-quality remote interpretation when display technologies meet ACR/AAPM guidelines, despite that mammography and subtle abnormality detection may benefit from controlled reading room conditions.

Significant knowledge gaps persist, including a lack of consensus on network performance specifications, incomplete validation protocols for consumer-grade hardware, and insufficient calibration standards for specialized applications. Regulatory gaps include limited standardization in multi-jurisdictional licensure models and underdeveloped reimbursement frameworks for distributed care. Future studies require detailed, structured methodologies addressing technical optimization through comparative evaluations and clinical outcome studies assessing diagnostic accuracy, consistency, efficiency, and sensitivity. Multi-center studies with reproducible methodologies across diverse practice settings will be essential for developing universally applicable recommendations and informing future teleradiology standards.

Abbreviations

AAPM

American Association of Physicists in Medicine

ACR

American College of Radiology

AI

Artificial intelligence

CT

Computed tomography

DICOM

Digital Imaging and Communications in Medicine

GDPR

General data protection regulation

GPU

Graphics processing unit

GSDF

Grayscale standard display function

HIPAA

Health Insurance Portability and Accountability Act

IPsec

Internet protocol security

MFA

Multi-factor authentication

MRI

Magnetic resonance imaging

PACS

Picture Archiving and Communication Systems

QA

Quality assurance

RAM

Random access memory

UPS

Uninterruptible power supply

VPN

Virtual private network

Author contributions

Mustafa Sajjad M. Alhasan conceptualized, researched, wrote, and reviewed the manuscript in full; Ayman Sajjad M. Alhasan contributed to manuscript writing, review, and editing. Both authors read and approved the final manuscript.

Funding

M.S.A. is affiliated with Teleradiology Solutions. The remaining authors received no specific funding for this work.

Data availability

No new data were generated or analyzed in this review article.

Declarations

Ethics approval and consent to participate

This article is a narrative review and does not involve human participants, animals, or patient data. Therefore, ethical approval and consent to participate are not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Rosenkrantz AB, Hanna TN, Steenburg SD, Tarrant MJ, Pyatt RS, Friedberg EB (2019) The current state of teleradiology across the United States: a national survey of radiologists’ habits, attitudes, and perceptions on teleradiology practice. J Am Coll Radiol 16:1677–1687. 10.1016/j.jacr.2019.05.053 [DOI] [PubMed] [Google Scholar]
  • 2.Çoraplı M, Oktay C, Çil E, Çoraplı G, Bulut HT (2021) The role of teleradiology in the COVID-19 pandemic. Cerrahpaşa Med J 45:80–86. 10.5152/cjm.2021.21028
  • 3.Bass RZ, Smith AD, Langston MC, Frazier MB, Tridandapani S (2022) Trainee and faculty perceptions of remote PACS workstations and next steps in a large US academic medical institution. Curr Probl Diagn Radiol 51:146–151. 10.1067/j.cpradiol.2021.09.003 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.American College of Radiology (2022) ACR-AAPM-SIIM technical standard for electronic practice of medical imaging 2017. https://gravitas.acr.org/PPTS/GetDocumentView?docId=136 [DOI] [PMC free article] [PubMed]
  • 5.Brady AP, Loewe C, Brkljacic B et al (2025) Guidelines and recommendations for radiologist staffing, education and training. Insights Imaging 16:57. 10.1186/s13244-025-01926-6 [DOI] [PMC free article] [PubMed]
  • 6.Özmen MN, Dicle O, Şenol U, Aydıngöz Ü (2021) TSR guidelines for the practice of teleradiology: 2021 update. Diagn Interv Radiol 27:504–510 [DOI] [PMC free article] [PubMed]
  • 7.Quraishi MI, Rizvi AA, Heidel RE (2020) Off-site radiology workflow changes due to the coronavirus disease 2019 (COVID-19) pandemic. J Am Coll Radiol 17:878–881. 10.1016/j.jacr.2020.05.008 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.McRoy C, Patel L, Gaddam DS et al (2020) Radiology education in the time of COVID-19: a novel distance learning workstation experience for residents. Acad Radiol 27:1467–1474. 10.1016/j.acra.2020.08.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Miller DL, Kakar S, Jiang L, Spelic DC, Burk L (2021) The US Food and Drug Administration’s role in improving radiation dose management for medical X-ray imaging devices. Br J Radiol 94:20210373 [DOI] [PMC free article] [PubMed]
  • 10.Hirschorn DS, Krupinski EA, Flynn, MJ (2014) PACS displays: how to select the right display technology. J Am Coll Radiol 11:1270–1276 [DOI] [PubMed]
  • 11.Jorritsma W, Cnossen F, Dierckx RA, Oudkerk M, Van Ooijen PM (2016) Post-deployment usability evaluation of a radiology workstation. Int J Med Inform 85:28–35 [DOI] [PubMed]
  • 12.Al-Katib S, Dearden A, Al-Bassam N et al (2025) Optimizing radiology remote reading: leveraging technology to improve efficiency. Abdom Radiol (NY) 50:1353–1362 [DOI] [PubMed]
  • 13.Rubin GD (2017) Costing in radiology and health care: rationale, relativity, rudiments, and realities. Radiology 282:333–347 [DOI] [PubMed] [Google Scholar]
  • 14.Choi Y (2023) Rethinking Radiology Monitor Purchases: Cost vs. Quality. Axis Imaging News. October 30, 2023. https://axisimagingnews.com/radiology-products/rethinking-radiology-monitor-purchases-cost-vs-quality
  • 15.Ikushima Y, Tokurei S, Sato S et al (2022) Influence of monitor display resolution and displayed image size on the spatial resolution of ultra-high-resolution CT images: a phantom study. Radiol Phys Technol 15:147–155 [DOI] [PubMed] [Google Scholar]
  • 16.Çağlar B, Serin S (2021) The effect of different monitor use on radiography interpretation in emergency medicine. J Contemp Med 11:18–21
  • 17.Tokurei S, Ikushima Y, Takegami K, Okada M, Morishita J (2021) Medical display application for degraded image sharpness restoration based on the modulation transfer function: initial assessment for a five-megapixel mammography display monitor. Phys Eng Sci Med 44:581–589 [DOI] [PubMed] [Google Scholar]
  • 18.Laffranchi A, Cicero C, Lualdi M et al (2018) Different pixel pitch and maximum luminance of medical grade displays may result in different evaluations of digital radiography images. Radiol Med 123:586–592. 10.1007/s11547-018-0891-6 [DOI] [PubMed] [Google Scholar]
  • 19.Garcia I, Navarro M, Macêdo E, Jesus D, Leite H (2021) Response of luminance meters used for radiodiagnostic applications. J Phys Conf Ser 1826:012054. 10.1088/1742-6596/1826/1/012054 [Google Scholar]
  • 20.Martínez J, Baciu T, Artigues M et al (2019) Nuclear medicine: workplace monitoring and internal occupational exposure during a ventilation/perfusion single-photon emission tomography. Radiat Environ Biophys 58:407–415. 10.1007/s00411-019-00798-x [DOI] [PubMed] [Google Scholar]
  • 21.Bevins NB, Silosky MS, Badano A, Marsh RM, Flynn MJ, Walz-Flannigan AI (2020) Practical application of AAPM Report 270 in display quality assurance: a report of Task Group 270. Med Phys 47:920. 10.1002/mp.14227 [DOI] [PubMed] [Google Scholar]
  • 22.Silosky MS, Marsh RM, Scherzinger AL (2016) Imaging acquisition display performance: an evaluation and discussion of performance metrics and procedures. J Appl Clin Med Phys 17:334–341. 10.1120/jacmp.v17i4.6220 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Ruuge AE, Mahmood UA, Erdi YE (2017) The assessment and characterization of the built-in internal photometer of primary diagnostic monitors. J Appl Clin Med Phys 18:170–175. 10.1002/acm2.12047 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Baniassadi A, Yu W, Travison T, Day R, Lipsitz L, Manor B (2025) Home ambient temperature and self-reported attention in community-dwelling older adults. J Gerontol A. 10.1093/gerona/glae286 [DOI] [PMC free article] [PubMed]
  • 25.McEntee M, Gafoor S (2009) Ambient temperature variation affects radiological diagnostic performance. Proc SPIE. 10.1117/12.811555
  • 26.Pinto dos Santos D, Welter J, Emrich T et al (2017) Comparison of medical-grade and calibrated consumer-grade displays for diagnosis of subtle bone fissures. Eur Radiol 27:5049–5055. 10.1007/s00330-017-4923-8 [DOI] [PubMed] [Google Scholar]
  • 27.Ho E, Tran J, Fateri C et al (2024) Work-related musculoskeletal disorders affecting diagnostic radiologists and prophylactic physical therapy regimen. Curr Probl Diagn Radiol 53:527–532. 10.1067/j.cpradiol.2024.03.008 [DOI] [PubMed] [Google Scholar]
  • 28.Li Y, Xu H, Li R, Cheng M, Wen C (2024) Prevalence and risk factors of work-related musculoskeletal disorders among healthcare workers in medical radiation environments. Work 79:1867–1877. 10.3233/wor-230552 [DOI] [PubMed] [Google Scholar]
  • 29.Vahdatpour B, Khosravi S, Rahimi A, Sattari S, Mogtaderi A, Dabiri Sco F (2010) Work-related musculoskeletal disorders among radiologists in Isfahan: a cross-sectional study. Res J Biol Sci 5:664–669. 10.3923/rjbsci.2010.664.669 [Google Scholar]
  • 30.Jacquier-Bret J, Gorce P (2023) Prevalence of body area work-related musculoskeletal disorders among healthcare professionals: a systematic review. Int J Environ Res Public Health 20:841. 10.3390/ijerph20010841 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Vanhees L, Geladas N, Hansen D, et al (2012) Importance of characteristics and modalities of physical activity and exercise in cardiovascular health: recommendations from the EACPR. Part II. Eur J Prev Cardiol. 19:1005–1033. 10.1177/1741826711430926 [DOI] [PubMed]
  • 32.Mohan C SM (2018) Ergonomics in radiology—time to revisit. Indian J Radiol Imaging 28:271–272. 10.4103/ijri.IJRI_358_18 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Glover AM, Whitman GJ, Shin K (2022) Ergonomics in radiology: improving the work environment for radiologists. Curr Probl Diagn Radiol 51:680–685. 10.1067/j.cpradiol.2022.03.001 [DOI] [PubMed] [Google Scholar]
  • 34.Halpenny D, O'Driscoll D, Torreggiani WC (2012) Ocular health among radiologists in the age of PACS: is it time for our profession to open its eyes to this issue in light of existing European legislation? Br J Radiol 85:e1309–e1311. 10.1259/bjr/27200799 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Sze G, Bluth EI, Bender CE, Parikh JR (2017) Work-related injuries of radiologists and possible ergonomic solutions: recommendations from the ACR Commission on Human Resources. J Am Coll Radiol 14:1353–1358. 10.1016/j.jacr.2017.06.021 [DOI] [PubMed] [Google Scholar]
  • 36.Brusan A, Durmaz A, Ozturk C (2021) A workflow for ensuring DICOM compatibility during radiography device software development. J Digit Imaging 34:717–730. 10.1007/s10278-021-00458-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Salpea N (2022) Medical Image Segmentation: A Review of Modern Architectures. In Computer Vision – ECCV 2022 Workshops (Lecture Notes in Computer Science. 10.1007/978-3-031-25082-8_47
  • 38.Rashid M, Nasiri A, Hamidi S (2024) Chapter 18 - Uninterruptible Power Supplies. In: Rashid, M (ed) Power Electronics Handbook, 5th edn. pp 687–703. 10.1016/B978-0-323-99216-9.00044-5
  • 39.Fatehi M, Safdari R, Ghazisaeidi M, Jebraeily M, Habibi-Koolaee M (2015) Data standards in tele-radiology. Acta Inform Med 23:165. 10.5455/aim.2015.23.165-168 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Adambounou K, Adjenou V, Salam AP et al (2014) A low-cost tele-imaging platform for developing countries. Front Public Health 2:135. 10.3389/fpubh.2014.00135 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Noumeir R, Pambrun JF (2011) Using JPEG 2000 interactive protocol to stream a large image or a large image set. J Digit Imaging 24:833–843. 10.1007/s10278-010-9343-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Qureshi HN, Manalastas M, Ijaz A, Imran A, Liu Y, Al Kalaa MO (2022) Communication requirements in 5G-enabled healthcare applications: review and considerations. Healthcare. 10.3390/healthcare10020293 [DOI] [PMC free article] [PubMed]
  • 43.Deneke BB, Beyene AM, Haile EA (2024) Improving software defined network controllers in a multi-vendor environment. Heliyon. 10.1016/j.heliyon.2024.e26215 [DOI] [PMC free article] [PubMed]
  • 44.Garg V, Brewer J (2011) Telemedicine security: a systematic review. J Diab Sci Technol 5:768–777. 10.1177/193229681100500331 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Vassilacopoulos G, Chrissikopoulos V, Peppes D (1996) Network security in a telemedicine system. In: Horster P (ed) Communications and multimedia security II: proceedings of the IFIP TC6/TC11 international conference on communications and multimedia security at Essen, Germany, 23–24 September 1996. Springer, Boston, pp 108–115
  • 46.Guillén E Estupiñán P (2012) Analysis of security requirements in telemedicine networks. Retrieved from ResearchGate. https://www.researchgate.net/profile/Edward-Guillen/publication/266871076_Analysis_of_security_requirements_in_telemedicine_networks/links/55e7060408ae65b638994609/Analysis-of-security-requirements-in-telemedicine-networks.pdf?_tp=eyJjb250ZXh0Ijp7ImZpcnN0UGFnZSI6InB1YmxpY2F0aW9uIiwicGFnZSI6InB1YmxpY2F0aW9uIn19
  • 47.Fausett CM, Christovich MP, Parker, JM, Baker, JM, Keebler, JR (2021) Telemedicine Security: Challenges and Solutions. Proceedings of the International Symposium on Human Factors and Ergonomics in Health Care, 10:340–344. 10.1177/2327857921101241 (Original work published 2021)
  • 48.Shore MW, Rubin DL, Kahn CE (2012) Integration of imaging signs into RadLex. J Digit Imaging 25:50–55. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Datta S, Godfrey-Stovall J, Roberts K (2020) RadLex normalization in radiology reports. AMIA Annu Symp Proc 2020:338–347 [PMC free article] [PubMed] [Google Scholar]
  • 50.Wang Y, Mehrabi S, Sohn S, Atkinson EJ, Amin S, Liu H (2019) Natural language processing of radiology reports for identification of skeletal site-specific fractures. BMC Med Inform Decis Mak 19:73. 10.1186/s12911-019-0780-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Hussain S, Naseem U, Ali M et al (2024) TECRR: a benchmark dataset of radiological reports for BI-RADS classification with machine learning, deep learning, and large language model baselines. BMC Med Inform Decis Mak 24:310. 10.1186/s12911-024-02717-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Kawa J, Pyciński B, Smoliński M et al (2022) Design and implementation of a cloud PACS architecture. Sensors. 10.3390/s22218569 [DOI] [PMC free article] [PubMed]
  • 53.Silva LAB, Costa C, Oliveira JL (2012) A PACS archive architecture supported on cloud services. Int J Comput Assist Radiol Surg 7:349–358. 10.1007/s11548-011-0625-x [DOI] [PubMed] [Google Scholar]
  • 54.Andriole KP (2023) Picture archiving and communication systems: past, present, and future. J Med Imaging. 10.1117/1.Jmi.10.6.061405 [DOI] [PMC free article] [PubMed]
  • 55.Cohen RY, Sodickson AD (2023) An orchestration platform that puts radiologists in the driver’s seat of AI innovation: a methodological approach. J Digit Imaging 36:700–714. 10.1007/s10278-022-00649-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Mildenberger P, Eichelberg M, Martin E (2002) Introduction to the DICOM standard. Eur Radiol 12:920–927. 10.1007/s003300101100 [DOI] [PubMed] [Google Scholar]
  • 57.Loose R, Braunschweig R, Kotter E, Mildenberger P, Simmler R, Wucherer M (2009) [Compression of digital images in radiology—results of a consensus conference]. Rofo 181:32–37 [DOI] [PubMed]
  • 58.Pinto dos Santos D, Brodehl S, Baeßler B et al (2019) Structured report data can be used to develop deep learning algorithms: a proof of concept in ankle radiographs. Insights Imaging 10:93. 10.1186/s13244-019-0777-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.dos Santos DP, Friese C, Borggrefe J, Mildenberger P, Mahringer-Kunz A, Kloeckner R (2020) The impact of irreversible image data compression on post-processing algorithms in computed tomography. Diagn Interv Radiol 26:22. 10.5152/dir.2019.18245 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Jorg T, Kämpgen B, Feiler D et al (2023) Efficient structured reporting in radiology using an intelligent dialogue system based on speech recognition and natural language processing. Insights Imaging 14:47. 10.1186/s13244-023-01392-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Hammana I, Lepanto L, Poder T, Bellemare C, Ly MS (2015) Speech recognition in the radiology department: a systematic review. Health Inf Manag 44:4–10. 10.1177/183335831504400201 [DOI] [PubMed] [Google Scholar]
  • 62.McDonald RJ, Schwartz KM, Eckel LJ et al (2015) The effects of changes in utilization and technological advancements of cross-sectional imaging on radiologist workload. Acad Radiol 22:1191–1198. 10.1016/j.acra.2015.05.007 [DOI] [PubMed] [Google Scholar]
  • 63.Lee SH, Park J, Yang K, Min J, Choi J (2022) Accuracy of cloud-based speech recognition open application programming interface for medical terms of Korean. J Korean Med Sci. 10.3346/jkms.2022.37.e144 [DOI] [PMC free article] [PubMed]
  • 64.Hodgson T, Coiera E (2016) Risks and benefits of speech recognition for clinical documentation: a systematic review. J Am Med Inform Assoc 23:e169–e179. 10.1093/jamia/ocv152 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Aldosari B, Babsai R, Alanazi A, Aldosari H, Alanazi A (2023) The progress of speech recognition in health care: surgery as an example. Stud Health Technol Inform 305:414–418. 10.3233/shti230519 [DOI] [PubMed] [Google Scholar]
  • 66.Luo X, Zhou L, Adelgais, K et al (2025) Assessing the effectiveness of automatic speech recognition technology in emergency medicine settings: a comparative study of four AI-powered engines. J Healthc Inform Res 9:494–512. 10.1007/s41666-025-00193-w [DOI] [PMC free article] [PubMed]
  • 67.Jeon S, Kim MS (2022) End-to-end lip-reading open cloud-based speech architecture. Sensor (Basel). 10.3390/s22082938 [DOI] [PMC free article] [PubMed]
  • 68.Zhang J, Wu J, Qiu Y et al (2023) Intelligent speech technologies for transcription, disease diagnosis, and medical equipment interactive control in smart hospitals: a review. Comput Biol Med 153:106517. 10.1016/j.compbiomed.2022.106517 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69.Bosmans JML, Peremans L, Menni M, De Schepper AM, Duyck PO, Parizel PM (2012) Structured reporting: if, why, when, how-and at what expense? Results of a focus group meeting of radiology professionals from eight countries. Insights Imaging 3:295–302. 10.1007/s13244-012-0148-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Salg GA, Ganten M-K, Bucher AM et al (2021) A reporting and analysis framework for structured evaluation of COVID-19 clinical and imaging data. NPJ Digit Med 4:69. 10.1038/s41746-021-00439-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Ruotsalainen P (2010) Privacy and security in teleradiology. Eur J Radiol 73:31–35. 10.1016/j.ejrad.2009.10.018 [DOI] [PubMed] [Google Scholar]
  • 72.Madej M, Sąsiadek MJ (2023) The growing role of telemedicine—possibilities and regulations concerning teleradiology in Poland. Pol J Radiol 88:535–545. 10.5114/pjr.2023.133456 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Andriole KP (2014) Security of electronic medical information and patient privacy: what you need to know. J Am Coll Radiol 11:1212–1216. 10.1016/j.jacr.2014.09.011 [DOI] [PubMed] [Google Scholar]
  • 74.Moore W, Frye S (2019) Review of HIPAA, Part 1: History, protected health information, and privacy and security rules. J Nucl Med Technol 47:269–272. 10.2967/jnmt.119.227819 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

No new data were generated or analyzed in this review article.


Articles from Insights into Imaging are provided here courtesy of Springer