An affordable, water-based, community condensation particle counter

Susanne Hering a, Stavros Amanatidis a, Arantza Eiguren-Fernandez a, Gregory S Lewis a, Steven R Spielman a, David Pariseau b
PMCID: PMC12448066  NIHMSID: NIHMS2083674  PMID: 40979442

Abstract

A “Community Condensation Particle Counter” (cCPC) has been developed to provide an affordable monitor of airborne particle number concentrations. The cCPC is an expansion-type condensation particle counter that incorporates single particle counting to yield a direct measurement of number concentration. Particle number concentrations are derived from the detection of individual droplets exiting the cell during the expansion, combined with the pressure readings and the physical volume of the particle cell. Modeling and experiment confirm detection of particles as small as 4 nm, with >95% detection above 20 nm. For 12 days of ambient sampling two collocated cCPCs exhibit a pooled standard deviation of 3.5%. Comparison to a pair of benchtop instruments (ADI MAGIC CPCs) yields a correlation of R2=0.98 and a regression slope of 1.1. Laboratory studies at concentrations higher than 3×104 cm−3 for both sulfate and dioctyl sebacate show equally reduced response when compared to a versatile water CPC, but this was not observed in ambient aerosol sampling. Further research will be needed to resolve this discrepancy.

Graphical Abstract:

graphic file with name nihms-2083674-f0001.jpg

1. Introduction

With the increasing interest in community monitoring of air pollutants, there is a need to develop instrumentation that provides accurate and reliable data at reasonable cost. While low-cost sensors continue to evolve for monitoring of gaseous pollutants and estimating suspended particle mass, an affordable method for ultrafine particles remains elusive. Loosely defined as those particles with diameters below 100 nm, ultrafine particles are too small to be detected by optical devices commonly used. Moreover, there are many sources, both indoors and outdoors, leading to large spatial and temporal variability. Thus, the levels to which people are exposed are difficult to capture through traditional central site monitoring. Marginalized communities, often located near known ultrafine particle emission sources such as freeways, are found to have high concentrations relative to the urban average (Gani et al., 2021).

Although not explicitly regulated, ultrafine particles are of concern for human health. When inhaled, ultrafine particles are deposited more deeply in the lung, are cleared more slowly from the respiratory system, and are distributed more efficiently to the bloodstream and organs than larger particles (Health Effects Institute, 2013). They can exacerbate respiratory symptoms in patients with chronic airway diseases, and are associated with variable heart rates, higher systolic blood pressure, and brain cancers (e Oliveira et al., 2019; Cassee et al., 2013; Schraufnagel, 2020; Marval and Tronville, 2022), as well as increased respiratory and cardiovascular hospital admissions (Bergmann et al., 2023). The European Parliament recently identified ultrafine particles as a pollutant of “emerging concern” for which monitoring may be necessary (European Parliament, 2024). Yet, the body of epidemiological literature on the relationship between health and exposure is limited, mainly due to the lack of routine monitoring data. Wide-spread, continuous monitoring is needed to fill these gaps, but is not feasible without affordable, reliable instrumentation.

While abundant by number, the mass of suspended ultrafine particles is small. Generally, ultrafine particles are not a significant contributor to suspended fine particle mass (PM2.5), yet they dominate the number concentrations, contributing 94% on average in urban areas (Seinfeld and Pandis, 2016). Data presented by Sun et al. (2019) from 21 sites across Germany, variously characterized as urban, roadside, regional background, or alpine, show that the mean fraction of particle number between 10 and 200 nm comprised over 90% of the total number from 10 to 800 nm. These data indicate that for a wide range of environments, the total particle number is a good metric for ultrafine particle concentration.

For decades, condensation particle counters (CPCs) have served as the reference method for particle number concentration measurement in the ultrafine size range (McMurry, 2000). Most CPCs create a region of vapor supersaturation through convective diffusion or mixing of saturated flows with differing temperatures. Within the supersaturated region ultrafine particles grow by condensation to form micrometer-sized droplets that are subsequently counted optically. These CPCs are continuous flow instruments that span a wide range in particle sizes and concentrations. The limitation for community monitoring is their expense as well as the use of butanol by many CPCs.

Several electrometer-based methods have been developed that infer ultrafine particle concentrations from the net electrical charge captured by particles from a unipolar charger (Fierz et al. 2011, 2014; Marra et al., 2010). These instruments do not directly measure the number concentration, but rather a moment of the distribution. Some incorporate an electrostatic precipitator to remove the most mobile particles, from which information on particle size, and hence more accurate number concentrations may be inferred (Asbach et al., 2024). While not a direct measure, electrometer methods offer advantages of compact size, ruggedness and affordability. They have enabled studies of ultrafine particles indoors (e.g., Beko et al., 2013; Mazaheri et al., 2019; Isaxon et al., 2015), and in urban areas using mobile platforms, such as the tram-based network in Zurich (Mueller et al., 2016).

The Community Condensation Particle Counter (cCPC) reported here is designed to have the compact packaging and ruggedness of the electrometer-based measurements, while offering an affordable, direct measurement of particle number concentrations. The cCPC is an expansion type instrument, wherein the supersaturation required for particle activation is created by expansion of a saturated flow, as was done in the nineteenth century by Aitken and Coulier (McMurry, 2000). The expansion instrument of Skala (1963) was automated and shown to be rugged, as it operated for years at the South Pole (Bodhaine and Murphy, 1980). In contrast to that instrument, which assessed particle concentrations from light scattering from the cloud of particles formed, the cCPC presented here incorporates single particle counting to provide a direct and more accurate measurement of particle number. As compared to current CPCs, the expansion approach of the cCPC leads to a less costly instrument. Roughly speaking, the low-cost sensors in current use are about a factor of 100 less expensive than CPCs now available. The cCPC presented here is a factor of 10 less expensive, and as such is a first attempt to bridge this cost gap.

2. Methods

2.1. Instrument Description

The cCPC enlarges ultrafine particles to optically detectable sizes by rapid expansion of a water-vapor saturated air sample contained within a small chamber. The pseudo-adiabatic expansion cools the sample, creating a region of supersaturation and activating the particle condensational growth. During the expansion, air exits the chamber through an optical detector that counts the individual droplets within this flow. At the same time, the pressure in the chamber is measured, from which the air volume is calculated. In contrast to laminar flow and mixing instruments, no heating or cooling elements are required, as all needed cooling is produced by the rapid expansion.

Shown in Figure 1, the components of the cCPC are a particle cell, an optical detector, a pair of valves, an evacuation chamber, a flow limiting orifice and a pump. The particle cell consists of a 31 mm long, 19 mm diameter Nafion tube (TT-700, Permapure, www.permapure.com) supported by a 31 mm long, 18 mm ID perforated stainless-steel cage mounted immediately inside the Nafion. A water jacket surrounds the Nafion, and is sealed by means of O-rings that press against the end caps on either side of the particle cell. The particle cell is directly coupled to the optical detector (Particles Plus Inc., Stoughton, MA). The nozzle delivering the particles to the detector has been replaced by one with a gentle taper to ensure efficient delivery of micrometer-sized droplets. Solenoid valves (ASCO miniature valve, nominal 0.030” orifice, plastic body) are mounted at the inlet of the particle cell and at the exit of the optical chamber. A pressure sensor monitors the pressure in the particle cell during the expansion. The volume of the particle cell, based on the ID of the metal cage and inclusive of the volume from the inlet valve to the tip of the optics nozzle, is 8.9 cm3. As implemented for the experiments presented here, the expansion volume is ~5 cm3, and the downstream orifice is 0.5mm (0.019”). Flow is provided by a small (~2 Lmin−1 free flow) diaphragm pump. We also incorporated a 0.7 L min−1 transport flow to ensure representative delivery of aerosol from the upstream sampling line to the inlet.

Figure 1.

Figure 1.

Community Condensation Particle Counter (cCPC) schematic, showing Particle Cell with Nafion walls, water jacket, inlet and outlet valves, optical detector, expansion volume and optional transport flow line.

As illustrated in Figure 2, the cCPC follows a three-step cycle for each particle number concentration measurement. The flow path at each step is determined by the valve configuration. With both valves open, air is flushed through the cell at a flow rate around 1.2 L min−1. Next, with both valves closed, the flow through the cell stops, and the air sample is humidified by water vapor transport through the Nafion walls. After several seconds, the downstream valve is opened to partially evacuate the cell. The expansion rate is determined by the downstream expansion volume, flow restrictions (including optics nozzle, solenoid valve and flow control orifice) and pumping speed. By selecting an expansion volume that is comparable to the combined size of the particle cell and optical head gas volume, the initial expansion rate is rapid. This rapid expansion lowers the temperature and creates a region of water vapor supersaturation within the particle cell. The expansion continues at a slower rate, during which time condensationally enlarged particles are counted as they exit the cell through an optical detector. The air volume is determined from the pressure change and mechanical volume of the particle cell. To allow time for droplet growth, counting is delayed by several tens of milliseconds from the start of the expansion. This yields a direct measurement of particle number concentrations. Results reported here are for a 10 s cycle consisting of (1) 1.7 s flush step, (2) 8 s humidification step, and (3) 240 ms expansion step.

Figure 2.

Figure 2.

Illustration of the operational cycle of the cCPC.

2.2. Modeling Methods

A numerical model was developed in COMSOL Multiphysics to evaluate fields of the flow velocity, pressure, temperature, water vapor concentration and saturation ratio in the cCPC during the expansion step. The model input is a 2D-axisymmetric geometry consisting of the cylindrical Nafion-tube chamber (18 mm ID, 31 mm length) and a 22 mm-long tapered nozzle that leads to the optical detector. The model incorporates a time-dependent pressure profile applied as a pressure boundary condition at the nozzle outlet cross-section. This pressure profile is tuned to replicate the experimental expansion profile. Initial temperature and humidity in the cell are set at 25°C and 95% relative humidity, respectively. The calculation includes the combined effects of cooling from the expansion and warming due to heat transfer from the walls. As water transport across the Nafion walls is slow compared to the rapid expansion, the wall was modeled as a boundary with zero vapor flux, held at a fixed temperature of 25°C. With these assumptions, the resulting flow velocity, pressure, temperature and water vapor concentration are solved concurrently in a time-dependent simulation, from which the saturation ratio is calculated at each point within the cell over the expansion. Particle activation and growth is calculated with input from the transport study according to Seinfeld and Pandis (2016), assuming wettable particles. Calculations presented here are for low particle concentrations, i.e. condensational heat release and vapor depletion are not included.

2.3. Experimental Evaluation Methods

The size and concentration response of the cCPCs was evaluated using mono-mobility aerosols of ammonium sulfate and dioctyl sebacate. A 2-nm versatile water CPC (Aerosol Dynamics Inc., Berkeley, CA) served as reference. It was operated with conditioner and initiator stage temperatures of 2 and 90 °C to give a 2 nm lower 50% efficiency threshold (Hering et al., 2017; Kangasluoma et al., 2017). At the time of these calibration runs its performance was checked against an aerosol electrometer (Seadm, Valladolid, Spain) with 6–60 nm size-selected particles for both types of test aerosols. As a quality assurance check, a MAGIC CPC (Aerosol Dynamics Inc., Berkeley, CA) was operated alongside the cCPCs under test. The MAGIC CPCs are laminar flow, water-based instruments with a 5 nm lower detection threshold (Hering et al., 2019).

Aerosol was generated by atomization, diluted with dry air, charge-conditioned by means of a soft X-ray source (TSI Model 3087), and size-selected using a nano-column differential mobility analyzer (DMA, TSI Model 3085) operating with a recirculating sheath flow of 15 L min−1 and an aerosol flow of 4 L min−1 (non-diffusive resolution of 3.7). A second soft X-ray charge conditioner was deployed downstream of the DMA to remove the majority of the electrical charge on the size-selected aerosol to minimize particle losses in the plastic body of the cCPC inlet valve. The calibration aerosol was delivered to a 750 cm3 metal plenum from which the aerosol was sampled by the Community and benchtop CPCs. To accommodate flow perturbations caused by the cCPC expansion cycles, the DMA system was operated at slight overpressure, with excess flow exhausting the plenum through a large HEPA filter to the room. In addition, an orifice was placed between the DMA and downstream charge conditioner to further minimize perturbations to the DMA flows.

The accuracy for ambient air monitoring was tested through continuous operation of a pair of cCPC units running side-by-side with a pair of MAGIC CPCs. Reported here are results obtained with the pre-production prototypes manufactured by Particles Plus. Sampling was done at ADI’s laboratory in Berkeley, CA at a location 1 km from an Interstate Freeway.

3. Results

3.1. cCPC Modeling

Figure 3a shows the measured pressure in the cell of the cCPC during expansion. This profile is input into the model as the pressure boundary condition at the nozzle exit. Also shown is the corresponding modeled mean cell pressure. Figure 3b shows the evolution of the temperature within the particle cell for the pressure profile of Figure 3a, when the initial temperature of the air and the particle cell walls are both 25 °C, and the initial relative humidity is 95%. The cell temperature is controlled by the balance between heat transfer from the temperature walls and the cooling from the energy expended by the air in the expansion. The largest decrease occurs in the center of the cell, with the temperature cooling to 7.5 °C. The saturation ratio climbs to a maximum value of 2.1 (Figure 3c), corresponding to a Kelvin size of 3.1 nm. These figures also show the mean temperature and saturation ratio calculated as the mass-weighted mean, i.e., where the value at each point is weighted by the local air density. The change in these mean values is smaller, with the mass-weighted mean temperature within the cell dropping from 25 °C to 17 °C, and the mean saturation ratio plateau at 1.3. As the nozzle is designed to preferentially eject air from the core of the cell, the lines showing the minimum temperature and maximum saturation ratio (solid lines) are more representative of the conditions seen by the particles that are counted than are the mean values.

Figure 3.

Figure 3.

(a) Pressure at the nozzle exit, used as a boundary condition in the model, and the resulting modeled mean cell pressure, during the expansion. (b) Corresponding modeled mass-weighted mean and minimum cell temperature, and (c) modeled mass-weighted mean and maximum saturation ratio. All calculations are for initial conditions of 25°C and relative humidity of 95%.

Figure 4 shows snapshots of the calculated saturation ratio field in the cell during the expansion. The model includes the cylindrical cell and, at the top of each image, the nozzle that couples the cell to the optical detector from which the aerosol is extracted. The geometry is radially symmetric, with the centerline at the left edge of each image. At 50 ms after the onset of the expansion, the air within the core of the cell reaches a rather uniform saturation ratio, with nearly the entire volume having a Kelvin diameter value below 10 nm. Further into the expansion the saturation ratio in the core region of the cell continues to increase, with Kelvin diameters dropping below 4 nm, even as the saturation level near the walls decreases.

Figure 4.

Figure 4.

Evolution of the radial and axial dependence of the modeled saturation ratio (S) during the first 150 ms of the expansion for the pressure profile of Figure 3a. Contours for Kelvin diameter values of 4 nm and 10 nm are also shown. The Particle Cell centerline is along the left-hand edge of each sketch.

Particle activation and growth was also modeled, taking into account the nozzle design, flow streamlines, as well as the saturation ratio, temperature and pressure conditions along the particle trajectories. Our modeling showed that by choosing a nozzle inlet that is of the order of one-half the diameter of the cell, rather than the full cell diameter, the air volume that is directed to the optical detector tends to exclude the region nearest the particle cell walls where saturation ratio is lower. Results with this nozzle design are presented in Figure 5, which shows the fraction of particles exiting the nozzle into the optics head that have grown to a size of 1 μm or larger as a function of time from the onset of the expansion. Traces for particles with an initial size of 10 nm and 20 nm are nearly identical, with a delay of about 30 ms before the first droplets are detected. After this delay the detected fraction increases rapidly, reaching a plateau of 0.90. For the fraction of particles that are not detected, the model indicates the limitation is not insufficient growth, but rather that these particles were close to the walls and were not activated. For a 5 nm initial particle size, the modeling predicts a somewhat longer delay in the onset of activation followed by a comparatively slower increase in the detectable fraction response curve. This analysis indicates the time needed to produce detectable droplets should be taken into account in the deriving of particle number concentrations from the count and pressure data during the expansion.

Figure 5.

Figure 5.

Fraction of particles that activate and grow to a detectable size of 1 μm for 5, 10 and 20 nm seed particles as a function of the expansion time. Modeling is for the pressure profile of Figure 3a, with an initial relative humidity of 95%,.

3.2. Example Data of Expansion Profiles

The two measured parameters during an expansion are the pressure in the cell and the particle count, from which the particle number concentration is calculated. Figure 6 shows high time resolution data of these parameters for an example expansion. The cell pressure, shown in Figure 6a, follows the profile of Figure 3a, decreasing rapidly initially as air fills the downstream expansion volume, and subsequently dropping more slowly. The fraction of the air in the cell that exits during the expansion depends primarily on this pressure, as the modeling shows that the change in the absolute temperature of the air remaining is small, decreasing by about 2%. For our experimental data, we calculate an ambient-equivalent volume of air exiting the cell using the ideal gas law, the mechanical volume of the cell, and the measured pressure profile, ignoring the temperature change within the cell. Figure 6b presents this ambient-equivalent air flow rate, i.e. an air mass flow rate expressed in terms of the volume the air would have at the initial cell temperature and pressure. Overlayed on this air flow rate is the dead-time corrected particle count rate, defined as the number of counts divided by the time the optical signal is below the detection threshold. During the first 20 ms or so no particles are detected, analogous to the delay predicted by the modeling. The count then rises quickly and subsequently decays slowly. After about 40 ms, the air mass flow rate and dead-time corrected count rate align.

Figure 6.

Figure 6.

Measured profiles of (a) Particle Cell pressure, (b) number of droplets detected per ms, and air flow rate expressed in terms of the equivalent volume at ambient temperature and pressure, and (c) the indicated number concentration calculated from Equation (1).

Between any two points in the expansion, t1 and t2, an indicated number concentration Ni can be calculated as the sum of the number of particles detected divided by the mass of air expressed as the equivalent volume at ambient temperature and pressure, that has passed through the optical detector during the interval:

Ni=∑t1t2CiVcP1-P2/P0 Equation (1)

where: Ci: dead-time corrected count in each measurement interval t1: time that counting starts t2: time that counting ends P1: cell pressure immediately before t1 P2: cell pressure immediately after t2 P0: ambient pressure Vc: physical volume of the particle cell

Over that portion of the expansion when the count rate and air mass flow rate are proportional, that is when the traces of Figure 6b overlap, this calculation will give the same reading for the particle concentration. If instead the calculation were done over the entire expansion time, the indicated concentration would be low due to the volume of air exhausted during the time required to create detectable droplets.

Figure 6c illustrates the sensitivity of the indicated particle number concentration to the starting and ending times t1 and t2. In this graph, the “instantaneous reading” (open symbols) is the number concentration inferred at each 64 Hz reading. This instantaneous reading rises sharply from zero to a plateau value that is the particle number concentration. The cumulative values are calculated from a fixed starting point at t1=48ms and continuing to a final reading at t2 shown on the abscissa, i.e. between 55 ms and 200 ms. With this delay in the counting, the indicated particle concentration is not sensitive to the value selected for t2. For t1=48ms, the cumulative values for the particle concentration are within 1% for a t2 range between 100–240 ms.

For our cCPC we do not capture the detailed count and pressure readings for each individual expansion. Instead, we use fixed values for t1 and t2. The clock starts when the expansion valve opens. At t1 the system reads the pressure P1 and starts the counter to accumulate the dead-time corrected counts. At t2 the counter stops, the system reads the final pressure P2 and calculates the particle concentration from Equation (1). Exercising our model, we find that for values of t1>45ms, the isothermal assumption implicit in Equation (1) is accurate to within 2%. For the data presented here we have used consistent values of t1=48ms and t2=240ms.

3.3. Laboratory Evaluation

Over the course of the development of the cCPC, calibrations of various prototype configurations were obtained using monodispersed aerosols. Presented here is the particle size and particle concentration dependence for a pair of prototype cCPCs manufactured by Particles Plus Inc.

Figure 7 shows the size-dependent detection efficiency for ammonium sulfate aerosol, where the reference is the 2-nm versatile water CPC. For these tests the reference particle concentrations were below 104 cm−3. The cCPC particle concentrations were calculated using Equation (1). The cell volume, Vc, is taken from the mechanical drawings, and includes the volume between the inlet valve and the outlet of the nozzle into the optics head.

Figure 7.

Figure 7.

Size-dependent counting efficiency of a pair of cCPCs manufactured by Particles Plus Inc, as compared to a 2-nm versatile water-CPC, where concentrations from the cCPC are calculated from Equation (1). Also shown is comparison to a MAGIC Model 250 CPC.

The detection efficiency, derived from first principles calculation of Equation (1) is approximately 70% at 5 nm, and > 95% for particles above 20 nm. At the smaller sizes the efficiency is higher than for the MAGIC laminar flow CPC (Model 250, Aerosol Dynamics Inc.) which was tested alongside the cCPCs. We note that these tests were conducted using aerosol that was electrically charged-conditioned downstream of the DMA. When the aerosol was charged, i.e. sampled directly from the DMA, we found substantial losses, e.g. greater than 50% for <10 nm particles. These losses were substantially reduced when the inlet valve was replaced by one with a metal body (rather than plastic), but the metal valves were noisier and less reliable.

Figure 8 shows the concentration-dependent response for both ammonium sulfate and dioctyl sebacate particles with a mean mobility size of 34 nm. Shown is the ratio of the indicated particle concentration from the cCPC to that measured by the 2-nm versatile CPC, as a function of the input concentration. The cCPC data are uncorrected, that is, values are calculated from Equation (1). We observe a decrease in counting efficiency at higher particle concentrations, from 95% at low concentrations to 70% at 105 cm−3. The response is similar for both units, and for both the inorganic salt (ammonium sulfate) and the organic aerosol (dioctyl sebacate). Modeling of the instrument response at high concentrations indicates that condensational heating may lower the supersaturation levels, and thus reduce the instrument response. The consistency in response for both inorganic and organic aerosols indicates that a correction could be applied. However, as shown below, this trend is not observed when sampling ambient aerosols. For this reason, the high concentration response remains a topic of further investigation.

Figure 8.

Figure 8.

Concentration dependence of the response of the cCPC for broad distribution of ammonium sulfate and dioctyl sebacate aerosols centered at 34 nm.

3.4. Ambient Air Testing

To test the accuracy for sampling ambient air, a pair of cCPCs was operated unattended for over 12 days alongside two benchtop MAGIC CPCs. No maintenance was provided during these tests. Data recovery for all four instruments was 100%. Time series of these four instruments are presented in Figure 9, where particle concentrations have been averaged to one-hour data. Note the daily variations observed, with sharp midday maxima that exceeded 2×104 cm−3 on several days. The sampling site was our laboratory, located within 1 km of an interstate highway.

Figure 9.

Figure 9.

Time series of ambient data from two cCPCs, and two Magic CPCs operated side-by-side in Berkeley, California.

Figure 10 presents scatterplots of the paired units, comparing the two cCPCs and the two MAGIC units to each other. The two cCPCs are correlated with R2>0.99, and a regression slope of 1.0. The cCPC precision, evaluated as the coefficient of variation, was 3.5%. Similarly, the two MAGIC units are correlated, also with R2>0.99, and a regression slope near 1 and a coefficient of variation of 4.5%. Larger differences are seen when the two cCPCs are compared to the MAGIC units. Figure 11 compares each cCPC to the mean value reported by the two benchtop MAGIC CPCs. For each of the two cCPCs, regression against the mean value of the collocated MAGIC CPCs yielded R2=0.98 and a regression slope of 1.1. Over the measurement period, the mean concentration measured by the cCPCs was 15% higher than by the two MAGIC CPCs. In contrast to the laboratory data for the 34 nm test aerosols, for the mix of particle sizes seen in the ambient the cCPC trends toward reporting higher concentrations than the MAGIC CPCs as the ambient particle number levels rise. This could be due to the difference in the lower cutpoints, as the cCPC efficiently detects particles with diameters below 20 nm. Although size distribution data were not available for these comparisons, it is quite possible that peak concentrations are due to the presence of fresh sources, such as vehicular emissions, that are characterized by a larger prevalence of small (<20 nm) particles.

Figure 10.

Figure 10.

Scatterplots of the ambient sampling data of Figure 9 showing the consistency between a) the two cCPCs, and (b) the two MAGIC CPCs. The lines are one-to-one.

Figure 11.

Figure 11.

Scatterplot of the two cCPCs against the average value from the two collocated MAGIC CPCs. The line is one-to-one.

Further evaluations under field conditions are underway. The Desert Research Institute operated three of the prototype cCPCs alongside a MAGIC CPC over a four-month period at a downtown air quality monitoring station in Reno, Nevada (Wang et al., 2024). Colorado State University (Swanson et al., 2023) operated several early prototype instruments in an indoor study in the Denver, CO area, with occasional comparisons to the TSI-3789 CPC. In the Los Angeles area, three prototype cCPCs are currently operating alongside a TSI Model 3783 CPC at an ASCENT site (Ng et al., 2022). Even though the Reno site is situated 1370 m above sea level, the cCPC readings remained comparable to a collocated benchtop MAGIC instrument. Similarly, comparable results are also found in the Denver study. Further detail is presented in those manuscripts.

4. Conclusions

An affordable and robust condensation particle counter has been developed that targets the need for community monitoring of ultrafine particles. Although not in the price category of ‘low-cost’ sensors, it is roughly one-tenth the price of benchtop CPCs now available. Importantly, it provides a direct measurement that is more accurate than existing low-cost sensors. Much like the first condensation particle counters, the cCPC relies on a rapid expansion of a humidified air sample to cool the air and create a region of supersaturation. No heaters or coolers are required. In contrast to these earlier counters, the particle number concentrations are derived from direct, automated measurement of individual particles exiting the particle cell during the expansion. The corresponding air volume is derived from the change in the cell pressure and its volume, derived from mechanical drawings. This yields a “first principles” or “direct” measurement rather than an inferred concentration from the scattering signal from the cloud of particles formed.

Laboratory testing against a 2-nm versatile water CPC that was checked against an aerosol electrometer yielded detection efficiency values, with no calibration factors applied, of 70% detection efficiency at 5 nm, increasing to 95% at 20 nm. Field testing shows excellent data recovery (100% for testing reported here), and precision of 3.5% for collocated units. When compared to a benchtop water CPC with ~5 nm cut-off size, the data are correlated with R2=0.98 and a regression slope of 1.1.

Ambiguity exists for concentrations in excess of about 3×104 cm−3. In laboratory tests with 34 nm ammonium sulfate and dioctyl sebacate aerosols the response decreased relative to the versatile water CPC, with a 30% drop at 105 cm−3. This high-concentration roll-off is not seen in ambient aerosol measurements, perhaps because the particle size distribution at high concentrations shifts towards smaller particle sizes that are more readily detected by the cCPC than the benchtop units used as reference. Further ambient aerosol measurements in diverse environmental conditions, in conjunction with ultrafine particle size distribution measurements, are needed to address the high concentration response.

Funding

This work was supported by the National Institutes of Health Grant R44-ES031458

Footnotes

Disclaimer

Aerosol Dynamics Inc. has received a US patent on the underlying concepts of the community condensation particle counter, with foreign patents pending, and has granted a license to Particles Plus Inc.

References

  1. Asbach C, Todea AM and Kaminski H, 2024. Evaluation of a Partector Pro for atmospheric particle number size distribution and number concentration measurements at an urban background site. Aerosol Research, 2(1), pp.1–12., 10.5194/ar-2-1-2024. [DOI] [Google Scholar]
  2. Bekö G, Weschler CJ, Wierzbicka A, Karottki DG, Toftum J, Loft S and Clausen G, 2013. Ultrafine particles: exposure and source apportionment in 56 Danish homes. Environmental science & technology, 47(18), pp.10240–10248. [DOI] [PubMed] [Google Scholar]
  3. Bergmann ML, Andersen ZJ, Massling A, Kindler PA, Loft S, Amini H, Cole-Hunter T, Guo Y, Maric M, Nordstrøm C and Taghavi M, 2023. Short-term exposure to ultrafine particles and mortality and hospital admissions due to respiratory and cardiovascular diseases in Copenhagen, Denmark. Environmental Pollution, 336, p.122396. [DOI] [PubMed] [Google Scholar]
  4. Bodhaine B and Murphy M 1980. Calibration of an automatic condensation nuclei counter at the South Pole. Journal of aerosol science, 11(3), 305–312. [Google Scholar]
  5. Cassee FR, Héroux ME, Gerlofs-Nijland ME and Kelly FJ, 2013. Particulate matter beyond mass: recent health evidence on the role of fractions, chemical constituents and sources of emission. Inhalation toxicology, 25(14), pp.802–812. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. European Parliament. 2024. Briefing: EU Legislation in Progress Revision of EU air quality legislation: Setting a zero pollution objective for air
  7. Fierz M, Houle C, Steigmeier P and Burtscher HJAS, 2011. Design, calibration, and field performance of a miniature diffusion size classifier. Aerosol Science and Technology, 45(1), pp.1–10. [Google Scholar]
  8. Fierz M, Meier D, Steigmeier P and Burtscher H, 2014. Aerosol measurement by induced currents. Aerosol Science and Technology, 48(4), pp.350–357. [Google Scholar]
  9. Gani S, Chambliss SE, Messier KP, Lunden MM and Apte JS, 2021. Spatiotemporal profiles of ultrafine particles differ from other traffic-related air pollutants: Lessons from long-term measurements at fixed sites and mobile monitoring. Environmental Science: Atmospheres, 1(7), pp.558–568. 10.1039/D1EA00058F [DOI] [Google Scholar]
  10. HEI Review Panel on Ultrafine Particles. 2013. Understanding the Health Effects of Ambient Ultrafine Particles. HEI Perspectives 3. 2013, Health Effects Institute: Boston, MA. https://www.healtheffects.org/system/files/Perspectives3.pdf [Google Scholar]
  11. Hering SV, Lewis GS, Spielman SR, Eiguren-Fernandez A, Kreisberg NM, Kuang C and Attoui M, 2017. Detection near 1-nm with a laminar-flow, water-based condensation particle counter. Aerosol Science and Technology, 51(3), pp.354–362. [Google Scholar]
  12. Hering SV, Lewis GS, Spielman SR and Eiguren-Fernandez A, 2019. A MAGIC concept for self-sustained, water-based, ultrafine particle counting. Aerosol Science and Technology, 53(1), pp.63–72. [Google Scholar]
  13. Isaxon C, Gudmundsson A, Nordin EZ, Lönnblad L, Dahl A, Wieslander G, Bohgard M and Wierzbicka A, 2015. Contribution of indoor-generated particles to residential exposure. Atmospheric Environment, 106, pp.458–466. [Google Scholar]
  14. Kangasluoma J, Hering S, Picard D, Lewis G, Enroth J, Korhonen F, Kulmala M, Sellegri K, Attoui M and Petäjä T, 2017. Characterization of three new condensation particle counters for sub-3 nm particle detection during the Helsinki CPC workshop: the ADI versatile water CPC, TSI 3777 nano enhancer and boosted TSI 3010. Atmospheric Measurement Techniques, 10(6), pp.2271–2281. [Google Scholar]
  15. Marval J and Tronville P, 2022. Ultrafine particles: A review about their health effects, presence, generation, and measurement in indoor environments. Building and Environment, 216, p.108992. [Google Scholar]
  16. Marra J, Voetz M and Kiesling HJ, 2010. Monitor for detecting and assessing exposure to airborne nanoparticles. Journal of Nanoparticle Research, 12, pp.21–37. [Google Scholar]
  17. Mazaheri M, Lin W, Clifford S, Yue D, Zhai Y, Xu M, Rizza V and Morawska L, 2019. Characteristics of school children’s personal exposure to ultrafine particles in Heshan, Pearl River Delta, China–A pilot study. Environment international, 132, p.105134. [DOI] [PubMed] [Google Scholar]
  18. McMurry PH 2000. The history of condensation nucleus counters. Aerosol Science & Technology, 33(4), 297–322. [Google Scholar]
  19. Mueller MD, Hasenfratz D, Saukh O, Fierz M and Hueglin C, 2016. Statistical modelling of particle number concentration in Zurich at high spatio-temporal resolution utilizing data from a mobile sensor network. Atmospheric Environment, 126, pp.171–181. [Google Scholar]
  20. Ng NL, Dillner AM, Bahreini R, Russell AG, de La Beaujardière J, Flynn J, Gentner DR, Griffin R, Hawkins LN, Jimenez JL and Mao J, 2022, December. Atmospheric Science and Chemistry mEasurement NeTwork (ASCENT): A new ground-based high time-resolution air quality monitoring network. In AGU Fall Meeting Abstracts (Vol. 2022, pp. A55R–1360). [Google Scholar]
  21. e Oliveira JRDC, Base LH, de Abreu LC, Ferreira Filho C, Ferreira C and Morawska L, 2019. Ultrafine particles and children’s health: Literature review. Paediatric respiratory reviews, 32, pp.73–81. [DOI] [PubMed] [Google Scholar]
  22. Schraufnagel DE, 2020. The health effects of ultrafine particles. Experimental & molecular medicine, 52(3), pp.311–317. [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. Seinfeld JH and Pandis SN, 2016. Atmospheric chemistry and physics: from air pollution to climate change. John Wiley & Sons., 3rd Edition, ISBN: 978–1-118–94740-1 [Google Scholar]
  24. Skala GF, 1963. A New Instrument for the Continuous Measurement of Condensation Nuclei. Analytical Chemistry, 35(6), pp.702–706. [Google Scholar]
  25. Sun J, Birmili W, Hermann M, Tuch T, Weinhold K, Spindler G, Schladitz A, Bastian SA, Löschau G, Cyrys J and Gu J, 2019. Variability of black carbon mass concentrations, sub-micrometer particle number concentrations and size distributions: results of the German Ultrafine Aerosol Network ranging from city street to High Alpine locations. Atmospheric Environment, 202, pp.256–268. [Google Scholar]
  26. Swanson B, Stevens R, Clagett W, Good N, McMeeking G, Patterson K, Boyer K, Gomez O, Good K, and Carter E 2023. Indoor Air Quality and Ventilation Patterns in Diverse Occupational Facilities across Colorado, and Outdoor Pollution Effects. Presented at the Annual Meeting of the American Association for Aerosol Research. [Google Scholar]
  27. Wang X, Myers P, Claassen M, Chow J, Watson J, Hering S, Eiguren Fernandez A, Pariseau D, Timmons D, McMullen B, Petersen C and Vega F 2024. Field Evaluation of an Affordable Community Condensation Particle Counters, Optical Particle Counters, and a Low-Cost Particle Sensor. Submitted to Aerosol Science and Technology. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES