Table 8.
Technical Specifications for Recording Electronic Implants.
| Neural Dust | FF-WINeR | WiPRoC | BioBolt ECoG | Microscale Opto-Electronically Transduced Electrode (MOTE) | ENGINI | |
|---|---|---|---|---|---|---|
| Publication Identifier | [182] | [185] | [212] | [210] | [186] | [214] |
| Clinical Readiness | Academic Research | Academic Research | Academic Research | Academic Research | Academic Research | Academic Research |
| Transmission Type | Ultrasound | RF | RF | Galvanic | Optical | RF, Near-field |
| External Coil Size (Diameter) | none | 45 mm (diameter) | 24 mm | none | none | ― |
| Repeater Coil | none | 32 mm (diameter) | none | none | none | yes |
| On-Chip Coil Material | PVT | AWG wire coil | ― | ― | AlGaAs Photovoltaic | Au |
| On-Chip Coil Size (mm) | 0.5625 | 1.2 (diameter) | 3 × 3 | none | none | 3.5 × 3.5 |
| Powering Frequnecy | 1.85 MHz | 60 MHz | 144 MHz | none | 545 THz | 433 MHz |
| Down Link Rate (kbps) | 0.5 Mbps | ― | 15 | 10 | ― | 205 |
| Up Link Rate (kbps) | ― | ― | ― | ― | ― | ― |
| Transmission Scheme | amplitude shift keying | on-off keying | analog-shift keying | frequency shift keying | pulse-position modulation | frequency shift keying |
| Device Size (um) † | 800 × 3000 × 1000 | 1000 × 1000 | 3000 × 3000 | ≈ 11,000 × 11,000 × 11,000 | 250 × 57 | 4000 (diameter), 1000 (thickness) |
| CMOS Tech (um) | custom | 0.13 | 0.18 | 0.25 | 0.18 | 0.35 |
| Chip Power | 0.12 | 93 | 160 | 365 | 1 | 92 |
| On Board Battery | none | none | none | Li ion | none | none |
| Working Distance (cm) | 0.88 | 1.6 | 1 | direct contact | 0.6 | ― |
| Testing | Tested recording in rat gastrocnemius muscle and sciatic nerve, comparing to wires ground truth. | Tested in ex-vivo dissected lamb head (w/o bone and skin) | Bench testing only. | Tested in non-human primate. | Bench testing only. Successful recovery of a pre-recorded neural signal. | Anti-buckling insertion and system tested in Agarose Gel. |
Note: “―” represents specification not indicated