Table 8.

Technical Specifications for Recording Electronic Implants.

Neural Dust FF-WINeR WiPRoC BioBolt ECoG Microscale Opto-Electronically Transduced Electrode (MOTE) ENGINI
Publication Identifier [182] [185] [212] [210] [186] [214]
Clinical Readiness Academic Research Academic Research Academic Research Academic Research Academic Research Academic Research
Transmission Type Ultrasound RF RF Galvanic Optical RF, Near-field
External Coil Size (Diameter) none 45 mm (diameter) 24 mm none none
Repeater Coil none 32 mm (diameter) none none none yes
On-Chip Coil Material PVT AWG wire coil AlGaAs Photovoltaic Au
On-Chip Coil Size (mm) 0.5625 1.2 (diameter) 3 × 3 none none 3.5 × 3.5
Powering Frequnecy 1.85 MHz 60 MHz 144 MHz none 545 THz 433 MHz
Down Link Rate (kbps) 0.5 Mbps 15 10 205
Up Link Rate (kbps)
Transmission Scheme amplitude shift keying on-off keying analog-shift keying frequency shift keying pulse-position modulation frequency shift keying
Device Size (um) † 800 × 3000 × 1000 1000 × 1000 3000 × 3000 ≈ 11,000 × 11,000 × 11,000 250 × 57 4000 (diameter),
1000 (thickness)
CMOS Tech (um) custom 0.13 0.18 0.25 0.18 0.35
Chip Power 0.12 93 160 365 1 92
On Board Battery none none none Li ion none none
Working Distance (cm) 0.88 1.6 1 direct contact 0.6
Testing Tested recording in rat gastrocnemius muscle and sciatic nerve, comparing to wires ground truth. Tested in ex-vivo dissected lamb head (w/o bone and skin) Bench testing only. Tested in non-human primate. Bench testing only. Successful recovery of a pre-recorded neural signal. Anti-buckling insertion and system tested in Agarose Gel.

Note: “―” represents specification not indicated