TABLE 1.

Agreement between the oxacillin E-test, the ATB Staph, the Rapid ATB Staph, and the Vitek systems and the reference methods (PCR amplification of the mecA gene and oxacillin agar screening) for S. aureus and coagulase-negative staphylococcal isolates

Reference method Isolate No. of strains with concordant results (% agreement) for indicated system
E-test ATB Staph Rapid ATB Staph Vitek
PCR of mecA S. aureus 61 (95.3) 57 (89.0) 49 (76.6) 61 (95.3)
Coagulase-negative staphylococci 51 (67.1) 44 (57.9) a 61 (80.3)
Agar screening S. aureus 63 (98.4) 59 (92.2) 51 (79.7) 63 (98.4)
Coagulase-negative staphylococci 55 (72.4) 48 (63.2) 65 (85.5)
a