. 1998 Jan;36(1):94–99. doi: 10.1128/jcm.36.1.94-99.1998

TABLE 3.

Performance of EIA/DFA versus EIA/PCR and EIA/LCR for true-negative and true-positive specimen results

Testa No. of samples that were:
Sensitivity (%) Specificity (%) Positive predictive value (%) Negative predictive value (%) Agreement with true results (%)
True positives False negatives True negatives False positives
EIA/PCRb
 DFA 119 18 105 0 87 100 100 93 93
 PCR 129 8 105 0 94 100 100 93 97
EIA/LCRc
 DFA 103 7 125 0 94 100 100 94 97
 LCR 109 1 115 10 99 92 92 99 95
a

Test to which EIA/DFA was compared. 

b

Number of true-positive specimens, 137 (n = 242). 

c

Number of true-positive specimens, 110 (n = 235).