. 2005 Oct;43(10):5312–5315. doi: 10.1128/JCM.43.10.5312-5315.2005

TABLE 3.

Results of the parallel testing algorithm OraQuick-Determine-Instant-Screena

Rapid-test algorithm results Double EIA-Western blot algorithm results
No. of samples positive No. of samples negative Total no.
Positive 474 33 507
Negative 0 2,993 2,993
Total 474 3,026 3,500
a

Test order is given left to right. A total of 3,500 serum samples were randomly selected from the remaining panel of 14,500 samples (1,000 had been selected for the first phase of the study), with a resultant HIV-1 seroprevalence of 13.5%. Results of the parallel rapid-test algorithm are compared with those of serial EIA and confirmatory Western blotting with the resulting performance characteristics: sensitivity, 100% (95% CI, 100%); specificity, 98.9% (95% CI, 98.6 to 99.3); PPV, 93.5% (95% CI, 92.7 to 94.3); NPV, 100% (95% CI, 100%).