. 1998 Jan;36(1):311–314. doi: 10.1128/jcm.36.1.311-314.1998

TABLE 2.

Differences in log10 RNA concentration among estimates based on VQA or kit standards for determining nominal copy numbera

Kit comparison Kit standard-based difference
VQA standard-based difference
Mean Median SD P value No.b Mean Median SD P value No.b
Monitor vs ES bDNA 0.38 0.37 0.28 <0.001 42 0.04 0.03 0.32 0.42 42
Monitor vs NASBA-QT −0.18 −0.15 0.29 <0.001 45 0.02 0.03 0.39 0.76 45
NASBA-QT vs ES bDNA 0.57 0.56 0.39 <0.001 37 0.01 0.10 0.48 0.91 37
Monitor vs ES bDNA (all) 0.32 0.36 0.36 <0.001 311 0.04 0.04 0.37 0.11 293
Monitor vs ES bDNA (placebo) 0.31 0.34 0.38 <0.001 159 0.03 0.03 0.43 0.37 152
Monitor vs ES bDNA (zidovudine) 0.32 0.37 0.33 <0.001 152 0.04 0.04 0.30 0.13 141
a

The upper section represents the subset of samples used for a three-way comparison, while the lower section is the entire data set obtained with a two-way comparison. The P value is that obtained by testing the null hypothesis that the average difference was zero. 

b

Number of specimens.