Skip to main content
Quantitative Imaging in Medicine and Surgery logoLink to Quantitative Imaging in Medicine and Surgery
. 2025 Sep 22;15(10):9325–9337. doi: 10.21037/qims-2025-900

Feasibility of finger-pulse triggering as a substitute for electrocardiogram triggering in cardiac magnetic resonance T1 mapping: a prospective comparative study

Ziyi Pan 1,#, Zhaoxia Yang 2,#, Jinyang Wen 1, Lingping Ran 1, Xianghu Yan 1, Yi Luo 1, Lu Huang 1, Liming Xia 1, Dazhong Tang 1,
PMCID: PMC12514587  PMID: 41081235

Abstract

Background

Electrocardiogram (ECG) triggering in cardiac magnetic resonance (MR) faces limitations with high-field systems (≥3.0T) due to magnetohydrodynamic (MHD) artifacts. This study aimed to evaluate the feasibility of using finger-pulse triggering for cardiac MR T1 mapping by comparing image quality and native T1 quantification with conventional ECG triggering.

Methods

Patients who underwent cardiac MR examination between March and April 2024 were prospectively and consecutively enrolled in the study. All the patients underwent identical pre-contrast T1 mapping with both ECG- and pulse-triggering acquisitions using a modified Look-Locker inversion-recovery (MOLLI) 5(3)3 sequence, covering short-axis views at basal, mid, and apical slices of the left ventricle. Three radiologists independently evaluated image quality using the Likert scale (range, 1–5). Two radiologists manually delineated myocardial regions of interest (ROIs) on native T1 maps to quantify segmental native T1 values. Paired t-tests or Wilcoxon signed-rank tests were used to compare the characteristics derived from the two triggering images. Bland-Altman plots, Kendall’s W test, and intraclass correlation coefficients (ICCs) were used for the agreement analysis.

Results

A total of 15 participants were included in the study (mean age: 41±19 years; 60% male). ECG and finger-pulse triggering demonstrated equivalent diagnostic image quality (median scores: 4.0 vs. 4.0, P=0.655), with excellent interobserver agreement (ECG: Kendall’s W =0.861, P<0.001; pulse: Kendall’s W =0.838, P=0.001). There were no statistically significant differences in the native T1 values between the two methods (all P>0.05). Bland-Altman plots revealed that the differences between the two triggering methods for native T1 values mostly fell within the 95% confidence interval. Both triggering modalities demonstrated good to excellent interobserver and intraobserver agreement (ICC range, 0.768–0.936).

Conclusions

Finger-pulse triggering demonstrates comparable accuracy and reliability to ECG triggering in cardiac MR T1 mapping, offering a viable clinical alternative for patients with ECG distortion or triggering failure.

Keywords: Cardiac magnetic resonance (cardiac MR), T1 mapping, electrocardiogram triggering (ECG triggering), pulse triggering, magnetohydrodynamic effect (MHD effect)

Introduction

Cardiac magnetic resonance (MR) has emerged as the gold standard non-invasive imaging tool for comprehensively evaluating cardiac anatomy, function, and tissue characteristics (1). Advanced myocardial mapping techniques, particularly T1 mapping, enable quantitative tissue characterization through the analysis of T1 relaxation time variations, facilitating the detection of fibrosis, edema, amyloidosis, and lipid accumulation (2). This capability positions T1 mapping as a pivotal tool for early diagnosis and precision medicine in cardiovascular disease.

T1 mapping relies on co-registered images acquired at sequential time points during T1 recovery (3). Most protocols employ balanced steady-state free precession sequences, with the modified Look-Locker inversion-recovery (MOLLI) technique being the predominant method for myocardial T1 quantification (4,5). Recent advancements, including improved MOLLI, saturation recovery single-shot acquisition, along with their variants, have further enhanced clinical applicability of T1 mapping (5). Notably, the utility of T1 mapping extends beyond myocardial assessment. For example, cardiac MR-derived liver T1 mapping has been shown to be significantly correlated with right ventricular function and myocardial fibrosis in dilated cardiomyopathy, and thus has potential prognostic value (6).

Despite its diagnostic value, cardiac MR faces inherent challenges related to cardiac and respiratory motion. Current cardiac gating predominantly uses electrocardiogram (ECG) R-wave triggering to synchronize data acquisition with the cardiac cycle (7). However, ECG signals are susceptible to interference from three primary sources: gradient interference, radiofrequency (RF) interference, and the magnetohydrodynamic (MHD) effect (8,9). Gradient switching and changes in RF pulses induce voltage fluctuations in ECG leads, which distort ECG signals (8,10). Many studies have developed methods to generate synchronous signals to eliminate these artifacts (11,12). One study modeled gradient-induced interference as a linear time-invariant system and estimated its impulse response to precisely remove artifacts without signal loss (11). Another study employed parallel transmission with multi-channel coils arrays, incorporating RF shimming, k-space subsampling, self-calibration, and specific absorption rate constraints to optimize RF pulses. This strategy improved B1 (RF field) homogeneity, reduced system-related errors, and effectively suppressed RF-induced artifacts (12).

The MHD effect arises from the interaction between conductive blood flow and the static magnetic field (B0), generating voltages and currents perpendicular to both B0 and flow direction, which are superimposed on the ECG signal (13). This phenomenon is most pronounced during early systole, when rapid ventricular ejection directs blood into the aortic arch (14). The MHD-induced currents predominantly distort T-wave morphology, often causing T-wave widening or elevation exceeding the amplitude of the R wave (8,15). These alterations lead to the false identification of T waves as R waves, compromising acquisition triggering accuracy in cardiac MR (9). The magnitude of the MHD effect is strongly correlated with the magnetic field strength. Significant ECG distortions have been documented at 3T (16), with these effects becoming more pronounced in ultra-high field systems (8). Current mitigation strategies include vector ECG-based triggering (17) and independent component analysis algorithms (14). However, technical limitations and implementation complexity have hindered their widespread clinical adoption.

When ECG triggering fails, peripheral pulse triggering is a valuable alternative. Photoplethysmography (PPG)-based systems detect blood volume changes via optical absorption/scattering in fingertip capillaries (18). The PPG waveform’s systolic upslope is correlated with ventricular contraction, while its diastolic decay reflects relaxation, enabling cardiac cycle synchronization (19). Moreover, the wave peak of the finger pulse is relatively broad and minimally affected by the magnetic fields, resulting in improved stability (19). Clinical validation demonstrates that finger-pulse triggering achieves comparable accuracy to ECG triggering for left ventricular (LV) functional assessment in 3T cardiac MR (20), suggesting its potential as a useful alternative triggering modality.

Despite these advancements, no prior investigations have evaluated the feasibility of finger-pulse triggering for T1 mapping. This study conducted a head-to-head comparison of pre-contrast T1 mapping using ECG and finger-pulse triggering to evaluate the technical feasibility of pulse-triggering T1 mapping, assess the agreement between the two triggering methods, and explore clinical scenarios suitable for pulse triggering. We present this article in accordance with the STROBE reporting checklist (available at https://qims.amegroups.com/article/view/10.21037/qims-2025-900/rc).

Methods

Study population

This study was conducted in accordance with the Declaration of Helsinki and its subsequent amendments. The study was approved by the Institutional Review Board of Tongji Hospital (IRB No. TJ-IRB20230914). Written informed consent was obtained from all participants prior to enrollment.

A total of 18 consecutive patients were prospectively enrolled in the study between March and April 2024. The inclusion criteria were as follows: (I) a clinical indication for cardiac MR examination; (II) an adequate breath-holding capacity (i.e., an ability to follow breathing instructions and maintain a stable breath-hold for ≥15 seconds); and (III) the successful acquisition of both ECG- and pulse-triggered pre-contrast T1 maps. The exclusion criteria were as follows: (I) arrhythmia (e.g., atrial fibrillation) or a rapid heart rate (>100 beats/min); and/or (II) a peripheral circulation disorder. After the inclusion and exclusion criteria were applied, a total of 15 patients were included in the study.

Cardiac MR image acquisition

All patients underwent cardiac MR examination on a 3.0T magnetic resonance scanner (MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany), with an 18-channel body coil, a 32-channel spine-array coil, and prospective ECG gating and finger-pulse triggering. Before the scan, all the patients were instructed to hold their breath after exhalation as directed. Scanning was performed in the head-advanced supine position. Four magnetic resonance imaging (MRI)-compatible electrodes were placed according to the manufacturer’s recommended positions. ECG signals from leads I, II, and III were acquired via a non-magnetic amplifier, and the lead with the highest R-wave amplitude, lowest T-wave amplitude, and fewest artifacts was selected to trigger cardiac MR data acquisition. The finger pulse clip was placed on the left or right index finger, and triggered based on the highest pulse wave. The pulse-triggered acquisition compensated for pulse transit time (PTT) delay through the real-time adjustment of the acquisition window, applying a 200–250 ms temporal offset relative to the ECG R wave to ensure diastolic phase sampling. When the other parameters were fixed, the same slice localization was selected, and the ECG trigger and pulse trigger were performed in the basal, mid, and apical slices, and heart rate and respiratory stability were monitored during the process.

Based on a MOLLI sequence, pre-contrast T1 mapping images were acquired at LV short-axis basal, mid, and apical slices using a 5(3)3 acquisition protocol (21). Under this two-inversion protocol, images were acquired for 5 heartbeats following the first inversion, after which there was a 3-second recovery period, and images were then acquired for 3 heartbeats following the second inversion. Typical scanning parameters were as follows: slice thickness =5 mm, slice spacing =2 mm, repetition time (TR) =3.2 ms, echo time =1.2 ms, flip angle =35°, field of view (FOV) =360×360 mm2, FOV phase =100%, matrix size =256×192, base resolution =256, phase resolution =75%.

Image quality evaluation

The T1 mapping images were independently assessed by three radiologists (D.T., with 10 years of experience; Z.Y., with 7 years of experience; Z.P., with 2 years of experience), who were blinded to the patient clinical information, diagnosis, and acquisition methods (i.e., ECG triggering and finger-pulse triggering). All images were presented in randomized order and assessed using a Syngo MMWP VE40B workstation (MAGNETOM Skyra, Siemens Healthcare). The ordinal scores of the three radiologists were averaged.

Subjective general image quality was rated on a five-point Likert scale as follows: 5= excellent image quality, interpretable with minimal to no artifacts; 4= good image quality, interpretable with mild artifacts; 3= adequate image quality, interpretable with moderate artifacts; 2= fair image quality, interpretable with obvious artifacts; and 1= poor image quality, uninterpretable images (22).

Native T1 quantification

All the cardiac MR T1 mapping images were analyzed by two experienced radiologists blinded to the patients’ clinical history using CVI42 (version 5.14.0, Circle Cardiovascular Imaging, Calgary, Canada), and reviewed by one observer twice at an interval of 2 weeks to avoid bias. Following the American Heart Association (AHA) 17-segment model (23), we adopted a modified 16-segment approach that excluded the apical cap for LV segmentation. The radiologists performed semi-automatic contouring of the endocardial and epicardial borders, with manual adjustments as needed, followed by the manual placement of region of interest (ROI) markers for each segment. The myocardial ROI was delineated as far as possible to avoid the endocardium and epicardium, and the location of the measured segments and the size of the ROI were consistent. Notably, ROIs smaller than 20 pixels were avoided to reduce partial volume effects, and ensure measurement accuracy and reproducibility (1).

Statistical analysis

All the data analyses were performed using SPSS (version 29.0.1.0; IBM, Armonk, NY, USA). Graphs were generated using GraphPad Prism (version 10.2.0; GraphPad Software, San Diego, CA, USA) and MATLAB (version R2021b; MathWorks, Natick, MA, USA). The Shapiro-Wilk test was used to assess the normality of the continuous variables. The normally distributed variables are presented as the mean ± standard deviation, and were compared using the paired t-test. The non-normally distributed variables are presented as the median with interquartile range (IQR), and were compared using the Wilcoxon signed-rank test. The categorical variables are presented as the absolute number and percentage. A Bland-Altman analysis was performed to evaluate the agreements and differences of native T1 values of different triggering methods. Interobserver agreements in the subjective general image quality assessment were evaluated with Kendall’s W, which was categorized as poor (<0.40), fair (0.25-0.60), good (0.61–0.80), or excellent (>0.80). Intra- and interobserver agreements in native T1 mapping quantification were evaluated using the intraclass correlation coefficient (ICC), and were labeled as poor (<0.50), fair (0.50–0.75), good (0.76–0.90), or excellent (>0.90). A two-tailed P value <0.05 was considered statistically significant.

Results

Participant characteristics

A total of 18 participants met the inclusion criteria; however, three patients were excluded because of arrhythmia (n=1), rapid heart rate (>100 bpm, n=1), and peripheral circulation disorders (n=1), respectively. Thus, the final study sample comprised 15 participants (mean age: 41±19 years; 60% male). The clinical diagnoses of the participants included hypertrophic cardiomyopathy (n=6), myocarditis (n=3), systemic disease cardiac involvement (n=2), hypertensive heart disease (n=1), cardiac tumor (n=1), and no clinical finding reported (n=2). The demographic characteristics of the patients were collected from clinical records and are summarized in Table 1.

Table 1. Patients’ clinical characteristics (n=15).

Characteristics Values
Male 9 [60]
Age (years) 41±19 (10–66)
Height (cm) 162.6±8.3 (140–175)
Weight (kg) 63.8±12.9 (41–82)
Body surface area (m2) 1.69±0.21 (1.26–2.00)
Heart rate (bpm) 74.6±11.5 (55–92)
Diagnosis
   Hypertrophic cardiomyopathy 6
   Myocarditis 3
   Cardiac involvement of systemic disease 2
   Hypertensive heart disease 1
   Cardiac tumor 1
   Negative finding 2

Continuous normally distributed data are expressed as the mean ± standard deviation (range), and categorical data are expressed as the n [percentage] or n.

Image quality evaluation

The T1 mapping sequences were acquired using MOLLI 5(3)3 mode. ECG triggering and finger-pulse triggering showed comparable image quality, with identical median Likert scores [4.0 (IQR, 3.7–4.7) vs. 4.0 (IQR, 3.8–4.7), P=0.655], indicating no significant difference between the two methods. Representative T1 mapping images of ECG triggering and finger-pulse triggering are shown in Figure 1.

Figure 1.

Figure 1

Image examples of pre-contrast T1 mapping images acquired by ECG triggering and pulse triggering in three short-axis slices (basal, mid, and apical). ECG, electrocardiogram.

Native T1 quantification

There was no significant difference in the native T1 quantification between the ECG-triggering and pulse-triggering imaging of the 16 segments (Table 2) (all P>0.05). The mean native T1 values of the participants in three short-axis slices are displayed as bullseye plots in Figure 2, and the quantification results between the ECG triggering and finger-pulse triggering in all 16 myocardial segments were similar. Figure 3 shows the comparable native T1 values between both triggering methods through boxplot visualization, with the statistical analysis confirming no significant difference (P=0.081). The Bland-Altman scatter plots (Figure 4) illustrated that the differences between the two triggering methods for native T1 values mostly fell within the 95% confidence interval. The differences in basal anteroseptal (bias =0.27 ms) and basal anterolateral (bias =0.53 ms) were closer to zero compared to those in the other segments.

Table 2. Image quality assessment and native T1 quantification between ECG-triggering and finger pulse-triggering cardiac MR T1 mapping.

Parameter ECG triggering Pulse triggering P value
General image quality 4.0 (3.7, 4.7) 4.0 (3.8, 4.7) 0.655
Native T1 value quantification
   Basal
    Anterior 1,269.87±28.61 1,271.40±27.48 0.474
    Anteroseptal 1,290.40±36.04 1,290.13±31.56 0.911
    Inferoseptal 1,285.93±33.83 1,290.53±28.23 0.236
    Inferior 1,274.40±31.40 1,279.07±29.41 0.112
    Inferolateral 1,271.27±35.28 1,269.13±37.95 0.519
    Anterolateral 1,262.33±19.22 1,261.80±32.45 0.934
   Mid
    Anterior 1,269.20±43.07 1,271.53±36.96 0.637
    Anteroseptal 1,302.73±37.47 1,305.93±38.97 0.258
    Inferoseptal 1,296.67±28.67 1,295.27±29.98 0.590
    Inferior 1,262.20±36.94 1,265.53±38.36 0.332
    Inferolateral 1,250.93±29.96 1,254.53±34.39 0.310
    Anterolateral 1,283.53±32.52 1,282.27±30.83 0.401
   Apical
    Anterior 1,271.40±45.39 1,273.93±28.47 0.465
    Septal 1,291.13±28.47 1,292.33±30.26 0.531
    Inferior 1,279.27±25.71 1,280.67±28.13 0.543
    Lateral 1,278.47±42.85 1,279.47±38.43 0.652

Continuous normally distributed data are expressed as median (Q1, Q3) or mean± standard deviation. ECG, electrocardiogram; MR, magnetic resonance.

Figure 2.

Figure 2

The mean native T1 values of participants in three short-axis slices are displayed as bullseye plots (AHA-16-segment-model). AHA, American Heart Association; ECG, electrocardiogram.

Figure 3.

Figure 3

Comparison of native T1 values between ECG- and pulse-triggering acquisitions. The boxplots display the median (central line), interquartile range (box), and full data range (whiskers) from end-diastolic measurements. ns: (P=0.081). ECG, electrocardiogram; ns, not significant.

Figure 4.

Figure 4

Bland-Altman plots for native T1 values of 16 segments showing agreement between ECG triggering and pulse triggering. The red solid lines represent the average difference, and the black dotted lines represent the 95% confidence interval. ECG, electrocardiogram; SD, standard deviation.

Agreements of scoring and quantification

The interobserver agreement of the general image quality evaluation of the two triggering methods was excellent (ECG: Kendall’s W =0.861; pulse: Kendall’s W =0.838) (Table 3). The native T1 value of ECG triggering and pulse triggering demonstrated good to excellent inter- and intraobserver agreement (all ICC values >0.75). The detailed inter- and intraobserver agreements of the native T1 quantification results are shown in Table 4.

Table 3. Interobserver agreement for the image general quality scores for ECG and pulse triggering.

Group Image quality scores Kendall’s W P value
ECG triggering 0.861 <0.001
   Observer A 4 (3, 5)
   Observer B 4 (4, 5)
   Observer C 4 (4, 5)
Pulse triggering 0.838 0.001
   Observer A 4 (3, 5)
   Observer B 4 (4, 5)
   Observer C 4 (4, 5)

Continuous non-normally distributed data are expressed as median (Q1, Q3). ECG, electrocardiogram.

Table 4. Intra- and interobserver reproducibility for native T1 measurements assessed by ICCs with 95% confidence intervals.

Parameter ECG triggering Pulse triggering
Intraobserver Interobserver Intraobserver Interobserver
Basal
   Anterior 0.919 (0.777–0.972) 0.929 (0.803–0.976) 0.860 (0.633–0.951) 0.845 (0.600–0.945)
   Anteroseptal 0.903 (0.735–0.966) 0.918 (0.774–0.972) 0.909 (0.751–0.968) 0.900 (0.730–0.965)
   Inferoseptal 0.936 (0.820–0.978) 0.934 (0.817–0.978) 0.828 (0.561–0.939) 0.802 (0.507–0.929)
   Inferior 0.856 (0.624–0.949) 0.839 (0.586–0.943) 0.891 (0.706–0.962) 0.870 (0.656–0.954)
   Inferolateral 0.824 (0.553–0.937) 0.813 (0.530–0.933) 0.866 (0.648–0.953) 0.895 (0.717–0.964)
   Anterolateral 0.860 (0.633–0.951) 0.815 (0.533–0.934) 0.843 (0.596–0.945) 0.829 (0.564–0.939)
Mid
   Anterior 0.907 (0.746–0.968) 0.903 (0.735–0.966) 0.870 (0.657–0.954) 0.857 (0.626–0.949)
   Anteroseptal 0.913 (0.761–0.970) 0.807 (0.517–0.931) 0.912 (0.759–0.970) 0.908 (0.748–0.968)
   Inferoseptal 0.796 (0.495–0.927) 0.830 (0.567–0.940) 0.852 (0.616–0.948) 0.828 (0.561–0.939)
   Inferior 0.839 (0.586–0.943) 0.810 (0.523–0.932) 0.850 (0.611–0.947) 0.886 (0.695–0.960)
   Inferolateral 0.827 (0.560–0.938) 0.840 (0.589–0.943) 0.873 (0.664–0.995) 0.812 (0.528–0.933)
   Anterolateral 0.874 (0.665–0.956) 0.785 (0.473–0.922) 0.837 (0.581–0.942) 0.857 (0.626–0.949)
Apical
   Anterior 0.915 (0.766–0.971) 0.776 (0.453–0.919) 0.898 (0.725–0.965) 0.913 (0.762–0.970)
   Septal 0.814 (0.531–0.933) 0.783 (0.467–0.921) 0.894 (0.714–0.963) 0.878 (0.676–0.957)
   Inferior 0.830 (0.56–0.939) 0.841 (0.591–0.944) 0.840 (0.588–0.943) 0.768 (0.440–0.916)
   Lateral 0.877 (0.673–0.957) 0.853 (0.617–0.948) 0.859 (0.630–0.950) 0.847 (0.604–0.946)

ECG, electrocardiogram; ICCs, intraclass correlation coefficients.

Discussion

This prospective method-comparison study systematically evaluated the performance of pulse-triggering versus ECG-triggering T1-mapping protocols at 3.0T cardiac MR. Our study revealed three main findings: (I) pulse triggering T1 mapping was feasible; (II) the image quality between pulse- and ECG-triggering protocols was equivalent (Likert score 4, indicating “good” diagnostic quality); (III) there were no significant differences in myocardial native T1 quantification between the two triggering modalities.

Cardiac MR is recognized as the gold standard non-invasive imaging tool in cardiovascular disease and continues to expand its utility in clinical practice (1). Conventional cardiac MR acquisitions rely on cardiac gating to capture information at end-diastole or end-systole (7). However, as clinical MRI systems increasingly adopt higher field strengths, elevated electromagnetic interference susceptibility and amplified MHD effects impair ECG signal fidelity, creating an urgent demand for alternative cardiac triggering schemes (24). As previous studies have shown, many methods could replace ECG triggering, such as pulse triggering (20), Doppler ultrasound triggering (25), and acoustic triggering (26).

Peripheral pulse triggering offers a cost-effective solution through PPG. Both the rise and fall of the PPG signal are associated with cardiac contraction and cardiac relaxation, making it suitable for cardiac triggering (19). Unlike ECG signals that suffer from T-wave amplification artifacts due to MHD effects at 3T and beyond, pulse-triggering acquisition eliminates electromagnetic interference risks. This fundamental physical distinction ensures reliable cardiac cycle detection regardless of the static magnetic field strength. In addition, it requires only a fingertip sensor, which simplifies the procedure, and is especially suitable for children or agitated patients. Previous studies have validated the efficacy of pulse triggering for LV functional assessment at 3T cardiac MR (20). We compared images acquired using finger-pulse triggering with those obtained using ECG triggering, and found that finger-pulse triggering can be used as an alternative and effective triggering method in T1 mapping.

The PPG waveform originates from cardiac mechanical activity, where LV contraction propels blood flow through the arterial system, generating peripheral pulse waves detectable via optical sensing. The PPG waveform lags behind the R wave in ECG (27). The time difference between the R peak in the ECG signal and the corresponding peak in the peripheral signal is called the PTT (28). In clinical practice, after recalculating and dynamically adjusting the TR acquisition period, pulse-triggered T1 mapping is still acquired at end-diastole. The present results show that the errors are negligible when using the PPG method during resting conditions (18,27). Compared with ultrasound triggering and acoustic triggering, pulse triggering is simpler to operate and cheaper, and the R-wave delay time is shorter. Lang et al. identified pulse-triggered signals as an alternative to ECG-triggered signals in gated myocardial perfusion single-photon emission computed tomography imaging, noting that the differences between the methods were not clinically significant (29). The results of our analysis showed that there was no significant difference in the quantitative analysis of T1 mapping obtained by pulse triggering and ECG triggering.

Admittedly, pulse triggering shares an apparent common drawback of conventional ECG, ultrasound, and acoustic triggering (i.e., extra hardware is required for signal detection and processing). To overcome the constraint of using ancillary hardware, various self-gating (SG) methods have been proposed and shown to be feasible in cardiac MR (30-32), especially for cardiac arrhythmias and children’s cardiac imaging. SG is a technique in which changes in the raw MR data over time are used to form a signal that may be used to retrospectively synchronize data to the cardiac cycle. The SG signal is typically produced by sampling the center of k-space repeatedly during cine data acquisition by interleaving the collection of non-phase encoded k-space lines (kpe =0) (31). SG methods might be the most promising approach, especially for high-field magnets. However, due to its high algorithm complexity and low cost-effectiveness ratio, the popularization rate of SG in clinical practice is still limited.

Recent advances in ECG-free cardiac MRI techniques have enabled free-breathing myocardial tissue characterization. One study developed a free-breathing myocardial T1-mapping technique using the inversion-recovery radial fast low-angle shot and calibration-free motion-resolved model-based reconstruction, which had good accuracy and reproducibility (33). Another study introduced a free-running T1* mapping sequence that enables the non-ECG quantification of myocardial T1* changes during physiological exercise (34). Emerging free-running approaches now allow simultaneous T1/T2 mapping and cine imaging during free breathing, incorporating methods such as non-rigid cardiac motion correction for three-dimensional whole-heart imaging (35) and motion-corrected MR fingerprinting (36). These techniques eliminate the need for external synchronization (e.g., ECG gating), while maintaining diagnostic image quality during unconstrained respiratory and physiological motion conditions. Although these free-running technologies hold great promise, they remain limited by technical and computational challenges, and are still in the exploratory stages. This underscores the value of pulse triggering as a practical and interim solution.

In addition, in the presence of arrhythmias and heart rate variability, T1 mapping image quality is commonly sub-optimal and cannot meet clinical diagnosis requirements (37). Heart rate significantly affects MOLLI acquisition, particularly at lower rates (37,38). Compared with 3T or conventional MOLLI, 1.5T and MOLLI 5 s(3 s)3 s may better mitigate heart rate, respiratory, or cardiac motion influences (38,39). The saturation pulse-prepared heart rate-independent inversion-recovery sequence has been proposed as a heart rate-independent alternative, offering greater stability and shorter breath-hold requirements compared to conventional MOLLI sequences (37). T1 maps are obtained by acquiring multiple sample points on a longitudinal magnetization recovery curve after magnetization preparation (5). In bradycardia patients, the R-R interval is prolonged and the distribution of sampling points is sparse, which affects the accuracy of the exponential fitting and may increase the error of T1 mapping (38). Moreover, as the total sampling time increases, the difficulty of breath-holding may introduce respiratory artifacts. Hence, performing T1 mapping during systolic phase has recently been proposed in tachyarrhythmias (40), and has also demonstrated improved quality in patients with atrial fibrillation (41) and bradycardia (38). This study provided preliminary verification of the feasibility and stability of pulse triggering in cardiac MR T1 mapping. In the future, we will collect more data to explore the potential application of pulse-triggering T1 mapping in arrhythmias.

Limitations

The study had several limitations. First, this study was conducted at a single center, on a single magnet, from a single supplier, at a single field strength, and with a small sample size. Thus, the generalizability of the conclusions cannot be guaranteed, and future studies involving magnets of various field strengths from different suppliers and larger populations need to be conducted at multiple centers. Second, hematocrit values were not available for some patients in this study, so only native T1 values of myocardial tissue were validated (post-contrast T1 values were not validated), and no comprehensive assessment of extracellular volume (ECV) was conducted. Future studies will continue to determine whether pulse triggering can be used for the assessment of ECV. Third, our cohort excluded patients with severe peripheral vascular diseases or arrhythmias. We preliminarily validated the feasibility and stability of pulse triggering in T1-mapping scanning. Further studies need to be performed to evaluate whether pulse triggering is also robust in patients with arrhythmias and peripheral vascular diseases.

Conclusions

This prospective comparative investigation evaluated the interchangeability of peripheral pulse triggering versus conventional ECG triggering for T1-mapping acquisition on 3T cardiac MR, showing comparable T1 quantification accuracy. These findings support the use of pulse-triggering T1 mapping as a viable alternative and practical strategy for patients when ECG triggering is unfeasible. This method may also aid in follow-up and clinical interventions for patients in the future.

Supplementary

The article’s supplementary files as

qims-15-10-9325-rc.pdf (471.7KB, pdf)
DOI: 10.21037/qims-2025-900
qims-15-10-9325-coif.pdf (302.7KB, pdf)
DOI: 10.21037/qims-2025-900

Acknowledgments

None.

Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. This study was conducted in accordance with the Declaration of Helsinki and its subsequent amendments. The study was approved by the Institutional Review Board of Tongji Hospital (IRB No. TJ-IRB20230914). Written informed consent was obtained from all participants prior to enrollment.

Footnotes

Reporting Checklist: The authors have completed the STROBE reporting checklist. Available at https://qims.amegroups.com/article/view/10.21037/qims-2025-900/rc

Funding: None.

Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at https://qims.amegroups.com/article/view/10.21037/qims-2025-900/coif). The authors have no conflicts of interest to declare.

Data Sharing Statement

Available at https://qims.amegroups.com/article/view/10.21037/qims-2025-900/dss

qims-15-10-9325-dss.pdf (73.3KB, pdf)
DOI: 10.21037/qims-2025-900

References

  • 1.Schulz-Menger J, Bluemke DA, Bremerich J, Flamm SD, Fogel MA, Friedrich MG, Kim RJ, von Knobelsdorff-Brenkenhoff F, Kramer CM, Pennell DJ, Plein S, Nagel E. Standardized image interpretation and post-processing in cardiovascular magnetic resonance - 2020 update : Society for Cardiovascular Magnetic Resonance (SCMR): Board of Trustees Task Force on Standardized Post-Processing. J Cardiovasc Magn Reson 2020;22:19. 10.1186/s12968-020-00610-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Xu Z, Li W, Wang J, Wang F, Sun B, Xiang S, Luo X, Meng Y, Wang X, Wang X, Song J, Zhang M, Xu D, Zhou X, Ju Z, Sun J, Han Y, Chen Y. Reference ranges of myocardial T1 and T2 mapping in healthy Chinese adults: a multicenter 3T cardiovascular magnetic resonance study. J Cardiovasc Magn Reson 2023;25:64. 10.1186/s12968-023-00974-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Moon JC, Messroghli DR, Kellman P, Piechnik SK, Robson MD, Ugander M, Gatehouse PD, Arai AE, Friedrich MG, Neubauer S, Schulz-Menger J, Schelbert EB, Society for Cardiovascular Magnetic Resonance Imaging ; Cardiovascular Magnetic Resonance Working Group of the European Society of Cardiology. Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR Working Group of the European Society of Cardiology consensus statement. J Cardiovasc Magn Reson 2013;15:92. 10.1186/1532-429X-15-92 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Messroghli DR, Greiser A, Fröhlich M, Dietz R, Schulz-Menger J. Optimization and validation of a fully-integrated pulse sequence for modified look-locker inversion-recovery (MOLLI) T1 mapping of the heart. J Magn Reson Imaging 2007;26:1081-6. 10.1002/jmri.21119 [DOI] [PubMed] [Google Scholar]
  • 5.Taylor AJ, Salerno M, Dharmakumar R, Jerosch-Herold M. T1 Mapping: Basic Techniques and Clinical Applications. JACC Cardiovasc Imaging 2016;9:67-81. 10.1016/j.jcmg.2015.11.005 [DOI] [PubMed] [Google Scholar]
  • 6.Stojanovska J, Feng L, Gilani N. Editorial for "Liver T1 Mapping Derived From Cardiac Magnetic Resonance Imaging: A Potential Prognostic Marker in Idiopathic Dilated Cardiomyopathy". J Magn Reson Imaging 2024;60:686-7. 10.1002/jmri.29229 [DOI] [PubMed] [Google Scholar]
  • 7.Pan Y, Varghese J, Tong MS, Yildiz VO, Azzu A, Gatehouse P, Wage R, Nielles-Vallespin S, Pennell DJ, Jin N, Bacher M, Hayes C, Speier P, Simonetti OP. Two-center validation of Pilot Tone based cardiac triggering of a comprehensive cardiovascular magnetic resonance examination. Int J Cardiovasc Imaging 2024;40:261-73. 10.1007/s10554-023-03002-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Stäb D, Roessler J, O'Brien K, Hamilton-Craig C, Barth M. ECG Triggering in Ultra-High Field Cardiovascular MRI. Tomography 2016;2:167-74. 10.18383/j.tom.2016.00193 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Abi-Abdallah D, Robin V, Drochon A, Fokapu O. Alterations in human ECG due to the MagnetoHydroDynamic effect: a method for accurate R peak detection in the presence of high MHD artifacts. Annu Int Conf IEEE Eng Med Biol Soc 2007;2007:1842-5. 10.1109/IEMBS.2007.4352673 [DOI] [PubMed] [Google Scholar]
  • 10.Dos Reis JE, Soullié P, Oster J, Palmero Soler E, Petitmangin G, Felblinger J, Odille F. Reconstruction of the 12-lead ECG using a novel MR-compatible ECG sensor network. Magn Reson Med 2019;82:1929-45. 10.1002/mrm.27854 [DOI] [PubMed] [Google Scholar]
  • 11.Felblinger J, Slotboom J, Kreis R, Jung B, Boesch C. Restoration of electrophysiological signals distorted by inductive effects of magnetic field gradients during MR sequences. Magn Reson Med 1999;41:715-21. 10.1002/(sici)1522-2594(199904)41:4<715::aid-mrm9>3.0.co;2-7 [DOI] [PubMed] [Google Scholar]
  • 12.Katscher U, Börnert P. Parallel magnetic resonance imaging. Neurotherapeutics 2007;4:499-510. 10.1016/j.nurt.2007.04.011 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Togawa T, Okai O, Oshima M. Observation of blood flow E.M.F. in externally applied strong magnetic field by surface electrodes. Med Biol Eng 1967;5:169-70. 10.1007/BF02474505 [DOI] [PubMed] [Google Scholar]
  • 14.Krug JW, Rose G, Clifford GD, Oster J. ECG-based gating in ultra high field cardiovascular magnetic resonance using an independent component analysis approach. J Cardiovasc Magn Reson 2013;15:104. 10.1186/1532-429X-15-104 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Gupta A, Weeks AR, Richie SM. Simulation of elevated T-waves of an ECG inside a static magnetic field (MRI). IEEE Trans Biomed Eng 2008;55:1890-6. 10.1109/TBME.2008.919868 [DOI] [PubMed] [Google Scholar]
  • 16.Dietrich O, Reiser MF, Schoenberg SO. Artifacts in 3-T MRI: physical background and reduction strategies. Eur J Radiol 2008;65:29-35. 10.1016/j.ejrad.2007.11.005 [DOI] [PubMed] [Google Scholar]
  • 17.Hamilton-Craig C, Stäeb D, Al Najjar A, O'Brien K, Crawford W, Fletcher S, Barth M, Galloway G. 7-Tesla Functional Cardiovascular MR Using Vectorcardiographic Triggering-Overcoming the Magnetohydrodynamic Effect. Tomography 2021;7:323-32. 10.3390/tomography7030029 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Lu G, Yang F, Taylor JA, Stein JF. A comparison of photoplethysmography and ECG recording to analyse heart rate variability in healthy subjects. J Med Eng Technol 2009;33:634-41. 10.3109/03091900903150998 [DOI] [PubMed] [Google Scholar]
  • 19.Allen J. Photoplethysmography and its application in clinical physiological measurement. Physiol Meas 2007;28:R1-39. 10.1088/0967-3334/28/3/R01 [DOI] [PubMed] [Google Scholar]
  • 20.Sievers B, Wiesner M, Kiria N, Speiser U, Schoen S, Strasser RH. Influence of the trigger technique on ventricular function measurements using 3-Tesla magnetic resonance imaging: comparison of ECG versus pulse wave triggering. Acta Radiol 2011;52:385-92. 10.1258/ar.2011.100505 [DOI] [PubMed] [Google Scholar]
  • 21.Raman FS, Kawel-Boehm N, Gai N, Freed M, Han J, Liu CY, Lima JA, Bluemke DA, Liu S. Modified look-locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners. J Cardiovasc Magn Reson 2013;15:64. 10.1186/1532-429X-15-64 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Zucker EJ, Sandino CM, Kino A, Lai P, Vasanawala SS. Free-breathing Accelerated Cardiac MRI Using Deep Learning: Validation in Children and Young Adults. Radiology 2021;300:539-48. 10.1148/radiol.2021202624 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS, American Heart Association Writing Group on Myocardial Segmentation and Registration for Cardiac Imaging . Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation 2002;105:539-42. 10.1161/hc0402.102975 [DOI] [PubMed] [Google Scholar]
  • 24.Becker M, Frauenrath T, Hezel F, Krombach GA, Kremer U, Koppers B, Butenweg C, Goemmel A, Utting JF, Schulz-Menger J, Niendorf T. Comparison of left ventricular function assessment using phonocardiogram- and electrocardiogram-triggered 2D SSFP CINE MR imaging at 1.5 T and 3.0 T. Eur Radiol 2010;20:1344-55. 10.1007/s00330-009-1676-z [DOI] [PubMed] [Google Scholar]
  • 25.Kording F, Ruprecht C, Schoennagel B, Fehrs K, Yamamura J, Adam G, Goebel J, Nassenstein K, Maderwald S, Quick HH, Kraff O. Doppler ultrasound triggering for cardiac MRI at 7T. Magn Reson Med 2018;80:239-47. 10.1002/mrm.27032 [DOI] [PubMed] [Google Scholar]
  • 26.Frauenrath T, Hezel F, Renz W, d'Orth Tde G, Dieringer M, von Knobelsdorff-Brenkenhoff F, Prothmann M, Schulz Menger J, Niendorf T. Acoustic cardiac triggering: a practical solution for synchronization and gating of cardiovascular magnetic resonance at 7 Tesla. J Cardiovasc Magn Reson 2010;12:67. 10.1186/1532-429X-12-67 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Selvaraj N, Jaryal A, Santhosh J, Deepak KK, Anand S. Assessment of heart rate variability derived from finger-tip photoplethysmography as compared to electrocardiography. J Med Eng Technol 2008;32:479-84. 10.1080/03091900701781317 [DOI] [PubMed] [Google Scholar]
  • 28.Gesche H, Grosskurth D, Küchler G, Patzak A. Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method. Eur J Appl Physiol 2012;112:309-15. 10.1007/s00421-011-1983-3 [DOI] [PubMed] [Google Scholar]
  • 29.Lang O, Trojanova H, Balon HR, Kunikova I, Bilwachs M, Penicka M, Kaminek M, Myslivecek M. Pulse wave as an alternate signal for data synchronization during gated myocardial perfusion SPECT imaging. Clin Nucl Med 2011;36:762-6. 10.1097/RLU.0b013e318217aec5 [DOI] [PubMed] [Google Scholar]
  • 30.Larson AC, White RD, Laub G, McVeigh ER, Li D, Simonetti OP. Self-gated cardiac cine MRI. Magn Reson Med 2004;51:93-102. 10.1002/mrm.10664 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Nijm GM, Sahakian AV, Swiryn S, Carr JC, Sheehan JJ, Larson AC. Comparison of self-gated cine MRI retrospective cardiac synchronization algorithms. J Magn Reson Imaging 2008;28:767-72. 10.1002/jmri.21514 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Piekarski E, Chitiboi T, Ramb R, Feng L, Axel L. Use of self-gated radial cardiovascular magnetic resonance to detect and classify arrhythmias (atrial fibrillation and premature ventricular contraction). J Cardiovasc Magn Reson 2016;18:83. 10.1186/s12968-016-0306-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Wang X, Rosenzweig S, Roeloffs V, Blumenthal M, Scholand N, Tan Z, Holme HCM, Unterberg-Buchwald C, Hinkel R, Uecker M. Free-breathing myocardial T(1) mapping using inversion-recovery radial FLASH and motion-resolved model-based reconstruction. Magn Reson Med 2023;89:1368-84. 10.1002/mrm.29521 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Guo R, Qi H, Amyar A, Cai X, Kucukseymen S, Haji-Valizadeh H, Rodriguez J, Paskavitz A, Pierce P, Goddu B, Thompson RB, Nezafat R. Quantification of changes in myocardial T(1) * values with exercise cardiac MRI using a free-breathing non-electrocardiograph radial imaging. Magn Reson Med 2022;88:1720-33. 10.1002/mrm.29346 [DOI] [PubMed] [Google Scholar]
  • 35.Phair A, Cruz G, Qi H, Botnar RM, Prieto C. Free-running 3D whole-heart T(1) and T(2) mapping and cine MRI using low-rank reconstruction with non-rigid cardiac motion correction. Magn Reson Med 2023;89:217-32. 10.1002/mrm.29449 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Jaubert O, Cruz G, Bustin A, Schneider T, Koken P, Doneva M, Rueckert D, Botnar RM, Prieto C. Free-running cardiac magnetic resonance fingerprinting: Joint T1/T2 map and Cine imaging. Magn Reson Imaging 2020;68:173-82. 10.1016/j.mri.2020.02.005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Weingärtner S, Akçakaya M, Basha T, Kissinger KV, Goddu B, Berg S, Manning WJ, Nezafat R. Combined saturation/inversion recovery sequences for improved evaluation of scar and diffuse fibrosis in patients with arrhythmia or heart rate variability. Magn Reson Med 2014;71:1024-34. 10.1002/mrm.24761 [DOI] [PubMed] [Google Scholar]
  • 38.Amano Y, Omori Y, Yanagisawa F, Ando C, Shinoda N, Suzuki Y, Yamamoto H, Matsumoto N. Relationship between Measurement Errors in Myocardial T(1) Mapping and Heart Rate. Magn Reson Med Sci 2020;19:345-50. 10.2463/mrms.mp.2019-0166 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Kellman P, Hansen MS. T1-mapping in the heart: accuracy and precision. J Cardiovasc Magn Reson 2014;16:2. 10.1186/1532-429X-16-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Ferreira VM, Wijesurendra RS, Liu A, Greiser A, Casadei B, Robson MD, Neubauer S, Piechnik SK. Systolic ShMOLLI myocardial T1-mapping for improved robustness to partial-volume effects and applications in tachyarrhythmias. J Cardiovasc Magn Reson 2015;17:77. 10.1186/s12968-015-0182-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Zhao L, Li S, Ma X, Greiser A, Zhang T, An J, Bai R, Dong J, Fan Z. Systolic MOLLI T1 mapping with heart-rate-dependent pulse sequence sampling scheme is feasible in patients with atrial fibrillation. J Cardiovasc Magn Reson 2016;18:13. 10.1186/s12968-016-0232-7 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

The article’s supplementary files as

qims-15-10-9325-rc.pdf (471.7KB, pdf)
DOI: 10.21037/qims-2025-900
qims-15-10-9325-coif.pdf (302.7KB, pdf)
DOI: 10.21037/qims-2025-900

Data Availability Statement

Available at https://qims.amegroups.com/article/view/10.21037/qims-2025-900/dss

qims-15-10-9325-dss.pdf (73.3KB, pdf)
DOI: 10.21037/qims-2025-900

Articles from Quantitative Imaging in Medicine and Surgery are provided here courtesy of AME Publications

RESOURCES