Skip to main content
. 2025 Oct 1;36(3):225–235.

Table 2:

Adherence with analytical performance specifications for comparators (C-APS).

Sample Bias Imprecision C-APS met2
Device n original Corrected1 n CV original final
Venous
INT 184 +3.0% +0.0% 2059 0.3% No Yes
YSI 1 64 -2.3% +0.1% 706 0.2% No Yes
YSI 2 55 -1.5% +0.0% 630 0.3% Yes Yes
YSI 3 65 -1.7% +0.0% 724 0.6% Yes Yes
Capillary
CNX (vs. INT) 142 +2.8% +0.1% 2061 1.6% No Yes
CNX (vs. YSI) 142 +3.8% +0.2% 2061 1.6% No Yes

As bias was corrected retrospectively, compliance with C-APS is given before and after retrospective correction.

CGM: continuous glucose monitoring system, FDA: Food and Drug Administration, BGMS: blood glucose monitoring system

CV: coefficient of variation, INT: Cobas Integra 400 plus, YSI: YSI 2300 STAT Plus, CNX: Contour Next

1 For venous samples on INT and the three YSIs, bias was retrospectively corrected based on a split-sample approach with a higher-order method. For capillary CNX samples, bias correction was implemented using a split-sample approach using corrected capillary INT or YSI results (as indicated in the table).

2 “Original” indicates application of C-APS to original bias and imprecision, whereas “final” indicates application of C-APS to retrospectively corrected bias and imprecision.