Table 4:
Recommendations of the IFCC WG-CGM regarding the comparator measurement procedure in CGM performance evaluations.
| Category | Recommendation |
|---|---|
| Comparator sampling frequency | Every 15 minutes for 6 to 8 hours [11] |
| Comparator glucose concentration distribution | At least 7.5% of all values each covering the following situations: [11] a) BG < 70 mg/dL b) BG > 300 mg/dL, c) BG ≥ 70 mg/dL, RoC < -1 mg/dL/min, BG < 70 mg/dL within 30 min at current RoC, and d) BG ≤ 300 mg/dL, RoC > +1.5 mg/dL/min, BG > 250 mg/dL within 30 min at current RoC |
| Comparator sample type | Capillary samples |
| Comparator bias | Bias ≤2.1% across measuring range vs. method (or material) of higher metrological order,[23] retrospective correction encouraged even if criterion is met |
| Comparator imprecision | CV ≤2.4% across measuring range [23] |
| Comparator replicates | ≥2, to assess and reduce imprecision [27] and to detect outliers in participant samples |
BG: blood glucose concentration, CGM: continuous glucose monitoring, CV: coefficient of Variation, IFCC WG-CGM: Working Group on Continuous Glucose Monitoring of the International Federation of Clinical Chemistry and Laboratory Medicine, RoC: rate of change