A Novel Method in PET Image Reconstruction Using MRI Anatomical Priors

M Mehdi Khalighi 1, Christina B Young 2, Matthew G Spangler-Bickell 3, Timothy W Deller 4, Floris Jansen 5, Dawn Holley 6, Hillary Vossler 7, Moss Zhao 8, Feliks Kogan 9, Gary Steinberg 10, Elizabeth Mormino 11, Michael Moseley 12, Greg Zaharchuk 13
PMCID: PMC12599855  NIHMSID: NIHMS2121274  PMID: 41221102

Abstract

The current spatial resolution of PET images is 3–4 mm for whole body PET/MR. Anatomical MR images with higher resolution and superior image quality have been used in PET reconstruction to improve the image quality and spatial resolution; however, mismatches between MR priors and actual tracer distribution can hinder accuracy. A novel PET reconstruction with MR priors, Magnetic Resonance-guided Block Sequential Regularized Expectation Maximum (MRgBSREM), that is robust to mismatches between anatomical priors and true activity distribution is proposed. This method is evaluated in diverse clinical settings using various tracers: 18F-florbetaben (FBB) in 373 subjects from a dementia study, 18F-FDG in a patient with chronic ischemic stroke, 18F-NaF in a knee study, and 15O-water in a patient with Moyamoya disease. Reconstruction using MRgBSREM visually improved both spatial resolution and image quality in all studies. In the 18F-florbetaben study, it mitigated white-matter spill-in into gray-matter as well as gray-matter spill over to the adjacent tissues, potentially leading to more accurate measurement of FBB uptake in the gray-matter. Visual assessment suggests that the proposed PET reconstruction enhances spatial resolution, which may contribute to improved diagnostic accuracy, while it displays robustness to mismatches between MR priors and true activity distribution.

Index Terms—: Anatomical Priors, Image Reconstruction, MRI, PET, PET/MRI

I. INTRODUCTION

Positron Emission Tomography (PET) is a highly sensitive quantitative molecular imaging tool increasingly used in oncology and neurodegenerative disorder staging. PET images, however, often suffer from noise and low spatial resolution. There are different strategies in PET reconstruction methods to control image noise. Ordered Subset Expectation Maximization (OSEM) [1] is one of the most common PET image reconstruction methods used in clinical practice. It has a superior image quality relative to analytical reconstruction methods; however, to control the image noise [2], it is stopped early, and post filtering is applied. An alternative set of PET reconstruction algorithms based on the Penalized Maximum-Likelihood (PML) method [3] apply a penalty function, which controls for image quality, and adds it to a cost function, which is minimized by iterative reconstruction, allowing users to control the trade-off between noise and image resolution by adjusting few parameters.

The spatial resolution of PET images is limited by several physical factors, including positron range, photon non-collinearity, and the finite size of the detector elements. The current spatial resolution of PET images is ~3–4mm for whole body PET/MR scanners. One approach to control image noise and increase PET image spatial resolution simultaneously is to use anatomical information from other imaging modalities, such as Computed Tomography or Magnetic Resonance Imaging, within the PET reconstruction algorithm [9–18]. These scanners can acquire anatomical images with both higher resolution and superior signal to noise ratio (SNR), which can be incorporated into the PET reconstruction to improve the image quality. These methods use the anatomical image as a priori knowledge to modify the penalty function of a PML reconstruction method. For instance, in Bowsher’s method [9], the local differences of each image voxel with its most similar neighbors from the anatomical image is calculated to update the penalty function. The anatomical information is generally applied in the image space, e.g., the intensity of neighboring voxels on anatomical images or the boundaries between different anatomical regions is used to update the penalty function. A recent study [19] has used convolutional neural networks (CNN) for deep learned joint PET-MR reconstruction.

The increased SNR gained by using anatomical priors can be traded off with either a shorter scan-time or a lower injected dose which are very desirable in clinical settings; however, two main problems have hindered the implementation and adaptation of these methods. The first problem is patient motion. A recent study [20] has shown that motion is prevalent in most PET/MR brain studies by analyzing more than 500 exams. Having good alignment between PET and anatomical images is vital for the accurate application of anatomical priors, and since motion can cause misalignment between the PET and anatomical images, accurate motion correction is very important. Thus, the first step before using anatomical priors is to perform PET motion correction, for example the data-driven approach demonstrated in [21]. The second problem is the mismatch between anatomical priors and the true activity distribution, which can lead to either suppression of a true PET signal or introduction of artifactual signal. Several studies have tried to tackle this problem with different techniques [10–14]. In this study we have developed a novel method [22–23] based on a feature approach to incorporate anatomical priors in PET image reconstruction, where the location of a voxel or its neighbors does not play a role directly, and instead both functional and anatomical images are used to construct the feature space. This will allow us to control the mismatch between anatomical priors and the true activity distribution. The application of this method is demonstrated in a stroke patient, and its effectiveness is investigated in an amyloid PET research study. The ability of this method to handle mismatches between anatomical images and the true activity distribution is tested and compared with Bowsher’s method in 3 different exams including two 18F-florbetaben brain exams and one 18F-NaF knee exam. Furthermore, it is shown in an 15O-water study that Bowsher’s method can provide superior spill-over correction where no mismatch exists between anatomical priors and the true activity distribution.

II. Framework

One type of penalized maximum-likelihood algorithm is based on the Relative Difference Prior (RDP) [4] and uses a Block Sequential Regularized Expectation Maximization (BSREM) [5–8] optimizer. The RDP applies activity-dependent smoothing to control noise at higher iterations and suppresses image noise in low-activity background regions. This relative difference penalty is calculated based on the difference between each voxel activity and its neighbors [4]. The correlation between each voxel and its neighbors is used to control image noise and that allows the utilization of a higher number of iterations to improve convergence and therefore image resolution. A weighting factor called β controls the smoothing. As β is increased, the image becomes smoother and will have a lower resolution.

We have applied the same framework to incorporate the anatomical priors into the image reconstruction by applying an additional penalty, which is calculated based on the relative difference between each voxel activity and other similar voxels found by anatomical priors. In this approach the first step is to find all similar voxels for each voxel. To determine similar voxels, we construct a multi-dimensional feature space using co-registered multi-parametric functional or anatomical images (e.g., MR images) and an initial PET reconstruction (PET seed) generated using the conventional OSEM algorithm. The feature space allows us to find voxels with similar characteristics across different modalities, even if their spatial locations are not directly adjacent. This approach helps control potential mismatches between anatomical priors and the true activity distribution. For example, voxels with similar anatomical intensities but significantly different uptake values will not be considered as neighbors in the feature space, preventing incorrect anatomical information from influencing the final PET image.

Fig. 1 shows an example of the feature space consisting of MRI brain T1 and T2 weighted images in addition to the PET seed. In this figure all images are normalized to have the same mean value over the brain tissue. Each blue dot represents a voxel from the image, and its location is determined by its normalized intensity values on the T1-weighted, T2-weighted, and PET seed images. The red sphere, which is enlarged for visualization purposes, shows the similarity neighborhood around a point located at its center. By including the PET seed images in the feature space, we can control the mismatches between the anatomical priors and the true activity distribution. For instance, if two voxels with similar anatomical intensities have a relatively large uptake difference, they will not be adjacent in the feature space and therefore the similarity in anatomical images will not impact the final reconstructed PET image. On the other hand, if the uptake difference between these voxels is relatively small (e.g., due to spill-in or spill-over effects), these voxels will fall into each other’s neighborhood in the feature space and the noise between them will be controlled by the penalized maximum-likelihood reconstruction method.

Fig. 1.

Fig. 1.

The feature space consisted of normalized 3D T1 IR-SPGR MR images on the x-axis, 3D CUBE T2-FLAIR MR images on the y-axis, and the PET seed images on the z-axis. The red sphere (enlarged for visual clarity) shows the similarity neighborhood around the voxel at its center. The voxels within this similarity neighborhood are used for penalized PET reconstruction in the same way that neighboring voxels have previously used for penalized PET recon.

In this proposed method any number of anatomical images can be used, and the PET seed is an under-converged image which can be reconstructed quickly by an OSEM method before the BSREM reconstruction gets started. The RDP penalty calculated based on similar voxels is combined with the penalty based on neighboring voxels in the original BSREM method, resulting in a new algorithm we call MR-guided BSREM (MRgBSREM). This algorithm leverages both spatial and anatomical information to control noise and improve image quality. The combined penalty function used in this method is shown in (1).

R(X)=−β∑j=1Nv∑k∈NjWjWkXj−Xk2Xj+Xk+γXj−Xk−βm∑j=1Nv∑l∈SjWjsWlsXj−Xl2Xj+Xl+γmXj−Xl (1)

where Nv is the total number of voxels in the image volume, Nj is the set of neighbors of voxel j derived from the image space, Sj is the set of neighbors of voxel j derived from the feature space, Xj is the intensity of voxel j, Xk is the intensity of voxel j’s neighbors in the image space, Xl is the intensity of voxel j’s neighbors in the feature space, wj, wk and wjs, wks are penalty weights for voxels in the image space and in the feature space, respectively. γ and γm control the level of edge-preservation.

Fig. 2 illustrates the concept of a similarity neighborhood. It depicts a set of T1-weighted and T2-weighted MR images alongside their corresponding PET seed image. A voxel (j) within the gray matter is outlined with a yellow border. Ten of its neighbors in the feature space (l ∈ Sj) are marked by red borders. While the voxels adjacent to voxel j represent its neighbors in the image space (k ∈ Nj), they are not explicitly marked in this figure.

Fig. 2.

Fig. 2.

Shows a set of normalized (a) T1-weighted and (b) T2-weighted MR images, along with their corresponding (c) PET seed image, used to construct the feature space as illustrated in Fig. 1. A voxel within the gray matter is highlighted with a yellow border, and ten of its neighbors in the feature space are marked with red borders. The region within the blue border is magnified and displayed across six different slices. Note that other neighboring voxels in the feature space are not visible in this view as they fall outside the region marked by the blue borders.

The penalty function consists of two main terms that address different aspects of the reconstruction process. The first term penalizes image noise. A weighting factor, β, controls the strength of this penalty. This term ensures that neighboring voxels in the reconstructed image have similar intensities, reducing noise. The second term penalizes the difference between voxels identified as similar by MR priors in the feature space. This term promotes consistency between the reconstructed image and the information provided by the MR priors. Another weighting factor, βm, controls the strength of this term. Overall, the combined penalty function helps achieve two goals: reducing image noise and ensuring adherence to the prior information from the MR data.

As seen in (1), the penalty function for similar voxels in the feature space mirrors the penalty function used for adjacent voxels in the image space. The similarity neighborhood in the feature space is shown by the red sphere in Fig. 1. The radius of this neighborhood is dynamically adjusted for each voxel to include only the 26 most similar voxels, analogous to the 26-voxel neighborhood used in the image space (i.e., 3×3 cube around a central voxel).

One of the main keys of MRgBSREM lies in its decoupling of PET information for mismatch handling from the MR priors used for anatomical guidance. Previous methods (e.g., [14]) that incorporate both PET and MR data in their iterative reconstruction pipelines, utilize the PET data itself at each iteration. In contrast, MRgBSREM employs a pre-processed PET image (PET seed), which is independent of the MR priors. This PET seed is then integrated with the MR data to form a feature space. This feature space guides the reconstruction process while simultaneously controlling for potential mismatches between the reconstructed PET image and the underlying anatomical information provided by the MR priors.

III. Methods

The studies used in this work were approved by Stanford’s Institutional Review Board (IRB) and all subjects provided written informed consent prior to the imaging session. All studies were performed on a 3.0T PET/MR scanner (SIGNA, GE HealthCare, Waukesha, WI). All PET images were reconstructed with MR-based attenuation correction, scatter correction, randoms correction and dead-time correction. For all brain studies a Zero TE (ZTE) sequence was used for attenuation correction [24]. Wherever mentioned, the MR 3D T1 IR-SPGR and 3D CUBE T2-FLAIR images were acquired based on Alzheimer’s Disease Neuroimaging Initiative 3 (ADNI3) protocol.

A subject with chronic ischemic stroke was injected with 300 MBq of 18F-FDG and underwent a 60-minute brain scan 45 minutes post injection. 3D T1 IR-FSPGR and 3D CUBE T2-FLAIR images were acquired simultaneously with PET. Motion was measured by co-registering PET short frames [21]. The PET images were reconstructed using 1 mm isotropic voxels over a 256×256×247 matrix. The PET images were reconstructed in list mode with 1) OSEM method with 28 subsets and 4 iterations using 3 mm full-width half-maximum (FWHM) Gaussian spatial filter and standard Z-filter, 2) same OSEM method as in (1) but with motion correction, 3) BSREM method with motion correction and β=75, and 4) MRgBSREM method with β=50 and βm=25 using 3D T1 IR-FSPGR and 3D CUBE T2-FLAIR images as priors.

In the amyloid study, 373 subjects of whom half were cognitively impaired, were injected with 300 MBq of 18F-florbetaben and underwent a 40-min brain scan 70 min after injection and a 20-min frame between 90–110 minutes after the injection was chosen for PET reconstruction. 3D T1 IR-FSPGR and 3D CUBE T2-FLAIR images were acquired simultaneously with PET. PET images were first reconstructed using our clinical standard which is OSEM (28 subsets and 4 iterations using 3 mm FWHM spatial filter and standard Z-filter over 300 mm field-of-view (FOV) on a 256×256×89 matrix with 2.78 mm slice thickness, and with point-spread function (PSF) correction). Subject motion was measured by co-reregistering PET short frames [25], with 0.5 million events in each frame, and PET images were reconstructed using MRgBSREM method with 1 mm isotropic voxels over 256×256×247 matrix, β=50 and βm=25 using 3D T1 IR-FSPGR and 3D CUBE T2-FLAIR images as priors. Standardized uptake value ratios (SUVRs) were calculated using a global cortical composite ROI divided by the whole cerebellum for both OSEM and MRgBSREM recons. The global cortical composite ROI was created with the ADNI protocol and includes volume-weighted frontal regions, anterior/posterior cingulate, lateral parietal, and lateral temporal regions. Half of the subjects were selected randomly and their PET images reconstructed with MRgBSREM method were reviewed by a trained reader for quality assurance. To investigate the robustness of MRgBSREM to mismatches between anatomical priors and the true activity distribution and comparing it with the Bowsher’s method [9], the 3D T1 IR-FSPGR and 3D CUBE T2-FLAIR images of one subject from the amyloid study were mixed in the sagittal plane to create artificial line artifacts in the axial plane. This artificially created set of MR images was then used as priors in both MRgBSREM and Bowsher’s method for comparison using the same β and βm values.

A subject was injected with 90 MBq of 18F-NaF after 8 flights of stair ascent/descent exercise and underwent a 30 min PET/MR knee exam immediately after the injection. Bilateral sagittal quantitative Double-Echo Steady-State (qDESS) MR images [26] were acquired with the following parameters: Repetition Time (TR) 18ms, first Echo Time (TE1) 6ms, second Echo Time (TE2) 30ms, 240 slices with 1.5 mm thickness, FOV 16 cm, Flip Angle (FA) 20 degree, and acquisition matrix 320×320. The PET images were reconstructed with 1 mm isotropic voxels using the BSREM method with β=75, MRgBSREM method with β=50 and βm=25 using the first echo of MR DESS images and 2-point Dixon fat/water images in the MR Attenuation Correction (MRAC) series as priors. For comparison, the PET images were also reconstructed with Bowsher’s method with matched parameters and priors.

A patient with Moyamoya disease, characterized by the narrowing or blockage of the carotid arteries in the brain, was injected with 850 MBq of 15O-water and underwent a 10 min scan immediately after the injection. MR angiogram images were acquired using a sagittal 3D TOF SPGR sequence with hypersense acceleration factor of 2, slice thickness 1.2 mm, TR 24ms, TE 3.4ms FOV 22 cm, FA 15 degree and acquisition matrix 400×400. To visualize arterial perfusion in the neck and brain, an early 20-second frame was selected for reconstruction using different methods: standard BSREM method with β=75, MRgBSREM method with βm=50 and βm=25 using the MR angiography images as priors, and Bowsher’s method with matched parameters to MRgBSREM method for comparison.

IV. RESULTS

Fig. 1 shows an example of a 3D feature space used in MRgBSREM, which consists of normalized 3D T1 IR-SPGR MR voxels on the x-axis, 3D CUBE T2-FLAIR MR voxels on the y-axis and the PET seed voxels on the z-axis. There is no limit to the number of anatomical series which can be used in MRgBSREM method. The red sphere is a visual representation of the similarity neighborhood around a specific point in feature space and is enlarged for clarity. Voxels within this sphere are considered similar to the central voxel in terms of their anatomical and functional characteristics. These voxels are used for penalized PET reconstruction in the same way as the neighboring voxels in image space but with a different weighting factor, i.e., βm vs. β.

Fig 2 provides a visual demonstration of how similar voxels, identified by defining neighboring voxels within the feature space, appear on the actual MRI and PET images. A voxel located within the gray matter, highlighted by a yellow border, is selected. Ten similar voxels across six slices are identified and marked with red borders. These red-bordered voxels exhibit similar intensity values to the original, yellow-bordered voxel across all three images: T1-weighted and T2-weighted MR images, as well as the PET seed image reconstructed using the OSEM method with 28 subsets and 1 iteration. Other neighboring voxels in the feature space are not visible in this view as they fall outside the region marked by the blue borders.

Fig. 3 showcases the application of the MRgBSREM method in a patient with chronic ischemic stroke and high motion during the PET scan. The motion estimate in the top left corner reveals significant head movement (16 mm for virtual tracker points located 70 mm away from the center of the brain in anterior and posterior directions), exceeding typical thresholds [21]. Fig. 3(a) and 3(b) present the anatomical MR priors used for subsequent reconstructions. 3D T1 IR-FSPGR MR images in 3(a) provide gray/white matter differentiation and the 3D CUBE T2-FLAIR images in 3(b) highlight cerebrospinal fluid and anatomical details for improved spatial localization. Fig. 3(c) shows the standard clinical PET image reconstructed with OSEM, exhibiting limitations in both spatial resolution and image quality. 3(d) shows motion-corrected PET image obtained using OSEM with motion information, leading to improved spatial resolution compared to 3(c). 3(e) shows the BSREM-reconstructed PET image with further enhanced spatial resolution and image Signal-to-Noise Ratio (SNR) compared to both 3(c) and 3(d). 3(f) is the final PET image reconstructed with MRgBSREM, incorporating both motion correction and MR priors. This image achieves the highest apparent spatial resolution among all methods, revealing fine anatomical details.

Fig. 3.

Fig. 3.

Images from a subject with chronic ischemic stroke, injected with 300 MBq of 18F-FDG who underwent a 60-minute brain scan 45 minutes post injection. Top left shows the patient motion during the exam. The displacement of anterior and posterior tracking points (70 mm away from the center of the brain) show that they have moved about 16 mm from their original position. (a) shows MRI 3D CUBE T2-FLAIR images, (b) shows MRI 3D T1 IR-FSPGR images, (c) shows PET images reconstructed with OSEM algorithm with 28 subsets and 3 iterations, (d) shows same PET recon as in (c) with motion correction, (e) shows PET images reconstructed with BSREM with β=75 and motion correction, and (f) shows PET images reconstructed with MRgBSREM with β=50 and βm=25 using MR priors shown in (a) and (b). The PET images in (c) to (f) show the evolution of PET images and how SNR and spatial resolution are improved by motion correction, penalized PET reconstruction, and MR-guided priors. Images in (f) reconstructed with MRgBSREM incorporating motion correction, penalized PET reconstruction and MR priors show the best SNR and spatial resolution.

Fig. 4 focuses on PET reconstruction methods applied to images from a single subject in an amyloid study with 2 negative and 1 positive visual reads of the standard OSEM reconstruction. This showcases the impact of incorporating anatomical MR priors to address the challenge of spill-in from white-matter affecting gray-matter uptake. Fig. 4(a) and 4(b) display anatomical MR priors used in the subsequent reconstructions. 3D T1 IR-FSPGR MR images shown in 4(a) provide contrast between gray-matter and white-matter, crucial for distinguishing the location of amyloid deposits, since amyloid scans are only considered positive if there is uptake in the cortex. 3D CUBE T2-FLAIR MR images shown in 4(b) highlight cerebrospinal fluid and anatomical details, aiding in spatial localization. Fig. 4(c) shows OSEM PET which is a standard clinical method, showcasing reasonable image detail but with limitations in spatial resolution. Fig. 4(d–g) present PET images reconstructed using MRgBSREM with β=75 and varying βm values of 0, 5, 15, and 25, respectively. These reconstructions utilized the MRI images depicted in Fig. 4(a) and 4(b) as anatomical priors. Fig. 4(d–g) demonstrate that progressively increasing the βm parameter from 0 to 25 enhances the spatial resolution and SNR of the PET images. This improvement is achieved by increasingly controlling the noise between similar voxels, identified with the aid of the MR priors. Clearly, Fig 4(g) demonstrates the sharpest image detail and highest spatial resolution. MRgBSREM successfully addresses the spill-in issue in gray-matter. White-matter spill-over, potentially masking or mimicking amyloid signal in the gray-matter, is effectively suppressed. This is evident in the cleaner delineation of cortical regions compared to OSEM in Fig 4(c) and BSREM in Fig 4(d). By resolving spill-in, MRgBSREM paves the way for more reliable detection and characterization of amyloid burden, potentially improving diagnostic accuracy and clinical decision-making.

Fig. 4.

Fig. 4.

Images from a subject injected with 300 MBq of 18F-florbetaben who underwent a 20 min exam, 90 min post injection. (a) MRI 3D T1 IR-FSPGR images. (b) 3D CUBE T2-FLAIR images. (c) PET images reconstructed with OSEM algorithm with 28 subsets, and 3 iterations. (d-g) PET images reconstructed with MRgBSREM with β=75 and varying βm values of 0, 5, 15 and 25 respectively. MR priors shown in (a) and (b) were utilized. Fig 4d–g show how increasing the βm parameter from 0 to 25, controls the influence of MR priors on the final PET image. The enhanced SNR and resolution shown in (g) helps clarify that the tracer uptake in gray-matter is minimal, which is the indicator for a negative amyloid study.

The effectiveness of MRgBSREM was assessed by quantitatively measuring global cortical SUVR in 373 18F-florbetaben exams. Fig. 5(a) shows a high correlation between the global SUVRs obtained using OSEM and MRgBSREM methods, indicating a strong linear relationship between the two reconstruction methods. The Bland-Altman plot shown in Fig. 5(b) further details the agreement between OSEM and MRgBSREM. A lower mean difference in global SUVR is observed with MRgBSREM compared to OSEM because of white-matter spill-in correction using the MRgBSREM reconstruction method. For those exams with low gray-matter uptake, MRgBSREM tends to measure a lower global SUVR compared to OSEM most likely due to its effective removal of spill-in from white-matter. However, as gray-matter uptake increases, in addition to white-matter spill-in correction, MRgBSREM method also corrected for gray-matter spill-over, which increases the uptake in gray-matter, therefore leading to higher global SUVR values compared to OSEM.

Fig. 5.

Fig. 5.

Assessing the effectiveness of MRgBSREM by quantitatively measuring global cortical SUVR in 373 18F-florbetaben exams. (a) shows a high correlation between the cortical global SUVRs obtained using OSEM and MRgBSREM methods. Each marker represents a single exam. This figure indicates a strong linear relationship between the two reconstruction methods. The Bland-Altman plot shown in (b) further details the agreement between OSEM and MRgBSREM. For those exams with low gray-matter uptake, MRgBSREM tends to measure a lower global SUVR compared to OSEM most likely due to its effective removal of spill-in from white-matter. However, as gray-matter uptake increases, MRgBSREM starts to recover the spill-over effect in gray-matter, leading to a higher global SUVR compared to OSEM.

Fig. 6 shows PET and MR images from an amyloid positive patient with 3 positive visual reads. The anatomical MR priors including 3D T1 IR-FSPGR and 3D CUBE T2-FLAIR are shown in Fig. 6(a) and 6(b), respectively. A clinical PET recon with OSEM method is shown in Fig. 6(d) and a PET recon with BSREM method is shown in Fig. 6(g) for comparison. As expected, BSREM offers superior spatial resolution compared to OSEM. Fig. 6(e) and 6(h) showcase PET reconstructions utilizing the same MR priors from 6(a) and 6(b), processed with MRgBSREM and Bowsher’s methods, respectively. Both methods employ matched reconstruction parameters for a direct comparison. A closer look at Fig. 6(e) and 6(h) reveals that Bowsher’s method appears to incorporate the MR priors more effectively. This manifests in higher spatial resolution and sharper image details compared to MRgBSREM. This difference arises from how each method handles potential mismatches between the MR priors and the actual amyloid distribution in the PET data. MRgBSREM’s reliance on a PET seed image within the feature space restricts its ability to fully capitalize on the MR priors but enables it to take this important factor into consideration rather than merely confining the PET signal to the anatomical boundaries. To further illustrate this point, Fig. 6(c) shows an artificially crafted MR prior image created by combining aspects of the T1 and T2 weighted images from 6(a) and 6(b). This “mismatched” prior deliberately introduces potential discrepancies between the anatomical information and the true amyloid distribution. Fig. 6(f) and 6(i) showcase PET reconstructions using Bowsher’s and MRgBSREM methods, respectively, but now incorporating the “mismatched” anatomical prior from 6(c). The line artifacts present in the “mismatched” prior in 6(c) noticeably affect the Bowsher’s method reconstruction in 6(i), leading to image degradation. In contrast, the MRgBSREM reconstruction in 6(f) maintains its robustness to this mismatch, preserving a clean and informative image despite the deliberately introduced discrepancy.

Fig. 6.

Fig. 6.

Images from a subject injected with 300 MBq of 18F-florbetaben and underwent a 20 min exam, 90 min post injection. (a) shows MRI 3D T1 IR-FSPGR images, (b) shows 3D CUBE T2-FLAIR images, (c) shows MR images from mixing T1 and T2 images in sagittal plane, (d) shows PET images reconstructed with OSEM algorithm with 28 subsets, 3 iterations and motion correction, (e) shows PET images reconstructed with MRgBSREM with β=50 and βm=25 using MR priors shown in (a) and (b), (f) shows PET images reconstructed with MRgBSREM with β=50 and βm=25 using MR priors shown in (c), (g) shows PET images reconstructed with BSREM with β=75 and motion correction, (h) shows PET images reconstructed with anatomy guided PET reconstruction (Bowsher’s method) with β=50 and βm= 25 using MR priors in (a) and (b), and (i) shows PET images reconstructed with anatomy guided PET reconstruction (Bowsher’s method) with β=50 and βm =25 using MR priors in (c). The PET images reconstructed with Bowsher’s method in (i) show the effect of line artifacts in MR priors in (c). Contrary to the Bowsher’s method, PET images reconstructed with MRgBSREM method in (f) show robustness to mismatches between PET and MR priors.

Fig. 7 compares the performance of MRgBSREM and Bowsher’s method in the presence of severe motion artifacts on a subject’s T1-weighted MR image. This subject was not able to hold still during the 3D T1 IR-FSPGR scan and the MR images appeared noisy and with severe motion artifact as shown in Fig 7(c). The PET images were reconstructed with motion correction using 7(a) OSEM algorithm with 28 subsets and 3 iterations and 7(b) BSREM algorithm with β=75. PET images were also reconstructed using 7(d) MRgBSREM method and 7(e) Bowsher’s method utilizing the MR priors shown in 7(c) and with matched parameters of β=50 and βm=25. While MRgBSREM demonstrates a degree of robustness to the poor quality of the MR priors in Fig. 7(c), exhibiting comparable image quality to the BSREM reconstruction in Fig. 7(b), it does not significantly improve image resolution or SNR due to the presence of severe motion artifacts in the MR images. In contrast, the PET image reconstructed using Bowsher’s method (Fig. 7(e)) exhibits lower SNR than the BSREM reconstruction and is also adversely affected by the motion artifacts present in the MR priors.

Fig. 7.

Fig. 7.

Imaging data from a subject injected with 300 MBq of 18F-florbetaben and undergoing a 20-minute PET scan 90 minutes post-injection. (a) PET images reconstructed using the OSEM algorithm with 28 subsets and 3 iterations (b) PET images reconstructed using the BSREM algorithm with β=75. (c) MRI 3D T1 IR-FSPGR images, exhibiting severe motion artifacts. (d) PET images reconstructed using MRgBSREM with β=50 and βm=25, utilizing MR priors from (c). (e) PET images reconstructed using anatomy-guided PET reconstruction (Bowsher’s method) with β=50 and βm=25, utilizing MR priors from (c). All PET images are reconstructed with motion correction. MRgBSREM demonstrates greater resilience to the motion artifacts present in the MR priors compared to Bowsher’s method. However, the presence of these artifacts limits the effectiveness of MRgBSREM in achieving improvements in image resolution and SNR.

Fig. 8 compares different PET reconstruction methods for a knee study examining 18F-NaF uptake after a bone-loading exercise. Fig. 8(a) shows the first echo of a sagittal bilateral qDESS MR series, providing anatomical detail of the knee. Both qDESS and 2-point Dixon fat/water MRAC images (not shown) [27] were used as priors in subsequent PET reconstructions. Fig. 8(b) presents the PET image reconstructed with the standard BSREM method after exercise. Note the clear uptake in the subchondral bone, highlighting areas of potential joint overload [28]. Fig. 8(c) and 8(d) showcase PET reconstructions using MRgBSREM and Bowsher’s method, respectively. Both utilize the same MR priors. However, a crucial difference arises in how they handle the mismatch between the dark appearance of subchondral bone in the MR images and the actual 18F-NaF uptake in this region. In Bowsher’s method, forcing the MR contrast directly onto the PET image leads to smearing of the lesions within the subchondral bone. This is evident in the less defined appearance of the uptake areas compared to BSREM. MRgBSREM, on the other hand, can handle the mismatch by incorporating a seed PET image into the feature space. This results in a PET image with sharper lesions like the BSREM image, despite the underlying mismatch in the MR priors. This figure emphasizes the limitations of using MR priors for PET reconstruction when there is a substantial mismatch between the MR and PET images. Nevertheless, it demonstrates the MRgBSREM method’s greater resilience to such misalignments compared to Bowsher’s method.

Fig. 8.

Fig. 8.

Images from a subject injected with 90 MBq of 18F-NaF after 8 flights of a stair ascent/descent exercise. The subject underwent a 30 min of dynamic PET acquisition immediately after the injection. (a) shows MR images acquired with the first echo of a bilateral qDESS sequence, (b) shows the PET images reconstructed with BSREM with β=75, (c) shows the PET images reconstructed with MRgBSREM with β=50 and βm=25 using MR priors in (a) and MRAC images, and (d) shows the PET images reconstructed with anatomy guided PET reconstruction (Bowsher’s method) with β=50 and βm=25 using MR priors in (a) and MRAC images. This figure highlights the limitations of using MR priors for PET reconstruction in scenarios with significant mismatch between the MR and PET images. However, it also demonstrates the superior handling of such mismatches by the MRgBSREM method compared to Bowsher’s method. The PET images show uptake in the subchondral bone while bone appears dark in MR images (arrow). This mismatch causes artifacts on PET images reconstructed with Bowsher’s method. The PET image reconstructed with the MRgBSREM method does not show this artifact because of the inclusion of the PET seed into the feature space.

Fig. 9 shows the MR and PET angiogram images of a subject in an 15O-water study. The angiogram images can be used to estimate the Arterial Input Function (AIF) which is used for cerebral blood flow (CBF) measurement [29]. For this application a high spatial resolution PET image which can resolve the spill-in and spill over problems is highly desirable. Using MR priors can be very helpful in improving spatial resolution and image SNR. This allows for a trade-off i.e., sacrificing some SNR gain for shorter image frames, thereby achieving higher temporal resolution in the AIF curve. Fig. 9(a) displays a Maximum Intensity Projection (MIP) of the MR angiogram in the coronal plane. Fig. 9(a) and 9(c) show the MIP of a PET angiogram reconstructed using the standard OSEM and BSREM methods, respectively, also in the coronal plane. Here, spill-over of activity to adjacent voxels of the carotid artery is evident. Fig. 9(d) presents the same frame reconstructed with the MRgBSREM method, incorporating the MR priors from 9(a). While image SNR improves and spill-over is partially corrected, Bowsher’s reconstruction in 9(e) demonstrates superior spill-over correction and even better SNR. This is because Bowsher’s method directly enforces MR priors onto the final PET image, whereas MRgBSREM is constrained by the PET seed image. In the case of an angiogram, where it is known that the signal must be confined in the anatomic structure, the Bowsher’s method that ignores the OSEM PET seed is expected to provide superior results.

Fig. 9.

Fig. 9.

Images from a subject injected with 850 MBq of 15O-water. The subject underwent a 10 min exam immediately after the injection. An early 20 s frame is reconstructed to visualize arteries in the neck and brain. (a) shows MIP MR angiography images with a 3D TOF SPGR sequence, (b) shows MIP PET images reconstructed using OSEM with 28 subsets and 3 iterations, (c) shows MIP PET images reconstructed using BSREM with β=75, (d) shows MIP PET images reconstructed with MRgBSREM with β=50 and βm=25 using MR priors in (a), and (e) shows MIP PET images reconstructed with anatomy guided PET reconstruction (Bowsher’s method) with β=50 and βm=25 using MR priors in (a). The PET images reconstructed with MRgBSREM method show increased SNR while the PET images reconstructed with Bowsher’s method show better resolution since they directly apply MR images contrast to the PET images. In the setting in which all of the activity must arise from a certain anatomic structure, Bowsher’s method may be preferred over MRgBSREM.

V. Discussion

To use anatomical priors in PET reconstruction, they must be reasonably matched with the PET data. Even though the anatomical MR priors are acquired on a simultaneous PET/MR scanner, the patient might have moved during the exam, and the anatomical MR priors should be co-registered with the motion-corrected PET image. Here, the PET image is reconstructed in list-mode, and each event is aligned with the attenuation correction map in the reconstruction process [21]. Therefore, the anatomical priors are also co-registered with the MR images used for attenuation correction before the reconstruction begins. Combining motion correction, penalized maximum likelihood (i.e., BSREM), and MR priors in the reconstruction pipeline offers a powerful approach for enhancing both spatial resolution and image quality. This improved quality can lead to more accurate diagnosis and potentially better clinical decision-making.

As mentioned in the Framework section, the radius of the neighborhood in the feature space is dynamically adjusted for each voxel to include only the 26 most similar voxels, analogous to the 26-voxel neighborhood used in the image space. By aligning the neighborhood sizes in feature and image space, we can directly compare the influence of the weighting parameters β (governing spatial regularization) and βm (governing MRI-guided regularization). This helps us to leverage existing expertise in the clinical use of BSREM reconstruction, where extensive optimization has been conducted on the β parameter [30–31]. Theoretically, the similarity neighborhood could include all voxels. However, the weight of each voxel (Wj or Wsj) is proportional to its distance from the original voxel in image space or feature space. Voxels farther from the original voxel contribute less to the penalty function, thus limiting their impact on image quality. This reduced contribution should be weighed against their computational cost.

The MR shading artifact can introduce errors in identifying similar voxels within the feature space, particularly when those voxels are spatially distant. To mitigate this issue, we limited the search radius for similar voxels, excluding those exceeding a specific distance threshold (e.g., > 30 mm).

The MRgBSREM method enhances robustness to mismatches between PET and MR images by incorporating a PET seed image during the identification of similar voxels. This prevents voxels with significantly different PET uptake from being considered similar solely based on their MR similarity. To control the influence of the PET seed on the feature space, a weighting factor can be applied. While this may seem like it limits the effect of anatomical priors, it is desirable given that there can be potential real mismatches between PET uptake and anatomical images, as shown in the knee example. Furthermore, the weighting of the anatomical priors, βm, can be adjusted to further fine-tune the effects of MR priors.

In principle, a single set of MRI images, such as 3D T1 IR-SPGR, is sufficient to construct a two-dimensional feature space in conjunction with the PET seed image, as it is used in Figure 7. If the second set of MR images provides similar tissue contrast information (e.g., delineating gray-matter, white-matter, and cerebrospinal fluid), it would not alter the similarity neighborhood within the feature space and, consequently, would not impact the final PET reconstruction.

In MRgBSREM, an additional computational step is required to identify similar voxels within the feature space. This step, however, needs to be performed only once. Subsequently, each iteration in MRgBSREM requires the calculation of two penalty terms, in contrast to the single penalty term calculated in BSREM. While the computational overhead associated with identifying similar voxels is relatively minor, the calculation of two penalty terms can nearly double the computational time per iteration compared to BSREM. However, there are strategies that can be employed to accelerate MRgBSREM. For example, the reconstruction process could initially proceed using the standard BSREM algorithm, followed by the inclusion of the penalty term related to similar voxels only in a subset of later iterations. This hybrid approach can effectively reduce the overall computational cost of MRgBSREM. In our implementation, it resulted in an approximately 30% increase in computation time compared to standard BSREM.

To ensure a direct and equitable comparison between Bowsher’s method and MRgBSREM, motion correction was implemented within the Bowsher’s method.

The global SUVR measured by MRgBSREM and OSEM reconstruction methods in the FBB study are similar as shown in Fig 5(a) as the global cortical composite ROI used to measure the global SUVR is a relatively large ROI. However, in early stages of Alzheimer’s disease, subtle amyloid uptake may occur in localized areas of the cortex which can be obscured in conventional OSEM reconstructions due to spill-in from adjacent white matter regions.

PET reconstruction incorporating MR priors enhances the accuracy of amyloid burden quantification in cases with ambiguous rating [32]. This is achieved by resolving spill-in effects from adjacent white matter voxels, which artificially inflate gray-matter uptake measures. With inconsistent ratings unfortunately common in amyloid PET reading, methods that improve the differentiation of intermediate amyloid levels can improve the quantitation of this modality.

While this study shows the application of MRgBSREM in brain and knee imaging, its potential extends beyond these anatomical areas. For instance, its applicability to prostate imaging has been previously explored in [23]. The primary challenges in applying MRgBSREM to other anatomical areas lie in motion correction and ensuring accurate image alignment between PET and anatomical priors. In the context of brain and knee imaging, these challenges are addressed by the assumption of rigid body motion.

Anatomical priors have been previously utilized in PET/CT imaging [33]. The MRgBSREM method, initially developed for PET/MR systems, can be adapted for use with PET/CT scanners by incorporating corresponding anatomical images. Misalignment between PET and anatomical priors, a critical concern, has been addressed in prior research, as detailed in [34].

VI. Conclusion

A penalized maximum likelihood algorithm based on the relative difference prior combined with using MR priors in the feature space can control the image noise and improve SNR and spatial resolution. The integration of MR priors within the feature space guides the algorithm toward anatomically plausible solutions, while the seed PET image ensures consistency with the underlying PET data. This synergistic combination can potentially lead to more accurate and reliable PET images.

Acknowledgment

The authors declare the following financial interests/personal relationships which may be considered as potential competing interests with the work reported in this paper: M.G. Spangler-Bickell and F. Jansen are employed by GE HealthCare. T.W. Deller was employed with GE HealthCare. F. Kogan and G. Zaharchuk receive research support from GE HealthCare.

This project was supported by National Institutes of Health (USA) P41-EB015891, R01-EB025220, R01-AR079431, R01-NS123025, R21-AG058859, R56-AG071558, the Stanford Alzheimer’s Disease Research Center (P30-AG06615), the Alzheimer’s Association (AARFD-21-849349), California Institute of Regenerative Medicine CLIN2-12379, American Heart Association (23SCEFIA1141920), GE HealthCare, and Life Molecular Imaging. This work involved human subjects in its research. Approval of all ethical and experimental procedures and protocols was granted by Stanford’s institutional review board (IRB).

Contributor Information

M. Mehdi Khalighi, Radiology Department, Stanford University..

Christina B. Young, Neurology Department, Stanford University.

Matthew G. Spangler-Bickell, PET/MR Engineering, GE HealthCare.

Timothy W. Deller, PET/MR Engineering, GE HealthCare, he is now at Oak Ridge National Laboratory.

Floris Jansen, PET/MR Engineering, GE HealthCare..

Dawn Holley, Radiology Department, Stanford University..

Hillary Vossler, Neurology Department, Stanford University..

Moss Zhao, Radiology and Neurology Departments, Stanford University..

Feliks Kogan, Radiology Department, Stanford University..

Gary Steinberg, Neurosurgery Department, Stanford University..

Elizabeth Mormino, Neurology Department, Stanford University..

Michael Moseley, Radiology Department, Stanford University..

Greg Zaharchuk, Radiology Department, Stanford University.

References

  • [1].Hudson HM, and Larkin RS (1994). Accelerated image reconstruction using ordered subsets of projection data. IEEE Trans. Med Imaging 13(4), pp. 601–609. [DOI] [PubMed] [Google Scholar]
  • [2].Liow JS, and Strother SC (1991). Practical tradeoffs between noise, quantitation, and number of iterations for maximum likelihood-based reconstructions. IEEE Trans Med Imaging. 10(4), pp. 563–571. [DOI] [PubMed] [Google Scholar]
  • [3].Qi J, and Leahy RM (2006). Iterative reconstruction techniques in emission computed. Phys Med Biol. 51(15), pp 541–78. [DOI] [PubMed] [Google Scholar]
  • [4].Nuyts JJ, Beque D, Dupont P, and Mortelmans L (2002). A concave prior penalizing relative differences for maximum-a-posteriori reconstruction in emission tomography. IEEE Transactions on Nuclear Science. 49(1), pp :56–60. [Google Scholar]
  • [5].Ahn S, and Fessler JA (2003). Globally convergent image reconstruction for emission tomography using relaxed ordered subsets algorithms. IEEE Transactions on Medical Imaging. 22(5), pp 613–626. [DOI] [PubMed] [Google Scholar]
  • [6].Ahn S, et al. (2015). Quantitative comparison of OSEM and penalized likelihood image reconstruction using relative difference penalties for clinical PET. Physics in Medicine & Biology. 60(15), pp 5733. [DOI] [PubMed] [Google Scholar]
  • [7].Passalaqua S, et al. (2014). Qualitative and quantitative evaluation of regularized PET image reconstruction. Journal of Nuclear Medicine. 55(supplement 1), pp 579–579. [Google Scholar]
  • [8].Lantos J, Mittra ES, Levin CS, and Iagaru A (2018). Standard OSEM vs. regularized PET image reconstruction: qualitative and quantitative comparison using phantom data and various clinical radiopharmaceuticals. Am J Nucl Med Mol Imaging. 8(2), pp 110–118. [PMC free article] [PubMed] [Google Scholar]
  • [9].Bowsher JE, et al. “Utilizing MRI information to estimate F18-FDG distributions in rat flank tumors,” in Proc. IEEE Nuclear Science Symp. and Medical Imaging Conf, 2004, pp 2488–92. [Google Scholar]
  • [10].Chen Y, Ma J, Feng Q, Luo L, Shi P, and Chen W (2008). Nonlocal Prior Bayesian Tomographic Reconstruction. J Math Imaging Vis. 30, pp133–146. [Google Scholar]
  • [11].Kazantsev D, Bousse A, Pedemonte S, Arridge SR, Hutton BF, and Ourselin S, “Edge preserving Bowsher prior with nonlocal weighting for 3D SPECT reconstruction,” in Proc. IEEE International Symposium on Biomedical Imaging, 2011, pp 1158–1161. [Google Scholar]
  • [12].Ehrhardt M, et al. (2016). PET Reconstruction with an Anatomical MRI Prior using Parallel Level Sets. IEEE Trans Med Imaging. 35(9), pp 2189–2199 [DOI] [PubMed] [Google Scholar]
  • [13].Mehranian A, and Reader AJ, “Multi-parametric MRI-guided PET image reconstruction,” in Proc. Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2016, pp 1–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [14].Bland J, et al. (2019). Intercomparison of MR-informed PET image reconstruction methods. Med Phys. 46(11), pp 5055–5074. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [15].Bai B, Li Q, and Leahy R (2013). MR Guided PET Image Reconstruction. Semin Nucl Med. 43(1), pp 30–44. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Schramm G, et al. (2018). Evaluation of Parallel Level Sets and Bowsher’s Method as Segmentation-Free Anatomical Priors for Time-of-Flight PET Reconstruction. IEEE Trans Med Imaging. 37(2), pp 590–603. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [17].Mehranian A, et al. (2017). PET image reconstruction using multi-parametric anato-functional priors. Phys. Med. Biol 62, pp 5975–6007. [DOI] [PubMed] [Google Scholar]
  • [18].Belzunce MA, Mehranian A, and Reader AJ (2019). Enhancement of Partial Volume Correction in MR-Guided PET Image Reconstruction by Using MRI Voxel Sizes. IEEE Trans Radiat Plasma Med Sci. 3(3), pp 315–326. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Corda-D’Incan G, Schnabel JA, Hammers A and Reader AJ (2023). Single-Modality Supervised Joint PET-MR Image Reconstruction. IEEE Trans Radiat Plasma Med Sci, 7(7), pp. 742–754. [Google Scholar]
  • [20].Khalighi MM, et al. “Studying Rigid Head Motion on PET/MR Brain Exams,” presented at ISMRM-SNMMI workshop on PET/MRI, Los Angles, Oct. 26–29, 2023. [Google Scholar]
  • [21].Spangler-Bickell M, Hurley S, Pirasteh A, Perlman S, Deller TW, McMillan A (2022). Evaluation of Data-Driven Rigid Motion Correction in Clinical Brain PET Imaging. Journal of Nuclear Medicine. 63(10), pp 604–1610. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Khalighi MM, Moseley ME, “Feature space based MR guided PET reconstruction,” US Patent 11,672,492, 2023.
  • [23].Khalighi MM, et al. (2020). High Quality Isotropic Whole-body PET Imaging Using MR Priors. J Nucl Med. 61(suppl), pp 1477. [Google Scholar]
  • [24].Delso G, et al. (2015). Clinical Evaluation of Zero-Echo-Time MR Imaging for the Segmentation of the Skull. Journal of Nuclear Medicine. 56 (3), pp 417–422. [DOI] [PubMed] [Google Scholar]
  • [25].Spangler-Bickell MG, et al. (2023). Optimizing the frame duration for data-driven rigid motion estimation in brain PET imaging. Medical Physics. 50, pp 5295–5295. [DOI] [PubMed] [Google Scholar]
  • [26].Staroswiecki E, Granlund KL, Alley MT, Gold GE GE, and Hargreaves BA (2012). Simultaneous estimation of T2 and apparent diffusion coefficient in human articular cartilage in vivo with a modified three-dimensional Double Echo Steady State (DESS) sequence at 3T. Magnetic Resonance in Medicine. 67(4), pp 1086–1096. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [27].Wollenweber SD, et al. (2013) Comparison of 4-class and continuous fat/water methods for whole-body, MR-based PET attenuation correction. IEEE Trans. Nucl. Sci 60:3391–3398. [Google Scholar]
  • [28].Watkins LE, et al. (2022). [18F]Sodium fluoride PET-MRI detects increased metabolic bone response to whole-joint loading stress in osteoarthritic knees. Osteoarthritis Cartilage. 30(11), pp 1515–1525. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].Khalighi MM, et al. (2018). Image-derived input function estimation on a TOF-enabled PET/MR for cerebral blood flow mapping. J Cereb Blood Flow Metab. 38(1), pp 126–135. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [30].Ribeiro D, Hallett W, and Tavares AAS (2021). Performance evaluation of the Q.Clear reconstruction framework versus conventional reconstruction algorithms for quantitative brain PET-MR studies. EJNMMI Phyics. 41(8). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [31].Teoh EJ, McGowan DR, Macpherson RE, Bradley KM, and Gleeson FV (2015). Phantom and Clinical Evaluation of the Bayesian Penalized Likelihood Reconstruction Algorithm Q.Clear on an LYSO PET/CT System. Journal of Nuclear Medicine. 56(9), pp 1447–1452. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [32].Khalighi MM, et al. “Enhancing the Diagnostic Accuracy of Amyloid PET: The Impact of MR-Guided PET Reconstruction,” medRxiv 2025.01.04.25319996; doi: 10.1101/2025.01.04.25319996 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [33].Cheng-Liao J and Qi J, (2011) PET image reconstruction with anatomical edge guided level set prior. Phys. Med. Biol 56, 6899. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [34].Tsai Y, et al. “Algorithms for Solving Misalignment Issues in Penalized PET/CT Reconstruction Using Anatomical Priors,” in Proc. IEEE Nuclear Science Symposium and Medical Imaging Conference Proceedings (NSS/MIC), 2018, pp 1–3. [Google Scholar]

RESOURCES