Skip to main content
. 2025 Oct 7;13(1):157–165. doi: 10.1159/000548849

Table 2.

BP, AP, AIx, and SEVR measured by both devices

​ CONNEQT pulse SphygmoCor XCEL p value
HR, beats/min 71.4±10.6 70.3±10.7 <0.001
bSBP, mm Hg 130.4±17.4 128.2±16.6 <0.001
bDBP, mm Hg 80.0±10.8 77.2±11.3 <0.001
bPP, mm Hg 50.4±13.0 51.0±12.2 0.22
cSBP, mm Hg 121.6±17.2 117.5±14.5 <0.001
cDBP, mm Hg 80.6±11.0 78.3±11.4 <0.001
cPP, mm Hg 41.0±12.7 39.2±9.5 <0.001
AP, mm Hg 11.5±6.2 12.8±6.6 <0.001
AIx, % 26.4±9.0 32.0±14.0 <0.001
SEVR, % 135.6±25.5 137.8±26.4 0.02