Table 2.
BP, AP, AIx, and SEVR measured by both devices
| | CONNEQT pulse | SphygmoCor XCEL | p value |
|---|---|---|---|
| HR, beats/min | 71.4±10.6 | 70.3±10.7 | <0.001 |
| bSBP, mm Hg | 130.4±17.4 | 128.2±16.6 | <0.001 |
| bDBP, mm Hg | 80.0±10.8 | 77.2±11.3 | <0.001 |
| bPP, mm Hg | 50.4±13.0 | 51.0±12.2 | 0.22 |
| cSBP, mm Hg | 121.6±17.2 | 117.5±14.5 | <0.001 |
| cDBP, mm Hg | 80.6±11.0 | 78.3±11.4 | <0.001 |
| cPP, mm Hg | 41.0±12.7 | 39.2±9.5 | <0.001 |
| AP, mm Hg | 11.5±6.2 | 12.8±6.6 | <0.001 |
| AIx, % | 26.4±9.0 | 32.0±14.0 | <0.001 |
| SEVR, % | 135.6±25.5 | 137.8±26.4 | 0.02 |