Abstract
Background
Bifurcation stenting of the coronary arteries represents a technical challenge that places patients at risk for poor outcomes. Existing techniques result in multiple stent layers at the neocarina, making for a challenging side branch recross and necessitating multiple kissing balloon inflations. This article presents a novel bifurcation stenting technique that provides full side branch coverage while maintaining carinal geometry, with an easier side branch recross and less technical complexity. Clinical outcomes at 1-year were investigated.
Methods
Feasibility of the technique was demonstrated in a wet heart model. A retrospective cohort study was then performed of patients undergoing jailed stent-balloon bifurcation stenting at a single tertiary academic center. The primary end point was a composite of target lesion revascularization (TLR), target lesion myocardial infarction (TLMI), and cardiac death at 1 year postprocedurally.
Results
The primary composite outcome occurred in 13 (7.9%) cases who underwent jailed stent-balloon bifurcation stenting. At 1 year, TLR occurred in 9 (5.4%) cases, TLMI in 3 (1.8%) cases, and cardiac death in 3 (1.8%) cases. In patients with left main bifurcation disease, the primary composite outcome occurred in 6 (9.5%) cases, TLR in 3 (4.8%) cases, TLMI in 2 (3.2%) cases, and cardiac death in 3 (4.8%) cases.
Conclusions
Jailed stent-balloon bifurcation stenting appears to be safe, feasible, and effective. Although not directly comparable, it appears to have similar clinical outcomes to existing bifurcation stenting techniques despite its reduced technical complexity.
Keywords: bifurcation stenting, complex bifurcation, percutaneous coronary intervention, upfront 2-stent strategy
Introduction
Bifurcation lesions account for 10% to 20% of percutaneous coronary interventions (PCIs).1, 2, 3 While a provisional stenting strategy is recommended for revascularization of most bifurcation lesions,4, 5, 6 an upfront 2-stent strategy is warranted when there is a high risk of side branch occlusion.5,7,8 This is particularly the case in bifurcation lesions involving the distal left main (LM).
Existing 2-stent techniques are technically challenging, and there is frequently a tradeoff between formation of a multistent-layered neocarina and a significant risk of missing the side branch ostium.9 Early studies of patients undergoing bifurcation stenting demonstrated a high rate of in-stent restenosis associated with missing the side branch ostium,10 so most 2-stent strategies favor hanging the side branch stent proximally into the main branch to ensure ostial coverage, thereby accepting formation of a multistent-layered neocarina. Subsequent optimization of the neocarina is challenging, and preservation of the geometry of both main branch and side branch ostia with well-apposed stents is critical to preventing target lesion failure.5,9 Kissing balloon inflation (KBI) has been found to be critical for creating a well-formed neocarina; however, KBI requires crossing through a main branch strut into the side branch ostium, which is frequently challenging because of multiple layers of stents, particularly in the crush family of techniques.5 Later iterations of the crush and culotte techniques sought to minimize the difficulty of recrossing into the side branch using a KBI before deployment of the main branch stent in addition to a finishing KBI (the “double kiss”).9 While effective, the added complexity of multiple KBIs increases the risk of operator error and makes the technique less accessible to operators with less experience or volume.
In this article, we report the single-center outcomes of the jailed stent-balloon technique, a novel bifurcation technique that provides full stent coverage of the side branch ostium and minimizes the number of stent layers at the neocarina, all in fewer steps and with an easier recross than existing bifurcation stenting techniques.
Description of the technique
An illustration of the technique is portrayed in the Central Illustration. The stents are positioned based on intravascular imaging, such that the proximal edge of the side branch stent is at the ostium of the side branch, while the main branch stent covers the prebifurcation and postbifurcation main branch. The degree of side branch stent protrusion into the main branch is dependent on the angle of the bifurcation. It is critical that the leading aspect of the side branch vessel (indicated by the asterisk in the Central Illustration) be fully covered by the side branch stent. It is ideal to have 0.0 mm of protruding stent at this leading aspect of the side branch, which necessitates the use of intravascular imaging (the authors use a manual ultrasound pullback technique, leveraging both dry and contrast-facilitated cineangiography to exactly mark the location of the side branch ostium on fluoroscopy). The degree of side branch stent hanging over the neocarina is subsequently a function of the bifurcation angle, with a 90° angle having 0.0 mm of overhang and a narrow bifurcation angle necessitating up to 1.0 mm of overhang or more in cases of a very large side branch diameter with an extremely narrow angle. The side branch stent is then deployed, but instead of removing the stent-deployment balloon, it is pulled back proximal to the stent edge of the undeployed main branch stent. The main branch stent is then deployed, jailing the side branch stent-deployment balloon. The jailed side branch stent-deployment balloon is then reexpanded to at least nominal pressures, correcting the deformation of the ostial side branch stent and preserving the carinal angle. The side branch stent-deployment balloon is then removed, keeping the side branch wire jailed in place. The main branch stent-deployment balloon is then reexpanded to correct the deformation of the main branch stent. Proximal optimization is then performed over the main branch wire. The side branch is then rewired, and a kissing balloon angioplasty is performed. Following this, intravascular imaging of both limbs is performed to ensure appropriate expansion and apposition. If eccentricity is found in the prebifurcation lesion, an additional proximal optimization balloon angioplasty can be performed.
Central Illustration.
Illustration of the jailed stent-balloon technique. The stents are positioned with the proximal edge of the side branch stent at the ostium of the side branch, while the main branch stent covers the pre- and postbifurcation main branch. The side branch stent is then deployed. The side branch stent-deployment balloon is pulled back to be 0.0-2.0 mm proximal to the proximal stent edge of the undeployed main branch stent. The main branch stent is then deployed, jailing the side branch stent-deployment balloon. The jailed side branch stent-deployment balloon is then reexpanded. After expansion, the balloon is removed. The main branch stent-deployment balloon is then reexpanded to initial deployment pressures. Proximal optimization is performed over the main branch wire. The side branch is then rewired, and a kissing balloon angioplasty is performed. ∗The degree of side branch stent protrusion into the main branch is dependent on the angle of the bifurcation. It is critical that the leading aspect of the side branch vessel (indicated by asterisk in the figure be fully covered by the side branch stent. It is ideal to have 0.0 mm of protruding stent at this leading aspect of the side branch, which necessitates the use of intravascular imaging. The degree of side branch stent hanging over the neocarina is subsequently a function of the bifurcation angle, with a 90° angle having 0.0 mm of overhang and a narrow bifurcation angle necessitating up to 1.0 mm of overhang or more in cases of a very large side branch diameter with an extremely narrow angle.
The jailed stent-balloon technique typically requires a 7F guiding catheter, but a modification of the technique allows for execution via a 6F catheter. See Supplemental Appendix S1 for further details. Additionally, an alteration of the technique is required when the length of the main branch stent proximal to the bifurcation exceeds the length of the side branch stent so that the jailed side branch stent-deployment balloon can be safely removed. See Supplemental Appendix S2 for further details.
Materials and methods
Feasibility
The jailed stent-balloon technique was executed in a wet heart model at the University of Minnesota Visible Heart Laboratory. Intracoronary videos (Supplemental Videos 1-5) were obtained to investigate carina geometry and stent deformation. They are available online (https://osf.io/x73bt/files/osfstorage).
Study design
This is a single-center, retrospective cohort study of patients undergoing PCI from 2013-2021. Patients were included in the cohort if they were aged ≥18 years at the time of intervention, had true bifurcation disease according to the Medina classification (1,1,1; 1,0,1; and 0,1,1), and had undergone bifurcation stenting with the jailed stent-balloon technique. Chart review was performed to obtain data on patient demographic characteristics, baseline medical comorbidities, procedural and lesion details, and clinical outcomes at 1 year postprocedurally. Patients whose outcomes were not available via chart review were interviewed via telephone. Medical records were obtained following interviews to verify outcome events.
Outcomes
The primary outcome was a composite of target lesion revascularization (TLR), target lesion myocardial infarction (TLMI), and cardiac death at 1 year postprocedurally. The TLMI was defined as a clinical presentation consistent with myocardial infarction (MI) according to the Fourth Universal Definition of MI (high sensitivity-troponin levels with at least 1 value >99th percentile of normal with symptoms of ischemia, new ischemic electrocardiogram changes, new pathological Q waves, or imaging evidence of new ischemia). All MIs were assumed to be TLMIs unless otherwise proven to be due to a nontarget vessel by angiogram or equivalent imaging modality. The etiology of death was adjudicated by the authors. In keeping with previous clinical trials investigating approaches to LM bifurcation revascularization, cases where the cause of death was unknown were assumed to be a cardiac death.7,11 Event adjudication was performed by the authors of the study. A prespecified, hypothesis-generating secondary analysis was performed to investigate how various patient, medical comorbidity, lesion, and procedural aspects were associated with the primary composite outcome.
Data analysis
Continuous data are presented as mean ± SD when normally distributed and median (IQR) when not. Categorical data are presented as frequency (percentage). Time-to-outcome events have been illustrated via cumulative incidence curves estimated via the Aalen-Johansen method. In the setting of competing events from noncardiac deaths, a Fine-Gray model was used to investigate which patient and procedural factors might be associated with the primary composite outcome. Owing to a limited sample size, only a subset of covariates could be included in the model, and these covariates were selected via a domain expert–driven approach. The variables included in the model were age, sex, diabetes, SYNTAX score, case acuity (elective vs urgent; there were no emergent cases), involvement of the LM bifurcation, use of mechanical support, use of atherectomy or lithotripsy, intervention on in-stent restenosis (ISR), main branch stent diameter, and side branch stent diameter. The proportional hazards assumption was verified using Schoenfeld residuals. Data were censored at 1 year, as beyond this period the independent censoring assumption might be violated. Extensive efforts were made to ensure that no missing data were present that might introduce bias into the Fine-Gray model. All data handling and analysis were performed in R software (version 4.3.1; R Core Team). Annotated statistical code is available publicly on GitHub (https://github.com/jeffreyshowtran/jailedStentBalloonTechnique/blob/main/JSB_statistialCode_redacted_v2.R). Deidentified data can be made available upon reasonable request to the corresponding author.
Results
Patient characteristics
During the study period, 165 patients underwent bifurcation stenting with the jailed stent-balloon technique. Over the 1-year follow-up period, zero patients were lost to follow-up. The mean age was 68 ± 11 years, and 119 (71%) were male. Diabetes was present in 78 (48%) patients, with 41 (25%) requiring home insulin use. Chronic kidney disease was present in 40 (24%) patients, with 14 (9%) having end-stage renal disease. Heart failure was present in 43 (26%) patients, with the median ejection fraction in patients with heart failure being 40% (27%-52%). Previous revascularization with PCI had occurred in 72 (43%) patients, and 25 (15%) patients had previously undergone coronary artery bypass grafting. Additional details on patient demographic characteristics and medical comorbidities can be found in Table 1.
Table 1.
Demographic characteristics and medical comorbidities.
| All patients (N = 165) | LM bifurcation disease (n = 63) | |
|---|---|---|
| Age, y | 68 ± 11 | 70 ± 13 |
| Male sex | 119 (71.3) | 50 (78.1) |
| Body mass index, kg/m2 | 29.1 ± 6.7 | 29.0 ± 6.8 |
| Smoking status | ||
| Nonsmoker | 99 (60.0) | 31 (48.4) |
| Current | 18 (10.9) | 6 (9.4) |
| Previous | 48 (29.1) | 27 (42.2) |
| Hypertension | 145 (86.8) | 56 (87.5) |
| Hyperlipidemia | 148 (88.6) | 55 (85.9) |
| Diabetes | 78 (46.7) | 34 (53.1) |
| Insulin use | 41 (24.8) | 18 (28.6) |
| Kidney disease | ||
| CKD | 40 (24.2) | 12 (18.8) |
| ESRD | 14 (8.5) | 7 (10.9) |
| COPD | 34 (20.6) | 16 (25.0) |
| Preexisting CAD | 54 (32.3) | 27 (42.2) |
| Previous PCI | 72 (43.1) | 36 (56.3) |
| Previous CABG | 25 (15.0) | 14 (21.9) |
| PAD | 26 (15.8) | 15 (23.4) |
| Heart failure | 43 (26.1) | 16 (25.0) |
| Ejection fraction in patients with heart failure, % | 40 (26.5-51.5) | 34 (27.5-46.3) |
Values are mean ± SD, n (%), or median (IQR).
CABG, coronary artery bypass grafting; CAD, coronary artery disease; CKD, chronic kidney disease defined by CKD stage 3a or greater; COPD, chronic obstructive pulmonary disease; ESRD, end-stage renal disease; LM, left main; PAD, peripheral artery disease; PCI, percutaneous coronary intervention.
Lesion characteristics
In terms of lesion characteristics, distal LM bifurcation disease was the target in 63 (38%) patients, and the majority (80%) of lesions were of Medina classification 1,1,1. A high SYNTAX score (>33) was present in 90 (56%) patients, an intermediate score (23-32) in 29 (18%), and a low score in 42 (26%). The main vessel was the LM in 67 (41%) patients, left anterior descending in 74 (45%) patients, left circumflex in 20 (12%) patients, and right coronary artery in 4 (2.4%) patients.
In patients who underwent intervention to the LM bifurcation, the SYNTAX score was high (>33) in the majority (94%) of cases. In 42 (67%) cases, the LM into left anterior descending left anterior descending was considered the main branch, whereas the left circumflex was considered the main branch in 21 (33%) cases. Additional lesion characterizations can be found in Table 2.
Table 2.
Lesion details.
| All patients (N = 165) | LM bifurcation disease (n = 63) | |
|---|---|---|
| Medina classification | ||
| 1,1,1 | 132 (80.0) | 53 (84.1) |
| 0,1,1 | 26 (15.8) | 7 (11.1) |
| 1,0,1 | 7 (4.2) | 3 (4.8) |
| SYNTAX score | ||
| Low (≤22) | 42 (26.1) | 1 (1.6) |
| Intermediate (23-32) | 29 (18.0) | 3 (4.8) |
| High (>33) | 90 (55.9) | 58 (93.6) |
| Main branch | ||
| LM/LAD | 44 (26.7) | 42 (66.7) |
| LM/LCx | 23 (13.9) | 21 (33.3) |
| LAD | 74 (44.9) | – |
| LCx | 20 (12.1) | – |
| RCA | 4 (2.4) | – |
| Side branch | ||
| LAD | 15 (9.1) | 21 (33.3) |
| LCx | 37 (22.4) | 42 (66.7) |
| Diagonal | 82 (49.7) | – |
| OM | 27 (16.4) | – |
| PDA | 3 (1.8) | – |
| PLB | 1 (0.6) | – |
Values are n (%).
LAD, left anterior descending; LCx, left circumflex; LM, left main; OM, obtuse marginal; PDA, posterior descending; PLB, posterolateral branch; RCA, right coronary artery.
Procedural characteristics
The majority of patients presented in a nonelective setting, with 7 (4%) presenting with ST-elevation MI, 46 (28%) with non–ST-elevation MI, and 63 (38%) with unstable angina; 51 (31%) of patients presented in an elective setting either with stable angina or for another elective indication. Atherectomy or lithotripsy was used in 42 (25%) cases. Mechanical support was used in 21 (13%) cases. ISR was intervened upon in 19 (11.5%) cases, with all ISR interventions resulting in the placement of a stent-in-stent. Intracoronary imaging with intravascular ultrasound was used in all (100%) cases. Angiographic success was achieved in 100% of cases. A procedural complication occurred in 9 (5%) cases, with the most common complication being iatrogenic coronary artery dissection requiring an additional stent placement. No reflow of the main vessel occurred in 1 (0.6%) case. There were no incidents of side branch occlusion or equipment entrapment.
In patients who underwent intervention to the LM bifurcation, atherectomy or lithotripsy was used in 31 (49%) cases, and mechanical support was used in 15 (24%) cases. Procedural complications occurred in 3 (5%) cases, again with the most common being iatrogenic dissection requiring additional stent placement. There were no incidents of no reflow of the main vessel, side branch occlusion, or equipment entrapment. Further procedural details can be found in Table 3.
Table 3.
Procedural details.
| All patients (N = 165) | LM bifurcation disease (n = 63) | |
|---|---|---|
| Clinical presentation | ||
| STEMI | 7 (4.2) | 1 (1.6) |
| NSTEMI | 46 (27.5) | 19 (29.7) |
| Unstable angina | 63 (37.7) | 25 (39.1) |
| Stable angina | 39 (23.4) | 16 (25.0) |
| Asymptomatic/other indication | 12 (7.2) | 3 (4.7) |
| Access | ||
| Femoral | 156 (94.6) | 57 (90.4) |
| Radial | 9 (5.4) | 6 (9.5) |
| Guidewire size | ||
| 6F | 3 (1.8) | 1 (1.6) |
| 7F | 136 (82.4) | 51 (81.0) |
| 8F | 26 (15.8) | 11 (17.4) |
| Procedural anticoagulation | ||
| Heparin | 153 (93.3) | 61 (96.8) |
| Bivalirudin | 11 (6.7) | 2 (3.2) |
| Atherectomy/lithotripsy used | 42 (25.4) | 31 (49.2) |
| Intracoronary imaging used | 165 (100) | 63 (100) |
| Intracoronary physiology used | 31 (18.9) | 10 (15.9) |
| In-stent restenosis | 19 (11.5) | 9 (14.3) |
| Mechanical support used | 21 (12.7) | 15 (23.8) |
| Main vessel stent type | ||
| Promus/SYNERGY | 33 (20) | 12 (19.1) |
| XIENCE | 66 (40) | 31 (49.2) |
| Resolute | 63 (38.2) | 19 (30.2) |
| Orsiro | 3 (1.8) | 1 (1.6) |
| Main vessel stent diameter, mm | 3.0 (2.75-3.5) | 3.5 (3.13-4.0) |
| Main vessel stenosis pre-PCI, % | 80 (70-90) | 87.5 (71.3-90) |
| Side vessel stent | ||
| Promus/SYNERGY | 32 (19.6) | 12 (19.1) |
| XIENCE | 65 (39.9) | 31 (49.2) |
| Resolute | 63 (38.7) | 19 (30.2) |
| Orsiro | 3 (1.8) | 1 (1.6) |
| Side vessel stent diameter, mm | 2.5 (2.5-3.0) | 3.0 (2.75-3.5) |
| Side vessel stenosis pre-PCI, % | 85 (75-90) | 90 (70-90) |
| Proximal optimization performed | 165 (100) | 63 (100) |
| Kissing balloon inflation performed | 165 (100) | 63 (100) |
| Angiographic successa | 165 (100) | 63 (100) |
| Closure | ||
| Radial compression band | 9 (5.4) | 6 (9.5) |
| Manual pressure | 20 (12.1) | 10 (15.9) |
| Angio-Seal | 128 (78.1) | 45 (71.4) |
| ProGlide | 4 (2.4) | 1 (1.6) |
| Celt | 3 (1.8) | 1 (1.6) |
| Fluoroscopy time, min | 31.5 ± 12.4 | 30.4 ± 11.9 |
| Contrast used, mL | 194.9 ± 77.1 | 187.5 ± 73.5 |
| Iatrogenic coronary dissectionb | 9 (5.4) | 3 (4.8) |
| No reflow | 1 (0.6) | 0 (0.0) |
| Side branch occlusion | 0 (0.0) | 0 (0.0) |
| Equipment entrapment | 0 (0.0) | 0 (0.0) |
| Major bleedingc | 4 (2.4) | 1 (1.6) |
| Postprocedural AKI | 5 (3.1) | 1 (1.6) |
Values are n (%), median (IQR), or mean ± SD.
AKI, acute kidney injury; NSTEMI, non–ST-elevation myocardial infarction; STEMI, ST-elevation myocardial infarction.
Angiographic success was defined as a residual stenosis of <30% with Thrombolysis in Myocardial Infarction (TIMI) flow 3 in both branches.
Coronary dissection is defined as a clinically significant iatrogenic dissection requiring placement of an additional stent.
Major bleeding is defined as Bleeding Academic Research Consortium (BARC) classification 2, 3, or 5.
Clinical outcomes
The primary composite outcome of TLR, TLMI, and cardiac death occurred in 13 (7.9%) cases who underwent jailed stent-balloon bifurcation stenting (Figure 1). TLR occurred in 9 (5.4%) cases, TLMI in 3 (1.8%) cases, and cardiac death in 3 (1.8%) cases.
Figure 1.
Cumulative incidence curves. Cumulative incidence curves generated via Aalen-Johansen estimates demonstrating cumulative events in all patients (left) and patients with left main (LM) bifurcation disease (right).
In patients with LM bifurcation disease, the primary composite outcome occurred in 6 (9.5%) cases (Figure 1). TLR occurred in 3 (4.8%) cases, TLMI in 2 (3.2%) cases, and cardiac death in 3 (4.8%) cases. Two patients with LM bifurcation disease experienced a simultaneous TLMI/TLR event. No other patients experienced multiple components of the composite outcome. One patient who had experienced a TLMI/TLR event died of noncardiac causes 211 days later. Based on the chart review, there was no direct link between the TLMI/TLR event and noncardiac death. The breakdown of clinical outcomes can be found in Table 4.
Table 4.
Clinical outcomes.
| Outcome | All patients (N = 165) | LM bifurcation disease (n = 63) |
|---|---|---|
| Composite outcome | 13 (7.9) | 6 (9.5) |
| TLR | 9 (5.4) | 3 (4.8) |
| TLMI | 3 (1.8) | 2 (3.2) |
| Cardiac death | 3 (1.8) | 3 (4.8) |
| Noncardiac death | 8 (4.8) | 5 (7.9) |
Values are n (%). The composite outcome consisted of TLR, TLMI, and cardiac death. Noncardiac death is included as a competing event. Two patients who underwent LM bifurcation revascularization experienced simultaneous TLMI/TLR events. No other patients experienced multiple components of the composite outcome. No formal hypothesis testing was conducted to evaluate for differences between patients with LM disease and all comers due to the nonrandomized nature of the data and the likely presence of confounders.
LM, left main; TLMI, target lesion myocardial infarction; TLR, target lesion revascularization.
In subgroup analysis, only interventions performed on lesions with ISR were found to have a statistically significant increase in the subhazard ratio (sub-HR) of a patient experiencing the composite outcome at 1 year (sub-HR, 7.4; 95% CI, 1.6-35). A high SYNTAX score (sub-HR, 0.45; 95% CI, 0.06-3.4, for a high SYNTAX score vs low), involvement of the LM bifurcation (sub-HR, 1.01; 95% CI, 0.15-6.7), mechanical support (sub-HR, 4.4; 95% CI, 0.97-20), and the use of atherectomy (sub-HR, 0.46; 95% CI, 0.09-2.4) were not found to have a statistically significant association with the composite outcome (Figure 2).
Figure 2.
Subgroup analysis. Forest plot demonstrating the subhazard ratios of various patient, medical comorbidity, lesion, and procedural aspects as they are associated with the 1-year composite outcome of target lesion revascularization, target lesion myocardial infarction, and cardiac death. Subhazard ratio estimates were obtained via a Fine-Gray model in the setting of competing events from noncardiac death. ISR, in-stent restenosis; LM, left main.
Discussion
Bifurcation stenting continues to present a technical challenge that places patients at risk for adverse outcomes.7,9 The ideal bifurcation stenting technique should provide full coverage of both main and side branches, maintain carina geometry, avoid malapposed or deformed stent struts, be applicable to most carina geometries, and be accessible to most operators.9 The jailed stent-balloon bifurcation technique fulfills these criteria. It is simpler to perform than other contemporary techniques and facilitates an easier side branch recross, while maintaining full side branch coverage and neocarina geometry.10 Despite its simpler execution, the jailed stent-balloon bifurcation stenting technique appears to be safe and feasible, with 100% angiographic success rate and periprocedural complication rates similar to other contemporary bifurcation stenting techniques.
Current contemporary bifurcation stent techniques struggle with carina deformation and procedural complexity.11 In both the crush and culotte family of techniques, multiple layers of overlapping stents at the carina cause neocarina deformation, which has been associated with rates of restenosis approximately 8.6% and rates of stent thrombosis approximately 3.3%.12,13 This has necessitated the use of KBIs both before deployment of the main branch stent and after deployment, increasing procedural complexity.5,9,13 Additionally, the critical nature of the final KBI mandates recrossing into the side branch through the main branch stent, which can be challenging and time consuming. In a single-center study, challenges were encountered 38% of the time recrossing into the side branch for the final KBI, 30% of which required a new guide wire and 15% of which required the assistance of a microcatheter.14
The jailed stent-balloon technique alleviates these challenges in 3 ways: (1) it leverages intravascular imaging to minimize the amount of side branch strut hanging into main branch, similar to the mini-crush and nanocrush techniques. This reduces the amount of deformed metal at the neocarina after main branch stent-deployment. (2) The balloon shaft of the jailed stent-balloon helps maintain carinal geometry, while the main branch stent is positioned and deployed. This group previously published on the jailed balloon provisional stenting technique, whereby a jailed balloon is kept across the side branch ostium while the main branch stent is deployed. After inflating this side branch balloon and then reinflating the main branch stent-deployment balloon, a 99% rate of side branch patency was achieved.15 The jailed stent-balloon bifurcation stenting technique is an extension of the jailed balloon provisional stenting technique, with the jailed side branch balloon a critical step in preventing carinal shift. (3) Reinflating the jailed side branch stent-balloon after deployment of the main branch stent performs a similar role to the pre-main branch stent-deployment KBI in DK crush and DK culotte, correcting side branch ostial stent deformation. This greatly simplifies side branch recross for the final KBI in much fewer steps. By addressing these challenges, the jailed stent-balloon technique makes bifurcation stenting more accessible to less experienced operators or those with lower volumes and reduces patients’ risk for poor outcomes.
Although not directly comparable, the jailed stent-balloon technique appears to have clinical outcomes similar to other bifurcation stenting techniques. For true LM bifurcation lesions, jailed stent-balloon bifurcation stenting had a 1-year prevalence of TLR of 4.8%, TLMI of 3.2%, and cardiac death of 4.8%. Based on previous clinical trials, DK crush has a 1-year prevalence of TLR of 2.4% to 3.8%, TLMI of 0.4% to 3.3%, and cardiac death of 1.0% to 1.3%.9,11,16 Culotte has as 1-year prevalence of TLR of 6.7%, TLMI of 5.3%, and cardiac death of 1.0%.16 DK nanocrush has a 1-year prevalence of TLR of 4.8%, TLMI of 0%, and cardiac death of 2.4%, albeit this was a small observational study of 42 patients.13 Further descriptive statistics can be found in Table 5.
Table 5.
Adverse clinical outcomes at 1 year in patients undergoing various upfront 2-stent bifurcation stenting techniques.
| Technique | n | TLR | TLMI | Cardiac death |
|---|---|---|---|---|
| Jailed stent-balloon | 63 | 3 (4.8) | 2 (3.2) | 3 (4.8) |
| DK Crush (in DKCRUSH-III) | 210 | 5 (2.4) | 7 (3.3) | 2 (1.0) |
| DK Crush (in DKCRUSH-V) | 240 | 9 (3.8) | 1 (0.4) | 3 (1.3) |
| Culotte (in DKCRUSH-III) | 209 | 14 (6.7) | 11 (5.3) | 2 (1.0) |
| DK nanocrush | 42 | 2 (4.8) | 0 (0.0) | 1 (2.4) |
Values are n (%). Use caution when comparing outcomes between techniques, as patient populations and study design differed significantly between each study. To avoid undue comparison, no inferential statistics were performed comparing different techniques.
TLMI, target lesion myocardial infarction; TLR, target lesion revascularization.
Notably, the studies from which these descriptive statistics are derived are vastly different in design and study population, so direct comparison should be cautioned. For example, in the studies investigating outcomes of patients undergoing classic DK crush and culotte stenting, patients with ISR or requiring atherectomy were excluded from the study, whereas in this study of jailed stent-balloon bifurcation stenting, 49.2% of patients required atherectomy and 14.3% had ISR. Interestingly, subgroup analysis did not find an association between adverse events at 1 year and the need for atherectomy or mechanical support, while there was a greater hazard of adverse events occurring in patients requiring an intervention for ISR. We include these data to demonstrate that event rates with the jailed stent-balloon technique fall within a reasonable range of other upfront 2-stent bifurcation stenting techniques, even through direct comparison is limited by differences in study population, design, adverse event adjudication, and end point definitions.
Limitations
This was a nonrandomized, retrospective cohort study, so it is not possible to directly compare the clinical outcomes of the jailed stent-balloon technique with those of other contemporary bifurcation stenting techniques. Future randomized clinical trials are necessary. Additionally, the generalizability of the technique to other operators is not clear, as, although the jailed stent-balloon technique has fewer steps than other techniques, all operators in this study are high-volume operators, performing >100 PCIs per year at a tertiary academic center. Finally, the follow-up period was limited at 1 year. Further studies are warranted to better understand long-term clinical outcomes.
Conclusion
The jailed stent-balloon bifurcation stenting technique appears to be safe and feasible, with reduced technical complexity compared with that of other 2-stent bifurcation stenting techniques. However, further studies are needed to ensure the efficacy of the technique relative to other contemporary approaches.
Acknowledgments
Declaration of competing interest
Jeffrey S. Tran is an associate statistical editor for the American Heart Association (Circulation: Heart Failure). Prashanth Thakker is a consultant for Philips, Biotronik, Amgen, and Shockwave Medical. Marc Sintek is a consultant for Philips, Edwards Lifesciences, Biotronik, and Shockwave Medical. Alan Zajarias is a consultant for Edwards Lifesciences, Boston Scientific, and Medtronic. John Lasala is a consultant for Boston Scientific, and Abbott. Jasvindar Singh is a consultant for Philips-Volcano, Boston Scientific, Abbott Vascular, Medtronic Vascular, Abbott Vascular/CSI, Penumbra, Shockwave Medical, and Abiomed. The other authors reported no financial interests.
Funding sources
This work was not supported by funding agencies in the public, commercial, or not-for-profit sectors.
Ethics statement and patient consent
The study was approved by the institutional review board (No. 202010224) with a waiver of consent given the retrospective nature of the study involving minimal risk.
Footnotes
To access the supplementary material accompanying this article, visit the online version of the Journal of the Society for Cardiovascular Angiography & Interventions at 10.1016/j.jscai.2025.103880.
Supplementary material
References
- 1.Al Suwaidi J., Yeh W., Cohen H.A., Detre K.M., Williams D.O., Holmes D.R., Jr. Immediate and one-year outcome in patients with coronary bifurcation lesions in the modern era (NHLBI dynamic registry) Am J Cardiol. 2001;87(10):1139–1144. doi: 10.1016/s0002-9149(01)01482-5. [DOI] [PubMed] [Google Scholar]
- 2.Fanaroff A.C., Zakroysky P., Wojdyla D., et al. Relationship between operator volume and long-term outcomes after percutaneous coronary intervention. Circulation. 2019;139(4):458–472. doi: 10.1161/CIRCULATIONAHA.117.033325. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Lassen J.F., Holm N.R., Banning A., et al. Percutaneous coronary intervention for coronary bifurcation disease: 11th consensus document from the European Bifurcation Club. EuroIntervention. 2016;12(1):38–46. doi: 10.4244/EIJV12I1A7. [DOI] [PubMed] [Google Scholar]
- 4.Maznyczka A., Arunothayaraj S., Egred M., et al. Bifurcation left main stenting with or without intracoronary imaging: Outcomes from the EBC MAIN trial. Catheter Cardiovasc Interv. 2023;102(3):415–429. doi: 10.1002/ccd.30785. [DOI] [PubMed] [Google Scholar]
- 5.Burzotta F., Louvard Y., Lassen J.F., et al. Percutaneous coronary intervention for bifurcation coronary lesions using optimised angiographic guidance: the 18th consensus document from the European Bifurcation Club. EuroIntervention. 2024;20(15):e915–e926. doi: 10.4244/EIJ-D-24-00160. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Hildick-Smith D., Behan M.W., Lassen J.F., et al. The EBC TWO Study (European Bifurcation Coronary TWO): a randomized comparison of provisional T-stenting versus a Systematic 2 stent culotte strategy in large caliber true bifurcations. Circ Cardiovasc Interv. 2016;9(9) doi: 10.1161/CIRCINTERVENTIONS.115.003643. [DOI] [PubMed] [Google Scholar]
- 7.Zhang J.J., Ye F., Xu K., et al. Multicentre, randomized comparison of two-stent and provisional stenting techniques in patients with complex coronary bifurcation lesions: the DEFINITION II trial. Eur Heart J. 2020;41(27):2523–2536. doi: 10.1093/eurheartj/ehaa543. [DOI] [PubMed] [Google Scholar]
- 8.Chen S.L., Sheiban I., Xu B., et al. Impact of the complexity of bifurcation lesions treated with drug-eluting stents: the DEFINITION study (Definitions and impact of complEx biFurcation lesIons on clinical outcomes after percutaNeous coronary IntervenTIOn using drug-eluting steNts) JACC Cardiovasc Interv. 2014;7(11):1266–1276. doi: 10.1016/j.jcin.2014.04.026. [DOI] [PubMed] [Google Scholar]
- 9.Moroni F., Shue-Min Yeh J., Attallah A., et al. Crush techniques for percutaneous coronary intervention of bifurcation lesions. EuroIntervention. 2022;18(1):71–82. doi: 10.4244/EIJ-D-21-00690. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Colombo A., Moses J.W., Morice M.C., et al. Randomized study to evaluate sirolimus-eluting stents implanted at coronary bifurcation lesions. Circulation. 2004;109(10):1244–1249. doi: 10.1161/01.CIR.0000118474.71662.E3. [DOI] [PubMed] [Google Scholar]
- 11.Chen S.L., Zhang J.J., Han Y., et al. Double kissing crush versus provisional stenting for left main distal bifurcation lesions: DKCRUSH-V randomized trial. J Am Coll Cardiol. 2017;70(21):2605–2617. doi: 10.1016/j.jacc.2017.09.1066. [DOI] [PubMed] [Google Scholar]
- 12.Ge L., Iakovou I., Cosgrave J., et al. Treatment of bifurcation lesions with two stents: one year angiographic and clinical follow up of crush versus T stenting. Heart. 2006;92(3):371–376. doi: 10.1136/hrt.2005.061531. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Ray S., Mukherjee P., Bandyopadhyay S., Karmakar S., Mitra S., Bhattacharjee P. A novel nano-crush technique for the management of coronary bifurcation lesions: in vitro bench test analysis and preliminary report on real-world clinical evaluation in patients with one-year angiographic follow-up. AsiaIntervention. 2019;5(1):41–51. doi: 10.4244/AIJ-D-18-00017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Simsek B., Kostantinis S., Karacsonyi J., et al. C-6: challenges and solutions in double kissing crush stenting technique in bifurcation percutaneous coronary intervention. J Soc Cardiovasc Angiogr Interv. 2022;1(3) doi: 10.1016/j.jscai.2022.100158. [DOI] [Google Scholar]
- 15.Singh J., Patel Y., Depta J.P., et al. A modified provisional stenting approach to coronary bifurcation lesions: clinical application of the “jailed-balloon technique.”. J Interv Cardiol. 2012;25(3):289–296. doi: 10.1111/j.1540-8183.2011.00716.x. [DOI] [PubMed] [Google Scholar]
- 16.Chen S.L., Xu B., Han Y.L., et al. Comparison of double kissing Crush versus Culotte stenting for unprotected distal left main bifurcation lesions: results from a multicenter, randomized, prospective DKCRUSH-III study. J Am Coll Cardiol. 2013;61(14):1482–1488. doi: 10.1016/j.jacc.2013.01.023. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.