Developments in image-based colorimetric analysis methods and applications of CIElab color space in pharmaceutical sciences: A narrative review

Sahar Marefat a,b, Ali Shayanfar c,⁎⁎, Farnaz Monajjemzadeh b,⁎
PMCID: PMC12634856  PMID: 41280507

Abstract

The evaluation of color in pharmaceutical materials and products serves as an essential method for assessing physical appearance, stability, and overall quality control. Any recognizable color change in pharmaceuticals may lead to failure in meeting quality objectives. Traditional visual examinations, while commonly used, are subjective and prone to inconsistency, making them inadequate for precise assessment. As a result, advanced instrumental techniques have gained prominence, with CIE (Commission Internationale de l'Eclairage) Lab color space being widely recognized for its accuracy and applicability. Developed by the International Commission on Illumination, the CIELab system characterizes color using three components: L (lightness), a⁎ (red-green axis), and b⁎ (yellow-blue axis), providing a quantitative, standardized approach for color measurement. This method has been extensively utilized in pharmaceutical research and industry for diverse applications, such as quality control, stability studies, and batch-to-batch consistency evaluations. The present review aims to discuss studies that have employed this method in pharmaceutical color assessment, and related quality control issues.

Keywords: CIELab, Tristimulus colorimetry, Stability, Photostability

Graphical abstract

Unlabelled Image

Highlights

  • •

    Introduces colorimetric analysis as a complementary tool in pharmaceutical formulation assessment.

  • •

    Emphasizes its potential for monitoring stability through visible chromatic shifts over time.

  • •

    Highlights tristimulus colorimetry, especially CIELab color space as a fast, low-cost, and widely accessible method.

  • •

    Suggests color change tracking as a non-invasive approach to detect adulteration in drug products.

  • •

    Positions colorimetric data within broader strategies for pharmaceutical quality assurance and control.

1. Introduction

Color plays a vital role in the pharmaceutical industry, influencing not only the visual characteristics of the product but also its quality and safety. The color of a medication affects patient compliance, as distinct shades help differentiate drugs, thereby reducing the risk of medication errors (Wirth, 1991). Additionally, changes in a drug's color can signal potential issues such as contamination, impurities, or degradation, making it essential to monitor and maintain color integrity throughout the product's shelf life (Yamazaki et al., 2010; Berberich et al., 2002). During stability testing, color changes are typically assessed through visual examination of the dosage form. While this method may be sufficient for end-product quality control, it is prone to observer subjectivity and inconsistency. Since direct visual evaluation depends on individual perception, alternative analytical approaches have been introduced (Wirth, 1991). Tristimulus colorimetry is a fundamental technique for quantifying color based on three primary components—X, Y, and Z values—that correspond to the human eye's response to color. This method follows the International Commission on Illumination (CIE) system, which standardizes color measurement to reflect human vision. CIE XYZ color space, commonly used in graphics, is derived from spectral curves known as color matching functions or the standard observer. It is preferred over other CIE spaces due to its advantageous numerical properties. Simple transformations, allow XYZ values to be converted into various widely used color spaces, including RGB, which is derived from the initials of the three additive primary colors: red, green, and blue (Fairman et al., 1997), YUV color space functions similarly to the human retina. The Y channel (luminance or “luma”) represents brightness, similar to how rod cells in the retina detect light intensity, especially in darkness (Podpora et al., 2014). Lab color space, also known as CIELab, is one of the most reliable tools for evaluating color variations in pharmaceuticals. Designed to align with human vision, this model offers a consistent and objective method for assessing color differences. By utilizing the Lab color space, manufacturers can ensure that their products maintain stable color quality, safeguarding both safety and efficacy. For defining this method in detail, it can be stated that the CIELab color space, also known as Lab, is a color model developed by CIE in 1976 (Flavier et al., 2022). It is designed to be a perceptually uniform space, meaning that a given numerical change corresponds to a similar perceived change in color. This makes it particularly useful for applications requiring precise color measurements and comparisons, such as in the pharmaceutical industries. In absolute color measurement, the a⁎ axis represents the opposition between green and red, with negative values indicating a shift toward green and positive values toward red. Similarly, the b⁎ axis corresponds to the contrast between blue and yellow, where negative values lean toward blue and positive ones toward yellow. Lastly, the L⁎ value ranges from 0, denoting ideal absolute black, to 100, which is signifying ideal absolute white (Durmus, 2020). Color alterations can subsequently be measured in relation to a control. One specific technique for determining the overall perception of color change is ΔE⁎, which is computed using the formula provided in Eq. 1 (Flavier et al., 2022). By using these values, manufacturers can objectively quantify color variations and ensure consistent color quality across different batches of products (Sakiroff et al., 2022). The color measurement technique is a non-destructive analytical method that allows for the assessment of optical properties without compromising the integrity of the sample (Dwiecki et al., 2022). This is particularly important in pharmaceuticals, where color change can indicate issues such as contamination, impurities, or degradation.

ΔE∗=L∗−L0∗2+a∗−a0∗2+b∗−b0∗2 (1)

Fig. 1 shows the basis for the parameters of this colorimetry. The CIELab color system has been part of the United States Pharmacopeia's general chapters since 1985 (Pharmacopeia, United States, 2024), and is widely used across various industries for its ability to quantify color differences in a way that aligns with human perception. In the textile industry, CIELab is essential for ensuring color consistency in fabrics, as even minor deviations can affect product quality (Zehra and Bashir, 2019). The food industry relies on CIELab to evaluate the appearance of products, such as, fruits, vegetables and packaged goods, where color is a key indicator of freshness and quality (Brühl and Unbehend, 2021). In the automotive sector, it helps maintain uniformity in paint colors across different batches, ensuring a seamless finish (Gómez et al., 2016). Additionally, the printing and packaging industries use this method to achieve accurate color reproduction, which is critical for brand identity and consumer appeal (Li and Meng, 2014).

Fig. 1.

Fig. 1

CIELab color space parameters adapted and redrawn based on (Ly et al., 2020) and (Bisulca et al., 2012).

This article attempts to review studies based on the CIELab color measurements to analyze color changes in pharmaceuticals.

2. Search strategy

A comprehensive literature search was conducted across PubMed, Google Scholar, and Scopus, covering the period from 1970 to 2024. Keywords such as “tristimulus colorimetry”, “CIELab color space”, “CIELab color measurement”, and “pharmaceuticals color evaluations” were used individually or in combination. The search yielded 32 relevant articles, all of which were selected for review and evaluation. The extracted data were categorized into four main groups: stability and quality control, pharmaceutical coating process evaluation, appearance and color uniformity analysis, and other utilities.

3. Stability and quality control

3.1. Stability

One of the most critical aspects of pharmaceutical evaluation is the assessment of drug stability (Alvani and Shayanfar, 2022; Khajir et al., 2021). The color of pharmaceutical products serves as a key parameter for stability evaluation. Conventional color assessment methods are often time-consuming and costly. Among the rapid techniques available, tristimulus colorimetry stands out as an efficient approach. In the following studies, the application of one of the most essential tristimulus evaluation methods, known as CIELab, has been examined for stability assessment (Wirth, 1991).

Some of these studies measured color changes to investigate the stability of drugs. In 1996, Stark et al., used CIELab values to evaluate color changes in four white solid dosage forms (theophylline tablets, cefoxitin sodium powder for injection, flucloxacillin sodium capsules, and captopril tablets) in the packaging provided by the manufacturer during stability testing. The average storage temperature ranged between 20 °C and 23 °C, while the relative humidity (RH) varied from 45 % to 70 % for long term studies. Stress conditions included 70, 55 and 40 °C, 50 % RH and 40 °C, 75 % RH for 84 and 168 days. They measured the tristimulus parameters to detect any discoloration, such as yellowing or browning, which could indicate degradation or instability of the product. This study revealed a linear relationship between color change and drug content for all except theophylline, where color changed with no degradation. They concluded that this type of color space tool provided a quantitative way to assess color changes and ensure the stability of the dosage forms over time (Stark et al., 1996).Years later, other studies were conducted on this topic, which will be discussed later. Berberich et al., in 2002, conducted a study on white uncoated tablets which contained an active ingredient with a structure of a piperidine derivative with different excipients (lactose, corn starch, magnesium stearate, and microcrystalline cellulose) and measured the color change by MicroColor® I, after storage at 25°, 40°, and 60 °C in glass bottles for 24 weeks. Along with CIELab method, they also used CI (color intensity) measurements in their work, which has been established to effectively characterize the discoloration kinetics, achieving satisfactory regression coefficients. CI values enable the calculation of discoloration kinetics using regression methods within reliable confidence intervals. Discoloration rates follow the Arrhenius equation, allowing estimation of activation energy. These values directly relate to the visual perception of tablets and help predict how long uncoated tablets remain white (Berberich et al., 2002).

Later in 2008, Rhee et al., used color change analysis techniques including CIELab color space Minolta CR-241 spectrocolorimeter to assess the stability of rabeprazole-excipient mixtures in the form of tablet at initial and 3 days of storage under the accelerated conditions (40 ± 2 °C and 75 % ± 5 RH) and stress condition (60 ± 2 °C), the device that they used excluded the spectral component, and was calibrated daily using a white reference and darkness to ensure accuracy. They compared the results obtained from this method with the results of a HPLC analysis obtained after the first, second, third, and fourth weeks (Rhee et al., 2008). A few years later, Zhou et al. used a Varian Cary 50 dispersive spectrometer (Agilent, Palo Alto, CA, USA) equipped with a transmittance stage and quartz cuvettes for UV–Vis transmittance spectra to measure color attributes based on three primary color values. To validate the effectiveness of tristimulus colorimetry, the authors compared the results with those obtained from visual inspection and spectrophotometry. The comparison demonstrated that this method offers greater accuracy, reproducibility, and reliability, making it a superior choice for color measurement in pharmaceutical applications. They also provided a detailed framework for implementing this method, including guidelines for instrument calibration, sample preparation, and data analysis (Zhou et al., 2011). Flavier and colleagues conducted a similar study on l-ascorbic acid powder and indigo carmine tablets (the excipients were microcrystalline cellulose, starch, and alpha-d-lactose) in 2022 and compared the results with long-term stability studies based on color analysis. For color measurements, the tablets were positioned in a wooden retainer aligned with the aperture, while l-ascorbic acid powdered samples were prepared by filling 0.924 ml vials to a 1-cm height. They used ColorQuest® XE Colorimeter paired with EasyMatch®QC software and ASAPprime® (version 6.0.1) to simulate the anticipated color variation in indigo carmine tablets and l-ascorbic acid powder, the ASAP method utilizes the moisture-adjusted Arrhenius equation (Eq. (2)) to construct the isoconversion model. Where k is the reciprocal of the time to reach the specification limit (isoconversion time), Ea is the activation energy A is the collision frequency, R is the gas constant, B is the humidity sensitivity factor, and T is the temperature. In an ASAP study, ΔE serves as a tool for predicting color variations over time, influenced by temperature and relative humidity (Flavier et al., 2022). In all three of these studies, it was found that the colorimetric methods provide acceptable, efficient and straightforward results for conducting stability studies and achieving reliable results (Flavier et al., 2022; Zhou et al., 2011; Rhee et al., 2008).

Lnk=lnA–Ea/RT+BRH (2)

In 2016, Mochizuki et al., evaluated color changes in acetaminophen solutions. Acetaminophen can be oxidized and hydrolyzed easily in the presence of oxygen and high temperature and light intense. In this study, the researchers used the tristimulus colorimetry by Konica Minolta CM-3500d spectrophotometer to evaluate the color change of the drug solutions. Each brown glass bottle of acetaminophen solution contains both dissolved oxygen (DO) within the liquid and gaseous oxygen in the headspace (HS). Variations in the levels of DO and HS oxygen may contribute to differences in the chemical transformations occurring during storage. To examine this effect, the impact of varying filling volumes—and consequently different HS oxygen levels—on color changes in the solution was investigated. Additionally, with controlled oxygen levels, a prediction method for the solution's color shifts was explored. To better understand the role of oxidation in these changes, two HS conditions (3 ml and 40 ml) were maintained in a stability chamber for observations. The device operated under the standard illuminant daylight source adhering to the CIELab color system. They found that oxygen levels significantly affect acetaminophen solution's color. Based on this study higher oxygen causes darker browning than lower oxygen levels at high temperature condition. They compared the kinetic data obtained from long-term color evaluations with experimental methods, each measurement was highly reproducible with a relative standard deviation of less than 0.2 %, thus, this instrument ensured accurate and consistent color analysis which revealed a rapid and suitable method for evaluating stability and performing quality control tests of acetaminophen in pharmaceutical solutions (Mochizuki and Takayama, 2016).

Sakiroff et al. in 2022, compared the results of spectrophotometric chromaticity measurements which includes tristimulus colorimetry compared with those observed through visual examination using European Pharmacopeia (Ph. Eur.) color reference solutions (Fig. 2).

Fig. 2.

Fig. 2

European Pharmacopeia color reference range solutions used for visual examination: (A) set of color B (brown) and (B) set of color Y (yellow) adopted from (Sakiroff et al., 2022). (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.)

According to Ph. Eur. color reference solutions are standardized liquid calibration materials used in pharmaceutical analysis to evaluate solution coloration. These reference solutions are prepared using three primary metal salts—ferric (III) chloride, cobalt (II) chloride, and copper (II) sulphate pentahydrate—blended in specific ratios to produce a range of standardized color. They play a crucial role in ensuring consistency and quality control in pharmaceutical formulations. Typically supplied in tamper-evident ampules, these solutions have a regulated shelf life and are accompanied by a certificate of analysis to meet compliance standards (European Pharmacopoeia, 2021; Pharmacopia, United State, 2024). In this study, they prepared samples of paracetamol and parenteral nutrition solutions, separately and subjected them to stress conditions by exposure to UVA radiation at 0.60 ± 0.10 W/m2 and white light with an average intensity of 6600 ± 200 lx and a certified climate-controlled chamber (Binder GmbH, Tuttlingen, Germany) maintained at 25 ± 1 °C with 60 ± 1 % relative humidity for up to 46 days. Color change recorded with spectrophotometric measurements by Jasco V-670 spectrometer (Jasco Corporation, Lisses, France) using a quartz measuring UV cell. This method detected color variations earlier and more precisely than visual examination in a double-blind randomized study in the case of lightening by an LED lamp emitting diffused white light (color rendering index of 85, 4000 K) and placed against a white paper background to enhance color contrast. They analysed the data using Stata software (Version 15, StataCorp, College Station, US) and demonstrated that it is a more accurate and sensitive method for evaluating color changes (Sakiroff et al., 2022).

3.2. Photostability

Light-induced degradation is a key factor affecting drug stability. Certain pharmaceutical compounds and excipients exhibit high sensitivity to light, leading to substantial deterioration that compromises both the quality and stability of the final formulation (Janga et al., 2018).

In 1978, Matsuda et al., measured the color of the coated and uncoated sulfisomidine tablets using a tristimulus colorimeter, which provided L, a, and b values in CIELab color space. They compared the color measurements of the coated tablets to those of the uncoated tablets which were stored under 400-w mercury vapor lamp and UV rays (300–400 nm) as shown in Fig. 3 and the temperature was maintained below 27 °C. The method was used to assess the effectiveness of the UV-absorbing film-coating on the surface of sulfisomidine tablets in terms of photostability and the extent of protection against color changes, thus reducing photolytic degradations (Matsuda and Nakanishi, 1978).

Fig. 3.

Fig. 3

Sample holder for evaluation color change adopted from (Matsuda and Nakanishi, 1978).

Ramadan et al. in 2006, investigated a surface treatment technique to enhance the photostability and solid-state properties of methyldopa, a photosensitive drug.

Methyldopa was treated with different solvents (ethanol, methanol, and isopropanol) and placed in Petri dishes and kept in a light stability cabinet with precise temperature control, set at 40 ± 1 °C and 75 ± 5 % relative humidity. Their upper surfaces were exposed to light from two LUMILUX Daylight fluorescent lamps (60 cm long, rapid-start type, 18 W), positioned 20 cm above each sample. The illuminance was maintained at 2600 lx, and the samples underwent a 10-day exposure, accumulating 1.2 million lx h, following ICH guidelines (Internatinal Council for Harmonisation, 2021). To evaluate the effects of temperature and humidity without light exposure, control disks were wrapped in aluminium foil for the duration of the study. Photostability test was conducted on solvent-treated methyldopa disks both with and without antioxidants. Superchroma integrating sphere-type color difference meter (BRAVE Instruments, Belgium) was used for measuring the color changes in the CIELab color space before and after surface treatment. Each measurements was averaged over three replicates to ensure precision, and visual evaluations by an observer panel consisting of three persons were also conducted. They found that certain tablet surface treatments by solvents could significantly reduce the rate of photodegradation, thereby enhancing the photostability. The study also examined how different antioxidants influenced methyldopa's photostability in ethanol over five days. Butylated hydroxytoluene did not improve ethanol's stabilizing effect, while ascorbic acid significantly enhanced methyldopa's photostability. Cysteine had a moderate stabilizing effect. Yellow discoloration appeared in samples treated with ascorbic acid, likely due to its oxidation, which increased b⁎ values after days. Ascorbic acid oxidized in solid form, causing tablet discoloration during storage under typical temperature and humidity conditions (Ramadan et al., 2006).

In a study conducted in 2009 by Sue-Chu et al., two different qualities (synthetic and biosynthetic) of riboflavin were prepared in both solid state and various tablet formulations prepared by different methods. They utilized the CIELab color space to quantify color changes under light exposure, providing an objective measure of riboflavin's photodegradation. Riboflavin tablets were stored under controlled light conditions to stimulate real world storage. After light exposure, the initially yellow tablets appeared green, possibly due to a loss of yellow coloration rather than the presence of a green component. However, none of the excipients exhibited a green color or changed after exposure. The most noticeable color shifts occurred in wet granulated tablets containing biosynthetic Riboflavin. When covered with a UV filter, the tablets showed a similar color change, indicating that visible light, rather than UV radiation, was responsible for the color change. Tablets made via wet granulation experienced the most significant color changes, whereas those produced by direct compression showed smaller shifts, independent of Riboflavin quality. Switching from wet granulation to direct compression appeared to help stabilize tablet color, especially for biosynthetic Riboflavin-containing samples. The green discoloration vanished when the tablets were dissolved in methanol or aqueous solutions, a finding supported by UV-VIS absorption measurements, which showed no spectral changes post-exposure. This was true for both the drug substance and the scraped-off colored layer, with exposed and non-exposed samples having identical normalized absorption spectra. They quantitatively evaluated the photoinduced color changes and obtained photoreactivity and stability (Sue-Chu et al., 2009).

Then in 2013, Baertschi et al., used CIELab color measurements to evaluate how light exposure affects the stability of injectable drugs during the clinical preparation and usage of the injectable dosage forms. They measured the values before and after light exposure. These color measurements helped them detect any photodegradation, which would manifest as changes in the CIELab values, indicating loss of stability and effectiveness. They later published their findings and proposed guidelines for photostability testing and labeling of injectable drugs under conditions of use as a commentary published in 2013 (Baertschi et al., 2013).

Last year, Matsushima et al., used amlodipine besylate tablets, a light sensitive drug, and exposed them to light to observe color changes. The core tablets were coated using a pan coater with excipients including d-Mannitol, microcrystalline cellulose, and magnesium stearate. Three different coating materials were applied: Hypromellose (HPMC), polyvinyl alcohol (PVA), and Kolli coat IR, a copolymer of ethylene glycol and vinyl alcohol. Ferric oxide was incorporated to enhance color differentiation in the coating and allow accurate measurement of ΔE⁎, as white coatings without ferric oxide tend to exhibit unclear color changes. Ferric oxide (FO) is a stable red pigment that remains chemically unchanged when exposed to light. The film-coated tablet was exposed to light from the D65 lamp at an intensity of 3500 lx inside the photostability chamber. The chamber's internal temperature was kept at 25 °C, while the relative humidity (RH) was regulated to 75 % with a saturated sodium chloride (NaCl) solution. By using CIELab color measurement by Nippon Daneshku SE6000 spectrophotometer, they found that titanium dioxide, a photocatalyst, can induce chemical changes in Hypromellose upon light exposure, leading to color changes. The color change in film-coated tablets containing only titanium dioxide (TiO2) increased rapidly in the first two weeks but slowed down afterward, similar to tablets containing both TiO2 and talc. However, tablets coated with only talc showed minimal color change. The degradation product (DP) was assessed, revealing that DP did not increase in tablets with TiO2 after 28 days of light exposure. Instead, an increase was observed in tablets coated with talc alone. Overall, DP levels in coated tablets were lower than those in uncoated ones. These findings suggest that TiO2 significantly impacts color changes due to light exposure while also being crucial for maintaining the photostability of amlodipine besilate. The study highlights the importance of selecting appropriate coating materials to prevent photodegradation while protecting the active ingredient from light (Matsushima et al., 2024). The Maillard reaction has previously been demonstrated for amlodipine (Ghaderi and Monajjemzadeh, 2020), and the united States Pharmacopeia also contains cautions regarding the use of various excipients with amlodipine (Pharmacopeia, United States, 2024). Our team has previously worked on Maillard reaction, it seems that further research is needed to determine whether the extent of Maillard reaction can be examined using this method (Monajjemzadeh et al., 2009).

3.3. Counterfeits

The incidence of counterfeit pharmaceuticals has surged in recent years, affecting both non-essential lifestyle products and critical medications. Beyond the serious risks to public health, pharmaceutical companies suffer significant financial losses and reputational damage. One of the fastest and most convenient methods for detecting counterfeit pharmaceutical products is color evaluation. The following study presents an application of the CIELab color space for counterfeit detection (Martino et al., 2010).

For quality control, a study was conducted by Al Haghbani et al. in 2016, on several pharmaceutical tablets. In this study, as well as in another study he conducted the following year on Viagra® tablets purchased from several online stores, they used the HunterLab ColorQuest® XE tristimulus colorimeter to measure the color differences, and concluded that this colorimetric method can be used to identify counterfeit pharmaceutical tablets. The instrument operated under standardized conditions, including daylight illumination and an observation angle, ensuring consistent and accurate results. Sildenafil tablets with different brands were securely positioned at the reflectance port using a spring-loaded sample clamp, with a specific viewing area, and a magnetic white ceramic disk ensured a uniform white background. They reported that a tablet characterized by a specific color, has distinct color parameters that differentiate it from counterfeit types. The results of these studies showed that the CIELab color space measurements can be an acceptable, fast, and accurate method for distinguishing counterfeit and genuine pharmaceutical tablets, especially in regions where accurate and expensive analytical methods are not available (Al Hagbani et al., 2016; Al Hagbani et al., 2017).

3.4. Quality control

In 2007, Siddiqui et al., explored the correlation between tablet surface color and tensile strength, aiming to establish a rapid, non-destructive quality control method for detecting deviations in tablet hardness. They hypothesized that compression pressure influences surface characteristics, which in turn affects colorimetric parameters measured by ColorQuest XE colorimeter. To validate the effectiveness of CIELab colorimetry, the study was compared with near-infrared spectroscopy (NIR) results, a widely used non-invasive method for assessing tablet hardness. While NIR provides accurate hardness measurements, it is more complex and costly. In contrast, CIELab colorimetry offers a more simple, cost-effective alternative that aligns with the FDA's Process Analytical Technology (PAT) initiative, which promotes real-time quality monitoring during pharmaceutical manufacturing (Siddiqui and Nazzal, 2007). Pack et al. in 2015, focused on the importance of modernizing color measurement practices in the pharmaceutical industries. By replacing subjective visual methods with quantitative tristimulus colorimetry, companies can improve the consistency and reliability of color assessments, enhance regulatory compliance by providing objective, data-driven results, and facilitate global harmonisation of color measurement standards. They collected the data using either the Hunter Labs UltrascanPro (Hunter Associates Laboratory, Reston, Virginia) or the BYK Gardner LCS IV (BYK-Gardner, Columbia, Maryland) (Pack et al., 2015).

A few years later, in 2020, Meszaros et al., used a Basler acA4112-30uc area scan color camera, equipped with a complementary metal-oxide-semiconductor (CMOS) sensor as shown in Fig. 4 for tristimulus colorimetry to predict the drug content of meloxicam tablets. The camera was linked to the computer via a USB 3.0 interface. Illumination was supplied by either a ring light featuring three rows of white light-emitting diodes for visible lighting or a UV lamp composed of four UV LEDs, emitting light within the 380–395 nm wavelength range. They developed and validated algorithms based on the CIELab color space. The results showed that the designed algorithms were successful in predicting the drug content accurately, demonstrating the potential of digital UV/visible imaging combined with tristimulus colorimetry as a rapid and reliable tool for tablet inspection (Mészáros et al., 2020).

Fig. 4.

Fig. 4

The design of the image-capture process adopted from (Mészáros et al., 2020).

In a study conducted in 2022 on Copper (ΙΙ) histidine injections, their color was examined using the NH 310 Portable Colorimeter (Sheenzhen ThreeNH Technology Co., Ltd., Sheenzhen, China) to determine the quality of the color.

The injections were kept in a refrigerator under controlled conditions at 4 °C (±0.1 °C), in an uncontrolled environment at 2–8 °C, or in a climatic chamber with regulated temperature and humidity settings (25 °C/60 % RH or 40 °C/75 % RH), at specific intervals. Color stability is an important quality parameter in formulations as well as in the quality of the inherently colored products. The colorimetric data in this study were compared with other quality control methods such as UV spectroscopy and the analyst's observations of color variations served as the basis for exploring the potential use of color measurement techniques within the CIELab color system which serves as an alternative to the previous methods. The results showed that this method could be a reliable, accurate, and non-invasive for quality control (Dwiecki et al., 2022).

4. Evaluation of pharmaceutical coating processes

In 2022, Palugan et al., applied different coatings to the tablets containing caffeine, using both titanium dioxide and alternative materials such as ready to use (R2U) coating formulations: AQP (AquaPolish® F White), AQR (Aquarius™ PRIME TF), FOR (ForCoat® White), INS (Instanute® IR White), NUT (Nutrafinish® White), SEP (Sepifilm® LP White TF), SHE (SheffCoat™ TF), SPE (Spectrablend® CC White) and VIV (Vivacoat® N White). With a tristimulus colorimeter, they measured the colors of these coated tablets specifically looking at L, a, and b values. They compared the color values from the titanium dioxide (TiO2)-coated tablets with those coated using alternative materials and analysed the data using one-way analysis of the variance (ANOVA). Tristimulus colorimetry allowed them to objectively measure and compare the color quality of different coatings to find suitable replacements for TiO2 (Palugan et al., 2022). One year later James et al., used the same method to analyze TiO2–containing and TiO2 free alternatives focusing on coverage and quality control and obtained acceptable results (James et al., 2023).

In 2019, Murillo et al., used CIELab colorimetry method to monitor and evaluate the color changes of placebo tablet cores throughout the coating process. The coating process lasted 130 min, with spraying paused every 10 min to randomly extract 21 samples. These intervals align closely with standard pharmaceutical industry standards for process documentation and traceability during tablet coating, enabling effective monitoring and error correction. Color measurements were performed on each batch of 21 samples taken at 10-min intervals using the DigiEye imaging system. This system features two light sources that replicate the CIE D65 illuminate and a specialized chamber that ensures diffuse, uniform lighting on the sample tray. The illumination setup minimizes minor irregularities, reduces shadows, and prevents glare on the surfaces of the samples, which, in this case, exhibit no visible gloss. They measured the color values at various stages, allowing them to assess the uniformity and consistency of the coating application. Through these measurements, the researchers were able to optimize the coating procedure and achieve acceptable final results (Murillo et al., 2019).

5. Appearance and color uniformity

Matsuda et al. in 1974, prepared a mixture of pharmaceutical powders such as lactose using colored particles to simulate different components. They used a tristimulus colorimeter to measure the color of the powder mixtures by obtaining L, a, and b values. Comparison of these color measurements enabled assessment of the uniformity and homogeneity of the mixing process possible. They looked for coloration differences that could indicate segregation or incomplete mixing by the color and color difference meter, model ND-K6B (Nippon Denshoku Kogyo, Tokyo). The tristimulus colorimeter data helped them identify areas where the mixing process could be improved to achieve a more uniform distribution of the Rose Bengal as the colored particles. This method enabled researchers to quantitatively evaluate the mixing process and ensure that the pharmaceutical powders were thoroughly mixed and remained uniform by the end of a certain manufacturing period (Matsuda et al., 1974).

Then, in 1975, Bogdansky explored the use of tristimulus colorimetry such as CIELab using a reflectance spectrophotometer fitted with a tungsten lamp to measure and analyze the surface color and color differences of tablet colorants. The study highlighted the importance of accurate color measurement in pharmaceutical tablets, as color can influence both consumer perception and product quality (Bogdansky, 1975).

Then in 1991, Wirth used L, a, and b values to assess the uniformity and consistency of the metoprolol tablets appearance in different strengths. White tablets stored for extended periods under accelerated conditions in hot and humid environments underwent color changes. In contrast, reference samples kept at 13 °C exhibited no discoloration. The measurements were conducted with the aid of a commercially available device from Dr. Lange GmbH (XENO COLOR LS 100).

The color of samples stored under various environmental conditions was visually assessed by multiple individuals, using reference samples from the same production batch stored at 13 °C. These subjective evaluations were recorded on a six-tier arbitrary scale, ranging from ‘no visible change’ (1) to ‘minute visible change’ (2), ‘small visible change’ (3), ‘moderate visible change’ (4), ‘off-white’ (5), and ‘change in color’ (6). In addition, the same tablets color characteristics were analysed using instrumental color measurement. This method provided a precise and objective way to evaluate tablet color, helping to ensuring consistent quality in pharmaceutical manufacturing, compared with visual inspection alone (Wirth, 1991).

Sucrose pellets were film-coated using a bottom-spray Wurster fluidized bed coater with a coating suspension containing hydroxypropyl methyl cellulose (film former), polyethylene glycol (plasticizer), and yellow iron oxide (color pigment) by Chan et al., in 2001. Four different methods were used to measure the color distribution over the pellet surfaces. These methods involved placing the pellets directly on the stage of the tristimulus colorimeter (Minolta CR-241) or using specially designed pellet sample holder. Color measurements were taken from multiple spots on each pellet to assess uniformity. The color uniformity of the color coat was indicated by the standard deviation of the color measurement values. By using tristimulus colorimetry, the researchers were able to quantitatively evaluate the uniformity of the color coating on the pellets, ensuring consistent quality in the coating process (Chan et al., 2001).

Several years later, in 2013, Hetrick et al., used the CIELab color space to measure and quantify the color of placebos that consisted of magnesium stearate, microcrystalline cellulose, starch, lactose monohydrate, and colloidal silicon dioxide, coated with six distinct colors, spanning shades from turquoise to dark blue. To measure color changes, the tablets were kept under controlled conditions for 21 days, with temperature and relative humidity settings as follows: 50 °C/75 % RH, 60 °C/40 % RH, 70 °C/75 % RH, 70 °C/10 % RH, and 80 °C/40 % RH. Colorimetric analysis was then conducted using a HunterLab LabScan XE, supported by EasyMatch QC software (HunterLab, Reston, VA). They correlated these measurements with visual ratings to remove subjectivity and conducted stability studies to rank the relative stability of color coatings. Their work provided a data-driven method for selecting and controlling the physical appearance of pharmaceutical products, enhancing quality and consistency (Hetrick et al., 2013).

After that, in 2014, Miyazaki et al., developed a novel method to determine the mass percentage of powdered medicine in mixtures without the need for expensive analytical equipment. The study focused on three specific drugs; Adona® powder 10 % (carbazochrome sodium sulfonate hydrate), Anginal® powder 12.5 % (dipyridamole), and Asverin® powder 10 % (tipepidinehibenzate), which were blended with lactose powder as a diluent, as seen in Fig. 5. Throughout the experiments, the room temperature was maintained between 20 °C and 22 °C, while relative humidity remained within the 35 %–45 % range. They captured digital photographs using a hand-held microscope (M3)  equipped  with  a  30 × magnification lens (30N) of the mixture of the powdered medicines with diluents. The color of the mixtures was analysed using CIELab color values. The results obtained from the image analysis were compared with those from traditional methods, such as UV spectroscopy to validate the accuracy and precision of the image analysis method. By using this method, the researchers were able to non-destructively and accurately assess the blending uniformity of the powdered mixtures, providing a reliable method for ensuring the correct proportions in pharmaceutical formulations (Miyazaki et al., 2014).

Fig. 5.

Fig. 5

Photos of A) Adona® B) Anginal® C) Asverin®, The variation in powder coloration is visually discernible adopted from (Miyazaki et al., 2014).

Yin et al., in 2017, used a ColorQuest® XE Colorimeter paired with EasyMatch®QC software for color measurements of protein solutions. Initially, six gravimetrically prepared Ph. Eur. color reference solutions were tested as unknown samples, ensuring that analysts remained unaware of their identities. Five analysts visually compared these samples against a set of thirty-seven Ph. Eur. color reference solutions using the visual assessment method and identified the closest match. Subsequently, the same samples were analysed using the HunterLab instrument. The study demonstrated that while the quantitative spectral method is comparable to the traditional visual assessment approach, it offers greater precision in determining color values (Yin et al., 2016).

6. Other utilities

In 1996, Akiyama et al., explored the use of a CM-503I, Minolta hand-held spectrophotometer for the quantitative evaluation of passive cutaneous anaphylaxis (PCA) which is a model used to study allergic reactions in the skin. In PCA reactions, vessel permeability at the IgE injection site is typically assessed by monitoring the leakage of Evans blue dye. The reaction's intensity is determined based on the amount of dye that escapes from the vessel at the injection site. To objectively quantify this dye leakage, researchers must carry out a series of labour-intensive and time-consuming procedures, including sacrificing the animals, excising the skin, extracting the dye from the tissue, and measuring its absorbance. In this study, the blue dye color at the injection site was analysed using a portable spectrophotometer.They found a good relationship between the dilution of sera and the ΔE values, which correlated well with the amount of Evans blue dye measured by conventional colorimetric methods (Akiyama et al., 1996).

In 2022, Peixoto et al., used a mobile device and recorded pictures to, evaluate several color spaces, including RGB, HSV and CIELab. They found that the coordinate a⁎ from the CIELab color space provided the highest sensitivity for detecting sulfamethoxazole. They also reported that this method enabled accurate and reliable colorimetric detection of antibiotic in environmental water samples using a mobile device (Peixoto et al., 2022).

Fan et al., in 2020, reviewed various color spaces, including RGB, CMYK, HSB/HSL, and CIELab and highlighted the advantages and challenges of using digital image colorimetry on smartphones. The widespread adoption of smartphones, the increasing reliance on digital images, and the continuous advancements in mobile applications have created new possibilities for developing rapid, affordable, and portable digital image colorimetry (DIC) for analysing various substances, including heavy metals, herbicides, pesticides, antibiotics, biological markers, natural compounds, and microorganisms. However, to ensure greater accuracy and reliability, further refinements are needed, particularly in standardizing geometric configurations and lighting conditions. They emphasized the potential of smartphones as a low-cost, rapid and sensitive tool for chemical analysis (Fan et al., 2021). Table 1 displays a summary of reviewed article.

Table 1.

Summary of reviewed articles on CIELab color evaluation.

Purpose Sample Compared with Analysis method Year Ref number Instrument
Stability testing theophylline tablets HPLC one-tailed t-test (P < 0.05) 1996 (Stark et al., 1996) reflectance spectrocolorimeter
Stability testing cefoxitin sodium powder for injection HPLC (R2 > 0.99, reproducibility<5 %) one-tailed t-test (P < 0.05) 1996 (Stark et al., 1996) reflectance spectrocolorimeter
Stability testing flucloxacillin sodium capsules HPLC one-tailed t-test (P < 0.05) 1996 (Stark et al., 1996) reflectance spectrocolorimeter
Stability testing captopril tablets HPLC one-tailed t-test (P < 0.05) 1996 (Stark et al., 1996) reflectance spectrocolorimeter
Stability testing piperidine derivative tablets not mentioned linear regression (regression coefficients >0.9) 2002 (Berberich et al., 2002) micro color® I, Dr. Lange; Colorflash Compakt, Optronik GmbH
Stability testing rabeprazole tablets HPLC (R2 > 0.9999, Intraday and interday precision: 0.71 and 0.57, Intraday and interday accuracy: 2.15 and 3.14) linear regression (correlation
coefficients >0.97)
2008 (Rhee et al., 2008) Minolta CR-241 spectrocolorimeter
Stability testing l-ascorbic acid powder Long term color change analysis Modeling⁎ (R2 = 0.97, Q2 = 0.96) 2022 (Flavier et al., 2022) ColorQuest® XE Colorimeter
Stability testing indigo carmine tablets Long term color change analysis Modeling⁎ (R2 = 0.98, Q2 = 0.93) 2022 (Flavier et al., 2022) ColorQuest® XE Colorimeter
Prediction of color changes acetaminophen solution Long term color change analysis linear regression (R2 > 0.98) 2016 (Mochizuki and Takayama, 2016) Konica Minolta CM-3500d spectrophotometer
Evaluating color changes paracetamol solution Visual examination linear mixed-effects model⁎⁎ 2022 (Sakiroff et al., 2022) Jasco V-670 spectrometer
Evaluating color changes parenteral nutrition solution Visual examination linear mixed-effects model⁎⁎ 2022 (Sakiroff et al., 2022) Jasco V-670 spectrometer
Photostability sulfisomidine tablet UV spectroscopy linear regression 1978 (Matsuda and Nakanishi, 1978) color difference meter equipped with a halogen lamp
Photostability methyldopa tablet HPLC, DSC, SEM linear regression 2016 (Ramadan et al., 2006) Superchroma integrating sphere-type color difference meter
Photostability riboflavin tablet UV-VIS absorption measurements – 2009 (Sue-Chu et al., 2009) not mentioned
Photostability amlodipine besylate tablet HPLC linear regression 2024 (Matsushima et al., 2024) Nippon Daneshku SE6000 spectrophotometer
Counterfeits Viagra® Visual examination Multivariate statistical methods⁎⁎⁎ (P < 0.005) 2016 (Al Hagbani et al., 2017) HunterLab ColorQuest XE colorimeter
Quality control Theophylline powder near infrared spectroscopy – 2007 (Siddiqui and Nazzal, 2007) ColorQuest XE colorimeter
Quality control Copper (ΙΙ) histidine injection UV spectroscopy linear regression 2022 (Dwiecki et al., 2022) NH 310 Portable Colorimeter
passive cutaneous anaphylaxis Skin color other operations linear regression (R2 = 0.934) 1996 (AKIYAMA et al. 1996) CM-503I, Minolta spectrophotometer
Comparison of Solution-Phase Transmittance
and Solid-Phase Diffuse Reflectance Methods
2,3-Dimethyl-2H-indazol-6-amine UV–Vis spectrometer method robustness, risk assessment, and quality-by-design 2011 (Zhou et al., 2011) Varian Cary 50 dispersive spectrometer
Color distribution analysis Sucrose pellets Visual examination standard deviation of the color measurement values (P = 0.000004) 2001 (Chan et al., 2001) Minolta CR-241 spectrocolorimeter
Method validation Protein solution European Pharmacopeia reference color solution standard deviation (ΔE2000⁎⁎⁎⁎ to the
mean 0.5)
2017 (Yin et al., 2016) HunterLab UltraScan VIS spectrophotometer
Evaluation of coating process Caffeine tablets Visual examination Average L⁎ ± SD 2023 (James et al., 2023) DataColor ColorReader
Evaluation of mixing process pharmaceutical powders such as lactose Visual examination linear regression (coefficient of variance<0.14 %) 1974 (MATSUDA et al. 1974) color and color difference meter, model ND-K6B
Evaluation of coating process Placebo white tablet CIEDE2000 linear regression (R2 = 0.996) 2019 (Murillo et al., 2019) DigiEye imaging system
Color uniformity Adona®, Anginal®, and Asverin® UV spectroscopy (R2 > 0.98, precision<2.85, accuracy<5) linear regression 2014 (Miyazaki et al., 2014) handy microscope
Drug content and particle size distribution meloxicam tablets UV spectroscopy linear regression (R2 > 0.94) 2020 (Mészáros et al., 2020) Basler acA4112-30uc area scan color camera, equipped with a CMOS sensor
Appearance of tablets Jasco V-670 spectrometer Visual examination linear regression⁎⁎⁎⁎ 1991 (Wirth, 1991) Dr. Lange GmbH
⁎

The model was used to extrapolate shelf life under real-world storage conditions. Q2: the cross-validated R2 is a sensitive indicator of how well the data fit the model, since it gauges how close each measured point is to the predicted value by running the model in the absence of that point (Consonni et al., 2010).

⁎⁎

Which are often used when measurements are taken over time or under different conditions from the same subjects or experimental units.

⁎⁎⁎

Multivariate statistical methods are analytical techniques used to examine and interpret data.

⁎⁎⁎⁎

ΔE2000 is an advanced color difference formula that quantifies perceptual color variation by adjusting for lightness, chroma, hue, and neutral tones, offering high accuracy in visually assessing small color differences (Luo et al., 2001).

7. CIELab accessibility and advantages over other methods

The CIELab color space has gained widespread acceptance in pharmaceutical and biomedical analysis owing to its perceptual uniformity and independence from specific devices (Flaviis, and Riccardo, and Giampiero Sacchetti., 2025). Unlike traditional color models such as RGB or CMYK, CIELab is structured to reflect uniform perceptual differences, whereby equal numerical changes correspond to consistent visual distinctions (Kaur and Kranthi, 2012). This characteristic makes this method suitable for detecting and quantifying subtle chromatic variations in pharmaceutical formulations as well as clinical diagnostics (Flavier et al., 2022). As a device-independent model, CIELab enables standardized color detection across diverse analytical platforms, which is essential for ensuring reproducibility in quality control procedures (Hetrick et al., 2013). Its comprehensive scope encompasses the full spectrum of human-visible colors, offering enhanced sensitivity for identifying minor shifts in drug composition or pathological specimens. The ΔE metric derived from this space provides a reliable quantitative measure of color differences, facilitating assessments of batch uniformity and product degradation over time (Chan et al., 2001). Despite its robust mathematical basis, the interpretation of CIELab's chromatic coordinates (a* and b*) may be challenging for practitioners lacking formal training in color science (Hanbury and Serra, 2002). Moreover, measurement accuracy can be affected by variables such as illumination and sample geometry, necessitating strict standardization protocols (Korifi et al., 2013). The adoption of CIELab color analysis has grown significantly due to integration of advanced colorimetric functions in current spectrophotometers and imaging technologies. Both open-source and registered software now offer robust support for CIELab computations, enabling widespread use across research laboratories and industrial settings. However, achieving optimal results still requires specialized knowledge in color science and access to properly calibrated instruments, which may limit its practical implementation (Hunt et al., 2021).

8. Conclusion

The present review indicates that tristimulus colorimetry is a precise and rapid analytical technique for evaluating various pharmaceutical and material properties. Its applications extend to stability assessment under controlled temperature and humidity conditions, photostability evaluation of light-sensitive pharmaceutical compounds, quality control and characterization of pharmaceutical formulations, and quantitative analysis of appearance and color uniformity. Additionally, this method facilitates the detection of counterfeit substances based on spectral color properties, providing a more objective alternative to traditional visual inspection, which is inherently subjective and prone to observer variability. Various instruments and software utilized by researchers for colorimetric analysis have also been reviewed, further enhancing the understanding and applicability of this technique. By converting color data into quantifiable metrics (e.g. CIELab), tristimulus colorimetry ensures reproducible and objective interpretation for diverse applications in pharmaceutical research as well as quality controlled quality assurance.

It is recommended is to conduct further evaluations on different tristimulus colorimetry techniques to systematically identify the advantages and limitations of each method. Additionally, future research could explore the application of this technique in other materials and assess its effectiveness in determining the stability of various pharmaceutical compounds under different conditions.

CRediT authorship contribution statement

Sahar Marefat: Writing – review & editing, Writing – original draft. Ali Shayanfar: Writing – review & editing, Project administration. Farnaz Monajjemzadeh: Writing – review & editing, Project administration.

Informed consent statement

Not applicable.

Institutional review board statement

This paper is a part of a PhD thesis, (No. 206) submitted by Sahar Marefat to the Faculty of Pharmacy, Tabriz University of Medical Sciences, and was approved by the research ethics committee of the same university (IR.TBZMED.PHARMACY.REC.1402.001).

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Declaration of competing interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Acknowledgments

This study was a part of the Ph.D. thesis of Sahar Marefat (no. 206) submitted to the Faculty of Pharmacy, Tabriz University of Medical Sciences.

Contributor Information

Ali Shayanfar, Email: Shayanfara@tbzmed.ac.ir.

Farnaz Monajjemzadeh, Email: Monaggemzadeh@tbzmed.ac.ir.

Data availability

No data was used for the research described in the article.

References

  1. Akiyama Hiroshi, Teshima Reiko, Akasaka Reiko, Fujimori Kannosuke, Goda Yukihiro, Sawada Jun-Ichi, Toyoda Masatake. Quantitative evaluation of passive cutaneous anaphylaxis (PCA) using a hand-held spectrophotometer. Biol. Pharm. Bull. 1996;19:1112–1114. doi: 10.1248/bpb.19.1112. [DOI] [PubMed] [Google Scholar]
  2. Alvani A., Shayanfar A. Solution stability of pharmaceutical cocrystals. Crystal Growth Design. 2022;22:6323–6337. [Google Scholar]
  3. Baertschi Steven W., Clapham David, Foti Chris, Jansen Patrick J., Kristensen Solveig, Reed Roberta, Templeton Allen C., Tonnesen Hanne Hjorth. Implications of in-use photostability: proposed guidance for photostability testing and labeling to support the administration of photosensitive pharmaceutical products, part 1: drug products administered by injection. J. Pharm. Sci. 2013;102:3888–3899. doi: 10.1002/jps.23717. [DOI] [PubMed] [Google Scholar]
  4. Berberich J., Dee K.-H., Hayauchi Y., Pörtner C. A new method to determine discoloration kinetics of uncoated white tablets occurring during stability testing—an application of instrumental color measurement in the development pharmaceutics. Int. J. Pharmaceut. 2002;234:55–66. doi: 10.1016/s0378-5173(01)00947-4. [DOI] [PubMed] [Google Scholar]
  5. Bisulca Christina, Nascimbene Paul C., Elkin Lisa, Grimaldi David A. Variation in the deterioration of fossil resins and implications for the conservation of fossils in amber. Am. Mus. Novit. 2012;2012:1–19. [Google Scholar]
  6. Bogdansky Frances M. Measurement of surface color and color difference of tablet colorants by tristimulus colorimetry. J. Pharm. Sci. 1975;64:323–328. doi: 10.1002/jps.2600640230. [DOI] [PubMed] [Google Scholar]
  7. Brühl Ludger, Unbehend Günter. Precise color communication by determination of the color of vegetable oils and fats in the CIELAB 1976 (L* a* b*) color space. Eur. J. Lipid Sci. Technol. 2021;123:2000329. [Google Scholar]
  8. Chan L.W., Chan W.Y., Heng P.W.S. An improved method for the measurement of colour uniformity in pellet coating. Int. J. Pharmaceut. 2001;213:63–74. doi: 10.1016/s0378-5173(00)00619-0. [DOI] [PubMed] [Google Scholar]
  9. Consonni Viviana, Ballabio Davide, Todeschini Roberto. Evaluation of model predictive ability by external validation techniques. J. Chemom. 2010;24:194–201. [Google Scholar]
  10. Durmus Dorukalp. CIELAB color space boundaries under theoretical spectra and 99 test color samples. Color. Res. Appl. 2020;45:796–802. [Google Scholar]
  11. Dwiecki Piotr Mariusz, Wróblewska Katarzyna Barbara, Krzywańska Justyna, Parmonik Agata, Muszalska-Kolos Izabela. Critical points in the methodology of preparing Copper (II) histidinate injections and their quality assessment applying color measurement. J. Pharm. Sci. 2022;111:2471–2480. doi: 10.1016/j.xphs.2022.03.015. [DOI] [PubMed] [Google Scholar]
  12. European Pharmacopoeia . 2021. Degree of Coloration of Liquids. [Google Scholar]
  13. Fairman Hugh S., Brill Michael H., Hemmendinger Henry. Color Research & Application: Endorsed by Inter-Society Color Council, the Colour Group (Great Britain) Vol. 22. Canadian Society for Color, Color Science Association of Japan, Dutch Society for the Study of Color, The Swedish Colour Centre Foundation, Colour Society of Australia, Centre Français de la Couleur; 1997. How the CIE 1931 color-matching functions were derived from Wright-Guild data; pp. 11–23. [Google Scholar]
  14. Fan Yajun, Li Jiawei, Guo Yaping, Xie Lianwu, Zhang Gui. Digital image colorimetry on smartphone for chemical analysis: a review. Measurement. 2021;171 [Google Scholar]
  15. Flavier Kristina, McLellan James, Botoy Teslin, Waterman Kenneth C. Accelerated shelf life modeling of appearance change in drug products using ASAP prime®. Pharm. Dev. Technol. 2022;27:740–748. doi: 10.1080/10837450.2022.2112223. [DOI] [PubMed] [Google Scholar]
  16. Flaviis De, Riccardo, and Giampiero Sacchetti. A 50-year theoretical gap on color difference in food science: critical insights and new perspectives. J. Food Sci. 2025;90 doi: 10.1111/1750-3841.70317. [DOI] [PubMed] [Google Scholar]
  17. Ghaderi Faranak, Monajjemzadeh Farnaz. Review of the physicochemical methods applied in the investigation of the maillard reaction in pharmaceutical preparations. J. Drug Deliv. Sci. Technol. 2020;55 [Google Scholar]
  18. Gómez Omar, Perales Esther, Chorro Elísabet, Burgos Francisco J., Viqueira Valentín, Vilaseca Meritxell, Martínez-Verdú Francisco M., Pujol Jaume. Visual and instrumental assessments of color differences in automotive coatings. Color. Res. Appl. 2016;41:384–391. [Google Scholar]
  19. Hagbani Al, Turki Mohammad T., Nutan Michael A., Veronin, and Sami Nazzal. 2016 32nd Southern Biomedical Engineering Conference (SBEC) IEEE Computer Society; 2016. The utility of colorimetry as a quality control tool for the identification of pharmaceutical tablets; p. 95. [Google Scholar]
  20. Hagbani Al, Turki Michael, Veronin A., Nutan Mohammad T., Nazzal Sami. Can the surface color of pharmaceutical tablets be used as a unique product identifier? J. Drug Deliv. Sci. Technol. 2017;37:141–146. [Google Scholar]
  21. Hanbury Allan, Serra Jean. Mathematical morphology in the CIELAB space. Image Analys. Stereol. 2002;21:201–206. [Google Scholar]
  22. Hetrick Evan M., Vannoy Jeffrey, Montgomery Laura L., Pack Brian W. Integrating tristimulus colorimetry into pharmaceutical development for color selection and physical appearance control: a quality-by-design approach. J. Pharm. Sci. 2013;102:2608–2621. doi: 10.1002/jps.23625. [DOI] [PubMed] [Google Scholar]
  23. Hunt Brady, Ruiz Alberto J., Pogue Brian W. Smartphone-based imaging systems for medical applications: a critical review. J. Biomed. Opt. 2021;26 doi: 10.1117/1.JBO.26.4.040902. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Internatinal Council for Harmonisation . ICH Secretariat; Geneva, Switzerland: 2021. “Draft Guideline for the Photostability Testing of New Drug Substances and Products.” in, 46. [Google Scholar]
  25. James William A., Stewart Sarah A., Lawrence Simon M. Comparison of titanium dioxide-containing and titanium dioxide-free film coating systems. Br. J. Pharm. 2023;8:S1–S2. [Google Scholar]
  26. Janga Karthik Yadav, King Tamara, Ji Nan, Sarabu Sandeep, Shadambikar Gauri, Sawant Sagar, Pengchong Xu, Repka Michael A., Narasimha S., Murthy. Photostability issues in pharmaceutical dosage forms and photostabilization. AAPS PharmSciTech. 2018;19:48–59. doi: 10.1208/s12249-017-0869-z. [DOI] [PubMed] [Google Scholar]
  27. Kaur Amanpreet, Kranthi B.V. Comparison between YCbCr color space and CIELab color space for skin color segmentation. Int. J. Appl. Inform. Syst. 2012;3:30–33. [Google Scholar]
  28. Khajir S., Shayanfar A., Monajjemzadeh F., Jouyban A. Crystal engineering of valproic acid and carbamazepine to improve hygroscopicity and dissolution profile. Drug Dev. Ind. Pharm. 2021;47:1674–1679. doi: 10.1080/03639045.2022.2045305. [DOI] [PubMed] [Google Scholar]
  29. Korifi Rabia, Le Dréau Yveline, Antinelli Jean-Francois, Valls Robert, Dupuy Nathalie. CIEL⁎ a⁎ b⁎ color space predictive models for colorimetry devices–Analysisof perfume quality. Talanta. 2013;104:58–66. doi: 10.1016/j.talanta.2012.11.026. [DOI] [PubMed] [Google Scholar]
  30. Li Zhi Jian, Meng Qing Jun. Research on the color-difference evaluation of food package printing matter based on the human visual characteristics. Appl. Mech. Mater. 2014;469:278–281. [Google Scholar]
  31. Luo M. Ronnier, Cui Guihua, Rigg Bryan. Color Research & Application: Endorsed by Inter-Society Color Council, the Colour Group (Great Britain) Vol. 26. Canadian Society for Color, Color Science Association of Japan, Dutch Society for the Study of Color, The Swedish Colour Centre Foundation, Colour Society of Australia, Centre Français de la Couleur; 2001. The development of the CIE 2000 colour-difference formula: CIEDE2000; pp. 340–350. [Google Scholar]
  32. Ly Bao Chau K., Dyer Ethan B., Feig Jessica L., Chien Anna L., Del Bino Sandra. Research techniques made simple: cutaneous colorimetry: a reliable technique for objective skin color measurement. J. Invest. Dermatol. 2020;140(3–12) doi: 10.1016/j.jid.2019.11.003. [DOI] [PubMed] [Google Scholar]
  33. Martino R., Malet-Martino M., Gilard V., Balayssac S. Counterfeit drugs: analytical techniques for their identification. Anal. Bioanal. Chem. 2010;398:77–92. doi: 10.1007/s00216-010-3748-y. [DOI] [PubMed] [Google Scholar]
  34. Matsuda Yoshihisa, Nakanishi Reiko. Stabilization of sulfisomidine tablets by use of film coating containing UV absorber: protection of coloration and photolytic degradation from exaggerated light. J. Pharm. Sci. 1978;67:196–201. doi: 10.1002/jps.2600670218. [DOI] [PubMed] [Google Scholar]
  35. Matsuda Yoshihisa, Shiromoto Sumiko, Takeda Yohko, Hayashi Shinichi, Okada Jutaro. Mixing of pharmaceutical powders examined by use of colored particles: discussions on coloration difference. Chem. Pharm. Bull. 1974;22:2803–2811. [Google Scholar]
  36. Matsushima Yuki, Hattori Masaki, Tanaka Akiko, Furubayashi Tomoyuki, Sakane Toshiyasu. Changes in tablet color due to light irradiation: photodegradation of the coating polymer, hypromellose, by titanium dioxide. AAPS PharmSciTech. 2024;25:26. doi: 10.1208/s12249-024-02732-x. [DOI] [PubMed] [Google Scholar]
  37. Mészáros Lilla Alexandra, Galata Dorián László, Madarász Lajos, Köte Ákos, Csorba Kristóf, Dávid Ádám Zoltán, Domokos András, Szabó Edina, Nagy Brigitta, Marosi György. Digital UV/VIS imaging: a rapid PAT tool for crushing strength, drug content and particle size distribution determination in tablets. Int. J. Pharmaceut. 2020;578 doi: 10.1016/j.ijpharm.2020.119174. [DOI] [PubMed] [Google Scholar]
  38. Miyazaki Yasunori, Miyawaki Kaoru, Uchino Tomonobu, Kagawa Yoshiyuki. Assessment of blending ratio of powdered medicine mixtures by image analysis. Chem. Pharm. Bull. 2014;62:322–327. doi: 10.1248/cpb.c13-00772. [DOI] [PubMed] [Google Scholar]
  39. Mochizuki Koji, Takayama Kozo. Prediction of color changes in acetaminophen solution using the time–temperature superposition principle. Drug Dev. Ind. Pharm. 2016;42:1050–1057. doi: 10.3109/03639045.2015.1107091. [DOI] [PubMed] [Google Scholar]
  40. Monajjemzadeh Farnaz, Hassanzadeh Davoud, Valizadeh Hadi, Siahi-Shadbad Mohammad R., Mojarrad Javid Shahbazi, Robertson Thomas, Roberts Michael S. Assessment of feasibility of Maillard reaction between baclofen and lactose by liquid chromatography and tandem mass spectrometry, application to pre formulation studies. AAPS PharmSciTech. 2009;10:649–659. doi: 10.1208/s12249-009-9248-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  41. Murillo Mac Arturo, Rodríguez-Pulido Francisco J., Heredia Francisco J., Melgosa Manuel, Pacheco Jorge, Vargas Ronny, Montero Ernesto, Gutiérrez Dionisio. Color evolution during a coating process of pharmaceutical tablet cores by random spraying. Color. Res. Appl. 2019;44:160–167. [Google Scholar]
  42. Pack Brian W., Montgomery Laura L., Hetrick Evan M. Modernization of physical appearance and solution color tests using quantitative tristimulus colorimetry: advantages, harmonization, and validation strategies. J. Pharm. Sci. 2015;104:3299–3313. doi: 10.1002/jps.24565. [DOI] [PubMed] [Google Scholar]
  43. Palugan Luca, Spoldi Mattia, Rizzuto Francesco, Guerra Norberto, Uboldi Marco, Cerea Matteo, Moutaharrik Saliha, Melocchi Alice, Gazzaniga Andrea, Zema Lucia. What’s next in the use of opacifiers for cosmetic coatings of solid dosage forms? Insights on current titanium dioxide alternatives. Int. J. Pharmaceut. 2022;616 doi: 10.1016/j.ijpharm.2022.121550. [DOI] [PubMed] [Google Scholar]
  44. Peixoto Patrícia S., Carvalho Pedro H., Machado Ana, Barreiros Luisa, Bordalo Adriano A., Oliveira Hélder P., Segundo Marcela A. Development of a screening method for sulfamethoxazole in environmental water by digital colorimetry using a mobile device. Chemosensors. 2022;10:25. [Google Scholar]
  45. Pharmacopeia, United States . United States Pharmacopeia Convention; Rockville, MD: 2024. United States Pharmacopeia and National Formulary. [Google Scholar]
  46. Pharmacopia, United State . 2024. USP General Chapter <1061>: Color-Instrumental Measurement. [Google Scholar]
  47. Podpora Michal, Korbas Grzegorz Pawel, Kawala-Janik Aleksandra. FedCSIS (Position Papers) Citeseer; 2014. YUV vs RGB-choosing a color space for human-machine interaction; pp. 29–34. [Google Scholar]
  48. Ramadan Alyaa, El-Massik Magda, El-Khordagui Labiba, Daabis Nabila, Hammouda Youssef. Surface treatment: a potential approach for enhancement of solid-state photostability. Int. J. Pharmaceut. 2006;307:141–149. doi: 10.1016/j.ijpharm.2005.09.030. [DOI] [PubMed] [Google Scholar]
  49. Rhee Yun-Seok, Park Chun-Woong, Shin Yoon-Sub, Kam Sung-Hoon, Lee Kyu-Hyun, Park Eun-Seok. Application of instrumental evaluation of color for the pre-formulation and formulation of rabeprazole. Int. J. Pharmaceut. 2008;350:122–129. doi: 10.1016/j.ijpharm.2007.08.039. [DOI] [PubMed] [Google Scholar]
  50. Sakiroff Lara-Malenka, Chennell Philip, Yessaad Mouloud, Pereira Bruno, Bouattour Yassine, Sautou Valérie. Evaluation of color changes during stability studies using spectrophotometric chromaticity measurements versus visual examination. Sci. Rep. 2022;12:8959. doi: 10.1038/s41598-022-13025-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  51. Siddiqui Akhtar, Nazzal Sami. Measurement of surface color as an expedient QC method for the detection of deviations in tablet hardness. Int. J. Pharmaceut. 2007;341:173–180. doi: 10.1016/j.ijpharm.2007.04.006. [DOI] [PubMed] [Google Scholar]
  52. Stark G., Fawcett J.P., Tucker I.G., Weatherall I.L. Instrumental evaluation of color of solid dosage forms during stability testing. Int. J. Pharmaceut. 1996;143:93–100. [Google Scholar]
  53. Sue-Chu M., Kristensen S., Tønnesen H.H. Photoinduced color changes in two different qualities of riboflavin in the solid state and in various tablet formulations photoreactivity of biologically active compounds. XX’, Die Pharmazie Int. J. Pharm. Sci. 2009;64:428–435. [PubMed] [Google Scholar]
  54. Wirth Michael. Instrumental color measurement: a method for judging the appearance of tablets. J. Pharm. Sci. 1991;80:1177–1179. doi: 10.1002/jps.2600801216. [DOI] [PubMed] [Google Scholar]
  55. Yamazaki Noriko, Taya Kumiko, Shimokawa Ken-ichi, Ishii Fumiyoshi. The most appropriate storage method in unit-dose package and correlation between color change and decomposition rate of aspirin tablets. Int. J. Pharmaceut. 2010;396:105–110. doi: 10.1016/j.ijpharm.2010.06.031. [DOI] [PubMed] [Google Scholar]
  56. Yin Jian, Swartz Trevor E., Zhang Jian, Patapoff Thomas W., Chen Bartolo, Marhoul Joseph, Shih Norman, Kabakoff Bruce, Rahimi Kimia. Validation of a spectral method for quantitative measurement of color in protein drug solutions. PDA J. Pharm. Sci. Technol. 2016;70:382–391. doi: 10.5731/pdajpst.2016.006494. [DOI] [PubMed] [Google Scholar]
  57. Zehra Zeenat, Bashir Muhammad Nauman. 2019 4th MEC International Conference on Big Data and Smart City (ICBDSC) IEEE; 2019. Color fastness grading system for textile industry using CIEL* a* b color space; pp. 1–5. [Google Scholar]
  58. Zhou Leon, Vogt Frederick G., Overstreet Patricia-Ann, Dougherty John T., Clawson Jacalyn S., Kord Alireza S. A systematic method development strategy for quantitative color measurement in drug substances, starting materials, and synthetic intermediates. J. Pharm. Innov. 2011;6:217–231. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

No data was used for the research described in the article.