ABSTRACT
Background
Despite recent advances in capillary blood collection technologies for implementation in a patient centric healthcare world, a direct comparison of various technologies has yet to be performed.
Research design and methods
A cross-sectional comparative study was performed by recruiting healthy subjects (n = 41) to participate in a collection comparison study consisting of four upper arm capillary collection technologies, a traditional fingerstick, and a venous collection. User experience (e.g. pain, preference), collection performance (e.g. sample volume and quality, collection time), and clinical accuracy with respect to venipuncture results were compared across technologies. In addition, healing was assessed by reaching out to subjects at two different time points following collection.
Results
User experience and collection performance results varied across the different capillary collection technologies and a single collection technology did not stand out as superior among those investigated. As correlative results were similar across all technologies, selection of a technology may be based on the intended use population and/or other factors.
Conclusions
Although there are still advancements to be made as well as additional studies to further evaluate the analyte equivalency of capillary and venous specimens, these technologies present a promising patient-centric option to provide clinical results and improve patient care.
KEYWORDS: Microsampling, capillary blood, lancet, patient centric sampling, remote sampling
1. Introduction
Venipuncture continues to be the standard in blood specimen collection despite the somewhat invasive nature of the procedure. With the growing interest in patient centric healthcare solutions, there have been significant innovations in capillary blood collection technologies (CCTs) [1–6]. In general, CCTs aim to reduce patient discomfort and collect smaller volumes of blood which consequently improves patient access, compliance, and allows the expansion of testing capabilities beyond traditional clinical settings and directly into the hands of patients. Notably, several upper arm sampling methods have been developed in recent years as incisions in the upper arm produce less pain than traditional fingerstick while still yielding sufficient sample volumes for many clinical tests [7,8].
Although each upper arm CCT is unique mechanically, the underlying principles are generally similar. A subject’s upper arm is first warmed to dilate the underlying vasculature and then cleansed with an alcohol wipe. Either before or after adherence of the CCT, the upper arm is lanced, and a small vacuum is generated around the wound to facilitate blood flow. Blood is collected via gravity into the attached sample collection tube for a pre-determined time or until a specified blood volume is obtained. The CCT is then removed and the resulting wound(s) bandaged appropriately. The sample collection tube is then detached from the CCT and sealed for additional processing and/or shipment prior to testing.
Despite the advances of CCTs and the similarities between them, a direct comparison of existing CCTs has not been performed. This study evaluates five CCTs (traditional fingerstick and four upper arm technologies) for user experience, collection performance, and clinical accuracy with respect to venipuncture results. By examining these technologies through both clinical and patient-centered perspectives, we aim to inform healthcare providers, technology developers, and regulatory bodies in optimizing CCTs that prioritize patient comfort while maintaining reliability and diagnostic accuracy comparable to FDA approved laboratory methods.
2. Materials and methods
2.1. Sample collection
All samples were collected from Labcorp employees by trained professionals within Labcorp facilities; self-collection was not performed for capillary samples. Following appropriate informed consent (WCG IRB Protocol # 520,100,174) and eligibility confirmation, venous samples were collected first using a serum separator tube (367983, BD, Franklin Lakes, NJ) with traditional phlebotomy. This was followed by the serial collection of capillary samples from four different upper arm collection technologies. RedDropTM One (RedDrop Dx, Fort Collins, CO), Tasso+ (Tasso, Inc. Seattle, WA), and Touch Activated Phlebotomy (TAP®) MicroSelect (YourBio Health, Medford, MA) are all FDA Class II 510 (k)-cleared lancets and will be referred to as RD1, T+, and TAP herein (Figure 1). The Labcorp Comfort DrawTM (CD) is a recently described CCT comprising a silicone bulb that enhances blood flow from conventional lancets by applying a vacuum [9]. The device is used by first activating the conventional lancet against the upper arm, then compressing the CD device over the lancet site, creating a vacuum and causing blood to flow into an attached small volume (<1 mL) tube. In this study, CD was used with a Gentleheel lancet (GRI-Alleset GHT10X50). Serum separator BD Microtainers® (BD 365,967) were used with each CCT to collect all capillary samples. All 41 subjects who participated in the study self-reported as healthy. At least 40 paired samples are recommended by the Clinical and Laboratory Standards Institute (CLSI) to assess sample equivalence [10].
Figure 1.

Images of the various upper arm capillary blood collection technologies. RD1 – RedDrop One, T+ – Tasso+, CD – Comfort Draw, TAP – TAP MicroSelect.
Upper arm collections were performed in random order for the four technologies with alternating arms to reduce variable heating effects between sequential collections. To perform the collection, the collection site (approximately 3 to 4 inches below the top of the shoulder) was first heated using an internally developed heat pack comprising an aqueous solution of sodium acetate and a metal activation disc in a vinyl pouch formulated to achieve peak skin temperature of 41–42°C. This consistent heating method was used for each collection technology to eliminate heating as a potential variable in comparing device performance. Following warming for 3 min, the upper arm was then cleansed using an alcohol wipe. After allowing the alcohol to dry for approximately 1 min, the lancets were adhered to the shoulder and activated, leading to a lancing of the skin and generation of a vacuum inside of the lancet (except the CD device, with wound and vacuum generation being separate steps). Devices were left on the arm until blood reached the upper line (representing 600 μL) or until 5 min elapsed. Note that three CCTs, from three separate subjects, exceeded the protocol time limit (up to 6 min allowed) but were included in the study dataset with the exception of CCT flow rate analysis. The upper line was selected for blood collection completion as opposed to the lower line (representing 400 μL) to enable more analytical measurements on each sample. Time of capillary blood collection was recorded as time from device activation (lancing) until the collection was stopped (device removal).
Following collection of capillary samples from the upper arm, a capillary sample was collected via fingerstick (FS) using a high flow rate lancet (BD 366,594). Warming, disinfection, and sample collection (time and volume) were performed in the same manner as upper arm collections; however, the collection process may have been stopped if the finger wound was no longer bleeding. If this did occur, no additional lancing was performed. Manipulation of the lanced finger (i.e., milking) was performed to facilitate blood flow.
2.2. Subject surveys
After each collection, subjects were asked to rate the pain felt for each collection technology on a scale of 0–10 (0 being no pain, 10 being the worst pain imaginable). Following collection of all samples, subjects were asked to preferentially rank order the six collection technologies (1 being the most favored, 6 being the least favored). No guidance was provided about what factors (e.g. speed, time) to consider when providing ranking. Survey results (on a per analysis basis) were excluded if the questionnaire was incomplete (e.g., a subject did not provide a rank for all collection technologies) or not completed properly (e.g., a subject provided the same rank to multiple collection technologies). Provided that subjects were available, follow-up surveys were performed in-person or via e-mail 3–10 days following collection (n = 38) and 17–32 days following collection (n = 39) to assess whether subjects had any ongoing pain or discomfort and if skin markings (e.g., scabbing or scarring) were still present at the site of each wound. Follow-up survey results were included in analysis whether the subject participated in one or both surveys. In addition to providing feedback related to the collection process, subjects were asked to provide demographic information, age, height, and weight.
2.3. Sample processing
All samples were centrifuged within 1 h of collection; venous samples were centrifuged at 1600 g for 10 min and BD Microtainers were centrifuged at 6000 g for 2 min following manufacturer recommendations. Following centrifugation, the combined mass of the whole blood sample and sample tube were measured for each sample. The resulting serum was then transferred to a microcentrifuge tube, and the combined mass was measured. The volume of capillary whole blood and serum was determined through appropriate background subtraction of the sample vessel and assuming densities of 1.060 and 1.018 mg/μL for whole blood and serum, respectively. Both capillary and venous serum were then stored at < −70°C until measurement.
2.4. Sample measurement and analysis
Prior to measurement, samples were thawed at room temperature (20–25°C) for at least 1 h. Capillary samples were transferred to sample cups appropriate for use with small volumes (Roche 05085713001) for measurement on Roche cobas® 8000 autoanalyzers. Sample quality was assessed by measuring each sample’s hemolysis index (Roche Serum Index Gen. 2 05172179190) which is a unitless measure (approximately equivalent to mg/dL) that estimates the amount of hemoglobin present in serum [11]. Measurements of common chemistry analytes were also performed as volume permitted (Supplementary Table S1).
Mean and standard deviation reported for pain scores, subject ranked preferences, blood volumes, serum volumes, collection times, blood flow rates, and hemolysis indices were determined with respect to each collection technology. Results were also visualized using a variety of methods. Correlation of venous measurements to measurements made on capillary samples obtained using each collection technology was assessed using a concordance correlation coefficient (CCC) which incorporates Pearson’s correlation coefficient and bias into a single numeric [12]. Confidence intervals (95%) were also generated for this statistic.
3. Results
3.1. Subject experience – reported pain, preference, and wound healing
Subject-reported pain scores are shown in Figure 2(A). As anticipated, due to the higher density of nerves in fingertips, the average pain score reported for fingerstick sampling (2.53) was higher than the other collection technologies (Table 1) [13,14]. Interestingly, the RD1 device produced a mean pain score (1.78) greater than the other upper arm devices (all less than 1) and venipuncture collections (1.51). Based on subject feedback and observation, this finding may be partly attributed to a psychosomatic response of the snapping sound generated upon device activation and/or the creation of a higher vacuum within this CCT (Supplementary Figure S2).
Figure 2.

A) Pain score reported by each subject on a scale of 0 to 10; mean and 95% confidence intervals are represented by the “X” and error bars, respectively, B) Cumulative subject preference rankings (1 to 6) for each collection technology, C) Percentage of subjects with skin markings present at an initial and secondary follow-up. Ven. – venous, FS – fingerstick, RD1 – RedDrop One, T+ – Tasso+, CD – Comfort Draw, TAP – TAP MicroSelect.
Table 1.
Results summary of the various analyses at time of collection. Ven. – venous, FS – fingerstick, RD1 – RedDrop One, T+ – Tasso+, CD – Comfort draw, TAP – TAP MicroSelect.
| Ven. | FS | RD1 | T+ | CD | TAP | ||
|---|---|---|---|---|---|---|---|
| Pain Reported (0–10 scale) |
Mean | 1.50 | 2.50 | 1.80 | 0.50 | 0.90 | 0.70 |
| SD | 1.40 | 1.90 | 1.80 | 1.00 | 0.90 | 1.10 | |
| n |
37 |
39 |
41 |
41 |
41 |
41 |
|
| Subject Preference (by rank) |
Mean | 3.4 | 5.5 | 3.7 | 2.2 | 3.2 | 3 |
| SD | 1.8 | 0.9 | 1.5 | 1.1 | 1.4 | 1.5 | |
| n |
39 |
39 |
39 |
39 |
39 |
39 |
|
| Blood Volume (μL) |
Mean | – | 473.8 | 601 | 533 | 522.3 | 329.7 |
| SD | – | 171.9 | 157.5 | 167.5 | 178.9 | 217.5 | |
| n |
– |
41 |
41 |
41 |
41 |
41 |
|
| Serum Volume (μL) |
Mean | – | 215.4 | 289 | 244.2 | 236.5 | 154.9 |
| SD | – | 85.3 | 89.3 | 86.6 | 88.1 | 106.1 | |
| n |
– |
41 |
41 |
41 |
41 |
41 |
|
| Collection Time (min) |
Mean | – | 2.4 | 3.8 | 3.6 | 3 | 4.2 |
| SD | – | 1.2 | 1.4 | 1.2 | 1.7 | 1.4 | |
| n |
– |
40 |
41 |
41 |
41 |
41 |
|
| Blood Flow Rate (µL/min) |
Mean | – | 264.8 | 213.2 | 180.8 | 300 | 126 |
| SD | – | 193.3 | 168.8 | 103.7 | 250 | 140.2 | |
| n |
– |
40 |
40 |
41 |
40 |
40 |
|
| Hemolysis Index (~mg/dL) |
Mean | 5.9 | 17.2 | 20.8 | 6.9 | 8.9 | 15.1 |
| SD | 5.3 | 17.6 | 10.8 | 7 | 7.5 | 6.5 | |
| n | 41 | 38 | 41 | 41 | 39 | 30 |
Figure 2(B) shows the subjects’ overall collection preference rankings. Fingerstick was the least preferred method with a mean ranking of 5.49 (out of 6) and ranked last 64.1% of the time (Table 1). The T+ device was the most preferred method, receiving a top ranking (1) by 35.9% of subjects and a mean ranking of 2.18. Although venipuncture, RD1, TAP, and CD collections produced preferences that ranged from 7.7 to 28.2%, the mean rankings for each were similar (3.41, 3.69, 3.21, and 3.03, respectively) with speed of the (entire) collection process, volume collected, and pain all being factors listed as reasons.
Follow-up assessments 3–10 days after collection (mean of 5.1 days) indicated that upper arm collection technologies tended to have a higher prevalence of skin markings present compared to venipuncture and fingerstick, with at least 55.3% of subjects having visible marks present where the upper arm wounds were created (Figure 2(C)). Comparatively, skin markings for venipuncture and fingerstick collections were much less prevalent (≤23.7%). However, ongoing pain/discomfort from fingerstick sampling was reported by 10.3% of subjects who participated in the follow-up. Ongoing pain/discomfort from the upper arm collection technologies produced was ≤5.1% with 0.0% of reporting subjects producing any ongoing pain or discomfort with the TAP device.
At the second follow-up 17–32 days post-collection (mean of 24.4 days), the skin markings created by the upper arm collection technologies had diminished with the prevalence reduced by at least half for the RD1, CD, and TAP devices (Figure 2(C)). The wound created by the T+ device (Supplementary Figure S1) was more likely to produce a skin mark that persisted on subjects, with 54.1% reporting skin markings present at the second follow-up. Skin markings present as a result of venipuncture and fingerstick collection were mostly absent at the second follow-up. No subjects reported any ongoing pain/discomfort for any collection technology at the second follow-up (data not shown).
3.2. Collection performance – collection and sample quality metrics
Blood and serum results obtained for each CCT can be found in Figure 3(A,B), respectively. The RD1 device produced greater volumes of blood and serum than any of the other collection technologies with mean volumes of 601.0 and 289.0 μL, respectively (Table 1). These results were statistically different (Supplementary Figure S3) than the volumes produced by the other CCTs which may be a result of the increased vacuum pressure achieved with the RD1 device versus other CCTs (Supplementary Figure S3). The TAP device produced volumes that were statistically lower than any of the other collection technologies with mean volumes of 329.7 and 154.9 μL, respectively. FS, T+, and CD collections produced blood and serum volumes that either produced a moderate level of significance (p < 0.05) with respect to each other or were not statistically different. No significant statistical difference was observed for blood volume collected based on device order or arm selected (Supplementary Figure S4). Additionally, mean blood volumes collected, across the CCTs, did not appear to be correlated to the subjects’ body mass index (Supplementary Figure S5).
Figure 3.

A) Blood volumes, B) Serum volumes, C) Collection times, and D) Resulting flow rates for each capillary collection technology with corresponding box plots. Mean values for each are represented by an “X.” FS – fingerstick, RD1 – RedDrop One, T+ – Tasso+, CD – Comfort Draw, TAP – TAP MicroSelect.
Investigation of the collection times yielded results that were not statistically different (Supplementary Figure S3) for the RD1, T+, and CD collection technologies with mean collection times of 3.8, 3.6, and 3.0 min, respectively (Table 1). TAP collections took an average of 4.2 min, which was statistically slower than the other collection technologies. Although FS sampling produced the shortest collection time with an average time of 2.4 min, this is likely skewed as several collections were stopped as a result of the subject no longer bleeding (8 of 41 subjects). Distribution of collection time results can be observed in Figure 3(C).
The resulting blood flow rates (Figure 3(D)) were determined by dividing the blood volume obtained (Figure 3(A)) by the observed collection times (Figure 3(C)). The three collections that were allowed to proceed beyond 5 min (Figure 3(C)) were excluded from this analysis. Results were not significantly different (Supplementary Figure S3) for collections performed using FS, RD1, T+, and CD technologies, but the flow rate observed with the TAP collection technology was statistically lower than the other collection technologies.
Investigation of the sample quality (i.e., hemolysis) observed by each collection technology (Figure 4) indicated that all collection technologies produced hemolysis indices that were acceptable for the majority of common clinical assays (Supplementary Table S1) [15]. Even so, all capillary collections produced hemolysis indices that were statistically higher than venous results with the exception of T+ samples (Supplementary Figure S6).
Figure 4.

Hemolysis indices observed for each collection technology with corresponding box plots and mean values represented by an “X.” Ven. – venous, FS – fingerstick, RD1 – RedDrop One, T+ – Tasso+, CD – Comfort Draw, TAP – TAP MicroSelect.
3.3. Clinical accuracy – concordance to venous serum measurements
CCCs generated by comparing measurements made using the CCT serum samples to venous serum samples can be found in Figure 5. All CCTs produced CCCs with overlapping confidence intervals for each analyte investigated. Several analytes (ALP, ALT, AST, BUN, cholesterol, CRP, and total bilirubin) produced substantial CCC values (>0.95) across all CCTs [8]. As observed previously, analytes with a narrow biological variation between subjects (<10%) produced lower correlation results (e.g., albumin) as previously reported [7,16,17]. Capillary serum creatinine produced lower CCC values despite the biological variation between subjects being >10% as it produced a bias with respect to venous serum measurements (approximately −8%, Supplementary Figure S15) which has previously been observed for healthy subjects [7]. Capillary serum carbon dioxide produced the lowest CCC values for all collection technologies as it has a narrow biological variation and also has been shown to have a bias with respect to venous measurements (likely as a result of capillary blood being a mixture of arterial and venous blood) [7]. Additional correlative analysis for each analyte can be found in Supplementary Figures S8 to S24 but overall the results observed were consistent with previous comparisons of results from venous and capillary samples [7,8,18,19]. Overall, 13 of 18 analytes investigated produced agreements consistently greater than 90% with respect to CLIA limits [20].
Figure 5.

CCC results observed for each capillary collection technology with respect to venous measurements. Error bars represent 95% confidence intervals while the value in parentheses represents the between subject biological variation for each analyte. FS – fingerstick, RD1 – RedDrop One, T+ – Tasso+, CD – Comfort Draw, TAP – TAP MicroSelect.
4. Discussion
This study compared four different upper arm blood collection technologies to each other as well as venipuncture and fingerstick sampling across three major areas: subject experience, collection performance, and clinical accuracy. While each upper arm collection technology may have performed slightly better or worse in each of the areas investigated, no single technology emerged as superior across all assessments. Generally, FS collections were the least preferred method likely due to the pain associated with lancing a fingertip. The RD1 lancet produced the highest average volume of blood and serum but also the highest mean hemolysis index among the CCTs albeit acceptable for most analytes. T+ produced the least amount of pain, was the most preferred, and produced the lowest mean hemolysis index among CCTs, but it also caused a much greater prevalence of long-lasting skin markings than any of the other CCTs. The CD technology produced a mean blood flow rate that was higher than the other CCTs but was not the lowest or highest ranked in any other category. The TAP device produced lower blood volumes than the other CCTs but also created wounds that healed more quickly than other upper arm CCTs.
It should be noted that the studies performed herein were standardized with respect to the heat packs used, heating time, and maximum collection time allowed. Results may have been different for each collection technology if specific manufacturer recommended procedures for each device were implemented. Another limitation of the study is that it employed professional sample collection for all specimens and thus does not capture the potential use case of remote (e.g., at home) self-collection of samples for clinical testing purposes. Self-collection, especially by a lay user who may or may not be healthy, will likely result in reduced performance characteristics for all of the CCTs investigated due to the limited experience and training with such devices [21,22]. The pitfalls of a potentially sub-optimal collection (e.g., low specimen volume) may be further compounded by the increased burden of transporting the sample to the laboratory for measurement. Ultimately, as more of the quality control process is passed onto less experienced individuals, there is an increased likelihood of the laboratory receiving samples of insufficient quality (e.g., excessive hemolysis) for testing purposes. Lastly, many users may not feel comfortable with the process of collecting their own sample (regardless of the ease of use) or may dislike having visible skin markings present for a period of time. It should also be noted that because only healthy subjects were recruited for this study, the analytical results did not cover the full measurement range for all of the analytes examined. A larger cohort that also includes unhealthy subjects might reveal subtler differences between collection methods that are not apparent in the current study. Nonetheless, the results of the study indicate that each of the upper arm collection technologies investigated has clinical utility for many of the analytical measurements tested here. Device selection for specific applications may depend on a variety of factors. Additionally, upper arm CCTs are relatively new, and these technologies are frequently being revised to improve performance, reliability, and comfort.
5. Conclusions
The data presented here provide a multi-faceted comparison for a selection of traditional and emerging blood collection technologies. Although this study represents a step forward with respect to the understanding and acceptance of patient-centric sampling solutions, further studies are still needed including more data assessing the analytical and clinical agreement between capillary and venous samples as well as investigation of the various collection technologies with respect to self-collection of samples in remote environments (e.g., at home). Further investigation into remote solutions for pre-analytical processing (e.g., sample centrifugation and shipping) will also be needed to ensure successful sample acquisition and measurement with remote collection.
Supplementary Material
Acknowledgments
The authors would like to thank RedDrop Dx and YourBio Health, Inc. for providing lancets and other materials that enabled the completion of these studies.
Portions of this work were previously presented at the Association for Diagnostics & Laboratory Medicine Annual Meeting in July of 2025.
Funding Statement
This paper was not funded.
Article highlights
Five different capillary collection technologies (one fingerstick and four upper arm) were investigated with respect to user experience, collection performance, and clinical accuracy as compared to traditional venous measurements.
Devices were evaluated on perceived pain, blood volume collected, serum volume collected, collection time, blood flow rate, hemolysis index, and post-collection skin markings.
Although fingerstick collections were the least preferred capillary collection technology, likely due to higher perceived pain, the upper arm collection technologies investigated each had their own advantages and disadvantages.
The various capillary collection technologies tested produced similar concordance with venous results with respect to the analytes tested.
The upper arm collection devices evaluated in this study may provide a patient-preferred alternative to traditional venipuncture or fingerstick collection for many routine blood tests.
Author contributions
Bradley B. Collier, Ulrich Y. Schaff, Greg J. Sommer, and Russell P. Grant conceptualized the study. Whitney C. Brandon, Matthew R. Chappell, Gabriella Iacovetti, and Nicolas Tokunaga were responsible for project administration including sample collections and/or data acquisition. Whitney C. Brandon was responsible for data curation. Bradley B. Collier was responsible formal analysis of the data and writing of the original draft. Ulrich Y. Schaff, Greg J. Sommer, and Russell P. Grant reviewed and edited the manuscript.
Disclosure statement
Bradley B. Collier, Matthew R. Chappell, Gabriella Iacovetti, Ulrich Y. Schaff, Greg J. Sommer, and Russell P. Grant are employees of Labcorp and participate in the employee stock purchase plan. Nicolas Tokunaga is an employee of Labcorp. Whitney C. Brandon is a former employee of Labcorp.
The authors have no other relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript apart from those disclosed.
All logos, trademarks, and brand names are the property of their respective owners.
No writing assistance was utilized in the production of this manuscript.
Reviewers Disclosure
Peer reviewers on this manuscript have no relevant financial or other relationships to disclose.
Ethical declaration
All samples were collected from Labcorp employees by trained professionals within Labcorp facilities; self-collection was not performed for capillary samples; following appropriate informed consent (WCG IRB Protocol # 520,100,174) and eligibility confirmation.
Informed consent was obtained from each subject that participated in the study and information that can potentially identify each subject is not available within the manuscript or the supplemental material.
Supplementary material
Supplemental data for this article can be accessed online at https://doi.org/10.1080/17576180.2025.2580284
References
Papers of special note have been highlighted as either of interest (•) or of considerable interest (••) to readers.
- 1.Pourafshar S, Parikh M, Abdallah B, et al. An assessment of individual preference for a novel capillary blood collection system. Patient Prefer Adherence. 2024;18:531–541. doi: 10.2147/PPA.S437969 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Thangavelu MU, Wouters B, Kindt A, et al. Blood microsampling technologies: innovations and applications in 2022. Anal Sci Adv. 2023;4(5–6):154–180. doi: 10.1002/ansa.202300011 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Poland DCW, Cobbaert CM.. Blood self-sampling devices: innovation, interpretation and implementation in total lab automation. Clin Chem Lab Med. 2025;63(1):3–13. doi: 10.1515/cclm-2024-0508 [DOI] [PubMed] [Google Scholar]; • A review of many of the existing capillary blood collection technologies available.
- 4.Lei BUW, Prow TW.. A review of microsampling techniques and their social impact. Biomed Microdevices. 2019;21(4). doi: 10.1007/s10544-019-0412-y [DOI] [PMC free article] [PubMed] [Google Scholar]; • Another review of existing capillary blood collection technologies as well as related technologies as well as the social impacts of these technologies.
- 5.Bossi E, Limo E, Pagani L, et al. Revolutionizing blood collection: innovations, applications, and the potential of microsampling technologies for monitoring metabolites and lipids. Metabolites. 2024;14(1). doi: 10.3390/metabo14010046 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Maass KF, Barfield MD, Ito M, et al. Leveraging patient-centric sampling for clinical drug development and decentralized clinical trials: promise to reality. Clin Transl Sci. 2022;15(12):2785–2795. doi: 10.1111/cts.13411 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Collier BB, Brandon WC, Chappell MR, et al. Maximizing microsampling: measurement of comprehensive metabolic and lipid panels using a novel capillary blood collection device. J Appl Lab Med. 2023;8(6):1115–1126. doi: 10.1093/jalm/jfad066 [DOI] [PubMed] [Google Scholar]; • Analysis of capillary sample results with respect to venous measurements.
- 8.Wickremsinhe E, Fantana A, Berthier E, et al. Standard venipuncture vs a capillary blood collection device for the prospective determination of abnormal liver chemistry. J Appl Lab Med. 2023;8(3):535–550. doi: 10.1093/jalm/jfac127 [DOI] [PubMed] [Google Scholar]; • Additional clinical analysis of capillary sample results as compared to venous measurements.
- 9.Schaff UY, Collier BB, Iacovetti G, et al. Evaluation of a simplified upper arm device for vacuum-assisted collection of capillary blood specimens. Diagnostics. 2025;15(15). doi: 10.3390/diagnostics15151935 [DOI] [PMC free article] [PubMed] [Google Scholar]; •• A recently developed capillary blood collection technology that is uses an off the shelf lancet.
- 10.CLSI . Assessment of equivalence or suitability of specimen types for medical laboratory measurement procedures. In: CLSI guideline EP35. 2nd ed. Wayne (PA): Clinical and Laboratory Standards Institute; 2025. [Google Scholar]
- 11.Roche . Si2: serum index gen. 2 (05172179190). 2024–02 (V7.0 English).
- 12.Lin L-K. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45(1):255–268. doi: 10.2307/2532051 [DOI] [PubMed] [Google Scholar]
- 13.Blicharz TM, Gong P, Bunner BM, et al. Microneedle-based device for the one-step painless collection of capillary blood samples. Nat Biomed Eng. 2018;2(3):151–157. doi: 10.1038/s41551-018-0194-1 [DOI] [PubMed] [Google Scholar]
- 14.Hoffman MSF, McKeage JW, Xu J, et al. Minimally invasive capillary blood sampling methods. Expert Rev Med Devices. 2023;20(1):5–16. doi: 10.1080/17434440.2023.2170783 [DOI] [PubMed] [Google Scholar]
- 15.Knezevic CE, Ness MA, Tsang PHT, et al. Establishing hemolysis and lipemia acceptance thresholds for clinical chemistry tests. Clin Chim Acta. 2020;510:459–465. doi: 10.1016/j.cca.2020.08.004 [DOI] [PubMed] [Google Scholar]
- 16.Aarsand AK, Fernandez-Call P, Webster C, et al. The EFLM biological variation database [Internet]. 2021. [cited 2025 Jul 21]. Available from: https://biologicalvariation.eu
- 17.Blumenfeld TA, Hertelendy WG, Ford SH. Simultaneously obtained skin-puncture serum, skin-puncture plasma, and venous serum compared, and effects of warming the skin before puncture. Clin Chem. 1977;23(9):1705–1710. doi: 10.1093/clinchem/23.9.1705 [DOI] [PubMed] [Google Scholar]
- 18.Maroto-García J, Deza S, Fuentes-Bullejos P, et al. Analysis of common biomarkers in capillary blood in routine clinical laboratory. Preanalytical and analytical comparison with venous blood. Diagnosis. 2023;10(3):281–297. doi: 10.1515/dx-2022-0126 [DOI] [PubMed] [Google Scholar]
- 19.Lee D, Rapp CG, Loureiro J, et al. Decentralized clinical trial design using blood microsampling technology for serum bioanalysis. Bioanalysis. 2023;15(21):1287–1303. doi: 10.4155/bio-2023-0136 [DOI] [PubMed] [Google Scholar]
- 20.Centers for Medicare & Medicaid Services (CMS) H, Centers for Disease Control and Prevention (CDC) H . Clinical laboratory improvement amendments of 1988 (CLIA) proficiency testing regulations related to analytes and acceptable performance. Fed Regist. 2022;87(131):41194–41242. [Google Scholar]
- 21.Boffel L, Van Mensel A, Pauwels J, et al. Self-sampling by adolescents at home: assessment of the feasibility to successfully collect blood microsamples by inexperienced individuals. AAPS J. 2024;26(4):75. doi: 10.1208/s12248-024-00947-1 [DOI] [PubMed] [Google Scholar]
- 22.Plebani M. Blood self-sampling: friend or foe? Clin Chem Lab Med. 2025;63(1):1–2. doi: 10.1515/cclm-2024-1162 [DOI] [PubMed] [Google Scholar]; •• An editorial investigating the potential benefits and harms of implementing direct to consumer testing.