Graphical abstract

Keywords: Head and neck cancer, Radiotherapy planning, Segmentation, computed tomography (CT), Photon-counting
Highlights
-
•
PCD-CT provides enhanced image quality, reduced noise and artifacts, and lower radiation dose compared with conventional CT.
-
•
This study is the first to investigate the use of PCD-CT for delineation purposes.
-
•
Accuracy of organ and target segmentation between PCD-CT and energy-integrating detector computed tomography (EID-CT) was compared.
-
•
The study established the feasibility of PCD-CT for radiotherapy segmentation.
-
•
PCD-CT-based delineation is at least comparable to conventional EID-CT.
Abstract
Background and purpose
Photon-counting detector computed tomography (PCD-CT) offers advancements over conventional energy-integrating detector computed tomography (EID-CT). This study is the first to investigate the use of PCD-CT for delineation purposes.
Materials and methods
Twelve patients with head-and-neck cancer were prospectively included. Delineation of clinical target volume (CTV) and organs at risk (OARs) was performed by three physicians using EID-CT (120 kVp) and PCD-CT at virtual monoenergetic levels of 140 keV and 50 keV. Segmentation accuracy was assessed and contouring-confidence was rated.
Results
The ratio of the mean computed tomography dose index (CTDI) for EID-CT/PCD-CT for all patients was 2.72. The study revealed high inter-observer agreement for the mandible, oral cavity, parotid glands, and submandibular glands, with DSCs exceeding 0.8 across both EID-CT and PCD-CT. No differences in delineation agreement were observed, while a significant improvement was found for the lips (p = 0.001) and oral cavity (p = 0.003) when combining 140 keV and 50 keV images in PCD-CT compared to EID-CT. PCD-CT outperformed EID-CT in terms of contouring confidence.
Conclusion
The study established the feasibility of PCD-CT for radiotherapy planning. PCD-CT-based delineation provides results comparable to EID-CT, with the added advantages of reduced radiation dose, improved image quality and higher delineation confidence. Significant improvements in delineating specific structures, such as the lips and oral cavity, were observed when combining multiple energy levels in PCD-CT.
1. Introduction
Photon-counting detector computed tomography (PCD-CT) represents a substantial advancement in computed tomography technology [1,2]. PCD-CT offers several advantages over energy-integrating detector computed tomography (EID-CT), including reduced electronic noise, enhanced contrast-to-noise ratio, greater radiation dose efficiency, superior spatial resolution, and the ability to perform simultaneous multi-energy acquisition at a single x-ray tube potential [[3], [4], [5], [6], [7], [8], [9], [10], [11], [12], [13]]. These features make PCD-CT a promising technology, offering superior imaging quality and enhanced diagnostic information [11,12]. Additionally, PCD-CT may require lower doses of contrast media compared to EID-CT [[3], [4], [5], [6], [7], [8], [9], [10], [11], [12], [13]]. The simultaneous multi-energy data acquisition allows for advanced processing techniques, such as the creation of virtual non-contrast (VNC) reconstructions without additional exposure [[13], [14], [15], [16], [17], [18]].
For radiotherapy planning the advantages of PCD-CT could be of utmost relevance. EID-CT scans, commonly used in radiotherapy planning, are limited by electronic noise and lower spatial resolution, which can impact the precision of tumor and organs at risk (OARs) delineation. PCD-CT addresses these limitations, potentially leading to more precise delineation. Moreover, studies have highlighted PCD-CT's ability to improve the characterization of artificial materials, such as dental fillings, implants, and osteosynthesis materials, common in patients undergoing radiotherapy.
Despite these promising features and pioneering studies on using PCD-CT for radiotherapy purposes [[19], [20], [21], [22]] there is currently only scarce evidence specifically addressing the use of clinically acquired PCD-CT for organ and target segmentation in radiation oncology, even less for the challenging scenario of head and neck cancer. The studies mentioned above could demonstrate improved organ segmentation in abdominal phantoms [19] and at least equal accuracy of the quantitative information to this obtained by EID-CT [20]. Several studies also reported on the feasibility of implementing Dual Energy CT (DECT) for radiotherapy segmentation and planning, a technique also capable of reconstructing virtual monoenergetic images (VMI) [23,24].
Primary goal of this study is to demonstrate the feasibility of implementing PCD-CT for delineation purposes in challenging, real-life head and neck cancer cases, and to assess whether it can achieve at least similar agreement among clinicians when contouring OARs and targets compared to EID-CT.
2. Materials and methods
2.1. Patient cohort
The patient cohort consisted of 12 patients with head-and-neck cancer who had target volumes near radio-opaque oral implants. The latest included dental-based implants, extensive metal fillings (including amalgam) or titan plate reconstructions of the jaw after tumor surgery being within the CTV or within 2 cm of the CTV (Supplementary Fig. S6). Inclusion criteria were: patients referred for routine radiotherapy, completion of the PCD-CT examination, age over 18 years, and consent to the use of personal health data via the signed hospital‘s General Consent form. Exclusion criteria included pregnant or breastfeeding patients, those without radio-opaque artificial oral implants, and patients with incomplete CT data.
2.2. Patient characteristics
In this study, nine patients were male, three were female, with an average age of 63.9 years. Eleven tumors were squamous cell carcinomas (SSC); one was undifferentiated. The primary tumor location was predominantly in the oral cavity, with one case having a submental nodal metastasis. The staging distribution among the patients varied: one was classified as Stage III, 11 as IVa-IVc. Regarding the intent of radiotherapy, nine patients received adjuvant, two definitive, and one received palliative radiotherapy. A summary of patient and tumor characteristics is provided in Table 1.
Table 1.
Patient and tumor characteristics.
| Patients characteristics | Number |
|---|---|
| Sex | |
|
9 (75 %) |
|
3 (25 %) |
| Age (years) | |
| mean | 63.9 |
| range | 45–90 |
| Tumor Histology | |
|
11 (91.67 %) |
|
1 (8.33 %) |
| Primary tumor location | |
|
11 (91.67 %) |
|
1 (8.33 %) |
| Stage | |
|
1 (8.33 %) |
|
7 (58.34 %) |
|
3 (25 %) |
|
1 (8.33 %) |
| Intent of Radiotherapy | |
|
2 (16.67 %) |
|
9 (75 %) |
|
1 (8.33 %) |
| Systemic treatment | |
|
6 (50 %) |
|
6 (50 %) |
| Type of systemic treatment | |
|
5 (42 %) |
|
1 (8 %) |
| Total | 12 (100 %) |
All data was collected prospectively following approval by the institutional review board (number: 2022–00676). Only data collected during the standard clinical diagnostic process was included, ensuring no additional burden on the patients, as the EID-CT is needed for planning anyway and a timely diagnostic imaging (in this case PCD-CT) is always used in our department. No further study-specific procedures were performed.
The workflow used in the image acquisition and data analysis is shown in Fig. 1.
Fig. 1.
Workflow of the analysis: PCD-CT and EID-CT were acquired within one week. Two PCD-CT reconstructions (50 and 140 keV) were selected. Three certified radiation oncologists performed delineations on the EID-CT and both PCD-CT reconstructions. Pairwise Dice scores, Hausdorff distance, and mean distance to agreement were calculated for the CTV and OARs to assess and compare the inter-expert agreement across PCD-CT 50 keV, PCD-CT 140 keV, and EID-CT.
Appointment A” was the day of the Photon-counting CT and “Appointment B” the one of the “conventional” planning CT. The two appointments had to be performed within one week from each other. All experts contoured all three images (PCD-CT 140 and 50 keV and EID-CT) for every patient on separate days. They were required to complete segmentation of one reconstruction type for all patients before proceeding to the next image type.
2.3. Imaging
All PCD-CT scans were performed at a Siemens NAEOTOM Alpha scanner (Siemens Healthineers) approved for clinical use in radiology. The patients were scanned using our institute standard radiology protocols, with 120 kVp, 500 ms exposure time, 200 mA tube current and on average 100 mAs exposure.
In PCD-CT, different reconstruction options are available, exploiting the multi-energy spectral information detected by the system. For this study, we selected two energy levels: 50 and 140 keV virtual mono-energetic images (VMI). Low-energy VMI (50 keV) offers good soft tissue and iodine contrast but can include much stronger artifacts for highly attenuating materials. On the other hand, high-energy VMI (110–140 keV) shows reduced soft tissue and iodine contrast, but in general less artifacts, also dependent on the specific material [[25], [26], [27]].
Different reconstruction algorithms among those available from Siemens Healthineers were evaluated. The Br40 kernel was selected for this study. The acronym is for 'Body regular', identifying a reconstruction kernel with slight edge enhancement, which is appropriate for delineation purposes. The '40' refers to the resolution index of the reconstruction, representing an intermediate resolution (“soft” image impression) that produced images optimal for soft tissue structures, as opposed to sharper kernels used for bones. As reference, we used the standard protocol for planning EID-CT (performed at a Siemens SOMATOM Definition AS) at our institute: a 120 kVp image (exposure 375 mAs, current 450 mA), performed on a Siemens Healthineers SOMATOM Definition AS scanner.
The resolution of the EID-CT, PCD-CT 140 keV, and PCD-CT 50 keV was 1.27–1.52 mm, 0.39–0.66, and 0.47–0.60, respectively. The slice thickness for EID-CT was always 2 mm, whereas for PCD-CT it ranged from 1.5 to 3 mm. Example images are shown in Fig. 2.
Fig. 2.
Head and neck images acquired with different CT techniques. From left to right: EID-CT, PCD-CT 50 keV, PCD-CT 140 keV.
2.4. Target and OARs delineation
As part of routine practice, target volumes, including Clinical Target Volume (CTV) and OARs, were delineated according to the guidelines [28], [29] on both the EID-CT and the PCD-CT images. Three experienced physicians performed these delineations using Varian Eclipse software.
The contoured OARs included the lips, mandible, oral cavity, pharyngeal constrictors, buccal mucosa, parotid glands and submandibular glands. Manual contours at 140 keV were created using only this energy level, while those at 50 keV were done by reviewing both 140 keV and 50 keV.
2.5. Contouring confidence rating
After performing contouring on both EID-CT and PCD-CT scans, all specialists evaluated the subjective contouring confidence of the following structures: 1) CTV, 2) mandible, 3) lips, and 4) other organs at risk, using a confidence scale ranging from 0 to 10. The mean values and range for each rating are reported.
2.6. Statistical analysis
For each region of interest (ROI), three segmentation metrics were calculated: Dice similarity coefficient (DSC), Hausdorff distance (HD), and mean distance to agreement (MDA). Because each patient was contoured using three different imaging modalities (EID-CT, PCD-CT at 140 keV, and PCD-CT 50 keV) and each case was contoured by three expert pairs, we used mixed-effects models with imaging modality as a fixed effect and patient and expert pair as random effects.
Since DSC values range from 0 to 1, we used a beta generalized linear mixed model with a logit link. Models were fit with the glmmTMB in R. Because HD and MDA are strictly positive and right-skewed, we used linear mixed-effects model with log-transformed outcomes, fit with the lmer package in R.
For each ROI and modality, estimated marginal means (EMM) were calculated. To control the family-wise error (FWER) within each ROI, p-values were adjusted using the Holm method.
3. Results
3.1. Dose exposure
We calculated the mean computed tomography dose index (CTDI) for EID-CT and PCD-CT across all patients. The ratio of mean CTDI (EID-CT/PCD-CT) was 2.72, indicating an overall lower radiation dose with PCD-CT for the majority of patients (see supplementary figures S5a-S5c).
3.2. Agreement among experts
Inter-expert agreement in terms of DSC was high for the mandible, oral cavity, parotid glands, and submandibular glands, with values exceeding 0.8 across all imaging modalities (Fig. 3; Supplementary Table S1). In contrast, the CTV, buccal mucosa, lips, and pharyngeal constrictor showed lower agreement, with DSC values averaging around 0.5. MDA and HD were highest for the CTV (approximately 6 mm and 25 mm, respectively; Fig. 3B-C; Supplementary Table S2-3). For most other ROIs, HD values were similar, ranging from 10 mm to 15 mm, while the submandibular glands had lower HD values ca. 6 mm). MDA was smallest for the submandibular glands (ca. 1 mm) and the mandible (0.7 – 0.8 mm).
Fig. 3.
Boxplots showing inter-expert delineation agreement across imaging modalities: EID-CT, PCD-CT 140 keV, and combined PCD-CT 50 keV + 140 keV. Agreement was quantified using (A) Dice Score, (B) Hausdorff Distance, and (C) mean distance to agreement.
Among all ROIs, the CTV showed the greatest variability in DSC, HD, and MDA. No significant differences in segmentation agreement were observed between the imaging modalities overall. The most significant modality-specific differences were found for the lips, mandible, and oral cavity. For example, segmentation agreement for the lips was higher in PCD-CT at 50 keV than in EID-CT in terms of DSC (p = 0.001) and MDA (p = 0.047). Similarly, oral cavity segmentation showed higher agreement in PCD-CT at 50 keV and PCD-CT at 140 keV compared with EID-CT in terms of DSC (p = 0.003 and p = 0.016), HD (p < 0.001), and MDA (p = 0.005 and p = 0.022). Complete information on segmentation metric values and p-values is available in Supplementary Tables S1-3.
3.3. Contouring confidence rating
The mean confidence ratings across all observers showed a trend for higher confidence for all contoured volumes when using PCD-CT, especially for the lips and the CTV. Notably, the subjective rating was higher for all segmented volumes. Specifically, the mean scores (range) for EID-CT were: 1) CTV 7.3 (range 5–9), 2) mandible 9.2 (range 6–10), 3) lips 6.9 (range 4–9), and 4) other organs at risk 8.1 (6–10). For PCD-CT, the mean scores (range) were: 1) CTV 8.2 (range 6–10), 2) mandible 9.7 (range 8–10), 3) lips 8.1 (range 5–10), and 4) other organs at risk 8.8 (6–10). Detailed results are provided in Supplementary Table S4.
4. Discussion
In this study, we prospectively evaluated the feasibility of using photon-counting CT (PCD-CT) for delineation in radiation oncology, focusing on the challenging clinical scenario of head-and-neck cancer, where several OARs and the CTV are located near oral implants that cause imaging artifacts.
PCD-CT offers potential advantages in reducing radiation dose and improving ROI delineation in head and neck cancer. Our study showed that the ratio of the mean CTDI between EID-CT and PCD-CT was 2.72, confirming the overall lower dose with PCD-CT for most patients. Similarly, Jungblut et al. [30] reported that PCD-CT reduced the radiation dose by 66 % compared to energy-integrating detector CT (EID-CT), while maintaining diagnostic image quality in interstitial lung disease, with superior image sharpness and reduced noise.
Our findings indicate that most ROIs did not show a statistically significant differences in inter-expert delineation agreement between EID-CT and PCD-CT 140 keV. However, combining 140 keV and 50 keV reconstructions in PCD-CT led to significant improvements in the delineation of certain OARs, such as the oral cavity and lips. Interestingly, some volumes, such as the buccal mucosa and lips, showed lower agreement in terms of DSC and MDA with PCD-CT at 50 keV compared to EID-CT, while Hausdorff Distance (HD) values remained similar across modalities. This result is not surprising, considering the proximity of these structures to metal artifacts and the superior artifact reduction achieved with PCD-CT at 140 keV [31]. Moreover, for some organs, such as the mandible and the submandibular glands, DSC values were significantly higher in EID-CT than with PCD-CT. These findings suggest that while PCD-CT, especially with combined energies, can enhance inter-expert agreement for specific OARs, conventional EID-CT may still provide higher agreement in some cases. Nevertheless, it is unclear if the small improvements in agreement observed for PCD-CT are clinically relevant in the radiotherapy workflow.
In the future, a PCD-CT-only workflow could enable a) acquisition of a single contrast-enhanced examination without the need for additional diagnostic imaging, b) virtual contrast subtraction to generate images more suitable for planning, c) lower radiation exposure, d) improved material characterization with more accurate dose calculation, and e) reduced artifacts. Future studies could further explore potential advantages in target segmentation by leveraging more reconstructions. Many of these features are also achievable with dual-energy CT (DECT), however at a cost of higher imaging dose; moreover, recent investigations are showing better performance of PCD-CT over DECT for certain applications, although further validation is required [32].
This study has several limitations, including the relatively small number of patients and physicians involved in the contouring process, as well as the fact that the two CT scans were performed at different appointments, which could have influenced patient positioning and anatomy. Furthermore, this study did not evaluate the ultra-high-resolution mode of PCD-CT, which provides a higher spatial resolution than the scan modes tested here [33]. Additionally, only two of many possible PCD-CT energies and reconstructions were investigated. Despite these limitations, to our knowledge this is the first prospective study to demonstrate PCD-CT-based OAR and CTV delineation results, possibly paving the way for future clinical use of PCD-CT in radiation therapy.
In conclusion this study demonstrates, for the first time in clinically obtained, real-world images, that PCD-CT provides delineation agreement at least comparable to conventional EID-CT in patients with head-and-neck cancer, even in challenging scenarios involving in-field metal artifacts. The potential advantages of PCD-CT, including reduced radiation dose and enhanced diagnostic image quality, underscore its promise in clinical applications.
CRediT authorship contribution statement
Niccolò Bertini: Conceptualization, Project administration, Data curation, Software, Formal analysis, Investigation, Methodology, Writing – original draft, Writing – review & editing. Hubert S. Gabryś: Conceptualization, Project administration, Data curation, Software, Formal analysis, Investigation, Methodology, Writing – original draft, Writing – review & editing. Hatem Alkadhi: Software, Resources, Methodology, Writing – review & editing. Lotte Wilke: Supervision, Data curation, Software, Methodology, Writing – review & editing. Patrick Wohlfahrt: Methodology, Writing – review & editing. Serena Psoroulas: Data curation, Writing – review & editing. Eugenia Vlaskou: Supervision, Data curation, Software, Methodology, Writing – review & editing. Laura Motisi: Supervision, Data curation, Software, Methodology, Writing – review & editing. Matthias Guckenberger: . Stephanie Tanadini-Lang: Supervision, Data curation, Software, Methodology, Writing – review & editing. Panagiotis Balermpas: Conceptualization, Project administration, Data curation, Software, Methodology, Writing – original draft, Writing – review & editing.
Funding
None.
Declaration of competing interest
The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: The department of radiation oncology, Zurich University Hospital, has a research agreement on photon counting CT with Siemens Healthineers. Patrick Wohlfahrt is an employee within the research and development team of Siemens Healthineers. The Department of Diagnostic and Interventional Radiology has research agreements with Bayer, Canon, Guerbet and Siemens. H.A. received speaker fee’s from Siemens. Niccolo’ Bertini, declares no conflicts of interest.
Acknowledgement
We thank all the staff of the Radiation Oncology Department of the University Hospital of Zurich for their support and especially Mohammad Varasteh from Radiation protection at the USZ, who helped with retrieving all the information on CTDIvol.
Footnotes
Supplementary data to this article can be found online at https://doi.org/10.1016/j.phro.2025.100864.
Appendix A. Supplementary data
The following are the Supplementary data to this article:
References
- 1.Greffier J., Viry A., Robert A., Khorsi M., Si-Mohamed S. Photon-counting CT systems: A technical review of current clinical possibilities. Diagn Interv Imaging. 2024 doi: 10.1016/j.diii.2024.09.002. Sep 20:S2211-5684(24)00195-5. Epub ahead of print. PMID: 39304365. [DOI] [PubMed] [Google Scholar]
- 2.Dane B., Froemming A., Schwartz F.R., Toshav A., Ramirez-Giraldo J.C., Ananthakrishnan L. Photon counting CT clinical adoption, integration, and workflow. Abdom Radiol (NY) 2024;49(12):4600–4609. doi: 10.1007/s00261-024-04503-5. Epub 2024 Jul 25 PMID: 39052057. [DOI] [PubMed] [Google Scholar]
- 3.Leng S., Bruesewitz M., Tao S., Rajendran K., Halaweish A.F., Campeau N.G., Fletcher J.G., McCollough C.H. Photon-counting detector CT: system design and clinical applications of an emerging technology. Radiographics. 2019;39(3):729–743. doi: 10.1148/rg.2019180115. PMID: 31059394; PMCID: PMC6542627. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Zanon C., Cademartiri F., Toniolo A., Bini C., Clemente A., Colacchio E.C., et al. Advantages of photon-counting detector CT in aortic imaging. Tomography. 2023;10(1):1–13. doi: 10.3390/tomography10010001. PMID: 38276249; PMCID: PMC10821336. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Khanungwanitkul K., Sliwicka O., Schwartz F.R. Abdominal applications of photon-counting CT. Br J Radiol. 2024 doi: 10.1093/bjr/tqae206. Epub ahead of print. PMID: 39392799. [DOI] [PubMed] [Google Scholar]
- 6.Cosset B., Sigovan M., Boccalini S., Farhat F., Douek P., Boussel L., et al. Bicolor K-edge spectral photon-counting CT imaging for the diagnosis of thoracic endoleaks: a dynamic phantom study. Diagn Interv Imaging. 2023;104(5):235–242. doi: 10.1016/j.diii.2022.12.003. Epub 2023 Jan 15 PMID: 36646587. [DOI] [PubMed] [Google Scholar]
- 7.Emrich T., O'Doherty J., Schoepf U.J., Suranyi P., Aquino G., Kloeckner R., et al. Reduced iodinated contrast media administration in coronary CT angiography on a clinical photon-counting detector CT system: a phantom study using a dynamic circulation model. Invest Radiol. 2023;58(2):148–155. doi: 10.1097/RLI.0000000000000911. Epub 2022 Sep 13 PMID: 36165932. [DOI] [PubMed] [Google Scholar]
- 8.Giantsoudi D., De Man B., Verburg J., Trofimov A., Jin Y., Wang G., et al. Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction. Phys Med Biol. 2017;62(8):R49–R80. doi: 10.1088/1361-6560/aa5293. Epub 2017 Mar 21 PMID: 28323641. [DOI] [PubMed] [Google Scholar]
- 9.Stieb S., McDonald B., Gronberg M., Engeseth G.M., He R., Fuller C.D. Imaging for target delineation and treatment planning in radiation oncology: current and emerging techniques. Hematol Oncol Clin North Am. 2019;33(6):963–975. doi: 10.1016/j.hoc.2019.08.008. Epub 2019 Sep 17. PMID: 31668214; PMCID: PMC7217094. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Kooiman J., Pasha S.M., Zondag W., Sijpkens Y.W., van der Molen A.J., Huisman M.V., et al. Meta-analysis: serum creatinine changes following contrast enhanced CT imaging. Eur J Radiol. 2012;81(10):2554–2561. doi: 10.1016/j.ejrad.2011.11.020. Epub 2011 Dec 15 PMID: 22177326. [DOI] [PubMed] [Google Scholar]
- 11.Taguchi K., Iwanczyk J.S. Vision 20/20: Single photon counting x-ray detectors in medical imaging. Med Phys. 2013;40 doi: 10.1118/1.4820371. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Douek P.C., Boccalini S., Oei E.H.G., Cormode D.P., Pourmorteza A., Boussel L., et al. Clinical applications of photon-counting CT: a review of pioneer studies and a glimpse into the future. Radiology. 2023;309(1) doi: 10.1148/radiol.222432. PMID: 37787672; PMCID: PMC10623209. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Sartoretti T., Wildberger J.E., Flohr T., Alkadhi H. Photon-counting detector CT: early clinical experience review. Br J Radiol. 2023;96(1147) doi: 10.1259/bjr.20220544. Epub 2023 Feb 10. PMID: 36744809; PMCID: PMC10321251. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Mergen V., Racine D., Jungblut L., Sartoretti T., Bickel S., Monnin P., et al. Virtual Noncontrast Abdominal Imaging with Photon-counting Detector CT. Radiology. 2022;305(1):107–115. doi: 10.1148/radiol.213260. Epub 2022 Jun 7 PMID: 35670712. [DOI] [PubMed] [Google Scholar]
- 15.Flohr T., Petersilka M., Henning A., Ulzheimer S., Ferda J., Schmidt B. Photon-counting CT review. Phys Med. 2020;79:126–136. doi: 10.1016/j.ejmp.2020.10.030. Epub 2020 Nov 26 PMID: 33249223. [DOI] [PubMed] [Google Scholar]
- 16.Dudás I., Schultz L., Benke M., Szücs Á., Kaposi P.N., Szijártó A., et al. The reliability of virtual non-contrast reconstructions of photon-counting detector CT scans in assessing abdominal organs. BMC Med Imaging. 2024;24(1):237. doi: 10.1186/s12880-024-01419-w. PMID: 39251996; PMCID: PMC11386360. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Wohlfahrt P., Möhler C., Hietschold V., Menkel S., Greilich S., Krause M., et al. Clinical implementation of dual-energy CT for proton treatment planning on pseudo-monoenergetic CT scans. Int J Radiat Oncol Biol Phys. 2017;97(2):427–434. doi: 10.1016/j.ijrobp.2016.10.022. Epub 2016 Oct 21 PMID: 28068248. [DOI] [PubMed] [Google Scholar]
- 18.Franka Risch, Decker Josua A., Popp Daniel, Sinzinger Andrea, Braun Franziska, Bette Stefanie, et al. Artifact reduction from dental material in photon-counting detector computed tomography data sets based on high-keV monoenergetic imaging and iterative metal artifact reduction reconstructions—can we combine the best of two worlds? Invest Radiol. 2023;58(9):691–696. doi: 10.1097/RLI.0000000000000967. [DOI] [PubMed] [Google Scholar]
- 19.Baek S., Ye D.H., Lee O. UNet-based multi-organ segmentation in photon counting CT using virtual monoenergetic images. Med Phys. 2025;52(1):481–488. doi: 10.1002/mp.17440. Epub 2024 Oct 7 PMID: 39374095. [DOI] [PubMed] [Google Scholar]
- 20.Hu G., Niepel K., Risch F., Kurz C., Würl M., Kröncke T., et al. Assessment of quantitative information for radiation therapy at a first-generation clinical photon-counting computed tomography scanner. Front Oncol. 2022;14(12) doi: 10.3389/fonc.2022.970299. PMID: 36185297; PMCID: PMC9515409. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Zimmerman J., Poludniowski G. Assessment of Photon-counting computed tomography for quantitative imaging in radiation therapy. Int J Radiat Oncol Biol Phys. 2025;121(5):1316–1327. doi: 10.1016/j.ijrobp.2024.11.069. Epub 2024 Nov 15 PMID: 39549761. [DOI] [PubMed] [Google Scholar]
- 22.Lustermans D., Fonseca G.P., Jeukens C., Taasti V.T., Parodi K., Landry G., Verhaegen F. Evaluating photon-counting computed tomography for quantitative material characteristics and material differentiation in radiotherapy. Phys Med Biol. 2025;70(10) doi: 10.1088/1361-6560/add3ba. PMID: 40315900. [DOI] [PubMed] [Google Scholar]
- 23.Wang T., Ghavidel B.B., Beitler J.J., Tang X., Lei Y., Curran W.J., Liu T., Yang X. Optimal virtual monoenergetic image in “TwinBeam” dual-energy CT for organs-at-risk delineation based on contrast-noise-ratio in head-and-neck radiotherapy. J Appl Clin Med Phys. 2019;20(2):121–128. doi: 10.1002/acm2.12539. Epub 2019 Jan 28. PMID: 30693665; PMCID: PMC6370994. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.van Elmpt W., Landry G., Das M., Verhaegen F. Dual energy CT in radiotherapy: current applications and future outlook. Radiother Oncol. 2016 Apr;119(1):137–144. doi: 10.1016/j.radonc.2016.02.026. Epub 2016 Mar 11 PMID: 26975241. [DOI] [PubMed] [Google Scholar]
- 25.Anhaus J.A., Schmidt S., Killermann P., Mahnken A., Hofmann C. Iterative metal artifact reduction on a clinical photon counting system-technical possibilities and reconstruction selection for optimal results dependent on the metal scenario. Phys Med Biol. 2022;67(11) doi: 10.1088/1361-6560/ac71f0. PMID: 35594851. [DOI] [PubMed] [Google Scholar]
- 26.Skornitzke S., Mergen V., Biederer J., Alkadhi H., Do T.D., Stiller W., et al. Metal artifact reduction in photon-counting detector CT: quantitative evaluation of artifact reduction techniques. Invest Radiol. 2024;59(6):442–449. doi: 10.1097/RLI.0000000000001036. Epub 2023 Oct 9 PMID: 37812482. [DOI] [PubMed] [Google Scholar]
- 27.Skornitzke Stephan, Mergen Victor, Biederer Jürgen, Alkadhi Hatem, Do Thuy D., et al. Metal artifact reduction in photon-counting detector CT: quantitative evaluation of artifact reduction techniques. Invest Radiol. 2024;59(6):442–449. doi: 10.1097/RLI.0000000000001036. [DOI] [PubMed] [Google Scholar]
- 28.Brouwer C.L., Steenbakkers R.J., Bourhis J., Budach W., Grau C., Grégoire V., et al. CT-based delineation of organs at risk in the head and neck region: DAHANCA, EORTC, GORTEC, HKNPCSG, NCIC CTG, NCRI, NRG Oncology and TROG consensus guidelines. Radiother Oncol. 2015;117(1):83–90. doi: 10.1016/j.radonc.2015.07.041. Epub 2015 Aug 13 PMID: 26277855. [DOI] [PubMed] [Google Scholar]
- 29.Grégoire V., Evans M., Le Q.T., Bourhis J., Budach V., Chen A., et al. Delineation of the primary tumour Clinical Target Volumes (CTV-P) in laryngeal, hypopharyngeal, oropharyngeal and oral cavity squamous cell carcinoma: AIRO, CACA, DAHANCA, EORTC, GEORCC, GORTEC, HKNPCSG, HNCIG, IAG-KHT, LPRHHT, NCIC CTG, NCRI, NRG Oncology, PHNS, SBRT, SOMERA, SRO, SSHNO. TROG consensus guidelines Radiother Oncol. 2018;126(1):3–24. doi: 10.1016/j.radonc.2017.10.016. Epub 2017 Nov 24 PMID: 29180076. [DOI] [PubMed] [Google Scholar]
- 30.Jungblut L., Euler A., von Spiczak J., Sartoretti T., Mergen V., Englmaier V., et al. Potential of photon-counting detector CT for radiation dose reduction for the assessment of interstitial lung disease in patients with systemic sclerosis. Invest Radiol. 2022;57(12):773–779. doi: 10.1097/RLI.0000000000000895. Epub 2022 May 28. PMID: 35640003; PMCID: PMC10184807. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Pallasch F.B., Rau A., Reisert M., Rau S., Diallo T., Stein T., Faby S., Bamberg F., Weiss J. Impact of different metal artifact reduction techniques in photon-counting computed tomography head and neck scans in patients with dental hardware. Eur Radiol. 2024;34(6):3742–3749. doi: 10.1007/s00330-023-10430-8. Epub 2023 Nov 16. PMID: 37968474; PMCID: PMC11166854. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Dane B., Ruff A., O'Donnell T., El-Ali A., Ginocchio L., Prabhu V., Megibow A. Photon-counting computed tomography versus energy-integrating dual-energy computed tomography: virtual noncontrast image quality comparison. J Comput Assist Tomogr. 2024;48(2):251–256. doi: 10.1097/RCT.0000000000001562. Epub 2023 Nov 17. PMID: 38013203. [DOI] [PubMed] [Google Scholar]
- 33.Mergen V., Sartoretti T., Baer-Beck M., Schmidt B., Petersilka M., Wildberger J.E., Euler A., Eberhard M., Alkadhi H. Ultra-high-resolution coronary CT angiography with photon-counting detector CT: feasibility and image characterization. Invest Radiol. 2022;57(12):780–788. doi: 10.1097/RLI.0000000000000897. Epub 2022 May 31. PMID: 35640019; PMCID: PMC10184822. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.