The carbon footprints of single-use and reusable medical devices: a systematic review

Amy Booth 1,✉, Monika Chowaniec 1, Sanay Goyal 1, Stuart Faulkner 1, Sara Shaw 1
PMCID: PMC12716512  PMID: 41419285

Abstract

Abstract

Objective

Medical devices account for approximately 6–10% of national health systems’ carbon footprints. The global use of single-use devices has increased, with implications for health systems’ climate impact. This systematic review aimed to synthesise global evidence on medical device carbon footprints, compare single-use and reusable devices and identify lifecycle carbon hotspots to inform policy and practice.

Design

We conducted a systematic review of carbon footprints of medical devices used in clinical settings, reported using Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 guidelines.

Data sources

We searched MEDLINE and Scopus, in 2022 and updated in 2025, and used citation tracking.

Eligibility criteria

English-language, primary research involving carbon modelling of medical devices used in clinical settings was included, with no date restrictions.

Data extraction and synthesis

Articles were screened, and data on carbon modelling methods, device footprints and lifecycle hotspots were extracted by two independent reviewers. Findings were synthesised in figures and tables, and narratively in text. The heterogeneity in carbon modelling approaches prevented quantitative synthesis.

Results

Of 5195 articles identified, 59 met inclusion criteria. Life cycle assessment was the main carbon modelling approach, though application and data quality varied. Carbon footprints of 61 devices were assessed, primarily in surgical (16), anaesthetic (8) and endoscopic (8) specialties. Reusable devices consistently had lower lifecycle footprints. Hotspots were production and manufacturing for single-use devices and reprocessing for reusables.

Conclusion

Reusable devices are preferable from a climate perspective, though efforts are needed to reduce reprocessing emissions. Co-ordinated interventions are required: policymakers can enable supportive regulation; manufacturers can improve device design; healthcare facilities can optimise reprocessing; and providers can prioritise reusable device procurement and use.

Keywords: Climate Change, Systematic Review, Health, Protocols & guidelines


STRENGTHS AND LIMITATIONS OF THIS STUDY.

  • This review employed a transparent, replicable search strategy developed with an expert research librarian and reported using Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 guidelines.

  • Inclusion criteria were clearly defined, with independent screening by multiple reviewers and conflict resolution by consensus, reducing selection bias.

  • Data extraction was conducted independently by two reviewers using a standardised form, increasing consistency and reliability.

  • The review was limited to English-language, peer-reviewed primary studies, potentially excluding relevant grey literature or non-English research.

  • Quantitative synthesis was not conducted due to substantial heterogeneity in carbon modelling methods, reporting standards and units across studies.

Introduction

Anthropogenic climate change, driven by greenhouse gas (GHG) emissions from human activity, poses a major global threat. The health impacts of climate change are well documented, including increases in cardiorespiratory diseases, vector-borne infections and malnutrition.1 2 While health systems are strained by climate-related illness, they also contribute to climate change, generating an estimated 5% of global GHG emissions.3 Approximately 6–10% of national health system emissions come from medical devices.3 4 In 2019, for example, the UK’s National Health Service attributed approximately 2520 kilo-tonnes of carbon dioxide equivalents (CO₂e) to medical devices.4

Medical devices are essential to modern healthcare and include instruments, appliances, software or materials used for diagnosis, treatment or care.5 In recent decades, health systems globally have become increasingly reliant on single-use devices, intended for one use and then discarded,6 rather than reusable devices that can be used across multiple patients and procedures.7,9 This reliance on single-use devices has implications for the carbon footprint and climate impact of medical devices.

The carbon footprint of medical devices is typically assessed using life cycle assessment (LCA), a methodological approach to evaluate environmental impacts, including GHG emissions, across all stages of a product’s lifecycle, from raw material extraction to disposal.10 Within LCAs, two approaches are primarily used for carbon modelling: bottom-up (process-based) and top-down (input–output-based). Bottom-up LCAs rely on detailed, process-specific data and are typically more precise, whereas top-down approaches, such as environmentally extended input–output (EEIO) models, estimate emissions based on financial expenditures, offering less granular insights.11 12 Published studies assessing the carbon footprint and hotspots (ie, lifecycle stages that contribute disproportionately to total emissions) of medical devices employ a mix of these approaches, reflecting the growing application of carbon modelling methods in healthcare. Despite a rising number of studies evaluating carbon footprints of medical devices, no comprehensive systematic review has yet synthesised this evidence to inform policy and practice.

Given the urgency to mitigate healthcare’s climate impact, and as literature on medical device carbon footprints expands, there is a need to synthesise this evidence. This paper aimed to (a) systematically review published evidence to summarise medical device carbon modelling approaches (b) report and compare published carbon footprints of single-use and reusable devices and (c) identify carbon hotspots across device lifecycles to inform opportunities for carbon reduction. Our goal was to inform policy and practice by highlighting the climate impact of medical devices, particularly the implications of single-use versus reusable products. Although this review focuses on carbon emissions, we anticipate that findings may have relevance across broader environmental concerns.

Box 1 summarises definitions used in this article.

Box 1. Climate, carbon modelling and medical device definitions used in article.

Anthropogenic climate change: Climate change resulting from human activity, primarily through the emission of greenhouse gases (GHGs) such as carbon dioxide and methane.

GHG emissions: Gases that trap heat in the atmosphere, contributing to global warming and climate change. Common GHGs include carbon dioxide (CO₂), methane (CH₄) and nitrous oxide (N₂O). In this review, we use carbon and GHG emissions synonymously.

Carbon footprint: The total amount of GHG emissions (expressed as CO₂ equivalents—CO₂e), caused directly and indirectly by an individual, activity, product or organisation.

Life Cycle Assessment (LCA): A method used to assess the environmental impact, including carbon footprint, of a product or system over its entire lifecycle, from raw materials to disposal.

Top-down carbon modelling: A method (eg, environmentally extended input–output (EEIO)) that estimates emissions based on the monetary cost of a product using sector-level emission factors.

Bottom-up carbon modelling: A detailed approach using raw process-specific data across a product’s lifecycle stages to calculate emissions.

Carbon hotspot: A stage or component in a product’s life cycle that contributes disproportionately to its total carbon footprint.

Medical device: Any instrument, appliance, software, material or other article intended for use in diagnosis, treatment or care.

Single-use medical device: A device intended for one-time use, typically made from plastic and discarded after use.

Reusable/reprocessed device: A device that undergoes processes after use to ensure its safe reuse across multiple patients or procedures, including cleaning, disinfection, sterilisation and related procedures as well as testing and restoring the technical and functional safety of the used device.

Methods

We conducted a systematic review of published literature on carbon footprints of medical devices and report the review in accordance with Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 guidelines.

Eligibility criteria

We included peer-reviewed, English-language, primary research articles that calculated the carbon footprint of medical devices used in clinical settings. All carbon modelling methodological approaches were eligible, and no date restrictions were applied. Studies were excluded if (a) they analysed a medical procedure or health service, rather than a medical device; (b) devices examined were used in public, community and home settings; our focus was on devices used in clinical settings where device procurement, use and reprocessing are centralised and amenable to policy and practice interventions; and (c) the publication was unavailable or not empirical research.

Information sources and search strategy

We searched MEDLINE and Scopus databases using a strategy developed with a senior research librarian (online supplemental table A.1). Search terms covered two domains: ‘medical devices’ and ‘carbon footprint’”, including MeSH terms where applicable. The search was initially conducted in 2022 and updated in January 2025. Additional sources included citation tracking and a pre-existing database of sustainability-focused healthcare literature. This database was curated by the lead author (AB) to map emerging evidence on environmental impacts of healthcare, including medical devices, as part of broader sustainable healthcare research and was used in this review to identify relevant studies not captured through standard searches.

Selection process

Titles and abstracts were screened independently by three reviewers (AB, MC, SG). All available full texts were obtained for articles that met inclusion criteria or required further clarification. Articles were managed using EndNote 20, with duplicates removed prior to screening. Disagreements on inclusion were resolved through collective review and discussion with a fourth reviewer (SS).

Data collection and analysis

Data were independently extracted by AB and MC using a standardised form adapted from similar reviews.13 14 Extracted data included: (a) study metadata (title, authors, year, setting); (b) medical device type; (c) carbon modelling approach (eg, boundaries, data sources, assumptions); (d) reported carbon footprint per functional unit (ie, quantified description of the function of the product) of single-use and reusable devices where available; (e) carbon hotspots; and (f) study conclusions. For studies assessing multiple environmental impacts, only carbon data were extracted.

Descriptive data, such as geographical location, device types, modelling approaches and footprint results, were summarised in tables and figures, with narrative synthesis in the text. To provide an overview of carbon modelling approaches, we employed, with permission, a quality assessment tool (online supplemental table A.2) developed by Rizan et al.13 This tool included criteria such as clarity of functional unit, boundary definitions, data sources, emission factors and transparency of assumptions which we report carbon modelling approaches against.

We report carbon footprint values and units as presented in the original studies, without standardised conversions, to reflect variation in reporting and preserve methodological assumptions. Due to heterogeneity in devices and modelling approaches, no quantitative synthesis was performed. Illustrative examples are included in text, such as comparisons of single-use and reusable device carbon footprints, and studies extrapolating results to national or global levels. Carbon hotspots were reported in studies across lifecycle stages in various formats (eg, without values, as CO₂e, or percentage of contribution). Where available, we recorded the top three hotspots per study, categorised under raw material extraction, production/manufacturing, packaging, transport, use, reprocessing and disposal, to ensure consistency and focus on the most significant reported emission sources.

Patient and public involvement

No patients or public were involved in this research.

Ethics statement

Ethical approval was not required for this study as it is a systematic review of previously published literature and does not involve human participants or the collection of new data.

Results

Overview of selected studies

A total of 5195 articles were identified, including 3776 from database searches and 1419 from the pre-curated sustainability library (figure 1). After removing 2227 duplicates, 2968 articles remained. Title and abstract screening excluded 2746 articles, 79 due to non-English language and 2667 for irrelevance, leaving 222 for full-text review. One article comparing single-use and reusable drapes was excluded as the full text could not be retrieved despite librarian support and the abstract information was insufficient.15 Of the 221 full texts assessed, 162 were excluded for being non-original research (n=111), non-English (n=1), irrelevant (n=49) or outside a clinical setting (n=1). 59 studies met inclusion criteria (see table 1 and online supplemental table A.4 for complete overview of included studies). These were published between 1996 and 2024, with a sharp rise from 2022 (figure 2). Most were conducted in the USA (n=22),16,37 followed by Australia (n=7),38,44 the UK (n=6),45,50 France (n=5)51,55 and Germany (n=4).56,59

Figure 1. PRIMSA flow diagram of included studies assessing the carbon footprint of medical devices. PRISMA, Preferred Reporting Items for Systematic Reviews and Meta-Analyses,

Figure 1

Table 1. Summary of nine studies included in the review (1996–2014), carbon modelling approaches employed and carbon footprint outcomes. Summary of remaining studies can be found in Table A.4.

Study, country Medical device studied (brand) Carbon modelling approach (guideline) Functional unit Carbon footprint Top three carbon hotspots (% or carbon dioxide equivalents/ CO2e if available) Study conclusions
Single-use Reusable Single-use Reusable
Kummerer et al,56 Germany Laparotomy pads
(not mentioned)
LCA (1980s ‘technical standard’) One laparotomy pad; results reported for 1000 pads 304.8 kg CO2; 225.9 g CO; 1637·9 g N₂O; 565.8 g NOₓ; per 1000 pads 157.1 kg CO₂; 163.9 g CO; 102.9 g N₂O; 432.2 g NOx per 1000 pads if reusable pads each used 15 times Production Washing Reusable pads have lower carbon footprint
Ison et al, 45 UK Suction receptacles
(not mentioned)
LCA (not mentioned) Average kilograms of waste from body fluids produced during 1 year of elective surgery at a district general hospital Results presented on logarithmic scale Results presented on logarithmic scale Not mentioned Washing Carbon footprint of single-use suction receptacles greater than reusable device
Carbon footprint of reusable affected by washing process
McGain et al, 38 Australia Anaesthetic drug trays
(Single-use: Chinese-made polyurethane. Reusable: Australian-made reusable nylon tray)
LCA (ISO 14040) One anaesthetic tray 126 g CO2 per tray
204 g CO2 per tray with cotton/paper
110 g CO2 per tray Polyurethane tray (111 g CO2)
Cotton gauze
(68 g CO2)
Polyurethane wrap (8 g CO2)
Washing (99 g CO2)
Drying (9 g CO2)
Nylon tray (2 g CO2)
Significant carbon savings by converting to reusable trays
Carbon savings exceed 50% in hospitals using gas co-generation compared with brown coal
Adding more cotton and paper to trays increases their carbon emissions
Eckelman et al,16 USA LMA
(Single-use: Unique, Hangzhou China. Reusable: Classic, Singapore)
LCA (ISO 14040) Maintenance of airway by 40 disposable LMAs or 40 uses of 1 reusable LMA 0.3 kg CO2e per maintenance of one airway 0.2 kg CO2e per maintenance of one airway Production and polymerisation of PVC Natural gas production and combustion to produce steam for autoclave machine Reusable LMAs have lower carbon footprint
Carbon footprint of reusable LMA can be reduced by bulk autoclaving, energy-efficient machines and ship transport
Grimmond et al,17 USA Sharps containers
(Single-use: BD Franklin Lakes. Reusable: Daniels Sharpsmarts Inc)
LCA (British Standards Institute) Conversion of single-use sharps containers to reusable for 500 uses over 12-month period, workload normalised per 100 occupied bed years 24.2 Mt CO₂e per 100 occupied bed years 4 Mt CO2e per 100 occupied bed years Manufacture (56.4%)
Transport (35.8%)
Disposal (7.8%)
Decanting and washing (52.5%)
Transport (25.5%) Manufacture (15.4%)
Reusable sharps containers reduced GWP by 83.5% over the study period
Electricity source can alter global warming potential by 15%
Total savings of 64 000 Mt CO2e if results expanded across the USA
McGain et al,39 Australia CVC insertion kits
(not mentioned)
LCA (ISO 14040) One CVC kit 407 g CO2e per kit
(European energy)
1211 g CO2e per kit (brown coal)
436 g CO2e per kit (hospital cogeneration)
764 g CO2e per kit (US energy)
572 g CO2e per kit (European energy)
Manufacturing of plastics (70%)
Steel (25%)
Sterilisation (70%) Reusable kit has a higher carbon footprint than single-use
The energy mix determines the carbon footprint
Ibbotson et al,57 Germany Surgical scissors
(not mentioned)
LCA (ISO 14040) 4500 use cycles of scissors during 18 years based on technical lifetime of the reusable product Individual carbon footprints not reported Individual carbon footprints not reported Material
Manufacturing
Raw material
Manufacturing
Use
Reusable scissors had the lowest carbon footprint; 11 and 52 times less than plastic and single-use stainless-steel scissors, respectively
Sorensen et al,70 Denmark Bedpans
(Single-use: Saniwaste System moulded cardboard; GoLoo polyethylene; stainless steel; Supercore T-499 McAirlaid superabsorbent inlay. Reusable: polyethylene)
Consequential LCA (Danish Environmental Design of Industrial Products) Use of one bedpan once for urinating and defecating while being hospitalised and in bed 0.15–0.2 kg CO₂e per use (cardboard)
0.1–0.15 kg CO₂e per use (polyethylene)
0.25–0.3 kg CO₂e per use (stainless steel)
0.25–0.3 kg CO₂e per use (polyethylene bedpan)
Waste incineration
Washing for the cardboard bedpan Incineration for polyethylene bedpan
Washing and disinfection of bedpan.
Gloves and disposable bag used to carry bedpans to washroom
Single-use polyethylene bedpan had lowest carbon footprint due to energy recovery opportunity
Energy, wastewater management and changed workflow contribute to differences in carbon footprint
Pourzahedi et al,18 USA Nanosilver-coated bandage (Acticoat 7) LCA (not mentioned) Not mentioned 130.04–130.06 kg CO₂e per bandage Not mentioned Silver nanoparticle production (130 kg CO2e) Bandage manufacture (0.04–0.05 kg CO2e)
Bandage disposal (0–0.01 kg CO2e)
Not mentioned Silver nanoparticle synthesis had highest carbon footprint, compared with production and incineration of the bandage

CO₂, carbon dioxide; CO₂e, carbon dioxide equivalent; CVC, central venous catheter; GWP, global warming potential; ISO, International Organization for Standardization; LCA, Life Cycle Assessment; LMA, laryngeal mask airway; Mt, Metric tonne; N₂O, nitrous oxide; NOx, nitrogen oxides; PVC, polyvinyl chloride.

Figure 2. Annual change in number of publications assessing the carbon footprint of medical devices.

Figure 2

Carbon footprints of 61 distinct medical devices were assessed (figure 3). These included surgical devices (n=16)2728 30 34 36 45 48 50 51 56,58 60 such as laparoscopic instruments and scrub caps; anaesthetic and airway management devices (n=8)16 21 35 38 40 42 43 52 including laryngoscopes; endoscopy devices (n=8)29 32 41 53 54 59 64 65 such as ureteroscopes and cystoscopes; infection control wear (n=6)2247 66,69 including face masks; imaging devices (n=4)23 37 including CT and ultrasound scans; patient hygiene products (n=4)31 33 70 71 such as bedpans; procedure kits (n=3)40 49 72 including central venous catheter insertion kits; obstetric and gynaecological devices (n=3)19 20 26 such as vaginal specula; diagnostic monitoring devices (n=2)25 73 including blood pressure cuffs; and other devices (n=7)17 18 24 44 46 74 75 such as intermittent compression sleeves.

Figure 3. Categories of medical devices with published carbon footprints (1996–2024).

Figure 3

Carbon modelling approaches

57 studies employed LCA approaches (table 1 and online supplemental table A.4), including consequential LCAs (modelling environmental impacts from change in practice), attributional LCAs (measuring direct environmental burdens) and process-based LCAs. The remaining two studies employed bottom-up carbon footprinting44 and product material analysis,62 respectively. A complicated landscape of LCA guidelines was evident across studies (box 2). 12 studies reported using ISO 14040,1619,21 25 38 5 cited ISO 1404434 48 50 61 74 and 11 reported both standards.22 31 33 42 51 53 59 63 66 68 69 Others included ISO 14025,60 ISO 9001/1400154 and ISO 14067.67 Additional frameworks included BSI,17 PAS 2050,46 PEF,47 IMPACT 2002+,72 Danish EDIP70 and GREET.36 Three studies mentioned using ‘standardised guidelines’ without naming them,41 56 73 and one used a literature-derived protocol.71 16 did not specify the guideline used.1823 24 26,30 32 35 37 43 45 49 52 64

Box 2. Overview of Life Cycle Assessment (LCA) frameworks used in studies.

ISO 14040 (1997, revised 2006)

Outlines the general principles and framework for conducting LCA.

ISO 14044 (2006)

Provides specific requirements, guidelines and methodologies for conducting LCAs, typically used alongside ISO 14040.

ISO 14025 (2006)

Standard for communicating quantified environmental information (eg, through Environmental Product Declarations or EPDs) based on LCA.

ISO 14067 (first published 2013, updated 2018)

Specifies principles, requirements and guidelines to quantify and report the carbon footprint of products.

ISO 14001 (first published 1996, updated 2004, latest version 2015)

Provides a framework for organisations to manage their environmental responsibilities through an Environmental Management System.

ISO 9001 (first published 1987, updated 2008, latest version 2015)

Globally recognised standard for quality management. It is sometimes referenced in studies related to environmental processes due to its emphasis on process control and continual improvement.

British Standards Institute (BSI) Guidelines UK national standards offering guidance on sustainable practices and environmental management.

Greenhouse Gas Protocol Publicly Available Specification (PAS) 2050 Global specification for assessing life cycle greenhouse gas emissions of goods and services, widely used in carbon footprinting.

Product Environmental Footprint (PEF) Methodology An EU-developed approach to measure the environmental performance of products across their lifecycle.

Impact assessment of chemical and particulate-related chemical toxicity (IMPACT) 2002+ Life cycle impact assessment method evaluating multiple environmental impacts of chemicals, including global warming potential.

Environmental Design of Industrial Products (EDIP) A Danish-developed methodology (1997) for assessing the environmental impacts of products across their life cycle.

Greenhouse gases, Regulated Emissions, and Energy use in Technologies (GREET) Model Developed by Argonne National Laboratory, US Department of Energy, estimates greenhouse gas and pollutant emissions related to transportation fuels and vehicle technologies.

SimaPro (n=31)1618 21 23 25 26 29 30 33,35 38 was the most commonly used LCA software (online supplemental table A.3), followed by OpenLCA (n=5)36 47 65 71 75 and Umberto (n=2).58 74 Other software included GaBi,68 WinPepi,24 the Packaging Industry Research Association’s Environmental Management System,45 a custom-built Excel-based model46 and an online carbon calculator.28 Sixteen studies did not report which software was used.17 19 20 22 27 31 32 41 44 49 56 59 62 64 67 76 Inventory data sources (eg, for emission factors) varied, with many studies using EcoInvent (n=38)1618,21 25 26 29 30 33; national or regional databases (n=20)17,1921 23 including the US Life Cycle Inventory, the UK’s Department for Environment, Food and Rural Affairs, the EU’s International Life Cycle Data system and China’s Life Cycle Database; and industry datasets (n=10).17 19 20 24 38 39 48 50 59 66 Eight studies did not specify their inventory source.28 32 36 41 45 49 52 76 Data collected for carbon modelling varied and included manufacturer-supplied information (eg, product material composition), manual measurements (eg, energy use, device weight), literature and device usage surveys.

Studies reported various carbon modelling system boundaries (online supplemental table A.3). Most studies reported including raw material extraction (n=45)1216 18,22 24; production and manufacturing (n=53)16,3538; packaging (n=17)16 18 21 22 24 29 30 32 35 38 39 58 61 68 71 73 77; transport (n=45)16,2426 27 29; use, referring specifically to the carbon footprint associated with the operation or clinical application of the device (n=37)1216 18,20 23 25; reprocessing of reusable devices, referring to the cleaning, disinfection and sterilisation of devices for safe reuse, typically on-site or by third-party contractors (n=38)1617 19 21 22 24,27 29; and disposal and end-of-life management (n=54).116,22 24

Common assumptions (online supplemental table A.3) included lifespan and reprocessing factors for reusable devices (eg, machine electricity source, machine load and water use per wash), transport mode for device distribution (eg, by air, sea or road) and clinical equivalency between single-use and reusable devices. Exclusions included infrastructure, machinery maintenance, minor components (eg, ink on packaging), staff transport and auxiliary materials (eg, gloves used during reprocessing). Reported limitations included database constraints, assumptions about user behaviour, narrow study scope and limited generalisability due to varying national electricity grids, procurement and waste management practices.

Carbon footprints of single-use and reusable medical devices

The carbon footprints of medical devices varied widely (table 1 and online supplemental table A.4) and were reported across diverse functional units such as per use, procedure, patient, provider, year or aggregated values. All studies expressed footprints in CO₂e, though units ranged from grams to tonnes. Single-use devices analysed per use ranged from 5·904 g CO₂e for staple firings for one surgical procedure62 to 130·06 kg CO₂e for use of one nano-silver-coated bandage.18 Reusable devices analysed per use ranged from 3 g CO₂e per use of a reusable microfibre cloth75 to 19·7 kg CO₂e for one MRI scan.23

Of the included studies, 47 directly compared single-use and reusable medical devices.1617 21 22 24,27 29 Of these, 39 (83%)1617 21 22 24,27 29 found reusable devices had a lower carbon footprint, often dependent on reuse thresholds (ie, reusables became more carbon-efficient after a certain number of uses). Studies reported various carbon savings from switching to reusable devices. For example, Grimmond et al17 found an 83·5% annual carbon reduction with reusable sharps containers; Vozzola et al27 reported a 66% reduction using reusable surgical gowns; and Burguburu et al51 found a 31% drop with reusable scrub suits. Eight studies17 28 29 48 50 61 63 69 extrapolated carbon savings to national or global scales. For example, Boberg et al61 projected a 360 tonne CO₂e saving if half of Germany’s laparoscopic cholecystectomies used reusable trocars; Rizan et al48 calculated a 396 tonne annual saving in England from hybrid laparoscopic equipment; and Cohen et al63 estimated global use of reusable head covers could reduce 10 000 tonnes.

Six studies40 43 53 58 67 70 found reusables had a higher footprint due to factors like raw material impacts (eg, cotton production for masks), fossil fuel–intensive energy grids driving up reprocessing emissions, carbon-intensive waste management strategies and operational efficiency from single-use devices. For example, Leiden et al58 showed single-use surgical equipment resulted in shorter operating times that lowered theatre energy use. Davis et al41 found comparable emissions for single-use and reusable duodenoscopes,41 while McGain et al39 reported varying results depending on the national energy grid.

Carbon hotspots of single-use and reusable medical devices

Carbon hotspots varied by device type (figure 4). For single-use devices, production and manufacturing were the most frequently cited contributors, identified among the top three in 36 studies.16,1821 22 24 Examples include manufacture of plastic and steel components in catheter kits (70% and 25%, respectively),39 single-use duodenoscopes (91–96%)29 and optical fibres in laryngoscopes (64%).52 Disposal and end-of-life processes were top hotspots in 13 studies,17 18 22 24 27 29 33 41 46 48 52 59 70 contributing, for instance, 14% of the footprint for cholecystectomy instruments.48 Transport was a major contributor in nine studies,24 25 33 46 48 53 60 69 75 while packaging and raw materials were hotspots in 22 27 61 68 75five and 25 33 47 48 60 67 68seven studies, respectively, with raw materials, for example, accounting for 40·5% of emissions from face masks.68 High production emissions were typically driven by frequent use of devices, while disposal emissions varied with waste treatment method (eg, incineration vs landfill).

Figure 4. Frequency of lifecycle carbon hotspot mentions for single-use and reusable devices.

Figure 4

For reusable devices, reprocessing, including washing, sterilisation, disinfection and laundry, was the most frequently reported carbon hotspot, appearing among the top three contributors in 36 studies.1617 21 22 24 26 27 29 31,33 38 For example, reprocessing reusable devices for reuse during the use phase contributed 52.5% of the carbon footprint of reusable sharps containers,17 99% for pulse oximeters73 and over 95% for reusable bronchoscopes.64 Reprocessing emissions varied with method (eg, autoclaving vs chemical disinfection), reprocessing batch size and national energy mix powering machines (eg, washing machines). Production and manufacturing were top contributors in 18 studies,1722,27 32 41 46 51 53 57 59 61 65 69 75 for example, representing 74% of the carbon footprint for flexible cystoscopes.53 Use-phase emissions were highlighted in five studies,23 26 29 51 57 for example, from electricity use of CT scans.23 Disposal and end-of-life processes were cited as key contributors in five studies,33 46 53 69 75 such as 12% of emissions from flexible cystoscopes.53 Transport appeared among the top three hotspots in five studies,17 24 26 46 53 particularly when off-site reprocessing was required. Packaging was cited in five studies21 27 32 52 61; while raw material extraction was a major contributor in two studies,33 57 such as cotton for reusable operating room bed covers.33

Discussion

Our systematic review provides a timely synthesis of evidence on carbon footprints of medical devices, offering three key contributions with implications for policy and practice. First, we make a methodological contribution by mapping the diverse carbon modelling approaches used across studies. While LCA was the predominant approach, its application varied widely in terms of system boundaries, data sources, assumptions and reporting units. This lack of standardisation limits comparability and highlights the need for more consistent yet adaptable methodologies that reflect the diversity of devices and settings. Second, we confirm that reusable devices generally have lower carbon footprints, especially when used extensively throughout their lifespan and reprocessed efficiently (eg, using energy-efficient machinery). However, this was not universally true. In some settings, reusables had higher footprints, highlighting the importance of device- and context-specific assessments rather than blanket assumptions. Third, by identifying recurring carbon hotspots, such as production and manufacturing for single-use devices and reprocessing for reusables, we pinpoint where carbon reduction efforts can be most effectively directed.

Application of LCAs varied considerably across studies included in our review. Particular challenges arose in modelling hybrid devices, such as endoscopes with both reusable and disposable components, and in accounting for diverse reprocessing pathways, including on-site versus third-party sterilisation. Further limitations stemmed from the lack of detailed emissions data for specific geographies, niche device components and sterilisation materials. Restricted access to proprietary data from medical device manufacturers also constrained bottom-up modelling approaches, leading many studies to rely on less accurate proxy data. Efforts are underway, nationally and internationally, to standardise carbon modelling in healthcare and expand emissions databases.79 These initiatives have the potential to enhance the accuracy, comparability and utility of carbon footprint data. However, given the diversity of device types and clinical settings, any standardised approach must retain flexibility and be sensitive to local conditions. Importantly, growing academic interest in device carbon modelling should not delay action. While robust data is useful, focusing excessively on measurement risks slowing the implementation of carbon reduction strategies.

Our review demonstrated a convincing climate case for prioritising reusable medical devices. Across diverse global contexts, reusable devices consistently exhibited lower carbon footprints. Alongside this, reusable devices offer additional environmental benefits, including waste reduction and long-term cost savings.25 38 39 In the few instances where single-use devices had lower emissions, this was attributable to contextual factors, such as energy-intensive local reprocessing systems or fossil fuel-dominated national electricity grids. In addition to these structural challenges, however, cultural and regulatory barriers constrain widespread adoption of reusable devices. Healthcare provider concerns about infection control, driven by industry marketing, ingrained habits and risk aversion, continue to limit uptake.80 These concerns persist despite a lack of strong evidence that single-use devices offer superior infection control.77 81 Regulatory complexity presents another obstacle. In some countries, for example, China, reprocessing remains prohibited.82 Encouraging, globally regulatory frameworks are evolving. Countries, including the USA, Canada, Japan and Australia, now permit the reprocessing of single-use devices.83 The European Union’s (EU’s) Medical Device Regulation similarly allows reprocessing under defined conditions, although national implementation may vary.84 For example, at the time of authorship, France is in consultation to determine if they will allow reprocessing; meanwhile, the Netherlands and Germany permit reprocessing, though the Netherlands imposes a more stringent administrative burden. Such divergence constrains economies of scale and limits the circularity potential of medical device reprocessing across EU markets. Meanwhile, in the UK, the Medicines and Healthcare products Regulatory Agency recognises re-manufacturing as a regulated process to restore devices for reuse.6 84 85

Beyond prioritising reusable devices, targeting carbon hotspots across device lifecycles offers further opportunities to reduce device emissions, requiring coordinated stakeholder action. Since production and manufacturing often drive emissions, there are opportunities for manufacturers to reduce footprints by optimising design and manufacturing processes, guided by circular economy principles.6 Facilities can invest in energy-efficient reprocessing and streamline reprocessing protocols (eg, bulk autoclaving). Healthcare providers can be trained to use devices efficiently, for example, by reducing overuse in procedural packs.86 Reducing disposal emissions requires collaboration with waste management systems to improve recycling and reprocessing. Supportive policy at national and local levels is needed to enable and incentivise use of reusable devices. Given the global nature of medical device supply chains, coordinated policy efforts could have a transformative impact on device sustainability.

This review supports prioritising reusable medical devices and encourages transitioning without waiting for complete carbon data on every product. A key exception is low-income and middle-income countries, where device carbon modelling is lacking. Given differing infrastructure and energy profiles, further context-specific research is needed to guide mitigation strategies. Future research should also focus on optimising lifecycle stages, such as improving reprocessing and disposal, and identifying policy, regulatory, economic or behavioural levers to support adoption. For example, although more than half of the studies incorporated reprocessing within their carbon modelling boundaries, these typically referred to on-site or small-scale processes rather than large-scale commercial reprocessing. As such, regional differences in reprocessing regulatory environments, market dynamics and large-scale reprocessing practices were not captured in this review and warrant future investigation. Gaps also remain in understanding industry-level mitigation efforts and infection control outcomes of single-use versus reusable devices. As carbon modelling becomes more common in healthcare, robust quality appraisal standards will be essential for interpreting emerging evidence. Finally, while our review primarily assessed the carbon footprint associated with medical devices, there is a need for future work to integrate a ‘handprint’ approach, that is, examining the procedural performance of the device, and how device use influences patient outcomes and may reduce the need for further, more carbon-intensive care if the device were not used. Such analysis would capture more holistic climate and clinical co-benefits and trade-offs. For example, the findings reported by Leiden et al,58 where single-use surgical equipment shortened operating times and reduced theatre energy use, illustrate how procedural efficiency can generate positive handprint effects that offset some or all of the carbon footprint of devices.

Our review has several limitations. Literature is concentrated in surgical and anaesthetic specialties and largely driven by a small group of researchers, limiting generalisability across device types. Most studies were conducted in high-income countries; differences in healthcare infrastructure, clinical practice and energy systems mean footprints in low-income and middle-income countries may differ. Although most studies used LCAs, methodological rigour varied, limiting comparability and preventing quantitative synthesis. No validated quality appraisal tool currently exists for healthcare carbon footprint studies; we adapted an existing framework, but its applicability to complex or mixed-use devices was limited. Lastly, our review focused on carbon emissions, limiting assessment of environmental trade-offs and co-benefits between single-use and reusable devices.

Conclusion

Our review demonstrates that carbon modelling of medical devices is increasingly represented in the academic literature; however, carbon modelling approaches are marked by methodological and data inconsistencies. Despite these inconsistencies, our review confirmed that reusable medical devices tend to have a lower carbon footprint than their single-use counterparts. We further identified carbon hotspots, which differ between device types. Production and manufacturing dominate emissions for single-use devices, while reprocessing contributes the most to emissions from reusable devices. The findings from this review support the preferential adoption of reusable devices and call for policy and practice interventions across the medical device lifecycle to reduce their climate impact. To achieve this, engagement is required by multiple stakeholders. Manufacturers must design and supply reusable products, healthcare facilities need to invest in and optimise reprocessing systems, procurers must prioritise sustainable purchasing, while healthcare providers should be supported to use reusable devices. Regulatory and policy frameworks that facilitate the safe reuse and reprocessing of medical devices are critical to supporting these practices. Addressing this global challenge also requires international cooperation across medical device supply chains. Future research should shift from generating additional carbon footprint data to exploring how existing evidence can inform policy, procurement and clinical practice to reduce the climate impact of medical devices.

Supplementary material

online supplemental file 1
bmjopen-15-12-s001.docx (154.4KB, docx)
DOI: 10.1136/bmjopen-2025-108446

Acknowledgements

We thank Nia Roberts for support with search strategy and database searches. We acknowledge the funding support from the National Institute for Health and Care Research Applied Research Collaboration Oxford and Thames Valley at Oxford Health NHS Foundation Trust.

Footnotes

Funding: Support for the review process was secured via two internships (MC, SG), supported by the National Institute for Health and Care Research (NIHR) Applied Research Collaboration Oxford and Thames Valley at Oxford Health NHS Foundation Trust. The latter additionally supported senior academic time (SS). The NIHR and NHS had no role in study design, data collection, data analysis, data interpretation or writing of the report.

Prepublication history and additional supplemental material for this paper are available online. To view these files, please visit the journal online (https://doi.org/10.1136/bmjopen-2025-108446).

Provenance and peer review: Not commissioned; externally peer reviewed.

Patient consent for publication: Not applicable.

Ethics approval: Not applicable.

Data availability free text: All data used in this systematic review are from publicly available sources and detailed in the manuscript and supplementary materials.

Patient and public involvement: Patients and/or the public were not involved in the design, or conduct, or reporting, or dissemination plans of this research.

Data availability statement

All data relevant to the study are included in the article or uploaded as supplementary information.

References

  • 1.Bayram H, Bauer AK, Abdalati W, et al. Environment, Global Climate Change, and Cardiopulmonary Health. Am J Respir Crit Care Med. 2017;195:718–24. doi: 10.1164/rccm.201604-0687PP. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Rocklöv J, Dubrow R. Climate change: an enduring challenge for vector-borne disease prevention and control. Nat Immunol. 2020;21:479–83. doi: 10.1038/s41590-020-0648-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Pichler P-P, Jaccard IS, Weisz U, et al. International comparison of health care carbon footprints. Environ Res Lett. 2019;14:064004. doi: 10.1088/1748-9326/ab19e1. [DOI] [Google Scholar]
  • 4.Tennison I, Roschnik S, Ashby B, et al. Health care’s response to climate change: a carbon footprint assessment of the NHS in England. Lancet Planet Health. 2021;5:e84–92. doi: 10.1016/S2542-5196(20)30271-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Medical devices: how to comply with the legal requirements in great britain gov.UK2022. 2022. https://www.gov.uk/guidance/medical-devices-how-to-comply-with-the-legal-requirements Available.
  • 6.MacNeill AJ, Hopf H, Khanuja A, et al. Transforming The Medical Device Industry: Road Map To A Circular Economy. Health Aff (Millwood) 2020;39:2088–97. doi: 10.1377/hlthaff.2020.01118. [DOI] [PubMed] [Google Scholar]
  • 7.Reprocessing of medical devices: European commission (topics of interest) https://health.ec.europa.eu/medical-devices-topics-interest/reprocessing-medical-devices_en n.d. Available.
  • 8.Gautam D, Sahney R. Reprocessing and reuse of single-use medical devices and the role of interprofessional collaboration: A literature review. Current Medicine Research and Practice . 2020;10:70–4. doi: 10.1016/j.cmrp.2020.03.001. [DOI] [Google Scholar]
  • 9.Shuman EK, Chenoweth CE. Reuse of medical devices: implications for infection control. Infect Dis Clin North Am. 2012;26:165–72. doi: 10.1016/j.idc.2011.09.010. [DOI] [PubMed] [Google Scholar]
  • 10.Beginner’s guide: how to do an lca for free, life cycle assessment tools and data: business energy. 2021. https://www.businessenergy.com/switching-services/lca-tools Available.
  • 11.Wiedmann T, Minx J. In: Ecological Economics Research Trends. Pertsova C, editor. Hauppauge NY, USA: Nova Science Publishers; 2008. A definition of ’Carbon footprint. [Google Scholar]
  • 12.Bolwig S, Gibbon P. Counting carbon in the marketplace: Part 1 - Overview paper. Global Forum on Trade. 2009 [Google Scholar]
  • 13.Rizan C, Steinbach I, Nicholson R, et al. The Carbon Footprint of Surgical Operations. Ann Surg. 2020;272:986–95. doi: 10.1097/SLA.0000000000003951. [DOI] [PubMed] [Google Scholar]
  • 14.Purohit A, Smith J, Hibble A. Does telemedicine reduce the carbon footprint of healthcare? A systematic review. Future Healthc J . 2021;8:e85–91. doi: 10.7861/fhj.2020-0080. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Mcdowell J. An environmental, economic, and health comparison of single-use and reusable surgical drapes and gowns. J-Global. 1993
  • 16.Eckelman M, Mosher M, Gonzalez A, et al. Comparative Life Cycle Assessment of Disposable and Reusable Laryngeal Mask Airways. Anesth Analg. 2012;114:1067–72. doi: 10.1213/ANE.0b013e31824f6959. [DOI] [PubMed] [Google Scholar]
  • 17.Grimmond T, Reiner S. Impact on carbon footprint: a life cycle assessment of disposable versus reusable sharps containers in a large US hospital. Waste Manag Res. 2012;30:639–42. doi: 10.1177/0734242X12450602. [DOI] [PubMed] [Google Scholar]
  • 18.Pourzahedi L, Eckelman MJ. Environmental Life Cycle Assessment of Nanosilver-Enabled Bandages. Environ Sci Technol. 2015;49:361–8. doi: 10.1021/es504655y. [DOI] [PubMed] [Google Scholar]
  • 19.Campion N, Thiel CL, Woods NC, et al. Sustainable healthcare and environmental life-cycle impacts of disposable supplies: a focus on disposable custom packs. J Clean Prod. 2015;94:46–55. doi: 10.1016/j.jclepro.2015.01.076. [DOI] [Google Scholar]
  • 20.Unger SR, Hottle TA, Hobbs SR, et al. Do single-use medical devices containing biopolymers reduce the environmental impacts of surgical procedures compared with their plastic equivalents? J Health Serv Res Policy. 2017;22:218–25. doi: 10.1177/1355819617705683. [DOI] [PubMed] [Google Scholar]
  • 21.Sherman JD, Raibley LA, Eckelman MJ. Life Cycle Assessment and Costing Methods for Device Procurement: Comparing Reusable and Single-Use Disposable Laryngoscopes. Anesth Analg. 2018;127:434–43. doi: 10.1213/ANE.0000000000002683. [DOI] [PubMed] [Google Scholar]
  • 22.Vozzola E, Overcash M, Griffing E. Environmental considerations in the selection of isolation gowns: A life cycle assessment of reusable and disposable alternatives. Am J Infect Control. 2018;46:881–6. doi: 10.1016/j.ajic.2018.02.002. [DOI] [PubMed] [Google Scholar]
  • 23.Martin M, Mohnke A, Lewis GM, et al. Environmental Impacts of Abdominal Imaging: A Pilot Investigation. J Am Coll Radiol. 2018;15:1385–93. doi: 10.1016/j.jacr.2018.07.015. [DOI] [PubMed] [Google Scholar]
  • 24.McPherson B, Sharip M, Grimmond T. The impact on life cycle carbon footprint of converting from disposable to reusable sharps containers in a large US hospital geographically distant from manufacturing and processing facilities. PeerJ. 2019;7:e6204. doi: 10.7717/peerj.6204. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Sanchez SA, Eckelman MJ, Sherman JD. Environmental and economic comparison of reusable and disposable blood pressure cuffs in multiple clinical settings. Resour Conserv Recycl. 2020;155:104643. doi: 10.1016/j.resconrec.2019.104643. [DOI] [Google Scholar]
  • 26.Donahue LM, Hilton S, Bell SG, et al. A comparative carbon footprint analysis of disposable and reusable vaginal specula. Am J Obstet Gynecol. 2020;223:225. doi: 10.1016/j.ajog.2020.02.007. [DOI] [PubMed] [Google Scholar]
  • 27.Vozzola E, Overcash M, Griffing E. An Environmental Analysis of Reusable and Disposable Surgical Gowns. AORN J. 2020;111:315–25. doi: 10.1002/aorn.12885. [DOI] [PubMed] [Google Scholar]
  • 28.Baxter NB, Yoon AP, Chung KC. Variability in the Use of Disposable Surgical Supplies: A Surgeon Survey and Life Cycle Analysis. J Hand Surg Am. 2021;46:1071–8. doi: 10.1016/j.jhsa.2021.05.027. [DOI] [PubMed] [Google Scholar]
  • 29.Le NNT, Hernandez LV, Vakil N, et al. Environmental and health outcomes of single-use versus reusable duodenoscopes. Gastrointest Endosc. 2022;96:1002–8. doi: 10.1016/j.gie.2022.06.014. [DOI] [PubMed] [Google Scholar]
  • 30.Agarwal D, Bharani T, Armand W, et al. Reusable scrub caps are cost-effective and help reduce the climate footprint of surgery. Langenbecks Arch Surg . 2023;408:358. doi: 10.1007/s00423-023-03107-9. [DOI] [PubMed] [Google Scholar]
  • 31.Griffing E, Overcash M. Reusable and Disposable Incontinence Underpads: Environmental Footprints as a Route for Decision Making to Decarbonize Health Care. J Nurs Care Qual. 2023;38:278–85. doi: 10.1097/NCQ.0000000000000697. [DOI] [PubMed] [Google Scholar]
  • 32.Kemble JP, Winoker JS, Patel SH, et al. Environmental impact of single-use and reusable flexible cystoscopes. BJU Int. 2023;131:617–22. doi: 10.1111/bju.15949. [DOI] [PubMed] [Google Scholar]
  • 33.Chang JH, Woo KP, Silva de Souza Lima Cano N, et al. Does reusable mean green? Comparison of the environmental impact of reusable operating room bed covers and lift sheets versus single-use. Surgeon. 2024;22:236–41. doi: 10.1016/j.surge.2024.05.003. [DOI] [PubMed] [Google Scholar]
  • 34.Donahue LM, Petit HJ, Thiel CL, et al. A Life Cycle Assessment of Reusable and Disposable Surgical Caps. Journal of Surgical Research. 2024;299:112–9. doi: 10.1016/j.jss.2024.04.007. [DOI] [PubMed] [Google Scholar]
  • 35.Kidane J, Thiel CL, Wang K, et al. A Comparison of Environmental Impacts Between Reusable and Disposable Flexible Laryngoscopes. Laryngoscope. 2025;135:1666–73. doi: 10.1002/lary.31927. [DOI] [PubMed] [Google Scholar]
  • 36.Lehane A, Sullivan GA, Dunn J, et al. Life Cycle Analysis of a Pediatric Surgical Kit-A Target to Reduce Operating Room Waste. JAMA Surg. 2024;159:1436–8. doi: 10.1001/jamasurg.2024.3823. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Esmaeili A, Twomey JM, Overcash MR, et al. Scope for energy improvement for hospital imaging services in the USA. J Health Serv Res Policy. 2015;20:67–73. doi: 10.1177/1355819614554845. [DOI] [PubMed] [Google Scholar]
  • 38.McGain F, McAlister S, McGavin A, et al. The financial and environmental costs of reusable and single-use plastic anaesthetic drug trays. Anaesth Intensive Care. 2010;38:538–44. doi: 10.1177/0310057X1003800320. [DOI] [PubMed] [Google Scholar]
  • 39.McGain F, McAlister S, McGavin A, et al. A life cycle assessment of reusable and single-use central venous catheter insertion kits. Anesth Analg. 2012;114:1073–80. doi: 10.1213/ANE.0b013e31824e9b69. [DOI] [PubMed] [Google Scholar]
  • 40.McGain F, Story D, Lim T, et al. Financial and environmental costs of reusable and single-use anaesthetic equipment. Br J Anaesth. 2017;118:862–9. doi: 10.1093/bja/aex098. [DOI] [PubMed] [Google Scholar]
  • 41.Davis NF, McGrath S, Quinlan M, et al. Carbon Footprint in Flexible Ureteroscopy: A Comparative Study on the Environmental Impact of Reusable and Single-Use Ureteroscopes. J Endourol. 2018;32:214–7. doi: 10.1089/end.2018.0001. [DOI] [PubMed] [Google Scholar]
  • 42.Lightfoot SJ, Grant T, Boyden A, et al. Single-use synthetic plastic and natural fibre anaesthetic drug trays: a comparative life cycle assessment of environmental impacts. Br J Anaesth. 2024;133:1465–77. doi: 10.1016/j.bja.2024.05.031. [DOI] [PubMed] [Google Scholar]
  • 43.Chen S, McAlister S, Colagiuri P, et al. Switching to reusable operating theatre equipment: lessons learnt from sterile light handle projects in two Australian hospitals. ANZ J Surg. 2025;95:216–21. doi: 10.1111/ans.19306. [DOI] [PubMed] [Google Scholar]
  • 44.Hansell L, Delaney A, Milross M, et al. Reducing unnecessary use of intermittent pneumatic compression in intensive care: A before-and-after pilot study with environmental perspective. Aust Crit Care. 2025;38:101125. doi: 10.1016/j.aucc.2024.09.010. [DOI] [PubMed] [Google Scholar]
  • 45.Ison E, Miller A. THE USE OF LCA TO INTRODUCE LIFE-CYCLE THINKING INTO DECISION-MAKING FOR THE PURCHASE OF MEDICAL DEVICES IN THE NHS. J Env Assmt Pol Mgmt. 2000;02:453–76. doi: 10.1142/S1464333200000497. [DOI] [Google Scholar]
  • 46.Grimmond TR, Bright A, Cadman J, et al. Before/after intervention study to determine impact on life-cycle carbon footprint of converting from single-use to reusable sharps containers in 40 UK NHS trusts. BMJ Open. 2021;11:e046200. doi: 10.1136/bmjopen-2020-046200. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Jamal H, Lyne A, Ashley P, et al. Non-sterile examination gloves and sterile surgical gloves: which are more sustainable? J Hosp Infect. 2021;118:87–95. doi: 10.1016/j.jhin.2021.10.001. [DOI] [PubMed] [Google Scholar]
  • 48.Rizan C, Bhutta MF. Environmental impact and life cycle financial cost of hybrid (reusable/single-use) instruments versus single-use equivalents in laparoscopic cholecystectomy. Surg Endosc . 2022;36:4067–78. doi: 10.1007/s00464-021-08728-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Byrne C, Pley C, Schorscher P, et al. A mixed-methods analysis of the climate impact, acceptability, feasibility and cost of switching from single-use pulp to reusable plastic trays in a large NHS trust. Future Healthc J. 2023;10:157–60. doi: 10.7861/fhj.2022-0129. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Rizan C. Environmental impact of hybrid (reusable/single-use) ports versus single-use equivalents in robotic surgery. J Robot Surg. 2024;18:155. doi: 10.1007/s11701-024-01899-6. [DOI] [PubMed] [Google Scholar]
  • 51.Burguburu A, Tanné C, Bosc K, et al. Comparative life cycle assessment of reusable and disposable scrub suits used in hospital operating rooms. Cleaner Environmental Systems . 2022;4:100068. doi: 10.1016/j.cesys.2021.100068. [DOI] [Google Scholar]
  • 52.Rouvière N, Chkair S, Auger F, et al. Reusable laryngoscope blades: a more eco-responsible and cost-effective alternative. Anaesthesia Critical Care & Pain Medicine . 2023;42:101276. doi: 10.1016/j.accpm.2023.101276. [DOI] [PubMed] [Google Scholar]
  • 53.Baboudjian M, Pradere B, Martin N, et al. Life Cycle Assessment of Reusable and Disposable Cystoscopes: A Path to Greener Urological Procedures. Eur Urol Focus. 2023;9:681–7. doi: 10.1016/j.euf.2022.12.006. [DOI] [PubMed] [Google Scholar]
  • 54.Pioche M, Pohl H, Cunha Neves JA, et al. Environmental impact of single-use versus reusable gastroscopes. Gut. 2024;73:1816–22. doi: 10.1136/gutjnl-2024-332293. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Schoen J, Chopra V. The Harm We Do: The Environmental Impact of Medicine. J Hosp Med. 2018;13:353–5. doi: 10.12788/jhm.2947. [DOI] [PubMed] [Google Scholar]
  • 56.Kümmerer K, Dettenkofer M, Scherrer M. Comparison of reusable and disposable laparatomy pads. Int J LCA. 1996;1:67–73. doi: 10.1007/BF02978647. [DOI] [Google Scholar]
  • 57.Ibbotson S, Dettmer T, Kara S, et al. Eco-efficiency of disposable and reusable surgical instruments—a scissors case. Int J Life Cycle Assess. 2013;18:1137–48. doi: 10.1007/s11367-013-0547-7. [DOI] [Google Scholar]
  • 58.Leiden A, Cerdas F, Noriega D, et al. Life cycle assessment of a disposable and a reusable surgery instrument set for spinal fusion surgeries. Resour Conserv Recycl. 2020;156:104704. doi: 10.1016/j.resconrec.2020.104704. [DOI] [Google Scholar]
  • 59.Thöne M, Lask J, Hennenlotter J, et al. Potential impacts to human health from climate change: A comparative life-cycle assessment of single-use versus reusable devices flexible ureteroscopes. Urolithiasis. 2024;52:166. doi: 10.1007/s00240-024-01664-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Quintana-Gallardo A, Del Rey R, González-Conca S, et al. The Environmental Impacts of Disposable Nonwoven Fabrics during the COVID-19 Pandemic: Case Study on the Francesc de Borja Hospital. Polymers (Basel) 2023;15:1130. doi: 10.3390/polym15051130. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Boberg L, Singh J, Montgomery A, et al. Environmental impact of single-use, reusable, and mixed trocar systems used for laparoscopic cholecystectomies. PLoS ONE. 2022;17:e0271601. doi: 10.1371/journal.pone.0271601. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Meissner M, Hafermann J, Silas U, et al. Evaluating the Environmental Impact of Single-Use and Multi-Use Surgical Staplers with Staple Line Buttressing in Laparoscopic Bariatric Surgery. Risk Manag Healthc Policy. 2023;16:1423–33. doi: 10.2147/RMHP.S415989. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Cohen ES, Djufri S, Bons S, et al. Environmental Impact Assessment of Reusable and Disposable Surgical Head Covers. JAMA Surg. 2023;158:1216–7. doi: 10.1001/jamasurg.2023.3863. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Massart N, Millet C, Beloeil H, et al. How green is my reusable bronchoscope? Anaesthesia Critical Care & Pain Medicine . 2024;43:101420. doi: 10.1016/j.accpm.2024.101420. [DOI] [PubMed] [Google Scholar]
  • 65.López-Muñoz P, Martín-Cabezuelo R, Lorenzo-Zúñiga V, et al. Environmental footprint and material composition comparison of single-use and reusable duodenoscopes. Endoscopy. 2025;57:116–23. doi: 10.1055/a-2364-1654. [DOI] [PubMed] [Google Scholar]
  • 66.Hiloidhari M, Bandyopadhyay S. Environmental footprints of disposable and reusable personal protective equipment ‒ a product life cycle approach for body coveralls. J Clean Prod. 2023;394:136166. doi: 10.1016/j.jclepro.2023.136166. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Luo Y, Yu M, Wu X, et al. Carbon footprint assessment of face masks in the context of the COVID-19 pandemic: Based on different protective performance and applicable scenarios. J Clean Prod. 2023;387:135854. doi: 10.1016/j.jclepro.2023.135854. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Atılgan Türkmen B. Life cycle environmental impacts of disposable medical masks. Environ Sci Pollut Res . 2022;29:25496–506. doi: 10.1007/s11356-021-17430-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69.van Straten B, Ligtelijn S, Droog L, et al. A life cycle assessment of reprocessing face masks during the Covid-19 pandemic. Sci Rep. 2021;11:17680. doi: 10.1038/s41598-021-97188-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Sørensen BL, Wenzel H. Life cycle assessment of alternative bedpans – a case of comparing disposable and reusable devices. J Clean Prod. 2014;83:70–9. doi: 10.1016/j.jclepro.2014.07.022. [DOI] [Google Scholar]
  • 71.Willskytt S, Tillman A-M. Resource efficiency of consumables – Life cycle assessment of incontinence products. Resour Conserv Recycl. 2019;144:13–23. doi: 10.1016/j.resconrec.2018.12.026. [DOI] [Google Scholar]
  • 72.Hemberg L, Wessberg N, Leire C, et al. Environmental impact of single-use and reusable items in central venous catheter insertion kits: a life cycle assessment. Intensive Care Med. 2023;49:662–4. doi: 10.1007/s00134-023-07078-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Duffy J, Slutzman JE, Thiel CL, et al. Sustainable Purchasing Practices: A Comparison of Single-use and Reusable Pulse Oximeters in the Emergency Department. WestJEM . 2023;24 doi: 10.5811/WESTJEM.58258. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 74.Lichtnegger S, Meissner M, Paolini F, et al. Comparative Life Cycle Assessment Between Single-Use and Reprocessed IPC Sleeves. Risk Manag Healthc Policy. 2023;16:2715–26. doi: 10.2147/RMHP.S439982. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 75.Maloney B, McKerlie T, Nasir M, et al. The environmental footprint of single-use versus reusable cloths for clinical surface decontamination: a life cycle approach. Journal of Hospital Infection . 2022;130:7–19. doi: 10.1016/j.jhin.2022.09.006. [DOI] [PubMed] [Google Scholar]
  • 76.Esmaeili A, McGuire C, Overcash M, et al. Environmental impact reduction as a new dimension for quality measurement of healthcare services. Int J Health Care Qual Assur. 2018;31:910–22. doi: 10.1108/IJHCQA-10-2016-0153. [DOI] [PubMed] [Google Scholar]
  • 77.Unger S, Landis A. Assessing the environmental, human health, and economic impacts of reprocessed medical devices in a Phoenix hospital’s supply chain. J Clean Prod. 2016;112:1995–2003. doi: 10.1016/j.jclepro.2015.07.144. [DOI] [Google Scholar]
  • 78.Campion N, Thiel CL, DeBlois J, et al. Life cycle assessment perspectives on delivering an infant in the US. Sci Total Environ. 2012;425:191–8. doi: 10.1016/j.scitotenv.2012.03.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 79.Health systems task force: sustainable markets initiative. 2024. https://www.sustainable-markets.org/taskforces/health-systems-taskforce Available.
  • 80.Grantcharov P, Ahmed S, Wac K, et al. Reprocessing and reuse of single-use medical devices: perceptions and concerns of relevant stakeholders toward current practices. Int J Evid Based Healthc. 2019;17:53–7. doi: 10.1097/XEB.0000000000000146. [DOI] [PubMed] [Google Scholar]
  • 81.Sherman JD, Hopf HW. Balancing Infection Control and Environmental Protection as a Matter of Patient Safety: The Case of Laryngoscope Handles. Anesthesia & Analgesia. 2018;127:576–9. doi: 10.1213/ANE.0000000000002759. [DOI] [PubMed] [Google Scholar]
  • 82.Wang D, Wu J. Reprocessing and reuse of single-use medical devices in China: a pilot survey. BMC Public Health. 2019;19:461. doi: 10.1186/s12889-019-6835-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 83.Global Regulatory Standards for “Single-Use” Medical Device Reprocessing and Remanufacturing. Association of Medical Device Reprocessors; 2022. [Google Scholar]
  • 84.Benedettini O. Green Servitization in the Single-Use Medical Device Industry: How Device OEMs Create Supply Chain Circularity through Reprocessing. Sustainability. 2022;14:12670. doi: 10.3390/su141912670. [DOI] [Google Scholar]
  • 85.Single-use medical devices . UK guidance on re-manufacturing. gov.uk; 2016. Agency mhpr. [Google Scholar]
  • 86.Wyssusek KH, Keys MT, van Zundert AAJ. Operating room greening initiatives - the old, the new, and the way forward: A narrative review. Waste Manag Res. 2019;37:3–19. doi: 10.1177/0734242X18793937. [DOI] [PubMed] [Google Scholar]

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.